CMC 156 DatasheetWM
CMC 156 DatasheetWM
Reference Manual
CMC 156
Technical Data
5 Technical Data
The following data refer to a CMC 156 test set with standard design, without
external amplifiers.
For test sets with the EP option (Extended Precision), technical data
regarding the power amplifiers are different from the corresponding data of
standard devices: therefore, these data can be found separately in chapter 6,
"Technical Data EP Option".
• Guaranteed values:
- In general:
Valid for 1 year from factory calibration, within 23 °C ± 5 °C
(73 °F ± 10 °F) at nominal value and a warm-up time of > 25 min.
- Guaranteed values of generator outputs:
Valid in the frequency range 10 – 100 Hz.
• Accuracy specifications:
Valid in the frequency range DC – 100 Hz.
Mains supply
Connector Plug acc. to IEC320
Voltage, 1-phase
Nominal voltage 110 to 240 Vac
Permissible range 99V to 264 Vac
Mains fuse T6.3 AH 250 V
Power consumption < 600 VA
Frequency
Nominal frequency 50 / 60 Hz
Permissible range 47 to 63 Hz
31
Test Set CMC 156
5.2 Outputs
The data given in the following table are valid for all generator outputs,
regardless of the amplifiers used (internal or external ones); the following
tables are valid only for equipment with internal amplifiers.
For the block diagram of the generator outputs available please refer to
Figure 3 on page 18.
1
If you purchased the option FL-6, the maximum output frequency is constrained to 599 Hz.
2
Valid for sinusoidal signals with 50 / 60 Hz.
32
Technical Data
1
Data for 3-phase systems are valid under symm. conditions (0 °, 120 °, 240 °).
2
For higher current or output power demands connection of an external amplifier is possible (e.g.
CMA current amplifier). The power data refers to an output current of 12.5 Aeff or 12.5ADC.
3
Single-phase model (operation with two phases in phase opposition):
two currents in series, additional adapter SPA156 recommended.
4
2 phases (2 x 10.5 A) in parallel for in-phase operation.
5
rd. = reading; rg. = range, whereat n % of rg. means: n % of upper range value (1.25 or 12.5 A).
6
Values at 50/60 Hz, 20 kHz measuring band width, nominal value and nominal load.
33
Test Set CMC 156
45
Ptyp in [VA]
40 Pgar in [VA]
35
30
25
P in VA
20
15
10
0
0 1 1 2 2 3 3 4 4 5 5 6 6 7 7 8 8 9 9 10 10 11 11 12 12 13
I in A
60
50 1-ph.
typ.
40
P in VA
1 ph.
guar.
30
20
10
0
10
11
12
13
14
15
16
17
18
19
20
21
0
1
2
3
4
5
6
7
8
9
I in A
Figure 13 Typical (typ) and guaranteed (guar.) output power of the current
outputs for single-phase operation (two outputs in parallel) with
standard design
34
Technical Data
1
If not indicated otherwise, the voltages quoted are L-N.
2
Refer to output power curve in Figure 14
3
Data for three-phase systems are valid under symmetrical conditions (0°, 120°, 240°).
4
rd. = reading; rg. = range, whereat n % of rg. means: n % of upper range value (125V).
5
Values at 50/60 Hz, 20 kHz measuring band width, nominal value and nominal load.
35
Test Set CMC 156
140
120
100
3 x sym. guar.
Power in VA
80
3 x sym. typ.
60
L2 only guar.
L2 only typ.
40
DC L2 only typ
20
0
0,00 25,00 50,00 75,00 100,00 125,00
V (L-N) in Vrms
Figure 14 Typical (Ptyp) and guaranteed (Pmin) output power of the voltage
amplifiers1 with standard design
6 outputs
Setting range 0 to 5 Vrms
Output current max. 2 mA
Resolution < 250 µV
Accuracy typ. error < 0.025 % guar. error < 0.05 %
2
Harmonic distortion typ. < 0.015 % guar. < 0.05 %
Short-circuit protection Unlimited against GND_A.
Isolation SELV isolated to all other potential groups.
Electrically connected to ground.
1
To obtain the specified power, in the test object parameters file (.TYP) in the CMC software the
corresponding voltage has to be set.
2
Values at nominal voltage (5V), 50/60 Hz and 20 kHz measuring band width.
36
Technical Data
The following diagram shows the limit curve for direct voltage (for alternative
voltage a max. power of 2000 VA is achieved).
Voltage in V / Power in W
300
250
200
V
150
P
100
50
0
0 1 2 3 4 5 6 7 8
Current in A
Figure 15 Limit curve of the relays of the binary outputs for direct voltage
37
Test Set CMC 156
CMC 156
Rear panel
Vexternal = 5 ...15V
22k
16V
6k8 47
38
Technical Data
5.3 Inputs
5.3.1 Binary Inputs
10 Binary inputs
Response time typ. 220 µs (debounced)
Sampling rate < 100 µs
Measurement accuracy See Table 18 and Table 19
Counter function
Counting frequency 3 kHz (per input)
Pulse width > 150 µs (for high and low signal)
Operating threshold for - Guaranteed 0:
potential-free operation input open or load against N > 100 k
- Guaranteed 1:
input short-circuited to N or
load against N < 20k
Operating threshold for 2 to 250 Vdc to be set in the software
potential-sensing
operation
Hysteresis 0.36V 0.2V
Input resistance 70k against reference potential (GND_BI)
Input capacity 10nF against reference potential (GND_BI)
Trigger criteria Switching of a potential-free contact or application
of direct voltage up to 250 V. Configurable
operating threshold.
Integration period 27 min
Connection 4 mm banana sockets
Isolation Functional isolation to power outputs and between
the two groups, with galvanic separation from each
other.
Reinforced isolation to all SELV interfaces and to
mains.
Configuration The binary inputs are configurable. Information
about binary input configuration is found in the
CMC software user's manual in Chapter 3
"Configuring the binary inputs”
Table 17 Binary inputs
39
Test Set CMC 156
40
Technical Data
2 Counter inputs
max. counting frequency 100 kHz
Pulse width >3 µs (high and low signal)
Operating threshold
pos. edge max. 8 V
neg. edge min. 4 V
Hysteresis typical: 2 V
rise, fall time < 1 ms
max. input voltage 30 V
Connector Socket "ext. Interf." (rear panel CMC 156)
Isolation SELV to all other potential groups of the test set.
Electrically connected to Gen. Out 7-12 and
ground.
CMC 156
Back panel
+ 12V
22k
Counter input 1..2
ext. Interface
47p
100k
41
Test Set CMC 156
1
Exceeding this value may lead to destruction in the device.
42
Technical Data
25-pole SUB-D-plug
Designation Host Interface
Use The Host Interface (parallel interface) serves as
connection from the CMC 156 test set to the
computer (LPT or USB).
If the computer is equipped with a parallel printer
port, too, use the 25-pole cable (VEHK0002) that
was supplied with the test set to connect the
CMC 156 to the computer.
If the computer is equipped with a USB port, use
an OMICRON CMUSB-P converter (ordering
number VEHZ2007) instead of the VEHK0002 to
connect the CMC 156 to the computer.
Isolation SELV isolated to all other potential groups;
electrically connected to ground.
Climate
Operating temperature 0 ... 50 °C (32 … 122 °F )
Storage and transport -25 ... +70 °C (-13 … 158 °F)
Humidity range 5 ... 95 % relative humidity; non condensing
Table 24 Climate
43
Test Set CMC 156
Dynamics
Vibration Tested according to IEC68-2-6 (operating mode)
frequency range 10 to 150 Hz; acceleration 2 g
continuous (20 m/s2); 10 sweeps, each axis.
Shock Tested according to IEC68-2-27 (operating mode)
15 g/11 ms, half-sine, each axis.
EMC
CE conformity The product complies to the normative document
about electromagnetic compatibility for
standardization of the laws of the member states of
the council of the European Union (EMC standard
89/336/EEC).
Emission
International EN 50081-2
Europe FCC Subpart B of Part 15 Class A
Susceptibility
International EN 50082-2:1992
Europe IEC 1000-4-2/3/4/6
44
Technical Data
5.6 Safety
45