[Data Sheet] R7_V3.03 Rev06
[Data Sheet] R7_V3.03 Rev06
Data Sheet
V3.03
Samsung Medison Bldg., 42, Teheran-ro 108-gil, Gangnam-gu, Seoul 135-851, Korea
TEL:(82)2-2194-1400 FAX:(82)2-556-9537
www.samsungmedison.com
Marketing Strategy Team
www.samsungmedison.com
Table of Contents
0 SPECIFICATION SUMMARY -------------------------------------------------------------- 1
1 GENERAL SPECIFICATION -------------------------------------------------------------- 3
1.1 Console Design ------------------------------------------------------------------ 3
1.2 Main Monitor ------------------------------------------------------------------ 3
1.3 Control Panel ------------------------------------------------------------------ 3
1.4 PC ------------------------------------------------------------------ 3
1.5 Electrical Specification ------------------------------------------------------------------ 3
2 System Specification -------------------------------------------------------------- 3
2.1 Application ------------------------------------------------------------------ 3
2.2 Preset ------------------------------------------------------------------ 3
2.3 Operation Mode ------------------------------------------------------------------ 4
2.4 Transducer Type ------------------------------------------------------------------ 4
2.5 System Standard Features ------------------------------------------------------------------ 4
2.6 System Option ------------------------------------------------------------------ 5
2.7 Display ------------------------------------------------------------------ 5
2.8 Language ------------------------------------------------------------------ 6
3 Processing -------------------------------------------------------------- 6
3.1 Pre-Processing ------------------------------------------------------------------ 6
3.2 Post-Processing ------------------------------------------------------------------ 7
3.3 Data Processing ------------------------------------------------------------------ 7
4 Connectivity -------------------------------------------------------------- 7
4.1 DICOM ------------------------------------------------------------------ 7
4.3 Peripheral Interface ------------------------------------------------------------------ 8
5 Scanning Parameters -------------------------------------------------------------- 8
5.1 2D Mode ------------------------------------------------------------------ 8
5.2 M Mode ------------------------------------------------------------------ 9
5.3 Color Mode ------------------------------------------------------------------ 9
5.4 PWD Mode ------------------------------------------------------------------ 9
5.5 CWD Mode ------------------------------------------------------------------ 9
5.6 PD Mode ------------------------------------------------------------------ 10
5.7 3D/4D Mode ------------------------------------------------------------------ 10
6 Probes -------------------------------------------------------------- 11
6.1 Linear Array ------------------------------------------------------------------ 11
6.2 Convex Array ------------------------------------------------------------------ 11
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
www.samsungmedison.com
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
www.samsungmedison.com
SPECIFICATION SUMMARY 19 inch LCD Monitor (LED Backlit unit) called “LCD
monitor” henceforth
IMAGING MODES
REAR PANEL INPUT/OUTPUT CONNECTIONS
2D imaging mode
VHS and SVHS VCR left and right audio
M imaging mode
B/W printer video and remote control
Color Doppler Imaging (CDI) mode
VGA monitor
Power Doppler Imaging (PDI) mode
Parallel port
Directional Power Doppler Imaging (DPDI) mode
USB
Pulse Wave (PW) Spectral Doppler imaging mode
LAN
Continuous Wave (CW) Spectral Doppler imaging mode
Monitor
Transmit focusing, maximum of eight points (four points
Microphone
simultaneously selectable)
e-motion marker (IPX7)
Digital dynamic receive focusing (continuous)
USER INTERFACE
PROBE CONNECTIONS
English, German, French, Spanish, Italian, Russian,
4 probe connectors (Including one CW prove connector)
Chinese, Portuguese
MAIN MONITOR
ELECTRICAL PARAMETERS
PRESSURE LIMITS
HUMIDITY LIMITS
TEMPERATURE LIMITS
Obstetrical
19 inches High Resolution Monitor with LED Backlit
Pediatric
Display Size: 1,280 X 1,024
Small Parts
Number of Color: 16.7M
TCD
Brightness Adjustment
Urology
Articulated Monitor Arm
Vascular
Interactive Dynamic Software Menu
Aorta
Alpha-Numeric KBD
Aortic arch
1 User Key
Arterial
4 Probe Holders
Bowel
Height Adjustment: 150mm
Breast
PC Carotid
Elbow / Wrist
Hand-Foot 3DC2-6
Micro-convex Array: CF4-9, C4-9, C4-9/10ED -. Cine loop: Max. 8,192 Lines
-. Diag. Physician
SYSTEM OPTIONS
-. Ref. Physician
4D -. Sonographer
3D XI™ -. Description
DICOM -. Indication
DynamicMR Preset
Panoramic Frequency
StressEcho Map
Gray Bar
DISPLAY Color Bar
Scale
Application
Index : MI, TIs, TIb, TIc
Auto ID Creation
Caliper
Date: 3 types (Selectable)
Measurement
-. YYYY-MM-DD
Indicator
-. MM-DD-YYYY
Pointer
-. DD-MM-YYYY
-. Depth
-. Power
PROCESSING
-. Scale (PRF)
-. Frequency -. Depth
-. Harmonic -. Frequency
-. Density -. Power
-. Power -. Scale(PRF)
-. TGC -. Sensitivity
PW Mode 3D / 4D Mode
-. Focal position -. 3D / 4D
3D Mode
POST-PROCESSING -. 3D
-. 3D XI™
B-Mode
-. Accept ROI
-. Chroma Map
-. Chroma Map
-. Frame Average
-. MagiCut™
-. Gain
-. Gray Map
DATA PROCESSING
-. Image Size
-. SDMR Cine
-. Read Zoom -. Function: Auto Run / Run / Stop / Cine Save / First Pos.
-. Doppler Map Image Viewer: Archived images from past patient exams
-. Chroma Map
-. Gain
DICOM Print
Dual live
PERIPHERAL INTERFACE
Dynamic Range: 50 ~ 200dB
-. Recording only Spatial Compound Index: Off, Low, Mid, High (Linear
-. Digital Color Video Printer: Sony UP-D23MD, Sony UP- Trapezoidal: On, Off (Linear probes only)
Update
Balance: 1 ~ 16
Filter: 0 ~ 3
Color Map: 1 ~ 14
Auto Calculation : On, Off
Density: Low, Middle1, Middle2, High
Color Format: Color + BW, Color only, BW only
CWD MODE
Sensitivity: 8 ~ 31
Frame Average: 0 ~ 9
Auto Calc.: On, Off
Frequency: Pen, Gen
Base line: -4 ~ 4
Gain: 0 ~ 100
Chroma Map: Off, 1 ~ 9, User 1 ~ 3
Dual Live
Display format
Invert
-. Up/down: 40/60, 50/50, 60/40
Power: 10 ~ 100
-. Side by Side
PRF: 0.6kHz ~ 14kHz
Dynamic range: 1 ~ 4
Sensitivity: 2 ~ 31
Gain: 0 ~ 100
Steer: Left/None/Right (Probe dependent)
Invert: On, Off
Velocity, Vel + Variance
Power: 10 ~100
Filter: 0 ~ 3
PRF: 1.53kHz0 ~ 45.83 kHz
Quick scan™: On, update
PWD MODE
Sweep speed: 60Hz ~ 360Hz -. Render Setup: Render mode1, Render mode2, TH High
Update -. Magicut™
Filter: 0 ~ 3 -. Palette
-. Layout
PD MODE -. Ref. Image
-. Ref.OH
Balance: 1 ~ 16
-. Preset
Color Map: 1 ~ 14
-. Palette
Density: Low, Middle1, Middle2, High
-. Post Processing
Color Format: Color + BW, Color only, BW only
Oblique View™
Sensitivity: 8 ~ 31
-. Single Static Line
Frame Average: 0 ~ 9
-. Dynamic Line
Frequency: Pen, Gen
-. Contour
Gain: 0 ~ 100
-. Multi Line
Dual Live
-. Multi Contour
Invert
-. Multi Parallel
Power: 10 ~ 100
-. Multi Plumb
PRF: 0.6kHz ~ 14kHz
-. Palette
Sensitivity: 2~ 31
-. Post Processing
Steer: None, Left, Right
Volume CT
Filter: 0 ~ 3
-. Cube
-. Cross
3D / 4D MODE
-. Select Surface
-. Ref. Image
3D
Preset
4D (Live 3D)
- Default / Surface / Skeleton / User1 ~ 5
3D XI™
ROI Size / ROI Position
-. MSV, Oblique View™, Volume CT™
Rendering Preset: Default, Surface, Skeleton, , User1 ~ 5
3D Cine
Scan quality: Low, Middle, High, Extreme
4D Cine
Volume Angle: 10 ~ 70° (Probe dependent)
MPR
-. Display Format: 2D, 2D/3D, VolumeCT, ROI 3D,
Fixed 3D
Safety Class: BF
LINEAR
L5-12/50EP
LN5-12 Center frequency : 7.5MHz
Center frequency : 8MHz Radius of curvature : Flat
Radius of curvature : Flat Field of view : 50mm
Field of view : 38.1mm Number of elements : 128
Number of elements : 128 Biopsy Guide : Available
Biopsy Guide : Available Application : MSK, Small Parts, Vascular
Application : MSK, Small Parts, Vascular Safety Class: BF
Safety Class: BF
CONVEX
L3-8 C2-5
Center frequency : 5.0MHz Center frequency : 3.4MHz
Radius of curvature : Flat Radius of curvature : 39.7mm
Field of view : 38.7mm Field of view : 45.7°
Number of elements : 128 Number of elements : 128
Biopsy Guide : Available Biopsy Guide : Available
Application : Abdomen, Small Parts, Vascular Application : Abdomen, OB, GYN
Safety Class: BF Safety Class: BF
L5-12/50 C2-8
Center frequency : 7.5MHz Center frequency : 4.9MHz
Radius of curvature : Flat Radius of curvature : 51mm
Field of view : 50mm Field of view : 68°
Number of elements : 128 Number of elements : 128
Biopsy Guide : Available Biopsy Guide : Available
Application : MSK, Small Parts, Vascular Application : Abdomen, OB, GYN
Safety Class: BF Safety Class: BF
HL5-12ED CF4-9
Center frequency : 7.5MHz Center frequency : 6.56MHz
Radius of curvature : Flat Radius of curvature : 14mm
Field of view : 38.4mm Field of view : 92°
Number of elements : 128 Number of elements : 128
Biopsy Guide : Available Biopsy Guide : Not Available
Safety Class: BF
C4-9, C4-9/10ED
Center frequency : 6.5MHz 3DC2-6
Radius of curvature : 14mm Center frequency : 3.8 MHz
Safety Class: BF
ENDOCAVITY
UE Vein
PN2-4 LE Artery
Testicle
SP3-8 Superficial
Angle
PENCIL
Line Trace
CW2.0
%StD
Center frequency: 2.0 MHz
Ellipse
Application : Cardiac
Area Trace
Safety Class: BF
Area %StD
1 Distance Volume
DP2B
3 Distance Volume
Center frequency: 2.0 MHz
Distance + Ellipse Volume
Application : Cardiac
Distance A/B
Safety Class: BF
Distance Velocity
Distance Trace
MEASUREMENT M Distance
Caliper OBSTETRICS
Obstetrics
Fetal Biometry
Fetal Echo
-. GS
Gynecology
-. MSD (D1, D2, D3)
-. CRL -. Ear
-. YS -. MP
-. OFD -. Pelvis
-. HC AFI
-. APD -. Q1
-. TAD -. Q2
-. AC -. Q3
-. FTA -. Q4
-. FL -. AFI Total
-. SL -. MVP
-. APTD Cervix
-. ThC -. HrtD ap
-. ULNA -. ThA
-. TIB -. HrtA
-. NF Renal Artery
-. NT Volume Flow
-. Lt. / Rt. Vp
-. A2C SV -. IVSd
-. A4C SV -. FS
2D Echo -. SV
-. MPA Diam -. EF
-. LA Diam -. Fetal HR
-. RA Diam MPA
-. RV Diam Duct A
-. IVS IVC
-. HrtC MV Inflow
-. ThC -. MV Inflow E
-. EF TV Inflow
CTAR -. TV Inflow E
-. All(D) -. TV Inflow A
-. ThD ap -. TV E/A
-. HrtD ap PLI
-. Uterus W
UE ARTERY
-. Uterus Volume
CAROTID
LE ARTERY
-. LVPWs
LE VEIN
-. RVAWs
General -. IVSs
-. LVIDs
RADIOLOGY -. LVPWs
-. RVAWs
Celica A
-. RVIDs
Splenic A
LV Vol. (Simpson)
Splenic Vol.
-. A4CVol.d
Hepatic A
-. A4C Vol.s
SMA
-. A2C Vol.d
IMA
-. A2C Vol.s -. LA
-. LVAs sax Ao / LA
LV Mass -. LA Diam
RV -. Ao Isth Diam
-. RVAd Ao / LA(M)
-. RVAWs -. All
-. RVIDs -. Ao Root
-. RV Major -. LA Diam
-. RV Minor -. LVPEP
-. RPA Diam RA
RV (M) -. RA Minor
-. RVAWd -. RAAd
-. RVIDd -. RAAs
-. RVAWs -. RAEDV
-. RVIDs -. RAESV
-. LA Major LVOT
-. LA Minor -. LVOT Diam.
-. LVOT AccT MV
RVOT -. MV Diam 1
-. RVOT ET -. MV Peak E
AV -. MV Peak A
-. AV Cusp -. MV AccT
-. AV Diam, -. MV DecT
-. AV Vmax -. MV ET
-. AV AccT -. MR VC Diam
-. AV ET -. MR Limited Trace
-. R-R Interval -. MR Manual Trace
-. AR VC Diam. MV(M)
-. AR ed Vmax -. D-E
TV -. PR PHT
-. TV Diam 1 -. PR DecT
-. TV Vmax -. All
-. TV Peak E -. MV TST
-. TV Peak A -. MV ET
-. TV PHT -. MV IVRT
-. TV VTI -. MV IVCT
-. TV DecT -. All
-. TR VC Diam -. All
PV Tissue Doppler
-. PR VC Diam -. LV Peak A’
-. PV Vmax Qp/Qs
-. PV DecT -. Sys. HR
-. PV ET -. Pulm. VTI
-. PR Auto Trace PE
-. PR Limited Trace -. PEd
UROLOGY TCD
Volume Flow
Hip Angle
SAFTEY/EMC
THYROID
- IEC/EN 62304:2006
Lt. / Rt. Volume - IEC/EN 60601-1-6:2010
Lt. / Rt. Flow - IEC/EN 62366:2007
- CAN/CSA 22.2 No.601.1-M90:1990, with R2003, with Output Display on the system monitor display of output
R2005 acoustic
SAFETY
RoHS
WEEE
of output acoustic
maximum output
ANTI-VIRUS SOLUTION