Principles of spectrophotometric methods-updated HT2016
Principles of spectrophotometric methods-updated HT2016
UV-Vis
Basic principles
Electomagnetic radiation
= c
Wavelength, ; frequency, ; speed of light, c
E = h
There are several other important techniques using electromagnetic radiation with wavelengths from
radio- to X-ray range (NMR, NIR, FTIR, XAFS, etc), although these will not be covered here.
The colour of a solution is the complement
of the colour of the light that it absorbs. The
colour we perceive depends not only on the
wavelength of light, but on its irradiance
(intensity).
Molecular orbital diagram of formaldehyde
Possible Electronic States Resulting
from an n → * transition a) excited
singlet state, S1 and b) excited triple
state, T1
What happens to the energy absorbed by the
molecule?
P
Transmittance T
P0
defined as the fraction of light passing through the sample cuvette,
often expressed as %.
P
Absorbance A log 0 log T
P
absorbance is often used since it is linearly proportional to concentration (see Box
17-1, page 397).
Beer´s law A bc
Where is the molar absorptivity (M-1cm-1), b the path length (cm) and c the
concentration (M)
• Monochromatic light
• Dilute solutions
Although UV-Vis uses electron transitions for measurements,
very broad absorption spectra are obtained due to vibration and
rotation band in molecules.
The relationship
I k P0 c
is valid for certain conditions and especially at low analyte
concentrations. At high concentrations so called “self-absorption” of
emitted light causes problems. k is a constant including, among other
things, the observation volume and the probability of transition at .
Applications of Spectrophotometry
UV-Vis spectrophotometry is not only important for quantitative chemical
analysis, but also a useful tool for fundamental studies in inorganic-, physical-
and biochemistry.
Chapter 19 describes several examples of this, and also the analysis of mixtures
of several analytes with overlapping spectra.
Immunoassays
Monochromators
Monochromators are used to select which wavelength interval to use for
measurements.
Concave mirrors are used to collimate (make parallel) the light beams before
hitting the grating and, after wavelength dispersion, to focus each wavelength
at a different point in the focal plane.
Only the wavelength interval that passes through the exit slit is used for
measurements.
By rotating the grating different wavelengths are projected on the exit slit.
Fig 20-5.Czerny-Turner grating monochromator.
The refraction grating is ruled with a series of closely spaced parallel groves
with distance d.
/ = nN
Where is wavelength, n is the diffraction order, and N is the
number of prooves of the grating that are illuminate.
Dispersion of grating:
/ = n/dcos
where is the angular dispersion and n is the difraction
order
The effect of changing slit width
The slit width is normally variable on monochromator instruments and can be changed
to change the resolution of the spectrometer to suit different applications.
Unfortunately, decreasing the slit width does not only increase resolution, but does also
decrease the light throughput of the monochromator. This deteriorates the signal-to-noise
ratio and thereby detection limit.
The wider the exit slit (in figure 20-5) the wider the band of wavelengths selected by the
monochromator.
A wide slit increases the energy reaching the detector and gives a high signal-to-noise
ratio, leading to good precision in measuring absorbance.
A monochromator bandwidth that is 1/5 as wide as the absorption peak generally gives
acceptably small distortion of the peak shape.
Detectors
Photomultiplier tube
Emits electrons from a photosensitive, negatively charged surface (the cathode)
when struck by visible light or ultraviolet radiation.
Electrons emitted from the photosensitive surface strike a second surface, called
dynode, which is positive with respect to the photosensitive emitter.
This process is repeated several times, so more than 10 6 electrons are finally
collected for each photon striking the first surface.
As a result extremely low light intensities are translated into measurable electric
signals.
Resolution depends on how closely spaced the diodes are and how
much dispersion is produced by the polychromator
Practical aspects