HEINE EN100®
Mounting plate
3 x fixing plugs U6
3 screws 4 x 35
Instrument locks with instruction leaflets S-277.04.06
Safety Information
Range of uses: The HEINE EN100® is designed exclusively for use as
a power source for medical instruments with 3.5V bulbs. The HEINE EN100® should
only be operated by trained medical staff.
General Safety: This unit needs no maintenance. Do not open the housing.
Repairs should only be carried out by qualified specialists. If the unit is damaged,
disconnect the mains cord immediately.
Safety in use: Ensure that the mains voltage specified on the instrument label
conforms to the local mains voltage. Do not use in the presence of an explosion
hazard or in a strong magnetic field.
Connecting instruments
The HEINE EN100® cord handles are fitted with HEINE automatic connectors for
use with all similarly – equipped HEINE instruments.
– Hold the handle in one hand and insert the instrument connector into the
guides in the handle head.
– Press the instrument into the head until it clicks audibly into place.
– To remove the instrument, rotate the upper conical ring with thumb and index
finger to the right and, with the other hand, pull out the instrument.
– Instruments can be locked to prevent unauthorised removal by fitting the
instrument locks according to the instruction leaflet S-277.04.06.
Installation
– Connect the unit to the mains with the mains cord.
– Switch on, the green light operates.
– The handles switch on automatically when removed from the unit.
– The control ring on the handle can be rotated to select the desired brightness.
– Both handles can be simultaneously used independent of each other.
Cleaning
Disconnect the mains cord!
Clean and disinfect using 70% Isopropyl alcohol.
Technical specification
Voltage Fuse
Input: 100 V~, 50 /60 Hz 2 x 400 mA slow
(see nameplate) 120 V~, 50 /60 Hz* 2 x 400 mA slow
127 V~, 50 /60 Hz 2 x 400 mA slow
230 V~, 50 /60 Hz 2 x 200 mA slow
240 V~, 50 /60 Hz* 2 x 200 mA slow
Current consumption: 40 VA
Output voltage: 3.5 V
Output current: max. 800 mA
Protection class: I
Operation: Continuous with intermittent loading
Application part: Type BF
Dimensions: 285 x 205 x 85 mm
Weight: 1700 g
Ambient temperature: +10 °C to + 40 °C
Rel. Humidity: 10 to 75 %
Air pressure: 700 to 1060hPa
Protection class: IP X0
General Conditions of Warranty
Instead of the statutory warranty time period of 2 years, HEINE will grant a guarantee of 5 years
from the date of the consignment of the goods ex works, concerning its equipment (excluding dis-
posables, e.g. bulbs, single-use articles, and rechargeable batteries). Date of consignment means
that HEINE hands the goods over to the transport carrier, freight forwarder or any other person
designated by the Customer for the transport of the goods without loading the collecting vehicle.
The guarantee covers irreproachable workmanship, on condition of the proper use of the equip-
ment and the observation of the operating instructions. During the warranty- and guarantee
time period, errors and deficiencies arising on the equipment will be rectified free of charge,
in so far as such are evidenced by defective materials, processing and/or constructional errors.
Should buyer complain of a material deficiency during the warranty time period, then the onus
of proof is always to be on the orderer, that the product was defective already upon receipt of
the goods. The statutory warranty and the guarantee do not apply to loss or damage caused by
wear and tear, negligent use, the non-employment of original HEINE components and/or spares
(in particular bulbs, as these have been especially developed for HEINE instruments in accordance
with the following criterions: colour temperature, useful service life, safety, optical quality and
performance. The statutory warranty and the guarantee do not apply to interventions by persons
not authorised by HEINE or when the operating instructions are not observed by the customer.
Any modification of a HEINE product with parts or additional parts which do not conform to the
original HEINE specification will invalidate the warranty for the correct function of the product and
further invalidate any warranty claims which result from such a change or modification. Further
claims, in particular claims for replacement of loss or damage, which are experienced otherwise
than directly on the HEINE product itself, are hereby excluded.