v_user
v_user
Blood Analysis
Setting Up The Blood Analysis Module..................... 101
Viewing Blood Analysis Results ................................ 102
BIS
Setting up BIS measurements ..................................... 108
Adjusting BIS measurements .............................. 108
CO2 40
ETCO2
CHANNEL #4 0
20
IMCO 2 0 NBP 19:47
AWRR 37 115/65
(81)
Alarm Vol 165 QRS Vol 150
Setup Reminders
Setup Reminders
Basic Troubleshooting
Problem Possible Cause Corrective Action
Wave is not on the Transducer is not plugged Plug in transducer
screen. in.
2
Problem Possible Cause Corrective Action
Wave is not on the The wave has not been Assign wave to a channel:
screen. assigned to a channel. 1. Press Monitor Setup → Display 1 Setup .
2. Press Select Item
(to select required channel).
3. Press Change Content repeatedly to select
desired wave.
(Change speed and # of waves on screen or wave
layout as required.)
Setup Reminders
Setup Reminders
4
Parameter Setup Reminders
▲
Press softkey with parameter name
(e.g., ECG ).
Setup Reminders
Setup Reminders
Control Keys
Alarms Procedures
– to turn all alarms off/on and individual alarms on/ – to perform Cardiac Output measurements, make ST
off, adjust alarm limits, show the alarm messages, go segment monitoring adjustments, review ST wave
to standby mode and other functions. segments, admit and discharge patients or end a
particular patient case and transfer patient data.
Monitor Setup
– to change settings such as screen displays and Module Setup
patient type, set up oxyCRG and recording and make – to set up and change settings for all modules that are
general configuration changes. plugged into the Agilent V24 or V26.
Trends/Calcs
– View data in graphs and tables, print reports and
enter calculation task windows and arrhythmia
functions.
Other Patients
– to view data and alarms from other beds.
6
Silence/Reset
– silences an alarm and, if cause has been eliminated,
resets it.
Suspend
– Switches all alarms on/off.
Main Screen
– returns you to the standard monitoring screen.
Delayed Record
– starts a delayed recording.
Realtime Record
– leads you to a selection window, from which you
can start a realtime preset recording or a realtime
selected recording.
, , ▲, ▼
▲
Arrow Keys –
▲
Setup Reminders
Setup Reminders
8
Battery
Battery
Battery
10
Warning
When connecting a non-medical device such as a printer to the RS-232 Interface of an Agilent V24CT
or V26CT, the monitor must be properly grounded, or the non-medical device properly isolated. This
must be done in order to prevent the possibility of an electrical shock hazard (to the user and or
patient) from excessive ground leakage currents from the non-medical device such as a printer, and to
maintain compliance with IEC 601-1-1 where applicable. This hazard is especially prevalent when the
Agilent V24CT or V26CT is operated in battery mode.
In normal AC line operation grounding the monitor is accomplished by ensuring that the AC line cord
is connected from the monitor to a properly grounded AC receptacle. Alternatively and additionally,
proper grounding may be achieved by attaching a safety grounded wire to the threaded safety ground
stud found on the rear of the monitor.
Isolation may be acheived by powering the printer through a safety isolation transfomer.
11
Battery
Battery
12
Recording
Recording
Recording
14
Vital Signs/Blood Recording
15
Recording
Recording
oxyCRG Episode
To make an oxyCRG Episode recording:
1. Press Alarms
2. Press Event Review
3. Select an event with the cursor.
4. Press oxyCRG Episode
5. Press Record Episode
16
Recording Summary
Delayed Realtime/Preset Realtime/Your Selection
17
Recording
Recording
1
The name of the actual recorder appears on the key. This can be PLUG-IN, Central, 4CH BS-1 or 4CH BS-2.
18
Alarms
▲
▲
1. - To adjust the SpO2 Desat alarm press and .
- The Extreme Bradycardia alarm difference is
adjustable in Configuration mode only.
19
Alarms
Alarms
20
Overview/Arrhythmia
21
Overview/
Arrhythmia
Overview/
Arrhythmia
22
Turning off the autmatic pop up of the
Arrhythmia Validation task window:
1. Press Trends/Calcs
2. Press Rcv Arrhyth -> No
Press Rcv Arrhyth -> Yes to turn pop up back on.
23
Overview/
Arrhythmia
Overview/
Arrhythmia
24
Admit / Discharge / End Case
25
Admit / Discharge /
End Case
Admit / Discharge /
End Case
26
Trends / Calcs
1. Press and .
▲
27
Trends / Calcs
Trends / Calcs
28
Using Graphs
Selecting Parameters for Viewing no symbol: the parameter is not selected for
To select parameters for viewing in graph trends: viewing in graph trends.
1. Press Trends/Calcs . 4. Press Graph Trends to view graphs.
2. Press Vital/Signs to display trends. Note — Selected parameter graphs are only available
upon entry to graph trends. When you press
3. Select parameters using ▲ and ▼ and Next Group , the default graph groups are
press On / Off / Combined for each selected displayed. To see the selected parameter graphs again,
parameter. The system will try to combine select Vital/Signs , and in the Vital Signs window,
parameters in one graph channel. If this fails, a select Graph Trends .
new graph channel will be assigned
automatically.
29
Trends / Calcs
Trends / Calcs
1. Press and .
▲
30
Printing Graphs Viewing Values in Table Form
To print graphs: To view values in table form:
1. Press Print Graphs . 1. Press Trends/Calcs .
2. Press Vital Signs .
▲
3. Press and to move time bar to the
▲
preferred time.
4. If selected time is not displayed, press
Zoom In Time .
31
Trends / Calcs
Trends / Calcs
32
Calculations
Performing Calculations b. Enter new value using handheld keypad.
To perform calculations: c. Press Confirm
1. Press Trends/Calcs .
To print the displayed calculations
2. Press Hemo Calc , Ventil Calc or
a. Press Print Calc
Oxygen Calc to display the Calculations
Task Window (Displayed values may have been
automatically or manually entered.)
Optionally: Press Resample Vitals and Viewing Collection Times
check that the values are representative of the To view collection times of non-calculated values:
patient state at the current time. If values are not 1. Press On/Off Data Time
representative press Resample Vitals again.
Press Confirm to store values without
calculation. Specifying a Calculation Time
3. Press Perform Calc to store any resampled To specify a calculation time
values and perform calculations with the
1. Press Select Item until calculation time is
displayed values
highlighted.
To change or enter an input value
▲
2. Press and to move selection to hours/
▲
a. Press ▲ and ▼ to select value. minutes
33
Trends / Calcs
Trends / Calcs
▲
▲
4. Press Confirm
5. Press Perform Calc To modify input values used to perform
calculations:
Viewing Ranges a. Highlight the time of the calculation to be
To view ranges of calculated values:
▲
modified using the and keys
▲
1. Press On/Off Ranges b. Press Original Calc
c. Enter modifications, as needed:
35
Trends / Calcs
Trends / Calcs
▲
▲
Calculating
To calculate rate, dose, amount or volume when all
inputs are complete:
1. Press Perform Calc
36
Marking and Reviewing Events
To mark an event:
1. Press Trends/Calcs
2. Press Mark Event
3. Press an event key (labelled A-D). The message
“Mark recorded” appears.
To review an event:
1. Press Trends/Calcs
2. Press Graph Trends
3. move the time bar along the Events line by
▲
▲
4. Press oxyCRG Episode
To view newer events in greater detail To view previous episodes
▲
Neonatal Event
Review
Neonatal Event
Review
40
oxyCRG
oxyCRG Measurements
41
oxyCRG
oxyCRG
42
Data Transfer
Transfer to Module
1. Plug the Data Transfer module into the rack. b. Updates:
Press Collect New Data to send only
2. Press monitor updates once a minute starting at the
current time.
Press Clear Yes/NO .
or (for non-english). Yes erases any data existing in the module
before the transfer begins.
3. Select the type of transfer. No appends new data to data currently
a. All Data: existing in the module. If the patient name
Press Transfer All Data . Any existing and ID do not match or if the monitor name
data in the module will be erased. The is blank, the Combine in Module Task
monitor data will be transferred, followed by Window will be displayed. Select the patient
monitor updates once a minute. name to be used in the module.
4. Press Confirm to start the transfer.
43
Data Transfer
Data Transfer
Blinking
• Rapid blinking of module light indicates a trans-
fer in progress.
• Slow blinking indicates that the module is col-
lecting new data.
1. Press
or (for non-english).
2. Press Confirm .
When the module is unplugged, it retains data for at
least one hour.
44
Quick Transfer
Cautions: Transferring Quickly
• Be sure that the patient in the monitor matches To transfer quickly:
the one in the module before performing a trans- 1. Plug the Data Transfer module into the rack.
fer to the monitor. To prevent accidental over-
writing of patient data or name, the module will 2. Press
ask you to choose a name/MRN if the monitor
name and module name don't match or if there is
no name/MRN. or
• Be sure to discharge the previous patient at cen-
tral before performing a transfer, or the name or
moved into the monitor from the module will be
overwritten by the name from the central moni-
or
tor.
Notes: (for non-english) to select transfer direction.
• After insertion, the Data Transfer Module 3. Press Confirm .
requires approximately 8 seconds to synchronize Remove the module when the transfer is
with the monitor. complete.
• If you are transferring data to the Component
Transport Monitor, erase any existing data in the
transport monitor before performing the transfer.
45
Data Transfer
Data Transfer
Transfer to Monitor
2. Press Confirm to start the transfer and
1. Plug the Data Transfer module into the rack. incorporate the module data into the
monitor's existing patient record.
2. Press c. If Discharge = Yes, the Discharge Task
Window will be displayed.
or (for non-english). 1. Press Confirm to discharge the
monitor patient first and then start the
3. Select whether or not to discharge the current transfer of data from the module.
patient data in the monitor before starting the
transfer.
Stopping a Transfer to Monitor
a. If Names Match, press Confirm to start
the transfer without discharging the patient in To Stop a Transfer to the Monitor:
the monitor. Unplug the Data Transfer module from the rack.
b. If Names Don’t Match (or either is blank) After a transfer is completed, the data in the module is
and Discharge = No, the “Combine in erased. If the transfer is interrupted, the module data
Monitor” Task Window will be displayed. will remain intact.
1. Select the patient name to be used in the
monitor.
46
ECG / RESP / ST
ECG Measurements
47
ECG / RESP / ST
ECG / RESP / ST
Adjusting the Size of the QRS Selecting the QRS Detection Method
To adjust the size of the QRS complex: To select the QRS detection method:
1. Press ECG on module. 1. Press ECG on module.
2. Press Adjust Size . 2. Press Auto/Man & Pacing .
PACED or NONPACED / MANUAL or AUTO.
Use AUTO except in rare cases where the
Selecting the Filter Setting monitor miscounts - if change to MANUAL,
To select the filter setting: adjust detection level so only QRS complexes
cross dotted threshold line.
1. Press ECG on module. (not available when arrhythmia is assigned).
2. Press Filter Mon/Diag .
FILTER - use only if excessive electrical Turning Off Channels 2 and 3
interference (can impede ECG analysis). To turn off channels 2 and 3:
MONITOR - use for normal monitoring
DIAG - use when diagnostic quality required 1. Press ECG on module.
(may increase false alarms). Press Setup Next ECG .
Press On/Off ECG-CH2 to highlight off.
(Channel 3 is turned off automatically.)
48
Respiration Measurements
49
ECG / RESP / ST
ECG / RESP / ST
ST Measurement Points
R-WAVE PEAK AT 0 MSEC
J POINT
T
P
DIFFERENCE =
ST VALUE
Q
S
ISOELECTRIC POINT ST MEASUREMENT POINT
DEFAULT = R – 80 MSEC DEFAULT = R + 108 MSEC
The ST measurement for each beat complex is the You can adjust both measurement points. Your
vertical difference between the points where the wave monitor is configured so that you set the ST Point in
intersects two measurement points: reference to: either the R-wave peak or the J Point,
1. The isoelectric point provides the baseline for the which is the transition between the QRS and the ST
measurement. segment. The ST point is referenced from the J point
at either +60 or +80 msec.
2. The ST point provides the other measurement
point.
50
ST Segment Monitoring
51
ECG / RESP / ST
ECG / RESP / ST
52
ECG Troubleshooting
Problem Possible Cause Corrective Action
No display of ECG System not plugged in or switched on. ECG Connect and switch on system. Check ECG or
or ECG/RESP module is not plugged in. ECG/RESP Module is plugged in correctly.
System not connected to satellite module Check that contrast control is correctly
rack. adjusted.
LEADS OFF message. ? ECG cable is not connected. One or more Check ECG cable is connected. Check that
instead of HR/RESP electrodes are detached. No gel on electrodes are not detached. Check that
numerics. INOP tone. electrodes. Strong solvent used for cleaning electrodes have enough gel. Change to a
skin. A 3-lead set is used and channels 2 and suitable skin cleaning solvent. Ensure that
3 are not switched off. channels 2 and 3 are switched off if a 3-lead set
is used.
ECG waveform is present One of the RESP electrodes is loose. No gel Check that RESP electrodes are not loose.
but the RESP channel on the electrodes. Patient’s breathing is too Check that electrodes have enough gel.
shows a flat line. shallow to be detected in auto mode. Reposition electrodes to try and optimize
signal. Change RESP detection mode from
auto to manual if necessary.
NO PULSE SOURCE The pulse source (PRESS or SpO 2/PLETH Check that module is plugged in. Check that
message (only if PULSE is Module) is unplugged, or the transducer is transducer is connected.
the active alarm parameter). not connected, or equipment malfunction.
53
ECG / RESP / ST
ECG / RESP / ST
54
Invasive Pressure
Pressure Measurements
55
Invasive Pressure
Invasive Pressure
56
Pressure Troubleshooting
Problem Possible Cause Corrective Action
Noise or fling in the pressure Movement of P.A. catheter tip. Reposition catheter (in accordance with
waveform Motion of pressure tubing. your Hospital Procedures Policy). Make
sure patient or ventilator tubing is not
touching pressure tubing.
Resonating Arterial Small air bubbles in, or very close Carefully flush the transducer and tubing
Waveform; to the transducer. when setting up. Debubble carefully after
each flush.
Dampened Arterial Thrombus formation or blood left Use syringe to withdraw air or particles in
Waveform in catheter following blood catheter, then flush the line with fresh
sampling. solution.
57
Invasive Pressure
Invasive Pressure
Catheter tip against the vessel wall. Reposition the catheter to relieve the spasm.
Catheter kinking or arteriospasm.
Tubing too long or too compliant. Shorten the tubing (recommended length
3-4 ft. or replace with large diameter stiff
tubing.
Abnormally Low Pressure Transducer level higher than heart. Check patient and transducer positions.
No Pressure If monitor turned on and there is no Replace transducer and cable. Check
pressure. The transducer or stopcocks.
connector wire could be broken.
58
NiBP
NBP Measurements
NiBP
NiBP
Stopping a Measurement
To STOP a measurement:
1. Press Stop NBP .
or
STOP on module.
60
NBP Troubleshooting
The following factors can make the NBP measurement unreliable or impossible to detect:
Patient movement: Movement, shivering and Heart rate extremes: Measurements can not be
convulsions may all interfere with the measurement made at a heart rate of less than 40 bpm or greater than
of the arterial pressure pulses. Measurement time will 300 bpm.
also be increased Incorrect cuff size: Using the incorrect cuff size
Cardiac arrhythmias: An irregular heart beat will will give erroneous readings
make the measurement unreliable or impossible. The Obese patients: A thick layer of fat surrounding a
measurement time will be increased. limb tends to dampen oscillations coming from the
Heart-lung machine: Measurements will not be artery, and stops them from reaching the cuff.
possible if the patient is connected to a heart-lung Accuracy may be lower than normal.
machine.
Pressure changes: Rapid changes in blood
pressure, while the arterial pressure pulses are being
analyzed, may make the measurement unreliable or
impossible.
Severe shock: Reduced blood flow to the
peripheries will cause reduced pulsation to the
arteries in conditions of severe shock or hypothermia.
This will make measurements unreliable.
61
NiBP
NiBP
62
SpO2/Pleth
SpO2/Pleth Measurements
63
SpO2/Pleth
SpO2/Pleth
64
SpO2 Troubleshooting
The following factors can cause errors in the measurement of the SpO2
65
SpO2/Pleth
SpO2/Pleth
66
Problem Possible Cause Corrective Action
Erratic readings, or Using inappropriate sensor for patient Select correct sensor based on patient weight,
screen messages: or clinical condition. application site, clinical condition.
SpO2 non pulsatile, Sensor applied to inappropriate site. Apply sensor to recommended site.
inaccurate or Weak
Signals Sensor applied incorrectly. Apply sensor to recommended site, check LED/
detector are lined up, check sensor still sticks.
Patient movement. Secure sensor cable with tape to prevent it from
coming off.
Poor sensor adhesion. Use new sensor if necessary. Do not use
additional tape.
Using too many adhesive dots. Do not use more than 3 dots for each LED and
detector. Do not use dots with sensors for which
they were not intended.
Tissue between LED and detector too Apply sensor to thinner or thicker tissue site.
thick or too thin.
Different values on Cardiac shunts. Check with blood sample. Accuracy of value
different extremities questionable, depending on shunt size.
67
SpO2/Pleth
SpO2/Pleth
68
CO2 & ssCO2
Adjustments to CO2
After calibrating the sensor, attach it to either: c. AWRR: set apnea delay time and low and
• mainstream - the patient's breathing circuit. high limit
• sidestream - the sidestream module 5. Press On/Off Alarms to turn alarms on.
(No next to parameter numeric.)
1. Press CO2 on module to bring up Task
Window. 6. Press Main Screen .
2. Press Change Scale to change size of The following waveform represents a normal
waveform. capnogram:
69
When to calibrate.
• If a new (or different) sensor is attached to the
module.
• When the Accuracy Check values displayed do
not coincide with the calstick value.
• Once every 1-2 weeks.
70
Performing an Accuracy Check If calstick and screen values are differ-
Note —This should be done once a day ent
1. Press CAL on module, and check calstick 1. Press ↓ Change Cal Value ↑ to set the
value on the screen. correct value.
2. Press Confirm .
If calstick and screen values are the 3. Go to Perform the calibration.
same
1. Place sensor on cell labelled 0.0 mmHg. Reading Perform the Calibration
on display should be within +0.3mmHg within 1
1. Place sensor on calstick cell. Wait 3 minutes.
minute.
2. Press Start Calibr. .
2. Place sensor on other cell. Reading on display
should be +1mmHg of calstick value within 1 3. Repeat process with other cell. (For both cells,
minute. values on screen should match values on
calstick.)
3. If the values match, you don’t need to calibrate;
Press Main Screen . 4. Press Main Screen .
4. If values don’t match, you need to calibrate. Press A more detailed version of this procedure can be
Start Calibr. . found in the Agilent CMS, V24 and V26 Series User’s
Reference Manual.
5. Go to Perform the calibration.
71
To Patient
Connects
to Module
Inspiratory
Airway Adapter Limb
and CO2 Sensor
Y-Piece
To Anesthesia
Machine
Expiratory
Limb
To Anesthesia
Machine
72
Sidestream Module, CO2 Transducer and Breathing Circuit
Bacterial Filter
CO 2 CAL
Transducer
Connects to Patient
73
74
CO2 and Sidestream Mode Troubleshooting
Problem Possible Cause Corrective Action
CO2 wave line Module not plugged in, transducer Check that module and accessories are plugged
disappears not connected, Equipment in. Sidestream - check pump switched on.
malfunction, sidestream pump off.
CO2 wave set to zero, Not in monitoring mode, calibration Change filter (may be blocked, check sample
other numerics taking place or failed, barometric tube (may be kinked). Check whether monitor in
displayed as -?- preesure out of range. calibration mode. Check that transducer is
placed on correct cal cell. Call service engineer
to check barometric pressure.
Wave Display Clipped Incorrect scale selected - complete Change wave scale size, by pressing CO2 on
wave display cannot be seen.
module then Change Scale .
75
76
tcpO2/tcpCO2
tcpO2/tcpCO2 Measurements
77
tcpO2/tcpCO2
tcpO2/tcpCO2
When to Calibrate:
• After you have fitted a new membrane on the
transducer.
• After you have changed the transducer operating
temperature.
• When INOP message “tcCAL Required” is dis-
played on the screen
78
Tc Gas Troubleshooting
Problem Possible Cause Corrective Action
CAL FAILED inop 1. Poor supply of calibration Check cal unit, tubing and tubing
gas. connections, then restart calibration.
79
tcpO2/tcpCO2
tcpO2/tcpCO2
Old electrolyte not fully removed Place transducer in cal chamber and turn on
during remembraning procedure. gas. Do not calibrate. If transducer does not
give stable reading after 5 minutes,
remembrane, taking care to clean out old
electrolyte. Check groove. Calibrate.
tcpO2 values too low or Capilliary samples taken Take arterial samples. Capilliary samples do
too high (Astrups, heelsticks). not correlate well with paO2.
Blood taken while patient Take another sample when patient is quiet.
agitated (crying, feeding,
suctioning, positioning).
80
Problem Possible Cause Corrective Action
tcpO2 values too low Hypoperfusion. No corrective action possible. tcpO2 levels
will underestimate paO2.
tcpO2 values too high Incorrect handling of blood Take another sample. Ensure that
sample. atmospheric oxygen does not enter sample.
tcpCO2 values too high Correction factor not ON. Either ask biomed or service to configure
correction ON, or take the rise of pCO2 due
to heating and metabolic CO2 production into
account.
tcpCO2 values too low Transducer improperly applied. Reapply fixation ring.
Air leaks. Reattach transducer.
81
tcpO2/tcpCO2
tcpO2/tcpCO2
82
Cardiac Output
5. When “Inject Now” appears, inject. (For 9. Press Confirm to store average in Trends.
accuracy start within 15 seconds - complete in 4 10.Press Main Screen .
seconds.)
To perform hemodynamic calculations:
1. Press Hemo Calc
83
Cardiac Output
Cardiac Output
84
Screen Message Possible Cause Corrective Action
1. Noisy Baseline Mechanical Ventilation Synchronize injections to end-expiration
2. Thermal Baseline Drift Administration of I.V. fluids; Try to stabilize patient temperature; Maximize
Rewarming after hypothermia, signal amplitude (as above); Increase number of
fever, shock measurements to be averaged.
85
Cardiac Output
Cardiac Output
7. Very Long Curve Injection too slow Inject again as soon as possible after the prompt.
9. Delayed injection Low stroke volume, Low flow rate Inject as soon as possible after prompt.
Low flow rates need 15-20 seconds to pass by
thermistor.
10. Injectate temperature too Incorrect positioning of injectate Make sure that probe is positioned in a
high probe solution identical to the one injected.
Hypothermic patient <34°C, 93.2°F. Use iced injectate with hypothermic patients.
Room temperature injectate (Remember to change comp. constant.)
86
Screen Message Possible Cause Corrective Action
11. Excessive Curve Height Injectate too cold Heat up injectate and start measurement
again
12. Curve Below Baseline Rewarming after hypothermia, Try to stabilize patient temperature and start
fever, shock measurement again.
87
Cardiac Output
Cardiac Output
88
Temp
Temperature Measurements
89
Temp
Temp
90
Anesthesia Gas Monitoring
91
Anesthesia Gas
Monitoring
Anesthesia Gas
Monitoring
s$JLOHQW M1026A
R
Airway Gases
2
3
4
1
92
Setting Up Anesthetic Gas Module with Watertrap (Option A02)
1. Insert a watertrap into the watertrap housing by 2. Switch on the Anesthetic Gas Module by
gently pushing it up and in. pressing the power-on switch.
3. Connect the gas sample tube to the water
separation filter.
Warning
Ensure that the connections are tight when fitting
the tube. Any leak in the system can result in
erroneous readings due to ambient air mixing with
patient gases.
93
Anesthesia Gas
Monitoring
Anesthesia Gas
Monitoring
1.
2.
3.
5.
7. 4.
6.
1. Watertrap Housing 5. 13902A/M16112A Airway Adapter
2. M1657A Watertrap 6. Connects to patient
3. Water Separation Filter 7. Connects to anesthesia machine.
4. M1658A Gas Sample Tube
94
Accessing the Airway Gas Task Window
To Access the Airway Gas Task Windows:
1. Press the Airway Gases set-up key on the
module or Module Setup followed by
Airway Gases to bring up the Overview
Task Window.
2. Select the airway gas task window you wish to
access by pressing:
Airway CO2 .
Airway N2O .
Airway Agent .
Airway O2 (optional).
3. Press the item you want to adjust: for example,
scales or alarms.
4. Press Main Screen when finished.
95
Anesthesia Gas
Monitoring
Anesthesia Gas
Monitoring
Standby Mode
96
Calibration
Span Calibration:
The Anesthetic Gas Module should be checked at
least once per year to determine if span calibration is
necessary.
97
Anesthesia Gas
Monitoring
Anesthesia Gas
Monitoring
98
Automatic agent selection Manual agent selection
To select automatic agent identification (the default): To manually select an anesthetic agent:
1. Press the Airway Gases set-up key on the 1. Press the Airway Gases set-up key on the
module or Module Setup followed by module or Module Setup followed by
Airway Gases to bring up the Overview Airway Gases to bring up the Overview
Task Window. Task Window.
2. Press Airway Agent 2. Press Airway Agent
3. Press Select Agent 3. Press Select Agent
or → to select the agent you want to
→
4. Select Auto for automatic agent identification 4. Use
5. Press Main Screen when finished. measure, for example ISO for isoflurane or
None if no specific anesthetic agent is being
measured.
The measured agents are as follows:
Anesthetic Agents Display Abbrev.
Halothane HAL
Isoflurane ISO
Enflurane ENF
Sevoflurane SEV
Desflurane DES
5. Press Main Screen when finished.
99
Anesthesia Gas
Monitoring
Anesthesia Gas
Monitoring
100
Blood Analysis
To enter the operator ID: To enter an O2 value or information in one of the free-
Press Select Oper.ID entry fields:
then either: Press Select O2/Field once for O2 entry
Select a pre-configured ID or
or: Press Select O2/Field twice to get to free
Press Manual Entry to enter your ID field entry then use the arrow keys to enter the
manually using the arrow keys. data.
101
Blood Analysis
Blood Analysis
102
Blood Analysis Troubleshooting
Screen Message Possible Cause Corrective Action
Sample Positioned Short The cartridge was underfilled. Try another cartridge.
of Fill mark (XX)1 The sample must reach the fill mark.
Insufficient Sample Either not enough sample in the sample Try another cartridge.
(XX)1 well of the cartridge or bubbles in the
sample.
Unable to Position The module did not detect movement of Try another cartridge.
Sample (XX)1 sample across sensors:
• the cartridge was not closed properly
• a clot in the sample prevented
movement, or
• aberrant cartridge
103
Blood Analysis
Blood Analysis
Cartridge not Inserted The cartridge or electronic simulator is Reinsert the cartridge or simulator. If
Properly not pushed in all the way. the problem recurs it may indicate a
module problem. Call your service
representative.
104
Screen Message Possible Cause Corrective Action
Temperature Out of The module is too warm or too cool If the module has been moved, leave it
Range because it has just been moved from, or for 30 minutes to reach room
is now in, a room that is too warm or too temperature before use. If not, check
cool. that room temperature is between 16
and 30 degrees Celsius.
Cartridge Error, Use This error can have various causes Try another cartridge. If one error
Other Cartridge (XX)1 related to the equipment, the sample or code appears repeatedly, the module
the user. may be defective. In this case, try
another module or contact your
service representative.
1
XX represents an error code which may assist the service representative to diagnose the problem.
105
Blood Analysis
Blood Analysis
106
BIS
BIS
Adjusting the size of the EEG wave Stopping a Cyclic Impedance Check
Press Change Scale Press Stop Check in the Cyclic Check Task
Window.
108
109
BIS
Agilent V24/26
Web Link
110
Adding a page to the Favorites Menu 2. Change the name when it becomes boxed.
1. Navigate to the page you want to add. 3. Press Enter . The name is now changed.
2. Click the Favorites icon on the toolbar.
Deleting a Favorites link
3. Click the Add button. Enter the name of the link
you want to create (i.e. TraceMaster Web) 1. Right click on the link in the favorites list and
select Delete from the menu.
4. Click OK . The link is now added to the
Favorites list.
111
Agilent V24/26
Web Link
Agilent V24/26
Web Link
Welcome Screen • System not plugged in or • Check the system connections and verify
does not display switched on. the display switch is in the correct posi-
when the link is tion.
selected. • The link to the Welcome page is • Check the link information by right click-
broken. ing on the link and selecting Properties.
Select the Web Document tab and verify
the URL box has the correct path. If not,
type the correct path and click OK.
112
Problem Possible Cause Corrective Action
No Welcome/ • The Favorite menu items have • Re-enter the Favorite menu application
application page(s) been erased. pages. Refer to the V24/26 Web Link
are listed in the Installation Manual for instructions if
Favorites menu. • The links that point to the necessary.
Favorites menu are broken. • Check the links by placing the cursor on
the link and right clicking and selecting
Properties. Select the Web Document tab
and verify the URL box has the correct
path. If not, type the correct path and
click OK.
113
Agilent V24/26
Web Link
Agilent V24/26
Web Link
Application’s page • The server that hosts that applica- • Check the other pages to see if they work.
shows an error when tion may be down. If they do, than that particular applica-
selected tion’s server may be experiencing trou-
ble. If others are also inaccessible, then
the V24/26 Web Link PC may need to be
rebooted.
Note — When you reboot the PC, it does
not effect the V24/26 monitor. (The PC
• The page is not configured cor- reboots in the background, indepedent of
rectly. the V24/26 monitor.)
• Follow the directions on the screen. If no
directions display, check each link by
right clicking the link and selecting Prop-
erties. Select the Web Document tab and
verify the URL box has the correct path.
If not, type the correct path and click OK.
Keyboard/Trackball • The cable that connects the key- • Verify the keyboard/trackball cable is
isn’t working board/trackball to the PC is loose. securely attached to the PC.
114
Problem Possible Cause Corrective Action
Monitor doesn’t • The cable that connects the PC to • Verify the PC and monitor cables are
display the correct the monitor is loose. securely attached to each other.
information
115
Agilent V24/26
Web Link
Index
A BIS 108 D
Absorption of light by other chemicals in the Blinking 44 Data Transfer 43
blood 65 Blood Analysis 101 Deactivating the Standby Mode 96
Accessing the Airway Gas Task Window 95 Bottom Row 7 DELAYED recording (normally ECG) 13
Activating the Standby Mode 96 Deleting a Favorites Link
Adding a page to the Favorites Menu C V24/26 Web Link 111
V24/26 Web Link 111 Calculating 36 Deriving Pulse From SpO2/Pleth 64
Adjusting Alarm Limit 55 Calculations 33 Discharging a Patient 26
Adjusting Anesthetic Gas Module Alarms Calibrating CO2 Measurements 70 Displaying Delayed Wave 22
100 Calibration 97 Displaying the Titration Table 36
Adjusting EEG measurements 108 Canceling a Report 35
Adjusting ST Alarms 52 Cardiac arrhythmias 61 E
Adjusting the Measurement Points 51 Cardiac Output 83 EASI 12-Lead ECG 47
Adjusting the Size of RESP Wave 49 Cardiac Output Troubleshooting 84 ECG / RESP / ST 47
Adjusting the Size of the QRS 47 Changing Alarm Limits 78 ECG Measurements 47
Adjusting Trigger Conditions 40 Changing Oxygen Channel 41 ECG Troubleshooting 53
Adjustments to CO2 69 Changing Repetition Time 59 Ending a Case 25, 37
Agilent V24/26 Web Link 110 Changing Site Time or Reset Timer 77 Expanding the Vertical Axis 30
Agilent V24/26 Web Link Troubleshooting Changing the Pressure Label 55
112 Changing the Recorder 14 F
Alarms 19 Changing the Reference Wave 52
fitting the gas sample tube to the watertrap
All Alarms On 20 Changing the Waveform Size 55
93
Anesthesia Gas Monitoring 91 Changing Transducer Temperature 77
Assigning an Unknown Entity Field 36 Changing Trend Priorities 28
Automatic agent identification 98 Charging Batteries 9 H
Automatic agent selection 99 CO2 69 Heart rate extremes 61
Automatic Zero Calibration 97 CO2 Transducer and Breathing Circuit 72 Heart-lung machine 61
Configuring to be the Alarm Recording 42 High levels of ambient light 65
B Control Keys 6 High levels of electrical interference 65
Basic Troubleshooting 2
Battery 9
117
I Overview Recording Multi-Lead ECG 47
If calstick and screen values are different 71 V24/26 Web Link 110 Recording Summary 17
If calstick and screen values are the same 71 Overview/Arrhythmia 21 Reduced arterial blood flow 65
Incorrect cuff size 61 oxyCRG 41 Respiration Measurements 49
Individual Alarms On/Off 20 oxyCRG Episode 16 Reviewing Calculations 34
Injection errors 84 oxyCRG Episodes 39
Instrument errors 84 oxyCRG Measurements 41 S
Invasive Pressure 55 Selecting Agent Identification 98
P Selecting Parameters for Viewing 29
M Parameter Setup Reminders 5 Selecting the Filter Setting 48
Making a Single Recording 15 Patient Data 25, 27 Selecting the QRS Detection Method 48
Making a Stat Measurement 59 Patient movement 61 Selecting the Required Lead 47
Making a Trended Recording 15 Perform the Calibration 71 Selecting The RESP Detection Method 49
Making an oxyCRG Recording 41 Performing Calculations 33 Sending/Receiving Alarms 21
Making Cardiac Output Measurements 83 Physiological reasons 84 Set/Change Alarm Limits 19
Making Drug Calculations 35 Pressure changes 61 Setting Up Anesthetic Gas Module 91
Making Recordings 13 Pressure Measurements 55 Setting Up Anesthetic Gas Module - Option
Manual agent selection 99 Pressure Troubleshooting 57 A02 93
Manual Zero Calibration 97 Printing Displayed Calculations 35 Setting Up Arrhythmia 22
Middle Row 6 Printing Graphs 31 Setting Up SpO2 63
Modifying an Input Field 35 Printing Scheduled Reports 32, 35 Setting Up ST Monitoring 51
Printing the Titration Table 36 Setting Up Temperature 89
N Printing Vital Signs 27 Setting Up The Blood Analysis Module 101
Setup Reminders 1
NBP 59
NBP Measurements 59 Q Severe shock 61
Quick Transfer 45 Showing Normal Ranges 35
NBP Troubleshooting 61
Sidestream Module, CO2 Transducer and
Neonatal Event Review 16, 39
R Breathing Circuit 73
Neonatal Events 39
Silence/Reset Alarms 19
REALTIME (configured) recording 13
Single Manual Measurement or Auto Cycle
O REALTIME (non-configured) recording 13
59
Obese patients 61 Recording 13
118
Index
Index
119