PM6685R
PM6685R
Frequency Counters
PM6685 & PM6685R
Operators Manual
4822 872 20101
May 2003 First Edition
II
Table of Contents
1 Preface 6 Measuring Functions
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 Introduction to This Chapter . . . . . . . . . . . . 6-2
Design Innovations . . . . . . . . . . . . . . . . . . . 1-2 Selecting Function . . . . . . . . . . . . . . . . . . . 6-2
New Fast SCPI Bus . . . . . . . . . . . . . . . . . . 1-4 Frequency . . . . . . . . . . . . . . . . . . . . . 6-3
Rubidium Counter. . . . . . . . . . . . . . . . . . . . 1-4 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Burst Frequency and PRF . . . . . . . . . . . . . 6-6
2 Safety Instructions AM Signals . . . . . . . . . . . . . . . . . . . . . . . . 6-10
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Ratio. . . . . . . . . . . . . . . . . . . . . . . . . 6-12
Safety Precautions . . . . . . . . . . . . . . . . . . . 2-2
Introduction . . . . . . . . . . . . . . . . . . . . . . . . 6-12
3 Preparation for Use Period . . . . . . . . . . . . . . . . . . . . . . . 6-13
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . 6-13
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . 3-2 Pulse Width and Duty Factor. . . . . 6-14
Introduction . . . . . . . . . . . . . . . . . . . . . . . 6-14
4 Using the Controls Pulse Width A . . . . . . . . . . . . . . . . . . . . . . 6-15
About This Chapter . . . . . . . . . . . . . . . . . . . 4-2 Duty Factor . . . . . . . . . . . . . . . . . . . . . . . . 6-15
The User Interface . . . . . . . . . . . . . . . . . . . 4-2 Totalize . . . . . . . . . . . . . . . . . . . . . . 6-16
Default Settings . . . . . . . . . . . . . . . . . . . . . 4-3 Totalize A Manual . . . . . . . . . . . . . . . . . . . 6-16
Basic Controls. . . . . . . . . . . . . . . . . . . . . . . 4-4
Input A Controls . . . . . . . . . . . . . . . . . . . . . 4-5 7 Measurement Control
Measurement Control Keys . . . . . . . . . . . . 4-6 About This Chapter . . . . . . . . . . . . . . . . . . . 7-2
Display Controls . . . . . . . . . . . . . . . . . . . . . 4-7 Measuring Time . . . . . . . . . . . . . . . . . . . . . 7-2
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8 Display Hold . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9 Arming . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . 4-10 Digit Blanking . . . . . . . . . . . . . . . . . . . . . . . 7-5
Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . 4-11 Controlling Measurement Timing . . 7-6
PM6685R Front Panel . . . . . . . . . . . . . . . 4-12 Arming Setup Time . . . . . . . . . . . . . . . . . . 7-12
Arming Examples . . . . . . . . . . . . . 7-13
5 Input Signal Conditioning Example #1: . . . . . . . . . . . . . . . . . . . . . . . 7-13
Introduction to this chapter . . . . . . . . . . . . . 5-2 Example #2. . . . . . . . . . . . . . . . . . . . . . . . 7-16
Input A . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2 Example #3. . . . . . . . . . . . . . . . . . . . . . . . 7-17
Input C . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Reducing Noise and Interference . . . . . . . . 5-6
III
8 Processing
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Nulling. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
9 Auxiliary Functions
About This Chapter. . . . . . . . . . . . . . . . . . . 9-2
Auxiliary Menu . . . . . . . . . . . . . . . . . . . . . . 9-2
10 Performance Check
General Information . . . . . . . . . . . . . . . . . 10-2
Recommended Test Equipment . . . . . . . . 10-2
Front Panel Controls. . . . . . . . . . . . . . . . . 10-3
Short Form Specification Test . . . . . . . . . 10-5
Rear Input/Output . . . . . . . . . . . . . . . . . . . 10-6
Measuring Functions . . . . . . . . . . . . . . . . 10-7
PM9621, PM9624 . . . . . . . . . . . . . . . . . . . 10-8
PM9623 . . . . . . . . . . . . . . . . . . . . . . . . . . 10-8
11 Preventive Maintenance
Calibration and Adjustment. . . . . . . . . . . . 11-2
Other Maintenance . . . . . . . . . . . . . . . . . . 11-4
12 Specifications
Measuring Functions . . . . . . . . . . . . . . . . 12-2
Input and Output Specifications . . . . . . . . 12-3
Auxiliary Functions. . . . . . . . . . . . . . . . . . . 12-4
GPIB . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-5
Battery Unit . . . . . . . . . . . . . . . . . . . . . . . . 12-6
Measurement Uncertainties . . . . . . . . . . . 12-7
Timebase Options. . . . . . . . . . . . . . . . . . . 12-8
General Specifications . . . . . . . . . . . . . . . 12-9
Ordering Information. . . . . . . . . . . . . . . . 12-10
13 Appendix
Appendix 1, Error Messages . . . . . . . . . . 13-2
14 Index
15 Service Centers
SERVICE CENTERS . . . . . . . . . . . . . . . . 15-II
IV
LIMITED WARRANTY & LIMITATION OF LIABILITY
Each Fluke product is warranted to be free from defects in material and workmanship under normal
use and service. The warranty period is one (1) year for the instrument and for its accessories. The
warranty period begins on the date of shipment. Parts, product repairs and services are warranted for
90 days. This warranty extends only to the original buyer or end-user customer of a Fluke autho-
rized reseller, and does not apply to fuses, disposable batteries or to any product which, in Fluke’s
opinion, has been misused, altered, neglected or damaged by accident or abnormal conditions of op-
eration or handling. Fluke warrants that software will operate substantially in accordance with its
functional specifications for 90 days and that it has been properly recorded on non-defective media.
Fluke does not warrant that software will be error free or operate without interruption.
Fluke authorized resellers shall extend this warranty on new and unused products to end-user cus-
tomers only but have no authority to extend a greater or different warranty on behalf of Fluke. War-
ranty support is available if product is purchased through a Fluke authorized sales outlet or Buyer
has paid the applicable international price. Fluke reserves the right to invoice Buyer for importation
costs of repair/replacement parts when product purchased in one country is submitted for repair in
another country.
Fluke’s warranty obligation is limited, at Fluke’s option, to refund of the purchase price, free of
charge repair, or replacement of a defective product which is returned to a Fluke authorized service
center within the warranty period.
To obtain warranty service, contact your nearest Fluke authorized service center or send the prod-
uct, with a description of the difficulty, postage and insurance prepaid (FOB Destination), to the
nearest Fluke authorized service center. Fluke assumes no risk for damage in transit. Following
warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If
Fluke determines that the failure was caused by misuse, alteration, accident or abnormal condition
of operation or handling, Fluke will provide an estimate of repair costs and obtain authorization be-
fore commencing the work. Following repair, the product will be returned to the Buyer transporta-
tion prepaid and the Buyer will be billed for the repair and return transportation charges (FOB
Shipping Point).
THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF
ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO
ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL
OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER
ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE
OR ANY OTHER THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclu-
sion or limitation of incidental or consequential damages, the limitations and exclusions of this war-
ranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforce-
able by a court of competent jurisdiction, such holding will not affect the validity or enforceability
of any other provision.
Fluke Corporation, P.O. Box 9090, Everett, WA 98206-9090 USA, or
Fluke Industrial B.V, P.O. Box 90, 7600 AB, Almelo, The Netherlands
V
DECLARATION OF CONFORMITY
for
Fluke
EN 61010-1 (1997)
Safety Requirements for Electrical Equipment for Measurement, Control, and
Laboratory Use
EN 61326/A1 (1998)
EMC Requirements for Electrical Equipment for Measurement, Control, and
Laboratory Use
The tests have been performed in a
typical configuration.
This Conformity is indicated by the symbol ,
i.e. “Conformité Européenne”.
VI
Chapter 1
Preface
Preface
No Mistakes
Introduction
You will soon find that your new counter is a
Your PM6685 Counter is designed to bring you delight to operate. One example is the backlight
a new dimension to portable and bench-top LCD that shows you measurement results, set-
counting. It offers significantly increased per- ting status, and operator messages. The AUTO
formance compared to traditional counters. The function triggers automatically on any input
counter offers the following advantages: waveform. A bus-learn mode simplifies GPIB
programming. With bus-learn mode, manual
– Ten digits of frequency resolution per sec-
counter settings can be transferred to the con-
ond and 250 ps resolution, as a result of
troller for later reprogramming. There is no
high-resolution interpolating reciprocal
need to learn code and syntax for each individ-
counting
ual counter setting if you are an occasional bus
– A 1.3 GHz or a 3.0 GHz input frequency user.
option
– A foolproof autotrigger function
1-2 Introduction
Preface
neously with the normal “digital” counting, the The counter’s microcomputer calculates the re-
counter makes analog measurements of the sult after completing all measurements, i.e., the
time between the start/stop trigger events and digital time measurement and the two interpo-
the next following clock pulse. This is done in lation measurements.
two identical circuits by charging an integrating
capacitor with a constant current, starting at the The result is that the ba-
trigger event. Charging is stopped at the leading sic “digital resolution”
edge of the first following clock pulse. The of ± 1 clock pulse (100
stored charge in the integrating capacitor repre- ns) is reduced to 0.25%
sents the time difference between the start trig- of a clock pulse cycle, or
ger event and the leading edge of the first fol- 250 ps.
lowing clock pulse. Since the measurement is synchronized with
When the “digital” part of the measurement is the input signal, the resolution for frequency
ready, the stored charges in both capacitors are measurements is very high and is independent
measured. The capacitors are discharged with a of frequency.
constant current, which is only 1/400:th of the The Counter has 10 display digits to ensure that
charging current, which means that the dis- the display does not restrict the display resolu-
charge time will be 400 times the charging time. tion. It also has an overflow function that lets
This 400-fold stretched time is digitally mea- you see digit 11 and 12.
sured by the counter itself, with adequate reso-
lution.
# o f d ig its
1 0 M e a s u r e tim e = 1 s
9 M e a s u r e tim e = 1 0 0 m s
8 M e a s u r e tim e = 1 0 m s
7 M e a s u r e tim e = 1 m s
1 k 1 0 k 1 0 0 k 1 M 1 0 M 1 0 0 M
F re q u e n c y (H z )
The counter is not only an extremely powerful The GPIB interface, the Analog Output in-
and versatile bench-top instrument, it also fea- cluded, makes up the PM9626/031 option that
tures extraordinary IEEE-488 bus properties. can be easily installed after purchase.
To ensure compatibility now and in the future,
the Counter incorporates the latest IEEE-488.2
bus standard and the internationally standard-
ized SCPI Command set (Standard Commands
for Programmable Instruments). The bus trans-
Rubidium Counter
fer rate of the Counter is up to 1,000 measure- The PM6685R contains an atomic reso-
ments over the IEEE-488 bus, and 1,600 mea- nance-controlled timebase (rubidium) that
surements per second to internal memory. This gives a new meaning to 10-12 digit measure-
very high measurement rate makes new mea- ments.
surements possible. For example, you can per-
form jitter analysis on several thousands of The battery unit PM9623 is unnecessary in this
pulse width measurements and capture them in model, nor can it be built in.
a second. Together with the IEEE-488 inter-
faces you get an extensive programming man- The PM6685R uses a larger cabinet due to the
ual that helps you understand SCPI and pro- size and power requirements of the rubidium
gramming of the counter. timebase. The standard outfit also includes a
fan.
You get an analog recorder output as standard
with all IEEE-488 interfaces. This output pro-
vides an analog signal proportional to the value
of any three consecutive display digits. The
output can be used for recordings of measure-
ments on a strip-chart recorder or as a feedback
signal to an analog control system.
The counter is easy to use in IEEE-488 bus en-
vironments. A built-in bus-learn mode makes it
possible to transfer the complete counter setting
to the controller after having made all
individual settings manually via the front panel.
The response can later be used to reprogram the
counter to the same settings. This eliminates the
need for the occasional user to learn all individ-
ual programming codes. Complete (manually
set) counter settings can also be stored in 20 in-
ternal memory locations and can easily be re-
called for use at a suitable opportunity. Another
user-friendly feature is macro-programming.
You can define your own mnemonics and de-
Safety Instructions
Safety Instructions
Symbols
2-2 Introduction
Safety Instructions
Disposal of Hazardous
Materials
If your counter has a rechargeable battery op-
tion, that option uses lead-acid batteries of a
type similar to automotive starter batteries.
When the batteries no longer work properly,
dispose of them at a battery recycling station.
You can of course return these batteries to Fluke
for recycling as well.
n Fuse
Unpacking A 1.6 A/250 V slow blow fuse is placed inside
Check that the shipment is complete and that no the counter. This fuse rating is used for the full
damage has occurred during transportation. If voltage range.
the contents are incomplete or damaged, file a
claim with the carrier immediately. Also notify
your local Fluke sales or service organization in
Fig. 3-1 1.6AT 5x20mm fuse
case repair or replacement may be required.
Charging
Identification
You must Charge the battery before use or stor-
Check marks on the rear panel show what op- age. The counter charges the battery automati-
tions are installed in your counter, see Figure cally when connected to line power, whether
3-2. the instrument is in standby or turned on.
Grounding
Installation Grounding faults in the line voltage
supply will make any instrument con-
nected to it dangerous. Before connecting any
Supply Voltage unit to the power line, you must make sure that
the protective ground functions correctly. Only
n Setting then can a unit be connected to the power line
The Counter may be connected to any AC sup- and only by using a three-wire line cord. No
ply with a voltage rating of 90 to 265Vrms, 45 to other method of grounding is permitted. Exten-
440 Hz. The counter automatically adjusts itself sion cords must always have a protective
to the input line voltage. ground conductor.
3-2 Unpacking
Preparation for Use
90V-265V
G D E
Installation 3-3
Preparation for Use
If you have ordered a 19 inch rackmount kit for – Remove the two plastic lids that cover the
your instrument, it has to be assembled after de- screw holes on the right and left side of
livery of the instrument. The rackmount kit
consists of the following:
– 2 brackets, (short, left; long, right)
– 4 screws, M5 x 8
– 4 screws, M6 x 8
WARNING: When you remove the
cover you will expose live parts
and accessible terminals which
can cause death.
the front panel.
WARNING: Capacitors inside the in-
strument can hold their charge – Push the instrument back into the cover.
even if the instrument has been – Turn it upside down
separated from all voltage – Install the two rear feet with the screws
sources. (A) to the rear panel.
n Assembling the Rackmount Kit – Fasten the brackets at the left and right
side with the screws included as illustrated
Make sure the power cord is disconnected from below.
the instrument.
Turn the instrument upside down.
– Loosen the two screws (A) at the rear feet. – Fasten the PM6685 or the PM6685R in the
rack via screws in the four rack mounting
– Grip the front panel and gently push at the holes
rear.
– Pull the instrument out of the cover. n Reversing the Rackmount Kit
– Remove the four feet from the cover. The instrument may also be mounted to the
Use a screwdriver as shown in the following il- right in the rack. To do so, first remove the plate
lustration or a pair of pliers to remove the on the long bracket and fasten it on the short
springs holding each foot, then push out the one, then perform the preceding steps.
feet.
3-4 Installation
Chapter 4
– Timeout.
About This Chapter – Scaling factor for the analog output (when
This chapter gives you a quick introduction to GPIB option is installed).
all the controls of the counter, the design of the The counter has no numerical keypad, so you
user interface, and front panel text. For the oc- must use the following keys:
casional user, the information in this chapter is
often sufficient to solve a measurement prob- n Coarse Adjustment
lem. – Press DATA ENTRY p or
FUNCTION u to increase, and
DATA ENTRY q or t FUNCTION
to decrease a value in 1-2-5 steps.
The User Interface n Fine Adjustment
Default Settings
PARAMETER VALUE/
SETTING
Input A:
Sensitivity AUTO
Trigger level AUTO
Impedance 1MΩ
Trigger slope Pos
Filter OFF
Arming:
Start OFF
Start Delay OFF
Stop OFF
Miscellaneous:
Function FREQ A
Null/offset OFF
Time out OFF
Measuring time 0.2s
Check OFF
Single cycle OFF
Analog output control OFF
Auxiliary functions All switched OFF
Blank LSD OFF
Basic Controls
Lit when the counter Press ON and the counter turns on and re- Selects measuring function
is off, but power is turns to its default (standard) setting. If
available to an oven you want to recall the settings you used
oscillator. before you turned off the counter, press
AUX MENU, then press ENTER twice.
Press STAND-BY to turn off the counter.
LOCAL/
PRESET
The instrument
will ask De-
fault? If you AUX
press ENTER
the counter
will return to
preset (default) W
settings (see W
page 4-3). W
W
If in Remote
mode: the
INPUT A INPUT C
counter
switches to lo- This input is used for all This input is used for high
cal operation. measuring functions except frequency measurements.
frequency C and Ratio C/A. The frequency range is
It measures signals with fre- printed above the connec-
quencies between 10 Hz and tor. This is an option and if
Up and running in no time! 300 MHz and levels between no connector is installed
30mVpp and 70Vpp. you do not have this func-
– Turn on the counter by pressing
tion. The C input is fully
the ON key.
automatic and no controls
– Connect a signal to the input. affect its performance.
– Select function with the FUNC-
TION key.
Now the counter automatically mea-
sures with optimum input settings.
Input A Controls
WAVEFORM AUTO SENSITIVITY
Offsets the trigger level, This key switches on the auto sensi- Increases or decreases the
when the AUTO function tivity and auto waveform compensa- sensitivity of the counter
is switched off, for sig- tion. when the AUTO function is
nals with: AUTO selects 33% of Vpp as sensi- switched off. The set sensitiv-
very low duty cycle tivity for input A, and it adjusts to ity is shown on the bar graph
the waveform more accurately than display.
normal duty cycle the three choices available via man- If you turn off AUTO by
ual setting. pressing the SENS keys, the
very high duty cycle Both the waveform and the SENS selections made by AUTO re-
keys switches off AUTO. main as fixed settings.
AUX
W
W
W
Filter 50 W SLOPE
Switches on and off the Switches between 50 W/1 MW The trigger slope can be set in the
100 kHz low pass filter. input impedance. AUX MENU; see chapter 10,
This filter removes high 1 MW allows the counter to “Auxiliary functions”.
frequency interference measure without loading down This selection is intended for neg-
when measuring on LF the measuring signal while ative pulse width and negative
sine wave signals. 50 W terminates the cables in duty factor measurements.
50 W systems, minimizing re- The negative slope of the wave-
flections and interference. form indicator flashes when the
counter triggers on negative
slopes.
EXT REF
Switches be-
tween the ex- AUX
W
W
W
Display Controls
Freezes the display until The NULL function stores the Each press on this key blanks
you press the key again. current result on the display, then out one display digit, starting
New measurements are shows all the following results as with the LSD.
armed every time MEAS deviation from that result. When all digits are blanked,
RESTART is pressed. You can read and change the the next press disables the
Starts/stops the measure- stored reference in the AUX blanking.
ment in TOT A MAN. MENU.
AUX
W
W
W
You can switch on and off The measurement result can con- If you press MENU, the
the display backlight in tain up to 12 digits. With the over- display shows all selectable
the AUX MENU. flow function you can see the two functions and the current
additional digits that normally are selections are blinking.
hidden.
Display
MEASURING NUMERICAL ENTER
FUNCTIONS PRESENTATION
Displayed when the instrument wants
The current mea- A 10-digit display used for you to confirm a selection by pressing
suring function is showing measuring results ENTER.
shown on the dis- and other values.
play.
The display always shows
If MENU is basic units (Hertz, counts or
pressed, all possible seconds) plus an exponent
selections are when necessary.
shown on the dis-
play and the current
setting is blinking.
Mantissa Exponent Unit Indicator
NULL
S OVERFLOW SINGLE
ON if the re- MEMORY BURST
PRF
sult is dis- ENTER AUX
ARM ARM
FREQ A FREQ C PER A RATIO A/E RATIO C/A P WIDTH A TOT A MAN DUTY F A STA STO
played relative
LEVEL A SENSITIVITY A
to a nulling REMOTE EXT REF
SRQ LO BAT
CHECK FILTER
OVERLOAD HOLD
150M W
NULL OFFSET
AUTO
TRIG
constant.
4-8 Display
Using the Controls
Display
S OVERFLOW SINGLE
MEMORY BURST
ENTER AUX PRF
ARM ARM
FREQ A FREQ C PER A RATIO A/E RATIO C/A P WIDTH A TOT A MAN DUTY F A STA STO
Display 4-9
Using the Controls
Rear Panel
GPIB OPTION
ANALOG OUT- GPIB ADDRESS GPIB CONNECTOR
PUT SWITCH
A standard IEEE 488.1
Outputs any 3 Selects address between 1 connector for connection
consecutive digits and 30. The display shows to a controller.
on the display as 0 the address every time the
to 4.98 V. counter is turned on.
90V-265V
G D E
A frequency reference out- If you have an in-house reference or want to run several instru-
put to be used by other in- ments on the same reference, connect 10 MHz here and select
strument. it by pressing the EXT REF key on the front.
Rear Panel
90V-265V
POWER SOCKET
88/IEC 625 INTERFACE
H1, T5, L4, SR1, The input can handle 90 to
C1, DT1, E2
628/85
—ON
PRIMARY FUSE
1.6AT
INSIDE 265 VAC and 45 to 440 Hz
without any range switching.
—OFF/EXT DC
697
12-24V DC INTERNAL
____
BATTERY
____
REF IN EXT ARM Made i n S weden
Just connect and go!
D E
F U N C T IO N M E A S U R IN G
T IM E
U N L O C K E D
H z
E N T E R A U X A U X
F R E Q A F R E Q C R A T IO A /E R A T IO C /A P W ID T H A T O T A M A N D U T Y F A M E N U M E N U S IN G L E
E X T R E F C H E C K
G A T E
F R E Q U E N C Y
L O C A L E X T
IN P U T A 1 0 m V 1 0 0 m V 1 V 1 0 V r m s A A D J U S T
P R E S E T R E F C H E C K F IL T E R 5 0 9 A U T O S E N S D C - 3 0 0 M H z C 0 .1 - 2 .7 G H z
M A X
M E A S U R E D A T A E N T R Y 1 2 V r m s -5 0 9
S T A N D - B Y O N R E S T A R T H O L D N U L L B L A N K E N T E R 3 5 0 V p -1 M 9
5 0 9
M A X 1 2 V rm s
F R E Q U E N C Y
A D J U S T
Input Signal
Conditioning
Input Signal Conditioning
IM P A U T O M A T IC F IX E D 1 0 0 k H z L P M IC R O -
A T T E N T IO N A C C O U P L F IL T E R S E N S C O N T R O L L E R
S L O P E ,
A A R M IN G ,
S Y N C ,
D E L A Y
E T C . N U L L ,
D IS P ,
H O L D
E T C
S E N S IT IV IT Y
W A V E F O R M
C O M P E N S A T IO N
D A C
M a n u A l
S E N S
O F F O N
A U T O
+
Auto sets the waveform compen-
sation much more accurately
than is possible manually. So it is
likely that the auto trigger can
handle signals that are impossi-
AUX
ble to trigger on manually.
W
W
+
auto function on. Switching Turn off AUTO:
- If the input frequency is <50 Hz
on the power always acti-
-If you measure AM signals
vates AUTO. -If you measure single shot phe-
The auto triggering function of the counter con- nomena
trols both the sensitivity (also called trigger
window or hysteresis) and the waveform com- Speed
pensation (also called trigger level offset, or The counter measures amplitude and calculates
duty factor compensation). Auto measures the sensitivity and trigger level offset rapidly. The
peak-to-peak levels of the input signal and sets typical time is 50 ms. If you use the counter in
the upper level of the hysteresis band to 66% an automatic test system and need faster mea-
and the lower level to 33% of that value (for surements, read about speed in chapter 19
pulse width and duty factor measurements, “How to Measure” in the GPIB Programming
both levels are set to 50%). AUTO accurately Manual.
sets the waveform compensation to compensate
for duty factors other than 50%. The waveform
H y s te r e s is b a n d ( S E N S )
T r ig g e r le v e l o ffs e t
T r ig g e r p o in ts
t
0 V
R e s e t p o in ts
Fig. 5-2 Parameters controlled by SENS and Waveform keys.
Input A 5-3
Input Signal Conditioning
+
to the last manual settings, and you can control It is much easier to set the cor-
the input amplifier by pressing the SENS and rect sensitivity if you know the
the WAVEFORM keys. signal level. Read the level on
the bar graph before switching
When to use manual sensitivity off AUTO.
– The most obvious use for manual settings
is for signals below 50 Hz, for which
AUTO does not function.
You can use the WAVE-
– When measuring on non-repetitive signals, FORM key to compensate
you also need to use manual trigger levels; a waveform in three steps:
see Chapter 6. (0 – 25% duty factor)
– When measuring AM signals. (25–75% duty factor)
– When measuring sine wave signals with (75–100% duty factor)
little noise, you may want to measure with The reason for this setting is that
a high sensitivity (narrow hysteresis band) non-symmetrical signals do not center on 0 V,
to reduce the trigger uncertainty. Trig- but get a dc shift after the ac coupling capacitor.
gering at or close to the middle of the sig- Signals with less than 50 % duty factor have a
The grayed fields show the position of the hysteresis band, and the circles are
trigger points and reset points
ac coupling capacitor
5-4 Input A
Input Signal Conditioning
Input A 5-5
Input Signal Conditioning
5-6 Input C
Input Signal Conditioning
n Stable readings
Measuring Functions
Measuring Functions
Frequency Measurements
– Frequency
AUX
W
W
– AM
W
Frequency
0 V
R e s e t p o in ts
A c tu a l G a te tg
T im e
Introduction 6-3
Measuring Functions
After the start of the set measuring time, the counter will show NO trig when triggering
counter synchronizes the beginning of the ac- has stopped.
tual gate time with the first trigger event (t1) of
the input signal it measures on. Signal detection is ON for all functions but
TOT A MAN, burst frequency, PRF, arming
In the same way, the counter synchronizes the and single measurements.
stop of the actual gate time with the input sig-
nal, after the set measuring time has elapsed. n Sample-Hold
The multi-register counting technique in the If the input signal disappears during the mea-
counter allows you to simultaneously measure surement, the counter will show NO SIGNAL.
the actual gate time (tg) and the number of cy- If you probe test points and you cannot simulta-
cles (n) that occurred during this gate time. neously view the display this will be a problem.
Thereafter, it calculates the frequency accord- In that case, press DISPL HOLD to activate
ing to Mr. Hertz’s definition: HOLD, put your probe tip on the test point and
n press MEAS RESTART. Now the counter
f= makes one measurement, and when you re-
tg
move the probe tip from the test point, the coun-
ter continues to display the measured value and
The counter measures the gate time, tg, with a
behaves like a voltmeter with a sample-hold
resolution of only 250 ps. independent of the
feature.
measured frequency. Consequently the use of
prescalers does not influence the quantization
n Time-Out
error. Therefore, the relative quantization error
is: 250 ps/tg. Mainly for GPIB use, you can manually select a
fixed time-out in the AUX MENU. The range
250 ps of the fixed time-out is 100 ms to 25.5s and the
Relative RMS quantization error =
tg default setting is OFF.
For a 1-second measuring time, this value is:
Select a time-out that is longer than the cycle
250 ps time of the lowest frequency you are going to
= 250* 10-12 = 2.5* 10-10
1s measure; multiply the time-out by the
prescaling factor of the input channel. When no
Except for very low frequencies, tg is nearly triggering has occurred during the time-out , the
identical with the set measuring time. counter will show Ti.Out.
6-4 Introduction
Measuring Functions
1 250 ps
Speed » + 0.001 still: .
tg tg
with AUTO turned OFF. The prescaling factors are:
Function Prescaling factor
n Frequency Average and Single
Cycle Measurements BURST A 1 (80 MHz)
FREQ A average* 4 (300 MHz)
To reduce the actual gate time or measuring ap-
*
erture, the counter has very short measuring PER A average 4 (300 MHz)
times and a measuring time called SINGLE. FREQ C (PM9624) 64 (3.0 GHz)
The latter means that the counter can measure FREQ C (PM9621) 512 (1.3 GHz)
during only one cycle of the input signal. In ap-
RATIO A/E 4/1 (300/80 MHz)
plications where the counter uses an input chan-
nel with a prescaler, this frequency divider RATIO C/A (PM9624) 64/4 (3.0/0.3 GHz)
causes the SINGLE measurement to last as RATIO C/A (PM9621) 512/4 (1.3/0.3 GHz)
many cycles as the division factor. TOTALIZE A 1 (100 MHz)
Introduction 6-5
Measuring Functions
measure a 0.1 Hz pulse with a non-prescaled etition frequency (PRF), that switches the CW
function like PERIOD SINGLE, the measure- signal on and off.
ment will require at least the duration of one cy-
cle, that is 10 seconds, and worst case nearly 20 With PM 6685’s built-in measurement control
seconds. The worst case is when a trigger event facilities, you can measure burst signals with-
takes place just before the beginning of a mea- out the external arming signals that traditional
suring time (Fig. 6-4). counters need. You can also make measure-
n HF Signals C W
+
When measuring Burst A or PRF
mance when measuring stable continuous RF. A, the maximum burst frequency
The prescalers have a nasty habit, and that is is 80 MHz and the minimum
that they are not stable and would like to number of cycles in a burst is
self-oscillate when there is no input signal pres- three.
ent. To prevent a prescaler from oscillating, the
prescaler incorporates a “go-detector.” The
go-detector continuously measures the level of n Triggering
the input signal and simply blocks the prescaler Bursts with a PRF above 50 Hz can be mea-
output when no signal, or a signal that is too sured with auto triggering on.
weak is present.
The out of sync error described under heading
N
&
T o c o u n tin g “Possible errors” on page 6- 8 may occur more
lo g ic
frequently when using autotrigger.
G o -d e te c to r
When PRF is below 50 Hz and when the gap
Fig. 6 -5 Go detector in the prescaler. between the bursts is very small, use manual
triggering.
B u rs t
S ig n a l
+ Internal burst synchronization
also works with the prescalers
PM 9621 and PM 9624, but with
S y n c . d e la y limited specifications.
Fig. 6-7 Set the sync delay so that it ex- – Select a measuring time that is shorter
pires in the gap between the than the burst duration minus two burst
bursts. frequency cycles or pulses.
– Press AUX MENU, select BURST, and
press ENTER.
n How Does PRF Work?
– Select channel A or C as measurement in-
The PRF is the number of bursts per second. put, and press ENTER.
This means that the counter must count one – Set a sync delay longer than the burst du-
pulse in each burst. ration and shorter than the burst repetition
period. See Figure 6 -7.
M e a s u re
– Press ENTER to measure.
surements on short bursts means using short reliable, at least two measurements must be
measuring times, giving a poorer resolution made, the first one for synchronizing the mea-
than normally achieved with the counter. surement and the second one for reading out the
measuring result.
n How Does the Sync Delay
Work? Autosync on Slow-Starting
The sync delay works as an internal start arm- Bursts
ing delay: it prevents the start of a new mea-
surement until the set sync delay has expired. Bursts may start a bit slowly, especially RF
See Figure 6-9. bursts. The result when measuring the burst fre-
quency will then be erroneous unless the first
few pulses are excluded from the measurement.
M e a s u re
A delay can be used that starts on the first nega-
tive slope of the burst, and expires when the set
arm start delay ends.
S y n c - D e la y n Preparations
– Switch off the AUTO function.
G a te T im e
– Check that the waveform compensation
Fig. 6 -9 Measuring the frequency of the shows .
carrier wave signal in a burst.
– Set a suitable sensitivity.
After the set measuring time has started, the n Switching on Burst
counter synchronizes the start of the measure- Measurements
ment with the second trigger event in the burst.
– Press AUX MENU.
This means that the measurement does not start
erroneously during the Burst Off duration or in- – Select BURST.
side the burst. – Press ENTER.
n Possible Errors
– Select A or C.
– Press ENTER when the display shows
Before the measurement has been synchronized
SYNC DELAY.
with the burst signal, the first measurement(s)
could start accidentally during the presence of a – Enter a synchronization delay that is lon-
burst. If this would happen and if the remaining ger than the burst duration but shorter than
burst duration is shorter than the set measuring the burst repetition time.
time, the readout of the first measurement will – Press ENTER.
be wrong. However, after this first measure-
– Press MEAS TIME and select 800 ns.
ment, a properly set start-arming sync delay
time will synchronize the next measurements. – Increase the measuring time until you get
the number of digits you want. Take care
In manually operated applications, this is not a not to increase it so much that the counter
problem. In automated test systems where the counts incorrectly
result of a single measurement sample must be
+
As long as burst is ON, the arm- The counter automatically disables the
ing delay is triggered by the burst go-detector when both start arming and
itself, not by the arming signal on FREQ C are active; thus, there is no special
input E. function to enable.
– Connect external SYNC to Input E. – Press SENS to select the highest sensi-
tivity and then decrease the sensitivity un-
– Use the FUNCTION key to select til the measurement result is no longer sta-
FREQ C. ble.
– Set a measuring time shorter than the burst – Increase the sensitivity a couple of dB’s
duration. and measure.
– Press the AUX MENU key, select ARM
START, and positive slope (ARM STA+).
A
– Press ENTER.
– Enter a delay time and press ENTER.
– Measure.
6-10 AM Signals
Measuring Functions
Modulating Frequency
The easiest way to measure the modulating fre-
quency is after demodulation. If no suitable de-
modulator is available, use the PRF function to
measure the modulation frequency in the same
way as when measuring Burst PRF.
– Press MEAS TIME and enter a measuring
time that gives you the resolution you
want.
– Switch off SINGLE.
– Press AUX MENU, select PRF, and press
ENTER.
– Select INPUT A as measurement input,
and press ENTER.
– Set a sync delay of approximately 75% of
the modulating period. See Figure 6-7.
– Press ENTER to leave the AUX MENU.
– Switch off AUTO.
– Press the WAVEFORM key to select sym-
metrical signal.
– Press SENS until the counter stops trig-
gering.
– Increase the sensitivity a couple of dB’s
and measure.
S y n c . d e la y
M e a s u r e d c y c le s
AM Signals 6-11
Measuring Functions
Ratio
Introduction
To find the ratio between two input frequencies,
the counter counts the cycles on two channels
simultaneously and divides the result on the pri-
mary channel by the result on the secondary
channel.
Ratio can be measured between A and E inputs
or between C and A inputs.
+
The input frequency range of in-
put E is limited to 80 MHz.
6-12 Introduction
Measuring Functions
Period
Introduction
From a measuring point of view, the period
function is identical to the frequency function.
This is because the period of a cyclic signal has
the reciprocal value of the frequency ( 1 f ).
number of cycles
f=
actual gate time
Introduction 6-13
Measuring Functions
Introduction Triggering
If AUTO is on when these functions are se-
lected, the counter adjusts the trigger level to
H y s te r e s is b a n d ( S E N S )
50 % and the sensitivity to maximum .
T r ig g e r le v e l o ffs e t
T r ig g e r p o in ts The set trigger level and trigger slope define the
t
start and stop triggering. The manual trigger
function of the counter allows only three trigger
0 V level settings. This limited selection makes it
R e s e t p o in ts possible to trigger correctly only on signals
with very steep slopes.
+
Always use AUTO or AUTO
ONCE when measuring Pulse
Fig. 6-13 Time is measured be- Width and Duty Factor.
tween the trigger point
and the reset point. Ac- n Hysteresis
curate measurements
are possible only if the In pulse width measurements the trigger hyster-
hysteresis band is nar- esis, among other things, causes measuring er-
row and centered rors. Actual triggering does not occur when the
around 50 % of the am- input signal crosses the trigger level at 50 per-
cent of the amplitude, but when the input signal
has crossed the entire hysteresis band. At maxi-
This counter can measure PULSE WIDTH and
mum sensitivity, the hysteresis band is as small
DUTY FACTOR.
as possible, about 10 mV.
6-14 Introduction
Measuring Functions
Pulse Width A
The counter measures pulse width on Input A.
Duty Factor
A duty factor (or duty cycle) measurement con-
sists of two measurements: one pulse width
measurement and one period measurement.
The duty factor is then calculated as:
Pulsewidth
Duty factor =
Period
Totalize
Totalize A Manual
This mode enables you to totalize (count) the
number of trigger events on channel A. Start
and stop of the totalizing is manually con-
trolled.
The counting capability is 1* 1014 events at rates
to 100 MHz.
FUNCTION Select TOT A MAN with the
FUNCTION key.
+
AUTO is switched off during
TOTALIZE. You must always set
manual trigger level.
Measurement Control
Measurement Control
+
If you select SINGLE, the set
start and stop of measurements and what you
Measuring Time becomes the
can obtain by doing that. The chapter starts by Display Time (time between
explaining the keys and the functions behind measurements).
them, then gives some theory, and ends with ac-
tual measurement examples. n Fine-Tuning the Measuring
Time
For times above 50 ms, you can set your own
measuring time as follows:
Measuring Time – Press MEAS TIME.
– Press SENS u and the parameter to be
set expands over the entire display.
– A cursor flashes to the left of the MSD
digit. Move this cursor to the digit you
want to change. Use the t SENS u keys.
– Change the value of the selected digit by
pressing the DATA ENTRY keys p and
q.
– Move the cursor to the next digit and re-
AUX
peat the procedure above until the display
shows the desired value. Then press
ENTER to confirm the selection.
W
W
W
W
Increasing the measuring time gives more dig- Use SINGLE when you want to measure on
its, but fewer measurements per second. single-shot phenomena or when you just want
fast results without the need for many digits.
To change the measuring time:
– Press the MEAS TIME key. The number of input periods in a SINGLE mea-
surement depends on the prescaler factor of the
– Increase/decrease the value by pressing the input and which function is selected as follows:
FUNCTION key.
– Confirm your selection by pressing
MEAS TIME again.
W
W
W
á Gate Indicator
The GATE LED is on when the counter is busy AUX External arming gives you
counting input cycles. MENU the opportunity to start and
stop a measurement when
an external qualifier event
occurs.
Display Hold Start and stop of the arming function can be set
independently to positive slope, negative slope,
DISPLAY Pressing DISPLAY HOLD or OFF.
HOLD freezes the result on the
display. The display is not Input E on the rear panel is the arming input.
frozen until one measure-
Arming is somewhat complicated, so study the
ment has been completed
examples later in this chapter to see what you
after DISPLAY HOLD has
can obtain by using it. There is normally no
been pressed.
need for arming except when measuring on
MEAS MEAS RESTART initiates a complex signals (non-continuous wave).
RESTART new measurement.
+
If you use start arming to arm an
average measurement, it only
controls the start of the first sam-
ple.
Stop Arming
Stop arming prevents the stop of a measure-
ment until the counter detects a level shift on
the E input. Combining Start and Stop Arming
7-4 Arming
Measurement Control
AUX
W
W
W
Controlling Measurement
Timing
The Measurement Process – The set measuring time has expired (in
frequency measurements, for example).
Basic Free-Running Measurements
– In SINGLE, the measurement stops imme-
Since the counter uses the reciprocal counting
diately when the input signal fulfils the
technique, it always synchronizes the start and
stop trigger conditions (which is normally
stop of the actual measuring period to the input
when it passes the trigger window the sec-
signal trigger events. In free-running mode a
ond time).
new measurement automatically starts when
the previous measurement is finished. This is
ideal for continuous wave signals. Measuring Time and
The start of a measurement takes place when
Measurement Rates
the following conditions have been met (in or- The set measuring time decides the length of a
der): measurement in all average types of measure-
ments. In a single-shot type of measurement,
– The counter has fully processed the previ-
however, the measuring time instead acts as a
ous measurement.
“display time” setting. For example, if a mea-
– If the counter makes SINGLE measure- suring time of 500 ms is set in a single period
ments, the display time (=set measuring measurement, and the period is 100 ns, the dis-
time) must have expired. play will show the result for 500 ms before the
– All preparations for a new measurement next measurement can start.
are made.
This is important to know when you want to
– The input signal triggers the counter’s make fast measurements, for example, when
measuring input. using the GPIB bus.
The measurement ends when the input signal
meets the stop trigger conditions. That happens
directly after one of the following events:
+
To get maximum measuring Resolution as Function of
speed, you should use the com-
bined benefits of SINGLE and Measuring Time
minimum measuring time. The quantization error and the number of digits
on the display mainly define the resolution of
The time between the stop of one measurement the counter, that is the least significant digit dis-
and the start of the next one can be below 1 ms played.
in free-running mode if you do the following:
– Do not use AUTO. As explained on page 6-4 under Reciprocal
Counting, the calculated frequency f is:
– Do not use NULL.
n
– Switch off the display via GPIB. f=
tg
Additional Control of Measurement
Start / Stop while the relative rms quantization
Free-running measurements may be easy to un- error = ±250 ps/tg .
derstand, but measurements can get more com- The counter calculates the mantissa of f with up
plex. to 15 digits. However, the number of justified
Besides input signal triggering, the start of a digits depends on the selected measuring time
measurement is further controlled by the fol- and the measured frequency, and is much more
lowing elements: limited.
– Manual MEAS RESTART, if DISPLAY The counter truncates irrelevant digits so that
HOLD is selected. the rms quantization resolution cannot change
the LSD (least-significant digit) more than
– GPIB triggering (<GET> or * TRG), if bus
± 2.5 units. This is when the displayed value is
triggering is selected.
99999999, and the quantization error is worst
– External arming signal, if Start Arming is case.
selected.
– Expired start arming delay, if Arming De-
lay is selected.
+ ±1 unit in 99999999 (=1E8)
means 10 times more relative
resolution than ±1 unit in
In addition to expired measuring time and stop 10000000 (=1E7), despite the
same number of digits.
signal triggering, the stop of measurement is
further controlled by: In practice, the quantization error is two to three
– External arming signal triggering, if Stop times better than the specified value, and the
Arming is selected. measured value can range from 10000000 to
GPIB triggering is described in the Program- 99999999. Therefore, in practice the
mer’s manual. Now let’s look deeper into the quantization uncertainty shown as instability in
concept of arming. the LSD can range from 0.25 to 2.5 LSD units.
R e la tiv e
Q u a n tiz a tio n Q U A N T IZ A T IO N R E S O L U T IO N
E rro r
-3
1 0
-4
1 0
-5
1 0
-6
1 0
-7
1 0
-8
1 0
-9
1 0
-1 0
1 0
-1 1
1 0
-7 -6 -5 -4 -3 -2 -1 0 + 1 + 2
1 0 1 0 1 0 1 0 1 0 1 0 1 0 1 0 1 0 1 0
M e a s u r in g T im e in s
If you use start arming and stop arming together n When Do I Use Stop Arming?
you get an externally controlled measuring
You normally use stop arming together with
time.
start arming. That means that the external
gating signal controls both the start and the stop
n Manual Arming
of the measurement. Such a gating signal can be
The counter has a manual arming function used to force the counter to measure the fre-
called DISPLAY HOLD. Here you manually quency of a pulsed RF signal. Here the position
arm individual measurements one-by-one by of the external gate must be inside a burst. See
pressing the RESTART key. Figure 7-4.
Use this manual arming mode to measure sin-
gle-shot phenomena, which are either triggered S ta rt A rm S to p A rm
E x t G a te
manually or occur at long intervals. Another
reason for using this manual arming could sim-
ply be to allow sufficient time to write down in- B u rs t
dividual results. S ig n a l
S IG N A L n o y e s
A U T O O N ? S H O W "N O S IG N A L "
D E T E C T E D ?
o u i n o
S H O W "N O T R IG "
D IS P L A Y y e s P R E S S
H O L D R E S T A R T
S T A R T y e s W A IT F O R y e s
D E L A Y ?
A R M IN G E X T . S IG N A L
n o
n o
W A IT F O R IN P U T W A IT P R E S E T T IM E
S IG N A L T O T R IG G E R
S T A R T O F
M E A S U R E M E N T
S T O P y e s W A IT F O R
A R M IN G ? E X T . S IG N A L
n o
E N D O F P R E S E T
M E A S U R IN G T IM E
W A IT F O R IN P U T
S IG N A L T O T R IG G E R
S T O P
M E A S U R E M E N T
P R O C E S S R E S U L T
& D IS P L A Y
Arming Examples
Example #1: The conditions for success are that the pulse
burst does not repeat itself more than 50 to 150
times per second. The duration of a pulse burst
Measuring Pulse Width in a (between first and last pulse) must be substan-
Pulse Burst tially less than the distance to the next burst.
In the first example we will measure the width
of pulse #1 in a repetitive pulse burst. In this ex- Do the following steps to perform auto syn-
ample, a synchronization signal (SYNC) with chronization without arming:
TTL levels is also available. See Figure 7-8. – Connect the burst signal to input A.
– Adjust the manual sensitivity and trigger
S y n c E
level until the burst signal triggers the
counter correctly.
– Use the FUNCTION key to select
In p u t A P WIDTH A.
ing time that almost equals the duration of a – Connect the burst signal to input A.
burst, the auto-synchronization will work.
– Adjust the manual sensitivity and trigger
level until the burst signal triggers the
counter correctly.
N o t U s e d
– Press AUX MENU, select ARM START,
and select arming on positive slope (ARM
STA+).
A
In p u t A
B C
– Use FUNCTION key to select PULSE
WIDTH.
D e la y
> 5 n s
In p u t A
In p u t A
Fig. 7 -9 Synchronization using start arm-
ing. Fig. 7-10 Synchronization using start
arming with time delay.
Example #2 A r m in g in p u t
Measuring Frequency in
Two-Tone Bursts
In p u t A o r C
Sonar bursts can consist of two different fre- s ig n a l
quencies with different durations. See Figure
7-12
S Y N C
M e a s u r in g tim e
In p u t A o r C
s ig n a l
M e a s u r in g tim e
7-16 Example #2
Measurement Control
Example #3 7-17
Measurement Control
How to Do a VCO Step Response When all 100 measurements have been made,
Profiling the results can be used to plot frequency versus
time. Note that the absolute accuracy of the
– with 100 samples during a time of
time scale is dependent on the input signal it-
10 ms, i.e., 100 ms between samples.
self. Although the measurements are armed at
This measurement scenario requires a repetitive 100 ms±100 ns intervals, the actual start of mea-
input step signal, and you have to repeat your surement is always synchronized to the first in-
measurement 100 times, taking one sample per put signal trigger event after arming.
switch period. Each new sample should be de-
layed 100 ms with respect to the previous one.
V o lta g e s te p
g e n e ra to r V C O
In p u t E , E X T A R M
In p u t A
7-18 Example #3
Chapter 8
Processing
Processing
Introduction Nulling
Two different ways to process a measurement
result are available: Averaging and Nulling.
Averaging
AUX
W
W
W
+
W
W
8-2 Introduction
Processing
Nulling 8-3
Processing
8-4 Nulling
Chapter 9
Auxiliary Functions
Auxiliary Functions
A U X
M E N U E N T E R
F U N C T IO N
D IS P L A Y B A C K L IG H T R E C A L L M E M O R Y
O N / O F F
N U L L S A V E M E M O R Y
R e a d /C h a n g e
D IS P L A Y O V E R F L O W P R O T E C T M E M O R Y
O N / O F F
A N A L O G O U T P U T T R IG G E R S L O P E
O N /O F F + C o n tr o l S c a lin g P o s it iv e o r n e g a tiv e
G P IB -A D D R E S S A R M IN G S T A R T A R M IN G D E L A Y
R e a d O N /O F F /D E L A Y T im e 2 0 0 n s to 1 .6 7 s
T IM E O U T A R M IN G S T O P
s e t T im e O u t, 1 0 0 m s to O N /O F F
2 5 .5 s a n d O F F
T E S T B U R S T C W
S e le c t S e lf T e s t A , C a n d O F F
S Y N C D E L A Y
T im e 2 0 0 n s to 1 .6 7 s
P R O G R A M ID E N T IF Y P R F
C o u n te r a n d G P IB p ro g ra m s A , C a n d O F F
Fig. 9-1 You will enter the AUX MENU at the same selection as you used the previous
time, except after power on. Then you enter the menu at RECALL.
+
FUNCTION key until the display shows The time-out is mainly used for
TEST. GPIB applications.
– Enter the test menu by pressing the
ENTER key. GPIB Address
Selections for internal self-tests are as follows: Shows and changes the
– TEST ALL (the four tests below in se- GPIB address. The new
quence) address is stored in
– TEST DISP (Display Test) non-volatile memory
and remains until
– TEST Logic (Measuring Logic) changed again via this
– TEST RA (RAM) menu, the address
– TEST RO (ROM) switches on the rear panel or via a bus com-
If any fault is detected, an error message will mand.
appear on the display and the program halts. This means that the address of the counter can
Possible error messages are as follows: differ from the address indicated by the
– Internal ROM test failed. switches on the rear panel.
– Internal RAM test failed. The hex-address
+
The last set address is the valid
where an error is detected is shown. address whether it is set via the
aux menu, the address switches
– Test of measuring logic failed. or a GPIB command.
+
The display test turns on all seg-
ments of the display for a visual The counter shows the used address during the
inspection. No failure is reported. power-up test.
Press ENTER to end the test.
Analog Output (GPIB option want the analog output to assume its maximum
voltage (5 V).
only)
The default mode of the analog output is OFF. Example:
You turn it ON/OFF and set the scaling factor – Take a measurement result, for instance:
under ANALOG OUT in the AUX MENU. 12.34567890E+6 Hz
– Write the value without exponent:
12345678.90 Hz
– Multiply this value by the scaling factor,
for instance 0.001.
12345.67890
– Take the fractional part of the result:
.67890
– This is the value that will determine the
output voltage, .00 will give 0 V and .99
will give 5 V. this means that “our” read-
ing will give .67890*5=3.3945 V.
This is output as 3.38 V due to the 0.02 V
resolution of the analog output.
n Scaling Factor
The scaling factor has two functions: Fig. 9-3 To use the shown decimal
point as reference; set the
– The exponent selects the display digits to exponent of the scaling fac-
be represented by the analog output. tor to the same value as the
– The significant figures set the reading to exponent of the measure-
ment result but with opposite
represent full scale.
sign.
The default scaling factor is 1 (1E0). This
means that the full scale value is 0.999, and the n Resolution
analog output converts the fraction (digits to the The analog output range is 0 to 5 V in 250 steps,
right of the decimal point) to a voltage. so one step is 0.02 V. If the scaling factor is 1,
one such step is taken each time the display
The scaling factor should be:
changes with X.004, and if the scaling factor is
1
Scaling factor = 4, one step is taken each time the display
full scale value changes with X.001.
where full scale value is the value for which you
The X in the above paragraph can be any digit Using OVERFLOW with Display Hold
and does not influence the output voltage. If the
When DISPLAY HOLD is active, The ENTER
display changes from 0.996 to 1.000, the volt-
key toggles OVERFLOW on/off. If the counter
age drops from 4.98 V to 0V. If the display
has measured a result with more than 10 digits
value increases further, the output voltage starts
and you press ENTER, the OVERFLOW an-
to increase again, see Figure 9-4.
nunciator will switch on and the display will
O u tp u t v o lta g e
show the additional digits. The counter is de-
S c a lin g fa c to r 1
4 .9 8 V signed this way because switching on OVER-
0 .0 0 4 FLOW in the AUX MENU will initiate a new
measurement. This is undesirable if you want to
2 0 m V
study single shot phenomena or long-term
totalizing.
0 .0 0 V
0 .0 0 0 1 .0 0 0 2 .0 0 0 D is p la y e d 14
0 .9 9 6 1 .9 9 6 v a lu e When totalizing, the result can be up to 10 ,
that is 14 digits, but overflow will not show
O u tp u t v o lta g e S c a lin g fa c to r 4
4 .9 8 V
more than 12 digits.
0 .0 0 1
2 0 m V
Null
The null function stores the result currently dis-
0 .0 0 V played, then shows all the following results as
D is p la y e d
deviations from that result.
0
00
v a lu e
0
0
0.
25
50
0.
0.
49
99
4
0.
0.
Fig. 9-4 Output voltage versus dis- ing the NULL key on the front panel. You can
played value for two different read and change the stored result in the AUX
scaling factors. MENU.
When NULL is ON:
Display Overflow – Press AUX MENU
Display overflow makes 12-digit measure- – Select NULL with the FUNCTION key.
ments possible . When OVERFLOW is on and – Press ENTER and the display will show
the measurement justifies 11 or 12 digits, the the stored value.
OVERFLOW annunciator turns on, and the
Change the value in 1-2-5 steps with the DATA
counter truncates one or two MSDs and shows
ENTRY keys, or select what digit you want to
one or two extra LSDs instead.
change with the SENS keys and change it with
You have to keep track of the decimal point the DATA ENTRY keys.
yourself, compare the values with overflow ON – When the display shows the desired value,
and OFF to see if one or two overflow digits are press ENTER.
shown when the OVERFLOW annunciator is
As default the null value is negative and sub-
ON.
tracted from the new readings, but you can also
enter a positive null value.
Display Light
You can turn the display backlight ON or OFF:
– Press AUX MENU.
– Select DISP LIGHT with the FUNCTION
key.
– Press ENTER.
– Select ON or OFF with the DATA EN-
TRY keys
– Press ENTER again to exit the aux menu.
Performance Check
Performance Check
General Recommended
Information Test Equipment
WARNING: Before turning on the in- Type of in- Required Suggested
strument Specifications Instrument
strument, ensure that it has been
LF Synthe- Square; Sine
installed in accordance with the in-
sizer up to 10 MHz
stallation instructions outlined in
Power 50 W PM9584/02
Chapter 3 of this manual. Splitter
T-piece
This performance procedure is intended to do
Termination 50 W PM9585
the following:
Reference 10 MHz Fluke 908,
– Check the instrument’s specification. oscillator ±0.1 Hz for or a counter
– Be used for incoming inspection to deter- standard oscil- with cali-
lator brated
mine the acceptability of newly purchased
PM9691
instruments and recently recalibrated in-
10 MHz Fluke 909,
struments. ±0.01 Hz for 910R,
– Check the necessity of recalibration after PM9691 and PM6685R or
the specified recalibration intervals. PM9692. PM6681R
+
The procedure does not check 10 MHz Fluke 910R
every facet of the instrument’s ±0.0001 Hz or Cesium
calibration; rather, it is concerned for PM6685R. Standard
primarily with those parts of the HF signal 500 MHz (no
instrument which are essential generator prescaler
for determining the function of option)
the instrument. 1.5 GHz (op-
tion PM9621)
It is not necessary to remove the cover of the in- 3.3 GHz (op-
strument to perform this procedure. tion PM9624)
Pulse Gen- 125 MHz
If the test is started less than 30 minutes after erator
turning on the instrument, results may be out of Oscillo- 350 MHz
specification, due to insufficient warm-up time. scope
S OVERFLOW SINGLE
MEMORY BURST
ENTER AUX PRF
ARM ARM
FREQ A FREQ C PER A RATIO A/E RATIO C/A P WIDTH A TOT A MAN DUTY F A STA STO
ENTRYq
The keyboard test verifies that the counter re- ENTER NO SIGNAL
sponds when you press any key. To check the MEASURE HOLD
function behind the keys, see the tests further HOLD
on in this chapter. MEASURE ----------
RESTART
Press the keys as described in the left column MEASURE NO SIGNAL
and look on the display for the text, as described HOLD
in the second column. Some keys change more SINGLE SINGLE ----------
text on the display than described here. The dis- tFUNCTION DUTY F A
play text mentioned here is the text mostly asso- tFUNCTION TOT A MAN 0
ciated with the selected key. FUNCTIONu DUTY F A ----------
FUNCTIONu FREQ A
For the instrument to respond correctly, this test AUX MENU RECALL
must be carried out in sequence, and you must PRESET DEFAULT? Default
start with the preset (power-up) setting. ENTER NO SIGNAL setting
NULL NULL
KEY(S) DISPLAY NOTE Pass
NULL NO SIGNAL
/Fail
CHECK 10.000000006 Start
STANDBY Display OFF Red
LED Hz * count-
ing
beside 6
the key BLANK 10.00000---
ON (3 times) Hz *
ON Backlight ON Self- MENU Displays all available
test functions, processes and
PRESET DEFAULT? Default input controls. Selected
items are blinking.
ENTER NO SIGNAL setting
PRESET DEFAULT? Default
EXT REF EXT REF
ENTER NO SIGNAL setting
Input A **
FILTER FILTER
50 W 50 W
Table 10-2 Keyboard test.
TRGLVL SENSITIVITY NO * The LSD digit may vary.
A TRIG
** MENU is not disabled by
TRGLVL
setting DEFAULT; press
menu again.
tSENS Bar graph:
(2 times) zzzzzzzz
SENSu Bar graph:
(2 times) zzzzzzzzzz
Others
PRESET DEFAULT? Default
ENTER NO SIGNAL setting
MEAS 200.0-3 s
TIME
DATA 500.0-3 s
ENTRYp
Reference Oscillators
Short Form
Crystal oscillators are affected by a number of
Specification Test external conditions such as ambient tempera-
ture and supply voltage but also by aging.
Therefore, it is hard to give limits for the al-
Sensitivity and Frequency lowed frequency deviation. You must decide
Range the limits depending on your application, and
– Press the PRESET key to activate the de- recalibrate the oscillator accordingly. See
fault setting mode (DEFAULT?). Then Chapter 11 Preventive Maintenance.
confirm by pressing ENTER.
Oscillator Max. tem- Max. ag- Max. ag-
– Turn off AUTO by pressing the TRGLVL perature ing per ing per
button to the left of AUTO. depen- month year
dence
– Select 50 W. (Hz) (Hz) (Hz)
100-300 20 -21
Prescaler Check 300-2500 10 -27
2500-2700 20 -21
To verify the specification of the HF input in the
2700-3000 100 -7
instrument:
– If your counter does not have an Input C Table 10-9 Sensitivity of PM9624
connector, skip this test.
– Connect the output of the signal generator
to the HF-input of the counter.
Preventive
Maintenance
Preventive Maintenance
* st st -10
After 1 year of operation. For 1 year add: < 3 x 10
Other Maintenance
Fan Replacement
The PM6685R is equipped with a fan, and if it
is operating in a 24h/day system, you need to
replace the fan every second year to maintain
high reliability. For part-time applications and
low ambient temperatures, an extended service
interval is acceptable.
Optional Battery-Pack
Replacement
The PM9623 rechargeable battery kit contains
two sealed lead-acid batteries. These batteries
are similar to ordinary automotive starter bat-
teries, but they are made maintenance free by
using a gel instead of liquid acid.
The performance of this kind of batteries de-
creases with time, and high ambient tempera-
tures accelerate this performance decrease.
We recommend that you run the capacity test in
the performance check chapter once a year to
ensure correct battery operating time. If your
counter is used in a hot environment, run this
test more often.
When the counter no longer passes the test, the
batteries must be replaced.
Specifications
Specifications
Measuring Single
Event counting on input A with manual start Automatic setting of input signal condition-
and stop. ing circuits for optimum triggering on differ-
ent amplitudes and waveforms.
Range: 0 to 1014 Frequency: Minimum 50 Hz
Frequency Range: 0 to 100 MHz Sensitivity Range: 10 mVrms to 25Vrms
Signal Monitor:
Input and Output A bar graph displays actual input signal level
in 3 dB steps, 10mVrms to 10Vrms
Specifications Low Pass Filter:
100 kHz nominal 3 dB point. Minimum
Input A 40 dB attenuation at 1 MHz.
Damage Level:
Frequency
Range: 10 Hz to 300 MHz 1 MW: 350 V (dc + ac peak) at dc to
Coupling: AC 440 Hz, falling to 12V rms at
Impedance: 1 MW//25 pF or 50W, VSWR 2:1 1 MHz and above
Connector: BNC 50 W: 12Vrms
Sensitivity:
Sinewave: 10 mVrms , 10 Hz to 50 MHz Input C (PM9621)
15 mVrms , 50 MHz to 100 MHz
20 mVrms, 100 MHz to 150 MHz Frequency Range: 70 MHz to 1.3 GHz
30 mVrms, 150 MHz to 200 MHz Prescaling Factor: 512
50 mVrms, 200 MHz to 300 MHz
Operating Input Voltage Range:
Pulse: 50 mVpp, 3 ns minimum pulse
70 to 900 MHz: 10 mV rms to 12 Vrms
width
900 to 1100 MHz: 15 mV rms to 12 Vrms
Dynamic Range: 30 mVpp to 70Vpp 1100 to 1300 MHz: 40 mV rms to 12 Vrms
Manual Trigger: Amplitude Modulation:
Sens.
DC to 0.1 MHz: Up to 94% depth
Range: 10 mVrms to 10Vrms , variable in
0.1 to 6 MHz: Up to 85% depth
3 dB steps, indicated on a bar
graph. Minimum signal must exceed minimum op-
Trigger erating input voltage.
Level: Selectable for optimum trigger-
ing on waveforms with duty Impedance: 50 W nominal
factors <0.25 , 0.25 to 0.75 and Coupling: AC
>0.75 VSWR: <2:1
Trigger Positive or negative Max. Voltage
Slope: Without Damage: 12Vrms, pin-diode prot.
Connector: BNC
Check: Applies 10 MHz to the mea- Analog Output: Select digits and scaling
suring logic factor
Display: LCD with high-luminance Display
backlight Backlight: On/Off
Number of
Digits: 10 plus exponent
Blanking: Least significant digits can be
Bar Graph:
blanked.
Displays input signal level or GPIB
sensitivity setting in 3 dB steps
from 10 mVrms to 10 Vrms (Option PM9626/031)
Programmable
Auxiliary Menu: Functions: All front panel and AUX
Functions available from AUX MENU or via MENU functions
GPIB (PM9626/031): Compatibility: IEEE 488.2-1987,
SCPI 1991.0
Save/Recall: 20 complete instrument set- Interface
tings. 10 settings can be user Functions: SH1, AH1, T6, L4, SR1,
protected RL1, DC1, DT1, E2
GPIB-Addr: Read and temporarily change Max. Meas.
via front panel keys. (Set new Rate to Inter-
address on rear panel switch.) nal Memory: 200 to 1600 readings/s, de-
Burst pending on measurement
Frequency: Input A + C, set synchroniza- function and internal data
tion delay time format
PRF: Input A + C, set synchroniza- Internal Mem-
tion delay time ory Size: 764 to 2600 readings, de-
Trigg. Slope: Positive or negative pending on measurement
Arm. Start: Positive or negative slope, set function and internal data
start arming delay time format
Max. Bus
Arming Stop: Positive or negative slope Transfer Rate
Null: Read and change stored offset from Internal
frequency Memory: 150 to 1000 readings/s, de-
Display pending on internal data for-
Overflow: Display of the 11th and 12th mat and output data format
digits Data Output
Test: Select self-tests Format: ASCII, IEEE double preci-
Program Ver- sion floating point
sion: Display instrument and GPIB Time Out: Off or 100 ms to 25.5 s in
program versions 100 ms steps
Time Out: OFF or 100 ms to 25.5 s in
100 ms steps
GPIB 12-5
Specifications
Temperature
Operating: 0 °C to +40 °C
Storage: -40 °C to +50 °C
Weight: 1.5 kg (3.3 lb)
Measurement Uncertainties
Mea- Random Uncertainty rms Systematic Uncertainty LSD Displayed
suring
func-
tion
Freq- ± Time Base Error ´ Freq (Per )
uency/ E q2 + 2 ´ (Trigger Error ) 2 250 ps ´ Freq (Per )
Period ± ´ Freq (Per )
±
250 ps
´ Freq (Per ) Measuring Time
Measuring Time
Meas.Time
Ratio
f1 /f2 ± ( presc. factor ) 2 + 2 ´ ( f1 ´ Trig . Err. of f 2 ) 2 Prescaler Factor
f 2 ´ Meas. Time
f 2 ´ Meas. Time
Pulse
Width ±Time Base Err. ´ Pulse Width
(Auto ± E q2 + (Start TE ) 2 + (Stop TE ) 2 ±0.5 ´ Transition Time 100 ps
Trig- ±1.5 ns
ger)
Duty
Factor ± E q2 + (Start TE ) 2 + (Stop TE ) 2 ´ Freq. ± (0.5 ´ TranisitionTime + 1 ´ 10 -6
1.5 ns) ´ Frequency
n Systematic Uncertainties
Timebase Options
Product Family PM6685 /PM6685R series
Model:
PM6685 PM6685 PM6685 PM6685R
Option: Standard PM9691 PM9692
Timebase type: UCXO OCXO OCXO Rubidium
Uncertainty due to:
- Cal. adjustment tolerance, at
+23 oC ± 3oC < 1 x 10
-6
< 2 x 10
-8
< 5 x 10
-9
< 5 x 10
-11
- Aging:
-10 1) -10 4)
per 24 hr. n.a. < 5 x 10 < 1 x 10 not specified
per month < 5 x 10 -7 < 1 x 10
-8
< 3 x 10
-9
< 5 x 10 -11 2)
-6 -8 -8 -10 3)
per year < 5 x 10 < 7.5 x 10 < 2 x 10 < 2 x 10
- Temp. Variation:
0o C -50oC, < 1 x 10
-5
< 5 x 10
-9
< 2.5 x 10
-9
< 3 x 10
-10
n Power Requirements
AC Voltage
PM6685: 90 to 265Vrms , 45 to 440 Hz
PM6685R: 90 to 265Vrms , 47 to 63 Hz
DC Voltage
PM6685: Option PM 9623: Internal
battery or external DC
source 12-24 V, max 2A
Timebase Options
PM 6685/_ 1_ Standard Timebase
Appendix
Appendix
Appendix 1, Error
Messages Err. UFLO
Underflow: A math operation in the counter
If the counter detects an internal error or an in- caused an underflow error.
valid setting, it shows an error message on the
nO bUS
display. This appendix lists all possible error
messages. No Bus: No GPIB interface is installed.
Index
Index
! input E . . . . . . . . . . . . . . . . . . . . . 4-11
±1 cycle count error . . . . . . . . . . . . . . . 6-3 manually . . . . . . . . . . . . . . . . . . . . 7-9
50ohm/1Mohm key . . . . . . . . . . . . . . . 4-5 selection . . . . . . . . . . . . . . . . . . . . 7-3
setup time . . . . . . . . . . . . . . . . . . 7-12
A start . . . . . . . . . . . . . . . . . . . . . . . . 9-5
AC coupling . . . . . . . . . . . . . . . . . . . . 5-5 stop . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Address sync. delay . . . . . . . . . . . . . . . . . . 7-9
Auto
Reading the GPIBAddress . . . . . . 9-6
Address switch . . . . . . . . . . . . . . . . . 4-10 once . . . . . . . . . . . . . . . . . . . . . . . 5-5
time to determine levels . . . . . . . . 5-3
Adjustments
time-out . . . . . . . . . . . . . . . . . . . . . 6-4
timebase . . . . . . . . . . . . . . . . . . . 11-2
Auto trigger
Air flow . . . . . . . . . . . . . . . . . . . . . . . . 3-3
levels . . . . . . . . . . . . . . . . . . . . . . . 5-3
AM modulated signals . . . . . . . . . . . . 6-10
Aux
Analog
indication . . . . . . . . . . . . . . . . . . . . 4-8
connector . . . . . . . . . . . . . . . . . . 4-10
menu description. . . . . . . . . 9-2 - 9-10
Analog output
menu key. . . . . . . . . . . . . . . . . . . . 4-6
connector. . . . . . . . . . . . . . . . . . . 4-11
AUX
specification . . . . . . . . . . . . . . . . 12-6
annunciator . . . . . . . . . . . . . . . . . . 9-2
using . . . . . . . . . . . . . . . . . . . . . . . 9-7
Averaging . . . . . . . . . . . . . . . . . . . 6-5, 8-2
Aperture
SEE Measuring time B
Arming Battery
delay . . . . . . . . . . . . . . . . . . . 7-4 - 9-5 disposal . . . . . . . . . . . . . . . . . . . . . 2-3
description . . . . . . . . . . . . . . . . . . . 7-8 test . . . . . . . . . . . . . . . . . . . . . . . 10-8
examples. . . . . . . . . . . . . . . . . . . 7-13
Blank digits . . . . . . . . . . . . . . . . . . . . . 7-5
indication . . . . . . . . . . . . . . . . . . . . 4-8
!- B II
Index
III C-E
Index
Examples Function
arming . . . . . . . . . . . . . . . . . . . . . 7-13 indication . . . . . . . . . . . . . . . . . . . . 4-8
Ext Ref period . . . . . . . . . . . . . . . . . . . . . 6-13
input . . . . . . . . . . . . . . . . . . . . . . 4-10 ratio . . . . . . . . . . . . . . . . . . . . . . . 6-12
key . . . . . . . . . . . . . . . . . . . . . . . . 4-6 selection . . . . . . . . . . . . . . . . . . . . 4-4
output test . . . . . . . . . . . . . . . . . . 10-6 Functions hidden
specification . . . . . . . . . . . . . . . . 12-4 SEE Aux menu
External Arming Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
input test . . . . . . . . . . . . . . . . . . . 10-7
specification . . . . . . . . . . . . . . . . 12-4 G
External Gate Gate
specification . . . . . . . . . . . . . . . . 12-4 indicator. . . . . . . . . . . . . . . . . . . . . 7-3
Go-detector . . . . . . . . . . . . . . . . . . . . . 6-9
F GPIB
Fan Address switch . . . . . . . . . . . . . . 4-10
replacement. . . . . . . . . . . . . . . . . 11-4 connector . . . . . . . . . . . . . . . . . . 4-10
Filter reading the address. . . . . . . . . . . . 9-6
analog LP . . . . . . . . . . . . . . . . . . . 5-5 specification . . . . . . . . . . . . . . . . 12-5
characteristics . . . . . . . . . . . . . . . . 5-5 Grounding . . . . . . . . . . . . . . . . . . . . . . 3-2
key . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Firmware release H
SEE Program version Handle . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Fixed Hidden functions
timeout . . . . . . . . . . . . . . . . . . . . . 6-4 SEE Aux menu
Fold-down support . . . . . . . . . . . . . . . 3-3 Hiding digits
Free-running SEE Digit blanking
measurements . . . . . . . . . . . 7-6, 7-17 Hold display . . . . . . . . . . . . . . . . . . . . 4-7
Freezing the display Hysteresis . . . . . . . . . . . . . . . . . 5-7, 6-14
SEE Display hold SEE ALSO Sensitivity
Frequency. . . . . . . . . . . . . . . . . . . . . . 6-3
I
burst . . . . . . . . . . . . . . . . . . . . . . . 9-5
Identification . . . . . . . . . . . . . . . . . . . . 3-2
burst PRF . . . . . . . . . . . . . . . . . . . 9-6
impedance
range test . . . . . . . . . . . . . . . . . . 10-5
ratio . . . . . . . . . . . . . . . . . . . . . . . 6-12 key . . . . . . . . . . . . . . . . . . . . . . . . 4-5
specification . . . . . . . . . . . . . . . . 12-2 Impedance
uncertainty. . . . . . . . . . . . . . . . . . 12-7 setting . . . . . . . . . . . . . . . . . . . . . . 5-5
Frequency versus time Input
SEE Profiling impedance. . . . . . . . . . . . . . . . . . . 4-5
F-I IV
Index
V K-O
Index
P-R VI
Index
Supply voltage
S setting . . . . . . . . . . . . . . . . . . . . . . 3-2
Safety requirements . . . . . . . . . . . . . . I-VI
Support
Sample-hold . . . . . . . . . . . . . . . . . . . . 6-4
fold-down. . . . . . . . . . . . . . . . . . . . 3-3
Scaling factor
Suppression of noise
analog output . . . . . . . . . . . . . . . . 9-7
SEE Analog lowpass filter
Self-test . . . . . . . . . . . . . . . . . . . 9-6, 10-3
Synchronization of a measurement . . . 6-3
SENS key . . . . . . . . . . . . . . . . . . . . . . 5-4
Syncronization delay
Sensitivity . . . . . . . . . . . . . . . . . . . . . 6-14
arming . . . . . . . . . . . . . . . . . . . . . . 7-9
SEE ALSO Hysteresis
manual setting . . . . . . . . . . . . . . . . 5-4 T
speed of auto setting . . . . . . . . . . . 5-3 Test Equipment . . . . . . . . . . . . . . . . . 10-2
test . . . . . . . . . . . . . . . . . . . . . . . 10-5
Time
Setup time
set the measuring. . . . . . . . . . . . . 7-2
for arming . . . . . . . . . . . . . . . . . . 7-12
period . . . . . . . . . . . . . . . . . . . . . 6-13
Signal detection . . . . . . . . . . . . . . . . . 6-4 to set up arming . . . . . . . . . . . . . 7-12
Single . . . . . . . . . . . . . . 4-6, 6-5 - 7-6, 8-2 Time interval
Single-shot phenomena . . . . . . . . . . . 7-9 SEE Pulse width
Slope. . . . . . . . . . . . . . . . . . . . . . . . . . 4-5 Timebase
Speed calibration . . . . . . . . . . . . . . . . . . 11-2
arming . . . . . . . . . . . . . . . . . . . . . 7-12 Timeout
of measurement . . . . . . . . . . . . . . 6-4
fixed . . . . . . . . . . . . . . . . . . . . . . . 6-4
of the auto function . . . . . . . . . . . . 5-3 manual . . . . . . . . . . . . . . . . . . . . . 6-4
SRQ . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Time-out
Standard Oscillator
set . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
test . . . . . . . . . . . . . . . . . . . . . . . 10-5 Time-Out
Stand-by
auto . . . . . . . . . . . . . . . . . . . . . . . . 6-4
indicator. . . . . . . . . . . . . . . . . . . . . 4-4
Totalize . . . . . . . . . . . . . . . . . . . . . . . 6-16
key . . . . . . . . . . . . . . . . . . . . . . . . 4-4
specification . . . . . . . . . . . . . . . . 12-3
Start
Totalizing
a new measurement . . . . . . . . . . . 4-6
reset . . . . . . . . . . . . . . . . . . . . . . 6-16
arming. . . . . . . . . . . . . . . 7-3, 7-9, 9-5
start . . . . . . . . . . . . . . . . . . . . . . . 6-16
of a measurement . . . . . . . . . 7-6 - 7-7
stop . . . . . . . . . . . . . . . . . . . . . . . 6-16
totalizing . . . . . . . . . . . . . . . . . . . 6-16
Touch-hold
Stop
SEE Sample-hold
arming. . . . . . . . . . . . . . . 7-3, 7-9, 9-5
Transient profiling . . . . . . . . . . . . . . . 7-17
of measurement . . . . . . . . . . . . . . 7-7
totalizing . . . . . . . . . . . . . . . . . . . 6-16 Trigger
auto level. . . . . . . . . . . . . . . . . . . . 5-3
VII S-T
Index
error . . . . . . . . . . . . . . . . . . . . . . . 5-4
hysteresis . . . . . . . . . . . . . . . . . . . 5-7
slope selection . . . . . . . . . . . . . . . 4-5
uncertainty. . . . . . . . . . . . . . . . . . . 5-7
Trigger level . . . . . . . . . . . . . . . . . . . . 4-5
SEE ALSO sensitivity
Trigger Level
output test . . . . . . . . . . . . . . . . . . 10-7
U
Uncertainty
duty factor . . . . . . . . . . . . . . . . . . 12-7
frequency . . . . . . . . . . . . . . . . . . 12-7
period . . . . . . . . . . . . . . . . . . . . . 12-7
PM9691. . . . . . . . . . . . . . . . . . . . 12-8
PM9692. . . . . . . . . . . . . . . . . . . . 12-8
pulse width . . . . . . . . . . . . . . . . . 12-7
random . . . . . . . . . . . . . . . . . . . . 12-7
ratio . . . . . . . . . . . . . . . . . . . . . . . 12-7
rubidium reference . . . . . . . . . . . 12-8
standard timebase. . . . . . . . . . . . 12-8
systematic . . . . . . . . . . . . . . . . . . 12-7
Underflow . . . . . . . . . . . . . . . . . . . . . 13-2
V
VCO
step response profiling . . . . . . . . 7-18
Voltage, supply . . . . . . . . . . . . . . . . . . 3-2
W
Warm-up time . . . . . . . . . . . . . . . . . . 10-2
Warning statements . . . . . . . . . . . . . . 2-2
Waveform . . . . . . . . . . . . . . . . . . . . . . 4-5
how to set . . . . . . . . . . . . . . . . . . . 5-4
Width
of pulse . . . . . . . . . . . . . . . . . . . . 6-15
U-W VIII
Chapter 15
Service Centers
SERVICE CENTERS
http://www.fluke.com
+31-40-2675200 in Europe
+1-425-446-5500 from other countries
II