TLB_manual_EN (1)
TLB_manual_EN (1)
TLB
SYMBOLS
Here are the symbols used in the manual to draw the reader's attention:
Further information.
WARRANTY
24 months from the date of the delivery note. Warranty covers only failures of defective components (due to construction
defects or defects in materials) and includes replacement or repair of the components and related labor costs.
Warranty is automatically forfeited in the event of:
- tampering, deletion, removal of the identification label and/or serial number of the product
- misuse, transformation, alteration, repair of products not carried out by Laumas personnel
Laumas provides a 1-year warranty from the date of the delivery note on defects in material or manufacture of the battery.
The terminals indicated on the instrument’s wiring diagram to be connected to earth must have
the same potential as the weighed structure (same earthing pit or earthing system). If you are
unable to ensure this condition, connect with an earthing wire the terminals of the instrument
(including the terminal –SUPPLY) to the weighed structure.
The cell cable must be individually led to its panel input and not share a conduit with other cables;
connect it directly to the instrument terminal strip without breaking its route with support terminal strips.
Use “RC” filters on the instrument-driven solenoid valve and remote control switch coils.
Avoid inverters in the instrument panel; if inevitable, use special filters for the inverters and separate
them with sheet metal partitions.
The panel installer must provide electric protections for the instruments (fuses, door lock switch etc.).
It is advisable to leave the equipment always switched on to prevent the formation of condensation.
-1-
RECOMMENDATIONS FOR CORRECT INSTALLATION OF THE LOAD CELLS
In case of weighing with dynamic loads, the installer has to estimate the thrust speed, the acceleration,
the frequency, etc.
WELDING
Avoid welding with the load cells already installed. If this cannot be avoided, place the welder ground
clamp close to the required welding point to prevent sending current through the load cell body.
-2-
EARTHING THE WEIGHED STRUCTURE
By means of a copper wire with suitable cross-section, connect the cell upper support plate with the
lower support plate, then connect all the lower plates to a single earthing system. Electrostatic charges
accumulated because of the product rubbing against the pipes and the weighed container walls are
discharged to the ground without going through or damaging the load cells. Failure to implement a
proper earthing system might not affect the operation of the weighing system; this, however, does not
rule out the possibility that the cells and connected instrument may become damaged in the future. It is
forbidden to ensure earthing system continuity by using metal parts contained in the weighed structure.
-3-
LOAD CELL INPUT TEST (QUICK ACCESS)
3s
From the weight display, press for 3 seconds: the display shows the response signal of the load
cells expressed in mV with four decimals.
Example: a load cell with 2.000 mV/V sensitivity provides a response signal between 0 and 10 mV.
IF ONE OF THE ABOVE CONDITIONS IS NOT MET, PLEASE CONTACT THE TECHNICAL
ASSISTANCE SERVICE.
-4-
MAIN SPECIFICATIONS OF THE INSTRUMENT
Weight transmitter with 6-wire load cells input suitable for assembly on back panel fitted Omega/DIN
rail. Dimensions: 25x115x120 mm. Six-digit semi-alphanumeric display, 8 mm height, 7 segment.
Four-key keyboard.
RS485 serial port for connection to: PC/PLC up to 32 instruments (max 99 with line repeaters) by
ASCII Laumas or ModBus R.T.U. protocol, remote display.
Optional: integrated CANopen, DeviceNet, CC-Link, PROFIBUS DP, Modbus/TCP, Ethernet TCP/IP,
Ethernet/IP, PROFINET IO, EtherCAT, POWERLINK, SERCOS III output.
TLB only: optoisolated 16 bit analog output (tension or current).
-5-
TECHNICAL SPECIFICATIONS
-6-
ELECTRICAL CONNECTIONS
BASIC INFORMATION
-7-
WIRING DIAGRAM
OUT COMMON
1 2 3 4 5 6
+12/24 VDC
-SIG
+SIG
-EX
+EX
+RS485
-RS485
0 VDC
OUT 1
OUT 2
OUT 3
13 14 15 16 17 18 19 20
TLB ONLY
1 2 3 4 5 6 7 8 9 10 11 12
IN 1
- COMMON
IN COMMON
-SIG
+SIG
IN 2
+REF
-REF
+ 0÷10 VDC
-EX
+EX
+ 0÷20 - 4÷20 mA
SH
TERMINALS LEGEND
-8-
TLB (other models)
OUTPUTS 12/24 VDC
max 115 VAC 150 mA RS485 SUPPLY
OUT COMMON
+12/24 VDC
+RS485
-RS485
0 VDC
OUT 1
OUT 2
OUT 3
15 16 17 18 19 20 21 22
1 2 3 4 5 6 IN 1 7 8 9 1 2 3 4 5 6
IN COMMON
-SIG
+SIG
-SIG
+SIG
IN 2
+REF
-REF
-EX
-EX
+EX
+EX
SH
TERMINALS LEGEND
-9-
LED AND KEYS FUNCTIONS
*) To activate the secondary LED function, during weight display press and hold down the keys
and (press immediately followed by ).
Into menus LEDs light up in sequence to indicate that it is not displaying a weight.
- 10 -
MENU MAP
Into menus changes are applied right after pressing the key (no further confirmation is required).
SETPOINT
… …
SYSTEM PARAMETERS
- 11 -
INSTRUMENT COMMISSIONING
Check that the display shows the weight and that when loading the load cells there is an increase in
weight. If there is not check and verify the connections and correct positioning of the load cells.
- If the instrument has already been theoretical CALIBRATED (plant system identification tag
present on the instrument and on the cover: load cell’s rated data already entered):
▫ Reset to zero (see section TARE WEIGHT ZERO SETTING).
▫ Check the calibration with sample weights and correct the indicated weight if necessary (see
section REAL CALIBRATION (WITH SAMPLE WEIGHTS)).
- If the instrument HAS NOT BEEN CALIBRATED (missing plant system identification tag)
proceed with calibration:
▫ If load cells data are unknown, follow the procedure in section REAL CALIBRATION (WITH
SAMPLE WEIGHTS).
▫ Enter the rated data of load cells following the procedure given in section THEORETICAL
CALIBRATION.
▫ Reset to zero (see section TARE WEIGHT ZERO SETTING).
▫ Check the calibration with sample weights and correct the indicated weight if necessary (see
section REAL CALIBRATION (WITH SAMPLE WEIGHTS)).
- If you use the analog output, set the desired analog output type and the full scale value (see
section ANALOG OUTPUT).
- If you use serial communication, set the related parameters (see section SERIAL
COMMUNICATION SETTING).
- If setpoint are used, set the required weight values and the relevant parameters (see sections
SETPOINT PROGRAMMING and OUTPUTS AND INPUTS CONFIGURATION).
- 12 -
PROGRAMMING OF SYSTEM PARAMETERS
From the weight display, press simultaneously keys and to access the parameter setting.
THEORETICAL CALIBRATION
- By modifying the theoretical full scale or the sensitivity, the real calibration is cancelled
and the theoretical calibration only is considered valid.
- If the theoretical full scale and the recalculated full scale in real calibration (see section
REAL CALIBRATION (WITH SAMPLE WEIGHTS)) are equal, this means that the
calibration currently in use is theoretical; if they are different, the calibration in use is the
real calibration based on sample weights.
- By modifying the theoretical full scale or the sensitivity, the system’s parameters
containing a weight value will be set to default values (setpoint, hysteresis, etc.).
- 13 -
MAXIMUM CAPACITY
: maximum displayable weight (from 0 to full scale; default: 0). When the weight exceeds this
value by 9 divisions, the display shows . To disable this function, set 0.
Perform this procedure after having set the THEORETICAL CALIBRATION data.
Use this function to set to zero the weight of the empty system after commissioning and then later on
to compensate zero variations due to the presence of product residues.
Procedure:
- Confirm the message by pressing .
- The weight value to be set to zero is displayed. In this phase all of the LEDs are flashing.
- Confirming once again, the weight is set to zero (the value is stored to the permanent memory).
- Press to display the value of the total weight reset by the instrument, given by the sum of all
of the previous zero settings.
- 14 -
REAL CALIBRATION (WITH SAMPLE WEIGHTS)
Load onto the weighing system a sample weight, which must be at least 50% of the maximum quantity
to be weighed.
By confirming the message the flashing value of the weight currently on the system is
displayed. In this phase all of the LEDs are off. Adjust the value on display by using the arrow keys if
necessary. After confirming, the new set weight will appear with all the LEDs flashing. After an
additional confirmation, the message will be restored and by repeatedly pressing the key
the weight will once again be displayed.
Example: for a system of maximum capacity 1000 kg and 1 kg division, two sample weights are
available, one of 500 kg and the other one of 300 kg. Load both weights onto the system and correct
the indicated weight to 800. Now remove the 300 kg weight, the system must show 500; remove the
500 kg weight too; the system must read zero. If this does not happen, it means that there is a
mechanical problem affecting the system linearity.
WARNING: identify and correct any mechanical problems before repeating the procedure.
- If theoretical full scale and recalculated full scale in real calibration are equal, it means that
the theoretical calibration is currently in use; otherwise, the real calibration based on
sample weights is in use.
- If the correction made changes the previous full scale for more than 20%, all the
parameters with settable weight values are reset to default values.
- 15 -
FILTER ON THE WEIGHT
The filter enables to stabilise a weight as long as its variations are smaller than the corresponding
“response time”. It is necessary to set this filter according to the type of application and to the full scale
value set.
ANTI PEAK
When the weight is stable, the anti peak filter removes any sudden disturbances with a maximum
duration of 1 second. Confirm the filter on the weight with and select one of the following options:
- : anti peak filter enabled (default);
- : anti peak filter disabled.
- 16 -
ZERO PARAMETERS
(from 0 to full scale; default: 300; considered decimals: 300 – 30.0 – 3.00 – 0.300): this parameter
indicates the maximum weight value resettable by external contact, keypad or serial protocol.
(from 0 to 10% of full scale; default: 0): if at switch-on the weight value is lower than the
value set in this parameter, the weight is reset. To disable this function, set 0.
ZERO TRACKING
(from 1 to 5, default: ): when the weight value is stable and, after a second, it deviates
from zero by a figure in divisions smaller or equal to the figure in divisions set in this parameter, the
weight is set to zero. To disable this function, set .
Example: if the parameter is set to 5 and is set to 2, the weight will be automatically
set to zero for variations smaller than or equal to 10 ( x ).
- 17 -
SETTING UNITS OF MEASURE
: kilograms
: grams
: tons
: pounds*
: newtons*
: litres*
: bars*
: atmospheres*
: pieces*
: newton metres*
: kilogram metres*
: other generic units of measure not included in the list*
For the units marked with * it’s possible to set also the display coefficient (parameter
, see the related section). To use is necessary to enable it, closing the
input (see section OUTPUTS AND INPUTS CONFIGURATION).
DISPLAY COEFFICIENT
If one of the inputs is set to mode (see section OUTPUTS AND INPUTS CONFIGURATION)
when the input is closed the value will be displayed modified according to the coefficient; when
the input is opened the standard weight display will be restored.
(max settable value: 99.9999; default: 1.0000): will have different meanings according to the
value set in , i.e. the selected unit of measure. (see section SETTING UNITS OF MEASURE).
- 18 -
If the unit of measure chosen is:
: pounds, the value set in will be multiplied by the weight value currently displayed;
: newton, the value set in will be multiplied by the weight value currently displayed;
: litres, in set the specific weight in kg/l, assuming that the system is calibrated in kg;
: bar, the value set in will be multiplied by the weight value currently displayed;
: atmosphere, the value set in will be multiplied by the weight value currently displayed;
: pieces, in set the weight of one piece;
: newton metres, the value set in will be multiplied by the weight value currently displayed;
: kilogram metres, the value set in will be multiplied by the weight value currently
displayed;
: generic unit of measure not included in the list, the value set in will be multiplied by
the weight value currently displayed.
WARNING: All other settings (setpoint, hysteresis, calibration ...) are expressed in weight
value. If you want to convert them to the new unit of measurement, perform one of the
following procedures for changing the system calibration.
The parameter must remain set to 1.0000.
- 19 -
OUTPUTS AND INPUTS CONFIGURATION
OUTPUTS
The outputs are set by default as follows: / / / / .
Possible operation modes:
- (normally open): the relay is de-energised and the contact is open when the weight is lower
than the programmed setpoint value; it closes when the weight is higher than or equal to the
programmed setpoint value.
- (normally closed): the relay is energised and the contact is closed when the weight is
lower than the programmed setpoint value; it opens when the weight is higher than or equal to the
programmed setpoint value.
- : the contact will switch on the basis of weight, according to setpoint (see section SETPOINT
PROGRAMMING).
- : the contact will not switch on the basis of weight, but is controlled by remote protocol
commands.
- : relay switching occurs when the weight is stable.
- : relay switching occurs when one of the following alarms is triggered: , ,
, , ; the operation mode is forced to (normally closed).
If the operation mode is selected, the following options are also active:
- : the contact will switch on the basis of gross weight.
- : the contact will switch on the basis of net weight (If the net function is not active, the contact
will switch on the basis of gross weight).
- : relay switching occurs for both positive and negative weight values.
- : relay switching occurs for positive weight values only.
- : relay switching occurs for negative weight values only.
- 20 -
Examples:
= / / / / = / / / /
= 500 kg; = 0 kg = 500 kg; = 0 kg
= / / / / = / / / /
= 500 kg; = 100 kg = 500 kg; = 0 kg
= / / / / = / / / /
= 500 kg; = 0 kg = 0 kg; = 0 kg
- 21 -
INPUTS
Default: input 1 = input 2 =
To perform a net operation (SEMI-AUTOMATIC TARE), close the NET/GROSS input or press the
key for less than 3 seconds. The instrument displays the net weight (just set to zero) and the NET
symbol lights up. To display the gross weight again, keep the NET/GROSS input closed or press
for 3 seconds.
This operation can be repeated many times by the operator to allow the loading of several products.
Example:
Put the box on the scale, the display shows the box weight; press , the display shows the net
weight to zero; introduce the product in the box, the display shows the product weight. This operation
can be repeated several times.
While the net weight is displayed, keep pressed to display the preset tare. When
the key is released the net weight will be displayed again.
The semi-automatic tare operation is not allowed if the gross weight is zero.
- 22 -
PRESET TARE (SUBTRACTIVE TARE DEVICE)
+
It is possible to manually set a preset tare value to be subtracted from the display
value provided that the ≤ max capacity condition is verified.
By default the instrument shows the last programmed preset tare value: to apply it press and then
.
After setting the tare value, going back to the weight display, the display shows the net weight
(subtracting the preset tare value) and the NET LED lights up to show that a tare has been entered.
To delete a preset tare and return to gross weight display, hold down for about 3 seconds or keep
the NET/GROSS input (if any) closed for the same length of time (3 seconds). The preset tare value
is set to zero. The NET LED is turned off when the gross weight is displayed once again.
While the net weight is displayed, keep pressed to display the gross weight. When
the key is released the net weight will be displayed again.
ALL THE SEMI-AUTOMATIC TARE (NET) AND PRESET TARE FUNCTIONS WILL BE
LOST WHEN THE INSTRUMENT IS TURNED OFF.
- 23 -
SEMI-AUTOMATIC ZERO (WEIGHT ZERO-SETTING FOR SMALL VARIATIONS)
By closing the SEMI-AUTOMATIC ZERO input, the weight is set to zero; alternatively, by pressing the
key, the message is displayed for 3 seconds, by pressing the weight is set to zero.
This function is only allowed if the weight is lower than the value (see section RESETTABLE
WEIGHT SETTING FOR SMALL WEIGHT CHANGES), otherwise the alarm appears and
the weight is not set to zero.
PEAK
By keeping the PEAK input closed the maximum weight value reached remains displayed. By opening
the input the current weight is displayed.
If you wish to use this input to view a sudden variation peak, set the FILTER ON THE
WEIGHT to 0.
- 24 -
ANALOG OUTPUT (TLB ONLY)
- : it selects the analog output type (4÷20 mA, 0÷20 mA, 0÷10 V, 0÷5 V, ±10 V, ±5 V;
default: 4÷20 mA).
For the output ±10 V and ±5 V a soldered jumper must be closed (the operation makes
the other analog outputs unusable):
▫ open the instrument;
▫ locate in the printed circuit board the soldered jumper highlighted in the picture below:
- : set the weight value for which you wish to obtain the minimum analog output value.
Only set a value different from zero if you wish to limit the analog output range; for instance:
for a full scale value of 10000 kg you require an 4 mA signal at 5000 kg and 20 mA at
10000 kg, in this case, instead of zero, set 5000 kg.
- : set the weight value for which you wish to obtain the maximum analog output value; it
must correspond to the value set in the PLC program (default: calibration full scale). E.g.: if I am
using a 4÷20 mA output and in the PLC program I wish to have 20 mA = 8000 kg, I will set the
parameter to 8000.
- 25 -
- : analog output correction to zero: if necessary adjust the analog output, allowing the PLC
to indicate 0. The sign “-“ can be set for the last digit on the left. E.g.: if I use a 4÷20 mA output
and, with the minimum analog setting, the PLC or tester read 4.1 mA, I must set the parameter to
3.9 to obtain 4.0 on the PLC or tester.
- : correction of analog output to full scale: if necessary permit modification of the analog
output by allowing PLC to indicate the value set in the parameter . E.g. if I am using a
4÷20 mA output with the analog set to full scale and the PLC or tester reads 19.9 mA, I must set
the parameter to 20.1 to get 20.0 on the PLC or tester.
Minimum and maximum values which can be set for zero and full scale corrections:
ANALOG OUTPUT TYPE Minimum Maximum
0÷10 V -0.150 10.200
0÷5 V -0.150 5.500
±10 V -10.300 10.200
±5 V -5.500 5.500
0÷20 mA -0.200 22.000
4÷20 mA -0.200 22.000
NOTE: the analog output may also be used in the opposite manner, i.e. the weight setting that
corresponds to the analog zero ( ) may be greater than the weight set for the analog full scale
( ). The analog output will increase towards full scale as the weight decreases; the analog
output will decrease as the weight increases.
For example:
= 10000 = 0 analog output 0÷10 V
All analog outputs of the instrument are ACTIVE and SINGLE ENDED type, therefore they
can be connected only to PASSIVE receiver devices. The minimum load allowed for
voltage outputs is 10 kohm, the maximum load allowed for current outputs is 300 ohm.
- 26 -
SERIAL COMMUNICATION SETTING
- : transmission speed (1200, 2400, 4800, 9600, 19200, 38400, 115200; default: 9600).
- : instrument address (from 1 to 99; default: 1).
- : maximum transmission frequency (10 – 20 – 30 – 40 – 50 – 60 – 70 – 80 – 100 –
200 – 300; default: 10); to be set when the transmission protocol is selected.
Maximum setting frequency ():
- 20 Hz with minimum baud rate 2400 baud.
- 40 Hz with minimum baud rate 4800 baud.
- 80 Hz with minimum baud rate 9600 baud.
- 100 Hz with minimum baud rate 19200 baud.
- 200 Hz with minimum baud rate 38400 baud.
- 300 Hz with minimum baud rate 38400 baud.
- 27 -
- : delay in milliseconds which elapses before the instrument replies (from 0 to
200 ms; default: 0).
- :
- : no parity (default).
- : even parity.
- : odd parity.
- : stop bit (1 – 2; default: 1).
For more information about protocols and methods of communication, request the
proper manual to technical assistance.
GND
GND
RS485 +
RS485 +
RS485 +
RS485 -
RS485 -
RS485 -
max 500 m
24 VDC
-
+
PC RS232
- 28 -
DIRECT CONNECTION BETWEEN RS485 AND RS232 WITHOUT CONVERTER
Since a two-wire RS485 output may be used directly on the RS-232 input of a PC or remote display,
it is possible to implement instrument connection to an RS-232 port in the following manner:
INSTRUMENT RS232
RS485 – → RXD
RS485 + → GND
This type of connection allows A SINGLE instrument to be used in a ONE WAY mode.
TEST
INFO MENU
- 29 -
SETPOINT PROGRAMMING
… …
- (from 0 to full scale; default: 0): Setpoint; relay switching occurs when the weight exceed the
value set in this parameter. The type of switching is settable (see section OUTPUTS AND INPUTS
CONFIGURATION).
- (from 0 to full scale; default: 0): Hysteresis, value to be subtracted from the setpoint to
obtain contact switching for decreasing weight. For example with a setpoint at 100 and hysteresis
at 10, the switching occurs at 90 for decreasing weight.
These values are set to zero if the calibration is changed significantly (see sections
THEORETICAL CALIBRATION and REAL CALIBRATION (WITH SAMPLE WEIGHTS)).
- 30 -
ALARMS
: the load cell is not connected or is incorrectly connected; the load cell signal exceeds
39 mV; the conversion electronics (AD converter) is malfunctioning; the load cell is a 4-wire
and there are no jumpers between EX- and REF- and between EX+ and REF+; the
references are not connected or are incorrectly connected.
: the weight display exceeds 110% of the full scale.
: internal instrument converter failure; check load cell connections, if necessary contact
technical assistance.
: the weight exceeds the maximum capacity by 9 divisions.
: maximum displayable value exceeded (value higher than 999999 or lower than -999999).
: weight too high: zero setting not possible.
: this message appears in the sample weight setting, in real calibration, after the eighth
sample weight value has been entered.
: the value set for the parameter is beyond the permitted values; press to quit the setting
mode leaving the previous value unchanged. Examples: a number of decimals is selected
for full scale which exceeds the instrument's display potential; value above the maximum
setting value; the weight value set in sample weight verification does not match the detected
mV increase; the analog output correction goes beyond the permitted limits.
: lock active on menu item, keypad or display.
: it’s not possible to display properly the number because is greater than 999999 or less
than -999999.
: gross weight equal to zero: the semi-automatic tare operation cannot be performed.
: issues with the fieldbus device.
With an alarm the relays open and the analog outputs go to the lowest possible value
according to the following table:
RANGE 0÷20 mA 4÷20 mA 0÷5 V 0÷10 V ±10 V ±5 V
Output value -0.2 mA 3.5 mA -0.5 V -0.5 V 0V 0V
- 31 -
RESERVED FOR THE INSTALLER
MENU LOCKING
Through this procedure, it’s possible to block the access to any menu on the instrument.
Select the menu that you wish to lock:
press simultaneously for 3 seconds, the display shows
(the left point on the text indicates that this menu item is now locked). If the operator tries to enter this
menu, the access is denied and the display shows .
MENU UNLOCKING
press simultaneously for 3 seconds, the display shows
(the left point on the text is off to indicate that this menu item is unlocked).
press and simultaneously for 3 seconds: it is now possible to enter
and modify all menus including those which are locked. By returning to weight display, the menu lock
is restored.
Upon instrument power-on hold down the key until the display shows , then proceed as follows:
Constants restore (does not erase the calibration): confirm , use arrow keys to select ,
set code 6935 and confirm
Program selection: confirm and use the arrow keys to select the desired program
- : basic program, setpoint management only
- 32 -
Set the approval status:
- : not approved program
- : approved program, single interval*
- : approved program, multi-interval*
* Contact technical assistance to request the proper manual and the correct procedures for
approval, indicating mandatory hardware code and serial number (see section INSTRUMENT
COMMISSIONING).
Set the active regulation on the instrument (only if have not been set), see the standards
observed in the regions listed in the TECHNICAL SPECIFICATIONS section of the approved
manuals
- : approved program according to OIML
- : approved program according to NTEP
- : approved program according to INMETRO
If you do not have a specific manual for the newly set program, you can request it
to technical assistance.
Press immediately followed by and hold them down for about 5 seconds (this operation is also
possible via the MODBUS and ASCII protocols):
- : no lock.
- : keypad lock: if active, when a key is pressed the message is displayed for 3 seconds.
- : keypad and display lock: if active, the keypad is locked and the display shows the instrument
model (weight is not displayed); by pressing a key the display shows for 3 seconds.
- 33 -
DECLARATION OF CONFORMITY - EU
Email: [email protected] Fabbricante metrico Prot. N. 7340 Parma - R.E.A. PR N. 169833 - Reg. Imprese PR N.19393 - Registro Nazionale Pile
Web: http://www.laumas.com N. IT09060P00000982 - Registro A.E.E. N. IT08020000002494 - N. Mecc. PR 008385 - Cap. Sociale € 100.000 int. vers.
SISTEMA QUALITÀ CERTIFICATO UNI EN ISO 9001 - SISTEMA GESTIONE AMBIENTALE ISO 14001 - MODULO D: GARANZIA DELLA QUALITÀ DEL PROCESSO DI PRODUZIONE
Dichiarazione di Dichiariamo che il prodotto al quale la presente dichiarazione si riferisce è conforme alle norme di seguito
I
conformità citate.
Declaration of
GB We hereby declare that the product to which this declaration refers conforms with the following standards.
conformity
Declaración de Manifestamos en la presente que el producto al que se refiere esta declaración está de acuerdo con las
E
conformidad siguientes normas
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nachstehenden
D Konformitäts-erklärung
Normen übereinstimmt.
Déclaration de Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est
F
conformité conforme aux normes citées ci-après.
Tímto prohlašujeme, že výrobek, kterého se toto prohlášení týká, je v souladu s níže uvedenými
CZ Prohlášení o shode
normami.
Wij verklaren hiermede dat het product, waarop deze verklaring betrekking heeft, met de hierna vermelde
NL Conformiteit-verklaring
normen overeenstemt.
Declaração de Declaramos por meio da presente que o produto no qual se refere esta declaração, corresponde às
P
conformidade normas seguintes.
Niniejszym oświadczamy, że produkt, którego niniejsze oświadczenie dotyczy, jest zgodny z poniższymi
PL Deklaracja zgodności
normami.
Заявление о Мы заявляем, что продукт, к которому относится данная декларация, соответствует
RUS
соответствии перечисленным ниже нормам.
Models: TLB, TLB+/-10, TLB+/-5, TLB+05, TLB+010, TLB+020, TLB+420, TLB485, TLBCANOPEN,
TLBDEVICENET, TLBCCLINK, TLBPROFI, TLBMODBUSTCP, TLBETHETCP, TLBETHEIPN,
TLBPROFINETION, TLBETHERCAT, TLBPOWERLINK, TLBSERCOS
M 2014/31/EU
NAWI Directive
EN 45501:2015
OIML R76-1:2006
(only if “M” mark is applied)
- 34 -
DECLARATION OF CONFORMITY - UKCA
Email: [email protected] Fabbricante metrico Prot. N. 7340 Parma - R.E.A. PR N. 169833 - Reg. Imprese PR N.19393 - Registro Nazionale Pile
Web: http://www.laumas.com N. IT09060P00000982 - Registro A.E.E. N. IT08020000002494 - N. Mecc. PR 008385 - Cap. Sociale € 100.000 int. vers.
SISTEMA QUALITÀ CERTIFICATO UNI EN ISO 9001 - SISTEMA GESTIONE AMBIENTALE ISO 14001 - MODULO D: GARANZIA DELLA QUALITÀ DEL PROCESSO DI PRODUZIONE
Dichiarazione di Dichiariamo che il prodotto al quale la presente dichiarazione si riferisce è conforme alle norme di seguito
I
conformità citate.
Declaration of
GB We hereby declare that the product to which this declaration refers conforms with the following standards.
conformity
Declaración de Manifestamos en la presente que el producto al que se refiere esta declaración está de acuerdo con las
E
conformidad siguientes normas
Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den nachstehenden
D Konformitäts-erklärung
Normen übereinstimmt.
Déclaration de Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente déclaration, est
F
conformité conforme aux normes citées ci-après.
Tímto prohlašujeme, že výrobek, kterého se toto prohlášení týká, je v souladu s níže uvedenými
CZ Prohlášení o shode
normami.
Wij verklaren hiermede dat het product, waarop deze verklaring betrekking heeft, met de hierna vermelde
NL Conformiteit-verklaring
normen overeenstemt.
Declaração de Declaramos por meio da presente que o produto no qual se refere esta declaração, corresponde às
P
conformidade normas seguintes.
Niniejszym oświadczamy, że produkt, którego niniejsze oświadczenie dotyczy, jest zgodny z poniższymi
PL Deklaracja zgodności
normami.
Заявление о Мы заявляем, что продукт, к которому относится данная декларация, соответствует
RUS
соответствии перечисленным ниже нормам.
Models: TLB, TLB+/-10, TLB+/-5, TLB+05, TLB+010, TLB+020, TLB+420, TLB485, TLBCANOPEN,
TLBDEVICENET, TLBCCLINK, TLBPROFI, TLBMODBUSTCP, TLBETHETCP, TLBETHEIPN,
TLBPROFINETION, TLBETHERCAT, TLBPOWERLINK, TLBSERCOS
- 35 -
On our website www.laumas.com there are videos on the guidelines for correct installation of
weighing systems and video tutorials on configuring our transmitters and weight indicators.
All Laumas product manuals are available online. You can download the manuals in PDF format
from www.laumas.com by consulting the Products section or the Download Area.
Registration is required.