9.1-01 Installation Manual
9.1-01 Installation Manual
Contents
䉬 Notice ......................................................................................................... 2
䉬 Introduction ............................................................................................... 5
1. Overview ...................................................................................................13
1.1 System Configuration ...................................................................................... 13
1.1.1 Wiring and Piping Diagram .................................................................14
1.1.2 External Dimensions ...........................................................................18
1.2 Auxiliary Provisions ..........................................................................................26
1.3 Necessary Operation Conditions for Gases .................................................... 27
1.4 Standard Performance ......................................................................................28
1.5 Data Plate 具CENTELEC Atex directive典 ............................................................30
TI 11B03A03-03E
©Copyright Oct. 2001(YK)
2nd Edition Sep. 2002(YK)
Yokogawa Electric Corporation
<Toc> <Ind> <Rev> <Introduction> 2
䉬 Notice
This Manual is to provide the technical information for installing a GC1000 Mark II at
site.
When installing or checking up the installation of GC1000 Mark II, read this sentence
carefully.
WARNING
A WARNING sign denotes a hazard. It calls attention to procedure, practice,
condition or the like, which, if not correctly performed or adhered to, could result in
injury or death of personnel.
CAUTION
A CAUTION sign denotes a hazard. It calls attention to a procedure, practice,
condition or the like, which, if not correctly performed or adhered to, could result in
damage to or destruction of part or all of the product.
Media No. TI 11B03A03-03E 2nd Edition : Sept. 2002 (YK) TI 11B03A03-03E 2nd Edition : Sept. 30,2002-00
All Rights Reserved Copyright ©2001, Yokogawa Electric Corporation
<Toc> <Ind> <Rev> <Introduction> 3
IMPORTANT:
Indicates that operating the hardware or software in this manner may damage it or
lead to system failure.
NOTE:
Draws attention to information essential for understanding the operation and
features.
TIP:
Gives information that complements the present topic.
See Also:
Gives the reference locations for further information on the topic.
Alternating current
Indicates the power switch state “ON”.
3. If protection / safety circuits are to be used for the product or the system controlled
by it, they should be installed outside of the product.
4. When you replace parts or consumables of the product, use those specified by our
company.
5. Do not modify the product
䉬 Introduction
Thank you for purchasing the GC1000 Mark II process gas chromatograph.
This manual describes the installation of Model GC1000D / GC1000S / GC1000T /
GC1000E / GC1000W / GC1000C (Hereafter, it is abbreviated as GC1000 Mark II)
Process Gas Chromatograph.
Please read the following respective documents before installing and using the
GC1000 Mark II system.
2. Operation Data
Operation data is supplied with the operation manuals in the delivered package and
contains the following required to use the gas chromatographs.
● Process conditions and measurement range
● Instrument specifications and operating conditions
● Standard sample for calibration
● General Precautions
WARNING
(1) In order to analyze gases, process gas chromatographs use a sample of the
process gas and utility gases.
Since these gases are typically combustible, combustion-sustaining, toxic,
odorous, resolvable, polymerizing, or corrosive, refer to the “Safety Informa-
tion” in our approval drawings and others to ensure safety thoroughly
before using these analyzers.
(2) Up to two protection systems, each of which weighs approximately 10 kg, are
installed on top of the GC1000. Therefore, the center of gravity is higher than the
center of the analyzer body.
Take great care when carrying and installing (piping- wiring) the GC1000.
The GC1000 must be carried and installed very carefully (including piping
and wiring) by more than one person (at least four people are recom-
mended).
(3) Since the GC1000 are precision instruments, take care when handling not to
jolt of knock them.
(4) Use the GC1000 within the range of your purchase specifications.
Yokogawa assumes no responsibility for problems resulting from use by the
customer outside the purchase specifications.
If the GC1000 need to be modified or repaired, please contact your nearest
Yokogawa representative. Yokogawa assumes no responsibility for results where
the customer or any third party has attempted to modify or repair these products.
(5) When touching LCD Panel switches
When touching LCD Panel switches, first, discharge Electro Static Charge of the
body. Then, touch the LCD Panel switches. If not, LCD display may be changed
by Electro Static Discharge.
IMPORTANT:
(1) Read the attached instruction manual before operating the GC1000
(2) The instruments must be installed and operated according to the installation
manual, instruction manual, approval drawings, and operation data.
<GC1000D / GC1000S>
• JIS Expd IIB + H2 T1 (programmed-temperature oven 320˚C max., isothermal
oven 225˚C max., liquid-sample valve 250˚C max.)
• JIS Expd IIB + H2 T2 (programmed-temperature oven 225˚C max., isothermal
oven 225˚C max., liquid-sample valve 225˚C max.)
• JIS Expd IIB + H2 T3 (programmed-temperature oven 145˚C max., isothermal
oven 145˚C max., liquid-sample valve 145˚C max.)
• JIS Expd IIB + H2 T4 (programmed-temperature oven 95˚C max., isothermal oven
95˚C max., liquid-sample valve 95˚C max.)
<GC1000W / GC1000C>
CENELEC (ATEX directive) certified : Group II Category 2G
• EEx pd II B +H2 T1 (programmed-temperature oven 320°C max, isothermal oven
225°C max., liquid-sample valve 250°C max.)
• EEx pd II B +H2 T2 (programmed-temperature oven 225°C max, isothermal oven
225°C max., liquid-sample valve 225°C max.)
• EEx pd II B +H2 T3 (programmed-temperature oven 145°C max, isothermal oven
145°C max., liquid-sample valve 145°C max.)
• EEx pd II B +H2 T4 (programmed-temperature oven 95°C max, isothermal oven
95°C max., liquid-sample valve 95°C max.)
<GC1000T / GC1000E>
FM Type X purging and Explosionproof for CLI, DIV1, GPS B, C & D, NEMA3R
Type Y purging and Type X purging for CLI, DIV1, GPS B, C & D, NEMA3R
• T1 (programmed-temperature oven 320˚C max., isothermal oven 225˚C max.,
liquid-sample valve 250˚C max.)
• T2 (programmed-temperature oven 225˚C max., isothermal oven 225˚C max.,
liquid-sample valve 225˚C max.)
• T3 (programmed-temperature oven 145˚C max., isothermal oven 145˚C max.,
liquid-sample valve 145˚C max.)
• T4 (programmed-temperature oven 95˚C max., isothermal oven 95˚C max., liquid-
sample valve 95˚C max.)
(7) Override function (The analyzer with optional code "FM/CSA Type Y
purging" does not have this function.)
WARNING
• When the override function is used, Analyzer becomes an ignition source and the
high temperature and the high voltage part will be exposed.
• Please confirm that in the ambient atmosphere, the concentration of explosive
gases is less than the allowable limit, by using a gas detector.
To return to the normal operation, turn off “the override switch” and then close the door
as it was before turning on power.
In this analyzer, if the pressure of the pressurized / type X purged enclosure system
(oven, electronic section) drops while the power is on, the pressurized explosion pro-
tection section is activated to stop power supply. Therefore, in case of opening the door
of the oven or of the electronic section inadvertently, for maintenance, while the power
is on, the protection system is activated to cut off the power.
The “override function” intensively releases this function of protection system.
The override switch is installed in section.
WARNING
<CENELEC>
* Only trained persons may use this instrument in a hazardous location.
* Do not open when energized.
<FM>
For type X purging:
* This equipment contains components that operate at high temperature. The
equipment shall be deenergized for 60 minutes to permit those components to
cool before the enclosure is opened unless the area is demonstrated to be non-
hazardous at the time.
* Enclosure shall not be opened unless the area is known to be nonhazardous, or
unless all devices within have been de-energized.
* Power shall not be restored after enclosure has been opened until enclosure has
been purged for 12 minutes. (When the internal pressure is restored, the system
automatically purges over the 12 minutes, then turns on the power again.)
For explosionproof enclosure:
* Seal all conduits within 18 inches
* Open circuit before removing cover.
For type Y purging:
* Enclosure shall not be opened unless the area is known to be non-hazardous, or
unless all devices within have been de-energized. Power shall not be restored after
enclosure has been opened until enclosure has been purged for 12 minutes at
specified pressure indicated by the pressure gauge labeled “EL.BOX” in the
pressure and flow control section.
* Alarm shall be provided and connected to alarm contact output.
(a) The alarm shall generate a visual or audible signal that attracts attention
(b) The alarm shall be located at constantly attended location.
(c) Electrical alarms shall be approved for the location in which they are installed.
<CSA>
For type X purging:
* This equipment contains components that operate at high temperature. The
equipment shall be deenergized for 60 minutes to permit those components to
cool before the enclosure is opened unless the area is demonstrated to be non-
hazardous at the time.
* Enclosure shall not be opened unless the area is known to be non-hazardous, or
unless all devices within the enclosure have been de-energized. Power must not
be restored after enclosure has been opened until enclosure has been purged for
12 minutes at a flow rate of 0.05m3/min.
* Power will automatically be removed when purge pressure falls below 40 mm (1.6
in) of water column.
Take care not to generate mechanical spark when accessing to the instrument and
peripheral devices in hazardous locations.
Do not press prick the keyboard of LCD panel (operation and display section) using
such as knives and sticks.
1MΩ
F0001.EPS
(c) When servicing cards on the bench, place them on a conductive sheet
grounded via a 1 MΩ resistance, wearing a wrist strap as in (2) above. Keep
easily-chargeable plastic materials away from the bench.
(d) Never touch components mounted on the cards, the pattern side, connec-
tors, pin components, etc. with bare hands, unless using a wrist strap and a
conductive sheet.
(e) Wrist straps and conductive sheets are available from Yokogawa Engineering
Service (YSV).
1. Overview
A gas c hromatograph separates a gas mixtures into eac h component and seriall y
detects the components. The gas c hromatograph is a representative gas anal yzer
owing to its wide measurab le area, no interf erence , and its capability to measure the
various components at the same time .
This anal yzer can be installed up to tw o detector s. The combinations of these detector s
can be: a thermal conductivity detector (TCD), a flame ionization detector (FID) or a
flame photometric detector (FPD).
Since the method of handling the anal yzer diff ers slightl y accor ding to the detector s
installed, refer to the Instruction Man ual in accor dance with the method to be adopted.
Gas chromatograph system v aries accor ding to user's conditions. Therefore, please
note that this installing and operating the anal yzer.
Class D independent
grounding (100Ω or less)
Communlcation via personal computer
Transmission : Analysis results, Operation info., Operational conditions Communication via DCS RS-422 1200/2400/4800/
(RS-422, 19.2/38.4k bps ; for explosion protection,
Transmission : Analysis results 9600/19200bps for
an RS-422/RS-232C converter is provided) explosion protection,
Reception : Operation requests, Operational conditions change an RS-422/RS-232C
Reception : Operation requests
converter is provided.
1/2B SGPW
vent stack
Keep pipe support so as Rc1/2 or
not to bend by the weight 1/2NPT
of pipe.
Dehumidifier
Vent Header
Both ends are RC 1/2
FID/FPD air
∗
Process1
∗ ∗1 ∗2 ∗ ∗ ∗ ∗ ∗ ∗
∗
Liquid sample 2
•
(up to two streams) •
∗ 1/2B SGPW • Instrument air
Drain pipe
∗
•
• ∗1. 350 to 900kPa
Up to 31 streams ∗
2. 400 to 700kPa
FID hydrogen
FID hydrogen
Carrier gas 1
Carrier gas 1
Carrier gas 2
Carrier gas 2
(including standard sample)
∗ Drain tank
2B SGPW Analyzer base
sampling system
(Note 1)
Drain tank is used only for using FID/FPD. (spare) (spare) (spare)
This is not used for using TCD
Note1: As an analyzer base sampling system is provided, in many cases, most applications require no external sampling
equipment.
In addition, optimum sampling systems are prepared depending on various conditions. (For details, consult
Yokogawa. Optimal sampling systems will be offered.)
Note2: For piping air purging, use stainles steel pipe of 1/2 inch or more. For another piping, see subsection 1.1.2,
"External Dimensions".
Wiring cables, piping and installation materials marked with an * should be supplied by the user.
Note3: For I/O cables, see Subsection 2.2.3, “Recommended Cables”.
Note4: Fix venting pipes properly so that the dead load of the venting pipes does not apply to the assembling vents of
this analyzer.
Note5: The specification decides the number of flameproof enclosure. No enclosure is applied for the general purpose or
Y-purge type.
Note6: There is a case of analog output by via GCCU. GCCU is a computing unit to handle various types of signal such
as A/O, analog serial output, trend output, digital output and so on. Refer to GS 11B03S03-01E for detail.
F01.EPS
Class D independent
grounding (100Ω or less)
Communlcation via personal computer
Transmission : Analysis results, Operation info., Operational conditions Communication via DCS RS-422 1200/2400/4800/
(RS-422, 19.2/38.4k bps ; for explosion protection,
Transmission : Analysis results 9600/19200bps for
an RS-422/RS-232C converter is provided) explosion protection,
Reception : Operation requests, Operational conditions change an RS-422/RS-232C
Reception : Operation requests
converter is provided.
Dehumidifier
Gas sampling vent
1/8 inch sus pipes
Used for only gas sampling FID/FPD air
∗
Process1
∗ ∗1 ∗2 ∗ ∗ ∗ ∗ ∗ ∗
∗
2
Liquid sample •
•
(up to two streams) • Instrument air
∗
•
• ∗1. 350 to 900kPa
Up to 31 streams ∗
2. 400 to 700kPa
FID hydrogen
FID hydrogen
Carrier gas 1
Carrier gas 1
Carrier gas 2
Carrier gas 2
1B SGPW or (including standard sample)
1/2B SGPW ∗
Vent stack
∗ Header Analyzer base
sampling system
(Note 1)
2B SGPW (spare) (spare) (spare)
Note1: As an analyzer base sampling system is provided, in many cases, most applications require no external sampling
equipment.
In addition, optimum sampling systems are prepared depending on various conditions. (For details, consult
Yokogawa. Optimal sampling systems will be offered.)
Note2: For piping air purging, use stainles steel pipe of 1/2 inch or more. For another piping, see subsection 1.1.2,
"External Dimensions".
Wiring cables, piping and installation materials with an * should be supplied by the user.
Note3: For I/O cables, see subsection 2.2.3, “Recommended Cables”.
Note4: The specification decides the number of flameproof enclosure. No enclosure is applied for the general purpose or
Y-purge type.
Note5: There is a case of analog output by via GCCU. GCCU is a computing unit to handle various types of signal such
as A/O, analog serial output, trend output, digital output and so on. Refer to GS 11B03S03-01E for detail.
F02.EPS
Class D independent
grounding (100Ω or less)
Communlcation via personal computer
Transmission : Analysis results, Operation info., Operational conditions Communication via DCS RS-422 1200/2400/4800/
(RS-422, 19.2/38.4k bps ; for explosion protection, Transmission : Analysis results 9600/19200bps for
an RS-422/RS-232C converter is provided) explosion protection,
Reception : Operation requests, Operational conditions change Reception : Operation requests an RS-422/RS-232C
converter is provided.
1/2B SGPW
vent stack
Keep pipe support so as Rc1/2 or
not to bend by the weight 1/2NPT Sample inlet
of pipe. Sample vent
Dehumidifier
Vent Header
Both ends are RC 1/2
FID/FPD air
∗
Liquid sample ∗1 ∗2 ∗ ∗ ∗ ∗ ∗ ∗
(up to two streams)
∗ 1/2B SGPW Instrument air
Drain pipe ∗1. 350 to 900kPa
∗2. 400 to 700kPa
FID hydrogen
FID hydrogen
Carrier gas 1
Carrier gas 1
Carrier gas 2
Carrier gas 2
∗ Drain tank
2B SGPW
Drain tank is used only for using FID/FPD. (spare) (spare) (spare)
This is not used for using TCD
Note1: For piping air purging, use stainles steel pipe of 1/2 inch or more. For another piping, see subsection 1.1.2,
"External Dimensions".
Wiring cables, piping and installation materials marked with an * should be supplied by the user.
Note2: For I/O cables, see Subsection 2.2.3, “Recommended Cables”.
Note3: Fix venting pipes properly so that the dead load of the venting pipes does not apply to the assembling vents of
this analyzer.
Note4: The specification decides the number of flameproof enclosure. No enclosure is applied for the general purpose or
Y-purge type.
Note5: There is a case of analog output by via GCCU. GCCU is a computing unit to handle various types of signal such
as A/O, analog serial output, trend output, digital output and so on. Refer to GS 11B03S03-01E for detail.
Class D independent
grounding (100Ω or less)
Communlcation via personal computer
Transmission : Analysis results, Operation info., Operational conditions Communication via DCS RS-422 1200/2400/4800/
(RS-422, 19.2/38.4k bps ; for explosion protection,
Transmission : Analysis results 9600/19200bps for
an RS-422/RS-232C converter is provided) explosion protection,
Reception : Operation requests, Operational conditions change an RS-422/RS-232C
Reception : Operation requests
converter is provided.
Sample inlet
Sample vent
Dehumidifier
FID/FPD air
∗ ∗1 ∗2 ∗ ∗ ∗ ∗ ∗ ∗
Liquid sample
(up to two streams) Instrument air
∗1. 350 to 900kPa
∗2. 400 to 700kPa
FID hydrogen
FID hydrogen
Carrier gas 1
Carrier gas 1
Carrier gas 2
Carrier gas 2
1B SGPW or
1/2B SGPW ∗
∗ Header
Vent stack
Note1: For piping air purging, use stainles steel pipe of 1/2 inch or more. For another piping, see subsection 1.1.2,
"External Dimensions".
Wiring cables, piping and installation materials with an * should be supplied by the user.
Note2: For I/O cables, see subsection 2.2.3, “Recommended Cables”.
Note3: The specification decides the number of flameproof enclosure. No enclosure is applied for the general purpose or
Y-purge type.
Note4: There is a case of analog output by via GCCU. GCCU is a computing unit to handle various types of signal such
as A/O, analog serial output, trend output, digital output and so on. Refer to GS 11B03S03-01E for detail.
Connection to
L
L
electronics section
Packing 100
Sealing fitting 130
ground wiring
1.6mm2 or more
Protection system (100Ω max.)
1212
Liquid sampling valve 1012
Pressure
with vaporizer
Flow controller Standard gases, carrier gas,
and air for FID/FPD
6mm or 1/4"tube
(1642)
RC1/4 or 1/4NPT(F)
Steam Inlet (Note 1)
RC1/4 or 1/4NPT(F)
Steam drain Outlet (Note 1)
RC1/4 or 1/4NPT(F)
450
305
245
M12 screw
25±5
1000
F04.EPS
L
Packing 100
Sealing fitting 130
ground wiring
1.6mm2 or more
(100Ω max.)
Operation and
display section Electronics section
1212
with vaporizer
1012
Prescore Inlet/Outlet of sample gas,
Flow controller standard gases, carrier gas,
550
and air for FID/FPD
6mm or 1/4"tube
Isothermal oven,
programmed-temperature
oven
50
Air for valve driving and for
50
460
300 (616) 400
1000
(1316)
F06.EPS
Connection to
electronics section
ground wiring
1.6mm2 or more
Protection system (100Ω max.)
1212
1012
Liquid sampling valve
Pressure
with vaporizer
Flow control section Standard gases, carrier gas,
and air for FID/FPD
6mm or 1/4"tube
(1642)
RC1/4 or 1/4NPT(F)
Steam Inlet (Note 1)
RC1/4 or 1/4NPT(F)
Steam drain Outlet (Note 1)
RC1/4 or 1/4NPT(F)
450
305
245
M12 screw
1000
25±5
F08.EPS
ground wiring
1.6mm2 or more
(100Ω max.)
Operation and
display section Electronics section
1212
with vaporizer
1012
Prescore Inlet/Outlet of sample gas,
Flow control section standard gases, carrier gas,
550
and air for FID/FPD
6mm or 1/4"tube
Isothermal oven,
programmed-temperature
oven
50
Air for valve driving and for
50
460
300 (616) 400
20
1000
• Installation height :
(Space for analyzer maintenance)
For the best installation, provide ample
clearance on the top of the wiring fitting
to meet the wire-lead requirements.
(1316)
F10.EPS
616 355
32±5
Connection to
electronics section
ground wiring
1.6mm2 or more
Protection system (100Ω max.)
1212
1012
Liquid sampling valve
Pressure
with vaporizer
Flow control section Standard gases, carrier gas,
and air for FID/FPD
6mm or 1/4"tube
(1642)
RC1/4 or 1/4NPT(F)
Steam Inlet (Note 1)
RC1/4 or 1/4NPT(F)
Steam drain Outlet (Note 1)
RC1/4 or 1/4NPT(F)
450
305
M12 screw
1000
25±5
F12.EPS
616 355
Connection to
32±5
electronics section
Protection system
ground wiring
1.6mm2 or more
(100Ω max.)
Operation and
display section Electronics section
1212
with vaporizer
1012
Prescore Inlet/Outlet of sample gas,
Flow control section standard gases, carrier gas,
550
and air for FID/FPD
6mm or 1/4"tube
Isothermal oven,
programmed-temperature
oven
50
Air for valve driving and for
50
460
300 (616) 400
20
1000
• Installation height :
(Space for analyzer maintenance)
For the best installation, provide ample
clearance on the top of the wiring fitting
to meet the wire-lead requirements.
(1316)
F14.EPS
27
AIR OUT 1 23
AIR OUT 2
AIR OUT 3
AIR OUT 4
AIR OUT 5
AIR OUT 6
AIR OUT 7
(A.T.M. BALANCE)
AIR OUT 8
(A.T.M. BALANCE) 322
SAMPLE 1 IN
SAMPLE 1 VENT
SAMPLE 2 IN
SAMPLE 2 VENT
CARRIER 1
CARRIER 2
BURNER FUEL
BURNER AIR
40
PURGE AIR
Rc1/4 or NPT1/4
(4) Dehumidifier
Any moisture in the carrier gas attac k the columns. Theref ore, in the case of the carrier
gas contains moisture abo ve the de w point of – 60 ºC, it is recommended to use a
desiccant, suc h as a molecular sie ve, to pre vent deterioration of the columns.
(8) Pipings
The pipes f or the sample inlet, carrier gas inlet, and standar d gas inlet ; the air f or v alve
driving and air f or FID / FPD ; the FID / FPD h ydr ogen ; steam ; the sampling b ypass ;
and the diver se vents are pr ovided.
(9) Joints
The joints are used to connect pipes.
(7) Steam
This is required when a sample is steam - heated. Prepare a steam supply at the
pressure listed in “Operation Data. ”
(2) Emission
Compl ying Standar d; EN55011-Group 1 Class A (f or GC1000W/C)
Test Item Frequency Range Basic Standard
Mains terminal disturbance voltage 0.15 MHz - 30 MHz
EN55011-Group 1 Class A
Electromagnetic radiation disturbance 30 MHz - 1 GHz
EEx pd2B+H2 : Type of protection and temperature class. Maximum leakage flow rate
0.1 m3/min 0.1 m3/min
T4 or T3 or T2 or 1 from pressurized enclosure
2.1 Installation
2.1.1 Analyzer Installation
There are tw o types : Floor mounting and wall mounting. Install the c hromatograph
accor ding to the pr ocedure determined f or eac h type .
(3) Unpacking
The Model GC1000 Mark II weighs about 120 kg. Unpac k it near the installation site .
Use a transpor tation mac hine to mo ve it. It is at the position where the center of gra vity
of GC1000 Mark II is high. Handle it carefull y so that it does not fall.
CAUTION
Take great care when carr ying and installing (piping, wiring) the GC1000 Mark II. The
GC1000 Mark II m ust be carried and installed ver y carefull y (inc luding piping and
wiring) b y more than one per son (at least f our people are recommended).
(4) Installation
To securel y install the anal yzer on the floor , use anc hor bolts.
To securel y install the anal yzer on a wall, use n uts and bolts.
Note 1: Drain tank is used only for using FID/FPD. Piping after removing the handle for transmission of left side on analyzer.
And piping incline so as not to stop drain in pipe. . .
Note 2: Piping and installation materials marked with an * should be supplied by the user.
Note 3: Provide the top end of vent stack to rain - protected construction.
Figure 2.4 Example of Vent Stack Construction
2.2 Piping
2.2.1 General Notes on Piping Work
(1) Do not remo ve the b lind plugs, provided at the anal yzer piping connections to
prevent deterioration of the columns, until immediatel y before star ting piping w ork.
(2) Use an anti - corr osion material f or the pipes and pipe fittings.
(3) Avoid using pipes with too lar ge a diameter f or the sample inlet piping, for the
purpose of reducing dead time . However, use a little lar ger pipe f or the e xhaust line
so as not to appl y back pressure to the venting lines.
(4) Use pipes and pipe fittings free of interior contamination fr om grease , oil or other
substances. Before connecting those pipes, completel y air - pur ge their interior s.
(5) Carefully connect the pipes so that there is no leakage from the pipe connections
suc h as the joints.
(6) Use filter s or other appr opriate pipe fittings to pre vent dust, moisture , or other
foreign matter fr om g etting inside the anal yzer.
To GC1000 Mark II
(Spare)
Figure 2.5
• When this piping is to be implemented, completel y check that the insides of the
pipes and pipe fittings are not contaminated.
• Do the piping so that there is no leaka ge. Specificall y when h ydr ogen gas is used,
make sure there is no leaka ge because that is also required f or explosionpr oof .
CAUTION
Do not use solvents containing impurities suc h as non - v olatile components to c lean
the inside of the pipes. Otherwise , this will contaminate the inside of the pipes disallo w-
ing correct anal ysis. (To correct suc h a condition, replacement of the par ts in the ana-
lyzer may be required.) If it is necessar y to use a solvent f or c leaning, use highl y pure
acetone .
2.3 Wiring
2.3.1 General Cautions on Wiring
(1) Cautions on wiring for protection system
• There ma y be 0 to 2 enc losures f or Protection System depending on customer's
specifications.
a. Specifications f or no enc losure :
• Model GC1000/E / GC100T Y-pur ging
b. Specifications f or one enc losure :
• Model GC1000S / GC1000E / GC1000C
• No analog input
c. Specifications f or tw o enc losures :
Models GC1000 Mark II without the abo ve specifications
• If JIS explosionpr oof wiring connections are required, always use the sealing fittings
or pac king adapter supplied.
<JIS>
In case when e xplosion-pr oof application, the follo wing wiring pr ocedure m ust be taken.
<CENELEC>
WARNING
• All wiring shall compl y with local installation requirements and local electrical code .
• The cab le entr y devises shall be of a cer tified flamepr oof type , suitab le for the
conditions of use .
Electrical connection
The type of electrical connection is stamped near the electrical connection por t accor d-
ing to f ollo wing codes.
(Code)
• 3/4NPT female :N
• PG21 female :D
• G3/4 female :G
• ISO M25 x 1.5 feemale : none (Code is not stamped.)
<FM>
• All wiring shall compl y with National Electric Code ANSI/NFP A 70 and Local
Electric Codes.
• In hazar dous location, wiring f or explosionpr oof enc losure and type x pur ged
enclosure to be in conduit as sho wn in the Figure 2.7.
• In hazardous location, wiring f or type Y pur ged enc losure to be in conduit as
sho wn in Figure 2.8.
Nonhazardous Nonhazardous
Location Location
Equipment Equipment
Nonhazardous Locations
Hazardous Locations
Sealing Fitting
18 inches
(457mm) maximum
Conduit
GC1000 as close as possible not
Explosionproof more than 18 inches(457mm)
Enclosure
GC1000
Type X Purging
Figure 2.7
Nonhazardous
Location
Equipment
Nonhazardous Locations
Hazardous Locations
Sealing Fitting
Conduit
as close as possible not
more than 18 inches(457mm)
GC1000
Type Y Purging
Figure 2.8
Nonhazardous Nonhazardous
Location Location
Equipment Equipment
Nonhazardous Locations
Hazardous Locations
Sealing Fitting
50cm
maximum
Conduit
GC1000 as close as possible
Explosionproof not more than 50 cm
Enclosure
GC1000
Type X Purging
Figure 2.9
Nonhazardous
Location
Equipment
Nonhazardous Locations
Hazardous Locations
Sealing Fitting
GC1000
Type Y Purging
Figure 2.10
1. Analyzer with two enclosures of Protection system 2. Analyzer with one enclosure of Protection system
Figure 2.11
CAUTION
• Wire the po wer suppl y cab le keeping the distance of 1 cm or more fr om other
signal wires.
• Power suppl y cab le must conf orm to UL or CSA.
• Do wiring after y ou do pr otective gr ounding securel y.
*1: Solderless lugs vary in outside diameter depending on the type of the equipment for which
the power cable is used.
Note
AMP: Japan AMP Co., Ltd.
JST: JST Co., Ltd. (Insulator s 0.8 mm 2 or more in siz e are optionall y availab le.)
WARNING
<CENELEC>
The cab le entr y devices and the b lanking elements shall be of a cer tified
flamepr oof type , suitab le for the conditions of use and correctl y installed.
For external ear thing of bonding connection a cab le lug shall be used so that the
conductor is secured against loosening and twisting and that contact pressure is
permanentl y maintained.
Figure 2.13
Figure 2.14
It branc hes off with terminal [Heater po wer suppl y line and Electric cir cuit po wer line].
Wire for branc h is installed in the terminal at the time of shipping. Be sure to tighten it
when y ou wire it f or the po wer suppl y line . Remove wire f or branc h, and wire it when
Heater po wer suppl y line and Electric po wer suppl y line are made another po wer
supply system.
Figure 2.15
Figure 2.16
Figure 2.17
It branc hes off with terminal [Heater po wer suppl y line and Electr onic cir cuit po wer line].
Wire for branc h is installed in the terminal at the time of shipping. Be sure to tighten it
when y ou wire it f or the po wer suppl y line . Remove wire f or branc h, and wire it when
Heater po wer suppl y line and Electr onic cir cuit po wer suppl y line are made another
power suppl y system.
Figure 2.18
L N G L N G
Electric Heater
Circuit Power Power
Figure 2.19
L N G L N G
Electric Heater
Circuit Power Power
Figure 2.20
Earth terminal
Rear Panel
Ground
Figure 2.21
CAUTION
• For pr otective gr ounding of terminal connector -type de vices, connect to the
terminal.
• For gr ounding of de vices with function gr ounding terminals, connect to the
terminal.
CAUTION
• Always use solderless lugs with insulated co vering.
• Always use solderless lugs and crimp-on tool man ufactured b y the same man ufac-
turer .
• The crimp-on tool m ust be matc hed to the wire thic kness.
Note
AMP: Japan AMP Co., Ltd.
JST: JST Co., Ltd. (Insulator s 0.8 mm 2 or more in siz e are optionall y availab le.)
CAUTION
• All de vices, suc h as the signal line , relay terminal, relay, power suppl y, etc. to be
connected to contact input and contact output m ust conf orm to IEC1010 or CSA 1010
or CSA 950.
• Do wiring after y ou do pr otective gr ounding securel y.
Figure 2.23
Figure 2.24
Figure 2.25
Figure 2.26
Figure 2.27
+ 29 1 +
D/I 3 + 15
- 30 D/I 2
- 16
2 - D/I 1
+ 31 3 +
D/I 6 + 17
- 32 D/I 5
- 18
4 - D/I 4
33 5 +
+ 19
34 D/I 8
- 20
6 - D/I 7
RDA 35 7 RDA
RDA 21
RS-422 RDB 36 8 RDB RS-422
(DCS) SDA RS-422 RDB 22
37 (GCCU) 9 SDA (PC)
SDA 23
SDB 38 10 SDB
SDB 24
39 11
25
40 12
26
P.MONITOR + 41 13 + P.MONITOR
P.MONITOR + 27
(DC12V) - 42 (DC12V) -28
14 - (DC12V)
Figure 2.28
+ 53 1 +
D/I 3 + 27
- 54 D/I 2
- 28
2 - D/I 1
+ 55 3 +
D/I 6 + 29
- 56 D/I 5
- 30
4 - D/I 4
57 5 +
+ 31
58 D/I 8
- 32
6 - D/I 7
RDA 59 7 RDA
RDA 33
RS-422 RDB 60 8 RDB RS-422
(DCS) SDA RS-422 RDB 34
61 (GCCU) 9 SDA (PC)
SDA 35
SDB 62 10 SDB
SDB 36
63 11
37
64 12
38
NC65 13 NC
NC 39
D/O 3 COM 66 14 COM D/O 1
D/O 2 COM 40
NO 67 15 NO
NO 31
NC 68 16 NC
NC 42
D/O 6 COM 69 17 COM D/O 4
D/O 5 COM 43
NO 70 18 NO
NO 44
NC 71 19 NC
ANNUN- NC 45
CIATOR COM 72 D/O 8 COM 46
20 COM D/O 7
NO 73 21 NO
NO 47
74 22
48
+ 75 23 +
A/I 3 + 49 A/I 1
- 76 A/I 2
- 50
24 -
77 25 +
51 A/I 4
78
52
26 -
Figure 2.29
Terminal
Number
29 Analog input 1 +
30 Analog input 1 -
31 Analog input 2 +
32 Analog input 2 -
33 Analog input 3 +
34 Analog input 3 -
35 Analog input 4 +
36 Analog input 4 -
Figure 2.32
Terminal
Number Terminal Number
+ Analog input 3 75 + Analog input 2 49 23 Analog input 1 +
- Analog input 3 76 - Analog input 2 50 24 Analog input 1 -
25 Analog input 4 +
26 Analog input 4 -
Figure 2.33
Terminal
Number Terminal Number
+ Contact input 3 (53) 29 + Contact input 2 (27) 15 1 (1) Contact input 1 +
- Contact input 3 (54) 30 - Contact input 2 (28) 16 2 (2) Contact input 1 -
+ Contact input 6 (55) 31 + Contact input 5 (29) 17 3 (3) Contact input 4 +
- Contact input 6 (56) 32 - Contact input 5 (30) 18 4 (4) Contact input 4 -
+ Contact input 8 (31) 19 5 (5) Contact input 7 +
- Contact input 8 (32) 20 6 (6) Contact input 7 -
Terminal
Number Terminal Number
Send data A (GCCU) (59) 35 Send data A (DCS) (33) 21 7 (7) Send data A (personal computer)
Send data B (GCCU) (60) 36 Send data B (DCS) (34) 22 8 (8) Send data B (personal computer)
Receive data A (GCCU) (61) 37 Receive data A (DCS) (35) 23 9 (9) Receive data A (personal computer)
Receive data B (GCCU) (62) 38 Receive data B (DCS) (36) 24 10(10) Receive data B (personal computer)
Power off signal +(GCCU) 41 Power off signal + (DCS) 27 13 Power off signal + (personal computer)
Power off signal - (GCCU) 42 Power off signal - (DCS) 28 14 Power off signal - (personal computer)
Terminal Number
Analog output 19 + + + Analog output 1 +
19 1 -
Analog output 19 - - Analog output 1 -
Analog output 20 + + + Analog output 2 +
20 2 -
Analog output 20 - - Analog output 2 -
Analog output 21 + + + Analog output 3 +
21 3 -
Analog output 21 - - Analog output 3 -
Analog output 22 + + + Analog output 4 +
22 4 -
Analog output 22 - - Analog output 4 -
Analog output 23 + + + Analog output 5 +
23 5 -
Analog output 23 - - Analog output 5 -
Analog output 24 + + + Analog output 6 +
24 6 -
Analog output 24 - - Analog output 6 -
Analog output 25 + + + Analog output 7 +
25 7 -
Analog output 25 - - Analog output 7 -
Analog output 26 + + + Analog output 8 +
26 8 -
Analog output 26 - - Analog output 8 -
Analog output 27 + + + Analog output 9 +
27 9 -
Analog output 27 - - Analog output 9 -
Analog output 28 + + + Analog output 10 +
28 10 -
Analog output 28 - - Analog output 10 -
Analog output 29 + + + Analog output 11 +
29 11 -
Analog output 29 - - Analog output 11 -
Analog output 30 + + + Analog output 12 +
30 12 -
Analog output 30 - - Analog output 12 -
Analog output 31 + + + Analog output 13 +
31 13 -
Analog output 31 - - Analog output 13 -
Analog output 32 + + + Analog output 14 +
32 14 -
Analog output 32 - - Analog output 14 -
Analog output 33 + + + Analog output 15 +
33 15 -
Analog output 33 - - Analog output 15 -
Analog output 34 + + + Analog output 16 +
34 16 -
Analog output 34 - - Analog output 16 -
Analog output 35 + + + Analog output 17 +
35 17 -
Analog output 35 - - Analog output 17 -
Analog output 36 + + + Analog output 18 +
36 18 -
Analog output 36 - - Analog output 18 -
Figure 2.36
Terminal Number
- Analyzer bus A 1 5 Analyzer bus B -
+ Analyzer bus A 2 6 Analyzer bus B +
- Power off signal (A) 3 7 Power off signal (B) -
+ Power off signal (A) 4 8 Power off signal (B) +
Figure 2.37
䉬 Revision Record
● Manual No. : TI 11B03A03-03E
● Title : Process Gas Chromatograph GC1000 Mark II Installation Manual