Unit II NABL_2015
Unit II NABL_2015
Pravin Mali
2/27/2022 Sandeep Gadhwe 4
is an autonomous body
registered under Societies Act in1992
under the aegis of Department of
Science & Technology, Government of
India.
CERTIFICATION ACCREDITATION
2/27/2022 Sandeep Gadhwe 11
Customer
Laboratory
Others
Approval forAccreditation
(( by Chairman NABL)
Pravin Mali
and corrective actions
taken
Pravin Mali
Handling - receipt to dispatch-
Corrective and preventive Records.
actions Record. Quality Assurance Records.
Internal Audit Records Test reports/Calibration
Management Review Records. Certificates.
Retention Record for Documents
/Samples
Pravin Mali
When audit findings cast doubt on effectiveness of
operations or on the correctness or validity of lab. results,
laboratory shall take timely corrective action.
Pravin Mali
quality control activities, resources and staff training.
Findings from management reviews shall be recorded and
actions taken in agreed timescale
Pravin Mali
2/27/2022 Sandeep Gadhwe 40
Major Sources of
Uncertainties
I – Instrument ( 5.5)
1 – Instrumental Error
Acceptance Norm.
W – Work piece ( 5.7 , 5.8) 2– Uncertainty of calibration P – Person (5.2) / Procedure
1 – Non-homogeneity of test 3 – Instrumental Resolution. (5.4)
material 4 – Repeatability. ( Type A) 1 – Repeatability ( Type A)
2 – Instrumental Resolution (Skill of operator for positioning
(When instrument is under and measuring pressure etc.)
calibration) SWIPE
Pravin Mali
E – Environmental Factors (5.3)
S – Standard ( 5.5 & 5.6 ) Temperature control limits
1 – Un-compensated Dev. from Temperature variation.
nominal value.
Humidity.
2 – Uncertainty of calibration.
Cleanliness ( TypeA ).
3 – Non-homogeneity ( Type A).
Vibration ( Type A).
Pravin Mali
effectiveness of the training actions taken shall be
evaluated.
Job descriptions are required to be maintained for all
laboratory persons.
A formal system of authorization of personnel is required
for specific tasks such as performing tests, issue of
reports, giving opinions & interpretations
2/27/2022 Sandeep Gadhwe 42
The laboratory needs to ensure that facilities and
environmental conditions do not adversely affect tests or
calibrations.
Particular care is necessary for sampling and testing /
calibration at off-site locations
Some tests / calibrations require specified environmental
conditions. The laboratory must monitor that these
conditions are met and not proceed if the conditions are
Pravin Mali
not met.
In some types of tests, effective separation of Test /
calibration area may be required
Access to areas affecting quality should be controlled.
The laboratory should install measures for good house
keeping
2/27/2022 Sandeep Gadhwe 43
The laboratory should normally carry out tests or calibrations
using ‘Standard methods’. These are methods published in
national / International standards, reputable technical
organizations, or in relevant scientific texts or journals, or as
specified by the manufacturer of the equipment
Methods cover sampling, handling, transport, storage and
preparation of items to be tested and/or calibrated, and, where
appropriate, an estimation of the measurement uncertainty
Laboratory can also use self developed methods for its own use
Non-standard method, can be used for performing tests /
Pravin Mali
calibrations for customers but these need to be approved by him
Laboratory developed methods and non standard methods require
validation
Validation can be carried out through calibration with reference
materials, inter-lab comparisons, systematic evaluation of
influencing factors
Pravin Mali
The Equipment History record must include
Identification
Manufacturer’s name, type identification, sr. No. Etc.
Dates, results and copies of report and certificates of calibration, adjustment,
Acceptance criteria, and due date of next calibration.
Maintenance plan , date of maintenance.
Details of any damage , malfunctioning or repairs
Pravin Mali
calibration on previous results
Identification of the Calibration status through labeling of
equipment
Intermediate checks to maintain confidence in the
calibration status.
Safeguarding from adjustments which would invalidate the
tests and/or calibration results.
2/27/2022 Sandeep Gadhwe 46
All equipment used for tests / calibrations, having an effect on
validity of results shall be calibrated.
Laboratory should have established program and procedures for
calibration Program for calibration shall be designed to ensure
that measurements are traceable to the international system of
units (SI) by reference to national measurement standards
Calibration services from laboratories that can demonstrate
competence, measurement capability and traceability.
When traceability is not possible, other methods permitted:
Use of certified reference material provided by competent supplier &
Pravin Mali
specified methods.
Use of specified methods and/or consensus standards
Reference standards shall be calibrated, made traceable to SI
units,
Reference material should be traceable to SI units or certified
reference material
Intermediate checks for maintaining confidence in calibration
status of reference
Pravin Mali
2/27/2022 Sandeep Gadhwe 48
The laboratory should have a procedure for transportation
/ handling / storage / disposal of test / calibration items
In case any abnormalities / departures from normal
functioning or normal conditions are observed these shall
be recorded.
There should be a Procedure for avoiding deterioration,
loss or damage to Test / calibration item while it is in the
custody of the laboratory
Pravin Mali
2/27/2022 Sandeep Gadhwe 49
The laboratory must have an internal quality control
procedure for monitoring the validity of calibrations
undertaken.
Results data shall be recorded in such a way that trends
are detectable: where practicable statistical techniques
shall be applied.
Monitoring methods used:
Internal quality control using secondary reference materials
Participation in inter-laboratory comparison or proficiency testing
Pravin Mali
programme
Replicate testing
Retesting of returned items
Correlation of item using different characteristics of the item
Pravin Mali
Unique item identification, description and condition
Dates of item receipt, test or calibration conducted
Sampling plan and procedures used,
The test or calibration results
Identification of person(s) authorizing the test report or calibration
certificate
Details of environmental conditions during test / calibration
Uncertainty in case of calibration reports
2/27/2022 Sandeep Gadhwe 51
Opinions and interpretations can be given but shall
document the basis up on which the opinions and
interpretations have made.
Results of test performed by subcontractor shall be clearly
identified.
Electronic transmission of results can be done but
requirements of this standard shall be met
Pravin Mali
The format for report and certificate shall be designed to
accommodate each type of test and calibration
Amendment to the test and calibration certificate
Can be issued as a supplement to test / calibration report
Complete issue shall be uniquely identified and shall contain reference
to the original that it replaces.
Pravin Mali
Review of customer requests, tenders & Contracts.
Pravin Mali
Evaluate and ensure staff competency
Design and implementation of the QC program
Stopping work when QC criteria are violated
Attention to matters relating to impartiality and
confidentiality.
Ensure technical validity of the results
Liaison with the quality manager in matters of common
concern, affecting testing or calibration.
2/27/2022 Sandeep Gadhwe 54
The testing laboratory is obligated to
fulfil the requirements of ISO/IEC 17025 and to have
approved authorized signatory who bears the full
responsibility for proper execution of the test or
calibration and correctness of the data in the test report/
Pravin Mali
Calibration Certificates .
Lab has to have at least one approved authorized signatory
for each specific discipline.
Authorized signatories should Carry out regular measures
to assure the quality of test results are maintained.
Pravin Mali
Laboratories
NABL 103-Specific Guidelines for Chemical Testing
Laboratories
NABL 114-Guidelines for Food Testing Laboratories
documents given by NABL and must posses the