Astm D2726 D2726M 21
Astm D2726 D2726M 21
for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: D2726/D2726M − 21
1
D2726/D2726M − 21
taken out of the water, blotted quickly with a damp cloth towel, 6.3 Drying Oven, capable of maintaining a temperature of
and weighed in air. The difference between the two masses is 110 6 5 °C [230 6 9 °F].
used to measure the mass of an equal volume of water at 25 °C
6.4 Thermometer—The thermometer shall be one of the
[77 °F].
following:
4.2 This test method provides guidance for determination of 6.4.1 A liquid-in-glass partial or total immersion thermom-
the oven-dry or thoroughly dry mass of the specimen. The bulk eter of suitable range with subdivisions and maximum scale
specific gravity is calculated from these masses. Then the error of 0.5 °C [1.0 °F] which conforms to the requirements of
density is obtained by multiplying the specific gravity of the Specification E1. Calibrate the thermometer in accordance with
specimen by the density of the water. one of the methods in Test Method E77 or verify its original
5. Significance and Use calibration at the ice point (Notes 4 and 5).
5.1 The results obtained from this test method can be used NOTE 4—Practice E563 provides instructions on the preparation and use
to determine the unit weight of compacted dense asphalt of an ice-point bath as a reference temperature.
mixtures and in conjunction with Test Method D3203/ NOTE 5—If the thermometer does not read 0.0 6 0.5 °C [32.0 6 1.0 °F]
at the ice point, then the thermometer should be recalibrated.
D3203M, to obtain percent air voids. These values in turn may
be used in determining the relative degree of compaction. 6.4.2 A liquid-in-glass partial or total immersion thermom-
eter of suitable range with subdivisions and maximum scale
5.2 Since specific gravity has no units, it must be converted
error of 0.5 °C [1.0 °F] which conforms to the requirements of
to density in order to do calculations that require units. This
Specification E2251. Calibrate the thermometer in accordance
conversion is made by multiplying the specific gravity at a
with one the methods in Test Method E77 or verify its original
given temperature by the density of water at the same tempera-
calibration at the ice point (Notes 4 and 5).
ture.
6.4.3 A platinum resistance thermometer (PRT) with a probe
NOTE 1—The quality of the results produced by this standard are which conforms to the requirements of Specification E1137/
dependent on the competence of the personnel performing the procedure
and the capability, calibration, and maintenance of the equipment used.
E1137M. The PRT shall have a three- or four-wire connection
configuration and the overall sheath length shall be at least
iTeh Standards
Agencies that meet the criteria of Specification D3666 are generally
considered capable of competent and objective testing, sampling,
inspection, etc. Users of this standard are cautioned that compliance with
50 mm [2 in.] greater than the immersion depth. Calibrate the
PRT system (probe and readout) in accordance with Test
(https://standards.iteh.ai)
Specification D3666 alone does not completely ensure reliable results.
Reliable results depend on many factors; following the suggestions of
Specification D3666 or some similar acceptable guideline provides a
Methods E644 or verify its original calibration at the ice point
(Notes 4 and 5). Corrections shall be applied to ensure accurate
Document Preview
means of evaluating and controlling some of those factors. measurements within 0.5 °C [1.0 °F].
6.4.4 A metal-sheathed thermistor with a sensor substan-
6. Apparatus tially similar in construction to the PRT probe described in
6.1 Balance, with ample capacity, and with sufficient sensi- 6.4.3. Calibrate the thermistor system (sensor and readout) in
tivity to enable bulk specific gravities of theASTM specimensD2726/D2726M-21
to be accordance with Test Methods E644 or verify its original
calculated/catalog/standards/astm/63cc7553-9421-45ba-9959-b145a233fce6/astm-d2726-d2726m-21
to at least four significant figures, that is, to at least calibration at the ice point (Notes 4 and 5). Corrections shall be
three decimal places. It shall be equipped with a suitable applied to ensure accurate measurements within 0.5 °C
apparatus to permit weighing the specimen while suspended in [1.0 °F].
water. To avoid erroneous readings by undue displacement of
water, use wire or fish line of the smallest practical size to 7. Standardization
suspend the specimen and holder. Do not use chains, strings, or
7.1 Balance—Standardize at least once each twelve months.
sash cords. The balance shall conform to Guide D4753 as a
Class GP2 balance. 7.2 Temperature-Measuring Devices—Standardize at least
NOTE 2—Since there are no more significant figures in the quotient
once every twelve months.
(bulk specific gravity) than appear in either the dividend (the mass of the 7.3 Drying Oven—Standardize thermometric devices at
specimen in air) or in the divisor (the mass of the volume of water equal
to the volume of the specimen, obtained from the difference in weight of
least once every twelve months.
the saturated surface-dry specimen in air and in water), this means that the
NOTE 6—The terms of standardization, verification, calibration, and
balance must have a sensitivity capable of providing both mass values to
check are defined in Specification D3666 Section 3.
at least four figures. For example, a sensitivity of 0.1 g [0.0002 lb] would
provide four significant figures for mass in the range from 100.1 to 999.9 g
[0.221 to 2.204 lb]. 8. Sampling
6.2 Water Bath, capable of maintaining a temperature of 25 8.1 Specimens may be either laboratory-molded asphalt
6 1 °C [77 6 1.8 °F] for immersing the specimen in water mixtures or from asphalt pavements.
while suspended, equipped with an overflow outlet for main-
taining a constant water level. The use of an overflow outlet is 8.2 Obtain field samples in accordance with Practice D979/
mandatory. D979M.
NOTE 3—The water bath does not need to be a sophisticated device. 8.3 Pavement specimens shall be taken from pavements
Any method that maintains 25 6 1 °C [77 6 1.8 °F] can be used including with a core drill, diamond or carborundum saw, or by other
tempering, aquarium heaters, stirrers, or other devices. suitable means, in accordance with Practice D5361/D5361M.