A310-CO-Monitor
A310-CO-Monitor
MODEL A310
I INTRODUCTION
This Model A310 Carbon Monoxide (CO) Monitor and Alarm is a 115 V AC dust-
resistant instrument that can be surface-mounted as close as possible to the air
supply but outside any classified hazardous area.
It continuously monitors a compressed air sample connected to its sample inlet and it
gives an alarm when:
. the CO in the sample exceeds a preset level (adjustable, initially set at 10 PPM).
It indicates CO concentration on a digital display, and has a pilot light to verify that
the instrument is on and operating properly.
The instrument is assembled into a powder-coated sheet metal housing 4" x 6" x 3½”
overall. A c” FPT opening for introducing the sample gas is located in the right hand
side of the flow block at the bottom of housing. Visible through windows in the front
face are the LCD display showing ppm CO, plus pilot and alarm lights. The 3-wire
power cord plugs into a RF-filtered socket on the upper right side, and the alarm
buzzer extends through the top.
When the hinged and gasketed front door is opened after releasing the latch on
the right hand side, the interior is visible and accessible. Each major component
is described in this section.
The electronic circuit board occupies the main part of the housing, toward the
front. It is secured by two slotted bars, and it can be removed if necessary by
taking out two screws holding the lower bar. Electrical connections are by cables
which connect to sockets on the rear. On the front face are these components of
interest to the user:
1. Display is a liquid crystal type which reads from 0 to 199 PPM of CO in steps
of one digit.
7. NORMAL light, green, shows that power is being provided to the circuit and
that the system is in normal operation.
8. ALARM light, red, comes on and glows steadily when CO is detected above
the alarm point level. It also indicates a sensor failure or disconnection by
blinking on and off.
In normal operation, a slight flicker can be seen in the red light, a little faster
than once per second, as a confirmation of normal operation.
B. Power Supply
AC power enters the housing through an EMI filter socket in the upper right hand
corner of case. The AC power cord has a standard grounding-type plug for a
normal 115 AC power outlet and the instrument end mates with the filter socket.
C. Flow System
The air to be analyzed passes through the instrument under its own pressure.
The components it flows through, in sequence are:
1. Sample inlet, c” FPT size, to accommodate any suitable tubing fitting having
a c” male pipe connection.
A3 10-011305-3
2. Flow block, a bored aluminum block with a cavity for receiving and sealing
the CO sensor, is secured to the outer bottom of the case. A porous metal
flow restrictor reduces the pressure of the compressed air to atmospheric
pressure before it reaches the sensor.
The CO sensor seals to the block by means of an O-ring seal against the
bottom of the cavity.
Sensor plugs into four sockets on the face of a small circuit board which
connects by three wires to the main circuit board.
This small circuit board is held down by two screws which provide pressure
against the O-ring seal to prevent leakage around the sensor.
E. External Connections
Two terminal blocks are installed on the lower crossbar. The one on the left has
three screw terminals, for connection to the internal alarm relay, and the one on
the right has two screw terminals for a 4-20 mA signal circuit. Wires to each of
these two external circuits can enter the housing through the ½" conduit opening
in the left hand side of the case. Connections are as follows.
1. The internal relay has 10 amp SPDT contacts, which are brought to the
marked terminals C, NO and NC.
2. The 4-20 mA signal circuit is brought to the two terminals marked + and -.
Output signal is 4.0 mA at 0 ppm, 20.0 mA at 100 ppm.
F. External Components
2. Flow block, described above, is secured to the outside bottom of the case,
and an opening in the case allows the sensor to be inserted from inside the
case.
A3 10-120905-4
3. Calibration gas inlet fitting, on left-hand end of flow block, is a hose barb with
internal orifice that restricts flow of calibration gas as it enters the block
during calibration. A second opening on the rear of the block provides for
exhaust of the calibration gas to prevent pressure buildup during calibration.
4. Alarm cutoff switch, a toggle switch on the right hand side, when turned off
prevents the internal buzzer from sounding. It has no effect on the external
alarm circuit.
III. OPERATION
A. Placing in operation
1. Install the instrument in the desired location, close to the air supply and to a
source of power, and where it will be reasonably protected but readily seen.
Mount the housing by means of #10 screws through the flanges top and
bottom.
3. Connect the air supply to be monitored, to the c” FPT opening in the outer
face of flow block, using appropriate tubing and fittings.
4. Using the flowmeter and tube supplied with the calibration kit, attach the inlet
end of tube to the calibration fitting on left side of flow block. Turn on air
pressure.
5. While holding a finger over the small hole in rear of flow block, observe
reading on flowmeter. It should read at least 0.3 while held in vertical
position. This confirms that sample flow system is working.
7. When connected, the green light should come on and the red light and
buzzer may also come on momentarily. The green light should flash for
about 15 seconds. Display should then show a reading of 00 or close to it.
9. Connect a source of nitrogen or CO-free air to calibration gas inlet, using the
flowmeter supplied with calibration kit, and set flow to 1.0.
10. If reading is other than 00, then turn the ZERO potentiometer to obtain a
zero reading; clockwise rotation increases reading.
11. Disconnect the air sample from calibration inlet, connect a known sample of
10 ppm of CO in air or nitrogen, and again set the flow to 1.0.
Note: The alarm buzzer can be silenced by pressing the silence button on
lower right hand side. This will lock the audible alarm out for 4 minutes.
12. Verify that the display reading increases and the alarm light and buzzer
operate.
13. If reading does not reach the desired level corresponding to the known gas
concentration, adjust it using the SPAN potentiometer. Clockwise rotation
will increase reading.
14. Remove the CO sample and be sure the air line sample is connected to the
sample inlet.
Note: The above steps are needed only for the first operation and
periodically during the life of the instrument.
B Normal Operation
1. Instrument will analyze the sample and show CO content on the display, in
parts per million (PPM). The green NORMAL light will glow continuously and
the red ALARM light will flicker about once a second.
A3 10-011305-6
MODEL A310 CO MONITOR - AC TYPE
2. When the CO concentration exceeds the alarm point (initially set at 10 PPM)
the red ALARM light will come on steady, the green NORMAL light will go out
and the buzzer will sound a steady tone. The internal alarm relay will
activate, and any connected external circuits will respond accordingly.
3. When the CO concentration drops below the alarm setting, the indicators will
automatically return to normal.
C Abnormal Indications
If there is a discontinuity in one of the detector leads, the indicator lights will show
as follows:
1. Working Electrode (WE) line (red wire) - - red and green light blink on and off,
accompanied by buzzer. Display goes to SC.
2. Counter Electrode (CE) line (black wire) - - red and green light blink on and
off, accompanied by buzzer. Display goes to SC.
3. Reference Electrode (RE) line (blue wire) - - CO reading goes off scale
positive (1----), red lamp comes on steady, green lamp off and buzzer
sounds continuously.
IV MAINTENANCE
A. Routine Maintenance
1. Each day or period of operation, verify reading close to 00. Adjust if needed.
B. Sensor Replacement
3. Loosen and remove the two screws extending up from the bottom of flow
block, into the threaded nuts of the sensor circuit board. This will allow the
sensor with board to be lifted up and out of the block. Unplug sensor from
board.
4. Install new sensor in the same position, after making sure the O-ring is in
place at the bottom of the cavity. Insert and tighten the screws, which will pull
the sensor down into contact with the O-ring seal.
6. After two hours of operation, recheck zero and span adjustments as explained
in III. A.
7. Discard old detector cell, keeping in mind that it contains a small amount of
sulfuric acid.
The principal electronic components are all installed on the main printed circuit
board, which is retained within the housing by two slotted bars. To remove the
board, take out the two screws holding the lower bar, after which the board can be
pulled out and the wires unplugged at their connectors. Keep track of the proper
position of each connector.
D. Buzzer
Once the circuit board is removed, the buzzer can easily be taken out by
unscrewing the locking ring at the top of case.
E. Power Supply
Once the circuit board is removed, this gives access to the power supply on its
small subpanel board. First take out the EMI filter block by removing two screws,
unplugging the white and black wires and disconnecting the green wire at the screw
post. Then take out the two screws holding the board to the case, pull out the
board and power supply assembly.
F. Flow Restrictor
If sample flow rate as tested under III. A.4 declines below 0.25, the flow restrictor
may have become partially plugged. To replace:
1. Remove sensor.
4. Remove two screws holding flow block to bottom of housing. Take off flow
block.
5. Either return the block to the factory for rework, or proceed as follows:
6. Take off calibration gas inlet fitting, and inlet air fitting.
7. Using a 2" long 10-32 screw, inserted into the threaded opening, gradually
tighten the screw until it bears on the flow element and pushes it out through
the inlet air opening. (If it fails to come out as a unit, it will be necessary to
send the block to the factory for rework).
8. Press a new flow restrictor into the same hole. It can be pushed into the
recess in the block, by pressing or by driving with a ¼” brass or aluminum rod
and a hammer.
V CALIBRATION KIT
A field calibration kit is offered for use in accurate adjustment of the instrument. It
consists of one compressed gas cylinder containing 100 ppm CO in air, and one
containing 100% nitrogen, all in a carrying case complete with flow adjustment valve,
tubing, a flowmeter and a small screwdriver.
2. Connect the tubing to the flow adjustment valve and to the flowmeter. Connect the
flowmeter outlet tube to the flow block hose barb.
3. Install flow adjustment valve on nitrogen cylinder. This will produce a flow which
can be set to 0.6 scfh using the valve.
4. Watch display, and after reading stabilizes set reading to 00 using ZERO
potentiometer.
6. Set flow to 0.6 and adjust SPAN to give reading corresponding to the value marked
on the cylinder, about 100 ppm.
The following is a list of items that may need replacement during the life of the
instrument.
06-1105 Tubing, ¼”
07-6122 O-ring
17-0282K Cal gas inlet
17-0285K Flow restrictor, porous metal
35-6005K Flow block
47-4120K Sensor cable with circuit board
52-1020 Buzzer, replacement
57-A310 Circuit board, main
65-2400T Cell, CO detector
81-1231K Calibration Kit, consisting of: