B&K 1022 Instruction Manual - Copy 2
B&K 1022 Instruction Manual - Copy 2
lffll
,.._a...,
+ .{. ;
•.
~
•
··J:t...!,!l-
f..:lf-·~
T-to~
=
~·.
•w
• ;·
m '
~~~¥~~v
~
!T
""""'"'""'...._
•"~'
,,.
~
~,- Beat Frequency Oscillator
•
r~ ,
.
•
-. :,.
Type 1022
A signal generator covering the
range 20-20,000 Hz and designed
especially for acoustic and electro-
acoustic measurements. Logarithmic
sweep and automatic drive for
recording of frequency response
curves. The output signal can be
frequency modulated and controlled
by a compressor loop.
THE BEAT FREQUENCY OSCILLATOR
TYPE 1022
1. INTRODUCTION • • • . • . • • . • • • . . • • • • • • • • • • • • • . • . . . . • . . • 5
2. CONTROLS . • • • • • • • • • • • • • . • . • . • • • • . . • . • . • . . . • • . . . . . . 6
4. OPERATION ••••••••••••••••••••••••••••.•.••.•.•.•• 15
5. APPLICATIONS ••••••••••••••••••••.•.•.•...•..•••.•• 25
6. ACCESSORIES •...••.•••..•••.••••.•.•.•..•.•...•••.• 43
7. SPECIFICATIONS ••••.•••..•••••......•...•••••••.••• 46
1. INTRODUCTION
The 1022 works on the heterodyne principle using two high frequency
oscillators one of which operates at a fixed frequency while the frequency
of the other can be varied. The required audio frequency is then obtained
by mixing these frequencies to produce a beat frequency.
Remote control facilities are provided to start and stop the scanning and
the oscillator and for instance to lift the pen on the Level Recorder when
the oscillator is sweeping outside the frequency range of interest.
5
2... CONTROLS
FRONT PANEL
6
AUTOMATIC SCANNING: Connects the variable capacitor of the
B. F.0. to a worm drive which can 'be
connected to and driven by an external
motor.
FREQUENCY SCALE
ALIGNMENT: Fine and Coarse potentiometers to ad-
just the output frequency to coincide
with that indicated on the main scale.
7
COMPRESSOR INPUT: For connection of the signal from the
regulating transducer when automatic
regulation of the B.F.O. output is re-
quired. Input impedance 25 kU Maxi-
mum range of regulation 50 dB.
MECHANICAL DRIVE
CONNECTION: Located on both sides of the instrument
are sockets for the connection of an ex-
ternal mechanical drive for automatic
frequency sweep. The shaft connection
fits the Flexible Drive Shaft UB 0041
which forms the mechanical link be-
tween the B.F.O. and the Level Recor-
der 2305.
8
REAR PANEL
9
3. TECHNICAL DESCRIPTION
Lood
Output
Eat . 120kHz
Mod. Output
Input
Attenuotor
Remote
Control Output
Input
The fixed oscillator is a tuned LC type and works at 120 kHz. Incorporat-
ed in the tuned circuit is a variable capacitor which is operated by the
Frequency Increment knob. This allows exact frequency selection within
±50 Hz about any setting on the main scale. In addition this capacitor per-
mits frequencies down to 10 Hz to be obtained at the risk of slightly more
distortion.
10
The reactance circuit is needed to control the actual frequency deviation
(modulation swing) of the fixed oscillator by producing an inductive react-
ance across the oscillator's tuned collector. The deviation can be varied
linearly from 0 to ± 250 Hz where the sensitivity is approximately
5 mV/Hz.
The speed with which the variation in output level is regulated back to
normal depends on the setting of the Compressor Speed knob, but also
depends to a certain extent on the amount of the deviation from the normal
level. Compressor speeds of 30 - 100 - 300 - 1000 dB/sec. are available
and are determined by the integration time constant of the rectifying cir-
cuit.
The variable oscillator is of similar design to the fixed one except that it
has a variable capacitor in its tuned circuit to vary the oscillation frequency
between 100 and 120 kHz. A worm gear, connected to the capacitor
11
dB
Output Level
10
Distorted
Signal
MAX. METER
DEFLECTION
-10+------
-20
-30 +-------...
-40
-50+--------.....
-60
-70
-80
-90
-20 -12 -6 -3
dB re 1V
--r--r--+--+--+--+-~--~_,--;-~r--.-v
0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 Comp.
spindle, permits automatic tuning with the aid of an external motor such as
that of the Level Recorder 2305. A magnetic clutch, which can be remotely
controlled, is used for connection of the external drive to the capacitor
spindle.
12
When the capacitor is set to frequencies above 20kHz or below 20Hz the
signal from the fixed oscillator is blocked so that no output voltage is
obtained. This has the advantage that when automatic recordings are being
taken with the Level Recorder 2305, no unwanted curves appear on the
paper. The cut-off section can be made wider by means of the adjustable
cams fixed on the capacitor spindle (Fig.4.5). The Remote Control plug is
also needed to make certain connections inside the instrument. By th is
method the overall frequency range of the apparatus can be reduced to
about one octave. In applications where the 1022 is used with the Level
Recorder 2305 the adjustable cams can be made to operate the pen Iift ..
A 1000 Hz reference signal can be supplied at the output sockets when
the Reference Oscillator is connected and the scale pointer is set to the
position 1000 Hz Ref.signal. The use of this reference signal can be seen in
the operation chapter.
The signals from the fixed and variable oscillators are mixed and then
passed to a low-pass filter. The filter has a cut-off frequency of 50 kHz and
is used primarily to eliminate any 100- 120kHz or 220- 240kHz compo-
nents.
The filtered signal is then fed to the output amplifier. This will give an
output power of 2.5 W nominal load (i.e. the load indicated on the Match-
ing Impedance switch). The output amplifier is current limited to protect
the transistors.
The 1022 can be operated from 100, 115, 127, 150, 220,240 Volts AC
and 50 to 400Hz mains supplies and the maximum power consumption is of
the order of 26 W with full load.
o.J.
1 Watt
Load
7
~
c
~ 0·2
"""
"' t'-..... v
.............__
--
...........
___... _......, / l.-/
............
~ ............__
-=0 1-- -
__, ~t.10V
No load
..
; 0·1
.:
-;
c
0
z 10 20 30 50 100 2 3 5 1k 5 10k 20k
Frequency (Hz) _ , .
11oo'J5
14
4. OPERATION
PRELIMINARY ADJUSTMENTS
Before the instrument is used check the voltage selector on the rear panel
is set to the correct line voltage. If not, remove the central fuse and adjust
with a small coin or screwdriver.
A. CALIBRATION
3. Set main scale pointer to the frequency of the mains supply (e.g. 50 or
60 Hz) checking that the frequency increment scale is set to zero.
4. Set the meter deflection to higher than centre scale reading with OUT-
PUT LEVEL knob.
5. Press POWER FREQUENCY BEAT button and hold "in". At the same
time slowly adjust the FREQUENCY SCALE ALIGNMENT FINE knob
until a large fluctuation registers, slows up, and practically ceases on the
meter dial.
6. Set main scale pointer to 20Hz and subtract 20Hz with FREQUENCY
INCREMENT knob. If the meter deflection drops to zero then the B.F.O.
is·calibrated. If not, continue to point 7.
7. If the meter deflection does not drop to zero, adjust FREQUENCY SCA-
LE ALIGNMENT COARSE with a screwdriver until the meter deflection
does.
8. Reset the main scale pointer to the frequency of the mains supply. Reset
frequency increment scale to zero. Make a final adjustment with F RE-
QUENCY SCALE ALIGNMENT FINE to obtain the exact position of.
"slow beat" as in point 5. The B.F.O. is then calibrated.
15
B. OPERATION USING LOAD TERMINALS
2. Select suitable matching impedance for the load using MATCHING IMPE-
DANCE switch.
Note: The meter reading is correct only when the impedance of the load
is high compared to the son attenuator impedance.
D. FREQUENCY MODULATION
1. Set knobs:
Note: When the Beat Frequency Oscillator is used in conjunction with the
Level Recorder 2305 the writing speed of the Level Recorder should be
kept below the regulation speed of the compressor.
F. AUTOMATIC RECORDING
18
. '~'
~-~i
Object
under
test -f_7oo17
Potentiometer Range
dB
~~~~~~~~....____Remote Control
~Two-Channel Selector
......._____Drive Shaft II
Finger Wheel Z
Input Potentiometer_/'
Ref. Voltage Pushbutton
9. In order to synchronise the units, stop the paper so that the stylus
rests on the 10 Hz line.
10. Depress 100 mV reference button on Level Recorder and use INPUT
POTENTIOMETER to adjust stylus to a suitable level on the record-
ing paper.
11. Set pointer of 1022 to 1000Hz REF. SIGN.AL and engage magnetic
clutch by use of clutch switch. The units will then be synchronised so
that the distance between the 10 and 20 Hz marks on the paper
20
corresponds to the distance between the 1000Hz REF. SIGNAL and
20 Hz marks on the 1022.
12. Depress 1000Hz REF. SIGNAL button and adjust the output level
from the 1022 and/or the INPUT ATTENUATOR of the Level Re-
corder so that the stylus deflects to somewhere in the middle of the
paper.
G. REMOTE CONTROL
Several forms of remote control are possible with the 1022. A diagram
showing the connections of the Remote Control plug is shown in Fig.4.4.
Oscillator Stop
Internal
b External Modulation
21
Magnetic clutch
External Modulation
Oscillator Stop
For remote interruption of the output signal (stopping of the fixed oscil-
lator) terminal c should be connected to terminal f (ground). When reverbe-
ration measurements are made automaticaUy using the Level Recorder
2305, a switch in the Level Recorder can be used to ground terminal c. See
Applications. Room Reverberation Time.
Terminals d and e
Note
When delivered from the factory, the Oscillator is supplied with a 6-poled
plug containing the necessary connections for dead zone blocking.
The initial and/or the fincH part of the frequency range can be blocked by
means of the cams mounted on the spindle of the capacitor (Fig. 4.5.). With
full use of the cams the frequency range can be reduced to approximately
one octave in any part of the frequency range.
22
2. Remove rear panel.
3. Loosen knurled lock-nut (Fig.4.5.) sufficiently for the outer and inner
cams to be moved relative to the one in the middle.
4. Set the pointer of the main scale to the lowest frequency of the desired
frequency range.
5. Turn the inner cam to the right (seen from behind) until the contact is
activated.
6. If necessary tighten the lock-nut a little so that the cam is not displaced
when the main scale pointer is moved.
7. Set the pointer of the main scale to the highest frequency within the
desired frequency range and turn the outer cam to the left until the
contact is activated. Be careful that the inner cam does not move.
8. The cam in the middle should be rotated so that it does not interfere with
the frequency range in use.
9. Tighten the lock-nut, still being careful that the cams do not move.
Cam Discs
Knurled Nut
23
I. USE WITH 2020
To drive the 2020 from the 1022 it is necessary only to connect up the
120 kHz and 100-120 kHz sockets on the rear of the oscillator to the
corresponding sockets on the rear of the 2020. For further details of opera-
tion see the instruction manual of the 2020.
24
5. APPLICATIONS
The field of use of the Beat Frequency Oscillator Type 1022 is so exten-
sive that only a few of the possible applications are illustrated in the follow-
ing pages, these being classified into three sections:
A. ELECTRONIC MEASUREMENTS
Should the compressor circuit be used to regulate the output signal from
the Oscillator it is advisable to verify that the voltage at the COMPRESSOR
\~ INPUT is approximately to required 0.5 volt. When it is intended to use the
equipment for automatic recording of frequency characteristics, the input
of the Level Recorder may first be connected to the input of the compres-
sor, and a recording of the compressor input voltage made for the complete
frequency range in which measurements are to be taken. With the compres-
sor working correctly the resultant recording should be a straight line. If this
is the case the input to the Level Recorder can then be disconnected, and
the desired measurements carried out.
25
D D D D D D D D D D D D D D D C
llriloi6KJ<w
.-.-
~
:"i:
UB0041
;et
=·:!·:·:·!-... _;~•:·
1022
OO')'J
The only requirement the bridge must satisfy is that one diagonal point is
grounded as shown in Fig.5.2. This requires the bridge to be supplied from
1?oo'lo
26
the B.F.a. via a screened transformer e.g. TU 0005, the B.F.a. being
grounded at one terminal.
The output voltage from the amplifier under test should be measured
with an Audio Frequency Voltmeter 2409 or a Microphone Amplifier 2603
or a Measuring Amplifier 2606 an example of the arrangement being given
in Fig.5.3 .
•
Amplifier
27
B. ACOUSTICAL MEASUREMENTS
The loudspeaker under test should be fed with constant voltage or con-
stant current. (See Operation E). A constant current arrangement produces
a constant force on the diaphragm. Such an arrangement is shown in Fig.
5.4. Here the loudspeaker is fed from the B.F.O; section of an Automatic
Frequency Response Recorder 3308 via a series resistor. The voltage drop
across the resistor is then fed to the compressor input of the B. F.0. A
constant current will then be obtained in the circuit when the voltage drop
across the resistor is approximately 0.5 volt.
28
261 9 1.11.5 Absorbing material
~ P\JSI">l
29
2619+·4145 Loudspeaker
Microphone
1022
under test
@o
0027
' 1~1 ~·;
i .•
!e·~~.::
2606
17oo ~~~
0 010----.
10
...
100 -
1000 10000 (160/:111!1 A
I
C
C LA
LA •
30
must be small in comparison to the time constant determining the compres-
sor speed. Under normal circumstances these conditions are easily fulfilled.
One of the more important factors that determines the acoustical quality
of a room is its reverberation time and the 1022 has been designed with
certain special features to aid such measurements. The compressor circuit
serves to keep the sound level in the room constant with frequency while
frequency modulation of the signal ensures that standing waves and inter-
ference phenomena are avoided. The latter feature helps to produce smooth
slopes to recorded decay curves whereas if pure sine waves were radiated in
the room, standing waves might produce uneveness.
Various set-ups can be used to measure reverberation time using the 1022
as an integral part. An arrangement whereby the decay curves are recorded
automatically using the Level Recorder 2305 is discussed here. The set-up
can be seen in Fig.5.8. The Frequency Spectrometer 2113 is chosen as
amplifier for the microphone since selective reception in 1/3 or 1/1 octave
bands will reduce background noise and increase the dynamic range of the
decay curves.
Decay curves throughout the complete audio range can be recorded auto-
matically at 1/3 octave intervals using this combination of equipment and
can be made on both frequency calibrated or nonfrequency calibrated pa-
per. In order to record the decay of sound in the room the sound source
must be disconnected at definite intervals by means of the oscillator stop.
Only the decay curve is required so the pen should be lifted from the paper
between decay curves and, as selective reception is used, the filters of the
Spectrometer also need to be switched between curves. The oscillator stop,
pen lift and the filter switching can all be performed simultaneously by
remote control from the two-channel selector of the Level Recorder. The
necessary connections between the different instruments are shown in Fig.
5.9 a. whilst the electrical circuit produced can be seen in Fig.5.9 b.
31
Fig.5.8. Set-up for the automatic recording of reverberation time
Frontal Connections
Ground o - - - - - - - - - - o Ground
(a)
JDrive Shaft
Lifting ~n
Magnet
Filter
Switch
C3
D1 81
Ground Ground
(b)
Fig.5.9. Electrical connection of instrumentation for automatic
recording of reverberation time
a) Connections between remote control plugs
b) Electrical circuit for remote control
32
Frequency Calibrated Paper
33
.
DDDDDDDDDODDDDDOOOODOOOOOOOOOODDODODODOODDDODOOOOOO
lrii~I**JO .w•ll- e.r.r•JJ- INII6l;..
73
~'7.:2QQ::-~~"
""~.,_, _
w•. ,.,~.,.,Ji.E.a
... 10
a.t.:~F,..w..l...oo
10 oo
QP 0123 10 100 1000 10000 ··~~.;",~: ~ .:_ "'
1000t/s
f= Potentiometer Range : 50 315C/s 4000C7s ~ 100~C/s __
~Rectifier : RMS
f= Lower Limiting Freq. 50 T• O.Jsec.
r= Writing Speed: 400
f=: Paper Speed : · 100 T• D.45sec.
~Drive Shaft Speed : 12
~ Loop Length: 490mm
~Measuring room: Information Deportment
f= Dote: 20-1-61 B.E. B.
f 618H'
1. The Protractor is held so that the printing is readable. The proper section
is chosen and its left limiting line (thick diagonal) is placed on top of the
portion of the recorded decay curve to be measured, and in such a man-
ner that the centre of the Protractor coincides with one of the horizontal
lines on the recording paper. See Fig.5.13.
If paper speeds other than 10 and 30 mm/sec. have been used, the deter-
mined reverberation times should be multiplied or divided by factors of 10.
34
Example:
50 dB Range Potentiometer.
Paper Speed 100 mm/sec.: Use the section "50 dB 10 mm/sec."
and divide the measured result by 10,
see also Fig.5.13.
( 6 (85'6 .
In each of the two rooms separated by the wall is placed a Type 4145
Condenser Microphone individually coupled to a Preamplifier Type 2619.
Two extension cables connect the microphone units with the Two-Channel
Microphone Selector 4408. The Microphone Selector is remotely controlled
by the two-channel switching device, which is "built-in" to the Level Recor-
der portion of the Automatic Frequency Response Recorder 3308. A 50 dB
Range Potentiometer can be used, the 1022 should be frequency modulated
and the loudspeaker (or loudspeakers) placed so that a sound field, as dif-
fuse and isotropic as possible, is built up.
35
receiver room transmitter room
ing Amplifier 2606, the different sound levels picked up in the two rooms
are taken alternately and amplified before being fed to the Level Recorder.
The result is that two independent curves are automatically reproduced on
the recording paper, enabling the sound level difference between the two
sides of the wall to be read off in decibels. Such a recording is shown in
Fig.5.15. The sound absorbed by the receiving room must be taken into
account.
The Heterodyne Slave Filter 2020 has very narrow bandwidths and is
ideal for such measurements. It can also be directly controlled by the 1022.
36
ooooooooooooooooooooooooooooooooooooooooooooooooooo
-
lrO:!.!.!I•r+
""-
_...__ -"""'-
w. ,....Sc-.._
s...: _
QP 0123
10
10 100
211110
~ ..,.,.A • c
WI . - A I C L-.
la
f6(070
4145
100-120kHz
120kHz
17oo lf7
37
loooooooooooooooooooooooooooooo~
+
BrDel & KJ .. r &ool & KJGt
"- 50 2.5
Mem.lringObjoct: _ _ _
db db
Sound
transmission 4020
anallsls
USing
2020
Slave F1lter
3.16 Hz 30 15
Bandwidth
20 10
R.c;No:
Dalt:
Sign.:
Roct; RMS
Z.OIM; 10 5
LU..Fr; 20
Pot: so
Wr. Sp.: 8
Papo< Sp.: 0.1
Muhtplyfr.q. Scaleby;_1_ 0 0
50 100 200 500 1000 2000 5000 10000
QP1123 100 1000 10 000
Note: To ensure that the compressor circuit is controlled from the funda-
mental of the sound source, another 2020 could be used with the 2606 in
the compressor loop.
Another use of the Heterodyne Slave Filter with the 1022 is shown
below. The 2020 can be used for frequency rejection whereby it will reject
frequencies by over 40 dB over a 3.15 Hz bandwidth.
38
~DDDDDDDDDDDDDDDDDDDDODDDDDDDDD~
Bru~jcar • Brooi&K;-
50 25
Meosuri"'l Objoct: _ _ _
db db
Sound
transmission 40 20
recording
Without
Filtering
30 15
20 10
Roc:; No.:
Dolo:
Sign.:
Rod"
z.... t..." 10 5
L Lim. F" 20
Poo.: 50
Wr. Sp.: 8
Popor Sp.: 0.1
Multiply freq. Scolo by"_ 1 - o o-
50 100 200 500 1000 2000 5000 10000
QP 1123 100 1000 10000
The test environment is provided by the Type 4212 Hearing Aid Test
Box, which gives practically free-field conditions over the frequency range
150 Hz to 5 kHz. The receiver of the hearing aid under test is excited by a
loudspeaker inside the anechoic chamber of the test box. Sound pressure at
the position of the receiver is maintained constant by a regulating condenser
microphone which provides feedback to the compressor circuit of the oscil-
lator supplying the loudspeaker signal. The hearing aid output is coupled to
an artificial ear as required by the I EC (Recommendation 118). The artifi-
cial ear contains a pressure response microphone which detects the output
from the earphone. Using the Slave Filters "Output" (L.F .) and "Rejection"
outputs the respective frequency and harmonic analysis spectrograms can be
recorded on the same frequency calibrated chart using the Level Recorder.
39
le ve l Recorde r
2305
Measur i ng Ampl if i er
2606
BFO
1022
100 - 120 kH z
120 kHz
Pr
p-x 100%
Where "Pr" is the RMS harmonic sound pressure level excluding funda-
mental and "P" is the overall RMS sound pressure level.
j+
Alternatively this can be expressed by:
~
22 + p32 + p4 2 + • - • - - - • -
X 100%
Harmonic Distortion= p + p + p + p ___ _
12 22 32 42
where P1 =amplitude of fundamental sound pressure
Pn = amplitude of the n th harmonic.
Note: When the "Rejection" output is used for harmonic analysis the
compressor feedback voltage can be taken from the "Output" of the same
instrument, since both "Rejection" and ,;Output" circuits function simul-
taneously.
40
C. MECHANICAL MEASUREMENTS
2623
41
To measure the strain on the test object a resistance strain gage should be
used and a Strain Gage Apparatus 1516 will be found ideal as the measuring
bridge. The output voltage from the Strain Gage Apparatus can be fed
directly to a Level Recorder with linear potentiometer for automatic record-
ing.
DODOODDODDDDODODOODODOOODD)
50
db
20
...
10
42
6. ACCESSORIES
The 2020 is driven electrically from the 1022 which is in turn driven by a
flexible shaft from the Level Recorder 2305. Hence for narrow band ana-
lyses the 10:1 Gear UG 3000 is recommended (Fig.6.1.). Generators 1022
or 1024 connect directly to this gear, which is driven from DRIVE SHAFT
II of the 2305 Level Recorder via the Flexible Shaft UB 0041, as shown in
Fig.6.2. Thus the mechanical loading on the Flexible Shaft is considerably
reduced and the sweep speed regulation is improved by a factor of 10 or
more.
To 1022 or 1024
To UB0041
Fig.6.1. The 10:1 Gear UG 3000
43
Note: For correct synchronization of paper and sweep speed the syn-
chronizing Gear Lever X on the 2305 (Fig.6.2.) should be in its outer
position and the knobs PAPER SPEED and DRIVE SHAFT SPEED should
be set as in TABLE I
TABLE I
~··· L____J • :. • ~~ :i ·
"'- - -·- ···:....;.· '
"'~:~"·· @
100-120 kHz
To
Slave Filter
2020
-17a1o 7
44
OUTPUT TRANSFORMER TU 0005
45
7. SPECIFICATIONS
Frequency Range 20-20.000 Hz.
Frequency Scales
Distortion
Frequency Deviation
46
Frequency Modulation
Oscillator Stop
Attenuator Output
Load Output
Compressor Input
For the requirements of this socket, see Remote Control, Operation chap-
ter.
47
Accuracy of Attenuators
Power Requirements
Cabinets:
Type 1022A
Type 10228
Similar to Type 1022A but the instrument and the metal case are housed
in a mahogany cabinet with cover. In this cabinet it is easy to transport the
instrument.
Type 1022C
48
Dimensions:
Accessories Supplied
1 Power Cord
1 Flexible Drive Shaft UB 0041
1 Remote Control Plug JP 4722
3 Fuses VF 0009
2 Scale Lamps VS 1271
2 Scale Lamps VS 1273
49
BROEL & KJ.lER instruments cover the whole field of sound and vibration
measurements. The main groups are: