L8_UV-VIS Spectrophotometry
L8_UV-VIS Spectrophotometry
Microwave
Infra-red
Visible
Ultraviolet
X-rays
Gamma rays
Long Wavelength Short
Low Energy High
Low Frequency High
EFFECT OF ENERGY ON A MOLECULE
ELECTROMAGNETIC SPECTRUM
ENERGY
1.2 x105 1.2 x107 12000 310 150 0.12 0.0012
( kJ/mol)
Electronic excitation
e-
FREQUENCY
(Hz) 1020 1018 1016 1014 1012 108
visible
Cosmic γ x Ultra Radio
Infrared Microwave
rays rays rays violet waves
WAVELENGTH
(m) 10-12 10-11 10-9 10-6 10-3 10-1
The UV Absorption process
•Any of these require that incoming photons match in energy the gap
corresponding to a transition from ground to excited state.
What are the π* π* π* Example
π* π* π* for a
simple
nature of these π* π* π* enone
n n n
absorptions? π π π
π
π π
-*; max=218 n-*; max=320
=11,000 =100
Example: → * transitions responsible for ethylene UV absorption at ~170 nm
calculated with ZINDO semi-empirical excited-states methods (Gaussian 03W):
h 170nm photon
• The amount of light transmitted with respect to the incident light is called
TRANSMITTANCE (T) ie.,
T= t
I
Io
• Sample can absorb all or none of the incident light and therefore transmittance
often quoted as a percentage eg.,
It
% T= X 100
Io
UV / VISIBLE SPECTROSCOPY - THEORY
ABSORBANCE A = - log10 T
It
A = - log10
Io
A = log10 Io
It
•Absorbance is logarithmic, meaning small changes in transmittance can result
in large changes in absorbance.
•As absorbance increases, transmittance decreases, indicating more light is
being absorbed by the sample.
For example:
•If T = 1 (100% transmittance), no light is absorbed, so A = 0.
•If T = 0.1 (10% transmittance), then A = 1.
•If T = 0.01 (1% transmittance), then A = 2.
• By plotting Absorbance vs wavelength an ABSORBANCE SPECTRUM is generated.
• With the advantage that absorbance measurements are usually linear with
Concentration, absorbance spectra are now used
THE LAWS OF SPECTROPHOTOMETRY
A cl
A = Ecl (A is a ratio and therefore has no units)
The constant E is called the MOLAR EXTINCTION COEFFICIENT
UV / VISIBLE SPECTROSCOPY - THEORY
UNITS OF THE MOLAR EXTINCTION COEFFICIENT
• CONCENTRATION (c) - Moles litre-1
• PATHLENGTH (l) - cm
A = Ecl Hence E= A
cl
E = 1 ˛
mole litre-1 x cm
E = mole-1 litre x cm -1
But 1 litre = 1000cm3
E = 1000 mole -1 cm3 x cm -1
Hence Units of E = 1000 cm2 mole -1
UV / VISIBLE SPECTROSCOPY - THEORY
ABSORBANCE AT 300nm
From measuring absorbance of sample x
x
Concentration of analyte in sample
can be obtained from the calibration curve x
x
E can be obtained from the slope of the
calibration curve for a given wavelength () x
CONCENTRATION (moles litre-1 )
Instrumentation
1. Radiation source
2. Collimating system
3. Sample holder/container
4. Detector
5. Amplifier/Read-out unit
Radiation Source (for UV radiations)
• Ideal characteristics of radiation source-
– Should be stable and no Fluctuations
– Provide high intensity incident radiations
– Should emit continuous spectrum/uniform intensity
– Inexpensive
– Shouldn’t show fatigue on continuous usages
1. Hydrogen Lamp
– A pair of electrodes in glass tube filled with Hydrogen at
low pressure. When current is applied at high voltage,
discharge of electrons occurs which excites Hydrogen gas
to high energy states. The electrons return to ground state
emitting radiation in 180-350nm range.
2. Deuterium Lamp
Radiation Source (for Visible radiations)
2. Prism
– Glass or quartz
3. Grating
– Large number of parallel grooves/lines (15000-30000
per inch) on highly polished aluminium surface.
Sample Holder
• Cells or cuvettes or test tubes are used for
handling liquid samples.
• For UV radiations
– Rectangular cuvettes (with or without lid) of glass with
quartz or fused silica windows
Detector
• The intensity of transmitted radiation by sample is collected on
detector and is measured.