BSBE_Package Insert of C4
BSBE_Package Insert of C4
Complement 4 Assay Kit 3.The production date and expiry date are shown on the
label .
(C4) APPLICABLE INSTRUMENT
This kit is theoretically suitable for all biochemistry
Method: Immunoturbidimetric Method analyzers and spectrophotometers covering the
wavelength range of 340 nm.
Cat . No. Packing Size Analyzer It is recommended to use this kit on a biochemistry
analyzer for testing according to laboratory conditions.
R1: 2×50 ml For Hitachi917&
EGS681M SAMPLE COLLECTION AND PREPARATION
R2: 2×10 ml OlympusAU640/400/600 Serum .
R1: 2×50 ml For Hitachi 717 Samples can be kept stable at 2-8℃ for 7 days.
EGB680M R2: 2×10 ml &ShimadzuCL7200/8000 ASSAY PROCEDURE
R1:30×4.3ml For Siemens Test Condition(Hitachi 917)
EGD681M R2: 6×4.3 ml Dupont/Siemens Behring Main
340nm Sample(S) 2.4 l
Series wavelength
For Mindray Secondary Reagent 1
EG15AG4 R1:1×25ml 200l
BS120/180/190/200/220/ wavelength (R1)
60BS R2: 1×5 ml
230/240/430/460/830 Reaction Reagent 2
37℃ 40l
R1:1×25ml temperature (R2)
EGG15AG For Semi Auto Analyser
R2: 1×5ml Two
Cuvette
INTENDED USE diameter
1cm Reaction type point
For in vitro the quantitative determination of end
complement 4 in human serum. Operate procedure
CLINICAL SIGNIFICANCE Add into Cuvette:
Complement is mainly synthesized in the liver, and Sample(S) 2.4l
complement C3 and C4 are the most frequently Reagent 1(R1) 200l
detected. C4 is an important component of the classical Mix well and incubate for 5 minutes at 37℃, read initial
activation pathway of complement, and its determination absorbance A1;
is helpful for the diagnosis and treatment of autoimmune
Reagent 2(R2) 40l
diseases such as SLE. Lower levels of C4 are seen in
autoimmune chronic active hepatitis, SLE, multiple Mix well and incubate for 5 minutes at 37℃, read final
sclerosis, rheumatoid arthritis, and LgA kidney disease. absorbance A2;
In SLE, the reduction of C4 is often earlier than that of Calculate △A=A2-A1
other complement components, and the recovery is later Note: Parameters above are only introduced with
in remission than that of other components. C4 values Hitachi 917 as an example. The parameters of different
were significantly lower in lupus nephritis than in non- biochemistry analyzers are slightly different. Please
lupus nephritis. Elevated levels of C4 are common in the read the manual carefully before setting parameters.
acute phase of rheumatic fever, perituberous arteritis, CALCULATION
dermatomyositis, myocardial infarction, Rditer's It is recommended to use Gcell C4 calibrator.
syndrome, and various types of polyarthritis. Calibrator traces to the international reference materials
ASSAY PRINCIPLES ERM-DA470k/IFCC.
Based on the reaction between C4 antibodies and C4 1. According to the requirements of the calibration
antigens, the test forms an immune complex and procedure in the operation manual of biochemistry
detects changes in turbidity at the wavelength of 340nm, analyzer, each laboratory establishes its own calibration
which is proportional to the content of C4 in the sample. procedure according to the specific conditions.
REAGENT COMPOSITION 2. The calibrated liquid shall be diluted proportionally.
The zero-point calibrated liquid shall be ultra-pure water
Contents Concentration 3. Five-point nonlinear calibration.
Reagent1 (R1) 4. Requirements for calibration and frequency: It is
recommended to calibrate at least once every two
Tris Buffer (pH=7.50) ≥10mmol/L
weeks. When the following situations occur, it is
Sodium Chloride ≥150mmol/L recommended to re-calibrate: change the reagent batch
Preservatives number, the indoor quality control runs out of control,
Polyethylene glycol (peg) the biochemistry analyzer carries out major
Reagent2 (R2) maintenance or replaces the main parts such as light
source or cuvette.
Tris Buffer (pH=7.50) ≥10mmol/L
QUALITY CONTROL
Anti C4 antibody -- It is recommended to use Gcell C4 control. The
Preservatives -- absorbance of quality control should be within the
STABILITY AND PREPARATION OF REAGENTS labeled value range. If the results deviate from the
1.Stable up to the expiry date when the reagent is scope, please find out the reason by following steps:
sealed and stored in dark at 2-8 ℃. 1. Check the parameter setting and light source.
2.Once opened, the reagents are stable for 28 days 2. Check the cleanliness of the cuvette and sampling
when refrigerated on the analyzer or refrigerator. needle.