The document outlines the calibration procedures for various laboratory instruments, including balances, UV-VIS spectrophotometers, fluorimeters, and HPLC systems, emphasizing the importance of accuracy and precision in measurements. It details the frequency of calibration, methods for ensuring instrument reliability, and the significance of adhering to standards set by organizations like the USP and Indian Pharmacopoeia. Additionally, it highlights the advantages of calibration such as improved quality assurance, reduced errors, and cost-effective auditing.
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0 ratings0% found this document useful (0 votes)
6 views
Calibration
The document outlines the calibration procedures for various laboratory instruments, including balances, UV-VIS spectrophotometers, fluorimeters, and HPLC systems, emphasizing the importance of accuracy and precision in measurements. It details the frequency of calibration, methods for ensuring instrument reliability, and the significance of adhering to standards set by organizations like the USP and Indian Pharmacopoeia. Additionally, it highlights the advantages of calibration such as improved quality assurance, reduced errors, and cost-effective auditing.
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 62
Calibration of Instruments
• ICH definition : The demonstration that a
particular instrument or device produces results within specified limits by comparison with those produced by a reference or traceable standard over an appropriate range of measurements. • The aim of the calibration program is to ensure that all measuring and testing equipment included in the program are calibrated within the manufacturers accuracy specifications or the tolerance required for the application. • Calibration may be called for: • A new instrument. • After an instrument has been repaired or modified. • When a specified time period has elapsed. • Before and/or after a critical measurement. • After an event, for example • After an instrument has had a shock, vibration, or has been exposed to an adverse condition which potentially may have put it out of calibration or damage it. • Sudden changes in weather. • Whenever observations appear questionable or instrument indications do not match the output of surrogate instruments. As specified by a requirement, e.g., customer specification, instrument manufacturer recommendation Advantages of Calibration • greater assurance of quality • Quality Control • Reduces Chances of Error • Cost Effective Auditing • Increase Consistency and Reliability Prevent and Predict Process Failures Electronic balance • USP General Chapter 41 "Balances" is mandatory and states the requirements for balances used for materials that must be accurately weighed. Weighing should be performed using a balance that is calibrated over the operating range and meets the requirements defined for repeatability and accuracy. • USP General Chapter 1251 "Weighing on an Analytical Balance" is a guideline applicable to balances used in all analytical procedures. • It provides information on installation and operational qualification (IQ/OQ), performance qualification and balance checks. • Weighing is a frequent step in analytical procedures, and the balance is an essential piece of laboratory equipment in most analyses. • In spite of this, weighing is a common source of error that can be difficult to detect in the final analytical results • The weighing procedure can be separated into three basic steps: ❑Planning ❑checking the balance and weighing the material. Balance Uncertainties • Check the sample, the balance, and the laboratory environment for the following causes of errors, and eliminate them: • 1. A balance door is open. • 2. Temperatures of the balance and the material to be weighed are not the same. • 3. The sample is losing or gaining weight. • 4. The balance has been recently moved but has not been allowed to equilibrate to its surroundings or has not been recalibrated. • 5. Air currents are present in the laboratory. • 6. Temperatures in the laboratory vary. • 7. The balance is not properly leveled. • 8. Laboratory operations are causing vibration.. • Typically, the weighing of a sample or standard is the first step in the analytical procedure, followed by dilution and subsequent analysis by techniques such as HPLC or qNMR. • Any error in the weighing step has the potential to propagate through the whole analytical process, causing inaccuracy in the final result. • To avoid this situation, the United States Pharmacopeia (USP) has set stringent requirements for balances that are used to weigh analytes for quantitative assessments. These requirements are designed to ensure that any uncertainty in weighing is small or even negligible within the analysis. Calibration • Frequency Internal calibration and accuracy = Daily • Precision: Once in a month
• 5.1 Operate the instrument as per respective
Standard Operating Procedure. • 5.2 Switch ‘ON’ the instrument. • 5.3 Switch ‘ON’ the balance. • 5.4 The display will blink with 8.8.8.8.8.8. • After few seconds, the display will show 0.00 g. • 5.6 If display is not stable, press the TARE key & wait till the display shows 0.00 g. • 5.7 Perform internal calibration Accuracy • Place 5 gm. on the weighing pan. • Note the weight. • Calculate the difference between the weight in certificate and observed weight. • Repeat the above steps using 50gm & 100 gm. weights. precision • Place 5 gm. in the weighing pan • Note the weight • Repeat the above two steps nine times. • Record the weight in Anenxure-I . • Calculate the Standard deviation. • Calculate the measurement uncertainty using following equation. • Measurement uncertainty = 2 X SD / Actual weight from certificate Calibration of UV-VIS spectrophotometer • Parameters for calibration: ➢Calibration for wavelength accuracy. ➢Calibration for absorbance measurement. ➢Gratings performance or stray light test. ➢Resolution power. Reference standards for UV VIS spectrsocpy Control of Wavelength • Weight accurately 1.0 gm of Holmium Oxide and dissolve it in 1.4 M Perchloric acid solution. Makeup to 25 ml with the same solvent. • Select the method file of CONTROL OF WAVELENGTH in the instrument. • After selecting the file press Reference button for baseline correction. • Then fill the Cuvette with 1.4M Perchloric acid and put in the sample cubicle and press reference to zero. • After auto zero put the Holmium perchlorate solution in sample cubicle then press start key. • Scan it and verify the wavelength using absorption maxima of Holmium Perchlorate solution. The permitted tolerance for control of wavelength calibration is given in below table. Control of absorbance • Dry a quantity of the Potassium dichromate by heating to constant weight at 130°C. • Weight accurately about 60 mg of dried potassium dichromate and dissolve it in 0.005M sulphuric acid solution. Make up to 1000 ml with the same solvent. Mark the solution as (A). (60ppm) • Place the solution A in sample cell holder. • Place blank 0.01M sulfuric acid in reference cell holder. • Measure the absorbance values at wavelength 235nm,257,313 and 350nm • absorption maxima of Potassium Dichromate solution at a different wavelength and calculate the absorbance, tolerance is given in below table. 3.stray light test • Gratings performance : • Prepare 1.2% potassium chloride solution in distilled water. • Distilled water is taken as reference. • Place the potassium chloride solution in sample cell holder. • Absorbance is measured at 195-220nm. • Absorbance must be greater than 2 at 198nm. 4. Resolution power: • Prepare 0.02%v/v solution of Toluene in Hexane UV • Record spectrum of a 0.02% v/v solution of toluene in hexane • Measure the absorbance of above solution at 266 nm and 269 nm using Hexane UV as blank solution. • The ratio of the maximum absorbance at about 269 nm to that at the minimum absorbance at about 266 nm should not be less than 1.5 unless otherwise specified in the monograph. Frequency : 3 Months Calibration of Fluorimeter Frequency 6 months + 3days • Prepare 1000ppm solution using reference standard Quinine sulphate. • dilute it further to prepare 1,2,3,4,5 ppm solution using 0.1N H2SO4 solution and measure the intensity using primary wavelength 366nm. • Plot a graph using conc at X axis and intensity at Y axis • Correlation coefficient should NLT 0.99 • When it is performed? • Most HPLC systems in pharmaceutical laboratories are calibrated every 6 to 12 months. ➢ Periods longer than 12 months are not recommended, while periods shorter than 3 months are deemed un-necessary. • ➢ It also required after annual maintenance (or) major repairs though only effected modules, and not the entire system need to be calibrated. Calibration of HPLC
Pump Calibration System Detector
calibration of injector precision calibration
Flow rate Injector Injection Detector Wavelength
accuracy linearity volume linearity accuracy accuracy • Calibration of Pump Flow Rate Accuracy: • ➢ fill all solvent reservoirs with HPLC grade water. • ➢ Set the flow rate to 0.5 ml/min. • ➢ Wait for about 15min to stabilize the system and ensure that the pressure is stable. • ➢ Insert the outlet tubing into a 10 ml volumetric flask and start the stop watch simultaneously. • ➢ Stop the stopwatch when the lower meniscus reaches the 10 ml mark on the flask. • ➢ Record the elapsed time. • ➢ Similarly check the flow for 1.0 ml/m and 2.0 ml/m. • ➢ The time taken to collect the water should be with in ± 2.0% of the actual value. Calibration of Injector Injector linearity • ➢ Duplicate injections of 10μg/ml caffeine solution are injected by setting 5μl, 10μl, 15μl and 20μl respectively one after one as injection volumes. • ➢ The mean peak areas for the above injected volumes are noted. • ➢ Calculate the correlation coefficient by using linearity curve and the value should be with in the acceptance criteria i.e., NLT 0.99 Calibration of Injector Injection volume Accuracy
• ➢ HPLC vial is filled with water (HPLC grade)
and weighed, it’s weight is recorded as W1 • ➢ The vial is placed in the injection tray, then inject six times. • After completion of six injections remove the vial and weigh, it’s weight is recorded as W2 • ➢ The injection volume is calculated using the formula, • Acceptance criteria: The mean injected should be 50.0±1.0 μl System Precision • ➢ Standard Preparation: Accurately weigh and transfer about 60mg of Caffeine into a 100ml volumetric flask. Dissolve and dilute to the volume with mobile phase. Transfer 10ml of this solution into a 100ml volumetric flask and dilute to the volume with mobile phase. • ➢ Procedure: Inject blank, followed by standard preparation in 6 replicates. • Note down the areas and retention times. • Now calculate the %RSD of retention time and peak areas for 6 replicates injections. • ➢ Acceptance criteria: The %RSD of retention time & peak area should be <1.0%. Calibration of Detector Detector Linearity • Prepare a standard stock solution of caffeine (1000μg/ml). • ➢ From the stock solution prepare solutions of concentration 1, 5, 10, 50, 100μg/ml respectively. • ➢ Duplicates of each concentration are injected. • ➢ The mean peak areas of all the concentrations are recorded. • ➢ Calculate the correlation coefficient value by using linearity curve and it should be within the acceptance criteria .i.e., 0.99. Calibration of Detector Wavelength Accuracy • ➢ The wavelength of detector is adjusted to 266nm. • ➢ Then inject duplicates of 10μg/ml solution of caffeine and then the peak area response is recorded. • ➢ The wavelength was increased successively at an increment of +1nm (267, 268, 269……, 276nm) and the peak area is recorded at all the wavelengths. • ➢ The maximum peak area response has to be obtained at 273±2nm (acceptance criteria). Flame photometer • Preparation of Standard Solution: • Weigh accurately 2.542 g of analytical reagent grade NaCl and transfer into 1 litre volumetric flask. Add 1.9g of analytical grade KCl and make upto 1000ml with double distilled water. This stock solution is successively diluted further to get working standard solution of 10, 20,40,60,80 and 100ppm. Keep highest conc 100 ppm in sample holder and adjust value of emission as 100. then record emission of sodium and potassium in flame photometer. • Indian Pharmacopoeia