Semi-Automatic Urine Analyzer Operator's Manual
Semi-Automatic Urine Analyzer Operator's Manual
: A/0
Operator’s Manual
Urine Analyzer
(Product Model Number:HT-115A)
This manual describes the functions, safety and operation of the semi-automatic urine
analyzer, in order to ensure that you can use the semi-automatic urine analyzer correctly and safely.
Please read all sections of this manual carefully before using this product.
After reading this manual, please keep it properly and refer to it whenever necessary. Thank
2. Version Information
The version number of this manual may be upgraded due to changes in software or technical
specifications. We will try our best to modify it at that time. However, there may still be some
3. Copyright
The copyright of this manual belongs to Qingdao Hightop Biotech Co., Ltd. Without the
permission of Qingdao Hightop Biotech Co., Ltd., this manual shall not be photographed, copied,
Copying or modifying the contents of this document without permission is considered illegal.
4.Applicable Objects
5. Product Information
Medical device registration Certificate No: Be sure to register with Lu machine 20202220107
Production License No.: Shandong Food and Drug Administration mechanical production
license20120133
Registered address: No.369 Hedong Road, Hi-tech Industrial Development Zone, Qingdao,
Urine Analyzer Version A/0
Shandong, China
Address of production: Add: No.369 Hedong Road, Hi-tech Industrial Development Zone,
6. Patent
The patent of semi-automatic urine analyzer is owned by Qingdao Hightop Biotech Co., Ltd.
7.Liability of manufacturer
The manufacturer is liable for the impact on the safety, reliability and performance of the
——The electrical facilities in the relevant rooms are in line with the relevant requirements;
If you really need to return the goods to Qingdao Hightop Biotech Co., Ltd., please contact
the after-sales Service department of Qingdao Hightop Biotech Co., Ltd., State the model number
and serial number of the product, and state the reason for return. Return requests will not be
Address: No.369 Hedong Road, Hi-tech Industrial Development Zone, Qingdao, Shandong,
China
Tel: 0532-58710705
Urine Analyzer Version A/0
Table of contents
Chapter 1 Safety and Precautions .......................................................................................................1
Appendix A ....................................................................................................................................... 37
Appendix B ....................................................................................................................................... 40
Appendix C ....................................................................................................................................... 42
Appendix D .......................................................................................................................................44
Appendix E ....................................................................................................................................... 46
Appendix F ....................................................................................................................................... 47
Manufacturer .....................................................................................................................................48
Urine Analyzer Version A/0
Before using the analyzer, the operator must be familiar with the safety instructions in the manual,
and follow the correct operation procedures. Users must pay attention to the information provided
in this manual. Only technicians specially trained by the manufacturer can carry out proper test
The instructions in this manual, there are three kinds of safety tips: danger, warning, attention,
before reading this manual, it is important to clearly understand these safety tips, the prompts are
defined as follows:
danger: If the situation described is unavoidable, it will cause personal injury to the operator,
warning:If the situation described is unavoidable, it may cause personal injury to the operator,
attention: If the situation is unavoidable, it may cause device damage, data loss, or system
chaos.
Type of analyzer: The noise does not reach the level that may cause harm (within the safe
During the use of the analyzer, the user shall ensure that the upper cover of the analyzer is
closed. The power supply of the grid must have a protective grounding terminal, and the
local distributor. Every component removed from the analyzer must be treated as a potential
infectious agent, with the manufacturer directing the parts, otherwise it may cause potential
hazards or lead to weakened protection. If the parts are not replaced according to this regulation,
the manufacturer shall not be responsible for the damage caused by the unauthorized operation.
1
Urine Analyzer Version A/0
Tools and instruments used in the inspection must be disinfected after use.
The user must ensure that all technical activities related to installation, inspection and
Operators have the obligation to comply with the local and national regulations on the
discharge and disposal of reagents, waste liquid, waste samples, consumables, etc.
Samples, quality control materials, and waste test strips are potentially biocontagious and
may irritate eyes, skin and mucous membranes. When handling relevant materials in the laboratory,
the operator should comply with the laboratory safety operation regulations and wear personal
protective equipment (such as gloves, masks, eye masks, protective clothing, etc.).
clothing or skin.
Follow the work procedures in the operating manual at all times. Before performing the
examination, it is necessary to check whether the patient information is entered correctly. Incorrect
input of operating procedures or patient information may result in incorrect interpretation of the
examination results.
Users are forbidden to make any changes to the analyzer, including the application software
system. Any unapproved modifications and repairs that violate the rules are not covered by the
warranty agreement.
The software can only meet the needs of users for regular tests, and for some special tests,
The user should prepare a maintenance plan and perform maintenance in strict accordance
For the use and storage of test strips and quality control products, please refer to the
operating instructions of the reagents and ensure that they are used within the validity period
2
Urine Analyzer Version A/0
Do not use test strips that are past their expiration date.
Do not use cleaning agents containing rosin oil, benzene and other substances to clean the
stain outside the instrument, because it may cause the change of color and shape, with a soft cloth
or wet cloth scrub, for serious stains, with diluted cleaning agent or alcohol clean.
When the outdoor temperature is low in winter, the instrument is placed at room temperature
Improper operation of the device may cause biological hazards and operation hazards. Please
strictly follow the protection instructions and operation instructions of the manual when using the
device.
If the storage period exceeds 5 years, contact the customer or manufacturer before
installation.
1.4 Warning
Do not unplug the power supply with wet hands, you may get an electric shock.
Do not step on, twist or pull the wires and cables that have been damaged. If the wires and
Verify that the input voltage meets the instrument requirements. Use fuses of specified
specifications.
Check the status of the analyzer switch in a before connecting the power cable.
3
Urine Analyzer Version A/0
Chapter 2 Summarize
2.1 Principle of Operation
According to the photoelectric colorimetric principle, the urine analyzer can measure the
content of biochemical components in urine through the color changes generated by the reaction
between the reagent region on the test strip and the biochemical components in urine.
When the test strip soaked with urine sample is placed on the test strip rack, the transmission
device of the instrument will transmit the test strip to the right below the detector, the reagent
blocks that have produced chemical reactions on the test strip are irradiated by the light source, the
reflected light is received by the detector. Each reagent block in the test strip reacts independently
with the corresponding components in the urine, showing different colors, and the shade of the
color is proportional to a certain component in the urine. The darker the color of each reagent
block after reaction, the larger the value of the absorbed light and the smaller the value of the
reflected light, the smaller the reflectance; conversely, the lighter the color, the smaller the value
of the absorbed light and the larger the value of the reflected light, the larger the reflectance, that
is to say, the darker the color is proportional to the concentration of various components in the
urine sample.
The white datum on the test bench is used to compensate for the error caused by the change
of urine color and instrument. The value of light reflected by each reagent region was compared
with the value of light reflected by the blank block, and the reflectance was calculated. The
reflectance.
T m Cr
R
Tr C m
In the formula:
R——Coefficient of reflection
Tr——The intensity of reflection of the test paper block against the reference light
4
Urine Analyzer Version A/0
Test Items: Urinogen, bilirubin, ketone bodies, blood, protein, glucose, white blood cells, nitrite,
Data storage capacity: 2000 test records and 200 quality control records
pressure:75kPa~106kPa;
Transmission rate:1200bps、9600bps
Power: 65VA
Weight: 6.5Kg
of biochemical components in human urine samples by medical institutions. Test items include
(Urochologen, bilirubin, ketone bodies, blood, protein, glucose, white blood cells, nitrite, pH,
specific gravity, ascorbic acid, microalbumin, creatinine, urinary calcium), provide reference for
5
Urine Analyzer Version A/0
2.5 Contraindications
Contraindications: No.
2.6 Mark
2.6.1 Host external marking
Biological contamination
risk of bioinfectivity.
6
Urine Analyzer Version A/0
care
packing box is 5
Figure2-1
① Display unit
② Propeller of propeller
③ Work table
④ Built-in printer
⑧ Air vent
7
Urine Analyzer Version A/0
8
Urine Analyzer Version A/0
Chapter 3 Installation
3.1 Environmental conditions for instrument installation
Please place the instrument on a stable, flat surface, and do not place it with the centrifuge
Do not place the instrument in a place where it can be affected by chemicals, corrosive gases
Do not use the device in a closed environment. Leave at least a 15cm distance between the
device and the wall or other surrounding devices to ensure proper ventilation and heat
dissipation.
Leave a distance of at least 15cm between the back of the instrument and the wall or other
surrounding devices to ensure that the power switch can be easily operated.
Ensure that the instrument is used in an environment with good heating and ventilation.
3.2 Unpacking
Take the semi-automatic urine analyzer and its accessories out of the packing box, the list is
as follows, check with the packing list, which shall prevail, and check whether the following parts
Description: Electrical schematics and list of components shall only be provided to qualified
9
Urine Analyzer Version A/0
2 User Manual
3 Work bench
4 Scrap box
5 Propeller
6 Power cable
7 Communication cable
11 Printing paper
3.3 Installation
3.3.1 Mounting workbench
Before installing the workbench, cut the nylon cable fastening the position of the propeller
support, remove the protective rubber pad, align the guide rail of the workbench with the guide
rail slot on the main engine, and push the workbench into the main engine in parallel, close to the
Figure 3-1
Hold the end of the propeller in hand, insert the propeller into the propeller mounting hole on the
sliding frame, and press the end of the propeller to the horizontal position, See Figure 3-2.
10
Urine Analyzer Version A/0
Figure 3-2
(a)Choose the width of 57 mm, diameter less than 45 mm thermal printing paper roll;
(b)Press the hook at the back of the printer cover, lift the cover upward, and remove the cover, as
(c)Put the printer paper roll into the paper roll box, loosen the paper roll, let the paper end close to
(d)Remove the printer press roller, press the printing paper about 10cm under the roller, and
reinstall the press roller; (Note: Do not offset the printer paper)
(e)The printing paper through the printer cover paper outlet, cover the printer cover;
Figure 3-3
A semi-automatic urine analyzer can be connected to a bar code reader to scan the bar code
on the sample tube. Take the bar code reader out of the packing box, connect the data cable of the
bar code reader, and connect the other end of the data cable (RS-232 interface) with the bar code
reader interface of the urine analyzer. Insert the bar code reader cable into the bar code reader
11
Urine Analyzer Version A/0
Figure 3-4
The semi-automatic urine analyzer can be connected to the computer and transmits data to
the computer.
Connect the end of the printer connection to the printer and plug the other end into the
Plug one end of the power cord into the power jack on the back of the urine analyzer and the
1. After the power switch is turned on, the propeller moves, and the display screen shows "the
2. After the self-check passes, the main menu is displayed, as shown in Figure 3-5.
Figure 3-5
12
Urine Analyzer Version A/0
3. Dip the strip completely into the urine sample, remove the strip and scrape the excess urine
4. Place the reagent area of the test strip upside down in the test strip area on the instrument
workbench (Figure 3-6).After the test, the built-in thermal printer prints the test report.
Do not touch any position on the touch screen during the instrument testing.
Figure 3-6
5. If the instrument does not display as expected, or the printed result is different from the
13
Urine Analyzer Version A/0
The user only needs to press the corresponding touch key to complete the setting operation.
Note: Do not touch the touch screen with hard or sharp objects.
4.2 Interface
4.2.1Test Main menu
Figure 4-1
In the picture:
On the home screen (Figure 4-1), you can perform the following operations:
14
Urine Analyzer Version A/0
Figure 4-2
You can set the test number as shown in Figure 4-2.Click the number key to enter the serial
number; Click the button to confirm the modification and return to the upper-level menu.
The serial number displayed on the screen represents the test record number of the next paper
to be tested.
After each successful test of the strip, the serial number is automatically increased by 1.
4.2.1.2 ID Number
The ID number is encoded by the sample scanned by the bar code reader. The semi-automatic
urine analyzer can be connected to a bar code reader via an RS-232 port. The ID number will be
displayed on the output report. The ID number can be scanned by a bar code reader or manually
entered.
Input method: Click ID in Figure 4-1. The following information is displayed (Figure 4-3):
Figure 4-3
15
Urine Analyzer Version A/0
The sequence number of the test corresponding to the first column, and the ID entered for that
column in the second column; If you continue to input the ID of No. 2, click "2", and the screen
Figure 4-4
Click the number key to enter the corresponding number, Click 、 move the cursor position,
Click to toggle the case of the input letter. After the input is complete, Click the button
Click the "Check calibration" in Figure 4-1 to enter the standard strip test interface, this
Click the "Run control" in Figure 4-1 to enter the QC solution test interface, this function will
Click the Figure 4-1“2020-09-11 13:01:05”, Enter the time setting screen (Figure 4-5).
16
Urine Analyzer Version A/0
Figure 4-5
Users can modify the date and time by pressing the numeric keys to enter the corresponding
numbers, input is complete, press "ENTER" to confirm, such as giving up the modification, then
In Figure 4-1, Click “Test memory”. The following information is displayed (Figure 4-6):
Figure 4-6
As shown in the figure above, there are 2 samples recorded with serial number, test time and
corresponding ID.Click the 、 ,you can turn the page to see, click the displayed content of each
line, you can view the specific test results of the record. Figure 4-7 below (14A test paper is used
as an example):
17
Urine Analyzer Version A/0
Figure 4-7
If the print is set to "Internal" or "External", click the , Printable test results, If "Computer
Port" is set to "On", the test results will be output to the computer at the same time;
Click the ,The instrument returns to the record list, as shown in Figure 4-6;
Click the ,The instrument directly returns to the main interface, as shown in Figure 4-1;
Click the 、 ,You can flip the page to view the sample records.
Click the “seq No.” in Figure 4-6. The following information is displayed (Figure 4-8):
Figure 4-8
The user can enter the seq No. through the number key, and the seq No. ranges from 0001 to
9999. After the input is complete, Click “ENTER’. If the machine stores this record, the record list
displayed on the screen is all the records under the same seq No. entered; Click the corresponding
line to view the detailed test results of this record; If the machine has no such record, "No record"
is displayed.
18
Urine Analyzer Version A/0
Figure 4-9
After the date is entered, As input”2020-04-08”,Click the Figure 4-9 ,The screen is
displayed, as shown in Figure 4-6,All the records on the screen are for April 8, 2020;Click the
corresponding line to view the detailed test results of this record; Click the unedit it, Return
to Figure 4-6.
4.2.1.6.4 Query by ID
In Figure 4-6, Click the ”ID” area, The screen display is as shown in Figure 4-4.After entering the
corresponding ID, Click the ,The record list displayed on the screen is all the records
under the same ID. Click the corresponding line to view the detailed test results of this record;
Click the
Click “Strip Type” to enter the test strip selection interface, as shown in Figure 4.2 Unedit it,
The instrument can hold 1000 QC records.Click the "Sample Record" area in the lower left
corner of Figure 4-6, The screen will switch to the query list of QC records.Click the "QC Record"
area in the lower left corner of the figure, the screen will switch back to the “sample record” query
list.
The user can set the language, unit, strip, printer, symbol, bar code (ID), PC interface, baud
19
Urine Analyzer Version A/0
options.
Figure 4-10
4.2.2.1.1 Language
Figure 4-11
The content after "Language" will be toggled with the options listed in the following table.
1 Chinese
2 English
Table 4-12
4.2.2.1.2 Unit
Click the "Unit" key and the content after "unit" will be toggled according to the options
listed in the following table.
1 Symbol
2 Conventional
3 SI
Table 4-13
4.2.2.1.3 Strip selection
Click the content after "Strip" to switch between the options listed in the following table.
1 11A
2 14A
20
Urine Analyzer Version A/0
Table 4-14
The content after " Strip " refers to the type of strip currently used. (See Appendix F for test
items of each strip)
"External" means that test results are output by an external printer connected to the instrument;
"Full on" means that test results can be output through both the built-in thermal printer and the
external printer connected with the instrument; "All Off" indicates that test results are not output
by either the thermal printer built into the instrument or the external printer connected to the
instrument.
In Figure 4-10, Click to turn the page, That is, enter the Settings menu 2 (Figure 4-16):
This interface contains the Settings of "symbol", "ID", "PC interface", "Abnormal flag" and so on.
Click the corresponding selection key, the content after the corresponding item will switch
21
Urine Analyzer Version A/0
Figure 4-16
Click the "Symbol" key and the "symbol" option toggles between "On" and "off"
22
Urine Analyzer Version A/0
4-24.
"PC interface- open", after testing each sample, the test results will be uploaded to the
computer data receiving software, otherwise no data will be sent to the computer;
"Abnormal flag-on" In the output test result, when the test value of a certain item is higher
than its abnormal value, the test result is marked with "*";"Abnormal flag - off", there is no
In the Figure 4-16, Click " " to turn the page,That is, enter the Settings menu 3( Figure
4-25):
23
Urine Analyzer Version A/0
This screen contains Settings such as “Baud rate”, “Date formate”, Set abnormal value”, and
Figure 4-25
“Baud rate” is used to set the instrument to be connected to the computer, Click “Baud rate”,
the content displayed thereafter will be switched in sequence according to the options listed in the
following table.
1 115200
2 9600
3 1200
Table 4- 26
Click “Date format”, What follows "Date format" toggles between the options in the
following table, The screen of the instrument and the date of the test result are set according to the
corresponding format.
1 YY/MM/DD
2 MM/DD/YY
3 DD/MM/YY
Table 4- 27
Click “Set abnormal value”, The following information is displayed (Figure 4-28):
24
Urine Analyzer Version A/0
Figure 4-28
In the Figure4-28, Users can set leukocyte, nitrite, urochologen, protein, blood, ketone body,
bilirubin, glucose. According to the corresponding setting gradient of each item, the following
contents will be switched according to the options listed in the following table.
Name
LEU NIT URO PRO BLD KET BIL GLU
Number
1 +-15cell/μL Pos 17μmol/L +-Trace Ca10Ery/μL 0.5mmol/L 17μmol/L 5.6mmol/L
Table 4-29
When one of the test results is higher than the set value in Figure 4-28, In the output, an "*"
is displayed before the item. Figure 4-30 shows the printed result.
Date:2020-06-09 15:47
Doctor No:0001
ID:– – – – – – – – – – – – –
URO Normal 3.4µmol/L
BIL Neg
KET Neg
* BLD 1+ Ca25 Ery/µL
PRO Neg
MCA 10mg/L
NIT Neg
* LEU 3+ >=Ca500Leu/µL
GLU Neg
SG >=1.030
PH <=5.0
VC >=0.6 mmol/L
Cr 4.4 mmol/L
A:C <3.4 mg/mmol
Normal
Figure 4-30
25
Urine Analyzer Version A/0
Click ‘‘Manufacture set up’’, The following information is displayed (Figure 4-31):
Figure 4-31
Only the correct password can be entered to set the internal parameters of the instrument.
26
Urine Analyzer Version A/0
Chapter 5 QC monitoring
5.1 Run control
In order to ensure the accuracy of test results, positive and negative urinalysis quality control
solution can be used for quality control monitoring under the following circumstances.
Test Method:
On the home screen (Figure 5-1), Click "Run control" to enter the Run control screen (Figure 5-2).
Figure 5-1
Figure 5-2
Place the test strip stained with urinalanalysis QC solution in the strip area of the workbench
(Figure5-3). After the instrument detects the presence of the test paper, the thruster pushes the test
27
Urine Analyzer Version A/0
paper to the test area (Figure 5-4).After the test is finished, output the test result and Click "Back"
to return to the main screen of the test (Figure 5-1). QC results refer to the reference value in the
QC fluid manual.
Figure5-3 Figure5-4
3. If the calibration strip is stained or damaged, please contact the supplier and do not
4. To ensure correct test results, it is recommended to test the instrument with a calibration
Test Method:
Click “Check calibration” on the home screen (Figure 5-1) to enter the Check calibration
Figure 5-5
Place the calibration strip in the check bar area on the instrument workbench (Figure 5-6),
28
Urine Analyzer Version A/0
Figure 5-6
The instrument is tested with calibration strip, and the results are printed after the test.
After the test, if the calibration passes, the measurement result will be "Calibration passed";
29
Urine Analyzer Version A/0
Since urine can be potentially infectious, please take protective measures during testing
Please dispose of urine samples and used discarded test strips in accordance with local
laboratory regulations.
This chapter describes the method of semi-automatic urine analyzer for routine urine test.
Please read Chapter 4, "Functions and Settings," in detail before you begin testing.
Figure 6-1
30
Urine Analyzer Version A/0
Note:
For specific operations, please read the test operation procedures in detail.
If the self-check fails, refer to the fault information table in Appendix D to rectify the
fault.
Note:
Do not place the instrument in direct sunlight during the test, so as not to affect the test
accuracy.
Please check the urine test paper type before the test to avoid incorrect test results due
Before starting the instrument, ensure that the workbench and the white datum are clean and
White datum
Figure 6-2
After the installation of the instrument, turn on the power switch and self-check the
31
Urine Analyzer Version A/0
Figure 6-3
The instrument tests 514 strips per hour, with a single strip measuring for 7 seconds. The
Step 1: After the self-test is passed, the propeller are in the initial position.
Step 2: The reagent area of the test strip is completely immersed in the fresh, fully mixed,
uncentrifuged urine sample book and removed. Scrape the side of the test strip along the wall of
Step 3: Lay the test paper flat on the workbench, making sure that the front end of the test
paper is in contact with the front wall of the workbench (Figure 6-4).After the instrument detects
the presence of the test paper, the propeller pushes the test paper to the test area (Figure 6-5).
Step 4: When the propeller returns to the original position (Figure 6-6), place another strip of
32
Urine Analyzer Version A/0
avoid touching the propeller during operation. The area above the propeller
Please place the test strip and standard strip when the propeller is in the initial
position. Do not operate during the movement of the thruster to avoid touching
33
Urine Analyzer Version A/0
Chapter 7 Maintenance
Do not spill liquid on the equipment or electronic parts to prevent damage to the
equipment.
Do not wipe the instrument with gasoline, paint thinner, benzene compounds and other
Do not wipe the bench with any material that will bruise the bench and the white base.
(1) Keep the instrument clean by wiping it with a soft dry cloth or a soft cloth dipped with a mild
stain remover.
(3) In order for the instrument to operate properly and provide accurate test results, the workbench
(4) Remove the propeller, workbench and stepper plate from the instrument, rinse with clean water,
(5) Wipe the white datum clean with a soft dry cloth.
34
Urine Analyzer Version A/0
(3)Use a cotton swab 0.1mol/L NaOH solution, crub thoroughly and rinse the remaining NaOH
(5)Reboot self-test.
Users are advised to clean the instrument in this way once a week.
7.3 Disinfection
(a)As datum, workbenches and stepper plate come into contact with urine samples, they should be
disinfected.
(d)Dip the workbench, propeller, stepper plate into the disinfectant and make sure the white datum
(f)After removal, rinse the disinfectant on the above items with water.
(h)Reboot self-test.
Repeated use of glutaraldehyde solution for a long time may result in slight
discoloration of the table, which does not affect the instrument test results.
Do not allow the disinfectant to come into contact with the white datum.
The service life of the product is 8 years, and the treatment of the product at the end of the
service life shall comply with the requirements of local laws and regulations.
35
Urine Analyzer Version A/0
1 Mainframe(HT-115A) 1 □
3 Power line 1 □
5 Conformity certificate 1 □
6 Instruction book 1 □
8 Scrap box 1 □
vibration and extrusion. Handle with care when handling, loading and unloading.
36
Urine Analyzer Version A/0
Appendix A
Semi-automatic urine analyzer and computer interface
The semi-automatic urine analyzer is connected to the computer through an RS-232 interface, The
optional
Check none
SI units, conventional units, symbol system communication format (bar code scanning off state):
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
STX CR LF
SP D a t e : × × × × - × × - × × SP × × : × × CR LF
SP N o . SP × × × × CR LF
37
Urine Analyzer Version A/0
SP(﹡) U B G SP × × × × × × × × × × × × × × × × CR LF
SP(﹡) B I L SP × × × × × × × × × × × × × × × × CR LF
SP(﹡) K E T SP × × × × × × × × × × × × × × × × CR LF
SP(﹡) B L D SP × × × × × × × × × × × × × × × × CR LF
SP(﹡) P R O SP × × × × × × × × × × × × × × × × CR LF
SP(﹡) M C A SP × × × × × × × × × × × × × × × × CR LF
SP(﹡) N I T SP × × × × × × × × × × × × × × × × CR LF
SP(﹡) L E U SP × × × × × × × × × × × × × × × × CR LF
SP(﹡) G L U SP × × × × × × × × × × × × × × × × CR LF
SP S G SP SP × × × × × × × × × × × × × × × × CR LF
SP p H SP SP × × × × × × × × × × × × × × × × CR LF
SP V C SP SP × × × × × × × × × × × × × × × × CR LF
SP C R E SP × × × × × × × × × × × × × × × × CR LF
SP A : C SP × × × × × × × × × × × × × × × × CR LF
SP × × × × × × × × × × × × × × × × CR LF
ETX
SI units, conventional units, symbol system communication format (bar code scanning open state):
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
STX CR LF
SP D a t e : × × × × - × × - × × SP × × : × × CR LF
SP N o . SP × × × × CR LF
SP I D : × × × × × × × × × × × × × × × CR LF
SP(﹡) U B G SP × × × × × × × × × × × × × × × × CR LF
SP(﹡) B I L SP × × × × × × × × × × × × × × × × CR LF
SP(﹡) K E T SP × × × × × × × × × × × × × × × × CR LF
SP(﹡) B L D SP × × × × × × × × × × × × × × × × CR LF
SP(﹡) P R O SP × × × × × × × × × × × × × × × × CR LF
SP(﹡) M C A SP × × × × × × × × × × × × × × × × CR LF
38
Urine Analyzer Version A/0
SP(﹡) N I T SP × × × × × × × × × × × × × × × × CR LF
SP(﹡) L E U SP × × × × × × × × × × × × × × × × CR LF
SP(﹡) G L U SP × × × × × × × × × × × × × × × × CR LF
SP S G SP SP × × × × × × × × × × × × × × × × CR LF
SP p H SP SP × × × × × × × × × × × × × × × × CR LF
SP V C SP SP × × × × × × × × × × × × × × × × CR LF
SP C R E SP × × × × × × × × × × × × × × × × CR LF
SP A : C SP × × × × × × × × × × × × × × × × CR LF
SP × × × × × × × × × × × × × × × × CR LF
ETX
Note:SP(*)---SPor*STX=0X02CR=0X0DLF=0X0A
39
Urine Analyzer Version A/0
Appendix B
Electromagnetic compatibility
Definition: EMC (Electromagnetic Compatibility) is the ability of a device to suppress
electromagnetic interference from other devices without causing similar electromagnetic radiation
Declaration: The equipment meets the emission and anti-interference requirements stipulated
Note:
This equipment is designed and tested according to the Class A equipment in GB4824.In the
home environment, the device may be subjected to radio interference, and protective measures
should be taken.
The use of devices that emit radio waves in the vicinity of the analyzer, such as cellular
phones, transceivers, mobile radio-controlled toys, etc., may cause instability in the analyzer's
function. The power to these devices should be cut off near the analyzer.
During installation, keep the analyzer away from other electronic equipment as far as
possible.
Cables provided or designed by the company must be used and connected according to
installation procedures.
Specified peripherals that can be connected to the analyzer should be used. Avoid using other
devices that are not specified. Otherwise, the EMC performance of the analyzer may be reduced.
Do not make any changes to this analyzer. The EMC performance of the analyzer will be
degraded if the user changes unilaterally. The changes of this analyzer include cable changes,
Make sure to tighten all screws after the repair. Loose screws may degrade EMC
performance.
②It is the responsibility of the user to ensure that the EMC environment of the equipment is in
40
Urine Analyzer Version A/0
①Electromagnetic interference can be mitigated by keeping other devices away from the analyzer.
④Electromagnetic interference can be reduced by changing the power channels of other devices.
41
Urine Analyzer Version A/0
Appendix C
Semi-automatic urine analyzer output value
Regular system of units SI of Units
Projec symbol symbol
Conventional The plus symbol
t system is system
system is on system is on system is off
off
Norma Norma
l 0.2mg/dL 0.2mg/dL l 3.4μmol/L 3.4μmol/L Normal
Norma 1mg/dL 1mg/dL Norma 17μmol/L 17μmol/L Normal
URO l 2mg/dL 2mg/dL l 34μmol/L 34μmol/L 1+
1+ 4mg/dL 4mg/dL 1+ 68μmol/L 68μmol/L 2+
2+ ≥8mg/dL ≥8mg/dL 2+ ≥135μmol/L ≥135μmol/L 3+
3+ 3+
Neg Neg Neg Neg Neg
1+ 1mg/dL 1mg/dL 1+ 17μmol/L 17μmol/L 1+
BIL
2+ 3mg/dL 3mg/dL 2+ 51μmol/L 51μmol/L 2+
3+ ≥6mg/dL ≥6mg/dL 3+ ≥103μmol/L ≥103μmol/L 3+
Neg Neg Neg Neg Neg
+- 5mg/dL 5mg/dL +- 0.5mmol/L 0.5mmol/L +-
KET 1+ 15mg/dL 15mg/dL 1+ 1.5mmol/L 1.5mmol/L 1+
2+ 40mg/dL 40mg/dL 2+ 3.9mmol/L 3.9mmol/L 2+
3+ ≥80mg/dL ≥80mg/dL 3+ ≥7.8mmol/L ≥7.8mmol/L 3+
Neg
Neg Neg Neg Neg
Ca10Ery/μL Ca10Ery/μL
+- +- Ca10Ery/μL Ca10Ery/μL +-
Ca25Ery/μL Ca25Ery/μL
BLD 1+ 1+ Ca25Ery/μL Ca25Ery/μL 1+
Ca80Ery/μL Ca80Ery/μL
2+ 2+ Ca80Ery/μL Ca80Ery/μL 2+
≥Ca200Ery/μ ≥Ca200Ery/μ
3+ 3+ ≥Ca200Ery/μL ≥Ca200Ery/μL 3+
L L
Neg Neg Neg Neg Neg
Trace Trace Trace Trace Trace
PRO 1+ 30mg/dL 30mg/dL 1+ 0.3g/L 0.3g/L 1+
2+ 100mg/dL 100mg/dL 2+ 1.0g/L 1.0g/L 2+
3+ ≥300mg/dL ≥300mg/dL 3+ ≥3.0g/L ≥3.0g/L 3+
10mg/L 10mg/L 10mg/L 10mg/L 10mg/L
30mg/L 30mg/L 30mg/L 30mg/L 30mg/L
MCA
80mg/L 80mg/L 80mg/L 80mg/L 80mg/L
150mg/L 150mg/L 150mg/L 150mg/L 150mg/L
Neg Neg Neg Neg Neg
NIT
Pos Pos Pos Pos Pos
Neg Neg Neg Neg Neg
LEU +- Ca15Leu/μL Ca15Leu/μL +- Ca15Leu/μL Ca15Leu/μL +-
1+ Ca70Leu/μL Ca70Leu/μL 1+ Ca70Leu/μL Ca70Leu/μL 1+
42
Urine Analyzer Version A/0
43
Urine Analyzer Version A/0
Appendix D
Fault Information Table
1. Reboot self-test.
The strip does not
2. Clean the workbench.
Screen display move to the test
5 3. Check that the test strip is selected correctly.
“Strip missing” position
4. Check the workbench and cover plate for foreign bodies.
5. If it still cannot be ruled out, please contact the supplier.
44
Urine Analyzer Version A/0
Ambient light is
abnormal. The
Screen display 1. Place the instrument away from strong light sources.
instrument is exposed
7 “Stray Light 2. Turn off flashing light sources.
to strong sunlight or
Warning” 3. Restart the machine.
near a flashing light
source
1. Check whether the type of strip used is consistent with
The type of strip is that set in the instrument.
Screen display
8 wrong 2. Check that the strip is fully dipped into the sample.
“Strip Error”
3. If it still cannot be ruled out, please contact the
supplier.
The strip optical
The thruster does 1. Check for direct sunlight on the work surface. If so,
detector can not
not perform the move the instrument away from direct sunlight.
9 determine the
normal push bar 2. If it still cannot be ruled out, please contact the
presence or absence
action supplier.
of test strips
Screen display
1. Reboot self-test.
10 “Test system Test system failure
2. Please contact the supplier.
error”
Screen display
The push bar system 1. Reboot self-test.
11 “Strip push system
is faulty 2. Please contact the supplier.
error”
or”
Screen display
The conveyor system 1. Reboot self-test.
12 “Strip moving
is faulty 2. Please contact the supplier.
error”
45
Urine Analyzer Version A/0
Appendix E
Test strip test items
HT-N11 Test Items:Vc,LEU,GLU,BIL,pH,BLD,SG,KET,PRO,NIT,URO
46
Urine Analyzer Version A/0
Appendix F
Product pollution control label
Table of contents of toxic or hazardous substances or elements
Plastic/polymer
○ ○ ○ ○ ○ ○
parts
Part of metal × ○ ○ ○ ○ ○
Circuit board
× ○ ○ ○ ○ ○
assembly*
The battery ○ ○ ○ ○ ○ ○
Solder × ○ ○ ○ ○ ○
*: Circuit board components include printed circuit boards and their components, such as resistors,
○: It means that the content of the toxic and harmful substances in all homogeneous materials of
the component is below the limit requirements stipulated in the SJ/T11363-2006 standard;
×: Indicates that the content of the toxic or hazardous substance in at least one homogeneous
material of the component exceeds the limit required by the SJ/T11363-2006 standard.
Note: This table indicates that the recorder may contain such substances.
47
Urine Analyzer Version A/0
Manufacturer
Qingdao Hightop Biotech Co., Ltd.
Add: No. 369 Hedong Road, Hi-tech Industrial Development Zone, Qingdao,
Shandong, China
Tel: 0086-532-58710705 Fax: 0086-532-58710706
Web: www.hightopbio.com E-mail: [email protected]
European representative
MedNet EC-REP GmbH
Borkstrasse 10, 48163 Muenster, Germany
48