HEALTHCARE CATALOGUE
Middle East Catalogue
GCE HEALTHCARE
CONTENT
HIGH PRESSURE REGULATORS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 HIGH PRESSURE GAS MANIFOLD ACCESSORIES . . . . . . . . . . . . . . . 69
MEDISELECT® II . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 COLLECTION PIPE LINE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
MEDIREG® II . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 HIGH PRESSURE HOSES. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
VARIMED . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 CONNECTING PIPES FOR CYLINDER MANIFOLDS . . . . . . . . . . . . . . .71
MMR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 CYLINDER RETAINING BRACKETS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
HOSPITAL WARD EQUIPMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11 CONNECTION PIPES FOR CYLINDER PACK MANIFOLDS . . . . . . . . .71
BLENDER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 GAS EVACUATION VALVE KIT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
MEDIMETER® . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 MEDICAL GAS TAIL PIPES / PIGTAILS . . . . . . . . . . . . . . . . . . . . . . . . . . 72
MEDIFLOW® II . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 CONNECTING COMPONENTS AND ACCESSORIES . . . . . . . . . . . . . 73
MEDIFLOW® ULTRA II (LOW PRESSURE REGULATOR). . . . . . . . . . . . 15 NON-RETURN VALVES FOR CONNECTION PIPES . . . . . . . . . . . . . . . 74
MEDIFLOWTEC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 HIGH PRESSURE FILTER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
VACUUM REGULATOR - MEDIEVAC+ . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 HIGH PRESSURE VALVES 300 BAR . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
SUCTION EJECTOR - MEDIEJECT II . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 BUNDLE CONNECTING POINT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
MEDICOLLECT AND MEDICOLLECT BAG . . . . . . . . . . . . . . . . . . . . . . . 20 HIGH PRESSURE LINE VALVES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
TROLLEYS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 INDICATION PANEL WITH CYLINDER SYMBOLS . . . . . . . . . . . . . . . . 75
LOW PRESSURE HOSES - MEDICONNECT . . . . . . . . . . . . . . . . . . . . . . 23 GAS PREHEATER GGP¯200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
GAS PROBES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 PRESSURE RELIEVE VALVE MEDICAL PIPELINE SYSTEMS . . . . . . . 77
RAILS AND CLAMPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 PRESSURE MONITOR, AVSU, GAS ALARMS AND VALVES. . . . . . 78
HOMECARE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 PRESSURE MONITOR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
ZEN-O LITE™ - PORTABLE OXYGEN CONCENTRATOR . . . . . . . . . . . 30 AREA VALVE SERVICE UNITS (AVSU) . . . . . . . . . . . . . . . . . . . . . . . . . . 80
ZEN-O™ - PORTABLE OXYGEN CONCENTRATOR . . . . . . . . . . . . . . . . 31 GAS ALARM - TMA9705-EHT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
M50 - OXYGEN CONCENTRATOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 GAS ALARM ¯ TMATFT03 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
OC-E SERIES - OXYGEN CONCENTRATORS . . . . . . . . . . . . . . . . . . . . 33 GAS ALARM ¯ GASMASTER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
ECOLITE® 4000 OXYGEN CONSERVER . . . . . . . . . . . . . . . . . . . . . . . . 34 GAS ALARM PANEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
ELITE - PNEUMATIC GAS CONSERVER . . . . . . . . . . . . . . . . . . . . . . . . . 36 GAS ALARM ¯ TMA6703 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
ACCESSORIE FOR OXYGEN THERAPY . . . . . . . . . . . . . . . . . . . . . . . . . 37 NON RETURN VALVE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
HUMIDIFICATION AND FILTRATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 MEDICAL SHUT OFF VALVES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
EMERGENCY EQUIPMENT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 TERMINAL UNITS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
EASE II . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 TERMINAL UNIT - MEDIUNIT (DIN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
MARS II - MANUAL AND AUTOMATIC RESUSCITATION SYSTEM . . 43 TERMINAL UNIT - MEDIUNIT (SS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
AMBULANCE PANEL II. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 TERMINAL UNIT - MEDIUNIT (BSI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
AMBULANCE PANEL SYSTEM - APS . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 TERMINAL UNIT - MEDIUNIT (CZ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
VALVES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 TERMINAL UNIT - MEDIUNIT (AFNOR) . . . . . . . . . . . . . . . . . . . . . . . . . . 96
MEDIVITAL® . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48 TU MU INSTALLATION TOOLS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
MEDIVITOP® . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 BHU TU COMPONENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
MEDICAL CYLINDER VALVES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 TERMINAL UNIT - AGSS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
GAS MANIFOLDS AND STABILIZERS . . . . . . . . . . . . . . . . . . . . . . . . 53 EMU MAINTENANCE VALVE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
GAS MANIFOLD MC25 (FLOW 25 M3/H) . . . . . . . . . . . . . . . . . . . . . . . . 54 TERMINAL UNIT - MC70 (SS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .101
GAS MANIFOLD MM40 - HP UNIT (FLOW 40 M3/H) . . . . . . . . . . . . . . 56 BED HEAD UNITS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
LINE REGULATOR (FLOW 40 M3/H) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 BHU - NEW YORK STYLE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
GAS MANIFOLD MM40 - STABILIZER (FLOW 40 M3/H) . . . . . . . . . . . 59 BHU - COPENHAGEN STYLE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
GAS MANIFOLD MM90 BHU - MALAGA STYLE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
- HP UNIT FULLY AUTOMATIC (FLOW 90 M3/H) . . . . . . . . . . . . . . . . . 60 CEILING ICU BEAMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
GAS MANIFOLD MM90 - HP UNIT (FLOW 90 M3/H) . . . . . . . . . . . . . . 61 MEDICAL PENDANTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
GAS MANIFOLD MM90 - STANDBY (BACKUP) (FLOW 90 M3/H) . . . 62 ANESTHESIA PENDANTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .113
GAS MANIFOLD MC80 - HP UNIT (FLOW 200 M3/H) . . . . . . . . . . . . . 63 SURGICAL PENDANTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .114
GAS MANIFOLD MC80 - STABILIZER (FLOW 200 M3/H) . . . . . . . . . . 64 ICU PENDANTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
GAS MANIFOLD DUPLEX - MC80 (FLOW 200 M3/H) . . . . . . . . . . . . . 65 MEDICAL GAS FITTINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .116
GAS MANIFOLD MC150 - STABILIZER (FLOW 360 M3/H) . . . . . . . . . 67 MUNSEN RINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
GAS MANIFOLD SIMPLEX MMR (FLOW 30 M3/H) . . . . . . . . . . . . . . . . 68
MEDISELECT® II
Rotating pressure gauge which Continuous flow between
allows convenient reading settings, in the unlikely event of
mechanism failure
360°
MEDIREG® II
360°
MEDISELECT® II
Mediselect II is a high pressure medical gas regulator with a quick connector and flow selector.
Mediselect II is designed with a continuous flow mechanism between flow settings, which allows patients to
receive gas therapy in the unlikely event of device failure.
Flow ranges*
0 to 2 lpm 0, 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8, 1, 1.5, 2
Outlet connection 9/16 UNF, M12×1,25, G3/8, G1/4 with hose nipple
O-rings EPDM
Regulatory status Complies with EN ISO 10524-1 (Pressure regulators for use with medical gases)
Complies with EN 1789 (Medical vehicles and their equipment - Road ambulances)
FLOW CURVE
MEDIREG® II
FEATURES / ADVANTAGES / BENEFITS
> Regulator with pressure outlet, constantly adjusted fl ow or with flowmeter
> Rotating pressure gauge which allows convenient reading.
> Ergonomic and streamlined design
> Easy cleaning surface
> Compact and user friendly
TECHNICAL DATA
O-rings EPDM
Regulatory status Complies with EN ISO 10524-1 (Pressure regulators for use with medical gases)
Complies with EN 1789 (Medical vehicles and their equipment - Road ambulances)
ACCESSORIES
Humidifier
MEDITECTM
The new MediTec medical high pressure regulator is a blend of innovation and GCE‘s heritage in the
manufacture of medical gas equipment. MediTec‘s simple, robust and ergnomic design includes a fl ow
selector and option for quick connector, its plastic guard makes it easy to clean.
FEATURES
> Innovative self centering feature that allows fl ow of gas in the unlikely event that the device fl ow selector is
placed between two settings
> Dual gauge scale - PSI and Bar
> Up to 25 l/min
> Service / No special maintenance or service, apart from the functional tests before use is needed.
> Low weight
> Option for quick connector available
> Suitable for use in emergency resuscitation and nebulisation
> Meets the requirements of ISO EN 10524-1
TECHNICAL DATA
0 to 15 lpm 0;1;2;3;4;5;6;7;9;12;15
0 to 25 lpm 0;1;2;3;4;5;6;7;9;12;15;25
Outlet connection: Hose nipple , optionally 9/16 UNF, M12×1,25, G3/8, G1/4
Plastics: Polyamide
O-rings: EPDM
Complies with EN ISO 10524-1 (Pressure regulators for use with medical gases)
Complies with EN 1789 (Medical vehicles and their equipment - Road ambulances)
VARIMED
Varimed is a single stage high pressure regulator with high flow capacity for use with medical gas cylinders.
Varimed can also be connected to other medical devices which include pressure monitors, anaesthetic
equipment etc.
TECHNICAL DATA
Weight 1,1 kg
Weight 2400 g
Capacity 30 m3/h at P1 = 200 bar (20 000 kPa) and P2 = 5 bar (500 kPa)
TECHNICAL DATA
Emergency flow (in the event that Air or O2 supply failing) <85 <15
Pressure drop at inlet pressure from 3,1 to 5,2 bar with a flow rate of 30l/min at 60% FIO2 <0,21 <0,14
Alarm sounds when Without regulator module on 3,3 off 4,2 on 3,3 off 4,2
supply pressure
drops With regulator module on 2,3 off 3,2 on 2,3 off 3,2
Gas inlet pressure 4,5 ± 0,5 bar: air and O2 differential should be with 0,7 bar
Connection types DISS inlets and outlets for O2 and/or NIST inlets for air and O2
Weight 1600 g
CONFIGURATIONS
FLOW-METERING DEVICE
MEDIMETER®
Medimeter® is a gas flow device intended for control and measurement of air and oxygen administered to
patients. Medimeter® flow devices are available in different regional gas connections.
FEATURES / ADVANTAGES / BENEFITS
> Flat surface float allows easy and safe reading of flow values by the users
> Ergonomic design, easy for cleaning
> Available with probe connector, rail mounting with a hose and twin versions
> Soft closing mechanism
> Robust float, resistant against impact
> New scale - better reading of flow values
TECHNICAL DATA
0 - 5 lpm
0 - 30 lpm
0 - 60 lpm
Outlet connection 9/16” UNF; M12×1.25; G3/8; G1/4 (with hose nipple)
O-rings EPDM
Body dimensions
ONEW
SIN Width 32 mm
60
W VER
Height 160 mm
Depth 60 mm
lpm
LO
Storage - 30 °C to + 60 °C
Operation - 20 °C to + 60 °C
DUO version with probe Rail clamp duo 50 mm NIST Rail clamp duo 25 mm NIST
FLOW SELECTORS
MEDIFLOW® II
MediFlow® is a flow selector intended for the administration of medical gases or for use during resuscitation
and for CPAP.
FEATURES / FEATURES / ADVANTAGES / BENEFITS
> Innovative self centering flow setting device with continuous flow between settings. In the unlikely event of
indent mechanism failure, the patient will still be supplied by medicinal gas
> Lateral and frontal reading of flow setting
> Enables to get up to 50 lpm for supplying machines (not for use directly to patient)
> 360° swivelling outlet - it enables better orintation of the tube (preventing from twisting)
TECHNICAL DATA
0 - 6 lpm
0 - 15 lpm
Flow ranges
0 - 25 lpm
0 - 50 lpm
Outlet connection 9/16” UNF; M12×1,25; G3/8; G 1/4 by probe or by hose connection
O-rings EPDM
Body dimensions
Width 32 mm
Height 160 mm
Depth 60 mm
Temperature range
Storage - 30 °C t o + 60 °C
Operation - 20 °C t o + 60 °C
Regulations Complies with EN 15 002 (Flow - metering devices for connection to terminal units)
Reail clamp Complies with Standard EN 1789 (Medical vehicles and their equipment)
version
* Flowrates expressed at 23°C and 101,3 kPa
MEDIFLOW® ULTRA II
MediFlow® Ultra is the new generation of medical gas flow selector device with built-in regulator. It covers
a comprehensive combination of inlet and outlet connections and o¤ers various options for all medical
applications, from neonatal care through to resuscitation.
FEATURES / FEATURES / ADVANTAGES / BENEFITS
> Built-in regulator provides a very stable and precise flow, independent of the pressure in the medical central
gas system or cylinder.
> Innovative self centering flow setting device with continuous flow between settings. In the unlikely event of
indent mechanism failure, the patient will still be supplied by medicinal gas.
> Lateral and frontal reading of flow settings.
> 360° swivelling outlet – it enables better orientation of the nasal cannula or oxygen mask towards
the patient (preventing from twisting).
> Higher number of flow disc holes increases treatment options. Extra flow setting of 25 l/min on the traditio-
nal 15 l/ min variant, allows use in resuscitation. The additional 7 l/min is intended for nebulization.
> Ergonomic and streamlined design.
TECHNICAL DATA
0 - 2 lpm 0 - 0.1- 0.2 - 0.3 - 0.4 - 0.5 - 0.6 - 0.7 - 0.8 - 1 - 1.5 - 2
DUO probe version
Flow ranges 0 - 6 lpm 0 - 0.25 - 0.5 - 0.75 - 1 - 1.5 - 2 - 2.5 - 3 - 4 - 5 - 6
0 - 25 lpm 0 - 1 - 2 - 3 - 4 - 5 - 6 - 7 - 9 - 12 - 15 - 25
Outlet connection 9/16 UNF; M12×1.25; G3/8; G1/4 with hose nipple
O-rings EPDM
Body dimensions
Diameter 39 mm
Length 77 mm
Weight 350 g
NEW GENERATION
OF MEDICAL CLICK STOP FLOW METERING DEVICE
MEDIFLOWTECTM
MediFlowTecTM is providing innovative design of valve body and body guard connected with already existing GCE design
features known from MediVital®, MediSelect® II and other GCE products. MediFlowTecTM is an alternative for MediFlow® Ultra or
MediFlow® which has less features but during development we highly focused on safety, robustness and fl ow accuracy.
A whole range of variants and approx 13 different country standards are available
TECHNICAL DATA
Gas: O2, Air
Inlet pressure range: 4-5 bar
Flow ranges: 0 to 2 lpm
0 to 6 lpm
0 to 15 lpm
0 to 25 lpm
Inlet connection: According to national standards
Outlet connection: Hose nipple or 9/16“UNF, M12×1,25, G 3/8, G 1/4
Body material: Brass
Control knob: Polyamide
O-rings: EPDM
Filter: Sintered bronze and stainless steel
Body dimensions:
Width: 32 mm
Height: 160 mm
Depth: 60 mm
Weight: 280 g (without connector)
Temperature range:
Storage: - 30 °C to + 60 °C
Operation: - 20 °C to + 60 °C
Regulatory status: Complies with Medical Devices Directive 93/42/EEC
Complies with EN 15002 (Low pressure regulators)
Complies with Standard EN 1789 (Medical vehicles and their equipment)
MRI compatible
* Flowrates expressed at 23°C and 101,3 kPa
MEDIFLOWTECTM
Angle of QC:
Nothing - straight
90 – 90°
120 – 120°
VACUUM REGULATOR
MEDIEVAC+
MediEvac+ is a variable control medical suction equipment powered from a vacuum source used for
vacuum regulation.
FEATURES / FEATURES / ADVANTAGES / BENEFITS
> Compact and lightweight medical vacuum regulator, which allows the user to effciently and safely control
suction therapy
> The suction level of the Medievac+ is regulated via an easy accessible, front mounted control knob
> A special feature of the Medievac+ on-off valve is easy resumption of the selected de-pressure value,
when the treatment is interrupted
> The Medievac+ gauge can easily be rotated, allowing the vacuum pressure to be clearly viewed by the
operator
> The gauge scale is colour coded in sections to display a clear indication of suction level
> Three versions of adjustable pressures are available to cover all therapy needs (-250, -600 and -1000 mbar)
> The -250 mbar version has a safety valve, which will automatically shut off to guarantee maximum protecti-
on of the patient, in the unlikely event of de-pressure increase
THE COMPACTNESS OF THE DEVICE OFFERS
• Fast connection to the vacuum source
• Quick and convenient mounting of accessories (for example safety jar)
• Good accessibility of other devices connected to close located terminal units
TECHNICAL DATA
ON-OFF function RED/ GREEN button
152
Height 133 mm; 260 mm
120
Width 63 mm
77
Depth 77 mm (without connector)
SUCTION EJECTOR
MEDIEJECT II
MediEject II is the suction ejector from GCE Healthcare that uses the venturi principle to generate vacuum.
The MediEject II has best in class performance in vacuum depth, gas consumption and noise level.
MediEject II is mainly used in healthcare organizations where a vacuum source is not available. MediEject
II can be used instead of a vacuum regulator and can be used for suction of blood, secretions and food
particles from the oral cavity, the nose/ pharyngeal space and the bronchial system.
MediEject II is available in wide range of inlet probe connectors that conforms to various international
standards like DIN, BSI, SS and also available in NIST versions
.FEATURES / FEATURES / ADVANTAGES / BENEFITS
> 10 years life time
> Maintenance free
> One knob for ON/OFF and Vacuum control system
> Ergonomic shape
> Easy to clean
> Excellent vacuum performances
> Low gas consumption
> Low noise level
150
> Available in hose and probe version
TECHNICAL DATA
70
Inlet pressure 4-5 bar (400 - 500 kPa)
52 Max gas consumption
25 lpm
at inlet 4 bar)
Dimensions
Total Width 70 mm
Depth (only body, no plate/
52 mm
clamp/probe)
Max Height 150 mm
Weight 0.550 kg
MDD 93/42/EEC-MDD 2007/47/EC
T-slide
Inlet
Rail clamp
NIST
Fixed hose
Fixed hose
version
version
Hose nipple 6 mmm
COLLECTING SYSTEM
MEDICOLLECT and MEDICOLLECT BAG
GCE MediCollect and its accessories is intended for collection of liquids or/and solid particles or/and gas from body cavities or
wound. It is intended to be used by healhcare professionals.
GCE Medicollect range o ers two main collecting systems including di erent sizes of jars and bags. The collecting systems are
air-tight, autoclavable and fitted with overfi ll protection.
The Medicollect collecting system for use with Medicollect bag removes the need to autoclave and associated costs.
The Medicollect bags are also available with Super Absorber which turns fl uid into the gel which helps
reduce the risk of contamination during handling and transportation.
FEATURES / BENEFITS
• Easy to clean
• Simple and easy to use
• Fitted with overfill protection
• Bag version fitted with a fi lter
• Bag version available with Super Absorber
TECHNICAL DATA
Collecting jar for use with bags Collecting jar for use without bags
(Snap-on lid) (Screw lid)
Jar: Polycarbonate (121°C) Polycarbonate (121°C), Polysulfone (134°C)
Polycarbonate,
Lid: Polypropylene
Material Thermoplastic elastamer (TPE)
Hose nipple: Nickel plated brass
Polyethylene
Bag: −
Anti-overflow arrangement: Integrated Ball float
Connector on source side Hose adapter for internal 09 Hose adapter
Connector on patient side: Hose adapter for internal 09 Hose adapter
Maximum applicable vacuum: -95 kPa (-950 mbar)
Minimum applicable vacuum:* -20 kPa (-100 mbar)
Maximum flow rate: 70 l/min 100 l/min
Scale interval: 50 ml (1000), 100 ml (2000) −
Storage conditions: -40 ± 2°C to +60 ±5°C and 10 to 100% RH
Operating conditions: -18 ± 2°C to +40 ±5°C and 40 to 70% RH
Medical device in risk class I
Regulatory status: + accessory following Medical Device Regulation (EU) 2017/745
EN ISO 10079-3 – Medical Suction Equipment.
Legal manufacturer: GCE, s.r.o., Žižkova 381, 583 01 Chotebor, Czech Republic
* for the correct function of the overfilling protection device
MEDICOLLECT
By connecting a vacuum source to the Medicollect collection container, it is possible to aspirate fl uids
and / or solid particles from the patient‘s body cavities. The liquid and / or solid particles are collected in a
collecting container or in a bag of the collecting container until it is completely fi lled or until the suction is
interrupted by the operator.
ACCESSORIES
Item no. Description Size (H×W×D) Weight Autoclave Durability
325111945 Medicollect Bag 1000 230x180x80 approx. 30 g − −
325111946 Medicollect Bag 2000 300×180×80 approx. 30 g − −
0727605 Medicollect Bag 1000 (SA*) 230×180×80 approx. 35 g − −
0727606 Medicollect Bag 2000 (SA*) 300×180×80 approx. 40 g − −
325111945 * super absorber
325111946
COLLECTING JARS FOR USE WITHOUT BAGS (SCREW LID)
Item no. Description Size (H×W×D) Weight Autoclave Durability
K291530 Medicollect 2000 Screw lid 302×135×130 545 g 121°C / 15´ 30 Autoclave cycles
K291540 Medicollect 1000 Screw lid 236×115×110 340 g 121°C / 15´ 30 Autoclave cycles
K291620 Medicollect 2000 Screw lid 302×135×130 545 g 134°C / 18´ 30 Autoclave cycles
K291619 Medicollect 1000 Screw lid 236×115×110 340 g 134°C / 18´ 30 Autoclave cycles
K291530 K291620
K291619
ASSOCIATED EQUIPMENT
Item no. Description
373234593 Suction hose adapter 9 mm
373234931 T-connection for suction
325113237 Silicone hose; 25m; 6×12
325113237
9435440 Silicone hose; 5m; 6×12
373234593
9435450 Silicone hose; 1m; 6×12
K006968 Manual vacuum stop device, non sterile for adults
0727435 Universal rail clamp 10 x 25 mm and 10 x 30 mm for T-slide application
0727435
K006968
TROLLEYS
Art. Nr. Description
325396136
ACCESSORIES
Art. Nr. Description Trolley
TECHNICAL DATA
Gas pressure O2, air, N2O, vacuum, CO2, N2O/O2 and Air 800
Wall 3 mm
Hardness (Shore A) 88 ± 5
Operation temperature - 20 °C to + 60 °C
Dimension
Art. Nr. Description Denomination Environment Lenght (m)
(inner/outer)
0731001 MediConnect NF O2 Antistatic 6.7/12.7 0,5
Low-pressure hoses are intended for the transfer of medical gases or vacuum, and/or the connection of two
medical products (e.g. a pressure regulator with an emergency ventilator).
> Maximum working pressure 14 bar.
> The hoses for medical breathing oxygen and nitrous oxide are marked with the chemical denomination of
the gas.
> The medical breathing air hose is marked ”Air”.
> Warranty time 2 years
Hospital Ward Equipment ⁄ Medical catalogue | 23
GCE HEALTHCARE
GAS PROBES
ANGLED PROBES
GCE gas probes are designed to provide a safe method of connecting medical equipment intended to deliver medical gases to
patients or to power gas driven medical devices.
GCE gas probes are components of various medical devices (e.g pressure regulators, hose assemblies, fl owmeters, lung
ventilators , anaesthesia workstations, etc ). They form a connection between the supply source and the medical device. Gas and
pressure specifi c connections are required to ensure the intended medical gas is safely delivered to the patient.
Numerous country specifi c standards are available, with 6 di erent versions to cover all possible requirements.
FEATURES / BENEFITS
• easy handling
• secure grip due to tallied body
• space-saving
• with anti-twist rotating hose barb as standard
STANDARD
Anti -twist hose guide due to 360° rotatable Angeled probe with tallied body shape Highly visible gas type identifier
hose nipple for a safe, comfortable handling
ADITIONAL OPTIONS
NRV for controlled gas shock relief Swivelling angeled probe for individual space saving
positioning in gas outlets
GAS PROBES
TECHNICAL DATA
STRAIGHT, THREAD
Inlet pressure 3 - 5 bar
Max. working pressure 6 bar for breathing / 12 bar for tools drive / -0,9 for vac
MEDICLAMP
The Mediclamp is universal and can be applied to all applicable rail systems; 1: 10 x 25 mm, 2: 10 x 30 mm
FEATURES / FEATURES / ADVANTAGES / BENEFITS
> Easy to clean
> Light in weight
> Spring activated
> Ergonomic shape
> MRI compatible
> Fulfi lls the Ambulance standards
MOUNTING OF CLAMP
VARIABLE INSERTS
PERMANENT CONNECTIONS T-SLIDE
Special aluminum plate for permanent T-slide insert made from
connection with the products like MediEject, aluminum for T-plate application
Duoline etc.
SPK33810004 0727435
SPK33810006
RAILS
ACCESSORIES
Art. Nr. Description Material
325112959 Distance 20 mm Stainless Steel (pack of 5)
325112960 Washer D 40 mm Stainless Steel (pack of 5)
325112961 End protection Plastic
ACCESSORIES
Art. Nr. Description Material
180901001P End protection (10 pcs) Plastic
182037404P Washer (10 pcs) Aluminium
329000223 Distance 20 mm Aluminium
HOMECARE
GCE HEALTHCARE
TECHNICAL DATA
Battery 249 mm × 97 mm × 235 mm
Size (W×D×H)
(9.8" × 3.8" × 9.25")
Weight 2.5 kg (5.5 lbs) without carry bag
AC adaptor: 100-240V AC (+/- 10%) 50-60 Hz in, 24V DC, 5.0A out
Power requirements
DC adaptor: 11.5 - 16V DC in, 24V, 5.0A out
Purity 87% - 96% at all settings
Maximum oxygen discharge pressure 20.5 psi
ZEN-O
Zen-O™ portable oxygen concentrator delivers up to 2 litres of oxygen in either pulse or continuous mode.
Zen-O™ is manufactured to meet the exacting standards of the European Medical Device directive and the
United States Food and Drug Administration.
OXYGEN CONCENTRATOR
M50
The lightweight, ergonomically designed M50 oxygen concentrator makes life simpler for patients that
require Long Term Oxygen Therapy (LTOT).
The M50 delivers high concentration of oxygen with flow rates of up to 5 litres per minute. Featuring a
compact design, the M50 is easy to move, store and maintain - with just one minute needed to replace
any part to be maintained by the user. It also comes with an integrated oxygen sensing device for
monitoring oxygen purity levels, and a low oxygen purity alarm to alert the patient if there is low oxygen
supply.
FEATURES
> Lightweight - a mere 15,2 kg
> Quiet operation
> Sleek, compact design with handle for easier movement
> Adjustable flow rate up to 5 litres per minute
> Fire safe oxygen outlet port for added security
> Low oxygen purity alarm for peace of mind
> Compatible with various bubble humidifiers
> Two year warranty
501010261 2 Stage Paper filter suitable for M50, OC-E80 and OC-E100
TECHNICAL DATA
OC-E SERIES
The OC-E80 and OC-E100 oxygen concentrators provide patients with up to 8 litres and 10 litres per minute
respectively. Thanks to the large LCD screen, patients can see a three colour bar oxygen concentration
reading to monitor supply levels in real time. The sleek design provides easy handling and, as one of the
quietest devices available, OC-E Series concentrators minimise disruption to patients’ day-to-day life.
FEATURES
> Sleek design for easy use
> Real time oxygen concentration display
> Large colour LCD screen
> Quiet operation
> Compatible with various bubble humidifiers
> Two year warranty
501010261 2 Stage Paper filter suitable for M50, OC-E80 and OC-E100
Weight 27 kg 27 kg
510W/230V/50Hz 650W/230V/50Hz
Average Power Consumption
380W/120V/60Hz 730W/120V/60Hz
Two stage paper filter Power failure, high pressure/low pressure. Low Oxygen Concentration,
Safety:
temperature, current overload or line surge shutdown, thermal switch,
Alarms & Features
40psig pressure relief valve
Air filter
(Located behind the cover)
ACCESSORIES
Art. Nr. Description
14090535
14090631
Settings Manual/Automatic
Low battery
No inhalation
Dimensions
Height 101 mm
Width 85 mm
Depth 32 mm
Ambient temperature
Relative humidity 25 % to 95 %
REGULATORY STATUS
The device meets the requirements of the Medical Device Directive 93/42/EEC relating to medical
devices, Class IIa.
TECHNICAL DATA
Quality assurance
1 137 6.51 3.25 2.17 1.42 1.08 2.17 1.08 0.46 0.34 0.23
1.7 137 11.38 5.49 3.52 2.54 1.56 3.53 1.56 1.18 0.58 0.39
2 137 13.42 6.51 4.34 3.25 2.17 4.34 2.17 1.31 1.09 0.46
2.7 137 18.29 9.14 6.09 4.37 3.04 6.10 3.05 2.03 1.32 1.02
9.4 137 64.23 32.11 21.27 16.05 10.43 21.28 10.44 7.09 5.22 3.35
0.5 200 5.00 2.30 1.40 1.15 0.50 1.40 0.50 0.33 0.25 0.17
1 200 10.00 5.00 3.20 2.30 1.40 3.20 1.40 1.07 0.50 0.33
1.7 200 17.00 8.30 5.40 4.15 2.50 5.40 2.50 1.53 1.25 0.57
2 200 20.00 10.00 6.40 5.00 3.20 6.40 3.20 2.13 1.40 1.07
2.7 200 27.00 13.30 9.00 6.45 4.30 9.00 4.30 3.00 2.15 1.30
9.4 200 94.00 47.00 31.20 23.30 15.40 31.20 15.40 10.27 7.50 5.13
NASAL CANNULA
Art. Nr. Description
14112132 Adult nasal cannula (2.1m), non-flared tip. Flow rate 1-7 LPM. Pack of 50
14090510 Adult nasal cannula (2.1m), flared tip. Flow rate 1-7 LPM pack of 50
14090513 Adult nasal cannula (2.1m). Flow rate 1-4 LPM. Pack of 50
WATER TRAPS
Art. Nr. Description
7001 Oxygen tubing water trap with threaded end cap. Pack of 25
MEDIWET
Art. Nr. Description Volume Connection Material
TECHNICAL DATA
Contents Only sterile water or boiled cold water
Dimensions Height 190 x 67 Width (incl hose nipple) x 57Ø mm
Weight Polysulphone version 115 g
Capacity 200 ml of water
Consumption 6 ml of water per hour at a gas flow of 10 l/min at 20 ºC
Jar autoclavable at 134 °C: polysulphone
Jar autoclavable at 121 °C: polycarbonate
Lid and outlet hose nipple: polypropylene
Material Inlet nut: Chromed brass
Diffuser: Polyethylene
Hose, flat gasket: Silicone
O-ring: EPDM
Outlet connection Tapered hose nipple for hose 6x9 mm (recommended length 2 m)
Cleaning Water, non abrasive detergent. N ever use solvents.
Autoclave Polysulphone version 134 ºC for 5 minutes
Special case Diffuser: exchange at every cleaning - do not sterilise!
Check that the humidifier is whole and air tight before use. Monthly exchange of
Maintenance gaskets is recommended. Exchange the diffuser when its microperforations no
longer exist.
Final product:
MW - Mediwet Brand:
G - GCE standard product
Jar volume (ml): S - special product variant
200
Autoclavable
temperature: Inlet connector (female):
Polycarbonate - 121°C M12x1.25F - female
Polysulphone - 134°C 9/16F - female
G3/8F - female
G1/4F - female
HUMID VENT 2
Art. Nr. Description
TRACH VENT
Art. Nr. Description
> Sterile
> Medium Hygroscopic and bacteriostatic microwell paper
> Suitable for adults and children
TRACH-VENT+
Art. Nr. Description
> Sterile
> Medium Hygroscopic and bacteriostatic microwell paper
> Moisture loss 8.4 mg H2O/l at Vt 500 ml
> Moisture Output 29.2 mg H2O/l at Vt 500 ml
EMERGENCY EQUIPMENT
GCE HEALTHCARE
Gas O₂ /N₂O
Dimensions 50 × 50 × 63 mm
Expiratory resistance Cracking pressure at zero flow +0,35 kPa at 120 l/min
Gas O₂
Weight 1300 g
Dimensiones 120 × 50 × 70 mm
Weight 175 g
Inspiratory resistance
AMBULANCE PANEL II
AMBULANCE PANEL II
The Ambulance Panel II is designed for permanent installation and use in road ambulances. Long
experience in this market has resulted in a product that gives the customer a lot of options and flexibility.
Due to the design and functions there are a big range of variants available. The functions and variants can
be combined upon customer’s request.
The modular concept means that same components can be used for many variants, resulting in short lead
times and full flexibility. The Ambulance Panel II is designed with same Shape for recessed and exposed.
RECESSED
Recessed mounting is when the ambulance panel is mounted in the wall. The inlet could either be mounted at
the end or at the back depending on the wall construction and the space behind the wall. The only part that
will be outside the wall is the quick connector.
EXPOSED
Exposed mounting is when the ambulance panel is mounted on the wall in the ambulance. The inlet could as
the recessed, also be located at the end or at the back. When using the end inlet the hoses are fastened on
the wall and fully visible. For some countries and regulations this is a must and the Ambulance Panel II has
the features complying to these requests. The panel is fastened easily by 4 screws. Depending on the wall
construction the fastening could either be done by a counter fastening frame or nuts and washers.
TECHNICAL DATA
Connection: G3/8"
Sketch of expossed mounted panel Sketch of recessed mounted panel Sketch of double recessed mounted panel
AMBULANCE REGULATORS
The Ambulance Panel System can be delivered with two types of ambulance regulators, Medireg and
Varimed. The Ambulance Panel System can also be used with combivalves*. Generally Medireg is for lower
capacity and Varimed for higher capacity and this capacity is determined by the equipment that is used in
the ambulance. The ambulance regulators are designed to withstand the conditions in the road ambulance
Medireg and are manufactured according to EN ISO 10524-1. The ambulance regulators are delivered with either
standard 3/8” connections or quick connectors. The Varimed and Medireg can be delivered with electronic
signal gauge to be connected to the ambulance monitoring or gas monitoring systems.
* Some restrictions regarding pressure monitoring when Ambulance Panel System includes electronic monitoring system.
Varimed
AMBULANCE HOSE
TECHNICAL DATA
Gas: O2, Air, N2O/O2
Cylinder pressure: 20-200 bar
Working pressure/flow: see separate information about regulators
Capacity Medireg: Recommended max 2 QC outlets
Capacity Varimed: Recommended max 5 QC outlets
QC outlet variants: SS, DIN, NF, BS, UNI, CZ
Capacity QC: 60 l/min*
Capacity FLS: 0-15, 0-25 l/min
Capacity suction: 25 l/min
Connection: G3/8” (female)
Connection hoses: G3/8”, G1/4”, SS, DIN, NF, BS, UNI, CZ
Hoses: antistatic, stainless steel reinforced
Working Temperature: -20C - +60C
Standards: EN 1789:2008
Manufacturer: GCE, s.r.o, Žižkova 381, 583 01 Chotebor, CZ
AP II O2 1x UN E S M O FS EJ G
Brand:
Ambulance Panel II G - GCE standard product
S - special product variant
Gases: Suction Ejector
O2 - with EJ (YES), without EJ (NO)
AIR
ENT (O2/N2O)
N2O Flowselector
VAC - with FS (YES), without FS (NO)
CG (O2, AIR, VAC, ...) Outlet
- with O (YES), without O (NO)
Number of QC Gauge
outlets: - with M (YES), without M (NO)
1
2 Type of QC standard: Manual switch
3 - with S (YES), without S (NO)
SS = Swedish standard
4
UN = Italian standard R = Recessed
NF = French standard DR = Double Recessed
DN = German standard E = Exposed
BS = British standard
CZ = Czech standard
DS = DISS
VALVES
GCE HEALTHCARE
USER FRIENDLY
> Suitable for use in Homecare, Emergency and Hospital applications
> Easy read Flow Selector and Gauge
> Shut off valve with clear open /closed status colour coded marking
> Ergonomic guard design allows easy handling of the cylinder package by all users
> Compact and lightweight design less than 1150 gr
> Easy clean guard material
HIGHEST SAFETY
> Testing in accordance with the following standards: EN ISO 10524-3, EN ISO 10297, ASTM G175 Pill test
> CE marked CE 2460 in accordance with MDD 93/42/EECand TPED 2010/35/EU
> Phthalate and Halogenated polymer free components
> MRI compatible up to Tesla 3
> For use with medical Oxygen, O2/N2O, Air, He/O2 up to 300 bar working pressure
TECHNICAL DATA
Gas O2, O2/N2O, Air, He/O2
Inlet pressure Up to 300 bar (4500 psi)
Outlet pressure 3,6 to 5,5 bar - acc. to EN ISO 10524-3 (or per customer specification)
RPV closing >3 bar
PRV opening and re-closing >5.5 bar
Flow ranges 0-6, 0-15 and 0-25 lpm
Metallic parts (gas wetted): brass
Elastomers: EPDM, silicon, PUR
Materials
Plastics: PA66, PEEK
* Standard Combivalve (flow control unit 0 - 15 l/min, flow outlet, DIN quick coupling pressure outlet) with guard.
All technical data are given for information only and are subject to modifications by the manufacturer.
MEDIVITOP®
MediVitop® is a pressure regulator integrated with a cylinder valve for use on medical gas cylinders.
USER FRIENDLY
> Suitable for use in Homecare, Emergency and Hospital applications
> Easy read Flow Selector and Gauge
The clever > Shut off valve with clear open /closed status colour coded marking
single touch solution > Ergonomic guard design allows easy handling of the cylinder package by all users
> Compact design, weight 1350 g
> Easy clean guard material
HIGHEST SAFETY
> Testing in accordance with the following standards: EN ISO 10524-3, EN ISO 10297, ASTM G175 Pill test
> In accordance with MDD 93/42/EEC and TPED 2010/35/EU
> Phthalate and Halogenated polymer free components
> MRI compatible up to Tesla 3
> For use with medical Oxygen, O2/N2O and Air up to 300
> bar working pressure
TECHNICAL DATA
Gas O2, O2/N2O, Air
Inlet pressure Up to 300 bar (4500 psi)
Outlet pressure 3,6 to 5,5 bar - acc. to EN ISO 10524-3 (or per customer specification)
RPV closing >3 bar
PRV opening and re-closing >5.5 bar
Flow ranges 0-15 and 0-25 lpm
Metallic parts (gas wetted): brass
Elastomers: EPDM, silicon, PUR
Materials
Plastics: PA66, PEEK
Springs (gas wetted): CuBe2, CuSn6
Dimensions (h×w×d) Height: 175 mm; Width: 141 mm; Depth: 123 mm
Weight* 1350 g
Inlet stem Tapered or parallel threads (17E, 25E, M18×1,5 or per customer specification)
Filling port ISO 5145, or per customer specification
Complies with MDD 93/42/EEC
Complies with TPED 2010/35/EU
Complies with EN ISO 10524-3
Regulatory status
Complies with EN ISO 10297
Complies with EN 1789
Complies with ASTM G175
Production in accordance with EN ISO 9001 and EN ISO 13485
Classification IIb
Manufacturer: GCE s.r.o, Žižkova 381, 583 01 Chotěboř, Czech Republic
* Standard Combivalve (flow control unit 0 - 15 l/min, flow outlet, DIN quick coupling pressure outlet) with guard.
All technical data are given for information only and are subject to modifications by the manufacturer.
RELATED PRODUCTS
Art. Nr. Description
0727421 Bed hanger (10 pcs)
0727427 Bed hanger with black plastic sleeves (10 pcs)
0727418 Humidifier holder 9/16 (10 pcs)
14112760 Connecting hose for humidifier holder 40 cm (1 pce)
0727422 Valving tool (1 pce)
on request Filling adaptor
Bed hanger Bed hanger - plastic Humidifier holder Connecting hose Valving tool Filling adaptor
Brand:
Final product: G - GCE standard product
MV – MediVital S - special product variant
MT - MediVitop
Outlet connector:
Nothing – standard fir tree
9/16 (including hose nipple)
Flow (lpm):
6
15 Outlet connector (QC):
25 AS – AS2902
Nothing* Gas: BS – BSI
O2 CZ – Czech
AIR DN – DIN
O2/N2O DS – DISS
He/O2 Inlet connector: Filling port connector:
NF – NF 90-116
0.750 – 0.750-16 UNF W24x2
NS – NIST
1.125 – 1.125-12 UNF W30x2
ON – Oxifar
14NGT - ½ 14 NGT W27x2
SS – SS8752430
17E – 17E CGA280
UN – UNI 9507
25E – 25E CGA540
Nothing – if no QC
18T – 18T CGA577
PV - purge valve (MT)
M18 – M18x1,5 G3/4
M30 – M30x2 G5/8
G5/8B
G5/8NL
G1/4
SI22 – SI22,91
W21.7
W21.8
TECHNICAL DATA
Materials Chrome plated brass (cylinder valves), Nickle plated brass (pin-index valves)
Burst disc 190, 216, 250, 300 bar, for CO2 and N2O, other gases optional
Classification IIb
OPTIONS OPTIONS
> Residual pressure device (RPD) > Residual pressure device (RPD)
> Burst disc > Burst disc
> Dip tube > Dip tube
> Hand wheel in different colours
> Customer logo on the hand wheel cap
OPTIONS
> Residual pressure device (RPD)
> Burst disc
> Dip tube
> Hand wheel plastic or aluminium
> Hand wheel with space for RF chip
> Customer logo on the hand wheel cap
SPECIFICATION
MC25 INCLUDES THE FOLLOWING COMPONENTS:
> Gas manifold MC25
> Gas alarm including power supply
> Evacuating kits for collecting pipe
> Shut-off valve for distribution line
> HP filters
> Collecting pipe for 2×1 cylinder
FOR A COMPLETE MC25 MANIFOLD ADD:
> High pressure collecting pipe set (high pressure valves, filters and non-return valves)
> High pressure hoses with safety wire
> Cylinder retaining brackets (included in gas cylinder collecting pipe set)
> Gas name sign
(For more information, please see accessories pages 17–21)
TECHNICAL DATA
Gases: O2, Air, N2, N2O, CO2, (all medical gases)
Nominal flow: 25 m3/h
Inlet nominal pressure: 200 bar (20 000 kPa)
Outlet nominal pressure: 4,5 bar (setting range 0,5–6 bar)
Inlet connection: W21,8×1/14"M
Outlet connection: G1/2"M + soldering piece pipe ø 10, ø 15 mm
Outlet pressure relieve valve: 6,8 bar
Intermediate pressure relieve valve: 17 bar
Pressure relieve valve pipe dimension: ø 15 mm
Regulatory status: Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
BASIC DIMENSIONS
450
Note!
485
Measurements in mm.
MM40 HP unit manifolds together with a stabilizer should be used as second and third source of gas in
systems with liquid gas tank. For hospitals without liquid gas tanks it is possible to use manifold MM40
HP unit together with a stabilizer as first and second source, and in connection with a third source (MM90
Standby) it will provide a final solution to fulfill ISO 7396-1 and national installation standards.
Manifolds are supplied with an alarm system which increases safety to maximal level and informs
the hospital personal about each non standard situation.
Gas Alarm is a standard accessory. The gas alarm gives visual and audible indications.
It acts as a surveillance system and sounds the alarm when the following happens
SPECIFICATION
MM40 INCLUDES THE FOLLOWING COMPONENTS:
> MM40 HP unit Manifold
> Gas alarm
> Purge valves
> HP filters
> Shut-off valve for distribution line to stabilizer
FOR A COMPLETE MM40 HP UNIT MANIFOLD ADD:
> Collecting pipe set (high pressure valves, non-return valves and high pressure components)
> Cylinder retaining brackets (included in gas cylinder collecting pipe set)
> High pressure hoses with safety wire
> Plug for close collecting pipeline
> Gas name sign
> Stabilizer
(For more information, please see accessories pages 17–21)
Item No. Denomination Gas PRV Signalization
0732307 MM40 – HP unit 2×1 O2, Air, N2 Standard Alarm TMA6703
0732308 MM40 – HP unit 2×1 O2, Air, N2 With pull-ring Alarm TMA6703
0727334* MM40 – HP unit 2×1 O2, Air, N2, N2O, CO2 Standard –
0732310 MM40 – HP unit 2×1 N2O, CO2 Standard Alarm TMA6703
0732311 MM40 – HP unit 2×1 N2O, CO2 With pull-ring Alarm TMA6703
0727371** MM40 – HP Unit 2x1 O2, Air, N2 Standard 4-20mA
*basic version without electric sensors
**ISO standard version without Leaking reserve as requested for HB370
TECHNICAL DATA
Gases: O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow: 40 m3/h
Inlet nominal pressure: 200 bar
Outlet nominal pressure: 12 bar (setting range 9–16 bar)
Inlet connection: W21,8×1/14"M
Outlet connection: G1/2"M + soldering piece pipe ø 10, ø 15 mm
Pressure relieve valve: 17 bar
Pressure relieve valve pipe dimension: ø 10 mm
Purge valves connection: W21,8×1/14"M
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Regulatory status:
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
BASIC DIMENSIONS
365
455
Note!
Measurements in mm.
SPECIFICATION
LINE REGULATOR INCLUDES THE FOLLOWING COMPONENTS:
> Line regulator
TECHNICAL DATA
Gases: O2, Air, Air–800, N2, N2O, CO2 (all medical gases)
Nominal flow: 40 m3/h
Inlet nominal pressure: 16 bar (1600 kPa)
Outlet nominal pressure: 4,5 bar (setting range 0,5–10 bar)
Inlet connection: G1/2"M + soldering piece pipe ø 12 mm
Outlet connection: G1/2"M + soldering piece pipe ø 12 mm
Pressure sensors: Optional (Pressure switches; Transmitter 4–20 mA)
Emergency QC inlet: Optional QC by national standards
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Regulatory status:
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present HTM 02-01
BASIC DIMENSIONS
206
151
83 Note!
Measurements in mm.
SPECIFICATION
MM40 STABILIZER INCLUDES THE FOLLOWING COMPONENTS:
> MM40 Stabilizer Manifold
ACCESSORIES
Item No. Denomination
COM001002 Lockable cover
For more information please contact our sales and product support.
Note!
Measurements in mm.
The gas alarm is able to communicate with other equipment through relays.
SPECIFICATION
MM90 INCLUDES THE FOLLOWING COMPONENTS:
> MM90 HP unit Manifold
> Gas alarm
> Purge valves
> HP filters
TECHNICAL DATA
Gases: O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow: 90 m3/h
Inlet nominal pressure: 200 bar (20 000 kPa)
Outlet nominal pressure: 9 bar (setting range 9–15 bar)
Inlet connection: W21,8×1/14"M
BASIC DIMENSIONS
Outlet connection: G3/4"F + soldering piece pipe ø 22 mm
530
Pressure relieve valve: 16 bar
Pressure relieve valve pipe dimension: ø 10 mm
Purge valves connection: W21,8×1/14"M + soldering piece pipe ø 10 mm
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Regulatory status:
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
455
Note!
Measurements in mm.
760
60 | Central gas supply systems ⁄ Gas manifolds and stabilizers
GCE HEALTHCARE
Gas Alarm is a standard accessory. The gas alarm gives a visual and audible indication. It surveils and the
alarm sounds when the following happens:
The gas alarm is able to communicate with other equipment through relays.
SPECIFICATION
MM90 INCLUDES THE FOLLOWING COMPONENTS:
> MM90 HP unit Manifold
> Gas alarm
> Purge valves
> HP filters
TECHNICAL DATA
Gases: O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow: 90 m3/h
Inlet nominal pressure: 200 bar (20 000 kPa)
BASIC DIMENSIONS Outlet nominal pressure: 9 bar (setting range 9–15 bar)
530 Inlet connection: W21,8×1/14"M
Outlet connection: G3/4"F + soldering piece pipe ø 22 mm
Pressure relieve valve: 16 bar
Pressure relieve valve dimension: ø 10 mm
Purge valves connection: W21,8×1/14"M + soldering piece pipe ø 10mm
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Regulatory status:
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
455
Note!
760 Measurements in mm.
SPECIFICATION
MM90 STANDBY INCLUDES THE FOLLOWING COMPONENTS:
> MM90 STANDBY Manifold with pressure switches
> Purge valves
> HP filters
FOR A COMPLETE MM90 STANDBY MANIFOLD ADD:
> Gas alarm
> Collecting pipe set (high pressure valves, and non-return valves / high pressure components)
> Cylinder retaining brackets (included in gas cylinder collecting pipe set)
> High pressure hoses with safety wire
> Plug for close collecting pipeline
> Gas name sign
> Stabilizer
TECHNICAL DATA
Gases: O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow: 90 m3/h
Inlet nominal pressure: 200 bar (20 000 kPa)
Outlet nominal pressure: 7 bar (setting range 7–15 bar)
Inlet connection: W21,8×1/14"M
Outlet connection: G3/4"F + soldering piece pipe ø 22 mm
Pressure relive valve: 16 bar
Pressure relive valve pipe dimension: ø 10 mm
Purge valves connection: W21,8×1/14"M + soldering piece pipe ø 10 mm
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Regulatory status:
BASIC DIMENSIONS Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
363
245 Note!
318 Measurements in mm.
MC80 HP UNIT contains two regulators with pressure relieve valves and it is connected to two various
cylinder banks using high pressure hoses. When the cylinder bank, which has been connected for
operation, has been emptied the other duty side is automatically connected.
Together with MC80 HP UNIT shall always be used gas alarm providing all alarms in compliance with the
standard EN ISO 7396-1.
SPECIFICATION
MC80 HP UNIT INCLUDES THE FOLLOWING COMPONENTS:
> MC80 HP Manifold
For variants not listed, please contact our sales and product support.
TECHNICAL DATA
Gases: O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow: 200 m3/h
Inlet nominal pressure: 200 bar (20 000 kPa)
Outlet nominal pressure: 12 bar (setting range 10–16 bar)
Inlet connection: W21,8×1/14"M
Outlet connection: G3/4"F
Pressure relive valve: 17 bar
Pressure relive valve pipe dimension: ø 10 mm
Purge valves connection: W21,8×1/14"M + pipe ø 15 mm
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Regulatory status:
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
580
The gas reduction unit MC80 – STABILIZER must always be installed in compliance with the standards
EN ISO 7396-1 and the appropriate national standards.
SPECIFICATION
MC80 STABILIZER INCLUDES THE FOLLOWING COMPONENTS:
> MC80 Stabilizer Manifold
TECHNICAL DATA
Gases: O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow: 200 m3/h
Inlet maximal pressure: 20 bar (2000 kPa)
Outlet nominal pressure: 4,5 bar (setting range 0,5–6 bar)
Inlet connection: G3/4"F
Outlet connection: G3/4"F
Pressure relieve valve: 6,8 bar
Pressure relieve valve pipe dimension: ø 15 mm
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Regulatory status:
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
BASIC DIMENSIONS
450
580
Note!
Measurements in mm.
2. MC80 - STABILIZER
The stabilizer makes the operating pressure in the distribution system remain constant. The module
contains two regulators with pressure relieve valves. Since the gas pressure is reduced in two steps the
drop in pressure, when changing from the operating cylinder to the other bank of cylinders, is kept to a
minimum. The unit is prepared for connection to a liquid oxygen supply tank (LOX).
The product is either equipped with an alarm system or a sensor only (as stated in column "Alarm" in the
product table below).
SPECIFICATION
DUPLEX (MC80) INCLUDES THE FOLLOWING COMPONENTS:
> MC80 – HP Unit
> MC80 – Stabilizer
> Sensors 4-20mA. Alarm based on item in column "Signalization"
> Evacuating kits for collecting pipe
> Shut-off valve for the distribution line
> HP filters
FOR A COMPLETE DUPLEX (MC80) STANDBY MANIFOLD ADD:
> Collecting pipe set (high pressure valves, and non-return valves, high pressure components)
> High pressure hoses with safety wire
> Gas name sign
> Connection pipe 90 degree
> Extension pipes if needed
TECHNICAL DATA
Gases: O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow: 200 m3/h
TECHNICAL DATA - HIGH PRESSURE UNIT MC80
Inlet nominal pressure: 200 bar (20 000 kPa)
Outlet nominal pressure: 12 bar (setting range 10–16 bar)
Inlet connection: W21,8×1/14"M
Outlet connection: G3/4"F
Pressure relieve valve: 17 bar
Pressure relieve valve pipe dimension: ø 10 mm
Purge valves connection: W21,8×1/14"M + pipe ø 15 mm
TECHNICAL DATA - STABILIZER MC80
Inlet maximal pressure: 20 bar (2000 kPa)
Outlet nominal pressure: 4,5 bar (setting range 0,5–6 bar)
Inlet connection: G3/4"F
Outlet connection: G3/4"F
Pressure relieve valve: 6,8 bar
Pressure relieve valve pipe dimension: ø 15 mm
BASIC DIMENSIONS
450
450
580
Note!
Measurements in mm.
580
The gas reduction unit MC150 - STABILIZER must always be installed in compliance with the standards
EN ISO 7396-1 and the appropriate national standards.
SPECIFICATION
MC150 STABILIZER INCLUDES THE FOLLOWING COMPONENTS:
> MC150 Stabilizer Manifold
TECHNICAL DATA
Gases: O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow: 360 m3/h
Inlet maximal pressure: 20 bar (2000 kPa)
Outlet nominal pressure: 4,5 bar (setting range 0,5–6 bar)
Inlet connection: 2× G1 1/2"F+soldering piece pipe ø 35 mm
Outlet connection: 2× G1 1/2"F+soldering piece pipe ø 35 mm
Pressure relieve valve: 6,8 bar
Pressure relieve valve pipe dimension: ø 35 mm
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Regulatory status:
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
BASIC DIMENSIONS
600
830
Note!
Measurements in mm.
The regulator is mounted in the collection unit. Each inlet connection has a filter,
a non-return valve and a shut-off valve. This arrangement makes it possible to use one
cylinder at a time.
In order to obtain a stabile outlet pressure this gas manifold is equipped with a preset
two-stage regulator. On the high pressure side of the regulator there is a contact gauge
SPECIFICATION
SIMPLEX MMR INCLUDES THE FOLLOWING COMPONENTS:
> Gas cylinder manifold Simplex MMR
> Collecting pipe Manyflow block for three hoses
> Gas evacuation kits for collecting pipe
FOR A COMPLETE SIMPLEX MMR ADD:
> Gas alarm
> Pressure relieve valve with manual activation
> High pressure hoses with safety wire
> Cylinder retaining brackets (included in gas cylinder collecting pipe set)
> Gas name signs
> Pressure relieve valve
TECHNICAL DATA
Gases: O2, Air, N2, Ar, N2O, CO2 (all medical gases)
Nominal flow: 30 m3/h
Inlet nominal pressure: 200 bar (20000 kPa)
Outlet nominal pressure: 5 bar (setting range 4–5 bar)
Inlet connection: W21,8×1/14"M
Outlet connection: G3/8"M
Pressure relieve valve: 6 bar
Pressure relieve valve pipe dimension: ø 8 mm
Purge valves connection: W21,8×1/14"M+pipe ø 15 mm
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Regulatory status:
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
BASIC DIMENSIONS
340
520
Note!
Measurements in mm.
HANDLING
The high pressure hose should be transported, stored, installed and maintained according to Instruction of
Use. Maximum life time after installation is 5 years.
For variants not listed, please contact our sales and product support.
TECHNICAL DATA
Tube: Acid-proof Stainless Steel (AISI 316)
Plait: Stainless Steel (AISI 304)
Wire: Stainless Steel (AISI 304)
Nut and tightening material: Acid-proof Stainless Steel (AISI 316)
Case and Oetiker: Stainless Steel (AISI 304)
Complies with Medical Devices Directive 93/42/EEC
Regulatory status: Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN ISO 21969 (High Pressure Flexible Connection)
215191074P 260 1
215191075P 550 2
TECHNICAL DATA
Tolerance: ±5mm (unless otherwise stated)
Strength test: 440 bar
Leak test: 230 bar
Pressure rated: 230 bar
Inlet/Outlet connection: BS, PIN INDEX, AFRNOR, DIN, UNI, CZ, SS
Complies with Medical Devices Directive 93/42/EEC
BS EN ISO 21969
Requirements: BS EN ISO 407
Oxygen cleand to ISO 15001
CE marked
Each batch of tailpipes is supplied with a certificate of conformity
215191044
200059835P
0777111
CONNECTING COMPONENTS AND ACCESSORIES
Item No. Description Inlet Outlet Position
215191010 Connection pipe 90° W21,8×1/14“ female W21,8×1/14“ male 1
2 215191077 Blind plug W21,8×1/14“ 2
311325111032P Cu sealing gasket (package of 10 pcs), 18×12,7×1,5 3
1
0984018050 Washers 1/2"x 2 mm DIN6 (10pcs) 3
14016927P Washers 1/2"x 2 mm PLAS DIN6 (50pcs) 3
3 215191085 T-piece W21,8×1/14“ female 2xW21,8 H 4
4 215191126 S–pipe W21,8 H W21,8 M 5
215191080 End plug with nut W21,8 H W21,8 H 6
5 200059835P Coupling nut W21,8 RH W21,8 LH 7
9387930P Connecting tube ODNCO2 50 mm DIN6 W21,8 H W21,8 H 8
6 9451080P Connecting tube ODNCO2 150 mm 300 bar W21,8 H W21,8 H 8
9387860P Connecting tube ODNCO2 290 mm DIN6 W21,8 H W21,8 H 8
7 9387870P Connecting tube OXY/INERT 830 mm W21,8 H W21,8 H 8
ADVANTAGE
Our line valves use Nylon or PEEK sealing seats according to gases, pressures or service temperature. It
allows a smooth closing, without forcing, without the generation of metallic particles. They are designed
for 100,000 cycles and backed by more than 50 years of manufacturing and nearly one million valves
manufactured per year.
Item No. Description Inlet Outlet
9561271 Line valve DN4 W21,8 M Der W21,8 M Der
0765278 Line valve DN4 W21,8 M Izq. W21,8 M Der
0777111 SOV DN4 W21,8×1/14" W21,8×1/14"LH
BV777105 BV 300 DN8 W21,8×1/14" W21,8×1/14"LH
BV777096 Valve 300 Bar O2 DN8 W21,8 M Der W21,8 M Der
ADVANTAGE
The GGP250 is designed for pressures up to 300 bar ( 4000 psi ). The highly efficient heating system uses
a high pressure filter to increase the heat exchange surface. Safety is ensured by use of a thermostat and
H28054801
safety switch which ensures temperatures never rise too high to cause damage to downstream equipment.
The IP65 protection ensures total user / operator safety. The GGP 250 is a reliable product which is
insensitive to frequency variations, and is CE certified.
TECHNICAL DATA
Voltage/Power: 230 V - 50 Hz Working temperature (environment): 40°C, +/-3°C
Power: 250 W Temperature switch: 80°C, +/-5°C
Gases: CO2, O2, N2O Dimensions: 90 x 150 mm
Protection class plug: IP44 Weight: Aprox 2,3 kg
Auxiliary port
G1 / 4 "plugged
MM90 MM40
Manifold MM90 Standby with manifold for 4 cylinders, shut off valve and outlet connection
215191044
215191085
200059835P F27310015
0777111
Panel MM40 for 4+2 cylinders with shut off valve and outlet connection
215191010 215191010
9625640 9625650 9625640
215191044 215191044
200059835P 200059835P
0777111 0777111
TECHNICAL DATA
Evacutaion flow: 200 m3/h
Evacutaion outlet pipe: ø 15 mm
Relief pressure: 6,8 bar
Material: brass, copper, stainless steel, rubber
Pressure class: PN16
Degreased for Oxygen use
Regulatory status:
no CE-marking
PRESSURE MONITOR
PRESSURE MONITOR
The pressure monitor makes sure that the lower distribution pressure for nitrous oxide compared to oxygen
is kept.
The lower nitrous oxide pressure will be maintained according to standards even when the emergency
supply is used through quick connectors or central emergency supply. The pressure monitor is equipped
with a digital pressure monitor unit monitoring the current gas pressures, and giving all the visual and
acoustic alarms required by standards. The signal to the gas alarm comes from pressure transmitters. The
visual and audible signals can be sent to a manned area, if it is required.
The following gases are under surveillance: breathing Oxygen, Nitrous Oxide, Air and instrument Air.
The alarm is indicated by an acoustic and visual signal at the same time as the exact cause of the alarm
is written on the display. This happens if the gas pressure rises above or sinks below the set maximum or
minimum limits respectively. The pressure monitor is also equipped with a bayonet coupling for breathing
oxygen, nitrous oxide, breathing air, and instrument air. When necessary, it is possible to connect spare
gas to these.
TECHNICAL DATA
Gases: O2, N2O, Air, Air–800, CO2, N2, VAC (all medical gases)
Number of gases: 2 to 4 valves (DN15)
4–5 bar (breathing gases)
Working pressure: 7–10 bar (instrumental gases)
(-0,4)–(-0,9) bar (vacuum)
Maximum pressure: 16 bar
Safety regulator capacity at 3 bar: 150 l/min
Tube dimension: ø 15×1 mm, ø 22×1 mm
Emergency QC inlets: QC by national standards
Pressure gauges: 0–16 bar
Pressure sensors: Transmitters 4–20 mA
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Regulatory status:
Complies with EN 60601-1-2 (Electromagnetic compatibility)
and present SIS HB 370
78 | Central gas supply systems ⁄ Pressure monitor, pressure watch, gas alarms and valves
GCE HEALTHCARE
BASIC DIMENSIONS
WALL
WALL
Central gas supply systems ⁄ Pressure monitor, pressure watch, gas alarms and valves | 79
GCE HEALTHCARE
AVSU SLIDE
According to the ISO 7396-1 so called Area Valve Service Units (AVSU) have to be installed and operationed at every functional area in medical
facility where patients are treated.
The AVSU SLIDE in pressure watch or shut-off valve box variants is used for monitoring and controlling the gas pressure from the source to the
point of use on department and/or room level. In case of malfunction, maintenance or repair of the Medical Central Gas System the gas fl ow
towards the point of use can be interrupted by closing the valves inside of the AVSU SLIDE. In case of emergency the emergency inlet points
AVSU SLIDE allow external feeding of life supporting gases. The integrated alarm will notify the medical and technical staff of too low or high
pressure in the Medical Central Gas System so they can take the appropriate actions.
The AVSU SLIDE can be equipped with a standard powder coated metal panel or a stylish
TECHNICAL DATA
O2, Air, N2O, CO2, O2/N2O, N2, Ar, UTL
(Medical) gases: AIR800 for driving surgical tools
VAC
Piping: DN15, DN22, DN28
Primary power supply: 230 V 50Hz AC (integrated power supply)
Secondary power supply: 24 V
Microswitch (indication on door and/or valve)
Signal device: Pressure Switch
Pressure transmitter 4 – 20mA
Gas Alarm Panel (Touch alarm)
Alarms: TMA9705 (LED alarm)
TMATFT03 (Touch alarm))
Complies with Medical Devices Directive 93/42/EEC
Regulatory status: Complies with EN ISO 7396-1 (Central Gas Supply Systems)
present SIS HB 370
Classification: IIb
80 | Central gas supply systems ⁄ Pressure monitor, pressure watch, gas alarms and valves
GCE HEALTHCARE
tempered glass front panel. The tempered glass panel front can be provided with individualized colours or prints to meet the architectural
requirements of today’s modern hospitals.
The AVSU SLIDE allows a great number of individual configurable parts to meet most of the specific needs and requirements of medical facilities,
such as:
• Number of gases
• Pipe diameters
• Front cover design
• Alarm unit models
• Fixed emergency piping
• QC inlet points (single or double)
• Universal box for recessed and exposed installation
• Small box sizes for space saving (100mm per gas line)
BASIC DIMENSIONS:
AVSU WITHOUT ALARM AND WITHOUT SENSORS
Gases L1 L2 L3 L4 L5 L6
1 or 2 310 310 118 200 93 280
3 310 407 118 297 93 280
4 310 503 118 393 93 280
5 310 600 118 490 93 280
6 310 696 118 586 93 280
Gases L1 L2 L3 L4 L5 L6
1 or 2 360 310 118 200 93 330
3 360 407 118 297 93 330
4 360 503 118 393 93 330
5 360 600 118 490 93 330
6 360 696 118 586 93 330
Central gas supply systems ⁄ Pressure monitor, pressure watch, gas alarms and valves | 81
GCE HEALTHCARE
Gases L1 L2 L3 L4 L5 L6
1 or 2 450 310 118 200 93 420
3 450 407 118 297 93 420
4 450 503 118 393 93 420
5 450 600 118 490 93 420
6 450 696 118 586 93 420
Gases L1 L2 L3 L4 L5 L6
1 or 2 650 310 153 200 128 620
3 650 407 153 297 128 620
82 | Central gas supply systems ⁄ Pressure monitor, pressure watch, gas alarms and valves
GCE HEALTHCARE
USED FOR:
> Pressure watches / Shut of valve boxes with sensors (4-20mA).
> Manifolds with sensors (4-20mA ) (Updated MC80)
> Any equipment using sensors (4-20mA) or NO/NC technology
ACCESSORIES
Item No. Denomination
SPK36410012 CABLE AND SOFTWARE FOR TMA9705
SPARE PARTS
Item No. Denomination
SPP36410008 Sp-ext.Battery for TMA9705-ETH
SPP36410004 Sp-alarm TMA9705-ETH GCE
Installation box SPP36410007 Sp-trafo 24VAC for alarms
SPP36410010 Sp-box exp.Ip40 for TMA9705-ETH
SPP36410012 Sp-box exp.Ip65 for TMA9705-ETH
SPP36410014 Sp-box recessed for TMA9705-ETH
TECHNICAL DATA
Power supply: 24 VAC 50/60Hz +-15% / 6 VA max
12 VDC External Backup Battery (Optional)
Trafo Inputs 16 Digital inputs (NO/NC)
6 Analogue inputs (4-20mA)
Output relay: 2 (1A Max)
Output power: +12V
Data transfer: RS485 Modbus RTU; Modbus TCP/IP
Cable use power: 0.75mm
Cable use switches: Modbus RTU
Alarm activation: Swedish, Norwegian, Danish, Finnish, English and Hungarian
Dimensions: 178mm x 160mm x 71mm
Housing: DIN box for 9 modules
ISO 7396-1 Medical gas pipeline systems
Standards:
ISO 60601-1-8 Medical electrical equipment
BASIC DIMENSIONS
71
160
Note!
Measurements in mm.
178
Central gas supply systems ⁄ Pressure monitor, pressure watch, gas alarms and valves | 83
GCE HEALTHCARE
APPLICATIONS:
> Alarm system to control the pressure of gases by digital and analog inputs
> Data are shown directly on the TFT touchscreen display
TECHNICAL DATA
Dimension: Box for external IP66
Dim: 200x120x60 mm
Compatible inputs: Up to 6 analog inputs (4-20 mA/RS485)
Up to 4 ON/OFF inputs
Compatible outputs: Ethernet, ModBus RS485
100-240 VAC / 50-60 Hz
Power supply:
24 VAC / 50-60 Hz
84 | Central gas supply systems ⁄ Pressure monitor, pressure watch, gas alarms and valves
GCE HEALTHCARE
network. The GASMASTER can manage up to 30 devices connected by a twisted shielded cable using
the MODBUS protocol. The configuration can be easily modified using the integrated touch monitor,
each input of each device can be configured in all their parameters (names, LEDs, sound signals, modbus
address, relays ...). The GASMASTER has also been designed to be integrated into a BMS (Building
Management System) system for monitoring its status and any alarms present.
For this purpose, using the RJ45 socket and connecting an Ethernet cable, is possible to interrogate
the GASMASTER using the MODBUS TCP/IP protocol.
The GASMASTER is also equipped with 2 relays which can be activated and configured, also for the
purpose of reporting the status of the GAS MASTER to an appropriately configured remote station.
The GASMASTER is also equipped with a back-up battery able to support its functioning in case the main
electric network ceases to supply current. The duration of this battery is limited to few hours.
The absence of current condition is however interpreted by the GASMASTER as an alarm condition, the
sound system will then operate and will cease once the main current source has been restored.
LEGEND CONNECTIONS
ModBus clamp
1 ModBus +
2 ModBus -
3 Not Used
4 Not Used
5 Not Used
Power Supply clamp
1 Power Supply 230V AC
2
NORMAL CONDITIONS:
When switched on, the system loads the stored configuration and performs a first cycle to check the
communication and the status of the peripherals. Once completed, the system shows the peripherals
connected.
Clicking on the desired peripheral button is possible to check the status of the same.
When one of the peripherals gives an alarm or when it is no longer reachable via MODBUS connection
cable, on the main page of the GASMASTER are shown the name of the device and which input is
in alarm. If several alarms or more inputs are alarmed simultaneously, the name of the alarms and
peripherals will appear sequentially.
To enter the configuration menu, press the corresponding button on the main screen and enter the
password. (default:"000000”).
Central gas supply systems ⁄ Pressure monitor, pressure watch, gas alarms and valves | 85
GCE HEALTHCARE
Pressing "Device Connected" button from the main menu you can enter the configuration
menu of the individual peripherals
Using the side scroll bar you can pass through the pages of the configuration and set the name of the
peripheral, Modbus address and bit stop, activation of the inputs, the association to the LEDs, relays and
ringtone.
Pushing the label of the input allows to set the transducer description, high threshold alarms, low
threshold alarms, fault alarms, ON OFF alarms.
Name of Device
Activating or
Deactivating
the device
MODBUS
address of
the device
The maximum distance that can be reached in optimal conditions for Modbus communication at 9600
baud is 1200m, but, due to interferences, it is suggested to reduce as much as possible this distance.
Pressing "Factory reset " button is possible to reset the device to the factory parameters
Important: be sure to exit the menu pressing the related pushbutton to avoid a device malfunctioning.
86 | Central gas supply systems ⁄ Pressure monitor, pressure watch, gas alarms and valves
GCE HEALTHCARE
recessed into a wall (fl ush-mounted) or mounted directly on the wall with an exposed box.
USED FOR:
> connect to the hospital’s BMS system via Modbus RTU, TCP/IP and a number of confi gurable digital outputs,
> add automatically confi gurable alarm replicators,
> monitor medical gas state from a mobile phone or other device with an internet browser with internet
access,
> record the pressure measurement history on a microSD card.
Item No. Denomination
SPK36410019 ALARM + trafo 24V + box exposed (PL/ EN/ RU/ FR)
SPK36410020 ALARM Scandinavian + trafo 24V + box exposed (EN/ SE/ DK/ NO/ FI)
SPK36410021 ALARM Scandinavian Replica + trafo 24V + box exposed (EN/SE/ DK/ NO/ FI
SPARE PARTS
Item No. Denomination
SPP36410022 SP alarm Scandinavian (EN/ SE/ DK/ NO/ FI)
SPP36410023 SP alarm Scandinavian Replica (EN/ SE/ DK/ NO/ FI)
SPP36410025 SP trafo 24V
SPP36410024 SP exposed box
TECHNICAL DATA
Gas: O2, N2, N2O, N800, N2O/O2 , H2/O2 , MA4, SA7, AIR, AIR800, AIR TECH, CO2 , VAC
Supply voltage: 24V DC +/- 10%
Power consumption: 10W
1-6 Programmable Channels
Inputs: 12x Digital lnput (Contact Manometers, Pressure Switches)
6x Analog lnput** (2-Wire 4-20mA Pressure Transmitters)
Communication with BMS: RS485 (MODBUS RTU)
Ethernet (MODBUS TCP/IP)
Potential Free Contacts
Integrated Web Server
History Record: microSD card slot (CSV)
Languages: PL/ EN/ RU/ FR/ SE/ DK/ NO/ FI
Alarm recessed dimensions: 115 x 190 x 70 mm
Alarm exposed dimensions: 144 x 220 x 80 mm
Alarm weight: 500g
IP protection class: IP40
Compliance with: RoHS Directive RoHS 2011/65/UE
Complies with Medical Devices Directive 93/42/EEC
Product classification: class IIb, in accordance with rule 9
1. Alarm
2. Mounting plate for expansion modules
3. I/O module
4. TCP communication module
5. Mounting ring.
6. Alarm housing for recessed installation
Central gas supply systems ⁄ Pressure monitor, pressure watch, gas alarms and valves | 87
GCE HEALTHCARE
USED FOR:
> Pressure watches / Shut of valve boxes with pressure switches or contact gauges.
> Manifolds with pressure switches or contact gauges
> Any equipment using NO/NC technology
Item No. Denomination
SPK36410005 Gas alarm TMA6703 GCE, trafo 24VAC, box exposed IP40
ACCESSORIES
Item No. Denomination
SPK36410011 CABLE AND SOFTWARE FOR TMA6703
SPARE PARTS
Item No. Denomination
SPP36410001 Sp-alarm TMA6703 GCE
Installation box
SPP36410007 Sp-trafo 24VAC for alarms
SPP36410009 Sp-box exp.IP40 for TMA6703
SPP36410011 Sp-box exp.IP65 for TMA6703
SPP36410013 Sp-box recessed for TMA6703
TECHNICAL DATA
BASIC DIMENSIONS
71
160
125 Note!
Measurements in mm.
88 | Central gas supply systems ⁄ Pressure monitor, pressure watch, gas alarms and valves
GCE HEALTHCARE
The non return valve unit consists of a non return valve (NRV) with
a flow direction arrow, lockable medical shut off valves, soldering
pieces, nuts and a gasspecific medical quick coupling (QC) for
medical breathing air. This design makes the NRV very easy to test.
The QC can also be used for checking the pressure, doing leak tests
and take gas samples. The NRV unit can also be delivered with QC
for instrumental air.
For more information please contact our sales and product support
TECHNICAL DATA
Opening pressure: 0,06 bar (6 kPa)
Pressure class: PN16
Degreasing for Oxygen use
Regulatory status:
no CE–marking
Central gas supply systems ⁄ Pressure monitor, pressure watch, gas alarms and valves | 89
GCE HEALTHCARE
SPARE PARTS
Item No. Denomination Valve Thread
A
TECHNICAL DATA
Gases: O2, Air, N2, Ar, N2O, CO2 (all medical gases)
Material valve housing: Nickel plated brass
Ball: Chrome plated brass
Stem: Nickel plated brass
Max working pressure: 40 bar (4000 kPa)
Tighten proof: (-1)–50 bar [(-100)–5000 kPa]
Complies with Medical Devices Directive 93/42/EEC
Regulatory status: Complies with EN ISO 7396-1 (Central Gas Supply System)
90 | Central gas supply systems ⁄ Termina Units Complies with EN 331 (Manually operated ball valves)
GCE HEALTHCARE
For variants not listed, please contact our sales and product support.
TECHNICAL DATA
Gases: O2, N2O, Air, CO2, N2, VAC
Dimensions: Height: 73 mm, Width: 73 mm, Depth: 63 mm
4–5 bar (breathing gases)
Working pressure: 7–10 bar (instrumental gases)
(-0,4)–(-0,9) bar (vacuum)
Maximum test pressure: 20 bar
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Regulatory status: Complies with EN ISO 9170-1 (Terminal units)
Complies with DIN 13260-2 (DIN gas specific connections)
present HTM 02-01
Note!
Measurements in mm.
SPARE PARTS
For variants not listed, please contact our sales and product support.
PEN version
TECHNICAL DATA
Gases: O2, N2O, Air, Air–800, CO2, N2, Ar, AGSS, VAC
Note! Dimensions: Height: 73 mm, Width: 73 mm, Depth: 63 mm
Measurements in mm. 4–5 bar (breathing gases)
Working pressure: 7–10 bar (instrumental gases)
(-0,4) – (-0,9) bar (vacuum)
Maximum test pressure: 20 bar
Complies with Medical Devices Directive 93/42/EEC
Split view Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Regulatory status: Complies with EN ISO 9170-1 (Terminal units)
Complies with SS 8752430 (SS gas specific connections)
present SIS HB 370 and HTM 02-01
For variants not listed, please contact our sales and product support.
PEN version
SPARE PARTS
Item No. Denomination Gas
SPK36810100 TU Quick coupler BSI AIR-800
SPK36810099 TU Quick coupler BSI O2/N2O
SPK36810098 TU Quick coupler BSI N2O
BASIC DIMENSIONS SPK36810097 TU Quick coupler BSI VAC
SPK36810096 TU Quick coupler BSI AIR
SPK36810095 TU Quick coupler BSI O2
TECHNICAL DATA
Gases: O2, N2O, O2/N2O, Air, Air-800, VAC
Dimensions: Height: 73 mm, Width: 73 mm, Depth: 63 mm
4–5 bar (breathing gases)
Working pressure: 7–10 bar (instrumental gases)
(-0,4)–(-0,9) bar (vacuum)
Maximum test pressure: 20 bar
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Regulatory status: Complies with EN ISO 9170-1 (Terminal units)
Complies with BS 5682 (BSI gas specific connections)
Note!
Measurements in mm. present HTM 02-01
For variants not listed, please contact our sales and product support.
SPARE PARTS
Item No. Denomination Gas
SPK36810106 TU Quick coupler CZ AIR-800
SPK36810105 TU Quick coupler CZ CO2
SPK36810104 TU Quick coupler CZ N2O
With pigtail
SPK36810103 TU Quick coupler CZ VAC
SPK36810102 TU Quick coupler CZ AIR
BASIC DIMENSIONS
SPK36810101 TU Quick coupler CZ O2
TECHNICAL DATA
Gases: O2, N2O, Air, Air-800, CO2, VAC
Dimensions: Height: 73 mm, Width: 73 mm, Depth: 63 mm
4–5 bar (breathing gases)
Working pressure: 7–10 bar (instrumental gases)
(-0,4)–(-0,9) bar (vacuum)
Maximum test pressure: 20 bar
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Regulatory status: Complies with EN ISO 9170-1 (Terminal units)
Complies with CSN 85 2762 (Czech gas specific connections)
present HTM 02-01
Note!
Measurements in mm.
PEN version
For variants not listed, please contact our sales and product support.
Maintenance valve
TECHNICAL DATA
Gases: O2, N2O, Air, Air–800,CO2, N2, VAC
Dimensions: Height: 73 mm, Width: 73 mm, Depth: 63 mm
Working pressure: 4–5 bar (breathing gases)
7–10 bar (instrumental gases)
(-0,4)–(-0,9) bar (vacuum)
Maximum test pressure: 20 bar
Regulatory status: Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN ISO 9170-1 (Terminal units)
Complies with NF S 90-116 (Afnor gas specific connections)
Complies with FD S 90-119 (Afnor Air-800 gas specific connection)
present HTM 02-01
TU MU INSTALLATION TOOLS
Item No. Denomination
1 - SPP36810163
2 - SPP36810164
3 - SPP36810165
The AGSS (Anaesthetic Gas Scavenging Systems) were created as the fi rst line of defence to protect
medical staff against exposure to anaesthetic (exhaust)gases.
The GCE AGSS units are designed in accordance with the ISO 7396-2 and ISO 9170-2 to provide this
protection in best possible way. The GCE AGSS is now available as part of the extensive GCE MCGS
product portfolio.
TECHNICAL DATA
Ejector feeding: Compressed or medical air - for AGSS active
Feeding pressure allowed:Air use: 4±0,5 bar typical (max 6 bar) - for AGSS active
Medical Air use: 20-25 l/min - for AGSS active
Nominal flow rate: up to 50 l/min - for AGSS active
Pressure drop: <150 mbar, 90 l/min - for AGSS active
80 N Typical
Axial connection force:
40 N Typical
Axial mechaical resistance of the connection: >500 N
Functioning temperature: -20°C to +60°C
Warehousing and transportation: -20°C to +70°C
Depression (4 bar-30 l/min): -7 mbar
Class IIa medical device according to MDD 93/42/EEC
ISO 7396-2
Regulatory status:
ISO 9170-2
CE certification CE0425 and DOC
The STANDARD model of VSP valve has an UNI or AFNOR or NIST (5359) outlet according to the
customer’s requirements; under the central body of the valve we put a valve that allows to scavenge
automatically the gas in case there is too much pressure (over-pressure valve).Standard pressure switch
high/low pressure is included.
When we connect the VSP valve, we also install a low pressure gauge connected to the alarm of the
central station. This gauge gives the possibility to check the value of the pressure inside the system.
Item No. Gas Inlet point
TECHNICAL DATA
Standard pressure in p1: 8 bar (800kPa )
Flow rate: >20 m³/h
Valve material: Brass alloy P-CuZn40Pb2 (sn 355N/mm²)
Production: Turning brass
Surface treatment: 0.25mm Chromed surface
Leakage absences test pressure: 11.2 bar
Seal test: No leakage
Discharge rate: > 25 m³/h
Inlet pipe: Set for pipe Ø16 ext.
Outlet pipe: Set for pipe Ø16 ext.
Welding alloy: DIN 8513 (BS. 450 Sn)
Ag 46%, Cu 27%, Zn 24%, Sn 3%, Cd 0%
Over pressure valve characteristics:
Inlet connection: M14x1.5 (inlet joint)
Outlet connection: M22x1.5 (outlet joint)
Pressure in p1: 8 bar (800 kPa )
Over pressure test valve: 12.8 bar
Leakage result at 1.6 P1: no loss of pressure
Drain flow rate: > 97 m³/h
All MC 70 gas outlets have the same design but different colour codings and labels for different gases and
of course gas specific non-interchangeable quick connection valves.
Special efforts have been made to make the maintenance of the gas outlets as easy as possible.
> No special tools
> Maintenance valve of ball-type
> Few components
Furthermore the MC70 are made according to standard SS EN 8752430 for quick connections and
international standard SS EN ISO 9170-1 for terminal units. This means that the gas components are non-
interchangeable in every maintenance connection point.
The gas outlet is delivered with separate packages for quick connection valve, valvebody, plastic cover
with name plate, push-release plate etc. To make installation easier, the valve body has a tightening plug
mounted for convenient pressure testing.
All necessary mounting details such as brackets, screws etc. are included in the packages. Detailed
instructions are also part of the delivery. When mounting the gas outlet in a recessed way the gas outlet
can be mounted either in the front wall or in the rear wall, depending on which is first set up. Recessed
and exposed instalation set is necessary orded separately.
INSTALLATION TOOL
Item No. Denomination
325197290 Combi tool
SERVICE KIT
Item No. Denomination
325197222 Sparepart kit
TECHNICAL DATA
Gases: O2, N2O, Air, Air–800, CO2, N2, Ar, AGSS, VAC
Dimensions: Diameter: 90 mm, Depth: 60 mm
4–5 bar (breathing gases)
Working pressure: 7–10 bar (instrumental gases)
(-0,4) – (-0,9) bar (vacuum)
Maximum pressure: 20 bar
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN ISO 9170-1 (Terminal units)
Regulatory status:
Complies with EN ISO 9170-2 (Terminal units for AGSS)
Complies with SS 8752430 (SS gas specific connections)
present SIS HB 370
LABELS MC70
Item No. Denomination Languages
548234A26760 Circular Label O2 White 85/55 TU SS –
325113069 Circular Label MEDICINSK OXYGEN White 85/55 TU SS SE
548234A26770 Circular Label N2O Blue 85/55 TU SS –
325113070 Circular Label DINITROGENOXID N2O Blue 85/55 TU SS DK
325113071 Circular Label MEDICINSK LUFT Black/White 85/55 TU SS SE
548234A26780 Circular Label Air Black/White 85/55 TU SS –
548234A37600 Circular Label Air–800 Black/White 85/55 TU SS –
325113074P Circular Label MEDICINSK KULDIOXID Grey 85/55 TU SS DK
548234A26790 Circular Label VAC Red 85/55 TU SS –
325113072 Circular Label VAC Red 85/55 TU SS DK
548234A26800 Circular Label VAC Yellow 85/55 TU SS –
548234A40850 Circular Label GASUTLOPP Blue/Brown 85/55 TU SS SE
548234A40860 Circular Label GASUDLØB Blue/Brown 85/55 TU SS DK
548234A40870 Circular Label GASSUTLØP Blue/Brown 85/55 TU SS NO
548234A40880 Circular Label KAASUJEN POISTO Blue/Brown 85/55 TU SS FI
548234A26810 Circular Label AGSS Purple 85/55 TU SS –
Central gas supply systems ⁄ Wall and ceiling suply units | 103
GCE HEALTHCARE
104 | Central gas supply systems ⁄ Wall and ceiling suply units
GCE HEALTHCARE
COPENHAGEN STYLE
Bed head unit designed for children’s wards where the patients will surely not be bored. With wide
choice of layouts, shapes and colours this BHU provides nicer and friendlier atmosphere in hospital
rooms. It comes with a variety of electrical, IT and lighting options including fluorescent or LED lights and
possibility of AmbientLight or ControlLight technology. Front panel is made of Corian® DuPont® what
enables 3D designs. Corian is a Solid Surface material very easy to clean what is of crucial importance in
hospitals.
SPECIFICATION
> Body made of aluminium profile mounted directly on the wall
> Front decorative cover made of Corian
> Length: 1500 mm
> Depth: 65 mm body + 24/36 mm front panel
> Height: depends on the unit’s shape
> Unit’s construction requires ceiling general lighting
Standard equipment for hospital rooms:
> Medical gases outlets: O2 and VAC
> Electrical sockets: 230V 3 pcs
> Equipotential socket: 1 pcs
> RJ45 cat.6 socket for IT: 1 pcs
> Indirect light 39W T5: switch on the unit
> Direct light 24W T5: switch on the unit
> Night light 2,8W LED: switch on the unit
Equipment options:
> Front decorative cover in any color (Corian color palette)
> Medical gases and vacuum outlets code types: DIN / BS / AFNOR / SS / others
> Lighting: T5 / 2G11 / LED
> Electrical sockets standards: German / French / British / others
> IT sockets: RJ-11 / RJ-12 / RJ-45 of cat.5e or higher
> Nurse call system (cut-out)
> Lighting switches on patient’s handset
Central gas supply systems ⁄ Wall and ceiling suply units | 105
GCE HEALTHCARE
MALAGA STYLE
This bed head unit is just what you need as this is you who decides about its design. A variety of
graphic inspirations and shapes are available. A highresolution printing on Corian® DuPont® panel or
on toughened glass results in exceptional colour depth and interaction of light. This unit can be fully
customized, assembled vertically or horizontally. The unit is particularly popular in plastic surgery clinics,
delivery wards and all the hospital areas where medical technology and beauty meet.
SPECIFICATION
> Body made of aluminium profile mounted directly on the wall
> Front decorative cover made of Corian or toughened glass
> 1 electrical and 1 gas installation input for double and multi-units
> Unit’s construction requires ceiling general lighting
> Length: 1600 mm
> Depth: 120 mm
> Height: 300 mm
106 | Central gas supply systems ⁄ Wall and ceiling suply units
GCE HEALTHCARE
1 - ceiling cover
2 - medical rails
3 - terminal units
4 - main beam
6 - electrical sockets
7 - lights
Central gas supply systems ⁄ Wall and ceiling suply units | 107
GCE HEALTHCARE
ICU TYPE L
A simple supply beam system applicable the hospital ICU ward. As an overhead solution it is far away
more functional in areas in where medical personnel needs access to the head of the bed. Coming with
standard fittings as medical gas outlets, electrical and data sockets and different type of lighting.
SPECIFICATION
> Body made of extruded aluminium profiles
> Front and side panels can be powder coated in any RAL colour
> Separate gas and electricity compartments and supply inlets
> Length: 2050, 2250, 2450, 2650, 2850 mm
> Height: 240 mm
> Depth: 370 mm
Sample configuration:
> Lenght: 2450 mm
> Beam powder coated in RAL 9002
> Medical gas outlets: O 2 pcs, AIR 2 pcs, VAC 2 pcs 2
> Electrical sockets: 230V 16 pcs
> Equipotential sockets: 16 pcs
> RJ45 cat. 6: 2 pcs
> General lighting 2x54W T5, lighting switch on the beam
> Direct/reading lighting 24W 2G11, lighting switch on the beam
> Night lighting 3W LED, lighting switch on the beam
Equipment options:
> Medical gas and vacuum outlets: DIN / BS / AFNOR / SS /others
> Electrical sockets standards: German / French / British / others
> IT sockets: RJ-11 / RJ-12 / RJ-45 of cat.6
> LED or fluorescent lighting
> Nurse call system (cut-out)
> Accessories: shelves, medical rails, IV poles, catheters baskets or wound care products carriers, suction
regulators, flowmeters and other
108 | Central gas supply systems ⁄ Wall and ceiling suply units
GCE HEALTHCARE
ICU TYPE M
Type M beam is an essential medical rescue aid equipment for the modern intensive care unit. It is mainly
composed of main beam frame, dry section and wet section. Its characteristics are reasonable dry and
wet separation structures where the wet area is equipped with the flexible infusion and dry - with sliding
glider c/w trays and drawers.
SPECIFICATION
> Body made of extruded aluminium profiles
> Front and side panels can be powder coated in any RAL colour
> Separate gas and electricity compartments and supply inlets
> Length: 2050, 2250, 2450, 2650, 2850 mm
> Height: 240 mm
> Depth: 370 mm
Sample configuration beam:
> Lenght: 2450 mm
> Beam powder coated in RAL 9002
> Medical gas outlets: O 2 pcs, AIR 2 pcs, VAC 2 pcs 2
> Electrical sockets: 230V 16 pcs
> Equipotential sockets: 16 pcs
> RJ45 cat. 6: 2 pcs
> General lighting 2x54W T5, lighting switch on the beam
> Direct/reading lighting 24W 2G11, lighting switch on the beam
> Night lighting 3W LED, lighting switch on the beam
Equipment options:
> Medical gas and vacuum outlets: DIN / BS / AFNOR / SS / others
> Electrical sockets standards: German / French / British / others
> IT sockets: RJ-11 / RJ-12 / RJ-45 of cat.6
> LED or fluorescent lighting
> Nurse call system (cut-out)
> Accessories: trays, medical rails, IV poles, catheters baskets or wound care products carriers,
suction regulators, flowmeters and other
Central gas supply systems ⁄ Wall and ceiling suply units | 109
GCE HEALTHCARE
ICU TYPE H
Type H beam - same as type M - is composed of main beam frame, dry section and wet section. The wet
and dry area adopts a pendant structure as they come as service heads with rotating arms. Service heads
come with fittings as medical gas outlets, electrical and data sockets and the cross beam is equipped with
a lighting treatment lamp in the middle. The rotating arm adopts a break to prevent drift.
SPECIFICATION
> Body made of extruded aluminium profiles
> Front and side panels can be powder coated in any RAL colour
> Separate gas and electricity compartments and supply inlets
> Length: 2050, 2250, 2450, 2650, 2850 mm
> Height: 240 mm
> Depth: 370 mm
Sample configuration beam:
> Lenght: 2450 mm
> Beam powder coated in RAL 9002
> RJ45 cat. 6: 2 pcs
> General lighting 2x54W T5, lighting switch on the beam
> Direct/reading lighting 24W 2G11, lighting switch on the beam
> Night lighting 3W LED, lighting switch on the beam
Equipment options:
> Medical gas and vacuum outlets: DIN / BS / AFNOR / SS /others
> Electrical sockets standards: German / French / British / others
> IT sockets: RJ-11 / RJ-12 / RJ-45 of cat.6
> LED or fluorescent lighting
> Nurse call system (cut-out)
> Accessories: shelves, medical rails, IV poles, catheters baskets or wound care products carriers,
suction regulators, flowmeters and others
110 | Central gas supply systems ⁄ Wall and ceiling suply units
GCE HEALTHCARE
MEDICAL PENDANTS
Medical ceiling pendants are crucial equipment in the operating rooms and ICUs. They are designed to allow an easy access to medical gases,
electrical and data sockets and other desired connections. Thanks to a great number of configurations medical devices can be positioned in a
user-friendly and ergonomic way. At GCE we understand modern ORs are multifunctional cross-disciplinary rooms and we offer ceiling units that
improve efficiency and provide the best care of the patients.
Central gas supply systems ⁄ Wall and ceiling suply units | 111
GCE HEALTHCARE
accessible for surgeons. The same way these types of arms enable better organization of surgeons
workspace and efficient workflows at the same time.
VERTICAL TRIANGLE
GAS OUTLETS NUMBERSERVICE
4 HEAD 6 8 10
Electrical sockets 8 [6] 12 [9] 15 [12] 18 [15]
IT sockets 2 3 4 6
Equipotential pins 8 8 16 16
112 | Central gas supply systems ⁄ Wall and ceiling suply units
GCE HEALTHCARE
ANESTHESIA PENDANTS
Anesthesia pendant is applied in operating rooms, endoscope rooms, anesthesia rooms. It is used to set up anesthesia working area, integrating
anesthesia machine, infusion pump, anesthesia information station and so on. It provides convenience of anesthesia operation to the greatest
extent. Thanks to a wide range of single or double arms with horizontal or vertical movement it covers the whole anesthesia area, can easily
locate the equipment in the ideal position, provide gas source, power supply, data communication interface, moreover cables for related
equipment needed in anesthesia area. Coming as standard with horizontal service heads can be customized to accommodate a large variety of
medical gas outlets and electricity.
Accessories: shelves, medical rails, IV poles, catheters baskets or wound care products carriers,
suction regulators, flowmeters and others
For more information and variants please contact our sales and product support
Central gas supply systems ⁄ Wall and ceiling suply units | 113
GCE HEALTHCARE
SURGICAL PENDANTS
Surgical pendant is designed to reduce OR tripping hazards and improve workflow, maximizing the surgical workspace. It reduces reliance on
carts and enables the operators to streamline the workflow of the operating room. Meeting all requirements of heavy duty systems in the medical
field surgical pendants perfectly match the demands of modern operating room. They come with vertical service heads equipped with medical
gas outlets, electrical and data sockets, shelves, drawers, medical rails, drip stands whereby its configurations can be fully customized. Thanks to
a wide range of single or double arms with horizontal or vertical movement they cover the whole surgical area.
Equipment options:
Additional gas outlets as: AIR-MOTOR; AIR-800; CO ; O /N O mixtures 2 2 2
Medical gas and vacuum outlets: DIN / BS / AFNOR / SS / others
German / French Electrical sockets standards: / British / others
IT sockets: RJ-11 / RJ-12 / RJ-45 of cat.6
Accessories: shelves, medical rails, IV poles, catheters baskets or wound care products carriers,
suction regulators, flowmeters and others
For more information and variants please contact our sales and product support
114 | Central gas supply systems ⁄ Wall and ceiling suply units
GCE HEALTHCARE
ICU PENDANTS
ICU Pendants are suitable for PICU - Pediatric ICU, NICU - Neonatal ICU, RICU - Respiratory ICU, SICU - Surgical ICU, MICU - Medical ICU.
The whole ICU setup can be exchanged with wet and dry sides. The patient is not disturbed and clinical staff has all what is needed. Designed
for the ICU and rescue room, the ICU medical pendant integrates most of the treatment and care equipment and supplies gas and electricity in a
centralized manner. Configuration can be fully personalized according to customer’s actual needs.
VSR 120 ICU PENDANT VSN 100 ICU PENDANT VSS 100 90 ICU PENDANT
Equipment options:
Additional gas outlets as: AIR-MOTOR; AIR-800; CO ; N O, AGSS, O /N O mixtures 2 2 2 2
Medical gas and vacuum outlets: DIN / BS / AFNOR / SS / others
Electrical sockets standards: German / French / British / others
IT sockets: RJ-11 / RJ-12 / RJ-45 of cat.6
Accessories: shelves, medical rails, IV poles, catheters baskets or wound care products carriers,
suction regulators, flowmeters and others
For more information and variants please contact our sales and product support
Central gas supply systems ⁄ Wall and ceiling suply units | 115
GCE HEALTHCARE
Specification for copper and copper alloy fittings with capillary ends for soldering and brazing for use with
copper tubes. BS EN 1254-1:1998
Specification for copper and copper alloy fittings with short ends for capillary brazing for use with copper
tubes. All fittings are supplied in individually sealed protective polythene bags and are specifically
designed for copper medical gas and vacuum systems.
Elbow 90° - FF
116 | Central gas supply systems ⁄ Wall and ceiling suply units
GCE HEALTHCARE
ELBOW 90° - FF
Item No. Dimensions (mm) d a x EQUAL TEE - FFF
Item No. Dimensions (mm) d a x
MEDEL012 12 16.2 7.6
MEDTE012 12 18 9
MEDEL015 22 28 12.6
MEDTE022 22 28 12
MEDEL028 28 35 16.6
MEDTE028 28 34 15
MEDEL035 35 46 23
MEDTE035 35 42 20
MEDEL042 42 55 28
MEDTE042 42 50 23
MEDEL054 54 70 38
MEDTE054 54 61.5 29.5
MEDEL067 67 83.5 50
MEDTE067 67 78 42
MEDEL076 76 91.5 58
MEDTE076 76 80 46
MEDEL108 108 127 79.5
MEDTE108 108 112 64
MEDEL133 133 127 79.5
MEDTE133 133 135 85
MEDEL159 159 160 100
MEDTE159 159 160 100
MEDEL219 219 200 130
MEDTE219 219 178 115
SERVICE TEMPERATURE**
Size Min -196°C 65°C 120°C 150°C 175°C Max 200°C
8mm 62.5 bar 62.5 bar 58.8 bar 46.7 bar 35.5 bar 23.3 bar
10mm 50.9 bar 50.9 bar 47.9 bar 38.0 bar 28.9 bar 19.0 bar
12mm 42.9 bar 42.9 bar 40.3 bar 32.0 bar 24.3 bar 16.0 bar
15mm 40.3 bar 40.3 bar 37.9 bar 30.1 bar 22.8 bar 15.0 bar
22mm 35.6 bar 35.6 bar 33.5 bar 26.6 bar 20.2 bar 13.3 bar
28mm 28.2 bar 28.2 bar 26.5 bar 21.1 bar 16.0 bar 10.5 bar
35mm 25.2 bar 25.2 bar 23.7 bar 18.8 bar 14.3 bar 9.4 bar
42mm 23.2 bar 23.2 bar 21.8 bar 17.3 bar 13.1 bar 8.6 bar
54mm 19.8 bar 19.8 bar 18.6 bar 14.7 bar 11.2 bar 7.4 bar
** Max hydraulic working pressure for LTC fittings when assembled with Lawton Tube’s EN13348 copper tube using hard solder
(brazing alloy) to EN 1044
SERVICE TEMPERATURE*
Size Min -196°C 65°C 120°C 150°C Max 200°C
67mm 18.6 bar 18.6 bar 17.5 bar 14.0 bar 6.9 bar
76mm 18.6 bar 18.6 bar 17.5 bar 14.0 bar 6.9 bar
108mm 17.2 bar 17.2 bar 16.2 bar 12.9 bar 6.4 bar
133mm 10.5 bar 10.5 bar 8.5 bar 7.7 bar 4.1 bar
159mm 11.7 bar 11.7 bar 9.7 bar 8.6 bar 4.6 bar
*** Max hydraulic working pressure for LTC fittings when assembled with Lawton Tube’s EN13348 copper tube using hard solder
(brazing alloy) to EN 1044
Central gas supply systems ⁄ Wall and ceiling suply units | 117
GCE HEALTHCARE
MUNSEN RINGS
The pipeline should be adequately supported at sufficient intervals in accordance with the below table to
prevent sagging or distortion. Supports for the surface-mounted pipework should provide clearance to
permit painting of the surface.
Where it is essential for pipes to cross electric cables or conduit, they should be support at intervals either
side of the crossing to prevent them from touching the cable or conduit. Supports should be of suitable
material or suitably treated to minimise corrosion and prevent electrolytic reaction between pipes and
supports.
> HTM 02 compliant
> Suitable for interior and external use
Sizes Material Thread
12mm - 159mm Brass M10
Note: Consideration should be given to additional supports near LVAs, elbows etc where the
potential effects of inadvertently applied torque can result in serve pipeline distortion or fracture
118 | Central gas supply systems ⁄ Wall and ceiling suply units
GCE HEALTHCARE
NOTES
GCE HEALTHCARE
NOTES
GCE HEALTHCARE
DELIVERY TEST 24. For parts, which have been repaired or replaced under Clause 21, the Seller shall have the same liability for
defects as for the original goods for a period of one year. For other parts of the goods the liability period referred to
5. Where a delivery test has been agreed, it shall, unless otherwise agreed, be carried out where the goods are in Clause 23 shall be extended only by the period during which the goods could not be used due to a defect for which
manufactured. the Seller is liable.
If technical requirements for the test have not been agreed, the test shall be carried out in accordance with general
practice in the industry concerned in the country where the goods are manufactured. 25. The Buyer shall notify the Seller in writing of a defect without undue delay after the defect has appeared and in
no case later than two weeks after the expiry of the liability period defined in Clause 23 as supplemented by Clause
6. The Seller shall notify the Buyer in writing of the delivery test in sufficient time to permit the Buyer to be present at 24. The notice shall contain a description of how the defect manifests itself. If the Buyer fails to notify the Seller in
the test. writing within the above time limits, he loses his right to make any claim in respect of the defect. If there is reason to
If the Buyer has received such notice, the test may be carried out even if the Buyer is not represented at the test. believe that the defect may cause damage, notice shall be given forthwith. If notice is not given forthwith, the Buyer
The Seller shall record the test. The test report shall be sent to the Buyer. The report shall, unless otherwise shown by loses the right to make any claim based on damage which occurs and which could have been avoided if such notice
the Buyer, be considered to correctly describe the execution of the test and its results. had been given.
7. If at the delivery test the goods are found not to be in accordance with the contract, the Seller shall as soon as 26. After receipt of a written notice under Clause 25, the Seller shall remedy the defect without undue delay. Within
possible ensure that the goods comply with the contract. If so required by the Buyer a new test shall thereafter be this limit the time for remedial work shall be chosen in order not to interfere unnecessarily with the Buyer’s activities.
carried out. The Buyer may not, however, require a new test if the defect was insignificant. The Seller shall bear the costs as specified in Clauses 21–32.
8. If no other division of the costs has been agreed, the Seller shall bear all costs for delivery tests carried out where Remedial work shall be carried out at the Buyer’s premises unless the Seller finds it appropriate to have the defective
the goods are manufactured. The Buyer shall, however, at such delivery tests bear all costs for his representatives, part or the goods sent to him for repair or replacement at his own premises.
including costs for travel and subsistence. The Seller shall carry out dismantling and re-installation of the part if this requires special knowledge. If such special
knowledge is not required, the Seller has fulfilled his obligations in respect of the defect when he delivers a duly
repaired or replaced part to the Buyer.
DELIVERY 27. If the Buyer gives such notice as referred to in Clause 25, and no defect is found for which the Seller is liable, the
Seller shall be entitled to compensation for the work and costs which he has incurred as a result of the notice.
9. Where a trade term has been agreed, it shall be interpreted in accordance with the INCOTERMS in force at the
formation of ´the contract. If no trade term is specifically agreed, the delivery shall be Ex Works.
28. If remedy of the defect requires intervention in other equipment than the goods, the Buyer shall be responsible for
any work or costs caused thereby.
TIME FOR DELIVERY. DELAY
10. If, instead of a fixed date for delivery, the parties have agreed on a period of time within which delivery shall take 29. All transports in connection with repair or replacement shall be at the Seller’s risk and expense.The Buyer shall
place, such period shall start to run at the formation of the contract. follow the Seller’s instructions regarding how the transport shall be carried out.
11. If the Seller finds that he will not be able to deliver the goods at the agreed time or if delay on his part seems likely, 30. The Buyer shall bear the increase in costs for remedying a defect which the Seller incurs when the goods are
he shall without undue delay notify the Buyer thereof in writing, stating the reason for the delay and if possible the located elsewhere than at the destination stated in the contract or – if no destination has been stated – the place of
time when delivery can be expected. If the Seller fails to give such notice, he shall, regardless of the provisions of delivery.
Clauses 13 and 14, reimburse the Buyer for any additional expenses, which the latter incurs and which he would have 31. Defective parts, which have been replaced under Clause 21, shall be placed at the Seller’s disposal and shall
avoided, had he received notice in time. become his property.
12. If delay in delivery is caused by a circumstance which under Clause 36 constitutes ground for relief or by an act or 32. If the Seller fails to fulfil his obligations under Clause 26 within a reasonable time, the Buyer may by written notice
omission on the part of the Buyer, including suspension by the Seller under Clause 18, the time for delivery shall be require him to do so within a final time. If the Seller fails to fulfil his obligations within that time limit, the Buyer may at
extended by a period, which is reasonable having regard to the circumstances in the case. The time for delivery shall his option:
be extended even if the reason for delay occurs after the originally agreed time for delivery. a) have the necessary remedial work carried out and/or have new parts manufactured at the Seller’s risk and expense,
provided that the Buyer proceeds in a reasonable manner, or
13. If the Seller fails to deliver the goods on time, the Buyer is entitled to liquidated damages from the date on which b) demand a reduction of the agreed purchase price not exceeding 15 per cent thereof.
delivery should have taken place. The liquidated damages shall be payable at a rate of 0.5 per cent of the agreed If the defect is substantial, the Buyer may instead terminate the contract by written notice to the Seller. The Buyer
price for each complete week of delay. If the delay concerns only a part of the goods, the liquidated damages shall shall also be entitled to such termination where the defect remains substantial after measures referred to in a). In case
be calculated on the part of the price which is properly attributable to the part of the goods which cannot be taken in of termination, the Buyer shall be entitled to compensation for the loss he has suffered. The compensation shall not,
use due to the delay. however, exceed 15 per cent of the agreed purchase price.
The liquidated damages shall not exceed 7.5 per cent of that part of the price on which it is calculated. The liquidated
damages become due at the Buyer’s written demand but not before all of the goods have been delivered or the
33. Regardless of the provisions of Clauses 21–32, the Seller shall have no liability for defects in any part of the goods
contract is terminated under Clause 14.
for more than two years from the start of the liability period referred to in Clause 23.
The Buyer loses his right to liquidated damages if he has not lodged a written claim for such damages within six
months after the time when delivery should have taken place.
34. The Seller shall have no liability for defects save as stipulated in Clauses 21–33. This applies to any loss the defect
may cause, such as loss of production, loss of profit and other consequential economic loss. This limitation of the
14. If the Buyer is entitled to maximum liquidated damages under Clause 13, and the goods are still not delivered, the Seller’s liability shall not apply, however, if he has been guilty of gross negligence.
Buyer may in writing demand delivery within a final reasonable period which shall not be less than one week.
If the Seller fails to deliver within such final period and this is not due to any circumstance for which the Buyer is
responsible, the Buyer may, by written notice to the Seller, terminate the contract in respect of that part of the goods LIABILITY FOR DAMAGE TO PROPERTY CAUSED BY THE GOODS
which cannot be taken in use due to the delay.
In case of such termination the Buyer shall also be entitled to compensation for the loss he suffers because of 35. The Buyer shall indemnify and hold the Seller harmless to the extent that the Seller incurs liability towards any
the Seller’s delay to the extent that the loss exceeds the maximum of liquidated damages which the Buyer may third party in respect of loss or damage for which the Seller is not liable towards the Buyer according to the second
claim under Clause 13. This compensation shall not exceed 7.5 per cent of that part of the price which is properly and third paragraphs of this Clause.
attributable to the part of the goods in respect of which the contract is terminated. The Seller shall have no liability for damage caused by the goods:
The Buyer shall also have the right to terminate the contract by written notice to the Seller if it is clear that there will be a) to any (movable or immovable) property, or consequential loss due to such damage, occurring while the goods are
a delay, which under Clause 13 would entitle the Buyer to maximum liquidated damages. In case of termination on this in the Buyer’s possession, or
ground the Buyer shall be entitled to both maximum liquidated damages and compensation under the third paragraph b) to products manufactured by the Buyer or to products of which the Buyer’s products form a part.
of this Clause. The above limitations of the Seller’s liability shall not apply if he has been guilty of gross negligence.
Except for liquidated damages under Clause 13 and termination of the contract with limited compensation under this If a third party lodges a claim for compensation against Seller or Buyer for loss or damage referred to in this Clause,
Clause 14, all other claims in respect of the Seller’s delay shall be excluded. the other party to the contract shall forthwith be notified thereof in writing.
This limitation of the Seller’s liability shall not apply, however, where the Seller has been guilty of gross negligence. The Seller and the Buyer shall be mutually obliged to let themselves be summoned to the court or arbitral tribunal
which examines claims against either of them based on damage or loss alleged to have been caused by the goods.
15. If the Buyer finds that he will be unable to accept delivery of the goods on the agreed date, or if delay on his part
The liability as between the Seller and the Buyer shall, however, always be settled by arbitration in accordance with
seems likely, he shall without undue delay notify the Seller thereof in writing stating the reason for the delay and, if
Clause 39.
possible, the time when he will be able to accept delivery.
If the Buyer fails to accept delivery on the agreed date, he shall nevertheless make any payment which is dependent GROUNDS FOR RELIEF (FORCE MAJEURE)
on delivery as if the goods in question had been delivered. The Seller shall arrange storage of the goods at the
Buyer’s risk and expense. If the Buyer so requires, the Seller shall insure the goods at the Buyer’s expense. 36. The following circumstances shall constitute grounds for relief if they impede the performance of the contract or
makes performance unreasonably onerous: industrial disputes and any other circumstance beyond the control of the
16. Unless the Buyer’s failure to accept delivery as referred to in Clause 15 is due to any such circumstance as parties, such as fire, war, mobilization or military call up of a comparable scope, requisition, seizure, trade and currency
described in Clause 36, the Seller may by written notice require the Buyer to accept delivery within a reasonable restrictions, insurrection and civil commotion, shortage of transport, general shortage of materials, restrictions in the
period. supply of power and defects or delays in deliveries by sub-contractors caused by any such circumstance as referred
If, for any reason for which the Seller is not responsible, the Buyer fails to accept delivery within such period, the to in this Clause.
Seller may, by written notice to the Buyer, terminate the contract in respect of that part of the goods which is ready The above described circumstances shall constitute grounds for relief only if their effect on the performance of the
for delivery but has not been delivered due to the Buyer’s default. The Seller shall then be entitled to compensation contract could not be foreseen at the time of formation of the contract.
for the loss he has suffered by reason of the Buyer’s default. The compensation shall not exceed that part of the price
37. The party wishing to claim relief under Clause 36 shall without delay notify the other party in writing on the
which is properly attributable to the part of the goods in respect of which the contract is terminated.
intervention and on the cessation of such circumstance.
If grounds for relief prevent the Buyer from fulfilling his obligations, he shall reimburse the expenses incurred by the
PAYMENT seller in securing and protecting the goods.
17. Unless otherwise agreed, the agreed purchase price, together with value added tax, if any, shall be invoiced with 38. Notwithstanding other provisions of these General Conditions, either party shall be entitled to terminate the
one third at the formation of the contract, one third when the Seller gives written notice that the bulk of the goods are contract by notice in writing to the other party, if performance of the contract is delayed more than six months by
ready for delivery. Final payment shall be invoiced at delivery of the goods. The invoiced amount becomes due 30 reason of any grounds for relief as described in Clause 36.
days after the date of the invoice.
18. If the Buyer fails to pay, the Seller shall be entitled to interest from the due date at the rate of interest determined DISPUTES. APPLICABLE LAW
by the law on late payments in the Seller’s country.
If the Buyer fails to pay by the due date, the Seller shall also, after having notified the Buyer in writing thereof, suspend 39. Disputes arising out of or in connection with the contract shall not be brought before the court, but shall be finally
performance of his contractual obligations until payment is made. settled by arbitration in accordance with the law on arbitration applicable in the Seller’s country.
40. All disputes arising out of the contract shall be judged according to the law of the Seller’s country.
GCE Group
www .gcehealthcare .com
www .gcegroup .com
ESAB Middle East F .Z .E, P.O.Box 8964 Jebel Ali Free Zone DUBAI, U.A.E
Art. Nr. CATHCME22. Alterations are subject to change without notice. 020922kk © GCE 2022
All photographs are illustrative only.