Application_Note_-_Functional_Changes_in_NX_3.0.24.00_-_NX_4.0.24.00
Application_Note_-_Functional_Changes_in_NX_3.0.24.00_-_NX_4.0.24.00
NX 24.00
Type 4406
This document describes the new customer features that were introduced with software
version NX 24.00.
Initial version
► Referenced documents
Document Reference
NX 24.00 - SB 01 - Release of NX 3.0.24.00 / 4.0.24.00 Document ID 80844814
Application Note “Functional Changes in NX 22.00, NX Document ID 68534240
22.00 SU1”
Application Note “Functional Changes in NX 23.00” Document ID 75753690
NX Service Manual Document ID 74737949
MUSICA Acquisition Workstation (NX) User Manual 4420T Library
Smartbox Service Manual Document ID 80983940
Smartbox - Service Bulletin No. 01 – Introduction of the Document ID 80982773
Smartbox to support AI pathology detection
► Manufacturer
Agfa NV
Published by
Agfa-Gevaert HealthCare GmbH
Max-Planck-Straße 1
82380 Peißenberg
Germany
Copyright 2023 Agfa NV
All rights reserved.
Technical modifications reserved.
Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium or its
affiliates.
All other trademarks mentioned in this document are held by Agfa NV or the respective
owners and are used in an editorial fashion with no intention of infringement.
Nothing contained in this legal notice nor in any text in this document shall be
construed as granting by implication, estoppel or otherwise, any license or right to use
any of the trademarks, service marks, trade names or logos appearing in this document
without the express prior written consent of their respective owner.
WARNING:
Improper operation or service activities may cause damage or injuries.
(1) Read the Generic Safety Directions prior to attempting any operation, repair or
maintenance task on the equipment. Refer to Document ID 11849633.
(2) Strictly observe all safety directions within the Generic Safety Directions and on
the product.
IMPORTANT:
The installation and service of the product(s) described herein is to be performed
by qualified personnel who are employed by Agfa NV or one of its affiliates
or who are otherwise authorized by Agfa NV or one of its affiliates to provide such
services.
LIST OF CONTENTS
Mammo DR Retrofit
CR Systems
DR Retrofit
DX-D 100
DX-D 300
DX-D 400
DX-D 500
DX-D 600
DR 100e
DR 100s
DR 400
DR 600
DR 800
Valory
Feature Reference
SmartPatientView 1.1
CriticalScan (Chest pathology detection)* 1.2
Collimation verification 1.3
SmartRotate for all body parts 1.4
Quicker annotation 1.5
Improved barcode scanner workflow 1.6
Improved grid line suppression 1.7
Compliance with EURATOM directive 1.8
Updated Dose Structured Report 1.9
CMS improvement 1.10
Orthogon “end of service life” 1.11
ISP hardening 1.12
= feature is not available
= feature is available
* AI supported chest pathology detection not supported for mobile Retrofit systems (with tablet / notebook) and in
general only in countries where the AI module “LUNIT Insight CXR” is registered.
The new features of XRDI 25.0 are explained in the XRDI Service Manual, Document
ID 72453884.
XRDI 26.0 does not have new customer features. For details of all solved software
issues and new features refer to the XRDI Service Manual, Document ID 72453884.
NOTE:
The NX version depends on the presence of the NX_MULTI_FRAME_IMAGING
license.
• License not present: Version is NX 3.0.24.00
• License present: Version is NX 4.0.24.00
The NX_MULTI_FRAME_IMAGING license enables dynamic imaging, a feature
currently supported by the DR 800 system only.
NX 3.0.24.00 / NX 4.0.24.00 is further abbreviated as “NX 24.00” if both software
versions are meant. The version is displayed in the About NX menu.
Figure 1: About NX
1.1 SmartPatientView
NOTE:
An overview of all SmartXR features is available in the following documents:
• DR 600 – Service Application Manual, Document ID 48556251
• DR 100s – Service Application Manual, Document ID 69476884
For DR 800, SmartPatientView and SmartRotate currently are the only SmartXR
functions.
NOTE:
For DR 100s and DR 600 it is possible to adjust the zoom by parameter X-Ray Device
settings > SmartDR Settings > Smart DR Context.
Figure 3
Once the image is taken, the “patient identification image” icon in the patient
identification pane turns from gray to white.
By clicking on this icon the “patient identification image” can be re-called.
Refer to Figure 4.
Figure 4
Figure 6
Figure 7
The “Patient Positioning Image” can be activated for each single exam in the
“Processing and Routing” pane of the exam tree.
Figure 8
Figure 9
It is possible to configure the “Patient Positioning Image” very efficiently via the Exam
Tree Configuration > Find and Replace Exposure Settings.
However, using the “Find and Replace Exposure Settings” function to set the archiving
function per exposure is not available. This has to be set manually, if required.
Figure 10
The feature to use AI (artificial intelligence) or any other means for automatic pathology
detection is not available.
*NOTE:
LUNIT Insight CXR is a deep learning based software that assists radiologists or
clinicians in the interpretation of chest X-ray images. Final responsibility of diagnosis
still lies with radiologists or clinicians. The analysis result contains localization of
suspicious areas in color or outline, abnormality scores reflecting the probability that
the detected lesion is abnormal, and text interpretation for the analysis result by each
finding.
Advantages:
• Radiographer can set a HPM (high priority marker) on patients with critical finding
to reduce the reporting delay, or call clinician according to local protocol to change
patient care pathway.
Limitations:
• Supports Chest PA and AP exposures.
• Currently for any patient who is 6 years or older.
• Released as an option for DR 100s, DR 400, DR 600, DR 800, DX-D 300,
DX-D 600, desktop PC based DR Retrofit systems and CR systems*. It will not be
released for DX-D 400, DX-D 500, Mammo DR Retrofit, and mobile systems
(except DR 100s).
• Released in countries only, where LUNIT Insight CXR is released. Contact your
local sales representative for information about whether Chest Pathology Detection
can be implemented in your country.
*only CR systems that are released for NX 24.00.
Refer to the NX 24.00 Release SB, Document ID 80844814.
Licensing:
This feature requires the license NX PATHOLOGY DETECTION LUNIT (MBA3H).
This license includes the following technical licenses:
• NX_PATHOLOGY_DETECTION
• NX_PATHOLOGY_DETECTION_LUNIT
• NX_AI_AUTO_ROTATE
• SmartBoxSoftware_SUB_BASE
Smartbox:
The additional computing power which is
required to perform the AI pathology
detection is provided by the Smartbox.
For DR 100s a special kit exists which
integrates the Smartbox into the system.
For more information to the Smartbox
refer to the Smartbox Service Manual,
Document ID 80983940 and the
Smartbox release Service Bulletin, Figure 13
Document ID 80982773.
Workflow:
When the criteria to allow AI Pathology Detection are fulfilled, the image is sent to the
Smartbox for processing by LUNIT Insight CXR. As a result, a secondary capture
overlay is sent to NX.
With release of NX 24.00 the criteria to allow AI Pathology Detection are as follows
(can change with later NX versions):
• Patient's Age ≥ minimum age
• Body Part Examined within supported Body Parts
• View Position within supported View Positions
Pathology Detection
DR Workflow
Select thumbnail
Start exposure
Image arrives on NX
Image reference
gets acknowledged by
NX backend process Image is sent to
LUNIT Insight CXR
Image is processed on Smartbox
incl. SmartRotate
Image is evaluated
Criteria to allow
AI Pathology Detection DICOM Secondary
fulfilled? yes
Capture is sent to NX
no
Image is displayed in NX
Image is displayed in viewer with Secondary
NX viewer Capture Overlay
Figure 14
NOTE:
The VL(Visual Light) and SC (Secondary Capture) images are sent to the selected
archive at the moment of a “Close and Send” action only. A “Send” action will only send
the X-ray image. It is possible to configure where the VL images are sent, but the SC
images are always sent to the primary automation archive. At the moment there is no
way to disable sending of SC images.
Figure 15
Configuration
Configuration of AI Pathology Detection comprises two parts:
• Configuration of LUNIT Insight CXR. This is described in the Smartbox Service
Manual, Document ID 80983940.
• Configuration of AI Pathology Detection behavior on NX: This is described in this
section.
To configure the AI device on NX, go to NX Service & Configuration Tool > Device
Configuration > AI Device and click Create.
Figure 16
Now the New buttons becomes active. Click on New to create a new AI device.
Figure 17
The Device Settings of the AI Device do not need any change. They fit to the default
settings of LUNIT Insight CXR on the Smartbox.
The Supported Pathologies pane defines the pathologies, which shall generate
operator attention, if this pathology is detected. This should be inline with the
“radiologic findings to analyze”, what is configured on the Smartbox.
Basic rule:
The pathologies which are activated on NX in the “Supported Pathologies” pane,
should also be selected in the “Select radiologic findings to analyze” menu in the
Smartbox Service Interface > DICOM gateway configuration.
Figure 18
For each exam, where AI Pathology Detection shall be applied, the following settings
are required:
1. Enable Smart Rotate in Exposure Settings. The reliability of the AI Pathology
Detection depends on the correct orientation. This setting is mandatory.
2. Enable module in Processing & Routing settings. This enables the AI
Pathology Detection module. This setting is mandatory.
3. Enable automation in Processing & Routing settings. This enables automatic
image evaluation on the Smartbox. If this is not active, the exposure can only
be sent manually for AI Pathology Detection. This setting is optional.
Figure 19
Figure 20
Figure 21
Advantage:
• Radiologist sees if the image
borders are the collimation borders
or if it was cropped digitally by the
radiographer. Figure 22
Benefits:
• Safety: QA manager / Radiographer team lead can verify if the X-ray image is
collimated properly for dose reduction purposes.
• Productivity: If a pathology is detected close to the border and deserves a larger
area, the radiologist knows if a retake is necessary or if he can ask for
uncropping to the radiographer.
• Regulatory: required in Germany.
Configuration on NX:
The white border is configured per exposure in the exam tree. The parameter is called
“Collimation border thickness”.
This border thickness has to be configured manually for each exposure.
By default it is “0” mm”.
It is not possible to use the “Find and Replace Exposure Settings” functions to adapt all
or several exposures at once.
Figure 23
Licensing:
This feature does not require an additional license. It is part of NX 24.00 by default.
Figure 24
Licensing:
This feature requires the technical license NX_AI_AUTO_ROTATE which is included in
the license NX SMARTDR AUTO ROTATE (MAE3Q).
For more details refer to document Functional Changes in NX 22.00, NX 22.00 SU1,
Document ID 68534240.
Figure 25
Figure 26
The text as well as the time format (Date / Date and Time / Time) is configured in the
NX Service & Configuration Tool > User Interface > General Settings > Custom
Markers.
Figure 27
Licensing:
The custom markers require the technical license NX_Advanced_Tools, which is
included in the NX Precision Tools (ES9BE_3.0) license.
Figure 28
Licensing:
The following license is required to enable the “COM port emulation mode”:
NX Barcode Reader license (MBGRK).
The hardware is ordered with the following sales order code:
XHV1L (BARCODE READER KIT (BLUETOOTH))
Figure 31
Figure 32
The Auto Open checkbox defines, how the system behaves in case a
patient is scheduled for more than one exam.
• With Auto open checkbox checked, the scheduled exam opens automatically, if
the configured attribute matches.
• With Auto open checkbox unchecked, the scheduled exams for this patient are
shown in the worklist. The user needs to select the scheduled exam manually.
For details refer to the NX Service Manual, Document ID 74737949, section 4.9.40,
Configuring a barcode scanner CS6080.
Figure 33
The usage of the barcode scanner is described in the “Barcode reader User Manual
4422A”.
The setup is described in NX Service Manual, Document ID 74737949, section 4.9.40,
Configuring a barcode scanner CS6080.
Limitations
The barcode scanner CS6080 is
delivered together with a Bluetooth
cradle using Bluetooth class I which
allows up to 100 m (330 feet) distance
between cradle and scanner.
A free USB Type A slot is required on
the NX.
If the NX is located in the control room,
the shielding of the control might prevent
a satisfying Bluetooth connection. In this
case consider mounting the cradle in the
examination room. Figure 34
The USB cable which is part of delivery has a length of 2 m and can be extended to a
maximum of 5 m (USB 2.0).
Figure 35
* XD+ Detectors = XD+10, XD+14, XD+17. De-activating NX_DR_ADVANCED_GLS is no option for XD 10,
XD 14, XD17.
With NX 23.00 the CMS for collection of hospital wide quality metrics was introduced.
This was achieved by distinguishing between:
• Rooms for Image Review
• Rooms for Dose and Repeat/Reject analysis
For details refer to Application Note “Functional Changes in NX 23.00”, Document ID
75753690.
Disadvantage: The database for Dose and Repeat/Reject analysis is limited to 10 GB.
In the worst case this can lead to OutOfMemory exception causing CMS connectivity
issues [PRB2000837].
NOTE:
How to read this table:
There are two general application types for a CMS:
CMS is either used for review of images (left column) or for review of quality metrics
(right column).
Depending on the application type, there are different requirements and technical
specifications for the CMS and in-room NX.
Example: If the CMS is used for review of quality metrics, the NX version of the CMS
must be ≥ NX 23.00. All in-room NX must be of version ≥ NX 21.00. All NX must be
configured the same dose type (lgm or EI). And so on ….
Condition • NX main version on CMS and in- • All NX must be configured for
room must be identical same dose type (lgm or EI)
• All in-room must be on the same
subnet as CMS
• All NX must be configured for
same dose type (lgm or EI)
Limitation • Max. 5 NX in-room can be • max. 850.000 dose records*
connected to the CMS • max. 850.000 reject reports*
Comment • CMS does not display the patient • NX / CMS can be on different
positioning image provided by subnets
SmartPatientView (option as of • No limitation on number of in-
NX 24.00 for DR 100s, DR 600 rooms
and DR 800) • When limit of dose records is
• CMS can be used for both, reached, the oldest records will
review of images and review of get replaced by newer ones
quality metrics • For NX < NX 23.00 on CMS,
statistics are deleted on the In-
Room NX after retrieval
• CMS can be used for both,
review of images and review of
quality metrics
*limit introduced as of NX 24.00
Figure 36
Definition of hardening
The purpose of system hardening is to eliminate as many ISP (Information Security
and Privacy) security risks as possible. This is typically done by removing all non-
essential software programs and utilities from the computer.
NOTE:
Difference between hardening and installing Windows hotfixes:
Hardening a Windows 10 computer means configuring the security settings. This
reduces opportunities for a virus, hacker, ransomware, or another kind of cyberattack.
By installing hotfixes, system files are installed to correct a functional or security issue.
2 Explanation of notes
Safety-relevant notes
Icon Signal word Situation
CAUTION: Hazardous situation which, if not avoided, can lead to a minor
injury to a user, engineer, patient or any other person.
Non-safety-relevant notes
Icon Name Type of information
IMPORTANT: Highlights very important actions which have to be carried out to
prevent malfunction.
3 Conventions
Style Use case Example
(1) Step 1 Step by step task description (1) Connect the cables.
(2) Step 2 (2) Mount the cover.
Bold Menu topics, keyboard keys, device Press F9 or double-click the Refresh
buttons, commands, and so on button.
Italic Emphasizing a word or indicating Do not insert the USB flash drive yet.
references in continuous text. Refer to section 10, Release
information.
Courier System messages When the success message
bold appears, close the window.