QuickTrace Software Manual
QuickTrace Software Manual
Software Manual
Part Number: 480114
QuickTrace Mercury Analyzer Software Manual
QuickTrace Mercury Analyzer Software Manual
REPRODUCTION SAFETY
All rights reserved. Reproduction or Instruments, accessories, components or
transmission of this document in whole or other associated materials may not be
in part, and by any means without the returned to CETAC Technologies if
express written consent of the copyright contaminated with biohazard or
owner or authorized agent is prohibited. radioactive materials, infectious agents, or
Requests for additional copies of this, or any other materials and/or conditions that
any other CETAC publication, can be filled could constitute a health or injury hazard
by contacting an authorized distributor or: to CETAC employees. Call Customer
CETAC Technologies Service and Support if there is any
Customer Service & Support question or doubt relative to
14306 Industrial Road decontamination requirements.
Omaha, Nebraska 68144, USA
Phone (800) 369-2822 (USA only) CAUTION and WARNING statements, as
Phone (402) 733-2829 applied in this document, shall be
Fax (402) 733-1932 interpreted consistent with the following
E-mail [email protected] context: CAUTION applies only to
potential property damage conditions;
DISCLOSURE WARNING applies to potential personal
injury conditions, in combination with or
This document contains CETAC exclusive of potential property damage.
proprietary data and is provided solely to
its customers for their express benefit of All user-serviceable components are
safe, efficient operation and maintenance specifically identified in this document as
of the product described herein. Use or such; the balance shall be assumed to
disclosure of CETAC proprietary data for require the expertise of a factory service
the purpose of manufacture or technician/engineer for adjustment,
reproduction of the item described herein, repair, replacement, modification, etc.
or any similar item, is prohibited, and Others not so qualified and performing
delivery of this document shall not these actions shall do so at their own risk.
constitute any license or implied Furthermore, never operate the
authorization to do so. instrument without first reading and
understanding the QuickTrace M-7500
REVISIONS Mercury Analyzer Operator’s Manual or
M-6000A Mercury Analyze Operator’s
CETAC Technologies strives to provide
Manual and QuickTrace Mercury
the scientific community with an
Analyzer Software Manual and ensuring
unparalleled combination of effective
the instrument is operated safely and
technology and continuing value. Modular
properly.
upgrades for existing instruments will
continue to be a prime consideration as
designs progress.
QuickTrace Mercury Analyzer Software Manual
ORIGINAL PACKAGING
Retain original factory packaging for
moves and factory return shipments.
Shipping in anything other than the
original fitted foam and container can
result in incidental damage from which
the purchaser will not be protected under
warranty.
WARNING
Under all conditions the user must observe safe laboratory
procedures during the operation of this product.
Notices and Compliance Declarations
This equipment generates, uses, and can Cet appareil numerique respecte les
radiate radio frequency energy and, if not limites de bruits radioelectriques
installed and used in accordance with the applicables aux appareils numeriques de
instructions, may cause harmful Classe A prescrites dans la norme sur le
interference to radio communications. materiel brouilleur: "Appareils
Operation of this equipment in a Numeriques," NMB-003 edictee par le
residential environment is likely to cause ministre des Communications.
harmful interference, in which case the
user will be required to correct the
interference at his own expense.
MODIFICATIONS
The FCC requires the user to be notified
that any changes or modifications made to
this device that are not expressly approved
by CETAC Technologies, Inc. may void the
user's authority to operate the equipment.
CABLES
Connections to this device must be made
with shielded cables with metallic RFI/EMI
connector hoods to maintain compliance
with FCC Rules and Regulations.
AD-1
Notices and Compliance Declarations
To reduce the risk of fire hazard and electrical shock, do not expose the unit to rain or
humidity. To reduce the risk of electrical shock, do not open the cabinet. All maintenance
CAUTION is to be performed by an Authorized CETAC Service Provider.
Protection provided by the equipment may be impaired if the equipment is used in a
manner not specified by the manufacturer.
CLEANING INSTRUCTIONS
To clean the exterior surfaces of the instrument, complete the following steps:
1 Shut down and unplug the instrument. 3 Repeat step 2, using a towel dampened
with clear water.
2 Wipe the instrument exterior surfaces
only using a towel dampened with a 4 Dry the instrument exterior using a dry
lab-grade cleaning agent. towel.
Do not allow any liquid to enter the instrument cabinet, or come into contact with
WARNING any electrical components. The instrument must be thoroughly dry before you
reconnect power, or turn the instrument on.
ENVIRONMENTAL
AD-2
Notices and Compliance Declarations
AVERTISSEMENT
TOUT CONTACT AVEC LES HAUTES
AVERTISSEMENT TENSIONS PEUT ENTRAINER LA MORT
OU DES BLESSURES SÉVÈRES. CE
POUR UNE PROTECTION CONTINUÉ
PANNEAU NE DOIT ÊTRE ENLEVE QUE
CONTRE LES RISQUES D’INCENDIE,
PAR UN RÉPARATEUR QUALIFIÉ.
REMPLACER UNIQUEMENT PAR DES
FUSIBLES DE MÊME TYPE ET
AMPÈRAGE.
AVERTISSEMENT
NE PAS GLISSER LA MAIN SOUS OU DERIERE LES
ECRANS THERMIQUES DU FOUR. GARDER LA
PORTE D'ACCES AU DEVANT DU BOITIER BIEN
FERMEE POUR ASSURER LA PROTECTION CONTRE
LES BRULURES
AVERTISSEMENT
TOUT CONTACT AVEC LES HAUTES
TENSIONS PEUT ENTRAINER LA MORT
OU DES BLESSURES SÉVÈRES. CE
PANNEAU NE DOIT ÊTRE ENLEVE QUE
PAR UN RÉPARATEUR QUALIFIÉ.
AVERTISSEMENT
TOUT CONTACT AVEC LES HAUTES
TENSIONS PEUT ENTRAINER LA MORT
OU DES BLESSURES SÉVÈRES. CE
PANNEAU NE DOIT ÊTRE ENLEVE QUE
PAR UN RÉPARATEUR QUALIFIÉ.
Notices and Compliance Declarations
AVERTISSEMENT
AVERTISSEMENT SURFACES CHAUDES, LAISSER LE
COUVERCLE HERMÉTIQUEMENT
TOUT CONTACT AVEC LES HAUTES
FERMÉ.
TENSIONS PEUT ENTRAINER LA MORT
POUR ACCÉDER, METTRE LA
OU DES BLESSURES SÉVÈRES. CE
TEMPÉRATURE DU FOUR À ZÉRO,
PANNEAU NE DOIT ÊTRE ENLEVE QUE
OUVRIR LE COUVERCLE ET LAISSER
PAR UN RÉPARATEUR QUALIFIÉ.
REFROIDIR 5 MINUTES AVANT DE
TOUCHER LA VERRERIE OU TOUTE
SURFACE MÉTALLIQUE INTÉRIEURE.
WARNING
HIGH LEAKAGE CURRENT -
ENSURE PROPER GROUNDING
AVERTISSEMENT
COURANT DE FUITE ÉLEVÉ — FORNIR
UNE MISE À LA TERRE EFFICACE.
AVERTISSEMENT
POUR LA PROTECTION PERMANENTE
CONTRE UN CHOC ÉLECTRIQUE, UNE
BRÛLURE DES YEUX (RADIATION UV)
OU DE LA PEAU, LAISSER LE
COUVERCLE HERMÉTIQUEMENT
FERMÉ LORSQUE L’APPAREIL EST SOUS
TENSION.
LAISSER REFROIDIR 5 MINUTES
(APPAREIL ÉTEINT) AVANT D’ENLEVER
LE COUVERCLE.
AD-4
Contents
QuickTrace Mercury Analyzer Software Manual
Contents
Contents
Preface....................................................................................... viii
ii
QuickTrace Mercury Analyzer Software Manual
Contents
iii
QuickTrace Mercury Analyzer Software Manual
Contents
iv
QuickTrace Mercury Analyzer Software Manual
Contents
7 Security 7-1
8 Troubleshooting 8-1
Glossary G-2
v
Preface
Preface
The QuickTrace Mercury Analyzer Software Manual provides an
overview and explains the theory of operation of the CETAC
QuickTrace Mercury Analyzer software.
WARNING
Before operating the QuickTrace M-7500 or the M-6000A and the
ASX-510 Auto Sampler, it is important to read the QuickTrace M-
7500 or M-6000A Mercury Analyzer Operator’s Manual and the
ASX-510 Auto Sampler Operator’s Manual.
Instructions
All step-by-step instructions begin with a “To” statement, and are then
numbered, as in the following example.
Menu Items
This book uses the following format for referring to menu items:
Settings / Communication
The text before the “slash” symbol is the name of the menu; the text
after the slash symbol is the menu choice.
ix
QuickTrace Mercury Analyzer Software Manual
Preface
Notes
Notes contain a reminder about the effect of particular actions. They
are indicated as follows:
Note:
Cautions
Cautions indicate situations that require immediate attention to
prevent harm to the Mercury Analyzer. Cautions are indicated as
follows:
Warnings
Warnings indicate situations that could cause bodily harm. Warnings
are indicated as follows:
WARNING
This example shows how a warning is displayed.
x
QuickTrace Mercury Analyzer Software Manual
Preface
xi
QuickTrace Mercury Analyzer Software Manual
Preface
e-mail: [email protected]
xii
1
Introduction
QuickTrace Mercury Analyzer Software Manual
Introduction
Introduction
The CETAC QuickTrace Hg software was developed to give the
Mercury Analyzer operator a comprehensive and multi-tasking
environment for method development, setup and analysis of samples,
and reporting/exporting data. Several important features and settings
are available, allowing the user to customize the software for their
particular applications.
1–2
QuickTrace Mercury Analyzer Software Manual
Introduction
• CD-ROM drive
1–3
QuickTrace Mercury Analyzer Software Manual
Introduction
It is necessary to have 2 free COM ports available, one each for the
Autosampler and the Mercury Analyzer.
3 Wait for the CD-ROM to initialize. The setup program should start
automatically; if it does not, run the setup.exe program on the CD.
5 After the installation is finished, in the under the Start Menu in the
Program Files|CETAC QuickTrace folder there will be a shortcut
labeled “Specify Installed Hardware”. Run this program.
1–4
QuickTrace Mercury Analyzer Software Manual
Introduction
1–5
2
Developing a Worksheet
To perform analyses, you must first open or create a Worksheet. You can
then use the Method Editor to develop the new Worksheet, or to modify
parameters in an existing Worksheet.
What is a Worksheet?
A Worksheet is a single file consisting of a method (a complete set of
operating parameters), a sequence (includes standards, samples and QCs),
and the resulting analysis data and signal profiles. Once a Worksheet has
been defined, you may use it as a template to create future worksheets.
This enables you to conveniently use the instrument repeatedly under a
fixed set of operating conditions without having to re-enter the parameters
each time. Several different template Worksheets are included with the
software to provide optimized operating parameters for different levels of
sensitivity.
Multi-tasking
More than one Worksheet may be open at a time, allowing you to multi-
task. For example, while running an analysis with one Worksheet, you
may develop the method for another Worksheet, or print a Report. Open
Worksheets will be displayed using bottom tabs. The “active” Worksheet,
which is the Worksheet you are currently working in, will be displayed as
the topmost tab.
2–2
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
Creating/Opening a Worksheet
To create a new Worksheet, click on the down arrow next to the Worksheet
icon. A drop-down box appears, presenting you with three choices: New,
New From (existing Worksheet), and Open.
New From is the recommended option when starting a new analysis. With
New From, you can create a new Worksheet from an existing Worksheet or
one of the provided Worksheet templates, which already contains a defined
set of operating parameters, calibration standards, and QC test solutions.
1 Select New, and a New Worksheet dialog window will appear (see
Figure 2-2).
2–3
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
4 Click on Save, and the Method Editor for the new Worksheet will
automatically open.
2–4
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
1 Select New From, and a dialog window will open (see Figure 2-3).
4 Click OK, and the new Worksheet will open to the Method Editor.
Note:
If you select Open, the Open Worksheet dialog window appears, allowing
you to select the desired Worksheet. The Worksheet will open to the
Sequence page.
Hint:
2–5
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
The File menu (above the main toolbar) contains additional Worksheet
options:
Note:
Close. Closes the Worksheet after prompting to save any pending changes.
2–6
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
Method Editor
Once you have created or opened a Worksheet, you can use the Method
Editor. To open Method Editor, click on the Method Editor icon located on
the main toolbar.
Conditions page
You can use the parameters on this page to define a Time Profile. The Time
Profile controls data collection during the sample analysis. It consists of the
Integration Read time at the top of the absorbance signal profile and 1 or 2-
point Baseline Correction times. The signal chart with colored bars
describing the Integration Read Time and the Baseline Correction times
makes it simple for you to create the Time Profile.
To display the Conditions page, click the Conditions tab in Method Editor.
2–7
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
Operating Conditions
Gas flow. This is the carrier gas flow rate for the analysis. Valid gas flow
rates range from 20 – 350 mL/min for the M-7500 Mercury Analyzer, and
35 – 350 mL/min for the M-6000A. See Tables 2-1a and 2-1b for suggested
optimized parameter ranges.
Pump Speed. This controls the rate that the sample and reagent solutions
will be pumped into the gas-liquid separator during analysis. Sample and
reagent flow rates depend on both the peristaltic pump type (M-7500 or M-
6000A) and the pump tubing. 100% speed recommended for both M-7500
and M-6000A.
2–8
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
Sipper Depth. This value determines how far the autosampler probe
moves down into the sample tube. A higher number will allow the probe to
move further down. A lower number will not allow the probe to reach the
bottom of the tube, which is useful in preventing uptake of particulates
which may collect in the tube bottom. Allowable range is 0 – 155 mm.
Sample Uptake Time. This is the time that the autosampler probe
remains in the sample tube, allowing uptake of liquid. Recommended
values range from 20 – 110 seconds, depending on the gas flow rate.
Rinse Time. This is the time that the autosampler probe remains in the
rinse solution, between samples. The sum of Sample Uptake time and
Rinse time equals the total analysis time for one sample. Recommended
values for Rinse Time range from 30 – 160 seconds, depending on the gas
flow rate.
Read Delay Time. This is the time that the system waits before it starts
to measure the signal. This delay allows the signal to stabilize before
measurement begins.
The read delay time is also the starting point of the Integration Read time,
indicated by a green bar on the Signal Profile chart. The sample analysis
stop point (the red bar) equals the Read Delay time plus the Replicate Read
Time multiplied by the number of Replicates.
Replicate Read Time. This is the length of time data is collected for a
single replicate.
Replicates are used by the software to determine %RSD for the selected
Integration Read time of the signal profile. Replicates are “slices” of the
Integration Read time. For example, a 10 second Integration Read time
may be sliced into five 2-second replicates. The more replicates, the more
accurate the %RSD.
2–9
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
A large %RSD indicates that the peak is abnormal, or that the Integration
read time and/or Baseline correction times are not positioned correctly.
2–10
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
2–11
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
Gas Flow
2–12
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
Liquid Flow
Sample and reagent flow rates are determined by the peristaltic pump
speed and the pump tubing diameters. To achieve a good %RSD for the
signal profile, sample and reagent flow rates are optimized to guarantee
complete coverage of the Gas-liquid separator (GLS) center post at all times.
See Tables 2-1a and 2-1b for suggested optimized pump speed settings.
Timing Bars
The Signal Profile chart uses colored bars to designate Start Read and End
Read times for the signal profile and the Baseline Drift Corrections. From
st
the left, The 1 Baseline Point is designated by teal blue and burgundy
bars. Next, green and red bars designate the Integration Read time on top
nd
of the signal peak. The Start and End Read times of the 2 Baseline Point
are designated by olive and purple bars.
During Time Profile setup, the timing bars may be moved by clicking the
left-mouse button on a bar and dragging it to the desired location.
Zoom Out. This demagnifies the chart. An area of the chart can be
magnified by selecting the desired area with the left mouse button. To
zoom, hold down the left mouse button and move the mouse right and down
to encompass the desired area. To unzoom, hold down the left mouse
button and move towards the upper left corner of the chart, or click on the
button.
2–13
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
Save. Saves new or edited parameters within the three sections of Method
Editor.
Print Chart. Prints the Signal Profile Chart on the Conditions page.
Before you create the Time Profile, the instrument must be prepared to run
samples. The instrument should be fully warmed up, with gas and liquid
flows established. (See the Operator’s Manual for setting up the Mercury
Analyzer.)
2–14
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
4 To set the Integration Read time, move the green and red bars to the
“flattest” portion at the top of the signal peak.
5 To set the 2nd Baseline correction point (if desired), move the olive and
purple bars to a flat area after the signal peak has completely washed out.
6 After the Signal Profile has been configured correctly, and operating
conditions have been selected, click Save from the File menu.
Standards page
You can use this page to define calibration standards and concentrations,
and to select units and the calibration equation. You can also define
calibration limits and tests.
Calibration Parameters
A Linear calibration uses a least squares curve fitting, with the formula y
= ax + b.
Weighted Fit. This option weights the points of the calibration curve so
that those with less error have a higher weighting. The line is then
constructed so that it will be closer to the points with higher weightings.
2–15
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
Weighted fit allows you to compensate for errors, by making them smaller
where the signal is most accurate.
2–16
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
Units. This field shows the concentration units for the calibration
standards. The units specified here will be used throughout the Worksheet.
Whenever a calibration is performed, the standards are run and the new
calibration replaces the previous calibration. If the Recalibration Rate is
set to zero, then the initial calibration will not be replaced during the
analysis. If the Recalibration Rate is set to 20, then for every 20 th sample,
the calibration standards will be rerun, and the calibration curve will be
recalculated.
Reslope Rate. This field determines how often a reslope of the calibration
is performed, by selecting the number of samples to be analyzed before
resloping the calibration.
Note:
2–17
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
Calibration Slope (Upper and Lower Limits). These values express the
maximum allowable change in the slope between successive pairs of
standards. When the slope changes by more than the specified percentages,
then the defined Calibration Error action will occur. To set these limits,
click on the up/down arrows to get the desired percentages.
Reslope Slope Test (Upper and Lower Limits). When the Calibration
slope differs from the previous Calibration by more than the specified
percentages, then the defined Calibration Error action will occur. To set
these limits, click on the up/down arrows to get the desired percentages.
Upon Calibration Error. This field determines what action to take when
the calibration fails. Stop will halt the analysis. Flag and continue will flag
the calibration curve but continue the analysis.
The columns under Standard Name and Tube will automatically be filled in
when the number of Calibration Standards is greater than zero. To
customize standard names, select Standard Names from the Tools menu
at the top of the Method Editor window. Enter the desired names and click
OK to save. The autosampler tube position may be changed by clicking the
ellipsis button under the Tube column, and selecting the desired tube.
Under Concentration, enter the appropriate concentration for each
standard.
Under the Tools menu of Method Editor is a feature called Define Units.
This allows the user to define customized units. Ten commonly used units
are available.
Unit Table. Displays a list of all the currently defined units. Highlight
the unit in the table to view its multiplier and divider values in the
corresponding fields.
2–18
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
Delete. Deletes the currently highlighted unit, if it is not being used in the
current worksheet. If the highlighted unit is being used, it cannot be
deleted.
Delete all. Deletes all units from the table, except those currently being
used in the worksheet.
Divider. Displays the divider value for the currently highlighted unit.
Add. Adds a new (customized) unit to the table. The unit must be defined
with a multiplier and divider before adding it to the table (see To define
customized units below).
To add a new unit, enter the unit name, the multiplier and divider and click
Add.
QC Tests page
You will use this page to enter QC solution concentrations and test
parameters, such as upper and lower limits which trigger failure actions by
the software. You may also enter detection limits and %RSD limits.
To display the QC Tests page, click on the QC Tests tab in Method Editor.
2–19
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
QC Test Solutions. This column specifies the QC Test Solutions that are
available, including several types of Check Standards, Check Blanks,
Spikes, and Duplicates.
2–20
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
2–21
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
2–22
QuickTrace Mercury Analyzer Software Manual
Using Method Editor
Specify Limits. Click on the box next to the desired tests, so that a check
appears. Enter in the appropriate parameters for each limit test.
2–23
3
Creating a Sequence
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
Creating a Sequence
Once the Worksheet method has been developed, you may set up the
Sequence of test solutions. The Sequence allows you to determine in
what order to run the samples, calibration standards, and quality
control tests.
This chapter will describe how to create a Sequence. You will configure
Autosampler racks, enter in solution information, and automate the
actions that will occur before, during, or after a sample run. These
sections are used to create a Sequence: Autosampler/Autodilutor setup,
Sequence screen, Sequence Editor, and Sequence Parameters.
The Sequence screen displays a table where you can manually enter
and/or edit sample information.
Autosampler/Autodilutor Setup
Clicking this button will pop up a window used to configure the
Autosampler and Autodilutor, and to define the types of racks in the
Autosampler tray.
3–2
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
Rack Definitions
The default setting is a 60 x 14mL rack (this means 60 14mL vials per
rack). There are 5 different rack types available. Select the
appropriate rack types from the drop-down boxes on the left. Changes
in the rack types are immediately reflected in the display.
3–3
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
Note:
The default Rack size used when creating a new Worksheet can be
customized from the Sequence tab of the Preferences menu.
CAUTION Failure of the analyst to use the correct rack sizes for the ASX-510
Autosampler can cause the sipper probe to jam against a solid object
like a rack or a sample tube. This may bend/damage the tip of the
sipper probe as well as other parts of the Autosampler. This type of
damage may not be covered in the warranty.
Start of dilution range. This is the first tube position that will be
used for diluting samples. Enter the appropriate tube position or use
the icon to click on the desired starting position on the Autosampler
guide. If you start the dilution range from a rack that was not
3–4
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
Dilute when. Select to dilute the sample either when the instrument
saturates or when the sample concentration is above the range of the
calibration.
Note:
Dilute samples by. Select the desired dilution ratio. For example, to
perform a 1:10 dilution, set the number to 10. Allowable values are 2 –
10.
If no more tubes. If you run out of dilution tubes during the analysis,
you have three choices: prompt for new tubes, continue undiluted, or
terminate analysis. Prompt for new tubes will pause the analysis and
display a message that new tubes are needed. Continue undiluted
allows the analysis to continue, but no more dilutions will be
3–5
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
performed. Stop analysis will stop the analysis when all available
dilution tubes have been used.
Sequence Editor
The Sequence Editor enables you to automatically generate a sequence
of mercury determinations, which may include samples, calibration
solutions and quality control tests. It consists of 3 tabs/pages:
Sequence, Auto QC, and Manual QC.
Sequence page
The Sequence page of the Sequence Editor is accessed by clicking on the
Sequence icon button. On this page, you will decide how many samples
are to be analyzed, and how to insert calibrations within the sample
run.
3–6
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
Note:
Starting Sample ID. This is the name you give to the sample(s). It
will be incremented automatically. Enter a name or use the default.
(You can edit the IDs from the Sample Table on the Sequence screen).
3–7
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
Inserting Calibrations
With this section, you will determine how to insert calibrations within
the sequence. Select 1 or more of the following options:
Note:
You must have a calibration curve for the Worksheet before you can
measure mercury in samples. It is recommended to calibrate at the
beginning of every sample batch analysis.
Note:
When setting recalibration rates, note that some of the QC tests are
treated as samples. For example, by default Duplicates, Dilutions and
Laboratory Control Samples are treated as samples, and are therefore
included in the sample count.
3–8
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
Note:
The default settings for these options can be set on the Sequence tab of
the Preferences dialog. The next time you create a worksheet, the
default options you set for the sequence will be implemented.
Generate a Sequence
If you are not using QC tests, you may click the Generate Sequence
button at the bottom right corner to display the sequence of samples
and calibration standards. This will immediately close the Sequence
Editor, so that you may view the Sample Table on the Sequence screen.
To add QC tests, use the Auto QC or Manual QC pages to select the test
intervals and error actions, covered in the following sections.
Auto QC page
You will use this page (Figure 3-3) to determine where to place QC
solutions within the automatically generated Sequence, and what test-
and-error actions will be performed. To display the Auto QC page,
select the Auto QC tab from the Sequence Editor.
The left column shows a list of the available QC test solutions. QCs will
be defined as either Calibration QCs (color-coded in yellow) or Sample
QCs (color-coded in green).
The right column displays the insertion points and error actions. From
this column, you can change the error action for each insertion point.
3–9
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
Note:
You can choose to include any or all of the available QC tests in the
current sequence by highlighting the test in the left column and
dragging it into the Calibration QC, Sample QC or Auto QC sections in
the right column. A more thorough explanation of the options available
for each follows:
3–10
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
Calibration QC
This section enables you to include Calibration QC tests in your
sequence, at one of the following five insertion points:
• Beginning of run
• Before re-calibration
• After re-calibration
• End of run
• Preparation Blank
• Method Blank
• Reagent Blank
• Equipment Blank
• Field Blank
3–11
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
Sample QC
This section enables you to include Sample QC tests in your sequence,
at one of the following two insertion points:
• On first sample
• QC Spike
• Matrix Spike
• Duplicate
• Stop analysis
3–12
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
2. Hold down the mouse and drag the test to the Calibration or
Sample QC Error Actions insertion point, in the right column. For
example, if it is a Sample QC test, drag it to the ‘On first sample’ or
‘On first sample after calibration’ position.
2. Click OK.
The QC interval will then be updated with the new selection. For
example, if you selected ‘Re-calibrate and repeat’, the action would read
‘After first calibration, re-calibrate and repeat upon error’.
Auto QC Actions
You may use this section to define your own error actions. Both
Calibration QCs and Sample QCs may be placed in this section.
You can also select what action will occur if an error results. The
following error actions will be available:
3–13
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
• Stop analysis
Note:
3–14
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
2. Set the interval between samples at which you want the Auto QC
action to be analyzed by typing a number in the ‘Every’ field, or by
using the up/down arrows. Select the action to occur in the event of
an error from the drop-down box.
3. Click OK. The action properties will be updated with the new
parameters.
Note:
When setting the Auto QC test interval, note that some of the QC tests
are treated as samples. For example, by default Duplicates, Serial
Dilutions and Laboratory Control Samples are treated as samples, and
are therefore included in the sample count.
To generate a Sequence
Once you have setup the QC test intervals and error actions, you may
click the Generate Sequence button at the bottom right corner to
display the sequence of samples, calibration standards, and QCs. This
will close the Sequence Editor, so that you may view the Sample Table
on the Sequence screen.
Manual QC page
You will use this page to set the error actions for QC test solutions that
are entered manually into the Sample table on the main Sequence
screen. Also, if you change a Sample to a QC on the Samples table, it
will be considered a manual insertion, and will use the error action as
described on this page.
To display the Manual QC page, select the Manual QC tab from the
Sequence Editor.
3–15
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
The left column shows a list of the available QC test solutions (see
Figure 3-4). The right column displays the error actions. From this
column, you can change the error action for each insertion point.
The table displays all of the QC tests that can be entered directly into
the Worksheet Sample table.
3–16
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
• stop analysis
3. Click OK. The error action you select will then be displayed in the
table.
To generate a Sequence
Once you have setup the QC test intervals and error actions, you may
click the Generate Sequence button at the bottom right corner to
display the sequence of samples, calibration standards, and QCs. This
will close the Sequence Editor, so that you may view the Sample Table
on the Sequence screen.
Sequence Screen
The Sequence screen is part of the main Worksheet window and may be
used to view/edit the sample Sequence, or to manually enter sample
labels, rack:tube positions, weights, volumes, and sample types.
The Sequence screen consists of the Sample Table and the AutoSampler
Guide.
3–17
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
3–18
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
Note:
The Autosampler Guide will only be visible on the Sequence page if the
Show Autosampler Guide option is selected in the Preferences menu.
The default view displays the Autosampler racks and tube color key,
however the appearance of the page depends upon the way it was last
configured. The option can also be selected by right-clicking on the
Sequence Sample table and selecting or deselecting the Show
Autosampler Guide option from the popup menu.
Sample Table
Tube. Displays the autosampler tube number for the sample. The first
number refers to the rack number of the sample, while the second
number relates to the position number within the rack. For example, a
tube number of 2:3 would indicate that the tube is in rack two of the
autosampler, at position three.
Sample Label. Lists the sample names. These labels are also
displayed in a non-editable column of the Analysis Sample table on the
Analysis Screen. Solutions that have been generated using the
Sequence Editor and appear in shaded cells are defined as
‘automatically generated’ samples. The Sample Label and Type for
these samples can be changed in the Sequence Editor, but their weight,
volume and dilution factors cannot be changed.
3–19
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
standards and QCS generated using the Sequence Editor, this option
cannot be changed.
3–20
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
Note:
Cells that cannot be directly edited from the Sample table appear to be
shaded. If the sample has been entered using the Sequence Editor, its
properties can only be changed in the Editor. Otherwise, no changes
can be made to shaded cells. Cells highlighted in white can be directly
edited from the Sample table simply by clicking in them and typing new
information or selecting an option from a drop-down menu.
Editing Cells
To copy the contents of a cell. Right click on the cell of interest and
select Copy from the drop-down menu. The contents will be copied to
the Windows clipboard.
To paste the contents of a cell. Right click on the cell of interest and
select Paste from the drop-down menu. The contents of the cell will be
overwritten with the contents of the Windows clipboard.
3–21
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
Note:
Rows may not be inserted before any samples that have already been
analyzed.
Autosampler Guide
The Autosampler Guide at the bottom of the Sequence screen provides
you with a visual representation of the layout. Please note that it is
NOT active: any changes you wish to make to the configuration must be
done using the Autosampler/Dilutor Setup page (see p.3-4 to 3-6).
Key to Tube Colors. Displays the tube colors used in the graphical
representation of the sample solutions. This will only be displayed if
the Show Autosampler Guide option is selected under the Options
menu, or by selecting Show Autosampler Guide from the right-button
menu on the Sequence Table.
Sequence Parameters
The Sequence Parameters dialog is accessible by selecting the Sequence
Parameters icon from the main screen toolbar. You can also access the
dialog by selecting the item from the Window menu above the main
toolbar.
3–22
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
This dialog window is used to set the actions to occur during and after
the run, and the defaults for report generation and other sequence
analysis behaviors.
Control
At End of Sequence
If you wish to put the instrument into standby mode at the end of a
sequence, check the desired options described below:
3–23
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
Lamp Off. The lamp will be turned off. (This option is not available
for M-6000A operators.)
Analysis Behaviors
Hint:
Instrumental Zeroing
Zero Before First Sample. The instrument will zero before running
the first sample.
3–24
QuickTrace Mercury Analyzer Software Manual
Creating a Sequence
Reports
Report, None. Reports will not print automatically.
Report, After Run. The report will print after the batch run is
completed.
Export To Text File. The report will be exported as a Text file (.txt).
There are several fields below that allow you to specify which
information should be exported. See Chapter 5: Report Options (p. 5-5
to 5-6) for descriptions of these fields.
3–25
4
Performing Analyses
QuickTrace Mercury Analyzer Software Manual
Performing Analyses
Performing Analyses
Once you have developed a Method and created a Sequence, the next
step in the software is to run samples.
This chapter will describe the procedures you will use to operate the
mercury analyzer through the QuickTrace software, using the toolbar
and the displays on the Analysis screen.
Before analyzing samples, make sure you have first accomplished these
steps:
4–2
QuickTrace Mercury Analyzer Software Manual
Performing Analyses
Analysis Screen
The Analysis screen is accessed by clicking on the Analysis button from
the main toolbar. This screen is where analysis controls and displays
are located. This chapter will describe how to start and stop analyses,
and how to use the Analysis screen displays.
4–3
QuickTrace Mercury Analyzer Software Manual
Performing Analyses
4–4
QuickTrace Mercury Analyzer Software Manual
Performing Analyses
Note:
If you stop during sample uptake, or before the rinse has been
completed, make sure you allow sufficient time in the rinse station
before re-starting the analysis.
4–5
QuickTrace Mercury Analyzer Software Manual
Performing Analyses
Note:
Using the Read Sample feature may invalidate your QC setup. If you
have stopped during a batch analysis to read a single sample, it will be
inserted into the Sequence after the last analyzed sample.
Note:
The Analysis screen shows you what is happening during the analysis,
allowing you to monitor the progress and results of the analysis. The
Analysis screen has four sections: the Sample Results Table, the Signal
Profile, the Calibration Chart, and the Worksheet Log.
4–6
QuickTrace Mercury Analyzer Software Manual
Performing Analyses
Conc. This field shows the concentration determined for each sample.
Flag. This field displays a flag if an error condition has occurred for
the sample.
4–7
QuickTrace Mercury Analyzer Software Manual
Performing Analyses
The Signal Profile chart displays a µAbsorbance Signal in real time for
the current sample being analyzed. There are several buttons on this
chart that are described below:
Copy Chart. This button will copy the Baseline chart to the
clipboard in bitmap format.
Calibration Chart
The Calibration displays the most recent calibration for the active
Worksheet. As a calibration is performed, each Calibration Standard
data point is placed on the chart after analysis of that standard. The
chart can be used to visually examine and edit the calibration curve.
The same buttons available on the Signal Profile chart are also on the
upper left of the Calibration chart (see above for descriptions).
4–8
QuickTrace Mercury Analyzer Software Manual
Performing Analyses
Worksheet Log
The Worksheet Log, located at the bottom left of the Analysis screen,
records each action in the Worksheet. These actions include saving a
Worksheet, analyzing a sample, and stopping an analysis.
From the right-click menu, you can select Export to save the log to
another location. To customize what is displayed, click Filter Log and
choose the desired option.
4–9
QuickTrace Mercury Analyzer Software Manual
Performing Analyses
4–10
5
Report Options
QuickTrace Mercury Analyzer Software Manual
Report Options
Report Options
This chapter contains information on setting up the report printout and
the available reporting and exporting options.
Reports
Once data has been collected, it may be examined, printed and/or
exported using the Reports window. Click on the Reports button on the
main toolbar to display the Reports window. There are two pages:
Worksheet and Settings. From both pages, the following print/export
options are available:
Export to File. This will allow you to create a .txt or .csv file,
and save it to the location of your choice.
5–2
QuickTrace Mercury Analyzer Software Manual
Report Options
Worksheet Page
On this page, you will select the Worksheet file for which you wish to
create a report, a print preview, or to export data. This page is set up
so that Worksheet files can be easily accessed from any location in the
Windows file system.
5–3
QuickTrace Mercury Analyzer Software Manual
Report Options
Settings page
The Settings page allows you to customize the report printout by
selecting solution information, report contents, and default number
format.
Note:
If you set the default report options you wish to use in Preferences
before you create a new Worksheet those settings will appear on the
Report Settings page.
5–4
QuickTrace Mercury Analyzer Software Manual
Report Options
Solution Information
% RSD. Displays the % RSD based on the replicates for the selected
Integration read time of the sample peak.
Online Dilution. Displays the Dilution factor used for the Worksheet
if an Auto-Dilutor is present.
Contents
5–5
QuickTrace Mercury Analyzer Software Manual
Report Options
5–6
6
Instrument Tools
QuickTrace Mercury Analyzer Software Manual
Instrument Tools
Instrument Tools
This chapter will describe how to use hardware controls and displays,
track maintenance for consumables, and maintain an instrument log.
Baseline Chart
The baseline chart is displayed at the bottom of the Autosampler and
the Analyzer hardware control pages. It mimics a strip chart recorder,
allowing you to visually examine the real-time absorbance signal.
6–2
QuickTrace Mercury Analyzer Software Manual
Instrument Tools
Autosampler/Autodilutor Controls
Move Sipper. This section contains controls that move the
autosampler sipper probe.
Park. Sends the sipper probe back into the rinse station, and will turn
on the rinse pump.
Move To. To move the sipper probe to a certain tube position, first
select the tube position by clicking on the … button, then click Move
To.
Up/Down. Click on Down to move the sipper probe down into the
tube, and Up to move the probe up. The box to the right designates
how far to move the probe down. A higher number will move the probe
further down.
Pump On/Off. Controls the rinse pump (at the back of the
autosampler), which cycles rinse solution in and out of the rinse station.
Racks 1-4. To change the rack size, click on the arrow and select the
appropriate size from the drop-down box. (The standard rack size is 60
x 14 mL.)
Autodilutor
Prime. Moves the dilutor probe to its rinse tube (position S:10, the last
position in the standards rack), and then fills and empties the pump
twice.
6–3
QuickTrace Mercury Analyzer Software Manual
Instrument Tools
The autodilutor syringe pump should not be run dry, or it may damage
CAUTION
the syringe pump. Be sure that the diluent bottle is in place, and that
the diluent tubing is down into the diluent solution.
Empty. Moves the dilutor probe to its rinse tube and empties the
syringe pump.
Analyzer Controls
Set Speed. Select the desired % Pump speed, and click Set Speed.
This setting will only be valid from this page. To use this setting in
your method, you must set the Pump speed in Method Editor.
Set Gas. Select the desired gas flow rate (0 – 350 mL/min), and click
Set Gas. This setting will only be valid from this page. To use this
setting in your method, you must set the Pump speed in Method Editor.
6–4
QuickTrace Mercury Analyzer Software Manual
Instrument Tools
you switch between Ar and N2, or if you re-install the software onto
your PC.
You will need a gas flow meter calibrated for the gas being used, in the
range of 0 – 400 mL/min. Connect the gas flow meter to the Carrier gas
supply port on the front of the analyzer (below the gas-liquid
separator). Make sure that gas pressure is set to 100 Psi. Click
Calibrate Gas Flow, and input values as prompted by the software.
Maintenance Tracking
This page is used to track the installation dates of consumables, and to
prompt you when to replace them.
Serial Numbers
Serial numbers for your Mercury Analyzer and Autosampler will be
displayed here if they were entered during the software installation.
(Otherwise, you can also locate the serial numbers on the back of your
instruments.) You may be asked for the serial numbers if you call
CETAC for service and/or technical support.
Wear Statistics
The wear statistics track the amount of time that consumables are in
use. The software will prompt you to replace consumables at
appropriate intervals. For the main consumable items (Mercury Lamp,
Nafion Dryer, and Pump Tubes), the display shows installation date,
hours in use, and when to replace. When a new item is installed, click
the Reset button to set the new installation date.
6–5
QuickTrace Mercury Analyzer Software Manual
Instrument Tools
Note:
Instrument Log
This page is used to log any information that you may want to track,
such as when the cell windows or the gas-liquid separator are cleaned.
You may also want to log any problems that occur, and what was done
to remedy the problem. This can be very helpful, especially if there are
multiple users of the Mercury Analyzer.
Note:
6–6
7
Security Options
QuickTrace Mercury Analyzer Software Manual
Security & Data Integrity
Security Options
This chapter describes how to install and use the security features.
Security
There are several levels of security available from the software to suit
many needs.
Default Security
The Users and Passwords administration window (Figure 7-1) from the
Control Panel allows individual accounts and groups to be
administered. (Under Windows NT 4.0 this functionality is accessed
through the Start|Administrative Tools|User Manager application.)
7–2
QuickTrace Mercury Analyzer Software Manual
Security & Data Integrity
7–3
QuickTrace Mercury Analyzer Software Manual
Security & Data Integrity
7–4
QuickTrace Mercury Analyzer Software Manual
Security & Data Integrity
User Accounts
Once security features are installed, user accounts which do not have
administrator privilege and are not members of one of the four groups
(see p.7-4) will not be able to run the QuickTrace software.
CETAC Service: These users have blanket access to the software and
all its features but cannot add user accounts or change access levels of
existing accounts.
CETAC Supervisor: These users can access any feature listed in this
software manual, except adding user accounts or changing the access
level of existing accounts.
CETAC Operator Level 2: These users can run samples and batches,
develop worksheets, and change most configuration options. They
cannot change QC limits or behavior. Operators with this level of
access cannot change a small number of policy type settings.
CETAC Operator Level 1: These users can run samples and batches
in existing worksheets. They can only create new worksheets from
templates. A very small number of additional operating settings can be
changed by users with this access level.
7–5
QuickTrace Mercury Analyzer Software Manual
Security & Data Integrity
In the left column, all accounts local to this machine are shown. In the
right column, the access level of each account is shown.
User accounts are created by clicking the Add button. Access levels are
modified by selecting the desired account and clicking Edit. In both
cases, the dialog in Figure 7-4 is displayed. User accounts that are
created using this tool are automatically given membership to the
group corresponding to their level of access to the QuickTrace features
(see page 7-4).
7–6
QuickTrace Mercury Analyzer Software Manual
Security & Data Integrity
Important Note:
7–7
8
Troubleshooting
QuickTrace Mercury Analyzer Software Manual
Troubleshooting
Troubleshooting
This chapter describes how to troubleshoot communication and
configuration problems between the software and the instruments. For
a complete list of hardware and performance issues, see the QuickTrace
M-7500 or the M-6000A Mercury Analyzer Operator’s Manual.
8–2
QuickTrace Mercury Analyzer Software Manual
Troubleshooting
Note:
8–3
QuickTrace Mercury Analyzer Software Manual
Troubleshooting
Figure 8-1.
8–4
QuickTrace Mercury Analyzer Software Manual
Troubleshooting
Figure 8-2.
If you have changed the COM port properties, you may have to sever
your connection and reconnect to have the new settings take effect.
This can be done using two buttons on the toolbar – the one with the
phone off the hook disconnects, the one with the phone on the hook
reconnects.
8–5
QuickTrace Mercury Analyzer Software Manual
Troubleshooting
The ASX-510 and the QuickTrace M-7500 (or M-6000A) will respond to
the command VER (capitalization is important).
8–6
QuickTrace Mercury Analyzer Software Manual
Troubleshooting
8–7
QuickTrace Mercury Analyzer Software Manual
Troubleshooting
8–8
QuickTrace Mercury Analyzer Software Manual
Troubleshooting
6 When the printer has been sent all the document information,
the Print Manager then deletes the temporary files and closes
itself.
At this point, the Print Manager closes, and hopefully the document
has printed. However, this is only the end of Window’s part of the
printing process. If you are using a network, there may be many more
places where a print can be lost even after the Print Manager has sent
it properly. If you are in doubt as to whether your computer is printing
correctly, connect a printer to the local port. If you receive a printout
from this local printer, and not from network printers, then the
problem is probably a network/PC configuration.
Hopefully, this information will help fix and avoid problems that occur
while printing.
If this does not work, call CETAC Technical Support at the number
listed in the Preface of this manual.
8–9
Glossary
QuickTrace Mercury Analyzer Software Manual
Glossary
Glossary
This book frequently uses the following terms:
Active When more than one worksheet is open, the
worksheet worksheet for which the bottom tab is topmost.
Abs Absorbance = (-log10 T or 2-LOG10 %T). Intensity
unit.
Algorithm Mathematical function used to convert a sample’s
measured absorbance into a concentration.
ADX-500 Online Auto Dilution accessory (optional)
ASX-510 ASX- 510 Auto Sampler.
Auto-zero The process of electronically setting the instrument
baseline to zero.
Baseline The process of correcting intensity for intensity due
correction to background. In this software, baseline
correction is performed by subtracting the intensity
measured with no sample in the sample cell from
the intensity measured with sample passing
through the sample cell.
Blank A solution that contains no Hg, used as a “zero”
point in the calibration.
Calibration The equation showing intensity (absorbance) vs.
curve concentration. The calibration curve is used to
determine the concentration of Hg in a sample.
CLP Contract Laboratory Protocol (analysis protocol of
U.S. EPA).
Cold Vapor Direct Atomic Absorption Spectrometric analysis
Direct AAS or (at 254 nm) of “head-space” gas from a stannous
CVAAS chloride or stannous sulfate reactor using neither
flame, nor plasma, nor furnace nor any other
electro-thermal atomizer.
G-2
QuickTrace Mercury Analyzer Software Manual
Glossary
G-3
QuickTrace Mercury Analyzer Software Manual
Glossary
G-4
QuickTrace Mercury Analyzer Software Manual
Glossary
G-5