1-1119D35e
1-1119D35e
PROFESSIONAL
& MICROMAT P
Operating Instructions
1 DESCRIPTION .........................................................................................................................4
CONNECTIONS ...............................................................................................................................7
5 INSTALLATION.......................................................................................................................8
5.1 SCOPE OF DELIVERY ...........................................................................................................8
5.2 INSTALLATION OF THE EQUIPMENT ........................................................................................8
5.2.1 Mains supply .....................................................................................................8
5.2.2 Microphone or a signal sensor connection .......................................................8
5.2.3 Printer connection .............................................................................................8
6 OPERATING ............................................................................................................................9
7 CONTROL KEYS.................................................................................................................. 10
7.1 level ROTARY KNOB.......................................................................................................10
7.2 cursor KEY ...................................................................................................................10
KEY 10
KEY 10
7.5 freeze KEY ...................................................................................................................10
7.6 print KEY .....................................................................................................................11
7.7 speaker KEY ................................................................................................................11
8 PARAMETERS...................................................................................................................... 11
8.1 PARAMETER SELECTION....................................................................................................11
8.2 BEAT MODE PARAMETER ..............................................................................................11
8.3 BEATS PER HOUR PARAMETER ......................................................................................12
8.4 MEAS. TIME PARAMETER ..............................................................................................12
8.5 LIFT ANGLE PARAMETER ...............................................................................................13
8.6 GRAPH RESOL. PARAMETER ..........................................................................................13
8.7 TEST MODE PARAMETER ...............................................................................................13
9 THE DIAGRAM ................................................................................................................... 15
11 ACOUSTIC CHECK.............................................................................................................. 16
17 ACCESSORIES...................................................................................................................... 28
18 CALIBRATION ...................................................................................................................... 28
By buying the WICOMETRE PROFESSIONAL you have chosen a test instrument which combines
the highest of technical standards with an operating convenience specifically designed for practical
applications.
If you operate it correctly, your new instrument will give you many years of service. We wish you
much pleasure and success in its application.
1 DESCRIPTION
The WICOMETRE PROFESSIONAL fulfills the highest requirements for testing mechanical
watches. Together with the automatic microphone MICROMAT P, it is an ideal item of test
equipment for final testing, rate adjustment and for laboratory test purposes.
• The diagram showing the watch beats is indispensable for trouble-shooting. The WICOMETRE
PROFESSIONAL records this diagram noiselessly on the LCD graphic screen.
• Selectable resolution for the diagram (magnifier).
• Integrated amplitude measurement. No additional appliance is necessary for this measurement
which is important for quality control.
• Automatic analysis of the diagram to provide a numerical display of the measured values for
rate, amplitude and beat error.
• Selectable measuring times for numerical measurements.
• Automatic gain control, manually adjustable.
• Automatic selection of the beat number for all the currently used beat numbers; manual
selection for any special beat numbers.
• Storage of the selected parameters after the equipment is switched off.
• Three different oscilloscope functions allow the presentation of the beat noises of the watch in a
graphic form. The source of an error can be directly recognized with the help of our booklet
giving details of watch movements and "Scope" functions.
• A connector is provided for the automatic microphone MICROMAT P. All the measuring
positions are programmable in different combinations.
• A beat-rate dependent stabilization time is introduced after each automatic change of position
before the start of a measuring cycle.
• The display can be stored which allows the diagram and the results to be observed and analyzed
for as long as desired.
• Printer-port for the printout of all numerical or graphical results.
• Automatic printout of the relevant results after the selected measuring time has elapsed.
• Built-in loudspeaker to hear the beat noise. Can be switched ON and OFF.
Basic operation of the WICOMETRE PROFESSIONAL is very simple as long as you are satisfied with
the elementary functions of the instrument. However, should you wish to avail yourself of the
numerous additional facilities offered by the WICOMETRE PROFESSIONAL, you should read the
following instruction manual carefully.
Please carefully read all the information given in this User’s Guide. It
provides you with important instructions about the use, safety and
maintenance of your instrument. Keep this manual in a safe place and if
the occasion should arise, pass it on to subsequent users.
The instrument may only be used for the designed purpose in accordance with this Operating
Instructions
THE MANUFACTURER,
Rear Panel
power
9.0 VAC Socket for the mains adapter.
1.2 A
printer Interface RS232 for printer connection. For the acquisition of the
output results and long time analysis, a PC may be connected.
BEFORE CONNECTING THE INSTRUMENT FOR THE FIRST TIME, CHECK THAT
THE MAINS ADAPTER UNIT VOLTAGE CORRESPONDS TO YOUR MAINS SUPPLY
VOLTAGE.
Connect the mains adapter unit to the corresponding socket for the mains adapter.
Caution! Check whether the voltage stated on the printer corresponds to the local mains
voltage.
Switch on the instrument with the level rotary knob and set the index line to I. The following
display appears after switching on:
Attention:
Instead of STAB-MICROMAT H-V/V-H 1T , the display may show STAB- MICROMAT H-V/V-H
2T. The differences are described in chapter 15.3 STABILIZATION TIME, page 24.
Thanks to the integral parameter memory, the last-programmed settings remain stored from the
last time the instrument was turned off.
Caution: The time base of the WICOMETRE PROFESSIONAL reaches its full stability about
2 minutes after being switched on (temperature stabilization of the quartz crystal).
If your instrument is equipped with the printer, do not forget to switch on the printer, too.
Your instrument is now ready to operate. Chapter 8 give details of how you can change the
parameters.
Clamp a watch onto the microphone, preferably in such a way that the crown lies between the
metal leads of the signal sensor, and then rotate the microphone to the required test position.
Please refer to chapter 15 OPERATING THE MICROMAT P if you intend using the automatic
multi-position microphone MICROMAT P
The sig LED flashes to the rhythm of the beats of the watch and indicates that a signal is being
received.
7.3 Ï KEY
The next higher value for a parameter is selected by pressing the Ï key as follows:
• step by step each time the key is briefly pressed
• continuously when the key is kept in the depressed position
• quickly if the Ï key is also pressed while the Ð key is in the depressed position.
7.4 Ð KEY
The next lower value for a parameter is selected by pressing the Ð key as follows:
• step by step each time the key is briefly pressed
• continuously when the key is kept in the depressed position
• quickly if the Ð key is also pressed while the Ï key is in the depressed position.
8 PARAMETERS
Use the key marked cursor to select a parameter which should be changed. The selected
parameter appears in inverse form (black background).
The BEATS PER HOUR parameter cannot be chosen if the parameter BEAT MODE is set to AUT
or FRQ as it is automatically selected.
The functions AUT, MAN, SEL or FRQ can be called with the Ï and Ð keys for the selection of
the beat number.
AUT Automatic selection: in this position the instrument analyses the beat number of the
watch which has to be measured. The following beat numbers are selected automatically:
12000, 14400, 18000, 21600, 28800 and 36000
MAN Manual selection: this position enables the less common beat numbers and those that
cannot be determined automatically to be selected for the BEATS PER HOUR parameter.
List of selectable beat-numbers
6000 7200 7380 7440 7800 9000 9100 10800 11880
12000 12342 12480 12600 13320 13440 13500 14000 14040
14160 14200 14280 14400 14520 14580 14760 14850 15000
15360 15600 16200 16320 16800 17196 17258 17280 17786
17897 18000 18049 18514 19332 19440 19800 20160 20222
20944 21000 21031 21306 21600 28800 36000
2. Check the watch to be regulated by reading the time over a period of 24 hours and noting
the deviation, e.g. +180 s/24h. Select the FRQ function. Once the WICOMETRE
PROFESSIONAL has calculated the beat number and the display is stable, select the SEL
function by means of the Ð key. The measured beat number is now stored. Regulate the
watch to give a reading of -180 s/24h. Go back to the FRQ function and the instrument
should show the correct beat number for the watch.
Because of the display principle used for the LCD-graphic screen, the following distinctions have to
be taken into account in comparison with diagram recording on paper strips:
The dots of the diagram are represented by the grid points (pixels) on the screen. These dots are
spaced at a distance of 0.5 mm. A sloping diagram is therefore always displayed as a staircase with
steps of 0.5 mm with the distance between the lines being a multiple of 0.5 mm only.
This effect occurs with high precision watches only and with a low diagram resolution. This effect
is normally obscured by the natural irregularities of the diagram. We therefore recommend working
with the highest possible resolution appropriate to the quality of the watch concerned.
As one dot is recorded for every beat of the watch and the dots have a fixed distance of 0.5 mm,
the recording speed as well as the slope of the diagram depend on the beat number of the
particular watch.
The horizontal recording speed is automatically reduced to ensure that the entire evaluated part of
the diagram remains visible during long-term measurements. This is achieved by not recording all
the watch beats but only every third, fifth, or seventh one.
The watch beats are always evaluated over a time period of 2 seconds for all numerical results.
If a longer measuring time (parameter MEAS. TIME) is set then the result will be the average
value of all the individual measurements. The average value is re-calculated continuously every 2 s.
Thus, for example: the last 5 measurements over 2 s periods are taken into account to determine
the average value for a measuring time of 10 s. The evaluated part of the diagram is marked by an
underscore.
Each diagram must be clean and even if the results obtained are to be significant and stable.
The longer the measuring time selected, the more stable the result displayed will be since
short-term rate irregularities will be compensated.
The beat noise can be checked for irregularities and stray noises by means of the built-in
loudspeaker.
The loudspeaker may be switched ON or OFF by pressing the speaker key.
An acoustic check is recommended whenever a clean diagram is not formed with the normal
amplification adjustment or when the diagram shows irregularities which are not easily explained.
The beat noise can also be analyzed in graphic form by using the 3 different SCOPE-functions.
(See chapter 13 OSCILLOSCOPE FUNCTIONS, page 17).
The following sections only apply if the instrument is equipped with a printer.
The oscilloscope function allows the acoustical beat noise of a watch to be presented in a graphic
form.
The beat noise is presented in the form of an envelope curve showing the maximum and minimum
signal amplitudes. The numeric results are not displayed in the SCOPE-functions.
The amplitude of the oscillograms can be changed with the level control.
This presentation makes errors in the escapement or stray noises visible immediately and with
some experience the watchmaker should be able to recognize the source of the error.
The oscillograms can also be stored for closer examination by pressing the freeze key and, if
required, they can be printed out.
13.1 SCOPE 1
The display covers a period of 20 ms with this mode of oscillogram presentation.
The entire beat noise is visible on the screen. Triggering is so arranged that a 'tick' and a 'tock' are
recorded alternately.
This type of presentation is suitable for analyzing beat noises.
0 5 10 15 20 ms
SCOPE1
SCOPE2
13.3 SCOPE 3
The display covers a period of 420 ms with this mode of oscillogram presentation.
Three to five consecutive beat noises are presented on the screen, according to the beat number.
The presentation starts alternately with a 'tick' and a 'tock'.
SCOPE3
The following information is intended to help understand the operation of the WICOMETRE
PROFESSIONAL and to interpret the display correctly.
• The first noise results when the ellipse on the balance wheel touches the fork (A). This beat is
time-wise very precise and is therefore used for producing the diagram as well as for calculating
the rate and beat error.
• The contact of the escape wheel with the entry pallet produces a second pulse (B). This noise is
very irregular and is not used for evaluation purposes.
• The third and most intense noise is produced by the escape wheel striking the exit pallet (C).
This noise is used for calculating the amplitude.
A B C
A
B
SCOPE1
Besides the above mentioned pulses, friction and stray noises also occur to a more or less degree.
It is important that the beginning of the first noise burst (A) is reliably detected for evaluation
purposes. If this first noise burst is weak for a particular watch, or if the watch produces strong
stray noises, then the amplification should be adjusted accordingly.
It should always be remembered that the beat noise is not composed of single sharp pulses but
rather of numerous mechanical noises which increase to a maximum and then decrease again. The
measured times therefore show a certain dependence on the amplification adjustment and on the
instantaneous signal form
Of course, the diagram not only shows the rate but also other time irregularities in the watch
beats, for example: beat errors, defective teeth on the escape wheel, etc.
14.5 AMPLITUDE
The time between the pulse A and pulse C of the beat noise is measured to calculate the amplitude
(see Figs. 6 and 7).
The balance wheel rotates through a particular angle between these two pulses. This so-called "lift
angle" is determined by the design of the watch movement and must be set as parameter. The
higher the amplitude of the balance wheel, the higher the speed with which it passes through this
lift angle and hence the shorter the time it needs to traverse this angle.
The amplitude can therefore be calculated from the time between pulse A and C of the beat noise
taking the beat number and lift angle into consideration. The WICOMETRE PROFESSIONAL always
calculates the average value of the amplitude over 2 seconds.
Example:
The measurement sequence can be interrupted at any point by pressing the reset button which
also causes the microphone to return to its starting position. Any measurement results already
obtained are displayed.
There is a slide switch on the back of the MICROMAT P which has two positions:
norm. Once a measurement sequence is over the start button has to be pressed again to
trigger a new one.
cont. Once a measurement sequence is over the next one is stared automatically after a
pause of 60 minutes.
18'000 19,2 s
19'800 17,45 s
21'600 16 s
28'800 12 s
36'000 9,5 s
All the results are automatically printed out in tabular form at the end of a measurement sequence
if a printer is connected to the instrument. The width of the paper strip is 70 mm.
Example of a printout:
R E F _____________________________
Measuring possibilities:
• Rate deviation, beat error and amplitude of mechanical watches. Diagram of the beat noises.
• Rate deviation of analogue quartz watches by means of an optional signal pick-up.
Beat number:
• Automatic determination of all common beat numbers. Manual selection for less common beat
numbers. Manual selection for any beat number in the range between 3,600 b/h to 36,000b/h.
Automatic determination of a unknown beat number.
Gain control:
• Automatic gain control. Manual control facility for watches with stray or unusual beat noises.
Diagram:
• Presentation of the diagram on LCD graphic display, resolution of 96 x 240 points.
• Time scale: selectable between 1 mm/ms to 10 mm/ms. Diagram recording speed: depends on
the beat number, 0.5 mm per watch beat. The recording speed is automatically slowed for long
measuring times so that the entire evaluated part of the diagram always remains visible.
• visible diagram length: 120 mm.
Rate measurement:
• Numeric display in s/24h.
• Selectable resolution : 1 s/24h or 0.1 s/24h. Measuring range: ±999 s/24h or ±99.9 s/24h. For
measurements of quartz watches by means with the EXPERT Q: resolution 0.01 s/24h.
Measuring range ± 9.9 s/24h.
Beat error:
• Numerical display in milliseconds.
• Resolution: 0.1 ms, measuring range: 9.9 ms.
Amplitude measurement:
• Numerical display in degrees.
• Resolution: 1o , measuring range: 80o to 360o.
• Lift angle: adjustable from 10o to 90o.
Freeze function:
• Enables the current screen-display to be stored for as long as required at any time.
Oscillogram:
• Presentation of the beat noise in form of an envelope curve of the amplitude values.
• Non-linear signal amplification for better display of low signal amplitudes.
• 3 different time deviations, 20 ms, 220 ms and 420 ms for full screen length.
Protocol print-out:
• Interface RS232: printer and PC (for long time analysis) connection.
• Printout of the numeric results.
• By using the MICROMAT P: print-out of the average and delta values of the programmed test
positions include the results.
• Graphic printout of the full screen presentation with diagram or oscillogram.
Time base:
• Pre-aged thermo-compensated high frequency quartz crystal time base, precision +/- 0.02 s/24h.
Display:
• LCD-graphic display, 128 x 240 dots.
Casing:
• Hardwood casing, scratch-proof aluminum front panel and steel base-plate with support.
• Dimensions: 260 x 175 x 80 mm.
• Weight: 2.2 kg without microphone.
Mains connection:
Plug supply adapter 230 V~ or 120 V~, 11VA.
DECLARATION OF CONFORMITY
• The equipment is in conformity with the following EC-Directives:
89/336/EWG EMC
EN 55011 : 1991 Emissions
EN 50082-1 : 1992 Immunity, public environments
89/392/EWG Machinery
EN 292 : 1991
18 CALIBRATION
We recommend that the instrument should be calibrated and checked once a year to guarantee its
measuring accuracy.
Please contact the customer service department at our head office or one of our representatives.