Datalogger Usb Humedad y Temperatura Ref. 15079667

The document provides operational instructions for an LCD display device that measures temperature and humidity, detailing setup, alarm functions, data logging, and memory management. It includes specifications for temperature and humidity ranges, battery requirements, and data download procedures. Additionally, it outlines warranty and service information, as well as installation instructions for a static suppressor to protect against radio frequency interference.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
25 views2 pages

Datalogger Usb Humedad y Temperatura Ref. 15079667

The document provides operational instructions for an LCD display device that measures temperature and humidity, detailing setup, alarm functions, data logging, and memory management. It includes specifications for temperature and humidity ranges, battery requirements, and data download procedures. Additionally, it outlines warranty and service information, as well as installation instructions for a static suppressor to protect against radio frequency interference.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

LCD DISPLAY: probe.

probe. Once ▲/▼button is released from long press, the step which alarmed, the alarm set point that was triggered,
LCD View Angle: 12 o’clock sizes are back to default. the date and time the channel reading went out of
h. All setup items, time, alarm settings are stored range, and the date and time the channel reading
FIGURE 1
in the device’s memory, and will not be lost if the returned to within range.
SIDE VIEW device is powered off. While running, the device
VIEW THE MEMORY CAPACITY/USAGE
stores time to memory once a minute.
1. Slide DISPLAY switch to MEM position
ALARM
ALARM 2. First line displays current percentage of memory full,
1. Once an alarm is triggered the corresponding the second line displays the number of days remaining
ON OFF
parameter will flash on LCD and alarm will beep. before memory is full at the current data logging
2. Alarm will continue to beep for 60 seconds and then interval, and the third line displays current date/time.
DISPLAY

MIN ALM MEM


MAX
will beep every 15 seconds. Note: For Excursion-Trac: the MEM symbol will become
SET/SEL 3. Press CLEAR button to clear alarm, if both LCD active on the display when the memory is full. Once
flashing and beeping on current display, INT or EXT. memory is full, the oldest data points will be overwritten
CLEAR If an alarm is triggered on both INT and EXT, press with new data points.
CLEAR to clear alarm on current display, then press For Memory-Loc: MEM symbol will become active on
°C/°F SET/SEL button to select other display, and press display when the memory is 95% full. Once memory is
CLEAR button to clear the alarm. full, device will stop logging new data points.
4. If temperature or humidity falls back to normal range,
CLEAR THE MEMORY
and then back to out of alarm range, alarm will be
1. Slide DISPLAY switch to MEM position.
triggered again.
For Excursion-Trac: Press and hold CLEAR button
SELECTING UNIT OF MEASURE: for 2 seconds until LCD displays “donEE”, then release
1. Slide DISPLAY switch to MIN/MAX or ALM. the button.
2. Press ▲(C/F) to select Celsius or Fahrenheit. For Memory-Loc: Memory cannot be cleared.
VIEW CURRENT MIN/MAX VIEW UNIQUE DEVICE ID NUMBER
1. Minimum values stored in memory are the minimum 1. Slide DISPLAY switch to MEM position.
TEMPERATURE: OPERATION INSTRUCTION:
readings measured since the minimum memory was 2. Press ▼button once, second line will display first eight
Range: 1. Setup device:
cleared. The maximum values stored in memory are digits of ID number, and ‘—’ is displayed at end.
Ambient: 0 to 65°C, a. Set-up Mode: Press and hold SET button for 3
the maximum readings measured since the maximum 3. Press ▼button again, second line will display last
External Probe: –20 to 70°C seconds, release button once YEAR digits start
memory was cleared. eight digits of ID number.
Resolution: 0.1°C flashing.
2. Min/Max values are not programmable; neither are 4. Press ▼button a third time to return to default display.
Accuracy: ±0.4°C b. Press ▲or ▼button to adjust to current year.
they stored in permanent memory. Min/Max values are
Recommended Temperature Operating Range: 0 to 65°C. c. Press and hold ▲or ▼button for 2 seconds, the DOWNLOAD DATA
reset upon power cycle.
value will increase or decrease faster. 1. USB download will only occur if AC Adaptor LCD
RELATIVE HUMIDITY: 3. Slide DISPLAY switch to MIN/MAX position.
d. Once YEAR is set, press and release SET button symbol is shown on display. Plug supplied AC adaptor
Range: 0 to 95% RH, non-condensing 4. Current Min/Max values are displayed.
(don’t hold SET button) to move forward to next into device to provide sufficient power for USB
Resolution: 0.1% RH 5. To clear current Min/Max values, while DISPLAY
setup item. operation.
Accuracy: ±3% RH between 5 to 75%, otherwise ±5% RH switch is in MIN/MAX position, press CLEAR button
e. Setup items move forward in the following order: 2. Data can be downloaded directly to a USB Mass
Recommended Humidity Operating Range: 20 to 80% RH once.
Year -> Month -> Day -> Hour -> Minute -> AM/ Storage Flash Drive. To begin download, insert USB
(non-condensing) over –10°C to 60°C.
PM -> Data Logging Interval -> Internal Humidity VIEW ALARM LIMITS flash drive into USB port located on right side of
BATTERY: 3 AAA Alkaline batteries (4.5V) Low Alarm -> Internal Humidity High Alarm -> 1. Slide DISPLAY switch to ALM position. device.
Internal Temperature Low Alarm -> Internal 2. Alarm setting are displayed on LCD. 3. Download will begin upon insert. USB LCD symbol will
DATA UPDATE FREQUENCY:
Temperature High Alarm -> External Humidity Low appear on display to indicate download has begun.
Temperature and humidity: updated every 10 seconds; ENABLE/DISABLE ALARMS
Alarm -> External Humidity High Alarm -> External Wait up to 60 seconds for USB LCD symbol to appear
Note: if a reading is out of operating range, the 1. Slide ALM switch to ON or OFF to enable or disable
Temperature Low Alarm -> External Temperature after insertion of flash drive. Flash drives with larger
corresponding location on LCD will display ‘----‘, and such the alarms.
High Alarm -> Exit Setup Mode.\ amounts of data stored on drive prior to downloading
out of range reading will not trigger alarm. 2. Alarms are enabled or disabled for all channels, can’t
f. While in Setup mode, press and hold SET button will take longer to begin the download.
be enabled or disabled individually.
Data Logging Frequency: 2 min by default, user- for 3 seconds to exit setup mode. Or while digits 4. Once download process is complete, USB LCD
adjustable between 1 min and 12 hours with 1 min step. are flashing, and no SET/▲/▼button pressed for DATA LOGGING OPERATION symbol will disappear from display and a short beep
15s, the device will exit Setup mode. 1. Device will continuously log measurements from all will sound. Do not remove USB drive until process is
Data Storage Capacity: Excursion-Trac: 525,504 data
g. While setting temperature and humidity alarms, channels in memory in user-specified intervals. Each complete.
records, 2 years for two-minute logging frequency
default step sizes are: data point contains humidity readings and temperature 5. Data transfer rate is approximately 200 data points per
Memory-Loc: 688,086 data records, 2 years and 7 months
Temperature: 0.1 readings. second.
for two-minute logging frequency.
Humidity: 1 2. Note: All data stored in %RH and Celsius. Date/time Note: DO NOT leave USB Mass Storage Flash Drive
VIEW CURRENT READING Data logging interval: 1 min format is MM/DD/YYYY. inserted into unit. Insert, DOWNLOAD, and then remove.
1. If readings are read from ambient sensor, INT LCD If press and hold ▲or ▼button for 2 seconds, the step 3. Note: DO Not LEAVE USB Mass Storage Flash Drive Device cannot continuously write to a USB.
symbol will appear on display; If readings are read sizes are changed to following for fast adjustment: inserted in the unit. Insert, DOWNLOAD, and then
REVIEW STORED DATA
from probe sensor, EXT LCD symbol will appear on Temperature: 1 remove. Device cannot continuously write to a USB.
1. Downloaded data is stored in a comma-delimited CSV
display. Humidity: 10 4. Device will also store the most recent 30 alarm events.
file on flash drive. Filename naming convention is
2. Press SET/SEL button to toggle between ambient and Data logging interval: 10 min Each alarm event data point contains channel number
“D1D2D3D4D5D6D7R1.CSV” where D1 through D7
are the last seven digits of the thermometer’s unique WARRANTY, SERVICE, OR RECALIBRATION STATIC SUPPRESSOR INSTALLATION
ID number and R1 is the revision of file starting with For warranty, service, or recalibration, contact: Static generated, radio frequency can affect any cable
letter “A”. TRACEABLE® PRODUCTS through the air or by physical contact. To protect against
2. If more than one file is written from same device to 12554 Old Galveston Rd. Suite B230 radio frequency, install a suppressor onto your ther-
USB flash drive, revision letter will be incremented to Webster, Texas 77598 USA mometers cable to absorb radio frequency as follows: TRACEABLE®
preserve previously downloaded files. Ph. 281 482-1714 • Fax 281 482-9448
3. Data file can be opened in any software package E-mail [email protected] • www.traceable.com EXCURSION-TRAC™
supporting comma-delimited files including
spreadsheet software (Excel) and text editors. Traceable® Products is ISO 9001:2015 Quality-  & MEMORY-LOC™
DATALOGGING
4. File will contain device unique ID number, the most Certified by DNV and ISO/IEC 17025:2005
recent 30 alarm events, and all stored temperature accredited as a Calibration Laboratory by A2LA. HYGROMETER
readings with date and time stamps. THERMOMETER
Note: All data stored in %RH, Celsius. Date/time format
is MM/DD/YYYY
INSTRUCTIONS
DISPLAY MESSAGE
1. If no buttons are pressed and --.-- appears on display, Lay the cable along the center of the
this indicates reading being measured is outside of the
suppressor with the connector to your left.
operating range of device, or sensor is damaged
LOW BATTERY DISPLAY
If battery falls below 3.4V, Low Battery symbol will display 
on LCD. Device checks battery voltage every two hours.

Loop the right end of the cable under the


suppressor and back up again laying the cable along
the center of the suppressor.

Carefully, snap the two halves together with the looped


cable routed through the center.

Cat. No. 6452


Traceable® is a registered trademark of Cole-Parmer.
This completes the installation of the suppressor.
Excursion-Trac™ and Memory-Loc™ are trademarks of
Cole-Parmer.
©2018 Traceable® Products. 92-6452-00 Rev. 1 020618