NO Nama Alat & Spesifikasi
NO Nama Alat & Spesifikasi
1. ADS II Anesthesia Machine with Touch Screen Ventilator Include Two Vaporizers
Type : ADS II
Merk : INFINIUM - USA
The ADS II is an advanced yet easy to use anesthesia workstation that provides accurate, pneumatically
driven and electronically controlled ventilation.
The ADS II has an ergonomic design that incorporates new technology and provides safe and effective
treatment options for the clinician. The ADS II includes Adult and Pediatric modes that provide patient-
appropriate default and ranges.
Gas Delivery :
Electronic flowmeters for O2, N2O, and air have safety control and LED displays. A mechanical total flowmeter
provides security backup.
Absorber :
BY-PASS soda–lime exchange during surgery. Auto heating & water collection.
Auto compensation of tidal volume & automatic AGSS.
Vaporizers :
Precise high quality vaporizers with automatic compensation for humidity, pressure, and flow.
TECHNICAL SPECIFICATIONS :
Physical Specifications :
Page 1
NO NAMA ALAT & SPESIFIKASI
- Dimensions and weight :
Dimensions (HxWxD) : 1450mm×1000mm×1330mm
Weight : 108 kg (Without vaporizer and cylinder)
- Top Shelf :
Weight limit : 34 kg
Width x Depth : 578×360 mm
- Work Surface :
Dimensions (HxWxD) : 827×557×311 mm
- Drawer (Internal Dimension) :
Dimensions (HxWxD) : 150×298×348 mm
Qunatity : Standard 2
- Caster :
Diameter : 125mm
Brakes : All four caster with brakes
Ventilation Specifications :
- Modes of Ventilation :
Ventilation mode : VCV SIMV(P)+PS
PCV CPAP/PSV,
PRVC Manual
SIMV(V)+PS Standby
- Compensation :
Circuit gas leakage compensation and automatic compliance compensation
- Ventilation Parameters :
Patient Type : Adult, Pediatric
Tidal volume : Pediatric: 10 ~ 300ml
Adult: 100 ~ 1500ml
(increments of 5 mL)
Pinsp : 5 ~ 70 cmH2O (increments of 1 cmH2O)
Plimit : 5 ~ 100 cmH2O (increments of 1 cmH2O)
Freq : Freq in SIMV mode: 1 ~ 40 bpm
Other modes: 4 ~ 100 bpm
(increments of 1 bpm)
I:E : 4:1 ~ 1:10 (increments of 0.5)
Tpause : OFF, 5% ~ 50% (increments of 5%)
Ti : 0.1 ~ 10 s (increments of 0.1s)
Flow trigger : 1 ~ 15 L/min (increments of 1 L/min)
Psupp : 5 ~ 60 cmH2O (increments of 1 cmH2O)
Page 2
NO NAMA ALAT & SPESIFIKASI
Ppeak : 0 ~ 100 cmH2O
Pmean : 0 ~ 100 cmH2O
Pplate : 0 ~ 100 cmH2O
I:E : 4:1 ~ 1:10
Rate : 0 ~100 bpm
Freqspont : 0 ~ 99 bpm
PEEP : 0 ~ 70 cmH2O
Resistance ® : 0 ~ 200 cmH2O/(L/s)
Compliance © : 0 ~ 200 ml/ cmH2O
- Control Accuracy :
Volume delivery : < 100 ml: ± 20 ml
≥100 ml: ± 20 ml or ± 15% of the set value, whichever is greater
Pressure delivery : ± 2.0 cmH2O or ± 10% of the set value, whichever is greater
PEEP delivery : ± 2.0 cmH2O or ±10% of the set value, whichever is greater
Flow trigger : ± 1.0 L/min or ±15% of the set value, whichever is greater
- Monitoring Accuracy :
Volume monitoring : < 100 ml: ± 20 ml
≥100 ml: ± 20 ml or ± 15% of the reading, whichever is greater
Pressure monitoring : ± 2.0 cmH2O or ± 10% of the reading, whichever is greater
PEEP monitoring : ± 2.0 cmH2O or ± 10% of the reading, whichever is greater
MV monitoring : 1L/min or ±15% of the reading, whichever is greater
- Trend Graph :
Continuous trend information for the latest 24 hours
- Log :
500 events storage, first in first out
- Alarm Settings :
Tidal volume High : 20 ~ 1500 mL, OFF
Low : OFF, 10 ~ 1500 ML
Minute volume High : 1 ~ 40 L/min, OFF
Low : OFF, 0 ~ 40 L/min
Airway pressure High : 1 ~ 100 cmH2O
Low : 0 ~ 99 cmH2O
Frequency High : 1 ~ 100BPM
Low : ~ 99BPM
FiO2 High : 21% ~ 100%,OFF
Low : OFF, 18% ~ 99%
Apnea alarm : 10 ~ 40s
Sustained airway pressure alarm : 15s
Negative pressure alarm : Paw < -10 cmH2O
Alarm silence countdown timer : 120 to 0 seconds
- System Language :
English, Spanish, Italian
Ventilator Component
- Flow Sensor :
Type : Variable orifice flow sensor
Position : Inspiratory and expiratory port
- Oxygen Sensor :
Type : Chemical
FiO2 displayed : 21% to 100%
Accuracy : ± (volume fraction of 2.5 % +2.5 % gas level)
Response time : ≤15 seconds
Page 3
NO NAMA ALAT & SPESIFIKASI
- Ventilator Screen :
Display type : Color TFT touch screen, rotatable
Display size : 15 inch
Resolution : 1024 x 768
Display parameter : All setting and alarm parameters (including Breath rate, I/E ratio,
Tidal volume, Minute volume, PEEP, MEAN, PEAK, PLAT,
and O2 concentration, EtCO2, N2O, Aesthesia gas concentration)
Display waveform : P-T, F-T, V-T, CO2-T
Loop : P-V, F-V and F-P
Timer : On screen timer
- Communication Port :
RS-232 connector, VGA x1
Vaporizer
Vaporizer brand : DRAGER Anesthetic Vaporizer or Penlon Vaporizer
Support agent : Halothane, Enflurane, Isoflurane, Sevoflurane
Position : Standard 2
Mounting mode : Selectatec®, with interlocking function
Fill method : Key fill, Pour fill, Quick fill
Electrical Specifications
- Power and Battery Backup :
Power input : 110 ~ 240 Vac, 50/60 Hz
Auxiliary electrical outlet : Up to 3 outlets (2A for each)
Battery backup : 60 min for 1 piece battery (powered by new fully-charged
batteries with 25℃ ambient temperature)
Battery type : Build-in Li-ion battery, 11.1 VDC, 7800 mAh
Safety feature : In case of electricity and battery failure, manual ventilation, gas delivery
and agent delivery are possible.
Pneumatic Specifications
- ACGO (Auxiliary Common Gas Outlet) :
Connector size : ISO 22 mm OD and 15 mm ID
- Pipeline Supply :
Gas type : O2, N2O, Air
Pipeline input range : 280 to 600 kPa
Pipeline connection : NIST
- Pipeline Supply Pressure Gauge :
Display type : Mechanical
Range : 0 to 1MPa
Accuracy : ± (4% of the full scale reading + 8% of the actual reading)
- Cylinder Supply :
Cylinder type : E-size Cylinder (US standard or UK standard)
O2 input range : 400 to 450 Mpa
N2O input range : 400 to 450 Mpa
Cylinder connection : Pin-Index Safety System (PISS)
Yoke configuration : O2, N2O
- Cylinder Supply Pressure Gauge :
Display type : Mechanical
O2 range : 0 to 25 Mpa
N2O range : 0 to 25 Mpa
Page 4
NO NAMA ALAT & SPESIFIKASI
Accuracy : ± (4% of the full scale reading + 8% of the actual reading)
- O2 Control :
Method : N2O shut off with loss of O2 pressure
O2 Flush : 25 ~ 75 L/min
- O2-N2O Link System :
Type : Mechanical
Range : O2 concentration not lower than 21%
- Auxiliary O2 Flowmeter :
Range : ~ 15 L/min
Indicator : Flow tube
- Electronic Flow Meter :
O2 flow range : 0~ 10 L/min
Air flow range : 0~ 10 L/min
N2O flow range : 0~ 10 L/min
Accuracy : between -10% and +10% of the indicated value (under 20℃ and
101.3 kPa, for flow between 10% and 100% of full scale)
±200mL/min(for flow below 10% of full scale)
Environmental Specifications
- Temperature - Operation : 10 ~ 40℃
Storage and -20 ~ 55℃
- :
transport
Relative humidity non-condensing)
- Operation : ≤ 80% R.H.
-
Storage and
- : ≤ 93% R.H.
transport
- Atmospheric pressure - Operation : 70 ~ 106 kPa
Storage and
- :
transport 50 ~ 106 kPa
- Electromagnetic Compatibility
Immunity : Complies with all requirements of IEC 60601-1-2
Emission : Complies with all requirements of IEC 60601-1-2
Page 5
NO NAMA ALAT & SPESIFIKASI
Accuracy : ± 10 cmH2O or ± 15% of the setting value, which is greater
Accessories Included :
- 2 Vaporizer
- Bellows
- Breathing Circuit
- Autoclavable CO2 Absorber
- Breathing Bag
- Regulator O2 / CNU O2
- Regulator Air / CNU Air
- Regulator N2O / CNU N2O
- Soda Lime
- Face Mask
- Reservoir Bag 3L
- Jackson Rees Disposable
- User Manual