NOTIFIED BODY: TUV PRODUCT SERVICE GMBH

EC CERTIFICATE NO.: G1 08 06 46736 003

POX-100BT

High Powered Mobile X-ray System

Product Manual (Operation & )


Product Manual: POX-100BT

TABLE OF CONTENTS

PART I. OPERATION MANUAL

1. Gener
General Description ---------------------------
---------------------------------------------------------
------------------------------------------
------------ 3
2. Notice ffo
or S
Saafe Us
Use -------------------------------------------------------------------- 5
3. Operating Environment --------------------------------------------------------------------- 6
4. Radia
Radiation Safety --------------------------------------------------------------------- 7
5. Compo
Composition --------------------------------------------------------------------- 8
6. Sp
Specification --------------------------------------------------------------------- 9
7. Name of Each Part --------------------------------------------------------------------- 16
8. Operation --------------------------------------------------------------------- 19
9. How to Operate --------------------------------------------------------------------- 21
10. Remote Control Function ----
---------
-------
---------
-------
---------
-------
--------
--------
-------
---------
-------
-------
-------
------ 25

PART II. SERVICE MANUAL


1. Notice
Notice -------------------------------------------------------------------- 29
2. Regu
Regular Maintenance
Maintenance ------------------------------------------------
----------------------- ----------------------------------------------
--------------------- 30
3. Block Diagram --------------------------------------------------------------------- 31
4. Calibr
Calibration Procedures -------------------------
-------------------------------------------------------
--------------------------------------------
-------------- 32
5 Error Code and Resolution -------------------------------------------------------
------------------------- --------------------------------------------
-------------- 37
6. Radia
Radiation Dose ---------------------------------------------------------
----------------------------- ----------------------------------------
------------ 39
7. Error Code and Resolution -------------------------
-----------------------------------------------------
--------------------------------------------
---------------- 40
8 Schema
Schematics -----------------------------------------------------
------------------------- --------------------------------------------
---------------- 42

2/47
Product Manual: POX-100BT

PART 1: OPERATION MANUAL

1. Gene
Generr al Des
Descri
cripti
ption
on

1) Outline

Thank you for purchasing POX-100BT, high powered mobile x-ray syste
syste m. POX-10
POX-100BT
0BT is the most
most
powerful high frequency x-ray machine. Our mobile systems are ripple free high frequency type
type with
40~50% stronger power output efficiency compared with conventional x-ray systems. Therefore, it can
be called as the perfect concept of radiography system providing easy operation and mobility
bility fo
forr
indoor and outdoor radiography.

2) Unique Features

-High frequency type x-ray generator, high efficiency


-Maximum output: 110kV / 100mA
-750 APR memories
-Remote control software
-Equipped digital reverse display function
-Electrical requirement: 220VAC, 50/60HZ

3) Product Manual

This consolidated manual consists of operating and service instruction. Product m anual provides
POX-100BT operating
operating instructions for x-ray practitioner and other medical personnel trained in the use
of x-ray equipment. Service man ual provides installat
llation
ion in struction
struction,, c
cal
alibration and routine
maintenance
maintenance for field service staffs.
This manual is not designed to replace or substitute for certified training in the radiological or medical
field. By reading this manual, the operator should be qualified in the use of this equipment.
This X-ray system is only to be used by licensed operators in diagnostic, operative and post-operative
procedures.

4) Notice For Safety

This manual provides guideline for safe use and o peration of


of the medical
medical radio
radiographic system. The
operator should be trained in the use of this equipm
equipment and refer to this man
manual in con
conjunction with
that training.
It is owner ’s responsibility to en
ensure
sure th
tha
at the system is operated
operated by properly trained and qualified

personnel who have obta from local, state, and federal authorities wher
obtained credentials from where required

3/47
Product Manual: POX-100BT

only.
Potential ha zards exist in the use of medical ele
elec
ctron
tronic
ic devi ces an
and X-ray systems. All user
users an
and
operators must understand the safet
safety and emer
emergency procedures, and oper
operating instructions written
in this manual

Poskom certifies that each system complies with the Radiation Control for Hea
Health Safety Act in Title
21, Chapter I, subchapter
subchapter J of the Code
Code of Fe
Federa
derall Regulation in the USA
USA and EN
EN 6060
6060 1, general
general

requirement
requirem ents
s for radiation
radiation prote
protectio
ction
n in diagnost
diagnostic
ic X-ray equip
equipmen
mentt for IEC 601-1
601-1 -3 in European
European
Nations.

1.EN60601-1:1990+A1:1993+A2:1995+A13:1996 Medical electrical equipment—


Part1: General requirements
requirements for safety and essential performance.
2.EN60601-2:2001/A1:2006 Medical electrical equipment—Part1-2:General
requirements
requirements for safety-collateral standard: Electromagnetic compatibil
compatibility-
ity-
Requirements and tests.
3.EN60601-2-32:1994 Medical electrical equipment—Part2:Particular
equipment—Part2:Particular requirements
for the safety of associated equipment of X-ray equipment.
4.EN60601-1-3:1994 Medical electrical equipment—Part1:General requirements for

safety—3.collateral standard: General requirements for radiation protection in


safety—3.collateral
Diagnostic X-ray equipment.
equipment.
5.EN6061-2-7:1998 Medical electrical equipme
equipment—Part2-7:Particu
nt—Part2-7:Particular
lar requirements
requirements
for the safety of high-voltage
high-voltage generators of diagnostic X-ray generators.
6.EN60601-2-28:1993 Medical electrical equipment—Part2:Particular
equipment—Part2:Particular requirements
Fo
Forr th
the
e saf
safe
ety of X-ray
X-ray sou
source
rce assem
assembli
blies
es and X-r
X-ray
ay tube assemblies for medica
ical
diagnosis.

ACOMA'S assume no respo


responsibility or liability
liability for after-sales operating and
and safety practices,
nor can it be respons
respon sible for persona
personal injury or damage resulting from
from misuse
misuse of its x-ray
systems. Please ad
address question
questions an
and com
comments reg
regardin
arding
g safety
safety issue to ACOMA'S servi
servicce
organization and/or
and/or each distributor in your region.

5) Cautions and Warnings

Possible haz
hazards are associated
associated with unsafe operation of x-ray equipments. These haz
hazards are
described on the following warning or caution signs with the safety precautions which should be taken.

CAUTION
The X-ray ra diation could be dangerous to patients and operators unless
following safety regulations are strictly observed.

4/47
Product Manual: POX-100BT

2. Notice for Safe Use

1) Equipment Direction

User’s safety, easy operation and product reliability are our major concern in designing this system of
generators. However,
However, it is advisable t o observe the following instructions and guidelines for your

additional safety in operating this equipment.


(1) This product should be operated only by or under the supervision of legally qualified persons.
(2) POX-100BT i s de signed f or the radiographi
radiographic
c u ses and n ot for fl uoroscopic o r other associated
applications.
(3) POX-100BT should be used for the diagnosis, not for the therapy.
(4) POX-100BT is specified ClassⅠtype B equipment under the IEC 601-1, 2 standards.
specified as Class

CAUTION
This equipment
equipment is not protective against ingress of water or liquid.

(5) Do not modify the equipment at your discretion and in case any modification is re quired
unavoidably, ask help of ACOMA'S or its authorized dealers for their service.
(6) This system has been
been calibrated for optimal
optimal operation.
(7) If you
you h ave any troubles with the equipm ent, tur
turnn off power immediately, and
and notify ACOMA'S or
its authorized dealers of their service and assistance.
(8) If you want
want connecting a device made by oth
other
er manufacturers to this equipment,
equipment, ask ACOMA'S or
its authorized dealers for their instructions.

2) Direction on Hazardous waste

This equipment and accessor


accessories
ies are
ar e to be disposed of safely after the life spa
pan
n of the
the mand
mand
national regulation must be observed. - This equipment and accessories are to be disposed
according to the in
inst
structions given by national regulations or ask a specialized waste service
service
company to dispose of it.

3) Direction on Electromagnetic Wave

“This e quipment has been tested and found to comply wit h the limit
limits for medical devices in IEC
60601-1-2:1994. These limit s are designed to provide reasonable protection against harmful
interference in a typical medical installatio
installation.
n.
This equipment generates
generates and can radiate radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to other de vices in the vicinity.
However, there is no gua rantee that
that interference wi
willll not occur in a particular installation. If this

equipment does cause harmful interference to other devices, which can be determined by turning

5/47
Product Manual: POX-100BT

the equipment off and on, tthe


he user is encourag
encouraged
ed to try to co rrect the interference by one or more
of the following measures:
- Reorient or relocate the receiving device.
- Increase the separation between the equipment.
- Connect the equipment into an outlet o n a circuit di fferent from tha
thatt to which the other device(s)
device(s)
are connected.
- Consult the manufacturer or field service technician for help.”

3. Working Environment

1) Maintenance Environment

For proper operation and maintenance, please avoid following places.


-Where the equipment is exposed to water vapor.
-Where the equipment is exposed to direct sunlight.
-Where the equipment is exposed to heavy dust.
-Where the equipment is exposed to high humidity.
-Where there is insufficient ventilation.
ventilation.

-Where the equipment is exposed to salty atmosphere.


-Where the equipment is exposed to chemicals or gas.

2) Operating Environment

Keep safe distance from strong vibration and maintain good environment
Operational Environment
Temperature range 10 C ~ 40 C ( 50℉ ~ 104℉)
° °

Humidity range 30% ~ 75%RH

Optimal Working Environment


Temperature range 17 C ~ 23 C ( 63℉ ~ 73℉)
° °

Humidity range 40% ~ 60%RH

Fo
Forr safe
safe sto
sto rage
rage a nd trans
transpo
port
rt atio
ation,
n, p
ple
leas
ase
e keep proper temperature, humidity and other
conditions

Storage and Transportation


Temperature
emperatur e range -25 C ~ +60 C (-1 3℉ ~ +140℉)
° °

Humidity range 10% ~ 95%RH

6/47
Product Manual: POX-100BT

4. Radiation Safety

CAUTION
The ionized radiation could be dangerous to patients and operators unless
following safety regulations are strictly observed

-Users and operators are requested to wear appropriate protecting gears and clothes.
-Users and operators are requested to be leave safe distance from radiant sources and secondary
radiation zones.
-Users and operators are requested to eliminate all unnecessary objects near the exposure zones.
-Users and operators are requested to leave at least 20cm away from the target.
-Users and operators are requested to use minimum kV and mAs values for experimental use.
-Users and operators are requested to be careful not to exceed the limited radiograms in the
exposure zone.

7/47
Product Manual: POX-100BT

5. Composition

POX-100BT is a radiolo gical device for medical


medical us es. It should be operated by qualified
qualified practitioners
practitioners
for radiographic diagnosis only. Users should comply with safety and health regulations concerning
the ionizing protection and the electrical/ mechanical safety of medical devices.

POX-100B
POX-10 0BT
T con
consis
sists
ts of follow
followin
in g parts
parts an d dev
devii ces
ces:: Foll
Foll owi
owing
ng ite
items
ms sho
shoul
uld
d be fir
firmly
mly fix
fixed
ed and
connected for regular functioning.
functioning.
1) Operation Console
2) Power Factor Controller Circuit & Inverter
I nverter
3) HV Tank including X-ray Tube
4) Collimator
5) Ha
Han
n d Swi
Switc
tch
h

Supplement
1) Operation & Service Manual
2) Power Plug

Hand Switch Remote Control

(Picture 1)
1st Stage = Prep Switch
nd
2 Stage = Exposure Switch

Hand Switch Holder

All of remote control


control functions are
are operated by depressing the
the first stage of the hand switch (ready).
(ready).

8/47
Product Manual: POX-100BT

6. Product Specifications

1) Electrical Features
Model POX-100BT
Input Power Battery
Output Power 5.0kW (Max.)

Generator Type Inverter


Switching Frequency 75kHz
Tube Voltage Current mAs
20mA 0.1mAs
40mA 0.2~0.3mAs
100mA 0.4~10mAs
40kV~50kV
80mA 12~32mAs
60mA 36~80 mAs
50mA 100mAs
20mA 0.1mAs
40mA 0.2~0.3mAs

51kV~60kV 80mA 0.4~8.0mAs


50mA 9.0~80mAs
40mA 100mAs
20mA 0.1mAs
40mA 0.2~0.3mAs
61kV~70kV 50mA 0.4~8.0mAs
Radiography
40mA 9.0~80mAs
30mA 100mAs
20mA 0.1mAs
40mA 0.2~0.3mAs
71kV~80kV 50mA 0.4~8.0mAs

40mA 9.0~21mAs
30mA 25~64mAs
20mA 0.1mAs
40mA 0.2~8.0mAs
81kV~100kV
30mA 9.0~30mAs
20mA 32~80mAs
20mA 0.1mAs
40mA 0.2~5.0mAs
101~110kV
30mA 6.4~12mAs

20mA 18~80mAs

Max. kV Deviation ±7 %
Max. mAs Deviation ±10 %

9/47
Product Manual: POX-100BT Posko m Co., Ltd.

Max. sec. Deviation ±10 %

Display 7-Segment & LED

Model ES24-12
Battery Battery Voltage 12VDC /24A X 4EA
Capacity 5.0AH / 20°C
Tube Model OX/110-15

Anode heat storage 30kJ


X-ray Tube Inherent Filtration 0.5mm Al
Target angle 15°
Focal Sport 0.6mm × 1.8mm
Total Filtration 3.0 mm Al
Battery Charger Input 110 / 220VAC, 50/60Hz
Battery Input 54VDC / 0.5A
Collimat Type Double slit type, Manual operation
or Min. X-ray Field Size 5cm X 5cm @ 1m
Max. X-ray Field Size 35cm X 35cm @ 65cm
Timer Push button illuminator with 30 sec timer
Lamp 24VDC 150W Halogen
FAN 24VDC / 0.07A
Inherent Filtration 0.5mm Al

2) Mechanical Features
Section SPCIFICATION
Output 5.0 ㎾P
Po
os kom
Generator POX-100BT PFC 6kW(380-400VDC)
6kW(380-400VD C) Poskom
Power Board POX-100BT
POX-100BT PWR Poskom
HV Tank POX-100BT TANK Poskom
kV & mA & mAs/sec digital display

-T
-Thi
hick
ckne
ness
ss(f
(for
or anim
anim al)
al) / Den
Den sity
sity(fo
(forr
human)
Console(OP) digitt al displa
digi display
y Poskom
-Prep. & Exposure LED
-Collimator lamp Switch
-APR Store Switch
Lamp: 12V100 Watt, halogen lamp,
Collimator Timer: 30sec
Ralco
(R72) X-ray Field Size: 43cm * 4 3cm @ 1m
SID

10/47
Product Manual: POX-100BT

Type: Fixed anode


Focal spot size: 0.6mm / 1.8mm
X-RAY TUBE Heat Storage: 30kJ
NAGO
(OX/110-15) Cooling rate: 250W
Inherent Filtration: 0.5mm Al
Target angel: 15°

3) Specification

Dimension
(1) Main Body

11/47
Product Manual: POX-100BT

4) Tube Specifications

(1) General Data


Operating Tube Voltage: OX/110-15 (CEI)
Tube kV range: 40 ~ 110kV
Focal spot size: 0.6mm x 1.8mm

D. Heat Storage: 30kJ


E. Cooling rate: 250W
F. Inherent Filtration: 0.5mm Al
G. Target angle: 15°

(2) Filament volt curve


A. Small Focus

B. Large Focus

12/47
Product Manual: POX-100BT

(3) Max. Output Curve


A. Small Focus

B. Large Focus

Anod
Anod e Heat
Heat Curve
Curve

3) Dimension

13/47
Product Manual: POX-100BT

(4) Bipolar Thermal Curve

4) Name of Major Parts

7.1 Name of Parts

(5) Dimension

14/47
Product Manual: POX-100BT

7. Name of Each Part

1) Main Console

1 17
16
2
15

3 14

4 13
5 12

11

10
6 9

7 8

[1] Exposure LED [10] Exposure Switch


[2] Ready LED [11] Density Up/Down Switch
[3] kV Up/Down Switch [12] sec/mAs Mode Select Switch

[4] APR Size Select Switch [13] Collimator Lamp Switch


[5] APR Mode Select Switch [14] sec/mAs Select Switch
[6] APR Store Switch [15] Battery Leftover Display
[7] APR Region Select Switch [16] Error LED
[8] APR Direction Select Switch [17] Wait LED
[9] Ready Switch - -

15/47
Product Manual: POX-100BT

2) Sub Console

1
2
3

4 8

5 6

[1] Wait LED [5] kV Up/Down Switch


[2] Exposure LED [6] mAs Up/Down Switch
[3] Ready LED [7] Collimator Lamp Switch
[4] kV Display [8] mAs Display

16/47
Product Manual: POX-100BT

Product Labels: Human (POX-100BT_H)

Warning: This X-ray unit maybe dangerous to patient and operator


unless safe exposure factors and operation instruction are observed

DIAGNOSTIC X RAY UNIT


Model: POX-100BT
Serial No: DATE:
Input Rating: 110/220V~, 50/60Hz, 5kVA
Rating: X-ray Tube Voltage: 40-110kV
X-ray Tube Current: 20-100mA
Total filtration: Min 3.0 mmAl
(Tube: 0.5 mmAl, Collimator: 0.5 mmAl, Al Filt: 2.0mmAl)
X-ray Tube : OX/110-15 Serial No:
Focal spot: 0.6mm x 1.8mm
ACOMA'S
R oom 405, Unitech Ville Venture T Town,
own, 1141-2, Baekseok-dong,
I lsandong-gu, Goyang-si Gyeonggi-do 41 411-722,
1-722, Korea
Take adequa
adequa te P rec
recaut
aution
ionss tto
o p rev
revent
ent the pos
possib
sib ili
ility
ty of a ny pe rso
rsons
ns
carelessly exposing themselves or others to radiation.

HIGH VOLTAGE TANK


MODEL: HV-100HF Serial No: DATE
DATE::
Tube Model: OX/110-15 Tube Serial No:
Focal spot: 0.6mm x 1.8mm
Inherent Filtration: 0.5mmAl
Rating: MAX. 110kV
110kV,, 40mA, 50kV
50kV,, 100mA
ACOMA'S
Room 405, Unitech Ville Venture Town, 1
1141-2,
141-2, Baekseok-dong,
Ilsandong-gu, Goyang-si Gyeonggi-do 411-722, Korea
Warning: Electric shock hazard do not remove cover

Reserve Servicing to qualified personnel.

BEAM LIMITING DEVICE COLLIMATOR)


MODEL: MCM-100BT Serial No:
Input Rating: 24VDC, 150W
Min. X-ray field size: Below 5m x 5m at SID 100cm
Max. X-ray field size: Below 35cm x 35cm at SID 65cm
Max. Tube Voltage: 110kV,
110kV, Inherent filtration: 0.5 mmAl
ACOMA'S
Room 405, Unitech Ville Venture Town 1
1141-2,
141-2, Baekseok-dong,
Ilsandong-gu, Goyang-si Gyeonggi-do 411-722, Korea

17/47
Product Manual: POX-100BT

8. Operation

1) Prep
Prep arati
aration
on

A. Wear a lea
lead
d ap
apron
ron for x-ray exposures.
exposures.
B. Leave at least 2 meters away from the unit or use e
extended
xtended ha
hand
nd switch cable in order to keep
safe distance.
C. Apply pro per beam size a nd techni cal values to produce the best radiographic results while
minimizing exposure dose.
D. Be careful about the decimal point on the mAs display in order to acquire intended mAs values.
E. For the
the in
indo
door
or radiography
radiography for the patient in t he b
bed,
ed, ask
ask visito rs to leave
leave the
the room
room for a wh ile
and keep safe distance from the patient.
F. Pay attention to proper management of the equipment and careful maintenance.
G. Cumulative radiation dose should not exceed the recommended amount. Ask the guide and
advice of experts for the re quirement
quirement of additional protective devices at frequent use of high kV
and mAs.

WARNING
Please check if the voltage and current of the power supply observe the
specifications written on t he system labels, attache
ached
d t o backsid
kside
e of machine
body. The levels of voltage should move within ±10% of the standard values.

CAUTION
Please check up all the line conne ction
tions
s before the operatio
operation
n and x-ray
exposures of machine.

2) Collimator Check-up

A. Place the loaded


loaded cassette on the patient’s
patient’s backsi
backside.
de.
B. Arrange the SID (Focal Spot to Image Receptor Distance) using a scale located on the side of the
collimator.
C. Turn on the collimator lamp and laser pointer by pushing collimator/laser on switch.
D. Adjust the size of the x-ray beam in acco
accordance
rdance with the film size using the x-ray field ad
adjustment
justment
knobs. Turn off the collimator and laser pointer after 30 seconds on the timer.

18/47
Product Manual: POX-100BT

3) Preheat

(1) Overview

All portable x-ray units re quire preheat procedure to protect the x-ray tube from an instant surge of
current and to secure long and safe operation of the system. Preheat of x-ray
x -ray tube is required under
following situation.
A. Initial use or installation
installation..
B. Non-use of the equipment over 1 month at low temperature
C. Temp
Tempera
eratur
ture
e of x-ray
x-ray tube
tube is b
below 0℃ (32℉). (Cold tube)
elow

(2) Method to Preheat

Following preheat procedure is recommended.


A. Install the equipment
equipment at a safe
safe place fo
forr radiation.
B. In case
case outd
outdoo
oorr tem
tempe
pera
ratu
ture
re is be
below 0℃ (32
low (32℉), move the equipment indoor and warm it up.
C. In order to preheat filament of x-ray tube, exposure x-rays in 15 second time intervals.

First: 3 exposures at 50 kV @ 0.2 mAs


Second: 3 exposures at 80 kV @ 5.0 mAs
Third: 3 exposures at 110 kV @ 5.0 mAs

CAUTION
Beep Sound
Sound will occur during x-ray exposu res. Do not
n ot press exp
exposure buttons
at wait mode after x-ray exposure is made.

19/47
Product Manual: POX-100BT

9. Operation Mode

1) General Mode
A. Open the carrying
carrying case.
B. Connect Power supply cable, communication cable and hand switch as following.

Power Supply Cable

Communication Cable

[OP Console] [Console Side] [Body Side]


Hand Switch Connection

20/47
Product Manual: POX-100BT

※ CAUTION※
Please use designated electronic parts (Power plug and Hand switch) and be
careful not to apply other components.

※ CAUTION※

and on power switch represent ON and OFF,


OFF,
respectively

C. Turn on the power on switch and initia


initiall conditions a
are
re displayed as
as follows.
(1) Main Console

(2) Sub Console

21/47
Product Manual: POX-100BT

D. Select kV settings by kV up/down switch. The speed of increasing or decreasing values can be
faster by keep pressing the buttons.
E. Select sec & mAs values by sec & mAs
mAs up/down switch. The sec segment displays the real
value of sec without separate table.
st
F. After setting the radiography conditions, make the x-ray at preposition by pressing the 1
stage of hand switch.
G. Press the 2nd step of hand switch and make exposure as soon as the ready LED is lighted on.
The exposure LED is lighted on during exposures.
H. The wait LED
LED is lighted
lighted off fo
forr 0.5 seconds
seconds after the x-ray exposures.
exposures.
I. Radiograph
Radiography
y conditions
conditions (kV & mAs) can be operated by sub console as well as the light
on/off of collimator lamp.

2) APR Mode
A. Select APR Switch.
B. Select the size of patient by APR Size Select Switch.
C. Select the subject region
r egion by APR region select switch.
D. Select the radiography direction by APR direction select switch.

E. Select wanted values by density up/down switch.


F. APR Storage
1) Users can save 30 (Body Size 2 X Region 5 X Direction 2) data for APR.
2) If you press Store Switch after changing kV/mA/Sec, the changed condition is saved.

※ NOTICE※
APR data ca n not be saved when OVER LED is lighted on or Density is not
Zero. In this case, change radiography condition
condit ion (kV, mA, Sec) and density
and resave.

※ NOTICE※
Under APR mode, the last
l ast saved technique values are automatically selected
and display and saving of data is available only at x-ray exposures.

※ NOTICE※
As explained at previous
previous pages, the exposure switch should be operated by 2
stages. It, however
however , instant exposure is available when
when the filament
filament board is
heated enough.

22/47
Product Manual: POX-100BT

23/47
Product Manual: POX-100BT

PART II. SERVICE MANUAL

1. NOTICE
1) If the output calibration is neede
needed after periodical inspection or repair, please observe following rules
and methods.

CAUTION
The ioniz
ionized radiation could be dang
dan gerous to the ope
operator u nless
nless f ollowing
ollowing
safety measures are strictly observed.

2) This manual is des


designed to ensur
ensure correct us
use and operation o f POX-100BT
POX-100BT.. Please rea
read all the
lines thoroughly before using this equipment.

3) Incorrect use and op


operation exceedi
exceedingng described conditions in this manua
manual may caus
cause damage of
the mac
machine and shorten its life. Parti
Particular attention
attention must be
be paid to all the w
warnings,
arnings, cautions and
notes incorporated here in.
4) This equipment should be used only by the legally qualified persons and practitioners.
5) Regarding the AC cable a
and
nd sof
software,
tware, only
only the one developed and supplied by ACOMA'S should be
used.
6) No responsibility is ta
taken by ACOMA'S for any infringement of pa
patients or other right s of the third
parties which many results from the use of this manual.
7) Always keep the manual at hand for your reference.

24/47
Product Manual: POX-100BT

2. Regular Maintenance

1) Overview
Portable x-ray equipment should be inspected regularly according to following instruction. .

2) Periodical Inspection

1) Every 6 Month
A. Check all visual displays:
displays: display
display readout win
windows
dows (FND), LEDs a
and
nd other indi
indication.
cation.
B. Check all functions:
functions: Operation switches, collimator,
collimator, la
laser
ser pointer and
and programme.
programme.
C. Check all status
status of calibration
calibration acco
according
rding to sec
section
tion 4 o
off service manual.
manual.
D. Test the thermal switch of x-ray tube inside the equipment.
Remove the connection to tube thermal swit ch and check the display of error message. Test
normal operation of x-ray exposure as well.
E. Execute additional test required by laws and official regulation.

2) Every 12 Month

A. Check damages
damages on the outer
outer case: Ope
Operation
ration console including buttons.
buttons.
B. Check
Check cab
cable
le con
conne
nections to main body: hand switch cable, collimator cable and other
connections.
C. Open the top cover of equipment and check visible damage: missing or loosened ground
connection,
D. Open the system cover and check the v isible damage:
damage: missing or loose groun d connections, oil
leakage, damaged cables and etc.

WARNING
This equipment should be handled and operated by the legally qualified persons
and practitioners only.

WARNING

Wait at least 5 minutes after turn off in order to discharge capacito


capacitors.
rs.

25/47
Product Manual: POX-100BT

3. Block Diagram v B
e
R

f
o
R
O
T 1
A
M
I
L t
L e
e
O h
K C S

N M
A
A R
G

T
P IA
M D 9
A 0
L K 0
2
C 4
,
V
O

H
L r 1
B e 1 t
T b 0 s
B m 0 u
0 u S gu
0 - A
N T
1
- t B ,
E n 0 y
e 0 a
IL
B - d
m 1
u X ir
X c F
O o O
P D P

:
e e 3 te
tli iz A a
T S D

D B
y 4
-1
A d
e
n
ig
te
a 8
-0
E
D
3 s 9
0
-
3
e 0
& 0
D
H
5 8
5 2
2
N 2
O 3
L
1

C 5
P

OP
M
A
L

0S
y
RO B y
+ B

0A
d d
T E V e e
V .5 B
2 v k
o c
1m
8 T r
NT R
5 p
p
A
e
h
C

OO D N
CM
T
0V R B
1 y

1k
B
V T
T E
.
.5 v
1 HR
8
5
e
R

2
0
-
1 8
5
2
N 2 e
O 3 t
1

C 5
a
D
6
1 0
-
2 2
N 1
3
C 12 3 4 5 6 5
3
V
P DD5
XXXNN1
T R E GG+

S
T
N
E
T
N
O
C

1 A
0
F 0 H H H H
1
yr
A A A A .
4
1 2 4
2 2 4
3 2 4
4 2 O
TV TV TV TV E D C B A N

tte
B2 B2 B2 B2
1 1 1 1

a
1 2
1 2
1 2
1 2

1
2
B
-
B
4
B
2

C
P
3

3
2
D
4

6
2
1
N
R 3 21

C C
4 20
3
P
A
-
6
O N8
5
C2 r
e
K
B
3
C 2 5
A *
J g
r
1
1
E
N
O
P a
N
C
H
P O h
B C
tt
1
-
4
B
Y
2

C
P
a
3
5
P 3
2 D 0
-
4

O
6
B O B 1
N
C
8
5
2
3
5
U
S
-
D B 1 23 4 5
1 2 3 4 5 67 8 9 0 1 23 4 5
1 1 11 1 1

W
/
2 2 S
2 E
W L
S D
N
1 1 A
H

H 2
W
/
S
C 점
접 1 G
T
I 조 WNI
SL
W 보
N 1
R
A
S O C
D
N
A E
L
H A
M
C
A
T G
E U
L L
IN P

26/47
Product Manual: POX-100BT

4. Calibration procedures

▣ Preparation
Following instrument and tools are required.
a. Tools: Philips screwdriver No.1 & 2, Straight slot screwdriver
b. Testers: DVM, Digital Oscilloscope (Tektronics TDS 3032 or other device with higher calibers.

1) Power Supply
-The power switch should be at off condition.
-Connect power cables to power consent.
-Power on the switch.

2) Calibration method is composed of three procedures.


-Setting and adjusting tube kV.
-Setting and adjusting tube mA
-Adjusting filament preheat.

A. Setting and adjusting tube kV (Pre shipment adjustment)


-Turn off the power switch.
-Connect Oscilloscope to HT Controller inside machine.
-Chanel 1: connect the probe of oscilloscope to TP11 (IP F/B) and GND (TP16, 17 or 18).
-Chanel 2: connect the probe of oscilloscope to TP10 (EP F/B) and GND (TP16. 17 or 18).

Chanel 1: mA, 1V = 10mA


Chanel 2: kV, 1V = 25kV

-Turn on the power switch.


-Set the conditio
itions
ns a s 4 0k
0kV
V and
and 0. 1m
1mAs
As (2 0mA) and observe mA waveforms at Channel 1 of
0mA) and
Oscilloscope.
-If the value at channel 1 is not 2V (20mA), adjust the mA output with VR1 of HT board. And observe if
the waveform of Channel 2 is 40kV (1.625V).
-Set the conditions as 1 00kV
00kV and 0.2 mAs (4 0mA)
an and observe mA
0mA) and waveforms at Channel 1 of
mA wavef
Oscilloscope.
-If the value at channel 1 is not 4 V (40m A), adjust the mA output with VR1 of HT board. And observe if
the waveform of Channel 2 is 100kV (4V).

※ CAUTION ※
As the kV value of channel
channel 2 is automatically a
adjusted
djusted by standard
standard parameters,
parameters,
Additional calibration
calibration is not required.

27/47
Product Manual: POX-100BT

B. Setting and adjusting tube mA


-Turn off the power switch.
-PCB HT board is located under control panel.
-Al
-Alll DIP
DIP swi
switt che
ches
s o n PCB
PCB HT b oard
oard are at o
off f co
con
n dition
ditions.
s. (N
(Note
ote:: DIP switch
switch i s located i n this
this pl
pl ace
normally.
-Chanel 1: connect the probe of oscilloscope to TP11 (IP F/B) and GND (TP16, 17 or 18).
-Chanel 2: connect the probe of oscilloscope to TP10 (EP F/B) and GND (TP16. 17 or 18).

Chanel 1: mA, 1V = 10mA


Chanel 2: kV,
kV, 1V = 25kV

-Turn on the power switch.


-Set the conditions as 4 0kV
0kV a
and
nd 0. 2m
2mAs
As (1 0m
0mA)
A) and
and observe mA wave
wavefforms at Channel 1 of
Oscilloscope.
-If the value at channel 1 is not 4V (40mA), adjust the mA output under calibration mode.
-If you press “kV Down”, “APR S”, “APR L” switch for 2 seconds, it is enter into calibration mode.

Calibration mode is displayed as below. “C00” on the kV segment is kV calibration code and “20” on
the mA segment is mA calibration value.
Refer to content of each menu at table 1.

28/47
Product Manual: POX-100BT

The kV display segment of FND is displayed as CXX.

Calibration Code Calibration Data

Press mA up switch (Refer to the OP sheet) and it displays as follows.


C01 is calibration code of tube current (40mA at 40-50kV) and 40 is calibration data of tube current.
.(40mA from 40 to 50kV)

-Select mA with mAs up/down switch (Refer to the OP sheet)


-Press the ‘Store Key’ longer than 1 second after application (Refer to the OP sheet)
-Exposure x-rays after saving the data

-If the measu


measu red data is n
not
ot 4V(40
4V(40 mA
mA),
), reset
reset t he cali
cali bra
bratio
tion
n dat a. Set the mA val
value
ue s as thi s
method.
-Table 1 represents calibration code and data of each tube current vales at 5kW.

C. Filament preheat calibration

-If the preheat calibration data or VR1 are set in wrong way, over current or low current can occurat X-ray
tube.
-There are two methods about preheat calibration. Calibration data can be changed by tube conditions.
-VR1 can be applied to overall tube
tube current. Filam
Filament preheat calibration data can be applied to each
tube current.
-Turn on the power switch.
-Under normal mode, press [‘S’+‘L’+‘kV_Down’] Key in case of human and [‘Bird’+‘Dog’+‘kV_Down’] Key
in case of veterinary over 2 seconds.

29/47
Product Manual: POX-100BT

⑦ kV Up/Down 스위치를 이용하여 관전류 조정모드에서 필라멘트 예열 조정모드로 변경한다.

P00 is filament preheat calibration code (preheat calibration code and data of
of 2
20
0mA) and 28 iis
s
calibration data.
-Apply the data by mAs up/down switch. (Refer to the OP sheet)
-Press the ‘Store’ key longer than 1 second (Refer to the OP sheet)
-Reset the calibration data if any over-shoot or under-shoot
under-shoot is observed.
Press kV up switch. (Refer to the OP sheet) and it displays as below.

-P01 is filament preheat calibration code and 30


3 0 is the cali
calibration
bration data (preheat calibration code and
data of 40mA)
-Apply the data by mAs up/down switch. (Refer to the OP sheet)
-Press the ‘Store’ Key longer than 1 second after application (Refer to the OP sheet)

-Exposure x-rays after saving the data.


-Reset the calibration data if any over-shoot or under-shoot
under-shoot is observed.
-Table 2 represents calibration code and data of each tube current vales at 3kW.
-Escape from calibration mode by press [‘M’=‘Cat’] Key one time.

[Human] [Veterinary]

※ Notice※
Calibration data can be changed by tube conditions.

※ Notice※
If you press [‘S’=‘Cat’] Key consecutively, it repeats P -> C -> L _ Mode.
[ P: Preheat, C: mA for 5kW, L: mA for 3kW ]

30/47
Product Manual: POX-100BT Posko m Co., Ltd.

5. Calibration data

[Table
[Table 1. mA calibration data for 5kW]
AC Power Code Data Tube kV Tube mA mAs 220V C00
220V C01 40 40kV ~ 50kV 40mA 0.2 ~ 0.3

220V C02 100 40kV ~ 50kV 100mA 0.4 ~ 10

220V C03 80 40kV ~ 50kV 80mA 12 ~ 32

220V C04 60 40kV ~ 50kV 60mA 36 ~ 200

220V C05 20 51kV ~ 60kV 20mA 0.1

220V C06 40 51kV ~ 60kV 40mA 0.2 ~ 0.3

220V C07 80 51kV ~ 60kV 80mA 0.4 ~ 8.0

220V C08 50 51kV ~ 60kV 50mA 9.0 ~ 100

220V C09 40 51kV ~ 60kV 40mA 125 ~ 200

220V C10 20 61kV ~ 70kV 20mA 0.1


220V C11 40 61kV ~ 70kV 40m 0.2 ~ 0.3

220V C12 50 61kV ~ 70kV 50mA 0.4 ~ 8.0

220V C13 40 61kV ~ 70kV 40mA 9.0 ~ 80

220V C14 30 61kV ~ 70kV 30mA 100 ~ 150

220V C15 20 71kV ~ 80kV 20mA 0.1

220V C16 40 71kV ~ 80kV 40mA 0.2 ~ 0.3

220V C17 50 71kV ~ 80kV 50mA 0.4 ~ 8.0

220V C18 40 71kV ~ 80kV 40mA 9.0 ~ 21

220V C19 30 71kV ~ 80kV 30mA 25 ~ 64

220V C20 20 81kV ~ 100kV 20mA 0.1

220V C21 40 81kV ~ 100kV 40mA 0.2 ~ 8.0


220V C22 30 81kV ~ 100kV 30mA 9.0 ~ 30

220V C23 20 81kV ~ 100kV 20mA 32 ~ 80

220V C24 20 101kV ~ 110kV 20mA 0.1

220V C25 40 101kV ~ 110k


110kV
V 40mA 0.2 ~ 5.0

220V C26 30 101kV ~ 110kV 30mA 6.4 ~ 12


220V C27 20 101kV ~ 110
110kV
kV 20mA 15 ~ 64

[Table 2. mA Calibration Data for 2.3kW]

AC Power Code Data Tube kV Tube mA mAs

31/47
Product Manual: POX-100BT

220V L00 20 40kV ~ 50kV 20mA 0.1

220V L01 40 40kV ~ 50kV 40mA 0.2 ~ 0.3

220V L02 60 40kV ~ 50kV 60mA 0.4 ~ 200

220V L03 20 40kV ~ 50kV 20mA 0.1

220V L04 40 40kV ~ 50kV 40mA 0.2 ~ 0.3

220V L05 50 51kV ~ 60kV 50mA 0.4 ~ 100

220V L06 40 51kV ~ 60kV 40mA 125 ~ 200

220V L07 20 61kV ~ 70kV 20mA 0.1

220V L08 40 61kV ~ 70kV 40mA 0.2 ~ 80

220V L09 30 61kV ~ 70kV 30mA 100 ~ 150

220V L10 20 71kV ~ 80kV 20mA 0.1

220V L1 1 30 71kV ~ 80kV 30mA 0.2 ~ 64

220V L12 20 81kV ~ 100kV 20mA 0.1

220V L13 30 81kV ~ 100kV 30mA 0.2 ~ 30

220V L14 20 81kV ~ 100kV 20mA 32 ~ 80

220V L15 20 101kV ~ 110kV 20mA 0.1 ~ 64

[T
[Table
able 3. Filament preheat calibration code and data]
AC Power Code Data Tub
Tube
e Current

220V P00 28 20mA

220V P01 30 30mA

220V P02 31 40mA

220V P03 31 50mA

220V P04 33 60mA

220V P05 35 80mA

220V P06 38 100mA

※ No tice
ice ※
Do not use the machine in case of frequent errors. Please make contact to
ACOMA"S service team or authorized dealers
dealers for trouble shootings.

32/47
Product Manual: POX-100BT

6. Radiation dose data

[Table
[Table 4. Radiation dose data]

29.204.3 T able: Me
Measured LEAKAGE R ADIATION in the LOADING STATE
ADIATION P

Conditions of LOADING Allowed limit of


of LEAKAGE

Points of Mode of CURRENT, (mA) or X-RAY TUBE LEAKAGE RADIATION

measurement operation CURRENT TIME VOLTAGE, (kV) RADIATION (mGy/h)


s PRODUCT, (mAs) (mGy/h)

Front - 20 mAs 110 1,0 0,24

Rear - 20 mAs 110 1,0 0,05

Left - 20 mAs 110 1,0 0,75

Right - 20 mAs 110 1,0 0,74

Top - 20 mAs 110 1,0 0,55

Bottom - 20 mAs 110 1,0 0,71

Front-left - 20 mAs 110 1,0 0,54

Front-right - 20 mAs 110 1,0 0,19

Rear-left - 20 mAs 110 1,0 0,03

Rear-right - 20 mAs 110 1,0 0,02

Supplementary information:

- Model: POX-100BT, Serial No.: PTH100-0809-001


- Test Instrument: RADCAL, 10x5-180/1015

33/47
Product Manual: POX-100BT

7. Error code and resolution

1) Error at stand-by

Without hand
hand switch operation, if hardware is over th an EP=1
EP=10kV,
0kV, IP=5mA,
IP=5mA, FIL=0.1V F/B Check, it is in
error.
Error Code Contents Measures
Err. 10 FILAMENT
FILAMENT FEEDBACK OVER Calibration or adjust VR.
Err. 11 kV FEEDBACK OVER Calibration or adjust VR.
Err. 12 mA FEEDBACK OVER Calibration or adjust VR.

2) Error at ready

With ready switch on, check the filam ent conditions and if hardware is over than FIL=0.35V F/B Check, it
is in error.
Error Code Contents Measures
Err. 14 FILAMENT
FILAMENT FEEDBACK LOW ERROR Calibration or adjust VR.

3) Error at Exposures
EP, lP Low Error: Lower than minimum Value (30kV).
EP,IP Over Error: Higher than maximum Value (120kV).
EP,IP State Low Error: Lower than output kV (-10kV)
EP,IP State Over Error: Higher than output kV (+10kV)

Error Code Contents Measures


Err. 15 kV FEEDBACK LOW DETECT ERROR AT EXPOSURE Calibration or adjust VR.
Err. 16 mA FEEDBACK LOW DETECT ERROR AT EXPOS
EXPOSURE
URE Calibration or adjust VR.
Err. 17 kV FEEDBACK OVER DETECT ERROR AT EXPOSURE Calibration or adjust VR.
Err. 18 mA FEEDBACK OVER DETECT ERROR AT EXPOSURE Calibration or adjust VR.
-1
-10kV
0kV k
kV
V FEEDBACK LOW DETECT ERROR AT Calibration or adjust VR.
Err. 19
EXPOSURE

+10kV
+10kV kV FEEDBACK
FEEDBACK LOW DETECT ERROR AT Calibration or adjust VR.
Err. 20
EXPOSURE

-1
-10kV
0kV mA FEEDBACK LOW DETECT ERROR AT Calibration or adjust VR.
Err. 22
EXPOSURE

Err. 23 +10kV
+10kV mA FEEDBACK LOW DETECT ERROR AT Calibration or adjust VR.

34/47
Product Manual: POX-100BT

EXPOSURE

4) Inter_Lock Error
Error Code Contents Measures
Err. 21 Tank Thermal S/W Check

35/47
PRODUCT MANUAL: POX-100BT

8. Schematics
1) Battery Charger

PRODUCT MANUAL: POX-100BT

2) DRV Board
2

PRODUCT MANUAL: POX-100BT

3) HT Controller

PRODUCT MANUAL: POX-100BT

4) HT Controller
4

PRODUCT MANUAL: POX-100BT

5) HT Controller

PRODUCT MANUAL: POX-100BT

6) HT Controller
6

PRODUCT MANUAL: POX-100BT

7) Console Board

PRODUCT MANUAL: POX-100BT

8) Console Board