ELECTRICAL DEPARTMENT
MODULE 2 NOTES
AC BRIDGES
AC bridge are similar to D.C. bridge in topology(way of connecting).It consists of four arm
AB,BC,CD and DA .Generally the impedance to be measured is connected between ‘A’ and ‘B’.
A detector is connected between ‘B’ and ’D’. The detector is used as null deflection instrument.
Some of the arms are variable element. By varying these elements, the potential values at ‘B’ and
‘D’ can be made equal. This is called balancing of the bridge.
. . . .
∴I 3 Z3 = I 4 Z 4 (2.3)
. .
I1 Z2
From Eqn. (2.2), we have ∴ = (2.4)
. .
I2 Z1
. .
I3 Z4
From Eqn. (2.3), we have ∴ = (2.5)
. .
I4 Z3
From equation -2.1, it can be seen that, equation -2.4 and equation-2.5 are equal.
. .
Z2 Z4
∴ =
. .
Z1 Z3
. . . .
∴Z1 Z 4 = Z 2 Z 3
∴ Z1 ∠θ1 Z 4 ∠θ 4 = Z 2 ∠θ 2 Z 3 ∠θ 3
⇒ Z1 Z 4 ∠θ1 + θ 4 = Z 2 Z 3 ∠θ 2 + θ 3
Z1 Z 4 = Z 2 Z 3
θ1 + θ 4 = θ 2 + θ 3
∗ For balance condition, magnitude on either side must be equal.
Summary
For balance condition,
. . . .
• I1 = I 3 , I 2 = I 4
• Z1 Z 4 = Z 2 Z 3
• θ1 + θ 4 = θ 2 + θ 3
. . . .
• E1 = E 2 & E3 = E4
• Vibration galvanometer
• Tuned amplifier
If the bridge is unbalanced the output of tuned amplifier is high. If the bridge is balanced,
output of amplifier is zero.
2.3 Measurements of inductance
The choke for which R1 and L1 have to measure connected between the points ‘A’ and
‘B’. In this method the unknown inductance is measured by comparing it with the standard
inductance.
( R1 + jXL1 ) R4 = ( R2 + jXL2 ) R3
( R1 + jwL1 ) R4 = ( R2 + jwL2 ) R3
R1R4 = R2 R3
R R
∴ R1 = 2 3 (2.6)
R4
wL1R4 = wL2 R3
L R
L1 = 2 3 (2.7)
R4
WL1 WL2 R3 R4
Q-factor of choke, Q = =
R1 R4 R2 R3
WL2
Q= (2.8)
R2
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
Advantages
Disadvantages
1
R4 ×
1 jwC 4
Z 4 = R4 || =
jwC 4 1
R4 +
jwC4
R4 R4
Z4 = = (2.10)
jwR4 C 4 + 1 1 + jwR4C 4
R4
( R1 + jwL1 ) × = R2 R3
1 + jwR4 C 4
( R1 + jwL1 ) R4 = R2 R3 (1 + jwR4C 4 )
R1R4 = R2 R3
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
R R
⇒ R1 = 2 3 (2.11)
R4
wL1R4 = wC 4 R4 R2 R3
L1 = C 4 R2 R3 (2.12)
Q-factor of choke,
WL1 R
Q= = w × C 4 R2 R3 × 4
R1 R2 R3
Q = wC4 R4 (2.13)
Advantages
Disadvantages
For measuring chokes with higher value of Q-factor, the value of C4 and R4 should be
higher. Higher values of standard resistance are very expensive. Therefore this bridge cannot be
used for higher value of Q-factor measurements.
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
.
E1 = I1R1 + jI1 X 1
. . .
E = E1 + E 3
. . I4
E 4 = I 4 R4 +
jwC 4
.
E 3 = I 3 R3
1 1 + jwR4C4
Z 4 = R4 + =
jwC4 jwC4
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
1 + jwR4C 4
( R1 + jwL1 )( ) = R2 R3
jwC 4
R1 − w 2 L1C 4 R4 = 0
R1 = w 2 L1C4 R4 (2.14)
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
C 4 R4 R1 + L1 = C 4 R2 R3
L1 = C4 R2 R3 − C4 R4 R1 (2.15)
Substituting the value of R1 fro eqn. 2.14 into eqn. 2.15, we have,
L1 = C 4 R2 R3 − C 4 R4 × w 2 L1C 4 R4
L1 = C4 R2 R3 − w 2 L1C4 2 R4 2
L1 (1 + w 2 L1C 4 2 R4 2 ) = C 4 R2 R3
C4 R2 R3
L1 = (2.16)
1 + w2 L1C4 2 R4 2
w 2C4 2 R2 R3 R4
R1 = (2.17)
1 + w 2C4 2 R4 2
wL1 w × C4 R2 R3 1 + w2C4 2 R4 2
Q= = ×
R1 1 + w2C4 2 R4 2 w2C4 2 R4 R2 R3
1
Q= (2.18)
wC4 R4
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
Advantages
Disadvantages
E1 = I1R1 + jI1 X1
I4 leads E4 by 900
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
. . .
E = E1 + E 3
. I2
E 2 = I 2 R2 +
jwC 2
. . . .
Balance condition, Z 1 Z 4 = Z 2 Z 3
1 jwC 2 R2 + 1
Z 2 = R2 + =
jwC2 jwC 2
1 (1 + jwR2C 2 ) × R3
∴ ( R1 + jwL1 ) × =
jwC 4 jwC 2
R1C2 = R3C4
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
RC
R1 = 3 4
C2
L1 = R2 R3C4
Q = wR2C2
Advantages
Expression for R1 and L1 are simple.
R1 and L1 are independent of Frequency.
Disadvantages
The Circuits used two capacitors.
Variable capacitor is costly.
Q-factor range is restricted.
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
.
E1 = I1 ( R1 + r1 ) + jI1 X 1
E3 = EC
. .
E 4 = I C r + EC
I 2 = I 4 + IC
− − −
E2+ E4 = E
− − −
E1 + E 3 = E
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
E1 = I1R11 + jI1 X 1
Step-2 I1 = I 3 , E3 is in phase with I 3 , From the circuit ,
E3 = EC , I C leads EC by 900
Step-3 E4 = I C r + EC
− − −
Step-4 Draw I 4 in phase with E4 , By KCL , I 2 = I 4 + I C
Step-5 Draw E2 in phase with I2
− − − − − −
Step-6 By KVL , E1 + E3 = E or E2 + E4 = E
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
Fig 2.12 Equivalent delta to star conversion for the loop MON
R4 × r jwCR 4 r
Z7 = =
1 1 + jwC ( R4 + r )
R4 + r +
jwc
1
R4 ×
jwC R4
Z6 = =
1 1 + jwC ( R4 + r )
R4 + r +
jwc
R4 jwCR4 r
( R11 + jwL1 ) × = R3 ( R2 + )
1 + jwC ( R4 + r ) 1 + jwC ( R4 + r )
R11R4 = R2 R3
( R1 + r1 ) R4 = R2 R3
R R
R1 = 2 3 − r1
R4
Comparing the imaginary term,
R
L1 = R3C 2 (r + R4 ) + r
R4
Advantages
Disadvantages
From the vector diagram, it can be seen that the angle between voltage and current is slightly less
than 900. The angle ‘ δ ’ is called loss angle.
D = wCR
1
D=
Q
sin δ δ
D = tan δ = ≅ For small value of ‘ δ ’ in radians
cos δ 1
D ≅ δ ≅ Loss Angle (‘ δ ’ must be in radian)
2.4.2 Desauty’s Bridge
C1= Unknown capacitance
At balance condition,
1 1
× R4 = × R3
jwC1 jwC 2
R4 R3
=
C1 C 2
R C
⇒ C1 = 4 2
R3
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
R11 = ( R1 + r1 )
R12 = ( R2 + r2 )
1 1
At balance condition, ( R11 + ) R4 = R3 ( R12 + )
jwC1 jwC 2
R4 R3
R11R4 + = R3 R12 + )
jwC1 jwC 2
R R1
R11 = 3 2
R4
( R2 + r2 ) R3
R + r1 =
R4
Comparing imaginary term,
R4 R
= 3
wC1 wC 2
R C
C1 = 4 2
R3
Dissipation factor D=wC1r1
Advantages
r1 and c1 are independent of frequency.
They are independent of each other.
Source need not be pure sine wave.
E1 = I1r1 − jI1 X 4
1 jwC1r1 + 1
Z1 = r1 + =
jwC1 jwC1
1
R4 ×
jwC 4 R4
Z4 = =
R4 +
1 1 + jwC 4 R4
jwC 4
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
. . . .
At balance condition, Z 1 Z 4 = Z 2 Z 3
1 + jwC1r1 R4 R3
× =
jwC1 1 + jwC 4 R4 jwC 2
R C
∴ C1 = 4 2
R3
C R
r1 = 4 3
C2
R C C R
D = wC1r1 = w × 4 2 × 4 3
R3 C2
∴ D = wC4 R4
Advantages
Disadvantages
Wein’s bridge is popularly used for measurements of frequency of frequency. In this bridge, the
value of all parameters are known. The source whose frequency has to measure is connected as
shown in the figure.
1 jwC1r1 + 1
Z1 = r1 + =
jwC1 jwC1
R2
Z2 =
1 + jwC 2 R2
. . . .
At balance condition, Z 1 Z 4 = Z 2 Z 3
jwC1r1 + 1 R2
× R4 = × R3
jwC1 1 + jwC 2 R2
[1 + jwC R 2
2 2 + jwC1r1 − w C1C 2 r1R2 = jwC1]R2 R3
R4
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
1 − w 2C1C 2 r1R2 = 0
w 2C1C 2 r1R2 = 1
1
w2 =
C1C2 r1R2
1 1
w= , f =
C1C 2 r1 R2 2Π C1C 2 r1 R2
NOTE
The above bridge can be used for measurements of capacitance. In such case, r1 and C1 are
unknown and frequency is known. By equating real terms, we will get R1 and C1. Similarly by
equating imaginary term, we will get another equation in terms of r1 and C1. It is only used for
measurements of Audio frequency.
A.F=20 HZ to 20 KHZ
R.F=>> 20 KHZ
Comparing imaginary term,
R R
wC 2 R2 + wC1r1 = wC1 2 3
R4
CR R
C2 R2 + C1r1 = 1 2 3 …………………………………..(2.19)
R4
1
C1 =
2
w C2 r1R2
Substituting in eqn. (2.19), we have
r1 R R
C2 R2 + = 2 3 C1
w2C2 r1R2 R4
R4
Multiplying in both sides, we have
R2 R3
R4 1 R
C2 R2 × + × 4 = C1
R2 R3 w2C2 R2 R2 R3
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
C R R4
C1 = 2 4 +
R3 w2C2 R22 R3
w 2C1r1C 2 R2 = 1
1 1
r1 = =
2
w C 2 R2C1 C R R4
w 2C 2 R2 2 4 +
R3 w 2C 2 R22 R3
1
=
w2C 2 R R
2 2 4 + R4
R3 R2 R3
1
∴ r1 =
R3 2 2 1
w C2 R2 +
R4 R2
R 1
∴ r1 = 3
R4 ( w 2C 2 R + 1 )
2 2
R2
(1) The high voltage supply is obtained from a transformer usually at 50 HZ.
Wagner earthing consists of ‘R’ and ‘C’ in series. The stray capacitance at node ‘B’ and ‘D’ are
CB, CD respectively. These Stray capacitances produced error in the measurements of ‘L’ and
‘C’. These error will predominant at high frequency. The error due to this capacitance can be
eliminated using wagner earthing arm.
Close the change over switch to the position (1) and obtained balanced. Now change the
switch to position (2) and obtained balance. This process has to repeat until balance is achieved
in both the position. In this condition the potential difference across each capacitor is zero.
Current drawn by this is zero. Therefore they do not have any effect on the measurements.
Precaution
The load inductance is eliminated by twisting the connecting the connecting lead.
A∈0∈ r
In the case of capacitive bridge, the connecting lead are kept apart.(Q C = )
d
In the case of inductive bridge, the various arm are magnetically screen.
In the case of capacitive bridge, the various arm are electro statically screen to reduced
the stray capacitance between various arm.
To avoid the problem of spike, an inter bridge transformer is used in between the source
and bridge.
The stray capacitance between the ends of detector to the ground, cause difficulty in
balancing as well as error in measurements. To avoid this problem, we use wagner
earthing device.
This is a sophisticated instrument. This works on the principle of PMMC meter. The only
difference is the type of suspension is used for this meter. Lamp and glass scale method is used
to obtain the deflection. A small mirror is attached to the moving system. Phosphorous bronze
wire is used for suspension.
When the D.C. voltage is applied to the terminals of moving coil, current flows through it. When
a current carrying coil kept in the magnetic field, produced by permanent magnet, it experiences
a force. The coil deflects and mirror deflects. The light spot on the glass scale also move. This
deflection is proportional to the current through the coil.
Q
i= , Q = it = ∫ idt
t
θ ∝ Q , deflection ∝ Charge
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
D.C. voltage is applied to the electromagnet through a variable resistance R1 and a reversing
switch. The voltage applied to the toroid can be reversed by changing the switch from position 2
to position ‘1’. Let the switch be in position ‘2’ initially. A constant current flows through the
toroid and a constant flux is established in the core of the magnet.
A search coil of few turns is provided on the toroid. The B.G. is connected to the search
coil through a current limiting resistance. When it is required to measure the flux, the switch is
changed from position ‘2’ to position ‘1’. Hence the flux reduced to zero and it starts increasing
in the reverse direction. The flux goes from + φ to - φ , in time ‘t’ second. An emf is induced in
the search coil, science the flux changes with time. This emf circulates a current through R2 and
B.G. The meter deflects. The switch is normally closed. It is opened when it is required to take
the reading.
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
The curve drawn with the current on the X-axis and the flux on the Y-axis, is called
magnetization characteristics. The shape of B-H curve is similar to shape of magnetization
characteristics. The residual magnetism present in the specimen can be removed as follows.
Close the switch ‘S2’ to protect the galvanometer, from high current. Change the switch
S1 from position ‘1’ to ‘2’ and vice versa for several times.
To start with the resistance ‘R1’ is kept at maximum resistance position. For a particular value of
current, the deflection of B.G. is noted. This process is repeated for various value of current. For
φ
each deflection flux can be calculated.( B = )
A
Magnetic field intensity value for various current can be calculated.().The B-H curve can be
plotted by using the value of ‘B’ and ‘H’.
In the case of no load test the reading of wattmeter is approximately equal to iron loss. Iron loss
depends on the emf induced in the winding. Science emf is directly proportional to flux. The
voltage applied to the pressure coil is V. The corresponding of wattmeter is ‘W’. The iron loss
WE
corresponding E is E= . The reading of the wattmeter includes the losses in the pressure
V
coil and copper loss of the winding S1. These loses have to be subtracted to get the actual iron
loss.
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
2.9 Galvanometers
D-Arsonval Galvanometer
Vibration Galvanometer
Ballistic C
Lamp and glass scale method is used to obtain the deflection. A small mirror is attached
to the moving system. Phosphors bronze is used for suspension.
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
When D.C. voltage is applied to the terminal of moving coil, current flows through it.
When current carrying coil is kept in the magnetic field produced by P.M. , it experiences a
force. The light spot on the glass scale also move. This deflection is proportional to the current
through the coil. This instrument can be used only with D.C. like PMMC meter.
The construction of this galvanometer is similar to the PMMC instrument except for the moving
system. The moving coil is suspended using two ivory bridge pieces. The tension of the system
can be varied by rotating the screw provided at the top suspension. The natural frequency can be
varied by varying the tension wire of the screw or varying the distance between ivory bridge
piece.
When A.C. current is passed through coil an alternating torque or vibration is produced. This
vibration is maximum if the natural frequency of moving system coincide with supply frequency.
Vibration is maximum, science resonance takes place. When the coil is vibrating , the mirror
oscillates and the dot moves back and front. This appears as a line on the glass scale. Vibration
galvanometer is used for null deflection of a dot appears on the scale. If the bridge is unbalanced,
a line appears on the scale
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
Example 2.2-In a low- Voltage Schering bridge designed for the measurement of
permittivity, the branch ‘ab’ consists of two electrodes between which the specimen under
test may be inserted, arm ‘bc’ is a non-reactive resistor R3 in parallel with a standard
capacitor C3, arm CD is a non-reactive resistor R4 in parallel with a standard capacitor C4,
arm ‘da’ is a standard air capacitor of capacitance C2. Without the specimen between the
electrode, balance is obtained with following values , C3=C4=120 pF, C2=150 pF,
R3=R4=5000Ω.With the specimen inserted, these values become C3=200 pF,C4=1000
pF,C2=900 pF and R3=R4=5000Ω. In such test w=5000 rad/sec. Find the relative
permittivity of the specimen?
R
C1 = C2 ( 4 )
R3
Let capacitance value C0, when without specimen dielectric.
Let the capacitance value CS when with the specimen dielectric.
R 5000
C0 = C 2 ( 4 ) = 150 × = 150 pF
R3 5000
R 5000
C S = C 2 ( 4 ) = 900 × = 900 pF
R3 5000
C S 900
εr = = =6
C0 150
Example 2.3- A specimen of iron stamping weighting 10 kg and having a area of 16.8 cm2 is
tested by an episten square. Each of the two winding S1 and S2 have 515 turns. A.C. voltage
of 50 HZ frequency is given to the primary. The current in the primary is 0.35 A. A
voltmeter connected to S2 indicates 250 V. Resistance of S1 and S2 each equal to 40 Ω.
Resistance of pressure coil is 80 kΩ. Calculate maximum flux density in the specimen and
iron loss/kg if the wattmeter indicates 80 watt?
CLASS NOTES ON ELECTRICAL & ELECTRONICS MEASUREMENTS
Soln- E = 4.44 fφ m N
E
Bm = = 1.3wb / m 2
4.44 fAN
RS E2
Iron loss= W (1 + )−
RP ( RS + RP )
40 250 2
= 80(1 + )− = 79.26 watt
80 × 10 3 (40 + 80 × 103 )