ECL760
ECL760
• TEST PRINCIPLE
• INSTALLATION
• SOFTWARE
• SERVICE
STRUCTURE
WORK STATION
INTERFACE AND EXTERNAL TUBING
Wash
sensor
Waste
sensor
Cleaning System:
• Measuring methods :
To grab cuvette from make-up place, incubation place and test place to
specified place to add up sample or reagent or make test or discard cuvette into
waste box after test; it also has even agitation performance with reaction liquid
and exploration for cuvette.
Specification
Test:
Coagulation reaction
Fibrinogen Stabilized Fibrin
E D D E D D E D D E D D E D D E D D
D D E D D E D D E D D E D D E D D E
100%
50%
dH
bH
0%
Clotting time Time
Reaction curve : Prothrombin Time with derived Fibrinogen
• Chromogenic Method:
Light transmission, Light resource is
purple light (405nm LED). It needs
405nm interference filter.
• Immunologic Method:
Light transmission, Light resource is
yellow light (575nm LED). It needs
575nm interference filter.
Hydraulics Wash Liquid path
Wash the sample probe’s
inner wall and outer wall.
at a constant pressure of
280mmHg±30mmHg
Code
Main control
Main board
board
Temp-control
board
Thermal Interface
printer board
Hydraulic assy.
Main control board
Multi-Channel Photo-
electric signal receiving Signal pre-amplifier board
and amplifying
Syringe、Pump、Vacuum pump、
Liquid path control Syringe valve、Cleaning valve
Read photoelectric signal transmitted from optical system component, and transmit
well-handled signal to main board and calculate test result;
Transfer drive voltage for injector, and control liquid filling and drainage from injector;
Create control signal and drive voltage for liquid path or gas path;
Inspect and test signal of sample rack, shielding cover or washing bottle and waste bottle.
Motor driving Module
Closed-loop control:
XYZ Motor Control Motor and photo-
coupler
Sample arm
interface board
Liquid level
detecting board Needle capacitor value
Liquid level
change convert to
detecting Electrical signal
Motor driving
Closed-loop control:
Module XYZ Motor Control Motor and photo-
coupler
Catcher arm big
interface board
Catcher arm small Cuvette detection Reflecting Photo-
interface board signal coupler
Mixer motor
switcher
Motor driving module mainly includes: 1. motor driving board 2. sampling arm
interface board 3. Liquid level detection board 4. Catcher arm big interface board 5.
Catcher arm small interface board
Function:Execute the command from the main control board, and make the sampling arm
and catcher arm move, Detect the liquid level by the sample needle, detect the cuvette by the
catcher, control the mixer motor.
Temperature control module
Reagent position (Closed-loop) Heating
Temp-control (Cold wire, thermistor
store)
2,Power output
Switch
power Power supply transition board
supply
Temp
Main board Main control board Motor driving board Fans
control
board
Touch Main
LCD LCD
screen Printer control
Interface
driving board board
board
Wires Connections Schematic
Main control board
Injector valve
Vacuum
pump
sample rack detection board
Sampling Y arm,
Catcher arm sensor
big conversion board
Sampling Z arm,
Catcher arm sensor
small conversion board
Temperature Control
Board
Mix motor
5 Temperature Control Board
Reagent agitation
motor
Incubation Position
Temperature sensor
Reagent Position
Temperature sensor
Temperature Control
peltier Board Heating membrane1
fan
Heating membrane2
Three-way signal
Four-way signal pre-amplify board Pre-amplify board
1, Instrument shielding
cover is plastic part, so it
can't bear outer force.
• Daily Maintenance
– The used cuvettes should be discarded.
– Disposal of waste fluid
– Wipe away the residual liquid on the reagent holder.
– Clean the sample/reagent needle.
• Weekly Maintenance
– Pipe cleaning
– Equipment cleaning
• Monthly Maintenance
– LED calibration
– Clean the powder or debris on the test hole and incubation hole.
• Quarterly Maintenance (three to six months)
– To maintain the machinery rail, clean the stain/dust, and then spray/coat with
special grease.
– Clean the interior of lotion bottle, and clean the filter on the outlet pipe of lotion
bottle for one time.
– Calibrate the mechanical position