james instruments_windsor_hp_probe_system_data_manual
james instruments_windsor_hp_probe_system_data_manual
Table of Contents
Introduction....................................................................................... 1
Applications (Intended Use) .................................................... 2
Features .................................................................................. 2
Instrument Contents List .................................................................. 3
Contents List ............................................................................ 4
Probes ..................................................................................... 7
Mohs’ Test Kit .......................................................................... 8
Scratch Testing ................................................................... 8
Power Driver Overview .................................................................... 9
Instrument Functions ............................................................. 10
Loading the Driver ................................................................. 10
Selecting Power Level ........................................................... 16
Driving Probes ....................................................................... 17
Flat Surfaces ..................................................................... 17
Vertical Surfaces ............................................................... 18
Curved Surfaces ............................................................... 18
Electronic Measuring Unit Overview .............................................. 19
Instrument Functions ............................................................. 20
Determining Results .............................................................. 20
Recording Results ................................................................. 22
Probe Removal ...................................................................... 23
Electronic Measuring Unit .............................................................. 25
Control Panel Buttons ............................................................ 25
Title Screen ........................................................................... 26
Enter Setup ............................................................................ 27
Show Settings ........................................................................ 28
Calibrate ................................................................................ 29
Set Clock ............................................................................... 31
Set Settings ........................................................................... 32
Download Tests ..................................................................... 34
Show Tests ............................................................................ 35
Run Test ................................................................................ 36
Measure ................................................................................. 38
Standard Power Table ................................................................... 39
Low Power Table............................................................................ 42
Software Installation ....................................................................... 45
Where to Probe .............................................................................. 52
Troubleshooting ............................................................................. 54
Contents
Introduction
1 www.ndtjames.com
Introduction
Features
• New electronic measuring system enhances accuracy
and efficiency
• Non-destructive and can be used with equal
effectiveness on fresh and mature concrete
• Measurement to 17,000 psi (110 MPa)
• Memory for data storage and uploading to PC
• Safe and reliable
• Does not allow for accidental discharge (even when
dropped) and does not have recoil
• Approved by municipalities in the United States, Asia,
and Europe, this system conforms to ASTM C-803, BS
1881 Part 207, ANSI A.10-3 ACI 347-78 and other
testing standards.
www.ndtjames.com 2
Instrument Contents List
3 www.ndtjames.com
Instrument Contents List
Contents List
Item # Definition
1 Barrel Brush/Rod – Used to clean contaminants out of
the barrel after firing and to position test probes to Low
Power position.
2 Power Driver – Used to drive the steel probe into the
surface of the concrete.
3 Withdrawal Kit – Used to withdraw the probe from the
concrete surface.
4 (3) Single Probe Measuring Plates – Used to place
over the probe and against the concrete to take an
electronic depth reading.
5 Electronic Measuring Unit – Used to electronically
measure exposed probe length, correlate this reading
with an appropriate compressive strength value and
store the values for later use.
6 (2) Single Probe Firing Templates – Used to place
the power driver on to fire probe into concrete.
7 (2) Driver Heads – Used to screw onto the threaded
end of the probe prior to firing.
8 (3) Single Probe Measuring Caps – Replaces driver
head on fired probes and used to take measurements
of depth penetration.
9 ®
Checklist – Checklist of items completed by a James
Instruments Inc. technician prior to packaging the unit.
10 Calibration Certificate – Certificate to confirm that the
instrument has been calibrated to meet or exceed
published specifications.
11 (2) AA Batteries – Used to power the Electronic
Measuring Unit.
12 Case – Used to carry Windsor® Probe equipment and
accessories.
www.ndtjames.com 4
Instrument Contents List
5 www.ndtjames.com
Instrument Contents List
Item # Definition
13 USB Cord – Serial cord used to connect the Windsor®
Probe to a PC to upload data.
14 Mohs’ Test Kit – Used to determine the Mohs’ value of
any aggregate (Scale of Hardness).
15 Wrench – Used with the withdrawal kit to extract the
steel probe from the concrete after firing.
16 CD Disk – Contains system upload program.
17 Whisk Broom – Used to clean the surface after driving
probes.
18 Three-Probe Template – Used to place or hang over
the first probe to determine the location of the
placement of additional probes. This ensures that any
subsequent probes are far enough apart from the initial
probe; so that if a weakness is created due to the initial
probe it won’t affect the reading of the subsequent fired
probes.
www.ndtjames.com 6
Instrument Contents List
Probes
Used with a precise charge, probes are fired into a concrete surface
to measure the amount of penetration. This penetration amount is
then used to determine the compressive strength of the concrete.
Two probe styles are available: one for lightweight, low density
concrete with air filled aggregate and the other probe for more
standard mix designs. Also, two standard power settings facilitate
testing fresh concrete as well as mature mixes.
Two power settings are available: low and standard power. The low
power is used where concrete strength is less than 3,000 psi (19.4
MPa). The newly designed Silver probes can be used for high
performance concrete with strength up to 17,000 psi (110 MPa).
The probes are made of a high strength steel alloy, heat treated
and annealed to achieve a hardness of Rockwell C 48. Special
machining of each probe eliminates stress concentrations.
The Gold probe has a 56% greater cross-sectional area than the
Silver probe. The Gold probe is recommended for lightweight
concrete - less than 125 lbs/cu. ft. (2,003 Kg/M3) in density. The
Silver probe is used with concrete having a density greater than
125 lbs/cu. ft. (2,003 Kg/M3).
Probes are consumable items and can be used one time only. For
additional probes, please contact NDT James® Instruments.
7 www.ndtjames.com
Instrument Contents List
www.ndtjames.com 8
Power Driver Overview
The power driver is a .32 caliber Smith and Wesson blank cartridge.
Upon the firing pin striking the cartridge, the resulting temperature
and pressure increase forces the probe out at a high velocity.
The following is an overview of all of the external features of the
Windsor® Probe Power Driver.
9 www.ndtjames.com
Power Driver Overview
Instrument Functions
Item # Definition
1 Barrel – Used to hold the probe, driver head, breech
plug and power load.
2 Breech Plug – Used to hold the power load in the
breech.
3 Handle – Used to hold and driver while firing and
houses the trigger to fire the probe.
4 Charge Extraction Tool – Used to remove the charge
from the breech.
www.ndtjames.com 10
Power Driver Overview
2. Pull the barrel away from the handle and fold the barrel
down.
11 www.ndtjames.com
Power Driver Overview
4. Prepare the test probe by screwing the driver head onto the
threaded end of the probe. Continue until it bottoms out and
cannot thread on anymore.
www.ndtjames.com 12
Power Driver Overview
5. Load the probe into the barrel by grasping the driver head and
with a twisting motion, insert the probe into the breech-end of
the driver barrel until it sits flush with the breech.
13 www.ndtjames.com
Power Driver Overview
6. Reinsert the breech plug so that it sits flush with the breech
surface.
www.ndtjames.com 14
Power Driver Overview
9. Fold the barrel up toward the handle and pull the two
together. Twist counterclockwise to close the barrel.
15 www.ndtjames.com
Power Driver Overview
Always use the low power range for testing concrete less than 28
days after placement or until the concrete has cured sufficiently to
cause loose probes (approximately 3800 to 4500 psi). If the probes
are not firmly embedded, change to Standard Power.
Warning! If the driver ever fails to fire, hold firmly against template
for 30 seconds, then remove the driver from the template, keep
pointed in a safe direction, open driver and remove the power load.
www.ndtjames.com 16
Power Driver Overview
Driving Probes
Individual probes, or the averages of only two probes, may not yield
results that accurately reflect the true strength of the concrete. One
is an indication, two a comparison, and three a statistically
significant result. This is also true of cores, cylinders, or any quality
control system.
Flat Surfaces
Place the single probe-locating template on concrete and drive
probe. Place the three-probe template over the first probe. Position
single probe-locating template in cut-away for the next two probes.
Warning! It is critical
that you fire with the
power driver
perpendicular to the
surface in order to
get a good reading.
Firing at an angle
can damage the
actuating template.
Also, be sure to wait
at least 3 minutes
between shots, to
allow the driver to
cool down.
Item # Definition
1 Power Driver
2 Single Probe Locating Template
3 Three Probe Template
17 www.ndtjames.com
Power Driver Overview
Vertical Surfaces
Once the first probe has been placed, the template will hang from
embedded probe while the driver assembly is reloaded.
Curved Surfaces
Use the single probe-locating template; place on concrete and drive
three probes individually. Probes must be set in groups of three. No
probe shall be located less than 7 inches (175 mm) from any other
probe, nor less than 4 inches (100 mm) from the edge of a concrete
surface. (ASTM C-803/C803M-03, Sect.8.1.1) (2010)
www.ndtjames.com 18
Electronic Measuring Unit Overview
19 www.ndtjames.com
Electronic Measuring Unit Overview
Instrument Functions
Item # Definition
1 Sensor bushing.
2 Connects to the USB port location to upload data onto
your PC.
3 2 AA battery location; unscrew open to replace.
4 Windsor® Probe display screen.
5 Windsor® Probe control panel.
6 Displays meter name, model number, and serial
number.
Determining Results
Determining results can be done manually with the optional depth
gauge or automatically with the WP-725 electronic measuring unit.
www.ndtjames.com 20
Electronic Measuring Unit Overview
Figure 16: Leveling the Measuring Plate with the Concrete Surface
21 www.ndtjames.com
Electronic Measuring Unit Overview
6. Push the WP-725 over the end of the cap until the bushing
touches the bottom gauge plate.
Recording Results
See p. 54 of this manual for a sample form for recording test
results. It is recommended that you keep this as a master copy and
photocopy it as needed. Every time the probe is used, one of these
forms should be completed.
www.ndtjames.com 22
Electronic Measuring Unit Overview
Probe Removal
Once the tests are completed and the measurements have been
taken, the probes can be removed.
2. Place the round withdrawal barrel over the probe with the flat
face down. The withdrawal barrel serves as a washer so that
the hexagonal nut can rotate freely.
Notice that the hexagonal nut has two different thread lengths
inside. This is designed to accommodate different levels of
probe penetration.
23 www.ndtjames.com
Electronic Measuring Unit Overview
4. With the provided wrench, turn the wrench clockwise to pull the
probe out of the concrete surface.
Figure 22: Pulling the Probe out of the Concrete Using a Wrench
www.ndtjames.com 24
Electronic Measuring Unit
Figure 23: Leveling the Measuring Plate with the Concrete Surface
25 www.ndtjames.com
Electronic Measuring Unit
Key Definition
The Power button. Pressing the power button on the
WP-725 should bring up the Title Screen on the
display. Depressing it again should turn-off the unit.
The Esc key. Pressing this at any time returns user to
the Title Screen and cancels any action.
The Enter key. Use this key to enter data into the unit.
Title Screen
When turned on, the title screen should display the following:
www.ndtjames.com 26
Electronic Measuring Unit
Item # Definition
1 Displays the current date in a day/month/year format.
2 Displays the current time in a 24-hour format.
3 Displays the number of stored tests.
Enter Setup
This function sets the parameters used by the measuring unit,
allowing the user to view and modify the settings, set the clock, and
calibrate the measuring device. The Setup function is displayed
when you press the Down Arrow key once from the main screen.
Pressing Enter when Enter Setup is displayed takes the user to the
Setup menu. By using the arrow keys the user can display the other
options of this menu.
27 www.ndtjames.com
Electronic Measuring Unit
Show Settings
Pressing Enter when Show Settings is displayed takes you to the
screen below.
Item # Definition
1 Power – Corresponds to the driver power that was
used. Low should be set for probes pushed two and
one half inches down the barrel. Standard should be
set if the probe was used in the standard manner. High
Perf should be set if this system is used on high
performance concrete.
2 Agg – This corresponds to aggregate size. The setting
for this will be either Mortar- M, 1 in.- S and 2 in.- L.
This should exceed the largest size aggregate in the
mix design.
3 Density – This corresponds to the density of the
concrete, either standard weight (Std Wght) or light
weight (Lgt Wght).
4 Hard – This corresponds to the Mohs’ number scale for
hardness of the aggregate in the mix design. NOTE:
This is not applicable to high performance and light
weight concrete mix designs. If you selected Lgt Wght,
you do not see a Mohs’ number but a measure of the
concrete density in lbs/cu. in. See “Set Settings” on p.
32.
www.ndtjames.com 28
Electronic Measuring Unit
Calibrate
29 www.ndtjames.com
Electronic Measuring Unit
This returns the user back to the main screen, and the unit is now
ready for use in the measure mode.
www.ndtjames.com 30
Electronic Measuring Unit
Set Clock
Pressing Enter when Set Clock is displayed will take you to a
screen where you can set the instrument clock.
Use this item to set the time and date. Use the arrow keys to adjust
the value under each cursor and the Enter key to validate.
31 www.ndtjames.com
Electronic Measuring Unit
Set Settings
Pressing Enter with this displayed brings the user through the
procedure to modify the settings used in correlating exposed probe
length to compressive strength. Use the arrow keys to change
selections, and use the Enter key to validate selections.
Setting Description
Set Power Use the arrow keys to select between low, standard
and high performance. Press “Enter” to select.
Set Density Use the arrow keys to select between standard
weight (Std Wght) and light weight (Lgt Wght)
concrete. Press “Enter” to choose. See the screen
on the next page for more details.
Set Mohs This is the hardness value of the aggregate.
Perform a scratch test to determine the Mohs
number, as on p. 8. The value is adjusted using the
arrow keys. Press “Enter” to accept the selection.
Set Units Use the arrow keys to select either PSI or MPa.
Press “Enter” to accept the selection. (Note: Older
versions may not have this adjustable option.)
Set Aggr Size The aggregate size needs to be specified. Adjust
the range using the arrow keys and use “Enter” to
validate. The setting for this will be either: Mortar -
M, 1 in.- S and 2 in.- L. (This setting should exceed
the largest size aggregate in the mix design.)
www.ndtjames.com 32
Electronic Measuring Unit
33 www.ndtjames.com
Electronic Measuring Unit
Download Tests
This function downloads data out of the USB port on backside of
unit next to the measuring cap bushing. It will also erase data saved
in the unit.
www.ndtjames.com 34
Electronic Measuring Unit
Show Tests
This function displays the data stored in the WP-725 on the display.
Press Enter to cycle through the tests.
35 www.ndtjames.com
Electronic Measuring Unit
Run Test
This function allows the user to quickly average three
measurements according to ASTM standards C-803. This will also
store all data involved in the test as well as the time and date of the
test. For storing single probe readings, simply store a single probe
as all three readings.
Press Enter during each reading to save the data, and move to the
next test. The average value of three tests will be calculated.
After each test, you will see a screen like the one on the next page.
www.ndtjames.com 36
Electronic Measuring Unit
Item # Definition
1 Displays the exposed probe length in inches.
2 Displays the test number. This number increments from
1 to 3 as you run/store each test result.
3 Displays the compressive strength in psi (or MPa).
4 Displays the power of the driver. Corresponds to the
driver power that is used. Low (L) should be set for
probes pushed two and one half inches down the
barrel. Standard (S) should be set if the probe was
used in the standard manner. High Perf (H) should be
set if this system is used on high performance
concrete.
After storing three test results, you will see another screen like the
screen above. In place of the test number (position # 2), you will
see “Ave” to indicate that the final result is the average of the three
tests. The strength value (psi or MPa), will be the average value of
the three test results.
37 www.ndtjames.com
Standard Power Table
Measure
This function is used simply for a quick reading of exposed probe
length and compressive strength. Data cannot be stored from this
function.
www.ndtjames.com 38
Standard Power Table
39 www.ndtjames.com
Standard Power Table
TEST WELL
BUILD WELL
40 www.ndtjames.com
Standard Power Table
41 www.ndtjames.com
Low Power Table
42 www.ndtjames.com
Low Power Table
TEST WELL
BU!LD WELL
3127 N. K dd e Ave .
Chltcoo, IL 60618-4545
USA
Pho11te USA: l-8(»426-6500
Pho11te 1,u' I: +l -773-463 -6565
FOY: +l - 773-463 -0009
Emo /I: ltJ/ af> l'tdr me .com
htW-IIW'M"DdCEPfflfi rom
www.ndtjames.com 44
Low Power Table
BUILD WELL
www.ndtjames.com 44
Software Installation
Software Installation
The PC software that has been developed for the Windsor® Probe
is for data upload only. To install the software (known as
Windsorlinx) on your PC follow these instructions.
45 www.ndtjames.com
Software Installation
www.ndtjames.com 46
Software Installation
47 www.ndtjames.com
Software Installation
www.ndtjames.com 48
Software Installation
49 www.ndtjames.com
Software Installation
www.ndtjames.com 50
Software Installation
51 www.ndtjames.com
Where to Probe
Where to Probe
www.ndtjames.com 52
Where to Probe
53 www.ndtjames.com
Troubleshooting
Troubleshooting
www.ndtjames.com 54
Troubleshooting
55 www.ndtjames.com
Maintenance
Maintenance
Although built for rugged use, the WP-500 does require some care
and precautions. Avoid excessive exposure to water. Although the
unit is splash-proof, full immersion in water should be avoided.
Beware of dirt and grime, both on the front panel and in the probe
measuring cap bushing. Excessive dust and grime can cause
damage to the face plate and care should be taken that dirt and
dust inside the measuring cap bushing does not damage any
mechanical components. Cleaning can be accomplished with a
damp rag.
The Windsor® Power Driver is a precision instrument and requires
reasonable maintenance to ensure long trouble-free service and to
give the best results. The following procedure should be followed:
www.ndtjames.com 56
Maintenance
The Driver unit (WP-500) includes two sealed sections; the handle
assembly and the barrel assembly. These sections should never be
disassembled in the field.
For cleaning, simply immerse in a pail of any cleaning solvent and
then let air-dry. Wipe exterior parts with WD-40 or other similar light
oil products.
In addition to the above steps, it is recommended that the
Windsor® Probe System be serviced and calibrated on an annual
basis.
57 www.ndtjames.com
Sample Test Form
www.ndtjames.com 58
Sample Test Form
59 www.ndtjames.com
Safety
Safety
1. The person using the tool must be trained in its use by the
manufacturer, be responsible, be over 18 years old and not
color blind.
2. If a misfire occurs, without withdrawing the tool wait 10
seconds then re-cock and fire again. If it still misfires, wait
another 10 seconds then withdraw the tool. In accordance
with the manufacturer’s instructions, remove the cartridge
and seek instructions from the cartridges supplier on how to
return it for destruction. CONTRACTORS MUST NOT
DESTROY OR DISPOSE OF CARTRIDGES
THEMSELVES.
3. Cartridges must be stored under lock and key in a cool, dry
store and only issued under controlled conditions.
4. The operator must wear suitable PPE, e.g. eye, head, and
ear protection. Eye protection shall conform to EN
166:2002 and ear protection shall conform to EN 458:2004.
Both eye and ear protection shall conform to directive
89/686/EEC.
5. The area must be cleared of all unnecessary personnel.
6. The tool must not be used in a situation where there is a
risk of fire or explosion.
7. Always use original spare parts. Modifications to the tool
are prohibited.
8. Never hold the nose of the tool against yourself or other
people.
9. Read the (Material) Safety Data Sheet before handling the
product. See (M)SDS page for more information on the
power loads.
www.ndtjames.com 60
Safety
PRECAUTIONS:
-Do not fire this product without a probe. Firing without a probe will
create a loud noise which can be damaging to hearing.
-Do not fire a probe into anything that is not concrete. It is possible
to activate the driver against a non-concrete object. Firing into
anything other than concrete can result in pinning the template and
probe into the material, or the probe not penetrating the material (if
too hard) and bouncing off. And thus, cause a safety issue for the
user.
-Do not fire this product closer than six inches from the edge of the
concrete slab. Firing too close to the edge can result in a section of
concrete breaking off and causing injury.
-Do not fire into concrete that is less than six inches thick. Firing
into too thin of a concrete slab can result in concrete chips being
ejected out of the other side, potentially causing injury.
-Do not fire into a round column that is less than 36” in diameter.
-The firing template will deform over time. If the inner hole of this
template is deformed it can cause probe breakage, scattering of
concrete chips or even allow the probe to pass through the
template. Therefore, it is advised a new firing template be
purchased every year to replace worn templates.
-Do not load more than one probe into the product at a time. This
could damage the unit.
-Do not load unless you are ready to use the product, and always
unload before work breaks or before storing the product at the end
of the day.
61 www.ndtjames.com
Safety
-Do not use the product if there are any noticeable defects. If the
®
product is not working properly, send it back to James Instruments
for repair. The condition of the driver should be checked by the
operator before every use and after each long work pause by visual
examination.
-Do not fire into spalled concrete. Firing into a surface that isn’t
reasonably flat causes more debris to shoot out from under the
template.
-Do not drill a pilot hole. This will yield invalid test results.
-Wait at least three minutes between firing two probes. The inside
of the barrel heats up every time it is fired and should be allowed to
cool.
-Only use product against a stable surface. Keep one hand on the
barrel and the other on the handle when firing.
-Make sure you are properly balanced and not leaning when firing
the unit.
-Do not fire into reinforcement bars (or rebar). Fire probes into
concrete only. Firing a probe into rebar has a similar effect as firing
into a material that is too hard (as mentioned above). Use a rebar
www.ndtjames.com 62
Safety
®
locator such as the James Instruments Rebarscope to locate
rebar.
(M)SDS for Power Load:
63 www.ndtjames.com
Safety
www.ndtjames.com 64
Specifications
Specifications
65 www.ndtjames.com
Warranty Information
Warranty Information
1. Contract
Unless otherwise stated all sales transactions are expressly subject to these
terms and conditions. Modification or additions will be recognized only if
®
accepted in writing by an authorized Officer of James Instruments Inc.
(hereinafter referred to as “James” or the “Company”), or an officially
designated representative. PROVISIONS OF BUYER'S PURCHASE ORDER
OR OTHER DOCUMENTS THAT ADD TO OR DIFFER FROM THESE TERMS
AND CONDITIONS ARE EXPRESSLY REJECTED. NO WAIVER OF THESE
TERMS AND CONDITIONS OR ACCEPTANCE OF OTHERS SHALL BE
CONSTRUED AS FAILURE OF THE COMPANY TO RAISE OBJECTIONS.
2. Warranties
The Company only warrants the equipment manufactured or supplied by the
®
Company as set forth herein. James makes no other warranties, either
expressed or implied (including without limitation, warranties as to
®
merchantability or fitness for a purpose). In no event shall James be liable for
any type of special, consequential, incidental, or penal damages, whether such
damages arise out of or are a result of breach of contract, warranty, negligence,
strict liability or otherwise. Warranty shall not apply where the equipment
manufactured or supplied has been subject to accident, alteration, misuse,
abuse, improper storage, packing, force majeure, improper operation,
installation, or servicing. In addition, the following shall constitute the sole and
®
exclusive remedies of Buyer for any breach by James of its warranty
hereunder.
a. New Products
®
James warrants the equipment manufactured or supplied by
®
James as set forth herein. This limited warranty can only be
®
exercised by the original purchaser of the equipment from James or
®
authorized James Agent and is not transferable to any subsequent
owner or party. This limited warranty gives you specific legal rights,
and you may also have other rights which vary from case to case.
®
i. For James Equipment
®
James warrants that James's equipment will be free from
defects in materials and workmanship for a period of twenty-
four (24) months on the electronic portion and six (6) months
on the mechanical portion from the date of shipment of
www.ndtjames.com 66
Warranty Information
®
equipment from James to Buyer. Should any defects be
found and reported by the Buyer during the applicable
limited warranty period, the defect will be corrected upon
® ®
return of the item to James . James will, during the
applicable new equipment warranty period, provide the
necessary replacement parts and labor to correct the defect.
67 www.ndtjames.com
Warranty Information
www.ndtjames.com 68
Warranty Information
three (3) months from the date repairs are completed either
®
at James or at the customer site. Should the defect for
which the repair work was performed reoccur within this
®
period, James will supply the necessary parts and labor
®
(repair at James facility) or parts (repair at Buyer facility)
required to repair the original equipment defect for which the
repair parts and labor were required. Additional repair
charges that may be incurred in conjunction with any repair
®
service warranty event will be invoiced at the James
customer service rates and policies in effect at the time of
the event.
Excluded are all consumable and wear and tear items such
as impact bodies, probes, connection cables, etc. These
items are subject to usual wear and tear during usage. Refer
to the Consumable Wear and Tear Item section of this
warranty document.
c. Warranty Claims
4. Notices
Notice by either the Company or Buyer will be made only by facsimile or similar
electronic transmission, effective on the first business day after confirmed
receipt, or by letter addressed to the) other party at its address as provided in
this Agreement, effective three (3) business days after deposit with the U.S.
69 www.ndtjames.com
Warranty Information
Postal Services, postage prepaid, or one (1) business day after deposit with a
recognized overnight express service.
5. Interpretation
Should any term or provision contained in the contract contravene or be invalid
under applicable law, the contract shall not fail by reason thereof but shall be
construed in the same manner as if such term or provision had not appeared
therein.
6. Assignability
Neither this contract nor any claim arising directly or indirectly out of or in
connection herewith shall be assignable by Buyer or by operation of law,
without the prior written consent of Company. This document shall be binding
upon and inure to the benefit of each party hereto and their respective
permitted successors and assigns.
7. Governing Law
This Agreement shall be governed by and construed in accordance with the
internal laws of the State of Illinois, without regard to its conflict of laws
provisions. Buyer and the Company expressly agree to submit to the personal
jurisdiction of the federal and/or stale courts silting in Chicago, Illinois, U.S.A.
and agree that such courts may be utilized if necessary to obtain injunctive or
any other relief. The Hague Convention and the United Nations Convention on
Contracts for the International Sale of Goods shall not apply to the construction
or interpretation of these Standard Terms and Conditions or affect any of its
provisions.
END.
www.ndtjames.com 70
Notes:
Repair Policy
Repair Policy
71 www.ndtjames.com
James Instruments Inc.
3727 N. Kedzie Ave.
Chicago, IL 60618-4503
USA
Tel: (773) 463-6565
Fax: (773) 463-0009
Purchase Date: _ _
Serial Number: _