Cro experiments
Cro experiments
To measure the voltage, frequency, time period, and phase angle of a given waveform using a
Cathode Ray Oscilloscope (CRO).
Apparatus Required:
Theory:
1. Voltage Measurement: The vertical deflection on the CRO screen corresponds to the
magnitude of the input signal. The voltage is measured by multiplying the vertical
displacement by the volts/division setting.
2. Time Period and Frequency Measurement: The time period is the time taken for one
complete cycle of the waveform, and it is measured by determining the horizontal
distance between two successive peaks (or zero crossings). Frequency is the reciprocal of
the time period.
3. Phase Angle Measurement: Phase angle is the difference in phase between two
waveforms (e.g., voltage and current) applied to the CRO. It is measured by finding the
time difference between corresponding points on the waveforms and calculating the angle
using the formula.
Experimental Procedure:
1. Connect the function generator to the CRO to provide the input signal. Set the desired
waveform (e.g., sine wave) with known amplitude and frequency.
2. Connect the CRO probe to the signal output.
3. Adjust the time base and vertical sensitivity (volts/div and time/div) on the CRO to
display a stable waveform.
1. Adjust the time base so that at least one complete cycle of the waveform is visible.
2. Measure the horizontal displacement (in divisions) between two consecutive peaks or
zero crossings.
3. Multiply the horizontal displacement by the time/div setting to obtain the time period T.
4. Calculate the frequency f using the formula:
1
f=𝑇
1. If two signals are available (e.g., voltage and current), apply them to the CRO’s dual-
channel mode.
2. Measure the horizontal displacement between the peaks (or zero crossings) of the two
waveforms.
3. Multiply the measured horizontal displacement by the time/div setting to obtain the time
difference Δt.
4. Calculate the phase angle ϕ using the formula:
Δ𝑡
ϕ= ×3600
T
Observation Table
Horizonta
Horizonta Time
Vertical l Phase
Calcula l Perio
S. Displacem Volts/ Time/ Freque Displacem Time Angle
ted Displacem d (T)
N ent Div Div ncy (f) ent for Differe (ϕ) in
Voltage ent in
o. (Y₍div₎) in (V/div) (s/div) in Hz Phase nce (Δt) Degre
(V) (X₍div₎) in Secon
Divisions Angle es
Divisions ds
(X₍div₎)
1
2
3
Calculations:
Precautions:
Ensure that the CRO settings (time/div and volts/div) are properly calibrated before
taking measurements.
Avoid touching the probe tips while taking measurements to prevent errors due to body
capacitance.
Make sure the waveform is stable before noting measurements.
Conclusion:
The voltage, frequency, time period, and phase angle of the given waveform were successfully
measured using a CRO, demonstrating the functionality of the oscilloscope in analyzing
electrical signals.
2nd Experiment
Objective:
To measure the rise time, fall time, and delay time of an electronic signal using a Cathode Ray
Oscilloscope (CRO).
Apparatus Required:
Theory:
Rise Time: The time required for a signal to change from 10% to 90% of its maximum
amplitude.
Fall Time: The time required for a signal to drop from 90% to 10% of its maximum
amplitude.
Delay Time: The time interval between the application of an input signal and the
response at the output.
Procedure:
Observations:
Record the measured rise time, fall time, and delay time from the CRO display for the given
waveform.
Results:
3rd Experiment
Objective:
To measure the time period, frequency, and average period of a signal using a frequency counter.
Apparatus Required:
1. Frequency Counter
2. Signal Source (Function Generator or Oscillator)
3. Connecting Wires
Theory:
Frequency (f): The number of cycles a signal completes per second, measured in Hertz
(Hz).
1
Time Period (T): The duration of one complete cycle of the signal, where T=𝑇
Average Period: The mean value of the time period measured over multiple cycles,
which helps to improve accuracy by reducing random measurement errors.
Procedure:
Calculations:
1
1. Time Period: T=𝑇
2. Average Period (Manual Calculation): Average Period=∑Ti / n.
o Where Ti is each individual time period measurement and n is the number of
measurements.
Observations:
Frequency (f) in Hz
Time Period (T) in seconds (s)
Average Period (if calculated manually or displayed directly)
Results:
Calculate the average period if necessary and verify that the measurements align with the
expected values for the input signal's frequency and time period.