Nuclear Medicine 02
Nuclear Medicine 02
P.B. ZANZONICO
Department of Medical Physics,
Memorial Sloan Kettering Cancer Center,
New York, United States of America
10.1. INTRODUCTION
287
CHAPTER 10
Anode
Photomultiplier
High voltage Bias voltage
tube
500–1500 V 10–500 V
Cathode
Photocathode Detector
material
X or ray
Crystal
X or ray
FIG. 10.1. Basic design and operating principles of (a) scintillation and (b) ionization
detectors.
At a bias voltage of 300 V, all of the primary electrons (i.e. the electrons
produced directly by ionization of the detector material by the incident radiation)
are collected at the anode and the detector signal is, thereby, maximized. Since
288
NON-IMAGING DETECTORS AND COUNTERS
109
106
there are no additional primary electrons to collect, increasing the bias voltage
further (up to 600 V) does not increase the signal. The 300–600 V range,
where the overall signal is equivalent to the number of primary electrons and,
therefore, proportional to the energy of the incident radiation, is called the
ionization chamber region. At a bias voltage of 600–900 V, however, the large
electrostatic force of attraction of the anode accelerates free electrons, as they
travel towards the anode, to sufficiently high speeds to eject additional orbital
electrons (i.e. secondary electrons) within the sensitive volume, contributing to
an increasing overall signal — the higher the voltage, the more energetic the
electrons and the more secondary electrons are added to the overall signal. The
number of electrons comprising the overall signal is, thus, proportional to the
289
CHAPTER 10
primary number of electrons and the energy of the incident radiation, and the
600–900 V range is, therefore, called the proportional counter region. As the bias
voltage is increased further, beyond 900 V (up to 1200 V), free electrons (primary
and secondary) are accelerated to very high speeds and strike the anode with
sufficient energy to eject additional electrons from the anode surface itself. These
tertiary electrons are, in turn, accelerated back to the anode surface and eject even
more electrons, effectively forming an electron ‘cloud’ over the anode surface and
yielding a constant overall signal even with further increase in the bias voltage.
The 900–1200 V range is called the Geiger counter (or Geiger–Müller) region.
Importantly, the magnitude of the charge represented by this electron cloud is
independent of the number of electrons initiating its formation. Therefore,
in contrast to ionization chamber and proportional counter signals, the Geiger
counter signal is independent of the energy of the incident radiation. Finally,
beyond a bias voltage of 1200 V, atoms within the detector material are ionized
even in the absence of ionizing radiation (i.e. undergo spontaneous ionization),
producing an artefactual signal; the voltage range beyond 1200 V is known as the
spontaneous discharge region.
Although the bias voltage is the principal difference among different types
of gas filled ionization detectors, there may be other differences. The sensitive
volume, for example, may or may not be sealed. Unsealed sensitive volumes
contain only air at atmospheric (ambient) pressure. For detectors with unsealed
volumes, the signal must be corrected by calculation for the difference between
the temperature and pressure at which the detector was calibrated (usually
standard temperature and pressure: 27°C and 760 mm Hg, respectively) and
the ambient conditions at the time of an actual measurement. For detectors with
sealed volumes, gases other than air (e.g. argon) may be used and the gas may be
pressurized, providing higher stopping power, and, therefore, higher sensitivity,
than detectors having a non-pressurized gas in the sensitive volume. In addition,
different geometric arrangements of the anode and cathode, such as parallel
plates (used in some ionization chambers), a wire along the axis of a cylinder
(used in Geiger counters), etc., may be used.
The functional properties and, therefore, the applications of the various
types of ionization detector — ionization chambers, proportional counters and
Geiger counters — are largely dictated by their respective bias voltage dependent
signal (Table 10.1). Ionization chambers are widely used in radiation therapy to
calibrate the output of therapy units and in nuclear medicine as dose calibrators
(i.e. devices used to assay radiopharmaceutical activities). The relatively
low sensitivity of ionization chambers is not a major disadvantage for such
applications, as the radiation intensities encountered are typically rather large.
The stability of the response is an important advantage, however, as it allows
the use of unconditioned AC electrical power (i.e. as provided by ordinary wall
290
NON-IMAGING DETECTORS AND COUNTERS
outlets). Proportional counters, because of their need for a stable bias voltage
and, therefore, specialized power supplies, are restricted to research applications
(e.g. in radiobiology) where both higher sensitivity and the capability of energy
discrimination may be advantageous. Proportional counters often employ an
unsealed, gas flow-through sensitive volume. Geiger counters, because of their
high sensitivity and stability with respect to voltage (allowing the use of a portable
power supply such as an ordinary battery), are widely used as survey meters to
measure ambient radiation levels and to detect radioactive contamination. For
such applications, sensitivity, and not energy discrimination, is critical. As with
dose calibrators, Geiger counters have sealed sensitive volumes, avoiding the
need for temperature–pressure corrections.
In addition to the more familiar gas filled ionization detectors, solid state
ionization detectors are now available. Such detectors are based on a family of
materials known as semiconductors. The pertinent difference among (crystalline)
solids — conductors, insulators and semiconductors — is related to the widths of
their respective electron ‘forbidden’ energy gaps. In a semiconductor, the highest
energy levels occupied by electrons are completely filled but the forbidden
gap is narrow enough (<2 eV) to allow radiative or even thermal excitation at
room temperature, thereby allowing a small number of electrons to cross the
gap and occupy energy levels among the otherwise empty upper energy levels.
Such electrons are mobile and, thus, can be collected by a bias voltage, with the
amplitude of the resulting signal being equivalent to the number of electrons
produced by the radiation and, therefore, proportional to the radiation energy.
Although many semiconductor materials have suitably large energy gaps (~2 eV),
techniques must be available to produce crystals relatively free of structural
defects. Defects (i.e. irregularities in the crystal lattice) can trap electrons
produced by radiation and, thus, reduce the total charge collected, degrading
the sensitivity and overall detector performance of semiconductors. Practical,
reasonably economical crystal growing techniques have been developed
for cadmium telluride (CdTe), cadmium zinc telluride (CZT) and mercuric
iodide (HgI2), and these detectors have been incorporated into commercial
intra-operative gamma probes and, on a limited basis, small field of view gamma
cameras.
291
CHAPTER 10
292
NON-IMAGING DETECTORS AND COUNTERS
the even larger positive voltage, ~400 V, on the second dynode. The impact of
these electrons on the second dynode surface ejects an additional three electrons,
on average, for each incident electron. Typically, a PMT has 10–12 such dynodes
(or stages), each ~100 V more positive than the preceding dynode, resulting in
an overall electron amplification factor of 310–312 for the entire PMT. At the last
anode, an output signal is generated. The irregularly shaped PMT output signal
is then shaped by a preamplifier and further amplified into a logic pulse that
can be further processed electronically. The resulting electrical pulses, whose
amplitudes (or ‘heights’) are proportional to the number of electrons produced
at the PMT photocathode are, therefore, also proportional to the energy of the
incident radiation. These pulses can then be sorted according to their respective
heights by an energy discriminator (also known as a pulse height analyser) and
those pulses with a pulse height (i.e. energy) within the preset photopeak energy
window (as indicated by the pair of dashed horizontal lines overlying the pulses
in Fig. 10.3) are counted by a timer/scaler.
Advantageous features of scintillation detectors include:
High mass density and effective atomic number maximize the crystal
stopping power (i.e. linear attenuation coefficient μ) and, therefore, sensitivity.
In addition, a higher atomic number crystal will have a higher proportion of
photoelectric than Compton interactions, thus facilitating energy discrimination
of photons which underwent scatter before entering the crystal. High light output
reduces statistical uncertainty (noise) in the scintillation and associated electronic
signal and, thus, improves energy resolution and scatter rejection. Other detector
considerations include:
293
CHAPTER 10
10.3.1. Sensitivity
Sensitivity (or efficiency) is the detected count rate per unit activity (e.g. in
counts per minute per megabecquerel). As the count rate detected from a given
activity is highly dependent on the source–detector geometry and intervening
media, characterization of sensitivity can be ambiguous. There are two distinct
components of overall sensitivity, geometric sensitivity and intrinsic sensitivity.
Geometric sensitivity is the fraction of emitted radiations which intersect, or strike,
the detector, that is, the fraction of the total solid angle subtended at the detector
by the source. It is, therefore, directly proportional to the radiation-sensitive
detector area and, for a point source, inversely proportional to the square of the
source–detector distance. Intrinsic sensitivity is the fraction of radiation striking
the detector which is stopped within the detector. Intrinsic sensitivity is directly
related to the detector thickness, effective atomic number and mass density, and
decreases with increasing photon energy, since higher energy photons are more
penetrating and are more likely to pass through a detector without interacting.
Characteristic X rays and γ rays are emitted from radioactively decaying
atoms with well defined discrete energies. Even in the absence of scatter, however,
output pulses from absorption of these radiations will appear to originate over a
range of energies, reflecting the relatively coarse energy resolution of the detector.
For this reason, many radiation detectors employ some sort of energy-selective
294
NON-IMAGING DETECTORS AND COUNTERS
counting using an energy range, or window, such that radiations are only counted
if their detected energies lie within that range (Figs. 10.3 and 10.4(a)). At least
for scintillation detectors, a so-called ‘20% photopeak energy window’, Eγ ± 10%
of Eγ, (e.g. 126–154 keV for the 140 keV γ ray of 99mTc) is employed, where Eγ
is the photopeak energy of the X ray or γ ray being counted. For such energy-
selective counting, overall sensitivity appears to increase as the photopeak energy
window is widened. However, this results in acceptance of more scattered as well
as primary (i.e. unscattered) radiations.
For each radionuclide and energy window (if applicable) for which a
particular detector is used, the detector should be calibrated, that is, its sensitivity
(e.g. in cpm/MBq) S determined, at installation and periodically thereafter:
Rg − R b
S= (10.1)
A 0e −∆t
where
Rg is the gross (i.e. total) count rate (cpm) of the radionuclide source (RS);
Rb is the background (BG), or blank, count rate (cpm);
A0 is the activity (MBq) of the radionuclide source at calibration;
λ is the physical decay constant (in month–1 or a–1, depending on the half-life)
of the calibration radionuclide;
and ∆t is the time interval (in months or years, respectively, again depending
on the half-life) between the calibration of the radionuclide and the current
measurement.
295
CHAPTER 10
Output
Preamplifier
signal
+ 1200 V
Anode
106:1e
+ 1100 V Amplifier
Photomultiplier tube
Magnetic
+ 1000 V
shielding
High ×106
+ 600 V
voltage
Dynodes + 500 V supply Energy
discriminator
+ 400 V
Focusing
+ 300 V Energy E0
grid
e
Photocathode
Timer/scaler
Entrance
window
Light Light pipe
photon Display
Reflective
inner surface Scintillator
of crystal housing crystal
X or ray
FIG. 10.3. The basic design and operating principle of photomultiplier tubes and scintillation
detectors.
296
NON-IMAGING DETECTORS AND COUNTERS
20% photopeak
energy window
662 keV
(b) 137Cs
100
Relative number of counts
80
Maximum height
60
DE 46
FWHM(%) = = ´ 100 = 7% FWHM
E = 46 keV
Eg 662
40
½ maximum
20 height
FIG. 10.4. (a) Energy spectrum for the 662 keV γ rays emitted by 137Cs, illustrating the
definition of energy resolution as the percentage full width at half maximum (FWHM) of the
photopeak energy Eγ. (b) Energy spectrum for the 140 keV γ rays emitted by 99mTc, illustrating
the contributions of primary (unscattered) and scattered radiation counts. In (a) and (b), the
energy spectra were obtained with a thallium-doped sodium iodide (NaI(Tl)) scintillation
detector.
297
CHAPTER 10
detector, however, even radiation which is not counted (i.e. which interacts
with the detector during the dead time of a previous event) prevents counting
of subsequent incoming radiations during the time interval corresponding to
its dead time. Geiger counters (with quenching gas) behave as non-paralysable
systems but most detectors, including scintillation detector based systems,
such as well counters, γ cameras and PET scanners, are paralysable. Modern
scintillation detectors generally incorporate automated algorithms to yield count
rates corrected for dead time count losses.
Non-paralysable
Paralysable
FIG. 10.5. The observed versus true count rates for paralysable and non-paralysable radiation
detectors. For paralysable detectors, the observed count rate increases to a maximum value
with increasing true count rate (e.g. with increasing activity) and then decreases as the true
count rate is further increased. For non-paralysable detectors, the observed count rate also
increases with increasing true count rate, asymptotically approaching a maximum value as the
true count rate is further increased. In both cases, the maximum observed count rate is directly
related to the detector’s dead time τ.
298
NON-IMAGING DETECTORS AND COUNTERS
(i.e. are operated at a relatively low potential difference between the anode
and cathode) and are designed for use where relatively high fluxes of X rays
and γ rays are encountered. The more familiar Geiger counters are operated at
a high potential difference (Fig. 10.2), providing a high electron amplification
factor and, thus, high sensitivity. Geiger counters are, therefore, well suited for
low level surveys, for example, checking for radioactive contamination. Both
cutie-pies and Geiger counters are generally calibrated in terms of exposure rate.
As an ionization chamber, the cutie-pie’s electron signal depends on the energy
of the detected X rays or γ rays and is, therefore, directly related to the exposure
for all radionuclides. For Geiger counters, on the other hand, signal pulses have
the same amplitude regardless of the energy of the incoming radiation. Thus,
Geiger counter calibration results apply only to the particular radionuclide(s)
used to calibrate the counter (see below). Solid state detectors employ a
non-air-equivalent crystal as the detection medium and, thus, cannot measure
exposure rates, only count rates.
1
The solid angle subtended at the centre of a sphere by the total surface of the sphere
is 4π steradians; a steradian is the unit of solid angle. A well-type detector configuration
approximates a point source completely surrounded by a detector, yielding a per cent geometric
efficiency of 100%, and is, therefore, referred to as a ‘4π’ counting geometry.
299
CHAPTER 10
terms of activity (e.g. MBq) using the measured isotope specific calibration
factor (cpm/MBq) (see Eq. (10.1)). Such devices are generally comprised of a
cylindrical scintillation crystal (most commonly, NaI(Tl)) with a circular bore
(well) for the sample drilled part-way into the crystal and backed by a PMT and
its associated electronics. An alternative design for well counters is the so-called
‘through-hole’ detection system in which the hole is drilled through the entire
crystal. The through-hole design facilitates sample exchange, and because
samples are centred lengthwise in the detector, yields a more constant response
for different sample volumes as well as slightly higher sensitivity than the well
counters. In both the well and through-hole designs, the crystal is surrounded by
thick lead shielding to minimize the background due to ambient radiation.
Scintillation counters are often equipped with a multichannel analyser
for energy (i.e. isotope) selective counting and an automatic sample changer
for automated counting of multiple samples. Importantly, because of their high
intrinsic and geometric efficiencies (resulting from the use of a thick crystal and
a well-type detector configuration, respectively), well counters are extremely
sensitive and, in fact, can reliably be used only for counting activities up to
~100 kBq; at higher activities, and even with dead time corrections applied,
dead time counting losses may still become prohibitive and the measured counts
inaccurate. Modern well counters often include an integrated computer which is
used to create and manage counting protocols (i.e. to specify the isotope, energy
window, counting interval, etc.), manage sample handling, and apply background,
decay, dead time and other corrections, and, thus, yield dead time-corrected net
count rate decay corrected to the start of the current counting session.
300
NON-IMAGING DETECTORS AND COUNTERS
Side view
Detector (crystal)
Detector thickness
Collimator
End view
FIG. 10.6. A typical intra-operative probe (Node Seeker 900, Intra Medical Imaging LLC,
Los Angeles, CA, United States of America). (a) Hand-held detector. (b) Control and display
unit which not only displays the current count rate but also often emits an audible signal, the
tone of which is related to the count rate, somewhat analogous to the audible signal produced
by some Geiger counters. (c) A diagram of the detector and collimator assembly of a typical
intra-operative probe, illustrating that the detector (crystal) is recessed from the collimator
aperture. (Courtesy of Intra Medical Imaging LLC, Los Angeles, CA, USA.)
301
CHAPTER 10
date, the few clinical studies directly comparing scintillation and semiconductor
intra-operative probes have not provided a clear choice between the two types of
probe.
Historically, organ uptake probes have been used almost exclusively for
measuring thyroid uptakes and are, thus, generally known as ‘thyroid’ uptake
probes.2 Thyroid uptake (i.e. the decay-corrected per cent of administered activity
in the thyroid) may be measured following oral administration of 131I-iodide,
123
I-iodide or 99mTc-pertechnetate. The uptake probe is a radionuclide counting
system consisting of a wide-aperture, diverging collimator, a NaI(Tl) crystal
(typically ~5 cm thick by ~5 cm in diameter), a PMT, a preamplifier, an amplifier,
an energy discriminator (i.e. an energy window) and a gantry (stand) (Figs 10.7(a)
and (b)). Commercially available thyroid uptake probes are generally supplied as
integrated, computerized systems with automated data acquisition and processing
capabilities, yielding results directly in terms of per cent uptake.
Each determination of the thyroid uptake includes measurement of the
thyroid (i.e. neck) count rate, the ‘thigh’ background count rate (measured
over the patient’s thigh and presumed to approximate the count contribution of
extra-thyroidal neck activity), the standard count rate (often counted in a neck
phantom simulating the thyroid/neck anatomy) and the ambient (i.e. ‘room’)
background, with a 1–5 min counting interval for each measurement. Based
on the foregoing measurements, and knowing the fraction of the administered
activity which is in the standard, the thyroid uptake is calculated as follows:
where
2
At one time, organ uptake probes were also used to measure kidney time–activity data
for the evaluation of renal function. In addition, organ uptake probes have been adapted to such
well counter applications as counting of blood samples and wipes.
302
nOn-IMAGInG DEtECtORs AnD COUntERs
FIG. 10.7. (a) A typical organ (‘thyroid’) uptake probe system, including an integrated
computer, set-up for a thyroid uptake measurement (AtomLab 950™ Thyroid Uptake System,
Biodex Medical Systems, Shirley, NY, USA). The rather large neck to collimator aperture
distance (typically of the order of 30 cm) should be noted. Although this reduces the overall
sensitivity of the measurement of the neck count rate, it serves to minimize the effect of the
exact size, shape and position of the thyroid, and the distribution of radioisotope within the
gland. (b) A diagram (side view) of the open, or ‘flat-field’, diverging collimator typically used
with thyroid uptake probes. (Courtesy of Biodex Medical Systems, Inc, Shirley, NY, USA.)
303
CHAPTER 10
A B P B 1/2
− × −
Total body activity (%) = t A t B t P t B × 100% (10.3)
1/2
A(0) − B(0) × P(0) − B(0)
t A(0) t B(0) t P(0) t B(0)
where
A and P are the anterior and posterior total body counts, respectively;
B is the room (background) counts;
tA, tP and tB are the counting intervals for anterior, posterior and room counts,
respectively;
As above, the total body activity may be corrected for radioactive decay
from the time of measurement to the time of administration (by multiplying
the right side of Eq. (10.3) by eλ∆t where λ is the physical decay constant of the
administered isotope and ∆t is the administration to measurement time interval).
304
NON-IMAGING DETECTORS AND COUNTERS
305
CHAPTER 10
306
NON-IMAGING DETECTORS AND COUNTERS
− ∆t
= A 0e Γ
D 2
(10.4)
d
where
and d is the distance between the reference source and the meter (Table 10.2).
3
At the installation of a dose calibrator, the geometry dependent response for 99mTc
must be measured and volume dependent (2–25 mL) correction factors relative to the ‘standard’
volume (e.g. 10 mL) derived. This procedure is required periodically following installation.
307
TABLE 10.2. LONG LIVED RADIONUCLIDES COMPRISING REFERENCE SOURCES FOR INSTRUMENTATION
308
QUALITY CONTROL
activity reading on each scale recorded; day to day readings should agree within
10%. for the accuracy test (sometimes also known as the energy linearity test), at
least two of the foregoing NisT-traceable reference sources are separately placed
in the dose calibrator and the activity reading on each scale recorded. for each
source, the measured activity on each scale and its current actual activity should
agree within 10%.
FIG. 10.8. Set of lead-lined plastic sleeves (CalicheckTM Dose Calibrator Linearity Test Kit,
Calicheck, Cleveland, OH, USA) for evaluation of dose calibrator linearity by the shield
method. The set is supplied with a 0.64 cm thick lead base, a colour coded unlined sleeve (to
provide an activity measurement equivalent to the zero time point measurement of the decay
method) and a six colour coded lead-lined sleeve providing attenuation factors nominally
equivalent to decay over 6, 12, 20, 30, 40 and 50 h, respectively. (Courtesy of Calicheck,
Cleveland, OH, USA.)
309
CHAPTER 10
the set of lead sleeves) between the source and the dose calibrator’s sensitive
volume, a decay-equivalent activity is measured for each sleeve. While the shield
method is much faster than the decay method for checking linearity (taking
minutes instead of days), an initial decay based calibration of the set of sleeves
is recommended to accurately determine the actual decay equivalence of each
shield.
The routine QC tests for well counters include checks of the photopeak
energy window (i.e. energy peaking) if the counter is equipped with a
multichannel analyser, background, constancy and efficiency (or sensitivity).
Prior to counting samples containing a particular radionuclide, the energy
spectrum should be checked to verify that the counter is properly ‘peaked’,
that is, that the radionuclide’s photopeak coincides with the preset photopeak
energy window4. For each photopeak energy window used, the background
count rate should be checked daily. Importantly, electronic noise as well as
ambient radiation levels, which may be relatively high and variable in a nuclear
medicine facility, will produce a non-zero and potentially fluctuating background
count rate. Furthermore, even trace contamination of the counting well will
produce inaccurately high count rate values. Accordingly, a ‘blank’ (i.e. an
empty counting tube or vial) should always be included to determine the current
background count. To check constancy, at least one NIST-traceable reference
source (Table 10.2) should likewise be counted each day; day to day net (i.e. gross
minus background) count rates should agree within 10%.
In addition, as noted above, for each radionuclide for which a particular
well counter is used, the counter should be calibrated — that is, its efficiency
(sensitivity) (in cpm/kBq) determined — at installation, annually and after any
repair (Eq. (10.1)).
In addition to daily battery and background checks (as done for survey
meters), QC tests of intra-operative probes should include a daily bias check
for both the primary and any backup battery to verify that bias voltage (or
high voltage) is within the acceptable range. As intra-operative probes may not
provide a display of the energy spectrum, it may not be possible to visually check
4
Isotope specific radionuclide counting or imaging with a scintillation detector is
commonly performed using a 20% photopeak energy window, equivalent to an energy range of
Eγ ± 10% where Eγ is the X ray or γ ray energy of the radionuclide.
310
NON-IMAGING DETECTORS AND COUNTERS
that the probe is properly peaked, that is, that the photopeak coincides with the
preset photopeak energy window. The lower counts or count rates resulting from
an inappropriate energy window may, therefore, go unnoticed. Thus, a long lived
reference source or set of reference sources (such as 57Co, 68Ge and/or 137Cs
(Table 10.2)) should be available for daily checks of count rate constancy; a
marked change (e.g. >±10%) in the net count rate from one day to the next may
indicate an inappropriate energy window setting or some other technical problem.
Ideally, the reference sources should each be incorporated into some sort of cap
that fits reproducibly over the probe so that spurious differences in count rates
due to variations in source–detector geometry are avoided.
BIBLIOGRAPHY
CHERRY, S.R., SORRENSON, J.A., PHELPS, M.E., Physics in Nuclear Medicine, 3rd edn,
Saunders, Philadelphia, PA (2003).
NINKOVIC, M.M., RAICEVIC, J.J., ANDROVIC, A., Air kerma rate constants for γ emitters
used most often in practice, Radiat. Prot. Dosimetry 115 (2005) 247–250.
ZANZONICO, P., Routine quality control of clinical nuclear medicine instrumentation: A brief
review, J. Nucl. Med. 49 (2008) 1114–1131.
ZANZONICO, P., HELLER, S., “Physics, instrumentation, and radiation protection”, Clinical
Nuclear Medicine (BIERSACK, H.J., FREEMAN, L.M., Eds), Springer Verlag, Berlin
Heidelberg (2007) 1–33.
311
CHAPTER 11
11.1. INTRODUCTION
312