Feeler Guage

The document details the calibration record for a Digimatic Caliper belonging to Mahmud Denims Ltd., conducted on February 4, 2015, with a due date for recalibration set for February 4, 2016. It includes observations of measurements, uncertainties, and standard equipment used, ensuring compliance with ISO 17025:2017. Additionally, it provides customer details for a calibration service request from Boxpak (Myanmar) Company Limited, along with calibration results for a feeler gauge conducted on June 13, 2023.

Uploaded by

Thiha Htun
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
8 views6 pages

Feeler Guage

The document details the calibration record for a Digimatic Caliper belonging to Mahmud Denims Ltd., conducted on February 4, 2015, with a due date for recalibration set for February 4, 2016. It includes observations of measurements, uncertainties, and standard equipment used, ensuring compliance with ISO 17025:2017. Additionally, it provides customer details for a calibration service request from Boxpak (Myanmar) Company Limited, along with calibration results for a feeler gauge conducted on June 13, 2023.

Uploaded by

Thiha Htun
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd
You are on page 1/ 6

OTS (PVT.) LTD.

DHAKA (BANGLADESH)
RECORD OF CALIBRATION AND ANALYSIS OF RESULTS

MAHMUD DENIMS LTD. Service Request No: 5011


Date of Calibration: 04/02/2015
Certificate No: 015/(18-03)/5011 Temp °C: 22.3
Due Date: 04/02/2016 RH%: 57.40

Details of Unit Under Calibration:


Model CD-8" ASX
Inst.Code AAL/DC/01
Description Digimatic Caliper
Manufacturer Mitutoyo
Serial Number 14285183
Range(mm) 01to50
Location of Calibration Laboratory
Readability (mm) 0.01
Visual Inspection OK

Details of Standard Equipment Used for Calibration:


SL No Name Manufacturer ID Number Certificate No. Validity Calibrated By
1 Check Master Mitutoyo OTS/CM/01 QSI/1024/15/02 12/12/2015 QSI

2 Granite Surface Plate Mitutoyo OTS/GSP/01 QSI/3749/12/12 18/11/2016 QSI

Calibration Procedure WI-05

OBSERVATION

UUC VALUE (mm)


SL No. UUC/STD READING (mm) Mean (mm) Uncertainty
1 2 3 4 5 6

1 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.000 ± 12.594180

2 10.00 9.99 9.99 10.00 9.99 10.00 10.00 9.995 ± 12.59418

3 30.00 29.99 29.99 29.99 30.00 29.99 29.99 29.992 ± 12.59418

4 50.00 50.00 49.99 49.99 49.99 50.00 49.99 49.993 ± 12.59418

5 70.00 70.00 69.99 70.00 70.00 70.00 70.00 69.998 ± 12.59418

6 90.00 89.99 90.00 90.00 90.00 89.99 89.99 89.995 ± 12.59418

7 110.00 110.00 110.00 109.99 110.00 110.00 110.00 109.998 ± 12.59418

8 130.00 130.00 130.00 130.00 129.99 130.00 129.99 129.997 ± 12.59418

9 150.00 150.00 150.00 150.00 150.00 149.99 150.00 149.998 ± 12.59418

10 200.00 200.00 200.01 200.01 200.01 200.01 200.01 200.008 ± 0.00000

11 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.000 ± 12.59418

12 10.00 10.00 10.00 9.99 10.00 10.00 10.00 9.998 ± 12.59418

13 30.00 29.99 30.00 30.00 30.00 29.99 30.00 29.997 ± 12.59418

14 50.00 49.99 49.98 50.00 50.00 49.99 49.99 49.992 ± 12.59418

15 70.00 69.99 69.99 70.00 70.00 69.99 69.99 69.993 ± 12.59418

16 90.00 90.00 90.00 90.00 90.00 90.01 90.00 90.002 ± 12.59418

17 110.00 110.00 110.00 109.99 109.99 110.00 109.99 109.995 ± 12.59418

18 130.00 130.00 130.00 130.00 130.00 129.99 130.00 129.998 ± 12.59418

Page 1 of 6
19 150.00 150.00 150.00 150.00 150.00 149.99 149.99 149.997 ± 12.59418

20 200.00 199.99 199.99 199.99 199.99 200.00 200.00 199.993 ± 0.00000

Uncertainty Calculation:
Reported Uncertainty will be: ± 12.59 µ

Form No: F02 (5.4/01) Issue No.02 Date of


Issue: 01/10/2012

Remarks:

Page 2 of 6
Certificate No. 023/(13-06)/4123
Issue Date 16/06/2023
Customer Details:

Name Boxpak (Myanmar) Company Limited.


Address Lot C-2 ,Zone A ,Thilawa Special Economic Zone , Thanlyin Township , Yangon
Myanmar.

Tel +95 943054230


E-mail [email protected]

Details of Unit Under Calibration (UUC):


Description Feeler Guage
Manufacturer N/P
Model/Type N/P
Serial Number BPM-006
ID No. N/P
Range/working Range Ref. On Obs.
Least Count Ref. On Obs.
Accuracy As Per Instrument
Location of Calibration Laboratory
Visual Inspection OK

Date of Inst. Receipt 13/06/2023


Date of Calibration 13/06/2023
Suggested Due Date 12/06/2024
Service Request No. 4123

Calibration Procedure The calibration has been performed under controlled conditions using OTS Laboratory reference
standards, which are periodically referenced to one or more of the primary standards traceable to
NPL/NIST or other national physical measures as equivalent to NIST. The calibration had been
performed in accordance with calibration procedure OTS. WI-05 M (Procedure based on IS-3651).

Calibration Result The details of standard equipment used for calibration & result of calibration are given in
page 2 & 3.

Conclusion For the status of measurements please refer to the guidance notes.

Environment: (certified against calibrated digital temperature & humidity meter)


Temperature (˚C) 25±4
Relative Humidity (%RH) <70

Change in temperature and relative humidity of the Laboratory during the calibration was less than 0.3°C per hour and
5.0% per 4 hours respectively.
This certificate is issued strictly in accordance with the requirements of ISO 17025:2017. All calibration equipments are
traceable to the International Standards . Documentary evidence is available upon request.

023/(13-06)/4123 Page 3 of 6
Details of Standard Equipment Used for Calibration:

Sl. No. Description Make Inst. Sl. /ID NO. Certificate No. Validity Calibrated By

023/(02-03)/OTS/
01 Digital Caliper Mitutoyo OTS/DC/04 01/03/2024 OTS
DC/04
Thermo 023/(20-03)/OTS/
02 Testo OTS/DL/01 19/03/2024 OTS
Hygrometer DL/01

Guidance Notes:
Status A The measurement is within tolerance, due allowances having been made for the uncertainty of the measurement.

Status B The measurement may be out of tolerance, due allowances having been made for the uncertainty of the measurement.

Status C The measurement is out of tolerance, due allowances having been made for the uncertainty of the measurement.

Status D No conclusion can be drawn, because the standard(s) do(es) not specify a tolerance for this measurement.

OBSERVATION:
Thickness: (verified against calibrated equipments)
Standard Value
Sl. No. UUC Value (mm) Error (mm) Tolerance (mm) Status Uncertainty (µm)
(mm)
01 0.05 0.05 0.00 N/S D
02 0.10 0.10 0.00 N/S D
03 0.20 0.20 0.00 N/S D
04 0.25 0.25 0.00 N/S D
05 0.30 0.30 0.00 N/S D
06 0.40 0.40 0.00 N/S D
±3.8
07 0.50 0.50 0.00 N/S D
08 0.60 0.60 0.00 N/S D
09 0.70 0.70 0.00 N/S D
10 0.80 0.80 0.00 N/S D
11 0.91 0.90 -0.01 N/S D
12 1.01 1.00 -0.01 N/S D

The overall uncertainty shall be calculated as per ISO “Guide to the Expression of Uncertainty in Measurement” (GUM). The
uncertainty represents an expanded uncertainty using a coverage factor k=2 to approximate a 95% confidence level.

023/(13-06)/4123 Page 4 of 6
Notes:
1. The values mentioned above are the mean readings.
2. No adjustment was done during the calibration.
3. Any section marked,“N/A” means Not Applicable,“N/P” means Not Provided,“N/R” means Not Readable “N/S” means Not Specified.

Calibrated By: Prepared By: Authorized By:

Zeyar Kyaw Myint Thiha Htun Animesh Nandy


(Calibration Engineer) (Calibration Engineer) (Asst. Technical Manager)

F01( 7.8/01/C) Rev No. 01 Date of Rev. :26/12/2018


***END***

023/(13-06)/4123 Page 5 of 6
Calculation of Measurment Uncertainty
Vernier Calipers

Type A evaluation
User Input Required
Measurment Record Formula
Sr. No. Measured Reading (mm) Master Instrument = CHECK MASTER
1 0.000 mm Uncertainty =+/- 1.9 (μm)
2 0.000 mm Accu. =+/- 5 (μm)
3 0.000 mm Thermal Expn.of Master = 11.5 µm/mtr/°c
4 0.000 mm
5 0.000 mm Expanded uncertainty :-
6 0.000 mm UE = +/- 12.59418 μm
Avarage = 0.000 mm with coverage factor k=2 for confidence level of 95 % and
Effective Degree of freedom = ∞
Standard Deviation = 0 mm

Degree of freedom = 5

Standard deviation of mean or standard uncertainty UA = 0.000000 (μm)