RECORD OF CALIBRATION AND ANALYSIS OF RESULTS
OBSERVATION
Page 1 of 6
19 150.00 150.00 150.00 150.00 150.00 149.99 149.99 149.997 ± 12.59418
Uncertainty Calculation:
Reported Uncertainty will be: ± 12.59 µ
Remarks:
Page 2 of 6
Certificate No. 023/(13-06)/4123
Issue Date 16/06/2023
Customer Details:
Calibration Procedure The calibration has been performed under controlled conditions using OTS Laboratory reference
standards, which are periodically referenced to one or more of the primary standards traceable to
NPL/NIST or other national physical measures as equivalent to NIST. The calibration had been
performed in accordance with calibration procedure OTS. WI-05 M (Procedure based on IS-3651).
Calibration Result The details of standard equipment used for calibration & result of calibration are given in
page 2 & 3.
Conclusion For the status of measurements please refer to the guidance notes.
Change in temperature and relative humidity of the Laboratory during the calibration was less than 0.3°C per hour and
5.0% per 4 hours respectively.
This certificate is issued strictly in accordance with the requirements of ISO 17025:2017. All calibration equipments are
traceable to the International Standards . Documentary evidence is available upon request.
023/(13-06)/4123 Page 3 of 6
Details of Standard Equipment Used for Calibration:
Sl. No. Description Make Inst. Sl. /ID NO. Certificate No. Validity Calibrated By
023/(02-03)/OTS/
01 Digital Caliper Mitutoyo OTS/DC/04 01/03/2024 OTS
DC/04
Thermo 023/(20-03)/OTS/
02 Testo OTS/DL/01 19/03/2024 OTS
Hygrometer DL/01
Guidance Notes:
Status A The measurement is within tolerance, due allowances having been made for the uncertainty of the measurement.
Status B The measurement may be out of tolerance, due allowances having been made for the uncertainty of the measurement.
Status C The measurement is out of tolerance, due allowances having been made for the uncertainty of the measurement.
Status D No conclusion can be drawn, because the standard(s) do(es) not specify a tolerance for this measurement.
OBSERVATION:
Thickness: (verified against calibrated equipments)
Standard Value
Sl. No. UUC Value (mm) Error (mm) Tolerance (mm) Status Uncertainty (µm)
(mm)
01 0.05 0.05 0.00 N/S D
02 0.10 0.10 0.00 N/S D
03 0.20 0.20 0.00 N/S D
04 0.25 0.25 0.00 N/S D
05 0.30 0.30 0.00 N/S D
06 0.40 0.40 0.00 N/S D
±3.8
07 0.50 0.50 0.00 N/S D
08 0.60 0.60 0.00 N/S D
09 0.70 0.70 0.00 N/S D
10 0.80 0.80 0.00 N/S D
11 0.91 0.90 -0.01 N/S D
12 1.01 1.00 -0.01 N/S D
The overall uncertainty shall be calculated as per ISO “Guide to the Expression of Uncertainty in Measurement” (GUM). The
uncertainty represents an expanded uncertainty using a coverage factor k=2 to approximate a 95% confidence level.
023/(13-06)/4123 Page 4 of 6
Notes:
1. The values mentioned above are the mean readings.
2. No adjustment was done during the calibration.
3. Any section marked,“N/A” means Not Applicable,“N/P” means Not Provided,“N/R” means Not Readable “N/S” means Not Specified.
023/(13-06)/4123 Page 5 of 6
Calculation of Measurment Uncertainty
Vernier Calipers
Type A evaluation
User Input Required
Measurment Record Formula
Sr. No. Measured Reading (mm) Master Instrument = CHECK MASTER
1 0.000 mm Uncertainty =+/- 1.9 (μm)
2 0.000 mm Accu. =+/- 5 (μm)
3 0.000 mm Thermal Expn.of Master = 11.5 µm/mtr/°c
4 0.000 mm
5 0.000 mm Expanded uncertainty :-
6 0.000 mm UE = +/- 12.59418 μm
Avarage = 0.000 mm with coverage factor k=2 for confidence level of 95 % and
Effective Degree of freedom = ∞
Standard Deviation = 0 mm
Degree of freedom = 5