Dimension of EMF
Dimension of EMF
F, V, W Series
Electromagnetic flowmeter
Full-bore flow sensors
Introduction
ABB’s full-bore FEF, FEV and FEW electromagnetic This User Guide should be used in conjunction with the
flowmeter sensors are available with either an following publications:
AquaMaster 3 or a WaterMaster transmitter. – AquaMaster 3 FET200 electromagnetic flowmeter
AquaMaster 3 and WaterMaster are a range of high transmitter User Guide (OI/FET200–EN)
performance electromagnetic flowmeters for the – WaterMaster FET100 electromagnetic flowmeter
measurement of electrically conductive fluids and are transmitter User Guide (OI/FET100–EN)
supplied as factory-configured and calibrated systems.
– WaterMaster FET100 electromagnetic flowmeter /
This User Guide provides end-user details for installation transmitter Hazardous areas ATEX / IECEx areas 2,
and connection. 21 and 22 User Guide (OI/FET100/ATEX–EN)
The Company
We are an established world force in the design and manufacture of instrumentation for industrial process control,
flow measurement, gas and liquid analysis and environmental applications.
As a part of ABB, a world leader in process automation technology, we offer customers application expertise, service
and support worldwide.
We are committed to teamwork, high quality manufacturing, advanced technology and unrivalled service and
support.
The quality, accuracy and performance of the Company’s products result from over 100 years experience, combined
with a continuous program of innovative design and development to incorporate the latest technology.
Quality Control
The UKAS Calibration Laboratory No. 0255 is just one of the ten flow calibration plants operated by the Company
and is indicative of our dedication to quality and accuracy.
0255
UKAS Calibration Laboratory No. 0255
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors
1 Safety ............................................................................................................................................... 2
1.1 Electrical Safety ............................................................................................................................................. 2
1.2 Symbols ........................................................................................................................................................ 2
1.3 Health & Safety ............................................................................................................................................. 3
4 Specification .................................................................................................................................. 18
OI/FEF/FEV/FEW–EN Rev. E 1
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 1 Safety
1 Safety
Information in this manual is intended only to assist our customers in the efficient operation of our
equipment. Use of this manual for any other purpose is specifically prohibited and its contents are not to be
reproduced in full or part without prior approval of the Technical Publications Department.
1.2 Symbols
One or more of the following symbols may appear on the equipment labelling:
Protective earth (ground) terminal Both direct and alternating current supply
2 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 1 Safety
Safety advice concerning the use of the equipment described in this manual or any relevant hazard
data sheets (where applicable) may be obtained from the Company address on the back cover,
together with servicing and spares information.
Warning.
System configuration must be carried out only by users or personnel with approved access rights
(user privileges).
Read all relevant sections of this guide before configuring the system or modifying system
parameters.
Install and use associated equipment in accordance with the relevant national and local
standards.
OI/FEF/FEV/FEW–EN Rev. E 3
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 2 Mechanical Installation
2 Mechanical Installation
2.1 Unpacking
Caution.
Inspect the flowmeter for damage before installation. Do NOT install a damaged or faulty
flowmeter.
When lifting the flowmeter, use the lifting hooks provided or sling under the body of the meter.
NEVER lift the flowmeter by the sensor cable terminal connection box – this will damage the
terminal connection box and invalidate the warranty.
Caution. Do NOT install the flowmeter in an area where a spillage of any substance could damage the
flowmeter.
4 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 2 Mechanical Installation
Caution. Do NOT install the flowmeter in a pipeline that may exert excessive movement and twisting
forces on the flowmeter, for example, vibration.
Caution. Do NOT subject the flowmeter to localized heat during installation. Remember – the flowmeter
is a measuring instrument.
Note. When installing an integral flowmeter, allow enough room to read the transmitter's display and
data label.
OI/FEF/FEV/FEW–EN Rev. E 5
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 2 Mechanical Installation
70 / 130 °C
(158 / 266 °F)
Max.
– 20 °C
(– 4 °F)
Min.
Caution. Ensure the flowmeter is operated within its specified temperature limits. Use flange seals
made from a material that is compatible with the fluid and fluid temperatures as required.
Min. Min.
Flow Direction
Note. Ensure the flow direction in the pipeline corresponds to the identification plate. The flowmeter
measures the flowrate in both directions. Forward flow is the factory setting. Experience has shown that,
in most installations, a straight upstream section 3 x pipe diameters in length and a straight downstream
section 2 x pipe diameters in length are sufficient. However, wherever possible, straight upstream and
downstream sections 5 x 2 pipe diameters in length are recommended.
Note. For accurate and reliable operation, install the flowmeter sensor in a position where it will be
completely full when in operation. The flowmeter will operate when not full, but indicated readings will not
be accurate.
6 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 2 Mechanical Installation
Supports
Caution. Do NOT install the flowmeter in a manner that will subject it to mechanical tension (torsion,
bending). If required, support the pipeline. Remember – the flowmeter is a measuring instrument.
Caution. Do NOT install the flowmeter in a position that will expose it to direct sunlight. Provide
appropriate sun protection if required.
IP68 (NEMA 6)
Caution. Ensure the level of IP rating is correct for the required meter installation application. IP68
(NEMA 6) rated meters can be continuously submerged.
OI/FEF/FEV/FEW–EN Rev. E 7
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 2 Mechanical Installation
Caution. When burying a flowmeter, ensure it can be found easily if required (for example, mark the
installation position with a post). Installing a protection plate above the meter is also recommended.
Note. For further advice when burying flowmeters, contact the ABB Service Organization.
Caution. Install all flowmeter cabling neatly. Installation within a conduit is recommended. Install the
cabling or conduit with a U-bend below the terminal connection box height to prevent water ingress into
the flowmeter sensor by capillary action.
8 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 2 Mechanical Installation
Fit Gaskets
Caution. Use correctly sized gaskets. Do NOT fit gaskets that will extend into the flow area – the
turbulence caused will adversely affect flowmeter accuracy.
Caution. If flowmeters are installed in adjacent pipelines, ensure they are at least 0.7 m (2.3 ft.) apart to
prevent the electromagnetic field from one meter affecting the other.
OI/FEF/FEV/FEW–EN Rev. E 9
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 3 Electrical Installation
3 Electrical Installation
3.1 Grounding
Caution. For safety reasons and optimum performance, the flowmeter, pipelines and medium must be
correctly bonded and grounded according to regulations.
Do not ground cathodically-protected pipelines to an external earth.
Note.
The flow sensor must not be connected to a ground spike.
For bonding connections use ³ 4mm2 (< 10AWG) cable.
AquaMaster 3 Transmitter 2
Insulating Pipe 2
10 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 3 Electrical Installation
OI/FEF/FEV/FEW–EN Rev. E 11
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 3 Electrical Installation
12 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 3 Electrical Installation
OI/FEF/FEV/FEW–EN Rev. E 13
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 3 Electrical Installation
Power Supply
14 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 3 Electrical Installation
To prepare the cable for connection at the transmitter and sensor terminal blocks:
1. Remove the outer cable insulation and Mylar® wrap.
2. Ensure the drain wire is sleeved.
3. Cut the cable connection wires to the lengths shown.
Caution.
Make connections only as shown.
Twist the screen wire of D1 / TFE + D2 with the outer screen drain wire and sleeve them green /
yellow.
Ensure the seal and mating surfaces are clean to maintain environmental rating.
Conduit connections must provide cable entry sealing.
Ensure cable glands are tightened after wiring. Do not overtighten the plastic cable glands to
avoid destroying their sealing properties. Initially, tighten finger-tight, then a further 1/2 to 3/4 turn
using a suitable spanner or wrench.
S1 Violet (Screen)
E1 Violet (*Signal)
E2 Blue (*Signal)
S2 Blue (Screen)
3 Green (Sleeve) Cut cables to
D2 Yellow 60 mm (2.35 in)
D1/TFE Orange
PE (Screen)
M2 Red
Screen to M1 Brown
Internal Earth
for NPT Variants **Drain Wire (Twisted with Screen Wire
from D1/TFE – Orange and D2 – Yellow)
for M20 Versions connect to PE
*Inner Wire
**For Cathodically Protected Systems connect the drain wire to terminal PE.
OI/FEF/FEV/FEW–EN Rev. E 15
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 3 Electrical Installation
Caution.
Twist the three screen wires together and sleeve them.
Keep cable pairs twisted.
Make connections only as shown.
Maintain Environmental Protection at all times.
Conduit connections must provide cable entry sealing.
7 Violet
6 Blue
5 Green
Cut cables to
4 Yellow
60 mm (2.35 in)
3 Orange
2 Red
16 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 3 Electrical Installation
Warning.
Potting materials are toxic. Read the manufacturers' instructions carefully before preparing the
potting material and use suitable safety precautions.
Power up and check all connections before potting.
The remote sensor terminal box connections must be potted immediately on completion to
prevent the ingress of moisture.
Do not overfill or allow the potting material to come into contact with 'O' rings or grooves.
Do not let potting material enter conduit (if used).
OI/FEF/FEV/FEW–EN Rev. E 17
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 4 Specification
4 Specification
WaterMaster optimized full-bore meter / full-bore meters flow performance (m3/h)
Standard Calibration – 0.4 % Class 2 High Accuracy Calibration – 0.2 % Class 1
DN Q4 Q3 Q0.4% Q2 Q1 Q0.2% Q2 Q1
10 3.1 2.5 0.167 0.013 0.008 0.31 0.02 0.012
15 7.88 6.3 0.42 0.032 0.02 0.79 0.05 0.03
20 12.5 10 0.67 0.05 0.032 1.25 0.08 0.05
25 20 16 1.1 0.08 0.05 2 0.13 0.08
32 31.25 25 1.67 0.13 0.08 3 0.20 0.13
40* 50 40 4.2 0.2 0.13 6 0.32 0.2
50* 79 63 4.2 0.32 0.20 7.9 0.5 0.32
65* 125 100 6.7 0.5 0.32 12.5 0.8 0.5
80* 200 160 10.7 0.81 0.51 16 1.3 0.8
100* 313 250 16.7 1.3 0.79 25 2 1.25
125* 313 250 16.7 1.3 0.79 25 2 1.25
150* 788 630 42 3.2 2.0 63 5 3.2
200* 1,250 1,000 67 5.1 3.2 100 8 5
250 2,000 1,600 107 8.1 5.1 160 13 8
300 3,125 2,500 167 12.7 7.9 250 20 12.5
350 5,000 4,000 267 20.3 12.7 400 32 20
400 5,000 4,000 267 20.3 12.7 400 32 20
450 7,875 6,300 420 32 20 630 50 32
500 7,875 6,300 420 32 20 630 50 32
600 12,500 10,000 667 51 32 1000 80 50
700 20,000 16,000 1600 102 64 1600 160 100
750 20,000 16,000 1600 102 64 1600 160 100
30 in (760) 20,000 16,000 1600 102 64 1600 160 100
800 20,000 16,000 1600 102 64 1600 160 100
900 31,250 25,000 2500 160 100 2500 250 156
1000 31,250 25,000 2500 160 100 2500 250 156
42 in 31,250 25,000 2500 160 100 2500 250 156
1100 31,250 25,000 2500 160 100 2500 250 156
1200 50,000 40,000 4000 256 160 4000 400 250
1350 78,750 63,000 6300 403 252 6300 630 394
1400 78,750 63,000 6300 403 252 6300 630 394
1500 78,750 63,000 6300 403 252 6300 630 394
60 in (1500) 78,750 63,000 6300 403 252 6300 630 394
1600 78,750 63,000 6300 403 252 6300 630 394
1650 78,750 63,000 6300 403 252 6300 630 394
1800 125,000 100,000 10000 640 400 10000 1000 625
1950 125,000 100,000 10000 640 400 10000 1000 625
2000 125,000 100,000 10000 640 400 10000 1000 625
2200 200,000 160,000 16000 1024 640 16000 1600 1000
2400 200,000 160,000 16000 1024 640 16000 1600 1000
* OIML R49 Certificate of Conformance to Class 1 and Class 2, with OIML R49 and MID versions available.
Note. OIML R49–1 allows Class 1 only for meters with Q3 ³ 100 m3/h. Meters outside this range have been tested and
conform to Class 1.
18 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 4 Specification
OI/FEF/FEV/FEW–EN Rev. E 19
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 4 Specification
Size Q4 Q3 Q(0.5%) Q2 Q1 Q2 Q1
m3 / h m3 / h m3 / h m3 / h m3 / h R m3 / h m3 / h R
mm in.
(GPM) (GPM) (GPM) (GPM) (GPM) (GPM) (GPM)
50 40 4 0.20 0.13 0.32 0.20
40 11/2 315 200
(220) (176) (18) (0.88) (0.57) (1.40) (0.88)
79 63 6.3 0.32 0.20 0.50 0.32
50 2 315 200
(247) (277) (28) (1.41) (0.88) (2.20) (1.41)
125 100 10 0.50 0.32 0.80 0.50
65 21/2 315 200
(550) (440) (44) (2.20) (1.41) (3.52) (2.20)
200 160 16 0.81 0.51 1.30 0.8
80 3 315 200
(880) (700) (70) (3.56) (2.24) (5.72) (3.52)
313 250 25 1.3 0.79 2.00 1.25
100 4 315 200
(1377) (1100) (110) (5.72) (3.47) (8.80) (5.5)
313 250 25 1.3 0.79 2.00 1.25
125 5 315 200
(1377) (1100) (110) (5.72) (3.47) (8.80) (5.5)
788 630 63 3.2 2.00 5.00 3.2
150 6 315 200
(3470) (2770) (277) (14.08) (8.80) (22.00) (14.08)
1250 1000 100 5.1 3.2 8.0 5.0
200 8 315 200
(5500) (4400) (440) (22.44) (14.08) (35.20) (22.00)
2000 1600 160 8.1 5.1 13.0 8.0
250 10 315 200
(8800) (7040) (700) (35.64) (22.44) (57.20) (35.20)
250 12.7 7.9 20.0 12.5
300 12 3125 (13750) 2500 (11007) 315 200
(1100) (55.88) (34.76) (88.00) (55.00)
400 20.3 12.7 32.0 20.0
350 14 5000 (22000) 4000 (17610) 315 200
(1760) (89.32) (55.88) (14.08) (88.00)
400 20.3 12.7 32.0 20.0
400 16 5000 (22000) 4000 (17610) 315 200
(1760) (89.32) (55.88) (14.08) (88.00)
630 32.0 20.0 50.0 32.0
450 18 7875 (34760) 6300 (27740) 315 200
(2770) (140.8) (88.00) (220.00) (140.8)
630 32.0 20.0 50.0 32.0
500 20 7875 (34650) 6300 (27740) 315 200
(2770) (140.8) (88.00) (220.00) (140.8)
1000 51.0 32.0 80.0 50.0
600 24 12500 (55000) 10000 (44000) 315 200
(4400) (224.4) (140.8) (352.0) (220.0)
20 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 4 Specification
Size Q4 Q3 Q(0.25%) Q2 Q1 Q2 Q1
m3 / h m3 / h m3 / h m3 / h m3 / h R m3 / h m3 / h R
mm in.
(GPM) (GPM) (GPM) (GPM) (GPM) (GPM) (GPM)
50 40 2.7 0.4 0.25 0.64 0.4
40 11/2 160 100
(220) (176) (11.8) (1.8) (1.1) (2.8) (1.8)
79 63 4.2 0.63 0.39 1.0 0.63
50 2 160 100
(247) (277) (18.5) (2.8) (1.71) (4.4) (2.8)
125 100 6.7 1.0 0.6 1.6 1.0
65 21/2 160 100
(550) (440) (29.5) (4.4) (1.41) (7.0) (4.4)
200 160 10.7 1.6 1.0 2.6 1.6
80 3 160 100
(880) (700) (47.0) (7.0) (4.4) (11.4) (7.0)
313 250 16.7 2.5 1.6 4.0 2.5
100 4 160 100
(1377) (1100) (73.5) (11.0) (7.0) (17.6) (11.0)
313 250 16.7 2.5 1.6 4.0 2.5
125 5 160 100
(1377) (1100) (73.5) (11.0) (7.0) (17.6) (11.0)
788 630 42.0 6.3 3.9 10.0 6.3
150 6 160 100
(3470) (2770) (184.8) (27.7) (17.1) (44.0) (27.7)
1250 1000 67.0 10.0 6.0 16.0 10.0
200 8 160 100
(5500) (4400) (294.8) (44.0) (26.4) (70.0) (44.0)
2000 1600 107.0 16.0 10.0 26.0 16.0
250 10 160 100
(8800) (7040) (470.8) (70.0) (44.0) (110.4) (70.0)
167.0 25.0 15.6 40.0 25.0
300 12 3125 (13750) 2500 (11007) 160 100
(734.8) (110.0) (68.6) (176.0) (110.0)
267.0 40.0 25.0 64.0 40.0
350 14 5000 (22000) 4000 (17610) 160 100
(1174.8) (176.0) (110.0) (281.6) (176.0)
267.0 40.0 25.0 64.0 40.0
400 16 5000 (22000) 4000 (17610) 160 100
(1174.8) (176.0) (110.0) (281.6) (176.0)
420.0 63.0 39.0 101.0 63.0
450 18 7875 (34760) 6300 (27740) 160 100
(184.8) (277.0) (171.6) (444.4) (277.0)
420.0 63.0 39.0 101.0 63.0
500 20 7875 (34650) 6300 (27740) 160 100
(184.8) (277.0) (171.6) (444.4) (277.0)
667.0 100.0 63.0 160.0 100.0
600 24 12500 (55000) 10000 (44000) 160 100
(2934.8) (440.0) (277.0) (704.0) (440.0)
OI/FEF/FEV/FEW–EN Rev. E 21
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 4 Specification
Functional Specification
Pressure limitations
As per flange rating – non approved
PN16 for OIML R49, MID Approved
Pressure equipment directive 97/23/EC
This product is applicable in networks for the supply, distribution and discharge of water and associated equipment and
is therefore exempt.
Temperature limitations
Ambient temperature
Remote transmitter –20 to 70 °C (–4 to 158 °F)
Integral transmitter –20 to 60 °C (–4 to 140 °F)
Medium temperature
°C (°F)
Code Lining Flange material Minimum Maximum
Carbon steel –10 (14) 90 (194)
FEF, FEW3 Hard rubber
Stainless steel –10 (14) 90 (194)
Carbon steel –10 (14) 130 (266)
FEW1 PTFE
Stainless steel –25 (–13) 130 (266)
Carbon steel –10 (14) 130 (266)
FEW3 PTFE
Stainless steel –10 (14) 130 (266)
Carbon steel –5 (23) 80 (176)
FEW3 Elastomer
Stainless steel –5 (23) 80 (176)
FEF Elastomer –6 (21) 70 (158)
Carbon steel
FEV Polypropylene –6 (21) 70 (158)
IP rating
IP68 (NEMA 6) to 7 m (20 ft.) depth
Note. Not sizes DN10 to DN32 (3/8 – 11/4 in. NB)
IP67 (NEMA 4X) – DN10 to DN32 (3/8 – 11/4 in. NB)
Buriable (sensor only)
FEV, FEF and FEW – DN450 to 2400 (18 to 96 in. NB) to 5 m (16 ft.) depth
Conductivity
>5µS cm–1
Transmitter mounting
Integral (not FEF) or remote
Electrical connections
20 mm glands
1
/2 in. NPT
20 mm armored glands
Sensor cable
ABB WaterMaster cable available in two forms – standard and armored
Maximum length 200 m (660 ft.)
22 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 4 Specification
Physical Specification
Wetted parts
Electrode material
Stainless steel 316 L / 316 Ti
Super-austenitic steel
Hastelloy® C-22 and Hastelloy C4
(other electrode materials available on request)
Potential equalizing rings
Minimum of 1 recommended
Lining material / potable water approvals
Pressure loss
OI/FEF/FEV/FEW–EN Rev. E 23
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 4 Specification
Non-wetted parts
Flange material
Housing material
DS/WM–EN Rev. T
24 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 5 Sensor dimensions
5 Sensor dimensions
5.1 FEW – DN10 to 125 (3/8 to 5 in. NB)
Dimensions in mm (in.)
77.8 104.5 98
(3.06) (4.11) (3.86)
Ø104
89 (4.1)
(3.5)
(7.91)
210
(3.1)
78
G E F
D
C
A
L
DN10 JIS10K 90 (3.54) 200 (7.87) 268 (10.55) 82 (3.23) 193 (7.6) 148 (5.83) 113 (4.45) 6 (13) 4 (9)
(3/8 in.)
PN10 to 40 90 (3.54)
JIS10K 95 (3.74)
Table 5.1 DN10 to 125 (3/8 to 5 in. NB) (FEW) dimensions / weights
OI/FEF/FEV/FEW–EN Rev. E 25
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 5 Sensor dimensions
DN25 PN10 to 40 115 (4.53) 200 (7.87) 268 (10.55) 82 (3.23) 193 (7.6) 148 (5.83) 113 (4.45) 9 (20) 7 (15)
(1 in.)
JIS5K 95 (3.74)
DN32 PN10 to 40 140 (5.51) 275 (10.83) 92 (3.62) 200 (7.87) 155 (6.10) 113 (4.45) 10 (22) 8 (18)
(11/4 in.)
JIS5K 115 (4.53)
DN50 PN10 to 40 165 (6.5) 281 (11.06) 97 (3.82) 206 (8.11) 161 (6.34) 115 (4.53) 12 (26) 10 (22)
(2 in.)
JIS5K 130 (5.12)
DN65 PN10 to 40 185 (7.28) 292 (11.50) 108 (4.25) 217 (8.54) 172 (6.77) 104 (4.09) 13 (29) 11 (24)
(21/2 in.)
JIS5K 155 (6.10)
DN80 PN10 to 40 200 (7.87) 292 (11.5) 108 (4.25) 217 (8.54) 172 (6.77) 104 (4.09) 17 (37) 15 (33)
(3 in.)
JIS5K 180 (7.09)
Table 5.1 DN10 to 125 (3/8 to 5 in. NB) (FEW) dimensions / weights (Continued)
26 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 5 Sensor dimensions
DN100 PN10 to 16 220 (8.66) 250 (9.84) 314 (12.36) 122 (4.8) 239 (9.41) 194 (7.64) 125 (4.92) 19 (42) 17 (37)
(4 in.)
PN25 to 40 235 (9.25) 23 (51) 21 (46)
DN125 PN10 to 16 250 (9.84) 324 (12.76) 130 (5.12) 249 (9.8) 204 (8.03) 125 (4.92) 22 (48) 20 (44)
(5 in.)
PN25 to 40 270 (10.63) 29 (64) 27 (59)
Table 5.1 DN10 to 125 (3/8 to 5 in. NB) (FEW) dimensions / weights (Continued)
OI/FEF/FEV/FEW–EN Rev. E 27
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 5 Sensor dimensions
Dimensions in mm (in.)
104.5 98 77.8
(4.11) (3.86) (3.06)
Ø104 89
(4.1) (3.5)
(7.91)
201
(3.1)
78
E F
G
DN150 PN10 to 16 285 (11.22) 300 (11.81) 371 (14.61) 146 (9.88) 296 (11.65) 251 (9.88) 166 (6.54) 33 (73) 31 (68)
(6 in.)
PN25 to 40 300 (11.81) 39 (86) 37 (81)
DN200 PN10 340 (13.39) 350 (13.78) 411 (16.18) 170 (6.69) 336 (13.23) 291 (11.46) 200 (7.87) 41 (90) 39 (86)
(8 in.)
PN16 340 (13.39)
28 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 5 Sensor dimensions
DN250 PN10 395 (15.55) 450 (17.72) 426 (16.77) 198 (7.80) 351 (13.82) 306 (12.05) 235 (9.62) 61 (134) 59 (130)
(10 in.)
PN16 405 (15.94) 65 (143) 63 (139)
DN300 PN10 445 (17.52) 500 (19.69) 449 (17.68) 228 (8.98) 374 (14.72) 329 (12.95) 272 (10.71) 74 (163) 72 (158)
(12 in.)
PN16 460 (18.11) 80 (176) 78 (172)
DN350 PN10 505 (19.88) 550 (21.65) 464 (18.27) 265 (10.43) 389 (15.31) 344 (13.54) 322 (12.68) 95 (209) 93 (205)
(14 in.)
PN16 520 (20.47) 110 (242) 108 (238)
DN400 PN10 565 (22.24) 600 (23.62) 506 (19.92) 265 (10.43) 431 (16.97) 386 (15.20) 322 (12.68) 103 (227) 101 (222)
(16 in.)
PN16 580 (22.83) 126 (277) 124 (273)
Table 5.2 DN150 to 400 (6 to 5 in. NB) (FEW) dimensions / weights (Continued)
OI/FEF/FEV/FEW–EN Rev. E 29
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 5 Sensor dimensions
Dimensions in mm (in.)
77.8 104.5 98
(3.06) (4.11) (3.86)
104
89
(4.1)
(3.5)
(7.91)
201
(3.1)
78
E F
G
A
L (1.0D)
L (1.3D)
DN450 PN10 615 (24.21) N/A 600 514 (20.24) 310 (12.20) 439 (17.28) 394 (15.51) 328 (12.91) 173 (381) 171 (377)
(18 in.) (23.62)
PN16 640 (25.20) 188 (414) 186 (410)
Table 5.3 DN450 to 2400 (18 to 96 in. NB) (FEW) dimensions / weights
30 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 5 Sensor dimensions
DN500 PN10 670 (26.38) N/A 600 514 (20.24) 310 (12.20) 439 (17.28) 394 (15.51) 367 (14.45) 190 (418) 188 (413)
(20 in.) (23.62)
PN16 715 (28.15) 240 (528) 238 (524)
ASME B16.5 CL300 775 (30.51) N/A 762 490 (1080) 488 (1076)
(30.00)
DN600 PN10 780 (30.71) N/A 800 565 (22.24) 361 (14.21) 490 (19.29) 445 (17.52) 469 (18.46) 284 (626) 282 (622)
(24 in.) (31.50)
PN16 840 (33.07) 318 (700) 316 (695)
DN700 JIS 5K 875 (34.45) 700 910 604 (23.77) 403 (15.87) 528 (20.79) 488 (19.21) 444 (17.48) 216 (475) 214 (471)
(28 in.) (35.83)
JIS 10K 905 (35.63) (27.56) 282 (620) 280 (616)
Table 5.3 DN450 to 2400 (18 to 96 in. NB) (FEW) dimensions / weights (Continued)
OI/FEF/FEV/FEW–EN Rev. E 31
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 5 Sensor dimensions
DN750 JIS 5K 945 (37.20) 750 990 630 (24.79) 429 (16.89) 554 (21.81) 514 (20.23) 444 (17.48) 251 (552) 249 (548)
(30 in.) (29.52) (38.98)
JIS 10K 970 (38.19) 327 (719) 325 (715)
DN800 JIS 5K 995 (39.17) 800 1040 654 (25.74) 453 (17.83) 578 (22.76) 538 (21.18) 542 (21.34) 280 (616) 278 (612)
(32 in.) (40.04)
JIS 10K 1020 (40.16) (31.49) 364 (801) 362 (796)
Table 5.3 DN450 to 2400 (18 to 96 in. NB) (FEW) dimensions / weights (Continued)
32 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 5 Sensor dimensions
DN900 JIS 5K 1095 (43.11) 900 1170 705 (27.7() 504 (19.84) 629 (24.76) 589 (23.19) 570 (22.44) 369 812) 367 (807)
(36 in.) (46.06)
JIS 10K 1120 (44.09) (35.43) 445 (979) 443 (975)
DN1000 JIS 5K 1195 (47.05) 1000 1300 755 (29.71) 554 (21.81) 679 (26.73) 639 (25.16) 624 (24.57) 441 (970) 439 (966)
(40 in.) (39.37) (51.18)
JIS 10K 1235 (48.62) 572 (1258) 570 (1254)
Table 5.3 DN450 to 2400 (18 to 96 in. NB) (FEW) dimensions / weights (Continued)
OI/FEF/FEV/FEW–EN Rev. E 33
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 5 Sensor dimensions
DN1050 AWWA C207 CLASS B 1346 (5299) 1050 1365 808 (31.82) 608 (23.92) 733 (28.84) 693 (27.28) 624 (24.57) 564 (1241) 562 (1236)
(42 in.) (41.33) (53.74)
AWWA C207 CLASS D 1346 (5299) 669 (1472) 667 (1467)
DN1100 JIS 5K 1305 (51.38) 1100 1430 510 (1122) 508 (1118)
(44 in.) (43.30) (56.30)
JIS 10K 1345 (52.95) 689 (1516) 687 (1511)
DN1200 JIS 5K 1420 (55.91) 1200 1560 860 (33.85) 659 (25.94) 784 (30.87) 744 (29.29) 802 (31.57) 651 (1432) 649 (1428)
(48 in.) (47.24) (61.42)
JIS 10K 1465 (57.68) 967 (2127) 965 (2123)
DN1350 AWWA C207 CLASS B 1683 (66.26) 1350 1755 955 (37.59) 754 (29.69) 879 (34.61) 839 (33.03) 902 (35.51 981 (2158) 979 (2154)
(54 in.) (53.15) (69.09)
AWWA C207 CLASS D 1683 (66.26) 1213 (2669) 1211 (2664)
DN1400 PN6 1630 (64.17) 1400 1820 1085 (2387) 1083 (2383)
(56 in.) (55.11) (71.65)
PN10 1675 (65.94) 1731 (3808) 1729 (3804)
Table 5.3 DN450 to 2400 (18 to 96 in. NB) (FEW) dimensions / weights (Continued)
34 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 5 Sensor dimensions
DN1500 JIS 5K 1730 (68.11) 1500 1950 1065 (41.92) 864 (34.02) 989 (38.94) 949 (37.36) 910 (35.83) 1029 (2264) 1027 (2259)
(60 in.) (59.05) (76.77)
JIS 10K 1795 (70.67) 1504 (3309) 1502 (3304)
DN1600 PN6 1830 (72.05) 1600 2080 1066 (41.96) 865 (34.06) 990 (38.98) 950 (37.4) 1000 (39.37) 1434 (3155) 1432 (3150)
(64 in.) (62.99) (81.89)
PN10 1915 (75.39) 2525 (5555) 2523 (5551)
DN1650 AWWA C207 CLASS B 2032 (80.00) N/A 2145 1116 (43.94) 915 (36.02) 1040 (40.94) 1000 (39.37) 1000 (39.37) 1504 (3309) 1502 (3304)
(66 in.) (84.45)
AWWA C207 CLASS D 2032 (80.00) 2025 (4455) 2023 (4451)
DN1800 PN6 2045 (80.51) N/A 2340 1181 (46.50) 980 (38.58) 1105 (43.50) 1065 (41.93) 1100 (43.31) 1853 (4077) 1851 (4072)
(72 in.) (92.13)
PN10 2115 (83.27) 3180 (6996) 3178 (6992)
DN1950 AWWA C207 CLASS B 2362 (92.99) N/A 2535 1291 (50.81) 1090 (42.91) 1215 (47.83) 1175 (46.26) 1180 (46.46) 2309 (5080) 2307 (5075)
(78 in.) (99.80)
AWWA C207 CLASS D 2362 (92.99) 3037 (6681) 3035 (6677)
DN2000 PN6 2265 (89.17) N/A 2600 2581 (5678) 2579 (5674)
(80 in.) (102.36)
PN10 2325 (91.54) 4254 (9359) 4252 (9354)
DN2100 AWWA C207 CLASS B 2534 (99.76) N/A 2730 1395 (54.91) 1194 (47.01) 1319 (51.93) 1279 (50.35) 1180 (46.46) 2641 (5810) 2639 (5806)
(84 in.) (107.48)
AWWA C207 CLASS D 2534 (99.76) 3487 (7671) 3485 (7667)
DN2200 PN6 2475 (97.44) N/A 2860 1330 (52.36) 3363 (7399) 3361 (7394)
(88 in.) (112.60)
PN10 2550 (100.39) 5795 5793
(12749) (12745)
DN2400 PN6 2685 (105.71 N/A 3120 1495 (58.85) 1294 (50.94) 1419 (55.87) 1379 (54.29) 1450 (57.09) 4100 (9020) 4098 (9016)
(96 in.) (122.83)
PN10 2760 (108.66) 6968 6966
(15330) (15325)
Table 5.3 DN450 to 2400 (18 to 96 in. NB) (FEW) dimensions / weights (Continued)
OI/FEF/FEV/FEW–EN Rev. E 35
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 5 Sensor dimensions
Dimensions in mm (in.)
77.8 104. 5 98
(3.06) (4.11) (3.86)
Ø104
(7.91)
89
201
(3.5) (4.1)
(3.1)
78
F
G E
X D
C
X
L
DN40 EN1092-1 PN10, PN40 150 (5.91) 200 (7.87) 260 (10.24) 30.4 (1.20) 185 (7.28) 138 (5.43) 30 (1.18) 12 (27) 11 (24)
(11/2 in.)
ASME B16.5 CLASS 150 127 (5.00)
DN50 EN1092-1 PN10, PN16 165 (6.50) 200 (7.87) 270 (10.63) 38.3 (1.51) 195 (7.68) 146 (5.75) 38 (1.50) 13 (29) 12 (27)
(2 in.)
ASME B16.5 CLASS 150 152.4 (6.00)
DN65 AS4087 PN14, PN16 165 (6.50) 200 (7.87) 275 (10.83) 45.2 (1.78) 200 (7.87) 152 (5.98) 48 (1.89) 15 (33) 14 (31)
(21/2 in.)
AS2129 TABLE C D E 165 (6.50)
1
Table 5.4 DN40 to 200 (1 /2 to 8 in. NB) (FEV) dimensions / weights
36 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 5 Sensor dimensions
DN80 EN1092-1 PN10, PN16 200 (7.87) 200 (7.87) 280 (11.02) 51.5 (2.03) 205 (8.07) 156 (6.14) 61 (2.40) 16 (36) 15 (33)
(3 in.)
ASME B16.5 CLASS 150 190 (7.48)
DN100 EN1092-1 PN10, PN16 220 (8.66) 250 (9.84) 320 (12.60) 63.75 (2.51) 245 (9.65) 196.8 (7.75) 70 (2.76) 19 (42) 18 (40)
(4 in.)
ASME B16.5 CLASS 150 228.6 (9.00)
DN125 EN1092-1 PN10, PN16 250 (9.84) 250 (9.84) 320 (12.60) 63.75 (2.51) 245 (9.65) 197 (7.76) 70 (2.76) 20 (44) 19 (42)
(5 in.)
ASME B16.5 CLASS 150 254 (10.00)
DN150 EN1092 PN10, PN16 285 (11.22) 300 (11.81) 340 (13.39) 84.4 (3.32) 265 (10.43) 217 (8.54) 103 (4.06) 32 (70) 31 (68)
(6 in.)
ASME B16.5 CLASS 150 279 (10.98)
DN200 EN1092-1 PN10 340 (13.39) 350 (13.78) 365 (14.37) 109.8 (4.32) 290 (11.42) 243 (9.57) 150 (5.91) 49 (108) 48 (105)
(8 in.)
EN1092-1 PN16 340 (13.39)
1
Table 5.4 DN40 to 200 (1 /2 to 8 in. NB) (FEV) dimensions / weights (Continued)
OI/FEF/FEV/FEW–EN Rev. E 37
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 5 Sensor dimensions
Dimensions in mm (in.)
Ø104
(4.1)
86 (3.4)
*Base
X
A
L
*Dimension C = centre line to base of flowmeter body
Dimensions in mm (in.)
DN250 (10 in.) ASME B16.5 CLASS 150 405 (15.94) 450 (17.72) 215 (8.46) 301 (11.85) 300 (11.81) 250 (9.84) 88 (194)
AS2129 TABLE C D
AS2129 TABLE E
AS2129 TABLE F
Table 5.5 DN250 to 600 (10 to 24 in. NB) (FEF) dimensions / weights
38 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 5 Sensor dimensions
Dimensions in mm (in.)
DN300 (12 in.) ASME B16.5 CLASS 150 485 (19.09) 500 (19.69) 231 (9.09) 317 (12.48) 352 (13.86) 300 (11.81) 128 (282)
DN350 (14 in.) ASME B16.5 CLASS 150 535 (21.06) 550 (21.65) 257.5 (10.14) 346 (13.62) 376 (14.80) 350 (13.78) 100 (220)
DN375 (15 in.) AS4087 PN14, PN16 550 (21.65) 550 (21.65) 257.5 (10.14) 346 (13.62) 376 (14.80) 350 (13.78) 115 (253)
DN400 (16 in.) ASME B16.5 CLASS 150 600 (23.62) 600 (23.62) 285 (11.22) 371 (14.61) 420 (16.54) 400 (15.75) 115 (253)
Table 5.5 DN250 to 600 (10 to 24 in. NB) (FEF) dimensions / weights (Continued)
OI/FEF/FEV/FEW–EN Rev. E 39
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 5 Sensor dimensions
Dimensions in mm (in.)
DN450 (18 in.) ASME B16.5 CLASS 150 635 (25.00) 700 (27.56) 317.5 (12.50) 402 (15.83) 480 (18.90) 450 (17.72) 160 (352)
DN500 (20 in.) ASME B16.5 CLASS 150 700 (27.56) 770 (30.31) 345 (13.58) 429 (16.89) 520 (20.47) 500 (19.69) 217 (455)
DN600 (24 in.) ASME B16.5 CLASS 150 815 (32.09) 920 (36.22) 387.5 (15.25) 472 (18.58) 610 (24.02) 600 (23.62) 315 (693)
Table 5.5 DN250 to 600 (10 to 24 in. NB) (FEF) dimensions / weights (Continued)
40 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeters | Full-bore flow sensors 6 Common accessories
6 Common accessories
Accessory Item Number
OI/FEF/FEV/FEW–EN Rev. E 41
F, V, W Series
Electromagnetic flowmeter
Notes
42 OI/FEF/FEV/FEW–EN Rev. E
F, V, W Series
Electromagnetic flowmeter
OI/FEF/FEV/FEW–EN Rev. E 43
F, V, W Series
Electromagnetic flowmeter
44 OI/FEF/FEV/FEW–EN Rev. E
Products and customer support
Automation Systems Customer support
For the following industries: We provide a comprehensive after sales service via a
— Chemical & Pharmaceutical Worldwide Service Organization. Contact one of the
— Food & Beverage following offices for details on your nearest Service and
— Manufacturing Repair Centre.
— Metals and Minerals
— Oil, Gas & Petrochemical UK
— Pulp and Paper ABB Limited
Tel: +44 (0)1453 826661
Drives and Motors Fax: +44 (0)1453 829671
— AC and 6 Drives, AC and DC Machines, AC Motors to
1kV USA
— Drive Systems ABB Inc.
— Force Measurement Tel: +1 215 674 6000
— Servo Drives Fax: +1 215 674 7183
Controllers & Recorders
— Single and Multi-loop Controllers
— Circular Chart and Strip Chart Recorders
— Paperless Recorders
— Process Indicators
Flexible Automation
— Industrial Robots and Robot Systems
Flow Measurement
— Electromagnetic Flowmeters
— Mass Flowmeters
— Turbine Flowmeters
— Wedge Flow Elements
Process Analytics
— Process Gas Analysis
— Systems Integration
Transmitters
— Pressure
— Temperature
— Level
— Interface Modules
www.abb.com