0% found this document useful (0 votes)
23 views

vc97-manual

The document is an operation manual for a digital multimeter, detailing its features, specifications, and measurement capabilities including DC/AC voltage, current, resistance, capacitance, frequency, temperature, and diode testing. It includes safety warnings, overload protection details, and instructions for use and maintenance. The multimeter is designed for high accuracy and performance, suitable for production lines, laboratories, and maintenance work.

Uploaded by

Ibrahim Khwaira
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
23 views

vc97-manual

The document is an operation manual for a digital multimeter, detailing its features, specifications, and measurement capabilities including DC/AC voltage, current, resistance, capacitance, frequency, temperature, and diode testing. It includes safety warnings, overload protection details, and instructions for use and maintenance. The multimeter is designed for high accuracy and performance, suitable for production lines, laboratories, and maintenance work.

Uploaded by

Ibrahim Khwaira
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Overload protection: 1000DC/750VAC RMS Overload protection: 250V DC/AC peak value

DIGITAL MULTIMETER Indication : mean value response (rms of sine wave) Diode and continuity Test
Resistance Range Description Test Condition
OPERATION MANUAL Range Accuracy Resolution
400Ω ±(0.8%+5d) 0.1Ω Display read approx. forward Forward DC Current
1. INTRODUCTION 4kΩ ±(0.8%+4d) 1Ω voltage of diode approx.0.5mA. Reversed
The instrument is a high performance, high accuracy, 3 3/4 digit, 42mm digit 40kΩ 10Ω voltage approx. 1.5V.
high LCD multi-meter for measuring DC and AC voltage, DC and AC current, 400kΩ 100Ω Buzzer sounds if resistance Open circuit voltage:
Resistance and Capacitance, Frequency, temperature, duty circle, Transistor, Diode 4MΩ 1kΩ Between terminals V/Ωand COM 0.5V
and Continuity test. Design parameter: unit indication, data hold (HOLD), is less than about70±30Ω.
40MΩ ±(1.2%+10d) 10kΩ
relatively measuring (REL), AUTO/MANUAL range, auto off and buzzer sounds, Overload Protection : 250V DC/AC peak value
Overload protection: 250V DC/AC peak value
etc. Warning: do not input voltage at the range for safety.
Open circuit voltage: 400mV
The Dual-slop A/D converter CMOS technology for auto-zero, polarity Transistor hFE Test
Note: at 400Ωrange, should short the test leads and measure the resistance of the wire,
selection and over-range indication. Full overload protection is provided. Because Range Displaying value Test condition
then, minus from measuring.
of its outstanding features, it is most suitable for use on production lines or for lab,
DC Current hFE NPN or 0~1000 Base Current approx.
R & D, maintenance and repair work.
Range Accuracy Resolution PNP 15uA, Vce approx. 1.5V
2.SPECIFICATIONS
400uA 0.1uA
2.1GENERAL SPECIFICATIONS Temperature
4000uA ±(1.0%+10d) 1uA
Display: 3 3/4digit LCD with a max readings of 3999 Range Displaying value Test condition
Measuring method: Dual-slop integrating A/D converter system. (-20-1000)℃ <400℃ ±(1.0%+5d) 1℃
40mA ±(1.2%+8d) 10uA
Sampling rate: Approx. 3times/second. ≥400℃ ±(1.5%+15d)
400mA 100uA
Max. Common Mode Voltage: 500V DC/AC RMS. Thermocouple: K type
Range select method: Automatic and Manual 10A ±(1.2%+10d) 10mA
Warning: do not input voltage at the range for safety.
Polarity: Automatic negative polarity indication. Overload protection: 0.5A/250V fuse, 10A/250V fuse. 3. FRONT PANEL DESCRIPTION
Over range indication: Only the MSD “ OL” display. Max measuring voltage: Full scale mA range: 1.2V; A range : 100 mV. 1.LCD
Low battery : The “ ” displays. Max input current: 10A (max. up to 15 seconds). 2.Function key
Safety standards : EMC/LVD. The meter is up to the standards of IEC1010 AC mV 2-1.HOLD Key
Pollution Degree 2, Over voltage categoryⅡor double insulationⅡ. Range Accuracy Resolution 2-2.REL Key
Operating environment: Temperature (0 ~ 40)℃, humidity<80%RH. 400mV ±(1.6%+8d) 0.1mV 2-3.Hz/DUTY Key
Storage environment: Temperature (-10~50)℃, humidity<80%RH. 2-4 “select” key
Power: Double, standard 1.5 volt battery. AAA 7# battery. AC Current 2-5.RANGE Key
Dimension: 190mm (H)×93.5mm (W) ×37mm (D). Range Accuracy Resolution 3.Rotary Switch
Weight: Approx.420g(including battery). 400uA ±(1.5%+10d) 0.1uA 4.Voltage, Resistance, Frequency socket.
2.2 ELECTRICAL SPECIFICATIONS 4000uA 1uA 5.COM input jack terminal
Accuracy is ±(percentage of reading + number of digit) at(23±5)℃,<75%RH. 40mA 10uA 6.Less than 400mA input jack terminal
DC Voltage 400mA 100uA 7.10A input jack terminal
Range Accuracy Resolution 10A ±(2.0+15d) 10mA 4.OPERATION
400mV ±(0.5%+4d) 0.1mV Overload protection: 0.5A/250V fuse, 10A/250V fuse 1. Check the 1.5-volt battery by setting the ROTARY Switch removes to OFF position.
4V 1mV Max measuring voltage: Full scale mA range: 1.2V A range: 100mV If the battery is weak, a sign will appear on the display. If this does not appear on
40V 10mV Max input current: 10A (up to 15 seconds).
400V 100mV Frequency response: 10A range: 40-100Hz; other range: 40-400Hz the display, proceed as below. See MAINTENANCE if the battery has to be replaced.
1000V ±(1.0%+6d) 1V Capacitance 2. The mark, or sign next to the lead jacks, is for warning that the input voltage or
Input impedance: 400mV range: more than 40MΩ other range: 10MΩ Range Accuracy Resolution current should not exceed the indicated values.
Overload protection: 1000V DC or AC peak value 4nF ±(2.5%+20d) 1pF This is to prevent from damaging the internal circuits.
DC mV: 40nF ±(3.5%+8d) 10pF 3. The function switch should be set to the range that you want to test before operation.
400nF 100pF 4.1 DC Voltage measurement
Range Accuracy Resolution 1) Insert the BLACK test lead to “COM” jack and RED test lead to the “V/Ω/Hz”
4uF 1nF
jack.
400mV ±(0.5%+4d) 0.1mV 40uF 10nF
2) Set the FUNCTION switch to “ ” range.
200uF ±(5.0%+10d) 100nF
3) The default range is Auto range, and "AUTO" is displayed. Pressing RANGE key
AC Voltage Overload protection :250V DC/AC peak value
switch to manual range, 400mV/4V/40V/400V/1000V can be selected.
Range Accuracy Resolution Frequency
4) Connect the test leads to the tested point, the voltage and polarity which
Range Accuracy Resolution connected with the red lead will appear on LCD.
4V ±(0.8%+10d) 1mV 100Hz ±(0.5%+10d) 0.01Hz Note:
40V 10mV 1000Hz 0.1Hz 1) Manual range, if LCD display "OL", over range is being indicated and the
10kHz 1Hz FUNCTION switch must be set to a higher range.
400V 100mV 100kHz 10Hz 2) Do not measure over 1000V, or, the meter will be damaged.
750V ±(1.0%+10d) 1V 1MHz 100Hz 3)Caution to avoid contact with high voltage circuits when measuring high voltage.
Input impedance: 400mV range: more than 40MΩ; other range: 10MΩ 30MHz 1kHz 4.2. AC voltage measurement
Frequency response: 750V range : 40-100Hz other range: 40-400Hz Input sensitivity: 1.0V 1) Insert the black test lead to “COM” jack and the red one to “V/Ω/Hz” jack.
2) Set the function switch to “ ” range. 4.7 Frequency measurement 6) Always ensure that the correct function and range is selected. If in doubt about
3) The default range is Auto range, and "AUTO" is displayed. Pressing RANGE 1) Connect test leads or shield cable to "COM" and "V/Ω/Hz " jack. the correct range to use, start with the highest and work downwards.
key switch to manual range, 400mV/4V/40V/400V/700V can be selected. 2) Set the FUNCTION switch to the "Hz" range, and connect test leads or cable 7) Extreme care should be taken when using the instrument to conjunction with a
4) Connect the test leads to the test point, the voltage of the two points which across the source load under test. current transformer connected to the terminals if an open circuit occurs.
connected with the leads will appear on LCD. 3) Press "Hz/Duty" to switch frequency/duty cycle, and display the reading of 8) Ensure that the test leads and probes are good condition with no damage to the
4.3 DC Current measurement frequency or duty cycle. insulation.
1) Insert the BLACK test lead to “COM” jack and RED test lead to the Note: 9) Take care not to exceed the over-load limits as given in the specification.
“mA”(max. 400mA) or "10A" jack (max. 10A). 1) Frequency rang have only auto range mode. 10) FUSE FOR REPLACEMENT MUST BE OF THE CORRECT TYPE AND
2) Set the FUNCTION switch to current range. Pressing “ ” key to select 2) Do not apply more than 250V DC/AC peak value to the input. Indication is RATING.
DC measure method, connect the leads across to the tested circuit, the current possible at voltage higher than 10V AC rms, but readings may be out of 11) Before opening the case of the instrument to replace battery or fuse, disconnect
value and polarity the red lead connect with will appear on LCD. specification. the test leads from any external circuit, set the selector switch to “OFF”
Note: 3) In noisy environment, it is preferable to use shield cable for measuring small position.
1) If the current range is unknown beforehand, set the FUNCTION switch to a signal. 6. MAINTENANCE
high range and work down. 4) Be caution to avoid contact with high tension circuits when measurement high Do not try to modify the inner circuit.
2) When only the figure "OL" is displayed over range is being indicated and the voltage. 1) Keep the multimeter dry. Keep the multi-meter away from dust and dirt
FUNCTION switch must be set to a higher range. 4.8 hFE measurement 2) Use and store the multi-meter only in normal temperature environments.
3) The max input current is 400mA,or 10A depending upon the jack used. 1) Set the function switch to hFE range. Temperature extremes can shorten the life of electronic devices, damage
Excessive current will blow the fuse. 2) Define the transistor is NPN or PNP type, insert the emitter, base and collector batteries, and distort or melt plastic part.
4.4. AC current measurement separately to the correct hole, the approx. value will be displayed on LCD. 3) Handle the multimeter gently and carefully. Dropping it can damage the circuit
1) Insert the BLACK test lead to “COM” jack and RED test lead to the mA”(max. 4.9 Diode and continuity Test boards and case and can cause the multi-meter to work improperly although the
400mA) or "10A" jack (max. 10A). 1) Connect the BLACK test lead to “COM” jack and RED test lead to the “VΩHz” holster can provide enough protection.
2) Set the FUNCTION switch to current range. Pressing“ ” key to select jack.. 4) Wipe the multi-meter with a damp cloth occasionally to keep it looking new. Do
AC measure method, connect the leads across to the tested circuit, the current 2) Set the FUNCTION switch to “ ” position. Press " DC/AC " key which not use harsh chemicals, cleaning solvents, or strong detergents to clean the
value will appear on LCD. select diode measure method. multi-meter.
Note: 3) Forward measure: Connect RED test lead to the positive of the test diode, 5) Take out off the battery if do not use for a long time. When LCD displays “ ”,
1) If the current range is unknown beforehand, set the FUNCTION switch to a BLACK test lead to the negative, then, reading of approx. forward voltage of the battery should be replaced.
high range and work down. this diode displays. a. Ensure the instrument is not connected to any external circuit. Set the selector
2) When only the figure "OL" is displayed over range is being indicated and the 4) Reverse measure: Connect BLACK test lead to the positive of the test diode, switch to
FUNCTION switch must be set to a higher range. RED test lead to the negative, the mark "OL" will be displayed. OFF position and remove the test leads from terminals.
3) The max input current is 400mA,or 10A depending upon the jack used. 5) Proper diode testing should include both steps. b. Remove the screw on the bottom case and lift the bottom case.
Excessive current will blow the fuse. 6) Press " " to select continuity mode. c. Remove the spent battery and replace it with a battery of the same type.
4.5 Resistance measurement 7) Connect the test probes to two points of circuit, if the resistance is lower than 6) Replace the fuse with same type and rating as the replacements.
1) Connect the BLACK test lead to “COM” jack and RED test lead to the approx. 50±20Ω. Buzzer sounds. NOTE:
“V/Ω/Hz” jack. Note: Do not input voltage at this range. 1) Do not input a voltage over 1000V DC/AC peak value.
2) Set the FUNCTION Switch to “Ω” range. 4.10 Temperature measurement 2) Do not measure voltage at current range, resistance range, diode and buzzer
3) Press "RANGE" to select Auto/Manual range 1. Set the function key to “℃”range. range.
4) If measuring small resistance, should short test leads first, press "REL" once, 2. Insert the cold-point of the thermocouple to “ K TEMP” hole, and the work- 3) Do not use the meter if the battery is not replaced well or the battery case is not
ensure measure value accuracy. point to the place wanted to take temperature, the value will be displayed on fixed.
Note: LCD. 4) Before replacing battery or fuse, release the test leads from the test point and
1) To use manual method, if the resistance range is unknown beforehand, set the NOTE: turn power off.
FUNCTION switch to a higher range and work down. 1. When the input terminal is in open circuit, will display the “ normal temp.”
2) If "OL" displays on LCD, it means over-range. When measuring resistance more 2. Do not change the thermocouple, or, the accuracy can not be secured.
than 1MΩ, the meter may take a few seconds to stabilize. This is normal for 3. Do not input voltage at this range. 601E-0097-000A
high resistance readings. 4.11 Data hold
3) When the input is not connected, i.e. at open circuit, the figure "OL" will be Press “Hold” key, the current data will display on LCD; Press the key again, will
displayed for the over range condition. cancel the hold function.
4) When checking in-circuit resistance, be sure the power has been switched off all 4.12 Auto power off
capacitors are fully discharged. 1) Stop working for 15mins, the instrument is auto off and into the sleep mode.
5) Do not input any voltage at this range. The buzzer will sound before power off. Press any key to turn on the power.
4.6 Capacitance measurement 2) Pressing “ ” key before turning on can cancel the function.
1) Set the FUNCTION switch to " " position. 5. WARNING
2) Press "REL" once to adjust to zero. 1) When measuring voltage ensure that instrument is not connected or switched to
3) Connect the tested capacitor to “COM”, “V/Ω/Hz” input sockets in accordance a current or resistance range, or to the diode check. Always ensure that the
to the leads (the polarity of the red lead is “+”), the value will be displayed on correct terminals are used for the type of measurement to be made.
LCD. 2) Take extreme care when measuring voltage above 50V, especially from sources
4) When measuring more than 4uF, it takes 15 seconds to stabilize. where high energy is existed.
Note: 3) Avoid making connections to “live” circuits whenever possible.
1) Capacitance range have only auto mode. 4) When making current measurements ensure that the circuit not “live ”before
2) Before measuring each time, must press "REL" to ensure measure accuracy. opening it in order to connect the test leads.
3) Units: 1uF=1000nF 1nF=1000pF 5) Before making resistance measurements or diode test, ensure that the circuit
4) Capacitors should be fully discharged to avoid damaging the meter. under test is de-energized.