Operation manual RA-915Lab
Operation manual RA-915Lab
OPERATION MANUAL
В0300-00-00-000 OM
2020
Read the instructions in this manual carefully before operating the RA-
915Lab MERCURY ANALYZER. Neither LUMEX INSTRUMENTS nor its
representatives will accept any liability for the damage caused by non-
observance of the requirements of this manual.
Contact us at [email protected]
Version 11
Created 16.03.2021
Operation manual RA-915Lab В0300-00-00-000 OM I
Danger:
Warning:
Caution:
Safety guidelines
Important safety precautions
Read these rules thoroughly before starting operation with the ana-
lyzer.
• Carefully study all sections of this operation manual and ana-
lyzer design.
• To avoid electrical shock, never work with the analyzer panels
taken off.
• Do not put extraneous objects inside the analyzer through its
ports.
• Use only the power supply, which is provided.
• Never use a defective power supply cord, do not put any ob-
jects on the power supply cord, and locate it in a manner to
avoid a trip hazard.
• External surfaces of the analyzer and silicone tube connected
to the outlet fitting on the back panel of the analyzer can be
heated up to 60…80°С during the analyzer operation. To avoid
skin burn, do not touch them until they fully cool down.
Operation manual RA-915Lab В0300-00-00-000 OM III
Contents
1 Introduction.............................................................................................. 1
1.1 General description ................................................................................ 1
1.2 Performance characteristics and specifications ..................................... 2
1.3 Operating conditions .............................................................................. 2
1.4 Analyzer design ...................................................................................... 2
1.5 Analyzer Description and Principle of Operation .................................... 4
1.5.1 Description of Analyzer ....................................................................... 4
1.5.2 Description of Autosampler ................................................................. 7
1.5.3 General Principles of Operation .......................................................... 7
5 Troubleshooting ...................................................................................... 1
7 Warranty ................................................................................................... 3
Annex A ........................................................................................................... 4
List of Figures I
List of Figures
1 Introduction
The components, which compose the basic delivery kit, are shown in
Table 1-1.
В0300-00-00-000 OM Section 1 Introduction 3
Table 1-1
№ Item Quantity, pcs
The following marks are located on the front panel of the analyzer
(Fig. 1-2):
• Analyzer name;
• Manufacturer’s trademark.
The nameplate with the following information is located on the back
panel of the analyzer (Fig. 1-3):
• Manufacturer’s trademark;
В0300-00-00-000 OM Section 1 Introduction 5
• Analyzer name;
• Analyzer serial number;
• Year of manufacture;
• Power supply parameters;
• CE certification marks.
The analyzer panels are displayed in the Fig. 1-2, Fig. 1-3 and Fig.
1-4.
1 – USB port; 2 – on/off power switch; 3 – power cord socket; 4 – air (or other gas) inlet
fitting; 5 – outlet fitting (for removal of gaseous products of thermal decomposition); 6 -
nameplate
a) b)
Fig. 1-4 - Right-hand (a) and left-hand (b) side panels of the analyzer
2 Analyzer operation
Caution:
Caution:
a) b)
Fig. 2-2 – Removing right-hand side panel of the analyzer
• Secure the spacer with the red locking screw inside the analyz-
er for storage as shown in Fig. 2-6;
14 Operation manual RA-915Lab В0300-00-00-000 OM
• Using the handle, move the optical unit back to the left until it
stops;
• Replace the right-hand side panel of the analyzer. To do this,
assemble the panel so that the juts on the panel install into the
hinges on the analyzer’ body (see Fig. 2-7), press the panel on
top until it clicks.
a)
b)
Fig. 2-8 – Disassembling the cover from the front panel of the analyzer
• Using the hex wrench supplied with the spares kit, completely
unscrew the red locking screw and take out the insert (Fig. 2-9).
16 Operation manual RA-915Lab В0300-00-00-000 OM
• Using the hex wrench, secure the insert with the locking screw
inside the analyzer for their storage (see Fig. 2-10);
atomizer inlet port (Fig. 2-11). The cover is fixed to the front
panel with magnets.
Fig. 2-11 – Assembling the cover on the front panel of the analyzer
Warning:
Caution:
and bring out the other end of the tube to the exhaust ventila-
tion.
Caution:
Connect the supplied power cord into the socket (pos. 3 in Fig. 1-3)
on the back panel of the analyzer. Connect the analyzer to an AC network
using the power cord.
When operating the analyzer with the autosampler, the power of both
instruments is via the power cord connected to the RA-915Lab analyzer.
2.1.7 Connection to PC
Connect one end of the supplied USB cable to the appropriate port of
the PC, and the other end – to the USB port on the back panel of the ana-
lyzer (pos. 1 in Fig. 1-3).
When operating the analyzer with the autosampler, connection of
both instruments to the PC is via the USB cable connected to the RA-
915Lab analyzer.
20 Operation manual RA-915Lab В0300-00-00-000 OM
• Switch the PC on. The operating system will load, and the
Windows desktop screen will appear;
• If you do not intend to use the analyzer for a long time, switch
it off using the switch on the back panel (pos. 2 in Fig. 1-3).
В0300-00-00-000 OM Section 2 Analyzer operation 21
Warning:
Switch the analyzer off using the switch on the back panel only after
the analyser has fully cooled down (the green indicator on the front
panel has turned off).
Caution:
Do not turn off or restart the PC with the RAPID software activated.
This can disrupt the software and lead to data loss.
This section contains basic instructions for operation with the analyz-
er in manual mode. General procedure is as follows:
1) Prepare the analyzer for operation and switch it on ac-
cording to par. 2.1;
2) Follow the appropriate instructions when preparing sam-
ples, selecting operation modes, and performing sample
analysis;
3) Open the Sample analysis window in the RAPID soft-
ware;
4) Select the atomizer operation mode in the drop-down list
in the RA-915Lab atomizer control section of the Sam-
ple analysis window;
5) Click the Start button in the RA-915Lab atomizer control
section. The atomizer heaters and the air pump will start
operating according to the settings of the initial stage of
the mode. The Stop button, which activates the dropdown
list for selecting operation mode, appears instead of the
Start button;
22 Operation manual RA-915Lab В0300-00-00-000 OM
Caution:
It takes 40 min to get the analyzer ready for operation after it started
heating the atomizer. Continue operation only after the specified time
has elapsed
Prior to sample analysis carry out the calibration of the analyzer for the
given sample type and specified working conditions (i.e. mode) in which
sample analysis will be performed. Calibration should be carried out us-
ing (certified) reference materials with the established (certified) mercury
content.
Danger:
While operating the analyzer, the atomizer inlet port must be closed.
Open the inlet port only when inserting the sample boat.
It is strictly forbidden to insert any objects into the heated atomizer,
except for the sample boat, as well as to touch the inlet port of the at-
omizer.
Failure to comply with these requirements may result in severe burns.
В0300-00-00-000 OM Section 2 Analyzer operation 23
a)
b)
Fig. 2-13 – Placing the sample boat on the holder
a) b)
Fig. 2-14 – Inserting the holder with the sample boat into the atomizer inlet
port
Danger:
After taking out the holder with the sample boat from the atomizer,
they remain hot for a while. Avoid touching them - use tweezers! Do
not place hot boat holder on any surface other than the tray for boats
carrying supplied with the analyzer.
13) Integrate the output signal and save the result of analysis
according to the RAPID Software User’s Manual;
14) Repeat steps 7 – 13 for the following samples.
For the basic instructions of operating the analyser with the au-
tosampler refer to the Autosampler Operation Manual.
В0300-00-00-000 OM Section 3 Analyzer Maintenance 25
3 Analyzer Maintenance
Danger:
Make sure that the analyzer has cooled to room temperature before
cleaning the quartz windows of the analytical cell
a) b)
Fig. 3-1 – Location of the special wrench in the analyzer
a)
b)
c)
Fig. 3-2 – Taking out the window mount from the analyzer
28 Operation manual RA-915Lab В0300-00-00-000 OM
a) b)
Fig. 3-4 – Installing the window into the mount
Warning:
The window mount installed in the left side of the analyzer (see Fig.
3-5 (a)) differs from the mount installed in the right side of the ana-
lyzer (see Fig. 3-5 (b)). Incorrect installation of the mounts will lead to
incorrect analyzer operation!
a) b)
Fig. 3-5 – Window mounts
10) Move the optical unit by the handle to the left until it stops;
11) Replace the right-hand side panel of the analyzer. To do
this, assemble the panel so that the juts on the panel in-
stall into the hinges on the analyzer’ body (see Fig. 2-7),
press the panel on top until it clicks;
12) Replace the special wrench into the analyzer case secur-
ing it with the butterfly screw (Fig. 3-1 (a));
13) Replace the left-hand side panel of the analyzer as de-
scribed in the 11th step.
30 Operation manual RA-915Lab В0300-00-00-000 OM
Danger:
Make sure that the analyzer has cooled to room temperature before
replacing the quartz windows of the analytical cell
Warning:
The window mount installed in the left side of the analyzer (see Fig.
3-5 (a)) differs from the mount installed in the right side of the ana-
lyzer (see Fig. 3-5 (b)). Incorrect installation of the mounts will lead to
incorrect analyzer operation!
Danger:
Make sure that the analyzer has cooled to room temperature before
cleaning the tray and atomizer inlet port
To clean the tray and the atomizer inlet port, proceed as follows:
3) Clean the tray located on the inside of the cover (see Fig.
3-6);
Danger:
Make sure that the analyzer has cooled to room temperature before
cleaning the analytical cell and the baffles.
7) Clean the baffle and the analytical cell using the cleaning
brush supplied in the spares kit;
8) Install the cleaned baffle so that the edges of its protru-
sions are deeper than the end of the cell, as shown in Fig.
3-8;
В0300-00-00-000 OM Section 3 Analyzer Maintenance 33
Fig. 3-7 – Taking out the baffle from the analytical cell
5) Take the spacer (Fig. 3-10) out, then remove the case of
the filtering element (Fig. 3-11);
Fig. 3-13 – The sealing ring for the case of the filtering element
At the end of your products and accessories life, it must not be dis-
carded as domestic waste. Ref: EU Directive 2012/19/EU on Waste Electri-
cal and Electronic Equipment Directive (WEEE). Please contact your dis-
tributor / supplier for further information. For the users outside of the EU
consult applicable regulations.
В0300-00-00-000 OM Section 5 Troubleshooting 1
5 Troubleshooting
7 Warranty
Annex A
(informative)
Physical Foundation and Principle of Operation
The RA-915Lab analyzer operation is based on differential Zeeman
atomic absorption spectrometry using high frequency modulation of light
polarization (ZAAS-HFM).
Mercury EDL lamp (1), which is placed in the gap between the poles
of the magnet, is excited by the high frequency generator (2). The light
successively passes through the polarization modulator (3) to the built-in
analytical cell (4) and is finally detected by a photo detector (5). The photo
detector signal arrives at the electronic signal-processing unit (6), where
the signal is separated at the modulation frequency and the analytical sig-
nal is formed. After that the information is displayed on the screen of a PC
(7).