QSG AR-6480 DWDualWave™ Arthroscopy Fluid Management System 0244-3-SUBr0 - en - US
QSG AR-6480 DWDualWave™ Arthroscopy Fluid Management System 0244-3-SUBr0 - en - US
Read all safety and set-up instructions in the Arthrex DualWave Arthroscopy Pump User’s Guide and
Directions For Use before using this this product. Please visit www.Arthrex.com for all product literature.
Additional languages are available on the Arthrex eDFU website located at www.arthrex.com/eDFU
Lavage Rinse
Pressure
Pressure Up Lavage: Timed increase of
Down
pressure for improved control of bleeding.
2
INFLOW
INFLOW TUBING SET-UP
1 2 3 4 5
Remove orange clip Insert connector into Open inflow door Insert green tubing Wrap tubing
pump pressure coupler hub around rollers
Inflow door
Rollers
6 7
Pump
pressure coupler
Clear tubing hub
8
Set Pressure
Boost Control
Run/Stop Menu
Running
Pressure
Reading
Flow Rate
Lavage
4
OUTFLOW
OUTFLOW TUBING SET-UP Outflow Boot
Luer Lock Shaver
Cut as needed
Waste
1 2
Connect to shaver Connect to Pass off Cassette and Red Waste Tubing
suction* independent cannula
Cannula
*Push connector onto shaver barb fully to ensure secure connection.
3 4 5 6 7
Open outflow door *Install outflow Close outflow door Outflow display Connect red waste
cassette tubing to waste system
*For easy set-up, move the rollers so that two wheels are flush. Extend the boot of the outflow cassette over the two wheels at the
same time. With one hand, hold the boot in place; and, with the second hand, pull down on all three tubes, pushing the cassette in
place. The cassette will be held in place by the edges of the outflow bezel.
5
INFLOW/OUTFLOW
INFLOW/OUTFLOW USER INTERFACE
Inflow/Outflow
Cannula Suction
Shaver Suction
Control
Control
Set Pressure
Run/Stop Menu
Running
Lavage Rinse
6
SHAVER DETECTION
SYNERGY SHAVER DETECTION AND HEADS UP DISPLAY
1 2
3 4
*For Synergy heads up display use AR-3200-1040 accessory cable kit to connect
DualWave to camera console. (Only 1 cable is needed to connect the DualWave to the
camera console.) Prior to using this system, see DFU 950-0052-XX for more detail. 7
SHAVER DETECTION
1 2
Shaver detect cord Plug into DualWave™
3 4
5 6
** Critical Failure ** Return to Arthrex® for repair. Does not pump 1. Check for error messages.
when activated 2. Open all tubing clamps and shut-off valves.
** Door Open ** 1. Roller housing is not secured – ensure the 3. Ensure the actual pressure is below the
locking lever is properly secured. target pressure.
2. If the failure persists, return to Arthrex for repair. 4. Check the tubing for pinches, kinks, or blockages.
5. If the failure persists, return to Arthrex for repair.
** Overpressure ** 1. Increase or open the outflow. Inadequate 1. Increase the pressure.
2. Manipulate the joint to a stress-free position. distention, liquid 2. Activate Lavage mode.
3. If the failure persists, return to Arthrex for repair. bloody or cloudy 3. Reduce the outflow.
4. Use high-flow cannulas.
5. If the failure persists, return to Arthrex for repair.
** Pressure Fault ** 1. Ensure adequate fluid supply.
2. Decrease the outflow. No (or inadequate) 1. Check for error messages.
3. Check the tubing for damage or pinches, kinks, flow 2. Check that all tubing clamps are open.
or blockages. 3. Check the settings for flow and pressure.
4. Check the tubing for the proper connections. 4. Check the tubing for pinches, kinks, or blockages.
5. Replace the tubing. 5. Check that the tubing seats correctly over
6. If the failure persists, return to Arthrex for repair. the rollers.
6. Verify use of high-flow cannulas.
** Tubing Out ** 1. If the tubing sensor indicator’s LED is red, the 7. If the failure persists, return to Arthrex for repair.
tubing is not properly connected.
2. Ensure the tubing sensor coupler is open, not No Shaver Detect Synergy Shaver Detection
seated. 1. Verify DualWave™ software is at least 1.8
3. Ensure that the tubing pressure line connector is 2. Verify Synergy Resection Console is at least 2.3.
seated completely. 3. Ensure the Synergy cable is connected.
4. Change the tubing. 4. Press stop and then run on DualWave while the
5. If the failure persists, return to Arthrex for repair. Synergy Resection Console is on.
5. Replace cable.
Console fails 1. Ensure no tubing is connected to the pump 6. If the failure persists, return to Arthrex for repair.
Self Diagnostic Test during the power on sequence. Universal Shaver Detection
2. If the failure persists, return to Arthrex for repair. 1. Ensure the universal shaver cable is connected.
2. Turn DualWave on and off while the shaver
Console will not 1. Check the AC mains power cord. system is powered on.
power up 2. Try an alternate power outlet. 3. Adjust the DELAY and/or THRESHOLD settings.
3. Check the AC mains fuses. 4. Isolate the power source.
4. If the failure persists, return to Arthrex for repair. 5. If the failure persists, return to Arthrex for repair.
9
CONTACT INFORMATION
CONTACT INFORMATION
Any serious incident that occurs in relation to the device should be reported to the manufacturer and to the health authority where the
incident occurred.
Erwin-Hielscher-Strasse 9
81249 München, Germany
Arthrex GmbH +49 89 90 90 05-0
arthrex.de
[email protected]
Ampèrestraat 9
5928 PE
Venlo, Netherlands
Arthrex Distribution Hub +31 88 712 9800
EMEA B.V. arthrex.nl
[email protected]
Unit 1 Bessemer Park
Shepcote Lane
Arthrex Ltd., Sheffield
S9 1DZ
United Kingdom
+44 0 114 232 9180
Hauptstrasse 16,
3186 Düdingen, Switzerland
confinis ch-rep ag, +41 26 494 8 494
Manufacturer’s Australian Suite 501, 20 Rodborough Road
Sponsor Frenchs Forest, NSW, 2086 Australia
Arthrex Australia Pty Ltd +1 800 950 637
arthrex.com.au
10 [email protected]
NOTES
NOTES
11
This description of technique is provided as an educational tool and clinical aid to assist properly licensed medical
professionals in the usage of specific Arthrex products. As part of this professional usage, the medical professional must
use their professional judgment in making any final determinations in product usage and technique. In doing so, the medical
professional should rely on their own training and experience and should conduct a thorough review of pertinent medical
literature and the product’s Directions For Use.