OCS-GP User Manual
OCS-GP User Manual
Main
Specifications ................................................................................................................ 1
Calibration..........................................................................................20
Conversion principle △
Non-linear <0.01%F.S
backlight)
Indicator :0℃~+40℃
1
Display:
N. ,G
10000 Stable
Zero
■Number: 00000
It is used to classify and count weighing data, and 5 digits can be input.
■Item number: 00
■Times: 000
Indicates the accumulated weighing times under the current interface information.
■ : Signal indication, when this sign is displayed, it means that the signal is
normal and can be weighed.
2
■Gross/Net:
■Stable:
Weighing data stable sign, when this character is displayed, it means that the current
weighing has been stabilized and can be carried out line storage printing operation, or
Keyboard function:
BL 0/F2
■ "SN"
Used to set the number. For multiple delivery units or individuals, different numbers
■ "CN"
■ "SEARCH"
■ "BL"
3
■ "1/Clear"
■ "2/TIME"
■ "3/Date"
■ "4/L.FL"
■ "5/AUTO"
■ "6/STAT"
■ "7/F1" key
■ "8/G.NET"
■ "9/R.P."
■ "0/F2" key
■ "PRETARE"
Used to enter the setting parameter menu and exit the function composite key.
4
■ "ENTER"
■ "SAVE/PRINT"
■ "ZERO"
■ "ON"
Power button
■ "OFF"
shutdown key
Parameter Setting
QQ-6.00
1.Print Setting 2.System Setting
Press the "1/Clear" key to enter the "Storage Print Settings" interface:
details
5
3.Automatic Save and Print 6. Print parameters
Concept explanation:
Set this item for the weighing object whose serial number and cargo number have
been input. After the weighing data is stable, press"SAVE/PRINT" , you can realize
Set this item for weighing objects whose number and item number are both zero.
Press " SAVE/PRINT" after the weighing data is stable to realize the required data
processing method.
If this item is set to valid, press the "5/AUTO" in the weighing state, the indicator
can switch to can be converted to automatically store printing status and the prompt
" A" is displayed at the same time. Printing is set to be valid, every weighing
After the weight data is stable for about a few seconds, the indicator will
automatically perform storage and print once, and print out the weighing
single record. If the print setting is invalid, only the data will be stored without
printing.
In the state of automatic save and print, press the "5/AUTO" , the indicator can be
converted to manual storage and printing status, and the prompt "A" goes out.
4. Form printing
If this option is selected, after the weight data is stable, press the "SAVE/PRINT", the
indicator will print a weighing list including all information will be produced, and the
6
5. Statistical details
If this item is selected, after entering the statistical interface, the data in the printing
The printer only prints the receipt of the total weight, and does not print the detailed
weighing record.
6. Print parameters
This parameter can be set to count the number of blank paper feeding lines after
printing.
After the instrument is stored and printed each time, the crane scale needs to return
The next lifting operation can be carried out without returning to zero. This item is
Due to the structural characteristics of the printer itself, every time a record is printed,
The record is not found. After this option is selected, the printer will feed the paper
4. Print Time
After this function is selected, the current time will be printed on the weighing sheet
7
In the “Save Print Setting” interface, press the “1/Clear” key, and the display will
display:
Use SN and CN
If this item is selected to be valid, after the weighing data is stable, press the
" SAVE/PRINT", and the printer will print out the record of the weighing, and save
the weighing data in the corresponding serial number and item number, the number of
If this item is selected to be valid, after the weighing data is stable, press the
" SAVE/PRINT", the indicator will save the data to the memory of the corresponding
number and item number, and add one to the number of times, but do not print
weighing record this time. (When you need to print, you can go to the statistical
Mention
■Only one option can be selected between “SAVE /PRINT” and “Save not Print” .
■ Press the corresponding number keys to select the desired item and press the
■ No serial number, article number means that both the serial number and the article
number are all "0".
■ The setting method is the same as the parameter setting method of serialized goods
number.
8
Zero Setting
In the "Parameter Setting" interface, press the "3/DATE" to enter the "Zero
Setting" :
Zero Setting
This parameter can set the range of zero setting for weighing the scale body when
Setting method:
Please press the "1/CLEAR" key, the cursor under the parameter number will display,
press the "SEARCH" key to make the number switch between "4, 10, 20, 40" values,
20%FS means 20% of the maximum weighing capacity (Generally, it should be set
to 20%).
This parameter can be used to set the range of the indicator to return to zero when
Setting method:
Please press the "2/TIME" , the cursor under the parameter number will display,
The number can be switched between "4, 10, 20, 40" and press the "ENTER" after
selection.
9
4%FS means 4% of the maximum weighing capacity (Generally, it should be set to
4%).
Communication Settings
1. Screen Display
If this function is selected, the instrument can display the current weight on the
filter settings
In the "Parameter Setting" interface, press the "5/AUTO" to enter the "Filter
Setting" :
■ Set this parameter according to the stable conditions on site, the worse the stable
conditions are, the larger the parameter is set.
■ The larger the filter parameter value, the slower the displayed data changes.
■ It can be set by pressing the corresponding number key, generally the parameter is
set to 3 or 4 levels, a total of 0-7 levels can be set.
Other settings
In the "Parameter Setting" interface, press the "6/STAT" to enter the "Other
Settings" interface:
Chinese settings
Chinese Setting
1.Chinese CN Setting
1. Chinese CN Setting
Press the "1/CLEAR" to enter the "Chinese CN Setting", and it will display:
10
Chinese CN Setting
Enter the cargo number, press the "ENTER" , the cursor jumps to the product name,
press the number keys to enter the Chinese character location code, press the
"ENTER" to confirm. If you want to enter the next cargo number, press the
and jump back to the cargo number, so that you can continue to enter the next item
number, and press the "STAT" after all input,print out the Chinese article number
Press the "2/TIME" to enter the print header setting item, the indicator displays:
At this time, press the number keys to input the Chinese character area code, and then
press the "ENTER" to confirm. Then press "STAT" to print out the print header
comparison table.
Mention:
■ You can enter up to three Chinese characters for the Chinese product name.
■ The Chinese goods name that has been entered, the corresponding Chinese will be
displayed in the goods number item when printing the statistical sheet later name.
■ The Chinese product name can be deleted, the operation method is: enter the
Chinese product number setting item, and the corresponding product will be
11
■ Up to 8 Chinese characters can be entered in the header of Chinese printing, which
The corresponding Chinese title is printed on the top line of the printed form.
■ In the process of inputting the area code, ifyou enter a wrong number, you can
This function can check the current battery voltage. Press the "2/TIME" in the
Subdivision
After this function is selected, press the "0/F2" when the meter is weighing, the
division value of the meter will change step by step within the set division value,
press once, and the division value on the interface will change once.
■ Press the "SN" in the weighing interface, a cursor will be displayed under the last
digit of the number, indicates that the bit is modifiable. At this time, press the number
keys to input the corresponding number value, each time you press the key, the
number moves one bit to the left, and a maximum of 5 digits can be input. After
■ In this serial number interface, after the weighing data is stable, press the
“SAVE/PRINT” , and the weighing data will be stored under this number. If the
printing status is valid, the indicator will print out the single weighing record.
12
■ Press the "CN" in the weighing interface, there is a cursor display under the last
digit of the article number, indicating that this digit is modifiable. At this time, press
the number keys to input the corresponding item number, each time the number key is
pressed, the number will move one bit to the left, and a maximum of 2 digits can be
■ In the interface of the item number, after the weighing data is stable, press the
" SAVE/PRINT" , and the data will be saved under the item number, if the printing
status is valid, the indicator will print out the single weighing record.
Mention:
If save and print is enabled in the save and print parameter settings (see P5-6 for
details), after each weighing data is stable, press the"SAVE/PRINT" , the printer will
print out one weight record of weighing. After a set of weighing is completed, press
the " STAT" and the printer will print out all detailed information of this group
Weighing list
---------------------
Number: 00123
Item No.: 02
---------------------
No. Time Weight
13
004 9:12 1000
■ If Save and Print is enabled in the save print parameter setting (see P6-7 for details),
every weighing data is stable, press the "SAVE/PRINT" , the weighing record will
■ When you need to print a statistical sheet, you can press the "STAT" to enter the
following interface:
At this time, you can press the number keys to select the statistical items, and press
the "ENTER" after selection, the printer will print out the format of the
corresponding statistics sheet is the same as the first statistical printing method.
3.STAT by DATE
14
Press the number keys to select the query item (three items can be combined for
searching), and press the "ENTER" after selection, The instrument displays the total
weight of the query items and the current memory status, and the weighing records
are displayed in reverse order, can view every the serial number, time and weight of
the record.
1 . Press "ENTER" on the first interface to view the current memory status of the
instrument, for example: display 80/2000, means: 80 is the current usage; 2000 is the
2. Press the "SEARCH" on the second interface, and the data will be searched
according to the cumulative serial number from large to small, and press " ENTER"
3. The "Total W" under the first interface is the sum of the weights of several
weighing that have been stored and printed before.
4. Specification input: In the first interface, press the "4/L.F." to display the following
interface:
15
"Specification: 000 0" means: the first 3 digits are the specification value, and the last
Press the "2/TIME" , input "025 1", and press the "Enter".
Press the "1/CLEAR" in the weighing interface, the indicator displays as follows:
1. Clear □
2. Total Clear □
1.Clear
Clear the last stored single weighing record of the serial number and cargo number
corresponding to the current interface. Press the number key 1 to select it and press
" ENTER", the indicator will return to the weighing interface, and the total time will
be reduced by one at the same time, and the list operation is successful.
Mention:
■If data is stored in the memory due to misoperation, the data can be cleared by using
the list function. There is no need to return to zero, and the storage and printing
■ When the number of times is "0", the list operation cannot be performed after
statistics.
2. Total clear
Clear all data in memory. Press the number key 2 to select this item, press "ENTER"
to return to the weighing interface, the interface displays the serial number, the item
16
number becomes zero, and the total time is reset to zero, indicating that the memory
installation steps:
1 . As shown in Figure 1, press the door opening wrench in the direction of the arrow.
2 . As shown in Figure 2, push the door opener outward to make the head roller jump
3. Open the cover of the paper compartment, load the printing paper, and pull out a
section (beyond a little tearing paper teeth), put the paper neatly, and pay attention to
the direction of the paper, as shown in Figure 3, the side with the liquid medicine
4 .Close the cover of the paper compartment, press the paper feed shaft of the print
head to align with the printing paper, press the paper feed shaft of the print head back
to the print head with a little force, and push the door opener back to reset.
5. Turn on the power of the printer, then press the LF key to rotate the machine head,
17
Use of chargers and batteries
when to charge
■ When the under-voltage sign on the battery sign on the display screen is displayed,
it means that it needs to be charged, please charge the meter as soon as possible.
When the under-voltage sign is on, the meter can still be used for 1 minute, and it will
■ During normal wireless signal reception and weighing process, the display screen of
the crane scale display shows: " Crane scale battery low voltage" means that the
battery power supply of the scale body is under-voltage and needs to be charged.
After the under-voltage display, it can be used for another 30 minutes, and the power
supply end of the scale body will automatically shut down after 30 minutes.
Charging time
■ The scale battery is a 6V/4Ah lead-acid battery, and the normal use time is about 55
hours.
■ The instrument battery is a 6V/4Ah lead-acid battery, and the normal use time is
about 40 hours.
■ The charger has three indication states: connected to the 220V power supply, the
indicator light is green; when charging, the indicator light is red; when the battery is
■ The charger is an intelligent charger, and it takes about 4 to 5 hours to fully charge
the battery. When the battery is fully charged, the smart charger will automatically
switch to a trickle state to protect the battery. In order to prolong the service life of the
battery, it is best to remove it as soon as possible after the battery is fully charged.
18
When charging the meter, please connect the 220V power supply to the
charger first, and then insert the charging plug into the charging socket of the meter
after the indicator light of the charger turns green to prevent the voltage surge from
following precautions:
■ Do not short-circuit the battery by entering water. Short-circuiting will damage the
battery.
■ Do not short-circuit the battery plug, otherwise the battery and the short-circuited
objects will be damaged.
■ If the battery is left unused for a long time, please take it out and charge it every
two months.
■ In order to prolong the service life of the lead-acid battery as much as possible,
please charge the battery after using it until it is under voltage. And every time you
Install the antenna and battery on the crane scale, turn on the switch, and the
working indicator light flashes quickly, indicating that the scale body is in the
working state.
19
Connect the random antenna to the antenna jack on the top of the handheld meter,
and press the "ON" button on the meter. If the communication is normal, the signal
sign of the instrument will light up, and after a few seconds, the current weighing of
the scale body will be displayed, indicating that the upper and lower ends are working
normally.
Note:
■ When the weighing data is stable, the meter will automatically shut down if there is
no operation or data change for 30 minutes. To use it again, you need to reboot.
Calibration:
20
Last, hanging the calibrate mass, when mass stable, wait 15 seconds, the
calibration is done. Return to the main interface for use
The safety factor has been considered in the design of this wireless crane scale.
Please do not exceed its maximum weighing capacity when using it. The maximum
weighing capacity has been marked on the obvious part of the crane scale. The
■ Please check the gauges, rings, hooks and other parts regularly every three months.
If any abnormality is found, please contact the company in time to ensure safe and
normal use.
■ Severe collision and long-term rain should be avoided. Special attention should be
paid to prevent the scale body from falling from the hook. Falling from a high place
■ Non- high temperature crane scales cannot work in high temperature environments.
It is strictly forbidden to conduct electric welding on the scale body, otherwise the
20
■ When the crane scale is in normal use, the antenna does not need to be removed
every day. Frequent disassembly and assembly of the antenna will easily cause wear
and tear on the antenna and socket. Pay attention to protect the antenna, do not collide
with water. In case of damage, please contact the supplier, do not replace the
Except for the printer and the battery, the rest of the parts of this machine
are guaranteed for one year without man- made damage.
■ The service life of the printer is directly related to the frequency of use, that is, the
printing volume. If the frequency of use is high, the use time will be short, and if the
frequency of use is low, the use time will be long. Therefore, the printer cannot use
■ The service life of the battery is directly related to the usage and charging habits.
The battery must be fully charged before it can be charged. When charging, the
Habits will reduce battery life. Therefore, the battery cannot be used as a warranty
deadline.
21