0% found this document useful (0 votes)
2 views

ch080

The document discusses implantable cardioverter defibrillators (ICDs), which detect and treat life-threatening arrhythmias by delivering high-voltage shocks to restore normal heart rhythm. It details the evolution of ICD technology, including advancements in pulse generators, electrode systems, and arrhythmia detection algorithms, emphasizing their effectiveness and patient comfort. Additionally, it addresses the integration of various therapeutic capabilities and the ongoing challenges in managing conditions like atrial fibrillation.

Uploaded by

carlos ali issa
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
2 views

ch080

The document discusses implantable cardioverter defibrillators (ICDs), which detect and treat life-threatening arrhythmias by delivering high-voltage shocks to restore normal heart rhythm. It details the evolution of ICD technology, including advancements in pulse generators, electrode systems, and arrhythmia detection algorithms, emphasizing their effectiveness and patient comfort. Additionally, it addresses the integration of various therapeutic capabilities and the ongoing challenges in managing conditions like atrial fibrillation.

Uploaded by

carlos ali issa
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 9

Duffin, E. G. “ Implantable Defibrillators.


The Biomedical Engineering Handbook: Second Edition.
Ed. Joseph D. Bronzino
Boca Raton: CRC Press LLC, 2000
80
Implantable
Defibrillators
80.1 Pulse Generators
80.2 Electrode Systems (“LEADS”)
80.3 Arrhythmia Detection
80.4 Arrhythmia Therapy
80.5 Implantable Monitoring
80.6 Follow-up
Edwin G. Duffin 80.7 Economics
Medtronic, Inc. 80.8 Conclusion

The implantable cardioverter defibrillator (ICD) is a therapeutic device that can detect ventricular
tachycardia or fibrillation and automatically deliver high-voltage (750 V) shocks that will restore normal
sinus rhythm. Advanced versions also provide low-voltage (5–10 V) pacing stimuli for painless termina-
tion of ventricular tachycardia and for management of bradyarrhythmias. The proven efficacy of the
automatic implantable defibrillator has placed it in the mainstream of therapies for the prevention of
sudden arrhythmic cardiac death.
The implantable defibrillator has evolved significantly since first appearing in 1980. The newest devices
can be implanted in the patient’s pectoral region and use electrodes that can be inserted transvenously,
eliminating the traumatic thoracotomy required for placement of the earlier epicardial electrode systems.
Transvenous systems provide rapid, minimally invasive implants with high assurance of success and
greater patient comfort. Advanced arrhythmia detection algorithms offer a high degree of sensitivity with
reasonable specificity, and extensive monitoring is provided to document performance and to facilitate
appropriate programming of arrhythmia detection and therapy parameters. Generator longevity can now
exceed 4 years, and the cost of providing this therapy is declining.

80.1 Pulse Generators


The implantable defibrillator consists of a primary battery, high-voltage capacitor bank, and sensing and
control circuitry housed in a hermetically sealed titanium case. Commercially available devices weigh
between 197 and 237 grams and range in volume from 113 to 145 cm3. Clinical trials are in progress on
devices with volumes ranging from 178 cm3 to 60 cm3 and weights between 275 and 104 grams. Further
size reductions will be achieved with the introduction of improved capacitor and integrated circuit
technologies and lead systems offering lower pacing and defibrillation thresholds. Progress should parallel
that made with antibradycardia pacemakers that have evolved from 250-gram, nonprogrammable, VOO
units with 600-µJ pacing outputs to 26-gram, multiprogrammable, DDDR units with dual 25-µJ outputs.
Implantable defibrillator circuitry must include an amplifier, to allow detection of the millivolt-range
cardiac electrogram signals; noninvasively programmable processing and control functions, to evaluate

© 2000 by CRC Press LLC


the sensed cardiac activity and to direct generation and delivery of the therapeutic energy; high-voltage
switching capability; dc-dc conversion functions to step up the low battery voltages; random access
memories, to store appropriate patient and device data; and radiofrequency telemetry systems, to allow
communication to and from the implanted device. Monolithic integrated circuits on hybridized substrates
have made it possible to accomplish these diverse functions in a commercially acceptable and highly
reliable form.
Defibrillators must convert battery voltages of approximately 6.5 V to the 600–750 V needed to
defibrillate the heart. Since the conversion process cannot directly supply this high voltage at current
strengths needed for defibrillation, charge is accumulated in relatively large (≈85–120µF effective capac-
itance) aluminum electrolytic capacitors that account for 20–30% of the volume of a typical defibrillator.
These capacitors must be charged periodically to prevent their dielectric from deteriorating. If this is not
done, the capacitors become electrically leaky, yielding excessively long charge times and delay of therapy.
Early defibrillators required that the patient return to the clinic periodically to have the capacitors
reformed, whereas newer devices do this automatically at preset or programmable times. Improved
capacitor technology, perhaps ceramic or thin-film, will eventually offer higher storage densities, greater
shape variability for denser component packaging, and freedom from the need to waste battery capacity
performing periodic reforming charges. Packaging density has already improved from 0.03 J/cm3 for
devices such as the early cardioverter to 0.43 J/cm3 with some investigational ICDs. Capacitors that allow
conformal shaping could readily increase this density to more than 0.6 J/cm3.
Power sources used in defibrillators must have sufficient capacity to provide 50–400 full energy charges
(≈34 J) and 3 to 5 years of bradycardia pacing and background circuit operation. They must have a very low
internal resistance in order to supply the relatively high currents needed to charge the defibrillation capacitors
in 5–15 s. This generally requires that the batteries have large surface area electrodes and use chemistries that
exhibit higher rates of internal discharge than those seen with the lithium iodide batteries used in pacemakers.
The most commonly used defibrillator battery chemistry is lithium silver vanadium oxide.

80.2 Electrode Systems (“Leads”)


Early implantable defibrillators utilized patch electrodes
(typically a titanium mesh electrode) placed on the surface
of the heart, requiring entry through the chest (Fig. 80.1).
This procedure is associated with approximately 3–4% peri-
operative mortality, significant hospitalization time and
complications, patient discomfort, and high costs. Although
subcostal, subxiphoid, and thoracoscopic techniques can
minimize the surgical procedure, the ultimate solution has
been development of fully transvenous lead systems with
acceptable defibrillation thresholds.
Currently available transvenous leads are constructed
much like pacemaker leads, using polyurethane or silicone
insulation and platinum-iridium electrode materials.
Acceptable thresholds are obtained in 67–95% of patients,
with mean defibrillation thresholds ranging from 10.9–18.1 J.
These lead systems use a combination of two or more elec-
trodes located in the right ventricular apex, the superior FIGURE 80.1 Epicardial ICD systems typi-
vena cava, the coronary sinus, and sometimes, a subcuta- cally use two or three large defibrillating patch
neous patch electrode is placed in the chest region. These electrodes placed on the epicardium of the left
leads offer advantages beyond the avoidance of major sur- and right ventricles and a pair of myocardial
gery. They are easier to remove should there be infections electrodes for detection and pacing. The gen-
or a need for lead system revision. The pacing thresholds erator is usually placed in the abdomen. (Copy-
of current transvenous defibrillation electrodes are typically right Medtronic, Inc. Used with permission.)

© 2000 by CRC Press LLC


FIGURE 80.2 The latest transvenous fibrillation systems employ a single catheter placed in the right ventricular
apex. In panel a, a single transvenous catheter provides defibrillation electrodes in the superior vena cava and in the
right ventricle. This catheter provides a single pace/sense electrode which is used in conjunction with the right
ventricular high-voltage defibrillation electrode for arrhythmia detection and antibradycardia/antitachycardia pacing
(a configuration that is sometimes referred to as integrated bipolar). With pulse generators small enough to be placed
in the pectoral region, defibrillation can be achieved by delivering energy between the generator housing and one
high-voltage electrode in the right ventricle (analogous to unipolar pacing) as is shown in panel b. This catheter
provided bipolar pace/sense electrodes for arrhythmia detection and antibradycardia/antitachycardia pacing. (Copy-
right Medtronic, Inc. Used with permission.)

0.96 ± 0.39 V, and the electrogram amplitudes are on the order of 16.4 ± 6.4 mV. The eventual application
of steroid-eluting materials in the leads should provide increased pacing efficiency with transvenous lead
systems, thereby reducing the current drain associated with pacing and extending pulse generator lon-
gevity.
Lead systems are being refined to simplify the implant procedures. One approach is the use of a single
catheter having a single right ventricular low-voltage electrode for pacing and detection, and a pair of
high-voltage defibrillation electrodes spaced for replacement in the right ventricle and in the superior
vena cava (Fig. 80.2a). A more recent approach parallels that used for unipolar pacemakers. A single
right-ventricular catheter having bipolar pace/sense electrodes and one right ventricular high-voltage
electrode is used in conjunction with a defibrillator housing that serves as the second high-voltage
electrode (Fig. 80.2b). Mean biphasic pulse defibrillation thresholds with the generator-electrode placed
in the patient’s left pectoral region are reported to be 9.8 ± 6.6 J (n = 102). This approach appears to be
practicable only with generators suitable for pectoral placement, but such devices will become increasingly
available.

80.3 Arrhythmia Detection


Most defibrillator detection algorithms rely primarily on heart rate to indicate the presence of a treatable
rhythm. Additional refinements sometimes include simple morphology assessments, as with the proba-
bility density function, and analysis of rhythm stability and rate of change in rate.
The probability density function evaluates the percentage of time that the filtered ventricular electrogram
spends in a window centered on the baseline. The rate-of-change-in-rate or onset evaluation discriminates
sinus tachycardia from ventricular tachycardia on the basis of the typically gradual acceleration of sinus
rhythms versus the relatively abrupt acceleration of many pathologic tachycardias. The rate stability

© 2000 by CRC Press LLC


function is designed to bar detection of tachyarrhythmias as long as the variation in ventricular rate
exceeds a physician-programmed tolerance, thereby reducing the likelihood of inappropriate therapy
delivery in response to atrial fibrillation. This concept appears to be one of the more successful detection
algorithm enhancements.
Because these additions to the detection algorithm reduce sensitivity, some defibrillator designs offer
a supplementary detection mode that will trigger therapy in response to any elevated ventricular rate of
prolonged duration. These extended-high-rate algorithms bypass all or portions of the normal detection
screening, resulting in low specificity for rhythms with prolonged elevated rates such as exercise-induced
sinus tachycardia. Consequently, use of such algorithms generally increases the incidence of inappropriate
therapies.
Improvements in arrhythmia detection specificity are desirable, but they must not decrease the excel-
lent sensitivity offered by current algorithms. The anticipated introduction of defibrillators incorporating
dual-chamber pacemaker capability will certainly help in this quest, since it will then be possible to use
atrial electrograms in the rhythm classification process. It would also be desirable to have a means of
evaluating the patient’s hemodynamic tolerance of the rhythm, so that the more comfortable pacing
sequences could be used as long as the patient was not syncopal yet branch quickly to a definitive shock
should the patient begin to lose consciousness.
Although various enhanced detection processes have been proposed, many have not been tested
clinically, in some cases because sufficient processing power was not available in implantable systems,
and in some cases because sensor technology was not yet ready for chronic implantation. Advances in
technology may eventually make some of these very elegant proposals practicable. Examples of proposed
detection enhancements include extended analyses of cardiac event timing (PR and RR stability, AV
interval variation, temporal distribution of atrial electrogram intervals and of ventricular electrogram
intervals, timing differences and/or coherency of multiple ventricular electrograms, ventricular response
to a provocative atrial extrastimuli), electrogram waveform analyses (paced depolarization integral,
morphology analyses of right ventricular or atrial electrograms), analyses of hemodynamic parameters
(right-ventricular pulsatile pressure, mean right atrial and mean right ventricular pressures, wedge
coronary sinus pressure, static right ventricular pressure, right atrial pressure, right ventricular stroke
volume, mixed venous oxygen saturation and mixed venous blood temperature, left ventricular imped-
ance, intramyocardial pressure gradient, aortic and pulmonary artery flow), and detection of physical
motion.
Because defibrillator designs are intentionally biased to overtreat in preference to the life-threatening
consequences associated with failure to treat, there is some incidence of inappropriate therapy delivery.
Unwarranted therapies are usually triggered by supraventricular tachyarrhythmias, especially atrial fibril-
lation, or sinus tachycardia associated with rates faster than the ventricular tachycardia detection rate
threshold. Additional causes include nonsustained ventricular tachycardia, oversensing of T waves, double
counting of R waves and pacing stimuli from brady pacemakers, and technical faults such as loose lead-
generator connections or lead fractures.
Despite the bias for high detection sensitivity, undersensing does occur. It has been shown to result
from inappropriate detection algorithm programming, such as an excessively high tachycardia detection
rate; inappropriate amplifier gain characteristics; and electrode designs that place the sensing terminals
too close to the high-voltage electrodes with a consequent reduction in electrogram amplitude following
shocks. Undersensing can also result in the induction of tachycardia should the amplifier gain control
algorithm result in undersensing of sinus rhythms.

80.4 Arrhythmia Therapy


Pioneering implantable defibrillators were capable only of defibrillation shocks. Subsequently, synchro-
nized cardioversion capability was added. Antibradycardia pacing had to be provided by implantation
of a standard pacemaker in addition to the defibrillator, and, if antitachycardia pacing was prescribed, it

© 2000 by CRC Press LLC


was necessary to use an antitachycardia pacemaker. Several currently marketed implantable defibrillators
offer integrated ventricular demand pacemaker function and tiered antiarrhythmia therapy (pacing/car-
dioversion/defibrillation). Various burst and ramp antitachycardia pacing algorithms are offered, and
they all seem to offer comparably high success rates. These expanded therapeutic capabilities improve
patient comfort by reducing the incidence of shocks in conscious patients, eliminate the problems and
discomfort associated with implantation of multiple devices, and contribute to a greater degree of success,
since the prescribed regimens can be carefully tailored to specific patient needs. Availability of devices
with antitachy pacing capability significantly increases the acceptability of the implantable defibrillator
for patients with ventricular tachycardia.
Human clinical trials have shown that biphasic defibrillation waveforms are more effective than
monophasic waveforms, and newer devices now incorporate this characteristic. Speculative explanations
for biphasic superiority include the large voltage change at the transition from the first to the second
phase or hyperpolarization of tissue and reactivation of sodium channels during the initial phase, with
resultant tissue conditioning that allows the second phase to more readily excite the myocardium.
Antitachycardia pacing and cardioversion are not uniformly successful. There is some incidence of
ventricular arrhythmia acceleration with antitachycardia pacing and cardioversion, and it is also not
unusual for cardioversion to induce atrial fibrillation that in turn triggers unwarranted therapies. An
ideal therapeutic solution would be one capable of preventing the occurrence of tachycardia altogether.
Prevention techniques have been investigated, among them the use of precisely timed subthreshold
stimuli, simultaneous stimulation at multiple sites, and pacing with elevated energies at the site of the
tachycardia, but none has yet proven practical.
The rudimentary VVI antibradycardia pacing provided by current defibrillators lacks rate responsive-
ness and atrial pacing capability. Consequently, some defibrillator patients require implantation of a
separate dual-chamber pacemaker for hemodynamic support. It is inevitable that future generations of
defibrillators will offer dual-chamber pacing capabilities.
Atrial fibrillation, occurring either as a consequence of defibrillator operation or as a natural progres-
sion in many defibrillator patients, is a major therapeutic challenge. It is certainly possible to adapt
implantable defibrillator technology to treat atrial fibrillation, but the challenge is to do so without
causing the patient undue discomfort. Biphasic waveform defibrillation of acutely induced atrial fibril-
lation has been demonstrated in humans with an 80% success rate at 0.4 J using epicardial electrodes.
Stand-alone atrial defibrillators are in development, and, if they are successful, it is likely that this
capability would be integrated into the mainstream ventricular defibrillators as well. However, most
conscious patients find shocks above 0.5 J to be very unpleasant, and it remains to be demonstrated that
a clinically acceptable energy level will be efficacious when applied with transvenous electrode systems
to spontaneously occurring atrial fibrillation. Moreover, a stand-alone atrial defibrillator either must
deliver an atrial shock with complete assurance of appropriate synchronization to ventricular activity or
must restrict the therapeutic energy delivery to atrial structures in order to prevent inadvertent induction
of a malignant ventricular arrhythmia.

80.5 Implantable Monitoring


Until recently, defibrillator data recording capabilities were quite limited, making it difficult to verify the
adequacy of arrhythmia detection and therapy settings. The latest devices record electrograms and
diagnostic channel data showing device behavior during multiple tachyarrhythmia episodes. These
devices also include counters (number of events detected, success and failure of each programmed therapy,
and so on) that present a broad, though less specific, overview of device behavior (Fig. 80.3). Monitoring
capability in some of the newest devices appears to be the equivalent of 32 Kbytes of random access
memory, allowing electrogram waveform records of approximately 2-min duration, with some oppor-
tunity for later expansion by judicious selection of sampling rates and data compression techniques.
Electrogram storage has proven useful for documenting false therapy delivery due to atrial fibrillation,

© 2000 by CRC Press LLC


FIGURE 80.3 Typical data recorded by an implantable defibrillator include stored intracardiac electrograms with
annotated markers indicating cardiac intervals, paced and sensed events, and device classification of events (TF =
fast tachycardia; TP = antitachy pacing stimulus; VS = sensed nontachy ventricular event). In the example, five rapid
pacing pulses convert a ventricular tachycardia with a cycle length of 340 ms into sinus rhythm with a cycle length
of 830 ms. In the lower portion of the figure is an example of the summary data collected by the ICD, showing
detailed counts of the performance of the various therapies (Rx) for ventricular tachycardia (VT), fast ventricular
(VTF), and ventricular (VF). (Copyright Medtronic, Inc. Used with permission.)

lead fractures, and sinus tachycardia, determining the triggers of arrhythmias; documenting rhythm
accelerations in response to therapies; and demonstrating appropriate device behavior when treating
asymptomatic rhythms.
Electrograms provide useful information by themselves, yet they cannot indicate how the device
interpreted cardiac activity. Increasingly, electrogram records are being supplemented with event markers
that indicate how the device is responding on a beat-by-beat basis. These records can include measure-
ments of the sensed and paced intervals, indication as to the specific detection zone an event falls in,
indication of charge initiation, and other device performance data.

80.6 Follow-up
Defibrillator patients and their devices require careful follow-up. In one study of 241 ICD patients with
epicardial lead systems, 53% of the patients experienced one or more complications during an average
exposure of 24 months. These complications included infection requiring device removal in 5%, post-
operative respiratory complications in 11%, postoperative bleeding and/or thrombosis in 4%, lead system
migration or disruption in 8%, and documented inappropriate therapy delivery, most commonly due to
atrial fibrillation, in 22%. A shorter study of eighty patients with transvenous defibrillator systems
reported no postoperative pulmonary complications, transient nerve injury (1%), asymptomatic subcla-
vian vein occlusion (2.5%), pericardial effusion (1%), subcutaneous patch pocket hematoma (5%), pulse
© 2000 by CRC Press LLC
generator pocket infection (1%), lead fracture (1%), and lead system dislodgement (10%). During a
mean follow-up period of 11 months, 7.5% of the patients in this series experienced inappropriate therapy
delivery, half for atrial fibrillation and the rest for sinus tachycardia.
Although routine follow-up can be accomplished in the clinic, detection and analysis of transient
events depends on the recording capabilities available in the devices or on the use of various external
monitoring equipment.

80.7 Economics
The annual cost of ICD therapy is dropping as a consequence of better longevity and simpler implantation
techniques. Early generators that lacked programmability, antibradycardia pacing capability, and event
recording had 62% survival at 18 months and 2% at 30 months. Some recent programmable designs
that include VVI pacing capability and considerable event storage exhibit 96.8% survival at 48 months.
It has been estimated that an increase in generator longevity from 2–5 years would lower the cost per
life-year saved by 55% in a hypothetical patient population with a 3-year sudden mortality of 28%. More
efficient energy conversion circuits and finer line-width integrated circuit technology with smaller, more
highly integrated circuits and reduced current drains will yield longer-lasting defibrillators while con-
tinuing the evolution to smaller volumes.
Cost of the implantation procedure is clearly declining as transvenous lead systems become common-
place. Total hospitalization duration, complication rates, and use of costly hospital operating rooms and
intensive care facilities all are reduced, providing significant financial benefits. One study reported
requiring half the intensive care unit time and a reduction in total hospitalization from 26 to 15 days
when comparing transvenous to epicardial approaches. Another center reported a mean hospitalization
stay of 6 days for patients receiving transvenous defibrillation systems.
Increasing sophistication of the implantable defibrillators paradoxically contributes to cost efficacy.
Incorporation of single-chamber brady pacing capability eliminates the cost of a separate pacemaker and
lead for those patients who need one. Eventually even dual-chamber pacing capability will be available.
Programmable detection and therapy features obviate the need for device replacement that was required
when fixed parameter devices proved to be inappropriately specified or too inflexible to adapt to a patient’s
physiologic changes.
Significant cost savings may be obtained by better patient selection criteria and processes, obviating
the need for extensive hospitalization and costly electrophysiologic studies prior to device implantation
in some patient groups. One frequently discussed issue is the prophylactic role that implantable defibril-
lators will or should play. Unless a means is found to build far less expensive devices that can be placed
with minimal time and facilities, the life-saving yield for prophylactic defibrillators will have to be high
if they are to be cost-effective. This remains an open issue.

80.8 Conclusion
The implantable defibrillator is now an established and powerful therapeutic tool. The transition to
pectoral implants with biphasic waveforms and efficient yet simple transvenous lead systems is simplifying
the implant procedure and drastically reducing the number of unpleasant VF inductions required to
demonstrate adequate system performance These advances are making the implantable defibrillator easier
to use, less costly, and more acceptable to patients and their physicians.

Acknowledgment
Portions of this text are derived from Duffin EG, Barold SS. 1994. Implantable cardioverter-defibrillators:
An overview and future directions, Chapter 28 of I Singer (ed), Implantable Cardioverter-Defibrillator,
and are used with permission of Futura Publishing Company, Inc.

© 2000 by CRC Press LLC


References
Josephson M, Wellens H (eds). 1992. Tachycardias: Mechanisms and Management. Mount Kisco, NY,
Futura Publishing.
Kappenberger L, Lindemans F (eds). 1992. Practical Aspects of Staged Therapy Defibrillators. Mount
Kisco, NY, Futura Publishing.
Singer I (ed). 1994. Implantable Cardioverter-Defibrillator. Mount Kisco, NY, Futura Publishing.
Tacker W (ed). 1994. Defibrillation of the Heart: ICD’s, AED’s, and Manual. St. Louis, Mosby.
PACE, 14:865. (Memorial issue on implantable defibrillators honoring Michel Mirowski.)

© 2000 by CRC Press LLC

You might also like