DSPEC0514 Pulse Oxymetre MD300 C15D Mode D Emploi en
DSPEC0514 Pulse Oxymetre MD300 C15D Mode D Emploi en
v. 28/06/2019 1
D.SPEC.05.14 Pulse oxymètre MD300-C15D - Mode d'emploi_EN
Oxycure concentrators
v. 28/06/2019 2
D.SPEC.05.14 Pulse oxymètre MD300-C15D - Mode d'emploi_EN
http://www.oxycure.be
e-mail : [email protected]
OPENING HOURS
From Monday to Friday 09.00 – 12.30 and 13.30 – 18.00
Notes
1. Technical interventions (repairs) are included in the price of the assistance of the leasing.
2. Deliveries of goods, outside the opening hours, will be charged.
v. 28/06/2019 3
D.SPEC.05.14 Pulse oxymètre MD300-C15D - Mode d'emploi_EN
Pulse Oximeter
USER MANUAL
Ver3.0C1/C4
General Description
Oxygen Saturation is a percentage of Oxyhemoglobin (HbO2) capacity, compounded with oxygen,
by all combinative hemoglobin (Hb) capacity in blood. In other words, it is consistency of
Oxyhemoglobin in blood. It is a very important parameter for the Respiratory Circulation System.
Many respiratory diseases can result in oxygen saturation being lowered in human blood.
Additionally, the following factors can reduce oxygen saturation: Automatic regulation of organ
dysfunction caused by Anesthesia, Intensive Postoperative Trauma, injuries caused by some
medical examinations. That situation might result in light-headedness, asthenia, and vomiting.
Therefore, it is very important to know the oxygen saturation of a patient so that doctors can find
problems in a timely manner.
The fingertip pulse oximeter features low power consumption, convenient operation and portability.
Place one fingertip into the photoelectric sensor for diagnosis and the pulse rate and oxygen
saturation will appear on the display. It has been proven in clinical experiments that it also features
high precision and repeatability.
Measurement Principle
Principle of the oximeter is as follows: A mathematical formula is established making use of Lambert
Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin (RHb) and
Oxyhemoglobin (HbO2) in red and near-infrared zones. Operation principle of the instrument:
Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse
Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm
red and 905nm near infrared light) can be focused onto a human nail tip through a clamping
finger-type sensor. A measured signal obtained by a photosensitive element, will be shown on the
oximeter’s display through process in electronic circuits and microprocessor.
Diagram of Operation Principle
1. Red and Infrared-ray Emission Tube
2. Red and Infrared-ray Receipt Tube
Precautions For Use
1. Before use, carefully read the manual.
2. Operation of the fingertip pulse oximeter may be affected by the use of an electrosurgical unit
(ESU).
3. The fingertip pulse oximeter must be able to measure the pulse properly to obtain an accurate
SpO2 measurement. Verify that nothing is hindering the pulse measurement before relying on
the SpO2 measurement.
4. Do not use the fingertip pulse oximeter in an MRI or CT environment.
5. Do not use the fingertip pulse oximeter in situations where alarms are required. The device has
no alarms. It is not for continuous monitoring.
6. Do not use the fingertip pulse oximeter in an explosive atmosphere.
7. The fingertip pulse oximeter is intended only as an adjunct in patient assessment. It must be
used in conjunction with other methods of assessing clinical signs and symptoms.
8. In order to ensure correct sensor alignment and skin integrity, the maximum application time at
a single site for our device should be less than half an hour.
9. Do not sterilize the device using autoclaving, ethylene oxide sterilizing, or immersing the device
v. 28/06/2019 4
D.SPEC.05.14 Pulse oxymètre MD300-C15D - Mode d'emploi_EN
Operation Instructions
1. Install two AAA batteries according to the Battery Installation instructions.
2. Place one of your fingers into the rubber opening of the pulse oximeter.
3. Press the switch button one time on front panel to turn the pulse oximeter on.
4. Keep your hands still for the reading. Do not shake your finger during the test. It is
recommended that you do not move your body while taking a reading.
5. Read the data from the display screen. There are two display modes. After turning on the pulse
oximeter, each time you press the power switch, the pulse oximeter will switch to another
display modes. (NOTE: only for the MD300C4 series)
Front Panel
The pulse bar less than 30% indicates signal inadequacy and the displayed SpO2 and pulse rate
value is potentially incorrect.
Battery Installation
1. Install two AAA batteries into the battery compartment. Match the plus (+) and minus (-) signs
in the compartment. If the polarities are not matched, damage may be caused to the oximeter.
2. Slide the battery door cover horizontally along the arrow shown as the picture.
Note:
Please remove the batteries if the pulse oximeter will not be used for long periods of time.
Please replace the battery when the power indicator starting flickering.
3. Remove the batteries if the oximeter is not operated for a long time.
4. It is best to store the product in -20℃~+55℃ and ≤93% humidity.
5. Keep in a dry place. Extreme moisture may affect oximeter lifetime and may cause damage.
6. Dispose of battery properly; follow any applicable local battery disposal laws.
Cleaning the fingertip pulse oximeter
Please use medical alcohol to clean the silicone touching the finger inside of oximeter with a soft
cloth dampened with 70% isopropyl alcohol. Also clean the being tested finger using alcohol before
and after each test.
Do not pour or spray liquids onto the oximeter, and do not allow any liquid to enter any openings in
the device. Allow the oximeter to dry thoroughly before reuse.
The fingertip pulse oximeter requires no routine calibration or maintenance other than replacement
of batteries.
The use life of the device is five years when it is used for 15 measurements every day
and 10 minutes per one measurement. Stop using and contact local service center if
one of the following cases occurs:
An error in the Possible Problems and solutions is displayed on screen.
The oximeter cannot be powered on in any case and not the reasons of battery.
There is a crack on the oximeter or damage on the display resulting readings cannot be identified;
the spring is invalid; or the key is unresponsive or unavailable.
Specifications
1. Display Type
LED display
2. SpO2
Display range: 0%~100%
Measurement range: 70%~100%
Accuracy: 70%~100%±2digits; 0%~69% no definition
Resolution: 1%
ARMS Value Analysis
Item 70--100 90--100 80--<90 70--<80
#pts 231 82 89 60
Bias 1.10 0.49 1.35 1.62
ARMS 1.68 1.09 1.77 2.14
Bland-Altman plot analysis of sampled data points on all subjects as below
A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor.
Clinical testing is used to establish the SpO2 accuracy. The measured arterial hemoglobin saturation
value (SpO2) of the sensors is compared to arterial hemoglobin oxygen (SaO2) value, determined
from blood samples with a laboratory CO-oximeter. The accuracy of the sensors in comparison to
the CO-oximeter samples measured over the SpO2 range of 70%~100%. Accuracy data is
calculated using the root-mean-squared (Arms value) for all subjects, per ISO 9919:2005, Medical
Electrical Equipment–Particular requirements for the basic safety and essential performance of
pulse oximeter equipment for medical use.
A functional tester is used to measure how accurately Fingertip Pulse Oximeter is reproducing the
v. 28/06/2019 7
D.SPEC.05.14 Pulse oxymètre MD300-C15D - Mode d'emploi_EN
8. Classification
According to the type of protection against electric shock: INTERNALLY POWERED EQUIPMENT;
According to the degree of protection against electric shock: TYPE BF APPLIED PART, (applied part:
the rubber hole of the device);
According to the degree of protection against ingress of water: IPX1
According to the mode of operation: CONTINUOUS OPERATION
Declaration
Guidance and Manufacturer’s declaration – electromagnetic emissions-
For all EQUIPMENT and SYSTEMS
Guidance and Manufacturer’s declaration - electromagnetic emission
The MD300C1 Pulse Oximeter is intended for use in the electromagnetic environment specified
below. The customer or the user of MD300C1 Pulse Oximeter should assure that it is used in such
an environment.
Emission test Compliance Electromagnetic Environment – guidance
RF emissions CISPR 11 Group 1 The MD300C1 Pulse Oximeter uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
v. 28/06/2019 8
D.SPEC.05.14 Pulse oxymètre MD300-C15D - Mode d'emploi_EN
electronic equipment.
RF emissions CISPR 11 Class B The pulse Oximeter (MD300C1) is suitable for use in all
Harmonic emissions Not establishments, including domestic establishments and
IEC 61000-3-2 Applicable those directly connected to the public low-voltage power
Voltage fluctuations/ Not supply network that supplies buildings used for domestic
flicker emissions Applicable purposes.
IEC 61000-3-3
Guidance and Manufacturer’s declaration – electromagnetic immunity-
For all EQUIPMENT and SYSTEMS
Guidance and Manufacturer’s declaration - electromagnetic immunity
The MD300C1 Pulse Oximeter is intended for use in the electromagnetic environment specified
below. The customer or the user of the MD300C1 Pulse Oximeter should assure that it is used in
such an environment.
Immunity test IEC 60601 Compliance Electromagnetic Environment –
test level Level guidance
Electrostatic +/- 6kV +/- 6kV Floors should be wood, concrete or
Discharge (ESD) contact contact ceramic tile. If floor are covered with
IEC 61000-4-2 +/- 8kV air +/- 8kV air synthetic material, the relative humidity
should be at least 30%.
Power frequency (50/60 3A/m 3A/m Power frequency magnetic fields should
Hz) magnetic field be at levels characteristics of a typical
IEC 61000-4-8 location in a typical commercial or
hospital environment.
Guidance and Manufacturer’s declaration – electromagnetic immunity-
For all EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and Manufacturer’s declaration - electromagnetic immunity
The MD300C1 Pulse Oximeter is intended for use in the electromagnetic environment specified
below. The customer or the user of the MD300C1 Pulse Oximeter should assure that it is used in
such an environment.
Immunity IEC 60601 Compliance Electromagnetic Environment – guidance
test test level Level
Radiated 3 V/m 3 V/m Portable and mobile RF communications equipment
RF 80 MHz to should be used no closer to any part of the Pulse
IEC 2.5 GHz Oximeter (MD300C1), including cables, than the
61000-4-3 recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
v. 28/06/2019 9
D.SPEC.05.14 Pulse oxymètre MD300-C15D - Mode d'emploi_EN
v. 28/06/2019 11
D.SPEC.05.14 Pulse oxymètre MD300-C15D - Mode d'emploi_EN
Box Content
1. Fingertip pulse oximeter
2. One lanyard
3. Two AAA batteries
4. One instruction manual
Applicable Models
MD300C1 series:
LED screen
MD300C1 MD300C12 MD300C13 MD300C15 MD300C16 MD300C17 MD300C19
MD300C1A MD300C1B MD300C1C MD300C1D MD300C1E MD300C1F MD300C1G
MD300C1H MD300C1I
LCD screen
MD300C15D MD300C15F MD300C150
MD300C4 series:
MD300C4 MD300C41 MD300C42 MD300C11
Notes:
1. The illustrations used in this manual may differ slightly from the appearance of the
actual product.
2. The specifications are subject to change without prior notice.
v. 28/06/2019 12