0% found this document useful (0 votes)
3 views1 page

2. FIT12019 Exercise Answer

This document is an instrument data sheet for a flow transmitter (DP) used in a project for PT XYZ in Bali. It includes specifications such as service conditions, instrument range, materials, and performance characteristics. The transmitter is designed for measuring saturated steam with specific parameters for flow, temperature, and pressure.

Uploaded by

Adi Kurdi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
3 views1 page

2. FIT12019 Exercise Answer

This document is an instrument data sheet for a flow transmitter (DP) used in a project for PT XYZ in Bali. It includes specifications such as service conditions, instrument range, materials, and performance characteristics. The transmitter is designed for measuring saturated steam with specific parameters for flow, temperature, and pressure.

Uploaded by

Adi Kurdi
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

BOOT CAMP EXERCISES

INSTRUMENT DATA SHEET NUMBER


Flow Transmitter - DP REV:
CLIENT : PT XYZ SITE: BALI
PROJECT NO : PROJECT XYZ PROJECT NO.: PROJECT
MAKE : PT XXZ REQUISITION NO.: 001
SUPPLIER : VENDOR ABC ORDER NO.: 001
REV
1 INSTRUMENT TAG NUMBER / DESCRIPTION FT12019
2 E.L.D. NO./EQUIPMENT or LINE NO. P&ID001
3 INSTRUMENT FUNCTION 250-STLS1-12762-BG55C-H70
4 FLUID NAME / STATE SATURATED STEAM
5 FLUID NAME / STATE (UPPER)
6 SERVICE CONDITIONS MAXIMUM NORMAL MINIMUM REMARKS
7 FLOW kg/h 5000 3000
8 TEMPERATURE °C 179 179
9 PRESSURE kPa a 900 900
10 DIFFERENTIAL PRESSURE kPa 25 25
11 LEVEL
12 DENSITY kg/m³ 5.153 5.153
13 DENSITY UPPER kg/m³ 5.153 5.153
14 VISCOSITY mPa s 0.015 0.015
15 VISCOSITY UPPER mPa s 0.015 0.015
16 OTHER PROPERTIES
17 DESIGN PRESSURE MPa g 16
18 DESIGN TEMPERATURE °C 120
19 VESSEL NOZZLE LOCATIONS: HIGH PRESSURE FLANGE TAPS
20 LOW PRESSURE FLANGE TAPS
21 TRANSMITTER EJX110A-EMS4G-91DEB
22 Model No. DP Flow Measurement
23 Service (Gauge/Absolute/Level/Flow/DP) Diaphragm
24 Element Type -100/100
25 Instrument Range kPa 0 - 25
26 Calibrated Range kPa 4-20 mA HART (Sqrt with DP) 24VDC (Loop Powered)
27 Output Signal / Power Supply Accuracy: +/-0.04% of span Repeatability: +/-0.04% of span
28 Performance Yes - Manufacturer Standard
29 Zero Suppression/Elevation Yes - Digital
30 Integral Indicator 2" pipe mount (304SS brackets and bolting)
31 Mounting N/A
32 Diaphragm Housing Extension TBC
33 Vessel Test Pressures High pressure: ½" NPTF
34 Process Connection Low pressure: ½" NPTF
35 Wetted parts: 316 Stainless Steel Enclosure/Body: Coated Aluminium
36 Materials : Capsule diaphragm: Hastelloy C-276 Bellows: N/A
37 Seals & gaskets: Teflon Capsule fluid: Fluorinated Oil
38 3-way s.s. manifold : 3 Way manifold Vent / drain valve: 316SS
39 Options : (AGCO) AX3TVIS-4-SSC
40 N/A - No remote diaphragm
41 REMOTE DIAPHRAGM Model No. Diaphragm: N/A Fill Fluid: N/A
42 Materials: Fill Fluid Density : N/A
43 Capillary Material: N/A Capillary length: N/A
44 Non Hazardous
45 Hazardous Area Classification N/A IP65
46 Explosion Protection Method / Ingress Protection N/A
47 Hazardous Area Certificate No. Screw Terminals M20 x 1.5
48 Electrical connections / Cable Entry
49 NOTES:
50
51
52
53
54
55
56
57 * denotes vendor to advise
58
59
60
61
62
63
64
65 REVISION :
66 STATUS:
67 DATE:
68 PREPARED BY:
69 CHECKED BY:
70 AUTHORISED BY:
71

You might also like