FUS 060 Handbook
FUS 060 Handbook
SITRANS F US
Ultrasonic flowmeter
Transmitter type FUS060
for use with sensor type SONO 3100,
SONO 3300 and SONOKIT 1- and 2-tracks
[ ]
English: http://www4.ad.siemens.de/WW/view/en/10806951/133300
SFIDK.PS.029.E2.02 A5E01204521
System overview This operating manual is for the transmitter part of the flowmeter consisting of sensors type SONO 3100,
SITRANS F US SONO 3300 and SONOKIT. These sensors have separate operating manuals.
FUS060 with
SONO 3100, Sensor Mild steel Stainless steel
SONOKIT Standard program mild steel
and SONO 3300 Extended program
SONO 3300 SONO 3100 SONOKIT
Connection EN 1092-1
examples ANSI B 16.5 150 lbs RF None
ANSI B 16.5 300 lbs RF
Approvals
ATEX ATEX
Transmitter FUS060
Approval ATEX
2 SFIDK.PS.029.E2.02
Contents 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1 Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 Precision measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.4 Literature overwiew . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.5 System overwiew . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.6 Technical data transmitter SITRANS FUS060 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.7 Dimensional drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.8.1 Principle of measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.8.2 Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3 Sitrans FUS060 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.1 Installing Sitrans FUS060 (only remote version) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2.1 Wall assembly (standard included) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2.2 Pipe assembly (accessory) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.3 Connecting the sensor cable to the transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.4 Mounting the cables into SITRANS FUS060 connection module . . . . . . . . . . . . . . . . . . . . 12
3.5 Installation of the transducer type SONO 3200 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.5.1 Electrical connection - transmitter to transducers (1-track) . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.5.2 Electrical connection - transmitter to transducers (2-tracks) . . . . . . . . . . . . . . . . . . . . . . . . 13
3.5.3 Electrical connection - allowed cable radius . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.6 Rotating the operating and display panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.7 Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4 Start up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.1 Electrical data connection - Ex-version (preliminary) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2 Operation (HART communication) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
5 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.1 Keypad and display lay-out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.2 Operating concept of the operating and display panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.3 Menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
5.4 Disabling or enabling access to settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.5 Operating examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
5.6 Setting option and factory setting in the display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
7 Trouble shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
9 Certificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
9.1 EC Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
9.2 ATEX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
SFIDK.PS.029.E2.02 3
Introduction 1
1.1 Preface
These instructions contain all the information required to commission and use the
SITRANS F US ultrasonic flowmeter transmitter type FUS060 with sensor type SONO 3100,
SONO 3300, SONOKIT 1- and 2-tracks.
1.2 Introduction
This manual only describes the FUS060 transmitter installation. The transmitter FUS060 with
a sensor (SONO 3100, SONO 3300 or SONOKIT) is designed for measuring the flow velocity
of liquids in full pipes. Satisfactory function of the ultrasonic flowmeter depends on a low sound
attenuation of the medium and a well-defined and stable flow profile.
The installation of an ultrasonic flowmeter system with transmitter type FUS060 is divided into
5 steps:
Siemens Flow Instruments SITRANS F US ultrasonic flowmeters are designed for measurement
of:
SITRANS F US ultrasonic flowmeters measure flow in standard volumetric and mass flow units.
Measurement is independent of changes in liquid temperature, density, pressure and conductivity.
This ultrasonic flowmeter is designed for use on single phase liquids.
Potential Hazards
Both transmitter and sensor must be grounded for optimal performance.
4 SFIDK.PS.029.E2.02
1.3 Precision measuring system
The ultrasonic flowmeter is a precision measuring system that is “user friendly”, but must be
installed in accordance with the instructions given in this handbook.
This operating manual is only for the transmitter part of the flowmeter system consisting of sensor
type SONO 3100, SONO 3300 or SONOKIT. These sensors have separate manuals.
SFIDK.PS.029.E2.02 5
1.6 Technical data transmitter SITRANS FUS060
Output
Analog output
Signal range 4...20 mA
Upper limit 20...22.5 mA, adjustable
Signal on alarm 3.6 mA, 22 mA or 24 mA
Load Max. 600 Ω; ≥ 230 Ω for HART communication
Only PROFIBUS PA version Analog output omitted, is replaced by digital PROFIBUS PA
interface
Digital output 1 - PROFIBUS PA and HART
Active or passive signal, can be configured Active: 24 V DC, ≤ 24 mA, Ri = 300 Ω (see following table)
with positive or negative logic Passive: open collector, 30 V DC, ≤ 200 mA (see following table)
For explosion protection with HART version Passive: open collector, 30 V DC, ≤ 100 mA (see following table)
ATEX II 2G Ex dem [ia/ib] T6/T4/T3
For explosion protection with PROFIBUS Only passive signals for digital output 1 (see following table)
ATEX II 2G Ex dem [ia/ib] T6/T4/T3
Output function, configurable Pulse output
Adjustable pulse significance ≤ 5,000 pulses/s
Adjustable pulse width ≥ 0.1 ms
Frequency output
fEND selectable up to 10 kHz
Error in measurement:
(at reference conditions)
Pulse output ≤ ± 0.5% of measured value at 0.5 m/s to 10 m/s or
≤ ± 0.25 / V [m/s]% of measured value at flow < 0.5 m/s
Analog output As pulse output plus ± 0.1% ± 20 µA
Repeatability ≤ ± 0.25% of measured value at 0.5 m/s to 10 m/s
Digital output 2 (Relay) - only HART version
Relay, NC or NO contact Switching capacity max. 5 W
Max. 50 V DC, max. DC 200 mA
Self-resetting fuse, Ri = 9 Ω
For explosion protection with HART version Max. 30 V DC, for DC max. 100 mA, for AC max. 50 mA
ATEX II 2G Ex dem [ia/ib] IIC T6/T4/T3 (cf. EC-Type Examination certificate)
Output function, configurable for: Alarm, flow direction, flow limits, sound velocity limits,
ultrasonic-amplitude limits, no function
Communication via analog output 4...20 mA
PC/laptop (PDM) or HART communicator
with SITRANS F flowmeter
Load with connection of coupling module 230...500 Ω
Load with connection of HART communicator 230...600 Ω
Cable 2-wire shield ≤ 3 km (≤ 1.86 miles)
Multi-core shield ≤ 1.5 km (≤ 0.93 miles)
Protocol HART, version 5.1
Communication via PROFIBUS PA interface Layers 1 + 2 according to PROFIBUS PA
Communication system according to IEC 1158-2
Layer 7 (protocol layer) according to PROFIBUS DP,
EN 50170 standard
Power supply Separate supply (four-wire device)
Permissible bus voltage 9...32 V
See certificates and approvals
Current consumption from bus 10 mA; ≤ 15 mA in event of error with electronic current
limiting
Electrical isolation Outputs electrically isolated from power supply and
from one another
6 SFIDK.PS.029.E2.02
Accuracy Typically depening on accuracy of installation conditions
Ambient temperature (transmitter) -20...+50 °C (-4...+122 °F)
In potentially explosive atmospheres Observe temperature class
Storage temperature -25...+80 °C (-13...+176 °F)
Enclosure rating IP65
Electromagnetic compatibility For use in industrial environments
Emission EN 61000-6-3 (Light industry)
Immunity EN 61000-6-2 (Industry)
Use shield cables for the outputs
Design
Transmitter only as separate version Transmitter is connected to the transducers via 3 ... 120 m
(9,8 ... 395 ft) long coaxial cables.
For ATEX versions mounted in the Ex zone only with 3 m
long cables.
Housing material Die-cast aluminum, painted
Electrical connection Cable glands: 2x M20 or 2x ½“-NPT; (M25 PROFIBUS-vers.)
Transducers: 2/4x M16 or 2/4x ½" NPT
Displays and controls
Display LCD, two lines with 16 characters each (backlight)
Multi-display: 2 freely-selectable values Flow, volume, mass flow, mass, flow velocity, speed of sound,
are displayed simultaneously in two lines ultrasonic signal information, current, frequency
Operation 4 infrared keys
Hierarchical menu prompting with codes
Power supply
Supply voltage 120...230 V AC ± 15% (50/60 Hz)
19...30 V DC/ 21...26 V AC
Power failure No effect for at least 1 period (> 20 ms)
Power consumption Approx. 8 VA / 6 W
Reference conditions
Process temperature 25 °C ± 5 °K (77 °F ± 9 °F)
Ambient temperature 25 °C ± 5 °K (77 °F ± 9 °F)
Installation conditions Upstream section > 20x DN and
downstream section > 5x DN
Warming-up time 30 min.
Certificates and approvals Explosion protection ATEX II 2G Ex dem [ia/ib] IIC T6/T4/T3
Digital output 1 - List of examples for output voltages and different loads
(frequency or pulse output; Profibus PA or HART version and electrical connection no. 5 and 6)
Note
For full performance at 10 KHz frequency on digital output 1, the capacitive load should not exceed 100 nF.
SFIDK.PS.029.E2.02 7
1.7 Dimensional drawings
8 SFIDK.PS.029.E2.02
1.8.1 Principle of measurement
A sound wave traveling in the same direction as the liquid flow arrives at point B from point A in
a shorter time than the sound wave traveling against the direction of flow (from point B to A).
The difference in sound transit time indicates the flow velocity in the pipe.
Since delay time is measured at short intervals both in and against flow direction, viscosity and
temperature have no influence on measurement accuracy.
In SITRANS F US flowmeters the two ultrasonic transducers are placed at an angle θ in relation
to the pipe axis. The transducers function as transmitters and receivers of the ultrasonic signals.
Measurement is performed by determining the time the ultrasonic signal takes to travel with and
against the flow. The principle can be expressed as follows:
This measuring principle offers the advantage that it is independent of variations in the actual
sound velocity of the liquid, i.e. independent of the temperature. Proportional factor K is
determined by wet calibration or “Auto” (SONOKIT only) in case of manual programming of
transducer angle, transducer distance, transducer displacement and pipe dimensions.
SFIDK.PS.029.E2.02 9
General safety instructions 2
• Installation, connection, commissioning and service must be carried out by personnel who
are qualified and authorized to do so.
• It is very important for any person working with the equipment to read and understand the
instructions and directions provided in this manual and follow instructions and directions
before using the equipment.
• Only personnel authorized and trained by the owner of the equipment may operate the
equipment.
• Installation personnel must ensure that the measuring system is correctly connected in
accordance with the connection diagram in this manual.
• For applications involving high working pressures or media that can be dangerous to
people, surroundings, equipment or other in the event of pipe fracture, Siemens recommends
taking precautions such as special placement, shielding or installation of a safety guard or
safety valve prior to installation of the sensor.
• Repair and service should be performed by approved Siemens Flow Instruments personnel
only.
Warning means that death, severe personal injury and/or substantial damage to property can
occur if the appropriate safety precautions are not observed.
) Note
is important information on the product itself, the handling of the product or the respective part
of the instruction manual to which particular attention should be paid.
10 SFIDK.PS.029.E2.02
SITRANS FUS060 3
3.2 Installation
3.2.1 Wall assembly (standard included)
The mounting bracket is suitable for wall assembly
only.
SFIDK.PS.029.E2.02 11
3.3 Connecting the sensor cables to the transmitter
When opening the lid you will have direct access
for checking the correct transducer cable con-
nections.
The cables to be connected to the transducers according to the schematics 3.5.1 or 3.5.2.
Enter the cable length in the transmitter.
Make sure the cables are equal length in order to avoid different signal delays in the signal
processing.
12 SFIDK.PS.029.E2.02
3.5 Installation of the transducer type SONO 3200
The flowmeter system consist of a sensor pipe and 4 ultrasonic transducers type SONO 3200.
The terminal box is fastened to the transducer holder by means of a union. Stress on the
connection wires must be avoided when passing them through the terminal box. The wire without
a spade terminal is connected to position 2. The wire with a spade terminal is connected to the
transducer housing by means of position 3. The inner wire of the transducer coax cable is
connected to position 1 and the shield wire coax cable is fixed to cable clip in position 4.
Note
When disassembling the terminal box the wires must be loosened from the connection board.
It is not necessary to remove the transducer cable.
Support the terminal box while loosening the union. Then remove the terminal box from the
transducer holder.
Warning
Turning the terminal box can damage the wires to the crystals.
Flow direction
Flow direction
SFIDK.PS.029.E2.02 13
3.5.3 Electrical connection - allowed cable radius
Lid lock
14 SFIDK.PS.029.E2.02
3.7 Warning
The pertinent regulations must be observed for electrical installation, in explosion risk areas in
particular:
• The rules and regulations governing the erection and operation of plants in explosion risks
areas (e.g. in Germany: Elex V, VDE 0165).
• The EC Type Examination Certificate - Do not install the device under voltage. Danger of
electric shock! -The version for power supply DC 19 to 30 V may only be connected to SELV
or PELV circuits.
The SITRANS FUS060 transmitter comply with the following electrical installation categories:
• Version for 115 to 230 V AC: Installation Category II
• Version for 24 V AC/DC: Installation Category II
General requirements:
• The power supply should be connected via a circuit breaker and fused via a fuse
(max. 4 A).
• Use cables with a cross section of at least 1.5 mm2 and double or reinforced insulation for
the power supply.
• Lay heat-resistant cables if high temperatures can occur on the housing, e.g. due to
conduction of heat by the sensor/metering tube. Lay the cables so that they do not come into
contact with the hot sensor/metering tube.
• Lay signal cables separately from cables with voltages > 60 V.
• Use cables with twisted wire pairs.
• Full HART 5.1 specifications only with shielded cables.
• Earth the transmitter housing.
SFIDK.PS.029.E2.02 15
PE connector - between the
cable glands (please connect
this PE connector)
Connection
In devices with protection type ATEX II 2G Ex dem [ia/ib] T6/T4/T3 the intrinsically safe circuits
must be separated from non-intrinsically safe circuits by certified safety isolators or barriers.
For reliable communication via the HART-protocol at least one load of 230 Ω must be available
in the signal circuit.
The FUS060 enclosure must be connected with a earth wire to the PE connector - see note in
figure above.
Warning
• If the device is energized (power supply, digital outputs in the case of external feeding) the
housing covers may not be unscrewed in areas where there is a risk of explosion.
• The lid over the terminals for the power supply may not be removed! Use only certified
measuring instruments!
16 SFIDK.PS.029.E2.02
Start up 4
The device is ready for operation immediately after switching on the power supply.
Be aware that a number of factory settings (e.g. max volume flow, angle of tracks, distance
between transducers, displacement of transducer from center line of pipe) are dimension
dependent. The settings are stored from factory in the transmitter for SONO 3100
and SONO 3300. For SONOKIT and spare part transmitters please enter values manually.
If there is still gas/air in the pipe (metering tube) or in the pipeline after assembly, a “flashing
F or D” can appear at the top right of the first line in the display. The failure signal is output at
the analog output.
Analog output
HART-version 4 to 20 mA Intrinsic safety
Connector 7 and 8 Ex ia IIC/IIB resp. Ex ib IIC/IIB
Max. values:
Uo = 15.8 V
Io = 64 mA
Po = 253 mW
Ri = 250 Ω
Linear characteristic
IIC IIB
Lo 0 0.5 2.5 8 0 1 5 30 mH
Co 400 2) 2) 0 2.8 2) 2) 0 nF
1) 1)
Max. values:
Ui = 23 V
Ii = 40mA
Pi = 70 mW
Ci = 24 nF
Li = negligible
SFIDK.PS.029.E2.02 17
Digital output 1
HART-version Intrinsic safety Ex ia IIC/IIB resp. Ex ib IIC/IIB
Connector 5 and 6
Max. values:
Uo = 7.2 V
Io = 2.2 mA
Po = 4 mW
Linear characteristic
Co = 105 nF
Lo = 1H
Max. values:
Ui = 30 V
Ii = 100 mA
Pi = 750 mW
Ci = 1 nF
Li = 108 uH
Digital ouput 1
PROFIBUS-PA-version Intrinsic safety Ex ia IIC/IIB resp. Ex ib IIC/IIB
Connector 5 and 6
Max. values:
Uo = 28.4 V
Io = 2.6 mA
Po = 19 mW
Linear characteristic
Co = 78 nF
Lo = 1H
Max. values:
Ui = 30 V
Ii = 100 mA
Pi = 750 mW
Ci = 1 nF
Li = 100 uH
18 SFIDK.PS.029.E2.02
Digital output 2 (relay output)
only HART-version Intrinsic safety Ex ia IIC/IIB resp. Ex ib IIC/IIB
Connector 3 and 4 only for connection to a certified intrinsic
current loop.
Max. values:
Ui = 30 V
Ii = 100 mA DC
Ii = 50 mA AC
Ci = 24 nF
Li = 73 uH
PA-ouput
PROFIBUS-PA-version FISCO DEVICE
Connector 7 and 8 Intrinsic safety Ex ia IIC/IIB resp. Ex ib IIC/IIB
only for connection to a certified intrinsic
current loop (e.g. FISCO-isolating power
supply) with following max. values:
FNICO DEVICE
Intrinsic safety Ex ic IIC/IIB resp. Ex nL IIB/IIC
resp. only for connection to a certificated
intrinsic current loop. (e.g. FNICO or
FISCO-isolating power supply) with following
max. values:
SFIDK.PS.029.E2.02 19
4.2 Operation (HART communication)
The device can be operated with:
• operating and display panel
• HART-Communicator
• PC/Laptop and software SIMATIC PDM
Warning
Do not use the coupling module in hazardous areas and do not connect to intrinsically safe
circuits.
20 SFIDK.PS.029.E2.02
Commissioning 5
Display (LCD)
2 lines à 16 characters backlight
Optical keypad
Infrared key
The individual functions and parameters for further groups are combined within these main
groups.
SFIDK.PS.029.E2.02 21
5.3 Menu structure
22 SFIDK.PS.029.E2.02
The available device functions and parameters are shown in the second line of the display.
Selection takes place with the and keys. These are shown in the second line of the
display. Enter the device function or the setting level of the parameters with
key (Enter function). It is possible to exit the selected function or setting level without
storage of the change by use key until the cursor key move to the last left position.
The currently valid setting appears after selecting the setting level of a parameter. If
programming is enabled the programmable value flashes in the second line of the display. The
parameter setting can be changed. There are two different types of data input:
• direct numerical input
• input from given table
In the numerical input, the and keys have the function of a cursor control. The selected
digit flashes. With the and keys the desired digit is set or the decimal point moved to the
right or left. Digits are input with carry over to the next highest place. After selecting the last digit,
the input is confirmed by use the key. The entered value is accepted if it is within the
permissible input range. In this case the user guidance returns to the selection menu for the
parameters of the group concerned. If the entered value is rejected, an error message briefly
appears on the display and then the previous setting. The value can be changed again.
Note
• If the and keys are operated permanently (finger left on the glass panel) the numeric
value or setting option is changed continuously in tabular selection. A numeric input carries
over to the next highest position.
• If an accidentally changed setting is wrong, it is possible to exit the menu item by actuating
the key several times (return to the next menu level up).
Note
• The programming is always enabled with code = 0 (factory setting).
SFIDK.PS.029.E2.02 23
5.5 Operating examples
Example 1
The physical unit for the flow value should be changed from m3/h to l/min. The starting point is
the multi-display.
24 SFIDK.PS.029.E2.02
Example 2
Change the pulse unit from m3/Imp to l/Imp and the pulse value from 10 m3/Imp to 10 l/Imp.
SFIDK.PS.029.E2.02 25
5.6 Setting option and factory setting in the display
Only menu items relevant for end users and simple service cases are listed in this table.
Menu items
Menu Device function, Description Factory setting Setting options Read or HART PROFIBUS
number parameter write
1. Display
1.1 Display parameter Setting of the actual display Read
parameter readings
1.1.1 Language Language in the display English english Write
deutsch
francais
español
1.1.2 Line 1 Parameter value in line 1 Flow Flow Write √ √
Total. net
Total. forward
Total. reverse
Flow velocity
Sound velocity
US-Amplitude
Analog output (HART only)
Frequency
1.1.3 Line 2 Parameter value in line 2 Total. net Flow Write √ √
Total. net
Total. forward
Total. reverse
Flow velocity
Sound velocity
US-Amplitude
Analog output (HART only)
Frequency
1.1.4 Display Flow Select presentation Physical Unit Physical units Write √ √
%
Diagram (Bar in %)
1.1.5 LCD lighting Switch LCD on/off (automatic Off On Write √ √
switch off after 10 min.) Off
1.1.6 Multi display Display of two measured Line 1: Flow Menu 1.1.2 Write √ √
values in line 2 Line 2: Totalizer Menu 1.1.3
1.2 Flow
1.2 Flow Flow in metering tube Actual metering value For unit see menu 3.1.1 Write √ √
1.3 Totalizer
1.3.1 Total. forward Totalizer forward Actual value Units selectable in Write √ √
menu 3.4.1
1.3.2 Total. reverse Totalizer reverse Actual value Units selectable in Write √ √
menu 3.5.1
1.3.3 Total net Totalizer difference Actual value Units selectable in Write √ √
Forward - reverse menu 3.6.1
1.3.4 Settings (all) All totalizers will be set to zero. No action No action Write √ √
(Independent setting of tota- Reset + stop
lizers in menu 3.4, 3.5, 3.6) Reset + start
1.4 Flow velocity
1.4 Flow velocity Flow velocity in metering tube Actual value m/s (range -12 to +12 m/s) Read √ √
1.5 Sound velocity
1.5 Sound velocity Sound velocity in the medium Actual value m/s Read √ √
1.6 Amplitude US
1.6 Amplitude US Relative ultrasonic amplitude Actual value Display in % Read √ √
of the received ultrasonic sig-
nals (reference water).
Depending on tracks activated
(up to 4 values).
1.7 Frequency
1.7 Frequency Frequency output value Actual value Hz Read √ √
26 SFIDK.PS.029.E2.02
Menu Device function, Description Factory setting Setting options Read or HART PROFIBUS
number parameter write
1.8 Current output
1.8 Current output Current value of the analogue Actual value mA Read √ √
output
2. Diagnostics
2.1 Device status
2.1 Device status Error message “OK” only read, see table in Read √ √
the chapter 6.2
2.2 Suppress error
2.2 Suppress error Switch off error message with None None Write √ √
(not stored) effect on: Unreliable flow
Analog output Measuring path
Digital output 1 disturbed
Digital output 2 Both
2.3 Device test
2.3.1 Self test Test device status Read √ √
2.3.2 Display test Visual LCD test Read √ √
2.4 Simulation
2.4.1 Flow Simulation of the flow Read
Value Simulation value. 0% -110…. + 110% Write
Affects all outputs and tota-
lizers.
Time Duration of simulation. End (no action) End (no action) Write
After expiration of this time the 10 min
normal measuring mode is 30 min
resumed. 60 min
2.4.2 Digital Simulation of the output signal End (no simulation) End (no simulation) Write
output 1 for digital output 1 0.1 Hz √ √
1 Hz
10 Hz
100 Hz
1 kHz
10 kHz
Alarm on
Alarm off
2.4.3 Digital Simulation of the output signal End (no simulation) End (no simulation) Write √
output 2 (only for digital output 2 (relay) Alarm on
HART version) Alarm off
2.4.4 Analog output Simulation of the output cur- 4 mA 4 mA Write √
(only HART rent of the analogue output 10 mA
version) 12 mA
20 mA
Error signal
3. Meas.
Function
3.1 Flow
3.1.1 Unit Physical units for volume flow, m3/h l/s, l/min, l/h, m3/s, Write √ √
mass flow m³/min, m³/h, m³/d,
Note: Whenever using mass Ml/d, ft³/s, ft³/min, ft³/h,
flow - actual density have to ft³/d, gal/s, gal/min,
be selected in menu 3.1.4 gal/h, gal/d, Mgal/d,
ImpGal/s, ImpGal/min,
ImpGal/h, ImpGal/d,
g/s, g/min, kg/s, kg/min,
kg/h, kg/d, ton/min,
ton/h, ton/d, lb/s,
lb/min, lb/h, lb/d,
LTon/d , STon/min,
STon/h, STon/d
3.1.2 Max. vol. flow Full scale value Depends on the See dimension table Write √ √
Note: Corresponds to 20 mA sensor size in sensor manual
and selected max. frequency
3.1.3 Limits Flow unit according to Read
menu 3.1.1
3.1.3.1 Lo alarm limit Lower alarm limit (the lower Depends on the See dimension table Write √ √
limit must be lower than the sensor size in sensor manual
upper limit)
SFIDK.PS.029.E2.02 27
Menu Device function, Description Factory setting Setting options Read or HART PROFIBUS
number parameter write
3.1.3.2 Hi alarm limit High alarm limit (the higher Depends on the See dimension table Write √ √
limit must be higher than sensor size in sensor manual
the lower limit )
3.1.3.3 Hysteresis Hysteresis for the limit in % 1% 0%...20% Write √
of the full scale value
Hysteresis for the limit in Depends on the 0…Q max/5 (20%) Write √
selected unit sensor size
3.1.4 Density Density of the media Read
Units Density unit kg/m³ g/cm³ Write √ √
g/l
kg/l
kg/m³
lb/ft³
lb/in³
lb/gal
lb/lmpgal
Density Input of the density value for + 1000.00 kg/m 3 200…5000 kg/m3 Write √ √
the arithmetic calculation of
the mass flow
3.1.5 Direction Read
3.1.5.1 Direction Determination of the main + Direction + Direction Write √ √
direction of flow (forward - Direction
direction)
3.1.5.2 Measurement Choise of measurement Forward only Forward only Write √ √
direction direction Forward + reverse
3.1.5.3 Hysteresis Setting the hysteresis for the 1% 0...20% Write √ √
flow direction related to the
full scale value
3.1.6 Low flow cut Determination of the switching 1% 0…20% Write √
point for the low flow cut-off.
Suppresion is related to the
full scale value
Determination of the switching Depends on the 0…Qmax/5 (20%) Write √
point for the low flow cut-off. sensor size
Suppresion is an absolute
value related to the sensor size
3.1.6 Flow damping Read
3.1.7 Time constant Selection of time constants for 5.0 sec 0.0…200 sec Write √ √
measuring variables
3.2 Sound velocity
3.2.1 Dynamic range Unit for the sound velocity m/s Read
3.2.2.1 Lower value Lower range value (LRV) for +600 m/s +200…upper value Write √ √
the sound velocity
3.2.2.2 Upper value Upper range value (URV) for +2000 m/s Lower value…2000 m/s Write √ √
the sound velocity
3.2.3.1 Lo alarm limit Lower alarm limit +200 m/s 200…2000 m/s Write √ √
3.2.3.2 High alarm limit Upper alarm limit +2000 m/s 200…2000 m/s Write √ √
3.2.3.3 Hysteresis Hysteresis for limit 5% 0…20% Write √
Hysteresis for limit +14 m/s 0…100 m/s Write √
3.2.4 Time constant Selection of time constant for 5s 0…200 s Write √ √
measuring variable
3.3 Amplitude US
3.3.1 Dynamic range Read
3.3.1.1 Lower value Lower range value for 1% 1% …upper range Write √ √
ultrasonic amplitude
3.3.1.2 Upper value Upper range value for 100% Lower range…150% Write √ √
ultrasonic amplitude
3.3.2 Limits Read
3.3.2.1 Lo alarm limit Lower alarm limit 1% 1%…150% Write √ √
3.3.2.2 Hi alarm limit Higher alarm limit 120% 1%…150% Write √ √
3.3.2.3 Hysteresis Hysteresis for limit 1% 0…20% Write √ √
3.3.3 Time Constant Filter time constant +10.00 s 0…200 s Write √ √
28 SFIDK.PS.029.E2.02
Menu Device function, Description Factory setting Setting options Read or HART PROFIBUS
number parameter write
3.4 Totalizer forward
3.4.1 Unit Physical volume unit or mass m³ l Write √ √
unit. hl
Note: Whenever using mass m³
flow the actual density have Ml
to be selected in menu 3.1.4 ft³
Gal
MGal
ImpGal
MImpGal
g
kg
Ton
lb
3.4.2 Hi alarm limit Upper limit of alarm +1.000.000.000 m³ 0-1.000.000.000 Write √ √
3.4.3 Setting forward Volume totalizer reset to “0” No action No action Write
and “stop/start” Reset + stop √ √
Reset + start
3.5 Total. reverse
3.5.1 Unit Physical volume unit or mass m³ l Write √ √
unit. hl
Note: Whenever using mass m³
flow the actual density have to Ml
be selected in menu 3.1.4 ft³
Gal
MGal
ImpGal
MImpGal
g
kg
Ton
lb
3.5.2 Lo alarm limit Value for the lower limit of -1.000.000.000 m³ 0-1.000.000.000 Write √ √
alarms
3.5.3 Setting reverse Volume totalizer reset to “0” No action No action Write √ √
and “stop/start” Reset + stop
Reset + start
3.6 Forward - reverse
volume
3.6.1 Total. net unit Physical volume unit or mass m³ l Write √ √
unit. hl
Note: Whenever using mass m3
flow the actual density have to Ml
be selected in menu 3.1.4 ft³
Gal
MGal
ImpGal
MImpGal
g
kg
Ton
lb
3.6.2 Lo Alarm Limit Value for the lower alarm limit -1.000.000.000 m³ 0-1.000.000.000 Write √ √
3.6.3 Hi Alarm Limit Value for the upper alarm limit +1.000.000.000 0-1.000.000.000 Write √ √
3.6.4 Hysteresis Hysteresis for the limits 0 0…3 Write √ √
3.6.5 Settings net Volume totalizer reset to “0” No action No action Write √ √
and “stop/start” Reset + stop
Reset + start
4. Device outputs
4.1 Analog output (only for HART version)
4.1.1 Analog output Assignment of a measured Flow Flow Write √
variable to the output signal Sound velocity
US-amplitude
SFIDK.PS.029.E2.02 29
Menu Device function, Description Factory setting Setting options Read or HART PROFIBUS
number parameter write
4.1.2 Current limit Upper current limit 20.5 mA 20.0 to 22.5 mA Write √
4.1.3 Error signal Output current in case of error. 3.6 mA 3.6 mA Write √
Use of “hold” in...s means 22 mA
current will be kept for x 24 mA
seconds. Hold: 5s, 20s, 40s, 60s,
120s, 240s,
hold continuous
4.1.4 Current output Actual current of the analog Read √
output in mA depending on
choice in menu 4.1.1
4.1.5 Flow range % Actual measurement in % of 0% Read √
selection in menu 4.1.1
4.1 Profibus PA version
4.1.1 Bus adress Profibus address 126 0…126 Read √
4.1.2 Ident-Nr Write √
4.1.3 Min Tsdr Min. “stationary delay” of the 11 bits Write √
responder
4.2 Digital output 1 Frequency or impulse signal
for flow or alarm signal
4.2.1 Function Assignment of output function Pulse (HART only) Pulse (HART only) Write √ √
Frequency
Alarm
Flow dir. forward
Flow min
Flow max
Flow min/max
Total forw. max
Total rev. min
Total net. min
Total net max
No function
4.2.2 Signal type Configure output: Passive-pos. Active-pos (HART only) Write √ √
Signal: active or passive Active-neg (HART only)
Logic: positive or negative Passive-pos
Passive-neg
4.2.3 Pulse value Read
Unit Physical unit per impulse m3/Imp l/Imp Write √ √
m³/Imp
Gal/Imp
ImpGal/Imp
kg/Imp
t/Imp
lb/Imp
Pulse rate Number of volume units per 1 unit/pulse 0…1000.0 Write √ √
impulse
4.2.4 Pulse width Setting the pulse width 0.10 ms 0.1…2000 ms Write √ √
(only active with “impulse“
selection in menu 4.2.1)
4.2.5 Full scale freq. Full scale frequency 10000 Hz 2…10 000 Hz Write √ √
4.3 Digital output 2 (only HART version)
(relay)
4.3.1 Function Assignment of output function Alarm Alarm Write √
Flow dir. forward
Flow min/max
Flow min
Flow max
Min/max US-vel
Max US-vel
Min US-vel
Min/max US-ampl
Max US-ampl
Min US-ampl
No function
4.3.2 Signal type Logic of signal output in case Contact opens Contact closes Write √
of function in menu 4.3.1 Contact opens
30 SFIDK.PS.029.E2.02
Menu Device function, Description Factory setting Setting options Read or HART PROFIBUS
number parameter write
5. Identification
5.1 Unit identity This unit
5.1.1 Tag Tag number on metering unit To be defined by user HART: max 8 charac. Write √ √
PA: max 32 charac.
5.1.2 Descriptor Tag number description To be defined by user HART: max 16 charac. Write √ √
PA: max 32 charac.
5.1.3 Message Messages to tag To be defined by user HART: max 32 charac. Write √ √
PA: max 32 charac.
5.2 Manuf. Ident.
5.2.1 Product type Order number Read only Factory set √ √
(MFLB order number) Read user
5.2.2 Serial number Serial number Read only Factory set √ √
Read user
5.2.3 Software rev. Software revision Read only Factory set √ √
Read user
5.2.4 Manuf. date Manufactoring date Read only Factory set √ √
DD.MM.YYYY Read user
5.2.5 Nominal size Sensor size Read only Factory set √ √
Read user
5.3 Device ID True identification in HART 30 Factory set √
(HART version) long address Read user
5.3 Device ID True identification 33113 (HEX 8159) Factory set √
(Profibus version) Read user
6. Service
6.1 Enter code Entering 4 digit code entered 0 0…9999 Read
in menu 6.2 in order to change According to setting in
user parameters menu 6.2
6.2 Customer Code Selection of a personal code. 0 0…9999 Write √ √
Code 0: User parameter are
not protected.
Code>0: User parameters are
protected under menu 6.1
6.3 Service code Only for service in the Read
Siemens factory √ √
6.4 Reset Re-start of unit without change Cancel Cancel Read √ √
of parameters
(Factory setting) Reset
6.5 Control values
6.5.1 Gain Gains of the single tracks 0-255 Read √ √
6.5.1.1 Gain up 1 Gain of the first track up 0-255 Read √ √
6.5.1.2 Gain dw 1 Gain of the first track down 0-255 Read √ √
6.5.1.3 Gain up 2 Gain of the second track up 0-255 Read √ √
6.5.1.4 Gain dw 2 Gain of the second track down 0-255 Read √ √
6.5.1.5 Gain up 3 Gain of the third track up 0-255 Read √ √
6.5.1.6 Gain dw 3 Gain of the third track down 0-255 Read √ √
6.5.1.7 Gain up 4 Gain of the fourth track up 0-255 Read √ √
6.5.1.8 Gain dw 4 Gain of the fourth track down 0-255 Read √ √
6.5.2 Amplitudes Amplitudes of the single tracks Read √ √
6.5.2.1 Ampl up 1 Amplitude of the first track up 0-255 Read √ √
optimal: 95-105
6.5.2.2 Ampl dw 1 Amplitude of the first track 0-255 Read √ √
down optimal: 95-105
6.5.2.3 Ampl up 2 Amplitude of the second track 0-255 Read √ √
up optimal: 95-105
6.5.2.4 Ampl dw 2 Amplitude of the second track 0-255 Read √ √
down optimal: 95-105
6.5.2.5 Ampl up 3 Amplitude of the third track up 0-255 Read √ √
optimal: 95-105
6.5.2.6 Ampl dw 3 Amplitude of the third track 0-255 Read √ √
down optimal: 95-105
6.5.2.7 Ampl up 4 Amplitude of the fourth track 0-255 Read √ √
up optimal: 95-105
SFIDK.PS.029.E2.02 31
Menu Device function, Description Factory setting Setting options Read or HART PROFIBUS
number parameter write
6.5.2.8 Ampl dw 4 Amplitude of the fourth track 0-255 Read √ √
down optimal: 95-105
6.5.3 Trigger/Level Trigger-level of the individual Read √ √
tracks
6.5.3.1 Trig up 1 Trigger-level of the first track 0-255 Read √ √
up
6.5.3.2 Trig dw 1 Trigger-level of the first track 0-255 Read √ √
down
6.5.3.3 Trig up 2 Trigger-level of the second 0-255 Read √ √
track up
6.5.3.4 Trig dw 2 Trigger-level of the second 0-255 Read √ √
track down
6.5.3.5 Trig up 3 Trigger-level of the third track 0-255 Read √ √
up
6.5.3.6 Trig dw 3 Trigger-level of the third track 0-255 Read √ √
down
6.5.3.7 Trig up 4 Trigger-level of the fourth track 0-255 Read √ √
up
6.5.3.8 Trig dw 4 Trigger-level of the fourth track 0-255 Read √ √
down
6.5.4 Error count % Read √ √
6.5.4.1 Error 1 Error totalizer in % for track 1 0-100% Read √ √
optimal: 0%
6.5.4.2 Error 2 Error totalizer in % for track 2 0-100% Read √ √
optimal: 0%
6.5.4.3 Error 3 Error totalizer in % for track 3 0-100% Read √ √
optimal: 0%
6.5.4.4 Error 4 Error totalizer in % for track 4 0-100% Read √ √
optimal: 0%
6.5.5 TOF up TIME OF FLIGHT (TOF) up- Read √ √
stream of the individual tracks
6.5.5.1 TOF up 1 TIME OF FLIGHT (TOF) of the Value in ns Read √ √
first track upstream
6.5.5.2 TOF up 2 TIME OF FLIGHT (TOF) of the Value in ns Read √ √
second track upstream
6.5.5.3 TOF up 3 TIME OF FLIGHT (TOF) of the Value in ns Read √ √
third track upstream
6.5.5.4 TOF up 4 TIME OF FLIGHT (TOF) of the Value in ns Read √ √
fourth track upstream
6.5.6 TOF down TIME OF FLIGHT (TOF) down- Read √ √
stream of the individual tracks
6.5.6.1 TOF down 1 TIME OF FLIGHT (TOF) of the Value in ns Read √ √
first track downstream
6.5.6.2 TOF down 2 TIME OF FLIGHT (TOF) of the Value in ns Read √ √
second track downstream
6.5.6.3 TOF down3 TIME OF FLIGHT (TOF) of the Value in ns Read √ √
third track downstream
6.5.6.4 TOF down 4 TIME OF FLIGHT (TOF) of the Value in ns Read √ √
fourth track downstream
6.5.7 Delta TOF TIME OF FLIGHT (TOF) Read
difference
6.5.7.1 Delta TOF 1 TIME OF FLIGHT (TOF) Value in ps Read √ √
difference of the first track
(up – down)
6.5.7.2 Delta TOF 2 TIME OF FLIGHT (TOF) Value in ps Read √ √
difference of the second track
(up – down)
6.5.7.3 Delta TOF 3 TIME OF FLIGHT (TOF) Value in ps Read √ √
difference of the third track
(up – down)
6.5.7.4 Delta TOF 4 TIME OF FLIGHT (TOF) Value in ps Read √ √
difference of the fourth track
(up – down)
32 SFIDK.PS.029.E2.02
Menu Device function, Description Factory setting Setting options Read or HART PROFIBUS
number parameter write
6.6 Zero Trim Zero point values for track
1 to track 4
6.6.1 Zero Pt.Offset Zero offset compensation Read
value for each track
6.6.1.1 Zr.Pt.Offset 1 Zero offset of the first track +x.xxx ns -50.000 ns ... +50.000 ns Write √ √
6.6.1.2 Zr.Pt.Offset 2 Zero offset of the second track +x.xxx ns -50.000 ns ... +50.000 ns Write √ √
6.6.1.3 Zr.Pt.Offset 3 Zero offset of the third track +x.xxx ns -50.000 ns ... +50.000 ns Write √ √
6.6.1.4 Zr.Pt.Offset 4 Zero offset of the fourth track +x.xxx ns -50.000 ns ... +50.000 ns Write √ √
6.6.2 Stand.Dev. Standard deviation value for Read
each track
6.6.2.1 Stand.Dev.1 Standart deviation of the track 1 +x.xxx ns Write √ √
6.6.2.2 Stand.Dev.2 Standart deviation of the track 2 +x.xxx ns Write √ √
6.6.2.3 Stand.Dev.3 Standart deviation of the track 3 +x.xxx ns Write √ √
6.6.2.4 Stand.Dev.4 Standart deviation of the track 4 +x.xxx ns Write √ √
6.6.3 Zero Trim Initiates an device specific Shows as “Zero status”: Write √ √
“Start zero tr.” adjustment cycle (zero trim ”start”, “running” and
processing), that determines “finished”
the true zero point value.
Only during no-flow process
conditions!
6.6.4 Undo zero tr. Undo the last zero trim After choice shows Write √ √
“Zero Trim”: “OK”
7. Sensor Param.
7.1 Sensor calib Sensor calibration Read
7.1.1 Calib choice Calibration choice Auto Wet Write √ √
Auto
7.1.2 App. param. The application parameters Read
7.1.2.1 Viscosity The viscosity of media 0.01 cm2/s 0.005…5 cm2/s Write √ √
7.1.2.2 Correct. fact. Gain compensation value for 1.0 0.000001…100 Write √ √
the flow sensor
7.1.2.3 Fl. offset comp. Offset compensation value for 0.0 (unit) Write √ √
the flow sensor
(in flow unit) (in the menu
3.1.1 closed unit)
7.1.2.4 Cable len. Total cable length for a track. Dependent on selection 0...500 m Write √ √
TOFKOR All tracks to have the same
length
7.1.3 Pipe geometry The pipe geometry Read
7.1.3.1 Pipe diameter The inside diameter of the pipe Dependent on selection 0.01…4.0 m Write √ √
7.1.3.2 Roughness The roughness of the inner 0.4 mm 0.01…10.0 mm Write √ √
pipe wall
7.1.4 Tracks The settings of individual Read
tracks
7.1.4.1 Track 1 Setting of track 1 Read √ √
7.1.4.1.1 Length 1 Length of track 1 Dependent of sensor >0 min value...8.0 m Write √ √
size
7.1.4.2.1 Length 2 Length of track 2 Dependent of sensor >0 min value...8.0 m Write √ √
size
SFIDK.PS.029.E2.02 33
Menu Device function, Description Factory setting Setting options Read or HART PROFIBUS
number parameter write
7.1.4.2.3 Displacement 2 The displacement of track 2 Dependent of sensor 0...1.5 m Write √ √
size
7.1.4.3.1 Length 3 Length of track 3 Dependent of sensor >0 min value...8.0 m Write √ √
size
7.1.4.4.1 Length 4 Length of track 4 Dependent of sensor >0 min value...8.0 m Write √ √
size
7.2 No. of tracks The number of active tracks 1-track or 2-tracks 1-track Write √ √
in the pipe 2-tracks
3-tracks
4-tracks
34 SFIDK.PS.029.E2.02
Device functions and parameters 6
The following sections describe only some of the device function parameters in the table that
need further explanations.
In the “1.3 Totalizer“ menu the total flow quantity is displayed. The display appears as a volume
or mass depending on the selection volume unit/mass unit in menu 3.4, 3.5, 3.6.
In the “3.4.3”, “3.5.3”, “3.6.5”, “Totalizer reset” menu the totalizer can be reset to zero - either with
stop of the totalizer or reset and on-going totalizer.
SFIDK.PS.029.E2.02 35
6.2 Main menu: Diagnostics (menu 2)
Device status, menu 2.1
2.1 Device status Error message “Ok” See failure table “Error message” below
The test-routines are systematically executed during the normal operation. The error is
indicated by a blinking notice on the display:
• D: Device error
• F: Process error
The error is described by the variable status_display (menu 2.1). The error messages and their
assignment to the display notice, analog output, and digital outputs are noted in the table.
36 SFIDK.PS.029.E2.02
6.3 Main menu: Measuring functions (menu 3)
The range and limit values can be displayed as follows:
Flow select (volume unit per time) or mass flow (mass unit per time).
The volume flow qv is converted to mass flow qm with the density entered (qm = qv * ρ in menu
“Density”).
Depending on the selection made, the quantities in the menu “1.3 Totalizer” are specified in
the function group “Display”.
Flow, sound velocity and amplitude US: Menu 3.1, 3.2 and 3.3
The measuring variables flow, sound velocity or ultrasonic amplitude can be assigned to the
analog output and the limit value message for digital output 2. The appropriate measuring
ranges and limit values are set in the subsequent sub-menus.
Every sub-menu is further divided for setting the subsequent values.
– Measuring range full scale value
– Measuring range start of scale value (omitted for flow, start of scale = 0)
– Limit value max (upper limit value)
– Limit value min (lower limit value)
– Hysteresis
– Damping (Filter time constant)
The selection made here defines the unit for several parameters simultaneously:
– for volume or mass flow: Measured value display, measuring range and limit value formation
– for volume or mass: Display, totalizer reading and pulse valency.
Hysteresis
The density of the medium is entered in this menu (step 2) to convert the volume flow to mass
flow (see also menu “3.1.1 Units”).
Note
Input of the density is unnecessary and ineffective when the “Volume flow” setting has been
selected in menu “3.1.1 units “.
SFIDK.PS.029.E2.02 37
Measurement direction, menu 3.1.5.2
SITRANS FUS060 is able to measure in both flow directions (forward and reverse) or in forward
direction only. The selection is made by the setting.
– forward + reverse
– forward only
If “forward” is parameterized, signals are only output or accumulated internally for a flow in this
direction.
Note
The low flow cut off suppression has no influence on signaling of the flow direction.
38 SFIDK.PS.029.E2.02
6.4 Main menu: Device output (menu 4)
The analog output is only active for the device with HART interface.
There are three types of measurements which can be assigned to the analog output:
• Flow
• Sound velocity
• US Amplitude
Current limiting
With the “Hold 5 s” setting, brief disturbances (e.g. air entrapped in the medium) can be bridged
for about 5 seconds without an error message via the analog signal.
The same applies for “Hold 20 s”, “Hold 40 s” etc.
The last valid measured value is output during the disturbance. If the disturbance lasts less than
5 s the current measured value is output at the end of the disturbance. If the disturbance persists
(e.g. disturbance lasts longer than 5 s), the output signal is set to 3.6 mA.
In the “Hold continuous” setting, the last valid value is output for as long as the disturbance
persists, helpful in the configuration is the possibility of selectively suppressing error messages,
see menu 2.2.
SFIDK.PS.029.E2.02 39
Digital output 1 menu 4.2
Function menu 4.2.1
The digital output 1 (implemented for both HART and PA interface) can indicate flow (impulse
or frequency) and some alarms/limits.
Volume/Impulse
Analog Input Function
Volume [m3]
Block 1 - Flow
Volume/Frequency
phys_dout1_select
In the device version with protection type ATEX II 2G Ex dem [ia/ib] IIC T6/T4/T3 with intrinsically
safe output signals:
Active
Passive
40 SFIDK.PS.029.E2.02
Signals with positive and negative logic can be generated (positive or negative pulses).
Figure explains the setting options.
Pulse output
The pulses are converted in a volume/impulse or mass/impulse block. There are two
parameters to be set for the impulse output in the menu which are used in this block:
• Pulse rate, menu 4.2.3: For example this value determines how many cubic meters have
to flow through to make one impulse. By setting of this value you can change the unit (unit/
impulses) to easier set the pulse volume.
• Pulse width, menu 4.2.4: The impulse width can be set in the range from 0.1 to 2000 ms.
The maximal impulse width is computed according to the set flow upper value.
The maximal frequency for impulse output is 5 kHz.
SFIDK.PS.029.E2.02 41
Function, menu 4.3.1
The digital output 2 (relay contact) can be configured for reporting the device status (device
fault, measuring error), the flow direction or a limit for flow, US velocity, US amplitude.
Note
The relay contact is open when “no function” is set.
Contact closes
Contact opens
Switching status of the relay
Digital output 2
The electronic fuse is tripped in the event of overloading. The recovery time of the fuse is a few
minutes.
42 SFIDK.PS.029.E2.02
6.5 Main menu: Identification (menu 5)
Unit identity, menu 5.1
Device-specific or tag-specific data can be called or entered in this menu.
Tag number, menu 5.1.1
Tag description, menu item 5.1.2
Tag message, menu item 5.1.3
Tag-specific data can be entered in these menu items. The and keys take on the function
of a cursor control here. The numbers, letters and text characters are selected with the and
keys.
If no text is stored, the end of text-character < flashes in the first segment when you enter these
menu items. By actuating the and key, the end of text-character moves one position to
the right and one character can be selected from the character set. The following characters
are available:
Output
The selected character is entered by actuating the key and the end of text character flashes
again. Another character can be selected with the and key. The text input is ended by
actuating the key when the end of text character flashes.
For texts longer than 16 characters the marks < and > in the first and last segment indicate that
there are further characters to the left and/or right of the displayed text section. These are
displayed by actuating the and keys.
Deleting texts: Select the end of text character from the character set with the and key
and then actuate the key. All entries to the right of the end of text character are deleted.
Note
If the and keys are actuated continuously (finger continuously touching the glass panel),
the characters are automatically scrolled.
SFIDK.PS.029.E2.02 43
6.6 Main menu: Service (menu 6)
Enter code, menu 6.1
Programming of the device by unauthorized persons can be disabled with a customer specific
code (range 0...9999) in the operating and display panel. By entering the appropriate code in
this menu item, the programming is enabled and the device settings can be changed.
- Customer code (from factory the customer code is set to “0”).
Note
- If the or key is actuated in the parameter setting level with the programming disabled,
the request for “Enter code” appears in the device display (when user code #0).
- The programming is disabled after returning to the display mode after 10 min (when user
code #0).
- Programming can be purposely disabled by entering any number, not the code number, in
this menu item (when user code #0).
44 SFIDK.PS.029.E2.02
Gain, menu 6.5.1
Every transducer gain - level can be read out.
These parameters are read only values in the range 0…255.
For SONOKIT flowmeter systems - measured values of the sensor pipe from the “ SONOKIT
- SENSOR GEOMETRY MEASURING REPORT” are used for programming the transmitter
FUS060.
Important
From factory all values having influence of the calibration date have been programmed with
nominal default values corresponding to ordered device.
Therefore new measurement data must to be entered into SITRANS FUS060 menu structure
item 7.1.
As we can see in the menu 7.1.1 - calibration choice, the switching between WET and AUTO
have a big influence to the calibration and calculation of the flow value.
When AUTO is selected the calibration constants will be taken from geometrical data entered
from measuring report.
With automatical calibration, there are several more input parameters, which will be combined
and calculated and have an output as the calibration factor. These formula output values for track
1, 2, 3, 4 are then read only parameters which are accessible via PDM only.
SFIDK.PS.029.E2.02 45
SONOKIT uses “AUTO” to determine the calibration constants based on input of:
- Pipe diameter
- Length from transducer front to transducer front of every track used
- Displacement of transducer from the center of pipe
- Average angle of every track used
- Length of used cable in one sound track
- Roughness
- Viscosity
Note
Automatic calibration can only be used for ordinary round shape pipes.
Nice to know:
When WET is selected, the calibration parameters will be taken from the factory calibration
process. These parameters will be stored in special WET calibration parameters (e. g. SONO
3100 and SONO 3300 sensors).
Note
Please use AUTO calibration mode for all SONOKIT systems.
Note
This menu 7.1.3.x is not for flowmeter systems with sensors SONO 3100 and SONO 3300.
Please don’t change these values. They are stored from factory side.
For SONOKIT: Data to be entered from “SENSOR GEOMETRY MEASURING REPORT”
46 SFIDK.PS.029.E2.02
Tracks, menu 7.1.4
This menu is for the parameters which are for the track settings. In the following table only the
parameters for track 1 are explained.
The parameters for track 2, 3, 4 will be handled like track 1.
Note
The following menus 7.1.4.1.x.x and 7.2 are not for flowmeter systems with sensors
SONO 3100 and SONO 3300. Please don’t change these values. They are stored from
factory side.
7.1.4.1.1 Length 1 The length of the first track 0.2 m >0 m ... 8.0 m
This parameter is the distance between the ultrasonic transducers, like illustrated in the picture.
The value will be necessary to calculate the ultrasonic velocity from the time of flight and the
distance between the transducers.
Note
These procedure is not for flowmeter systems with FUS060 transmitter who are
delivered or calibrated together with sensors SONO 3100 and SONO 3300. Please don’t
change these values. They are stored from factory side.
The following procedure describes the necessary parameters/data to be entered and the
sequence of entering them in order to configure the transmitter correctly for a given pipe
applications:
The pipe data are taken from the “SONOKIT” measurement report or from another information
/ data sheet of the exsisting senor / calibration report.
Measurement:
1. Enter the sensor diameter (menu 7.1.3.1).
2. Choose appropriate engineering unit for flow (menu 3.1.1).
3. Enter maximum volume flow (menu 3.1.2).
4. Choose “AUTO” for calibration choice (menu 7.1.1).
5. Enter the viscosity for the media, otherwise the default value (0.01 cm2/s for water 20 °C)
will be used (menu 7.1.2.1).
6. Enter roughness of the inner surface of the sensor pipe, otherwise the default value (0.4
mm) will be used (menu 7.1.3.2) (see also chapter 8.7).
7. Enter the measured transducer distance for each track (menu 7.1.4.1.1 (track 1) and menu
7.1.4.2.1 (track 2)).
8. Enter the measured angle for each track (menu 7.1.4.1.2, (track 1) and menu 7.1.4.2.2
(track 2)).
9. Enter the measured displacement for each track (menu 7.1.4.1.3, (track 1) and menu
7.1.4.2.3 (track 2)).
10. Check the number of tracks (menu 7.2). They are pre-set from the factory according to
ordered SONOKIT).
11. Go to the service menu and check that the control values for amplitude (menu 6.5.2.X) have
a value within the interval 90 – 110.
12. Go to the service menu and check that the control values for the error counters (menu
6.5.4.X) are 0.
13. Go to the service menu and check that the control values for time-of-flight (menu 6.5.5.X)
has a stable value that may vary in the nanosecond range (for example ±10ns).
14. Make sure that 0 flow conditions are applied and thereafter the 0 point calibration procedure
is activated by use of menu 6.6.3 (“Zero trim”).
15. The calculation of the flow can be corrected by use of the customer correction factor (menu
7.1.2.2) which is a scaling factor.
Based on the entered data the transmitter is capable of measuring and calculating the actual
flow. The accuracy of the system at this stage is dependent on the entered accuracy of the
geometry data.
48 SFIDK.PS.029.E2.02
Trouble shooting 7
Measuring path disturbed The medium in the measuring line is not permeable to
sound, this applies both when the pipe is filled with gas or
is empty. The measured value on the display is set to zero.
Cable is broken detached.
Flow measurement unreliable The measured values have a very high dispersion so that
errors in the measuring evaluation or disturbances in the
flow due to cavitation, twist or inhomogenity such as
bubbles or foreign bodies.
Measured values are displayed but an “F” flashes in the
right hand corner of the display.
Time-of-flight (menu 6.5.5.X Permanent disturbance inside If possible, make sure that the
and 6.5.6.X) for the active the pipe transducer window inside the
tracks is not stable pipe is not damaged or that no
permanent disturbance is present.
Error counters (menu 6.5.4.X)
not equal to 0
Flow not stable. A flashing Air bubbles or particles of any Make sure that the pipe line is
“F” appears in the display kind disturbing the measurement ventilated and that the concentra-
continuously/occasionally. tion of particles is limited to a level
Device status (menu 2.1): where normal measurement with
“Unreliable flow value” the meter is possible.
Gain (menus 6.5.1.X) for the Cable partly connected Make sure that the transducer
active tracks < 250 but is not cable is connected in the terminal
stable housing of the transmitter.
Error counters (menu 6.5.4.X) Ultrasonic echoes Make sure that the entered track
not equal to 0 lengths correspond with the
measurement report.
SFIDK.PS.029.E2.02 49
Error/Symptoms Diagnosis Solution/Comment
Device displays negative Interchange of transducer cable Make sure that the transducer
flow connections cables are connected correctly:
1A → 1A
1B → 1B
2A → 2A
2B → 2B
Wrong setting of measurement Make sure that the setting of menu
direction 3.1.5.1 is “+ direction”.
Current output does not work Internal electrical failure (open Start simulation on the current
circuit, short circuit etc.) output by use of menu 2.4.4.
The instrument for measuring the
current is coupled in series with
the output (across the terminals).
Pulse/Frequency output does Internal electrical failure (open Start simulation on the pulse/
not work circuit, short circuit etc.) frequency output by use of menu
2.4.2. The voltage probe is
coupled directly on the terminals.
Relay output does not work Internal electrical failure (open Start simulation on the relay
circuit, short circuit etc.) output by use of menu 2.4.3.
The instrument for measuring the
resistance is coupled directly on
the terminals.
Device displays 0 flow. Process flow value is very low Set the low flow cut off (menu
Device status: “OK” compared with the low flow cut 3.1.6) to 0.
off limit of the device
LCD display dark or partly Internal electrical failure (open Turn off power. Check the
dark (missing characters) circuit, short circuit etc.) connections for the power supply
PCBA and the flat cable for the
display. Start display test (menu
2.3.2) in case of partly dark display.
Operation with infrared keys Light interference Check that display is not dirty.
not possible Check that display lid is locked.
50 SFIDK.PS.029.E2.02
Maintenance and repair 8
The device fuse (slow blow 1.6 A/250 VH (switch off capacity 1500 A); license: UR, CSA, VDE)
is located on the power supply board.
The following messages are output in the event of a device error or measuring error:
– The output signal selected in menu item 3.4.3 is output at the analog output.
– A message is output at digital output 1 and 2 when the outputs are configured for the alarm
message.
– The message “D” appears in the display within the display menu for device errors or the
message “F” for measuring errors.
If you send in the ultrasonic flowmeter to Siemens for repairs, please enclose a note bearing
the following information.
Warning
All residue of the medium must be removed before sending in the device. This is particularly
necessary when the medium poses a health risk such as caustic, toxic, carcinogenic,
radioactive etc.
SFIDK.PS.029.E2.02 51
8.1 Technical information on the application
(please add sketch of installation)
52 SFIDK.PS.029.E2.02
Certificates 9
9.1 EC Declaration of Conformity
SFIDK.PS.029.E2.02 53
54 SFIDK.PS.029.E2.02
SFIDK.PS.029.E2.02 55
56 SFIDK.PS.029.E2.02
SFIDK.PS.029.E2.02 57
58 SFIDK.PS.029.E2.02
9.2 ATEX
SFIDK.PS.029.E2.02 59
We have checked the contents of this manual for agreement with the hardware The reproduction, transmission or use of this document or its contents is not permitted
and software described. Since deviations cannot be precluded entirely, we cannot without express written authority.
guarantee full agreement. However, the data in this manual are reviewed regularly Offenders will be liable for damages. All rights, including rights created by patent grant or
and any necessary corrections included in subsequent editions. Suggestions for registration of a utility model or design, are reserved.
improvement are always welcomed.
Technical data subject to change without prior notice. Copyright © Siemens AG 08.2007 All Rights Reserved