Manual 220ms
Manual 220ms
Printed in U.S.A.
03-954144-00:3
All rights reserved including the right of reproduction in whole or in part in any form. This document may be electronically reproduced, distributed, or printed in its entirety, provided this copyright and statement are attached. Any modification or any other reproduction, distribution, or use of this document or portions hereof is strictly prohibited without the express written permission of Varian, Inc. COPYRIGHT 2009 Varian, Inc. All rights reserved.
Contents
Introduction ...............................................................................................................................5 Functional Description .............................................................................................................9
Introduction ................................................................................................................................................................. 9 Gas Chromatograph (GC) ........................................................................................................................................ 11 Mass Spectrometer (MS) .......................................................................................................................................... 11 Cooling Fans ...................................................................................................................................................... 11 Vacuum System ................................................................................................................................................. 12 Transfer Line ...................................................................................................................................................... 14 Ion Trap Assembly ............................................................................................................................................. 16 Ion Gauge........................................................................................................................................................... 20 Electronic Assemblies .............................................................................................................................................. 21 Power Input Subsystem and Turbomolecular Pump Controller ......................................................................... 22 Main Power Circuit (210-MS and 220-MS) ........................................................................................................ 22 Power Board ...................................................................................................................................................... 23 RF Generator Assembly..................................................................................................................................... 24 Manifold Electronics Assembly .......................................................................................................................... 25 Foreline Pump .......................................................................................................................................................... 26 210-MS and 220-MS .......................................................................................................................................... 26 225-MS ............................................................................................................................................................... 26 Data System ............................................................................................................................................................. 27 Computer/Instrument Interface ................................................................................................................................. 27 Computer Hardware and Software Requirements ................................................................................................... 27 Autosampler.............................................................................................................................................................. 27
MS Maintenance ..................................................................................................................... 41
Foreline Pump (210, 220-MS only)...........................................................................................................................41 Checking Foreline Pump Oil .............................................................................................................................. 41 Purging Foreline Pump Oil ................................................................................................................................. 42 Changing Foreline Pump Oil .............................................................................................................................. 43 Flushing Pump Oil .............................................................................................................................................. 44 Changing the Oil Mist Cartridge (210, 220, and 225-MS) ........................................................................................ 44 DS-42 Oil Mist Eliminator ................................................................................................................................... 45 DS-102 Oil Mist Eliminator ................................................................................................................................. 46 Checking Cooling Fans ............................................................................................................................................ 47 Replacing the Turbomolecular Pump (210-MS and 220-MS) .................................................................................. 48 Replacing the Turbomolecular Pump (225-MS) ....................................................................................................... 50 Servicing the Ion Trap .............................................................................................................................................. 50 Turning Off the MS ................................................................................................................................................... 51 Retracting the Transfer Line ..................................................................................................................................... 51 Removing the Analyzer Assembly ............................................................................................................................ 52 Replacing the Electron Multiplier .............................................................................................................................. 53 Replacing the Filament(s) ......................................................................................................................................... 54 Removing the Ion Trap Oven ................................................................................................................................... 55 Cleaning the Trap Components................................................................................................................................ 56 Disassemble the Trap Components ................................................................................................................... 56 Cleaning the Trap Components ......................................................................................................................... 57 Cleaning Silica-Coated Electrodes .................................................................................................................... 59 Cleaning the Quartz or Silica-Coated Spacers .................................................................................................. 59 Reassembling the Trap ............................................................................................................................................ 59 Reinstalling the Trap Oven Assembly ...................................................................................................................... 60 Repositioning the Electron Multiplier ........................................................................................................................ 61 Reinstalling the Analyzer Assembly ......................................................................................................................... 61 Installing the Transfer Line ....................................................................................................................................... 61 Closing the Vent ....................................................................................................................................................... 62 Turning On the MS ................................................................................................................................................... 62 Baking Out the Trap.................................................................................................................................................. 62 Checking the Ion Trap Operation ............................................................................................................................. 62 Filling the Calibration Compound Vial ...................................................................................................................... 63 Moving the MS .......................................................................................................................................................... 63 210-MS or 220-MS ............................................................................................................................................. 63 225-MS ............................................................................................................................................................... 64
Troubleshooting ..................................................................................................................... 65
Isolating the Problem ................................................................................................................................................ 65 Checking the Data System ....................................................................................................................................... 65 Checking the GC ...................................................................................................................................................... 65 Checking the MS ...................................................................................................................................................... 65 Resolving Problems with Spectra ............................................................................................................................. 66 No Spectrum ...................................................................................................................................................... 66 Checking for an Open Filament ......................................................................................................................... 67 Checking the Turbomolecular Pump.................................................................................................................. 67 Checking the RF Adjustment ............................................................................................................................. 67 Checking the Method Parameters ..................................................................................................................... 67 Checking the Trap Assembly ............................................................................................................................. 68 Checking the Electronics.................................................................................................................................... 69 Loss of High Mass Peaks ......................................................................................................................................... 69 Missing Part of the Spectrum ................................................................................................................................... 69 Checking the RF Adjustment ............................................................................................................................. 70 Checking the RF Storage Level ......................................................................................................................... 70
Checking the Trap Temperature ........................................................................................................................ 70 Poor Resolution with Acceptable Air and Water Levels ........................................................................................... 70 Checking the Ion Content of the Trap ................................................................................................................ 71 Checking the Axial Modulation Setting .............................................................................................................. 71 High Baseline at High Masses.................................................................................................................................. 71 Trap Calibration Fails after Calibration Ions are Identified ....................................................................................... 72 Checking the Electron Multiplier Voltage ........................................................................................................... 72 Checking the Cal Gas Pressure ......................................................................................................................... 72 Checking for Leaks ................................................................................................................................................... 72 Setting Up for Leak Checking ............................................................................................................................ 73 Removing High Water Levels ................................................................................................................................... 77 Using Leak Detection Gas ........................................................................................................................................ 78 Repairing Large Air Leaks ........................................................................................................................................ 79 Repairing Small-to-Moderate Air Leaks.................................................................................................................... 79 Checking GC Connections ................................................................................................................................. 79 Removing the Capillary Column ............................................................................................................................... 80 Installing New Capillary Columns ............................................................................................................................. 81 Troubleshooting the GC ........................................................................................................................................... 82 Running the COLTEST Sample ............................................................................................................................... 83 Setting Up the Injector Conditions ..................................................................................................................... 83 Setting Up the Column ....................................................................................................................................... 83 Setting Up the Transfer Line and Trap-Temperature Conditions.......................................................................83 Setting Up a MS Acquisition Method ........................................................................................................................ 84 Troubleshooting Chromatographic Problems .................................................................................................... 85
Introduction
This manual contains hardware information for the Varian 210-MS, 220-MS, and 225-MS Ion Trap Mass Spectrometers. There are five chapters. The first chapter provides a functional description of the mass spectrometer (MS) and details of the instrument subsystems. The next chapter describes the installation and operation of the chemical ionization source. The third chapter contains MS maintenance procedures. The fourth chapter describes troubleshooting procedures. The final chapter provides information about related documents, instrument parts, and contacting Varian, Inc. The following identifies the components of the ion trap MS with the top cover off.
A B C D E
210-MS and 220-MS Top View Chemical Ionization (CI) Shutoff Valve Service Switch Transfer Line Heater Trap Heater Manifold Heater Cooling Fan (1 of 2) Turbomolecular Pump Transfer Line
A B C D
E F G H
F
225-MS Top View Chemical Ionization (CI) Shutoff Valve Service Switch Transfer Line Heater Trap Heater
A B C D
E F G
A F B C
210-MS, 220-MS, and 225-MS Front Panel A B C Cal Gas Adjust Vent Valve Cal Gas Vial D E F RF Coil Adjustment Screw Power Switch LED CI Cal Gas Adjust
Functional Description
Introduction
The 210-MS, 220-MS, and 225-MS GC/MS systems have four principal components: Gas chromatograph (GC) Mass spectrometer (MS) Data system (DS) Autosampler (optional)
The following figure is a block diagram of the 210-MS, 220-MS, and the 225-MS. A short, transfer line connects the GC and MS. The autosampler sits on top of the GC. Samples are injected manually or using the autosampler onto the capillary column through the GC injection port. The gas chromatograph separates the sample molecules. Effluent from the GC enters a fused silica capillary column, which goes through the transfer line and into the ion trap. The sample molecules undergo electron or chemical ionization before being analyzed according to their mass-to-charge ratios. The ions are detected by an electron multiplier, which produces a signal proportional to the number of ions detected. The electron multiplier passes the ion current signal to the system electronics, which in turn amplify the signal, digitize the result, and pass it on to the data system for further processing and display. See the figures that follow this one.
A B C
D E F
A C
A B C
D E
10
The analyzer compartment includes the following: Transfer line Vacuum manifold, which includes the ion trap Vacuum pump, controller, and turbo power supply RF coil and generator Pneumatics manifold Manifold Board Integrated Pumping Solution (225-MS only)
Cooling Fans
Two fans mounted on the rear panel of the spectrometer cool the unit. The analyzer compartment fan draws air from the back, blowing it directly on the bearing end of the turbomolecular pump in the analyzer compartment. The air then flows past the manifold electronics and out the front of the instrument. The turbomolecular pump controller supplies power to the analyzer compartment fan. The electronics section fan draws air from the back and blows it across the controller and power boards in the electronics compartment. To prevent hot air from the GC oven from affecting the MS, ensure that the 210-MS or the 220-MS is at least 15.3 cm (6 in.) away from a wall 225-MS is at least 25.4 cm (10 in.) away from a wall
Hot air from the GC oven does not affect the MS as long as the system is at least six inches from a wall. The power board supplies power to the electronics compartment fan.
11
Vacuum System
The vacuum system evacuates water vapor, air, and carrier gas from the MS ion trap assembly. Principal vacuum system components include Vacuum manifold Turbomolecular pump Foreline pump Vent valve Cal gas valve Chemical Ionization (CI) reagent gas valves
Vacuum Manifold
The vacuum manifold encloses the ion trap assembly. The vacuum manifold is a stainless-steel tube, which houses the analyzer. The turbomolecular vacuum pump, which evacuates the manifold, discharges into a foreline pump for the 210-MS and the 220-MS. The 225-MS has the integrated pumping solution, which includes the turbomolecular and the foreline pumps.
12
The vacuum manifold sits on top of the RF coil housing. The turbomolecular pump makes an airtight seal with the manifold, with a Viton O-ring. The ion trap assembly, which is suspended from the analyzer flange, extends into the body of the manifold. Another Viton O-ring makes an airtight seal between the manifold and the analyzer flange. Quick release tabs permit easy removal of the trap in the absence of vacuum. Eight electrical feed-throughs pass through the analyzer flange: One for the electron gate Three for the filament assembly Two for the axial modulation voltages applied to the filament and multiplier end cap electrodes of the ion trap assembly One for the high voltage to the electron multiplier cathode One for the ion current signal from the electron multiplier anode
A feed-through that passes through the underside of the manifold, provides radio frequency (RF) voltage to the ring electrode. An ion gauge monitors the pressure inside the manifold by generating and collecting ions from any gas present. The ion gauge also passes through the analyzer flange. The four additional vacuum manifold functions are: Transfer line CI reagent gas Introduction of the cal gas Venting
13
If the pump speed falls below 92% of its maximum operating speed, the TURBOMOLECULAR SPEED OK signal to the controller board turns off. The filament, electron multiplier, RF generator, CI reagent gas valve, and cal gas valve turn off automatically. This indicates a major air leak in the system or that the pump is too hot.
Pneumatics Manifold
The pneumatics manifold is an aluminum block mounted to the front of the vacuum manifold. It has two solenoid and two needle valves for the cal gas, and CI cal gas, the cal gas vial, and the vent valve. The vent valve, which is manually operated, connects to the atmosphere through the pneumatics manifold. A toggle arm on the front of the instrument opens and closes the vent valve. The calibration-gas-valve assembly consists of a metering needle valve, an ON/OFF solenoid-operated valve, and a glass vial containing the calibration liquid. The assembly sits directly behind the instruments door. The needle valve controls cal gas flow into the vacuum manifold through the solenoid valve. The calibration compound is perfluorotributylamine (PFTBA) or C12F27N, also known as fluorocarbon-43 (FC-43). A small glass vial attached to the valve assembly holds the compound. You set the flow of cal gas into the manifold manually using a needle valve. The data system controls the opening and closing of the solenoid-operated valve. Two solenoid valves control the flow of CI reagent gas into the manifold. The shutoff valve, which is near the rear panel, opens to let reagent gas flow into the instrument. The foreline pump removes a portion of the CI gas to prevent CI gas surges (pressure pulses). The gas flows through the shutoff valve through metering and solenoid operated valves before entering the vacuum manifold. The CI needle valve determines the split ratio of the reagent flow between the manifold and foreline pump. Turn the CI reagent gas valve on and off using System Control or Acquisition. Adjust the flow rate of the reagent gas into the manifold by means of a metering valve.
Transfer Line
A stainless steel tube transfer line couples the GC to the MS. The transfer line keeps the GC column warm as the column enters the MS. The transfer line is 12 cm (5 in.) long, and has a diameter of 4.1 cm (1.6 in.). One end enters a hole in the right side of the GC before passing into the GC oven. The transfer-line tip enters the vacuum manifold and goes into the ion trap.
The transfer line is hot. Ensure it is cool before touching it, or use protective gloves.
The transfer line is a stainless-steel weldment fitted with a center tube, a heat exchanger, and a boot. The heat exchanger, an aluminum cylinder, contains a cartridge heater and a thermocouple as the temperature sensor. The temperature sensor measures the temperature of the line. The cartridge heater heats the cylinder, which distributes heat evenly throughout the transfer line. The
14
boot of the transfer line, which attaches to the GC, prevents hot air leakage from the GC oven.
A B C D E F
G H I J K
Transfer Line
A bayonet mount secures the transfer line. Before removing the trap, push the bayonet mount gently as you twist it counterclockwise and pull it out. Make sure the transfer line extends out from the trap. NOTE: Not removing the transfer line before removing the trap may damage the trap heater post, quartz ring, or the transfer line tip or all. The power board supplies power to the cartridge heater through a transfer line heater cable. The heater cable projects out from one end of the transfer line. It plugs into a connector on the top of the power board panel. Set the transfer line temperature from the Temperature view in System Control. The maximum temperature of the transfer line is 350 C; the minimum temperature depends on the GC oven and trap temperatures. In general, set the transfer line temperature as much as 30 C below the maximum column operating temperature and not observe adverse chromatographic effects, such as, retention time shifts or peak broadening.
15
A B C D E F G
Heating Cable Boot Nut Ferrule Transfer Line/Alignment Tool Nose Nose Hole
H I J K L M N
O-ring Transfer Line Tip Heating Cable Slot Nose Clip Bayonet Mount Analyzer Assembly Tongue Analyzer Assembly Lock-Down Tabs Note: The tabs may be long or short
The following figure shows the ion trap assembly with its three electrodes, electron gate, and filament lens. NOTE: The Silica-Coated Spacers have a shiny finish on the inside surface.
16
A B C D E F G H I J K L M
A B C D E F G
Screw, 6/32, 4 places Clamping Plate Exit-End Cap Quartz or Silica-Coated Spacer, 2 places RF Ring Electrode Filament (entrance) End Cap Electron Gate
H I J K L M
Wave Washer Gate Conductor Trap Oven, T is located this side. Filament Assembly Filament Clip Screw
Trap Oven
The trap oven is a heated anodized aluminum block that maintains a uniform temperature for the trap electrodes. A heater post on the manifold flange generates the heat. A thermal well measures the oven temperature. The oven holds the ionization filaments and acts as a lens for focusing the ionizing electrons before they enter the trap.
Filament Assembly
The filament assembly is in the trap oven. It is connected to three feed-throughs on the manifold flange. The filament assembly consists of two filaments and a repeller plate. The two filaments are mounted side-by-side, with each filament approximately equidistant from the entrance hole of the ovens electron focusing lens. The MS only uses one filament at time; the extra filament is a back up in case the first one burns out.
17
Each filament is a rhenium wire. When heated by electric current, the filament produces electrons by thermionic emission. The filament emission current refers to the flow of emitted electrons from the filament. The magnitude of the filament emission current is set in Instrument Control and current settings range from 5 to 100 A. NOTE: The two filaments will probably not have the same net flow of electrons into the ion trap. Therefore, the signal amplitudes from two different filaments will be different. A typical difference is 2:1, but it may be as high as 5:1.
Electron Gate
The electron gate is a cylindrical electrode that controls the entry of electrons into the ion trap cavity. When electrons emitted from the heated filament are not required for ionization, the electron gate is held at a -150V dc potential. The electron gate sits inside the trap oven, in front of the lens and behind the entrance-end cap electrode. An anodization layer insulates it from the filamentend (entrance) cap. When the ion trap requires electrons, the electron gate potential changes from 150 to +150V dc. The gate potential remains positive for a variable length of time, e.g., from 10 sec to 65 msec. During this interval, the electrons are focused into the ion trap cavity with sufficient energy, usually, 50 to 80 eV, to achieve electron ionization of the sample molecules, or of the reagent gas molecules in the case of chemical ionization.
18
The filament-end cap, exit-end cap, and RF-ring electrodes have hyperbolic inner surfaces. Together, these electrodes form a cavity in which ionization, fragmentation, storage, and mass analysis take place. Energetic electrons enter the ion trap cavity through the filament-end cap using the electron gate. There are seven holes in the center of the exit-end cap electrode. Sample ions produced in the ion trap are ejected through these holes into the electron multiplier. Two identical quartz or silica-coated spacers separate the central ring electrode from the filament and exit-end cap. The trap oven and its clamping plate hold the electrodes and spacers in place. A cutout is provided in the quartz spacers and in the exit-end cap to allow the transfer line to enter the ion trap. The RF generator assembly provides high voltage RF that is applied to the RF ring electrode. Under the proper RF voltage, the ion trap electrodes create a three-dimensional, hyperbolic electric field. This field is capable of trapping the ions in stable, aperiodic orbits. As the RF voltage increases, however, the ion trajectories become unstable in increasing order of mass per charge. The ion trap ejects the ions and sends them to an electron multiplier for detection. During mass analysis, a supplementary RF voltage of 485 kHz is applied to the filament- and exit-end caps. This voltage, termed the axial modulation voltage, improves spectral mass resolution and analytical sensitivity. Other voltages may be applied between the end caps to implement such options as CI and MS/MS.
Electron Multiplier
The electron multiplier is at the exit-end cap electrode. It is in a pre-aligned position on a protective metal clip and can be replaced easily. The multiplier detects positive ions as the ion trap ejects them through the holes in the exit-end cap electrode. The continuous dynode electron multiplier consists of a leadoxide/glass, funnel-like resistor. A negative voltage of between -800 and -3000V is applied to the front end of the electron multiplier, which is the cathode. The back end of the cathode is held at ground potential, and is the anode.
19
A B C D E
Exit-End Cap Electron Multiplier Track Multiplier Signal Pin EM Grid Electron Multiplier Mount
F G H I
Multiplier Contacts Multiplier High Voltage Pin Transfer Line Alignment Transfer Line Entrance Hole
The negative voltage applied to the cathode attracts the positive ions ejected from the ion trap cavity. These ions strike the cathode with sufficient velocity to dislodge electrons from the inner curving surface of the cathode. The increasingly positive potential gradient draws the ejected electrons into the electron multiplier, further accelerating them in the process. Because the electron multiplier is curved, the ejected electrons strike the inner surface of the multiplier again, and more electrons are emitted. This configuration produces a cascade of electrons that is accelerated toward ground potential at the exit-end of the cathode. The anode collects the electrons, and passes the resulting ion current signal to the integrator circuit on the lower manifold board. The ion current signal is proportional to the total number of electrons that the ion trap ejects. Typically, the 5 voltage applied to the electron multiplier is adjusted until the gain is about 10 . Therefore, each ion that enters the electron multiplier generates approximately 105 electrons.
Ion Gauge
The optional ion gauge is based on the Bayard-Alpert gauge tube. The specifications for the gauge are commercially available gauges. Fixed pressure readings with nominally identical gauges may exhibit variations of 15%. The gauges usually have an accuracy of 25% in mid-range In general, the ion gauge has good repeatability. However, the ion gauge response depends on gas composition. A certain pressure of air and water give a different reading than that of Helium. The ion gauge is a rough indicator of vacuum conditions. It is not a precise quantitative tool. The gauge uses thoria-coated iridium (ThO-Ir) filaments. These filaments are burnout resistant, and therefore exhibit high tolerance to air and water in the vacuum manifold. There is a time delay associated with heating the filament. This
20
delay translates to a delay in determining whether a filament is open. To obtain a stable reading, wait 15 to 20 seconds after the filament is turned on. The ion gauge measures pressures between 10-6 and 10-2 Torr. A logarithmic amplifier amplifies the collector current, and the data system interprets this current as measured vacuum.
Electronic Assemblies
The electronic assemblies consist of the following: Power input subsystem and turbomolecular pump controller Power board MS Controller board Manifold electronics assembly RF generator board and RF coil.
The placement of the electronics minimizes the cable lengths between critical components. The MS controller and power boards are in an electronics enclosure separated from the analyzer section by a sheet metal bulkhead. The manifold electronics are enclosed above the analyzer. The RF generator attaches to the rear of the RF coil assembly.
DATA SYSTEM
USB
CONTROLLER BOARD
ETHERNET
CP-8400 AUTOSAMPLER
21
22
The turbo controller regulates the speed of the turbo pump. The controller provides turbo speed and startup power to the power board. The +24V dc power supply supplies power for the solenoid valves, electronics compartment fan, and the electron multiplier power supply.
WARNING:
SHOCK HAZARD
In the event of an emergency, shut off all power to the MS by turning the main power switch OFF.
Power Board
The power control board supplies power to all electronics components except the turbomolecular controller. It controls the heaters, ion trap and ion gauge filaments, and solenoid valves. NOTE: The switching power supply is protected by a 5A, non-time-delay fuse. The following switching power supplies are on the board: The + 5V dc power supply provides voltage to all digital circuits. The -15V and +15V dc power supplies provides voltages to the analog circuits on the power board and the manifold electronics assembly. The +20V and -20V dc power supplies provides the voltages to the controller and RF generator boards analog circuitry. The +60V dc power supply, provides unregulated +60V dc voltage to the RF generator board and trap heater. The +180V and -180V dc power supply provides voltage to the ion trap electron gate circuit and the ion gauge.
The following circuits are on the board: The trap and ion gauge filament control circuits that provide current to heat the filament and regulate the emission current from the filament. The trap-filament emission current is set between 5 and 100 A. Three heater control circuits that provide feedback control for the manifold, trap, and transfer-line heaters. The trap heater uses a proportional integral (PI) control circuit. Three solenoid control circuits that turn the cal gas, CI reagent gas, and CI shutoff valve solenoids on and off. The electron energy control circuits that controls the dc bias on both the ion trap and ion gauge filaments. The diagnostic multiplexer circuit that routes the voltage output of various components, and circuits on the power control board to the controller board. On the top edge of the power board are 12 monitor LEDs. When illuminated, these lights indicate that the voltages of the various circuits on the power board are at the correct levels, and that there are no faults. In idle mode, all LEDs, except the +180V, -180V, and trap filament are on. The LEDs for +180V, -180V, and the trap filament only turn on when the filaments are on.
23
RF Generator Assembly
The RF generator assembly consists of an RF generator circuit board, an RF detector circuit board, and the RF coil. A shielded housing beneath the vacuum manifold encloses the coil and RF detector circuit board. The RF generator circuit board is attached to the back of the shielded housing. The RF generator circuit board receives an analog signal from the controller board that is proportional to the current mass position in the scan, which is in turn proportional to the RF voltage applied to the ion trap. The RF detector circuit board sends a signal to the RF generator; this signal is proportional to the actual amount of RF voltage applied to the ion trap. The RF generator board compares the desired and actual RF voltages, and based on this feedback, adjusts the gain to modify the applied RF voltage amplifier to equal the desired RF voltage level. Since the high voltage required by the ion trap exceeds the capabilities of conventional electronic amplifiers, a resonant LC (inductor-capacitor) circuit consisting of the RF coil and the ion trap capacitance is used. At resonance, the RF voltage at the ion trap-end of the coil is about 100 times that at the RF generator circuit end of the coil.
DATA SYSTEM
CONTROLLER BOARD
24
MS Controller Board
The controller board controls the MS. The controller board communicates with the data system using the USB interface of the data system computer. The MS controller performs the following functions: Interprets instrument commands from the data system and produces a sequence of analog and digital signals that control the operation of circuits on other MS boards Collects analog and digital diagnostic data from other subsystems and transmits that information to the data system Filters, integrates and digitizes the ion current signal, and transmits the spectra to the data system Generates axial modulation waveforms, including waveforms used by CI, MS/MS, and SIS options
When powered up, the controllers processor runs a ROM resident program that initializes the board. The program permits the processor to receive information through the USB interface. When the data system is started, operating information downloads to the RAM memory of the controller board. The controller board then performs in response to the commands sent through the USB interface. NOTE: The controller board is accessed through two connectors on the rear panel of the instrument. J42 is a USB connection that the Data System. J43, a D-shell connector labeled, Remote Option, is a special research application and the GC start signal. When a mass spectrum is acquired, the data system downloads parameters, such as, electron multiplier voltage, scan range, and time, and ionization mode. This information is used to create a scan over the desired mass range. At the end of the scan, the data is sent to the data system for further processing and display.
25
The waveform generator can create waveforms over a wide range of frequencies and amplitudes. The waveform generator has the following: Dual-port RAM (256 Kbytes) to provide memory for single or multiple digitized waveforms. A selectable frequency generation clock (625 KHz, 1.25 MHz, or 2.5 MHz and a 15-bit variable-length counter to control timing. A 12-bit DAC, low-pass filter and amplifier to reconstruct waveforms. A variable operational frequency range that uses the high frequency transformer (12 to 500 KHz) or the low frequency transformer (200 Hz to 1.25 KHz). Two transformers, that apply the waveform output to the end cap electrodes.
NOTE: The waveform options, for example, CI, MS/MS, or SIS, require the waveform key(s). The key(s) is installed by the factory, or by a Varian Customer Support Representative.
Foreline Pump
The foreline pump has two purposes. The first is reducing the vacuum system pressure to a level that permits the operation of the high vacuum turbomolecular pumps. The second is maintaining the vacuum system pressure by removing the exhaust gases of the high vacuum pump.
If you use the 210-MS, 220-MS, or 225-MS to analyze hazardous materials, you must have an exhaust system for the foreline pump that complies with applicable safety and environmental regulations.
225-MS
The 225-MS has a foreline pump integrated into the module and does not need an external vacuum line or a power connection on the rear panel.
26
Data System
The data system (DS) has both hardware and software components. The hardware includes a computer/instrument interface, personal computer, video display monitor, and optionally, a printer. The software installed on the system includes programs that control the MS, that control the GC, that set system parameters automatically, and that oversee scancontrol, data-acquisition, and data processing For a complete description of software, refer to the MS Workstation Software Reference Manual.
Computer/Instrument Interface
The GCMS uses a universal serial bus (USB) interface. The USB is a standard computer/instrument communications link for all types of computers.
Autosampler
The autosamplers available are the Varian 8400, 8410, and CombiPAL AutoSamplers. For complete installation and operating instructions, please refer to the autosampler manual.
27
28
Chemical Ionization
Introduction
NOTE: CI mode is an MS option. If your system does not have this option, you cannot perform CI analyses. Chemical Ionization provides mass spectral data that complement electron ionization (EI) data for the analysis of complex compounds. In the standard CI mode of operation, a CI reagent gas is introduced into the ion trap analyzer from an external gas supply cylinder. The reagent gas is ionized by EI to form reagent ions. These reagent ions then ionize sample molecules that enter the ion trap with the helium carrier gas from the capillary column. The operation and adjustment of reagent gases for the standard CI option are described in the first part of this section. Two additional options allow the selection of certain liquids as sources for CI reagents. These are the Liquid CI Inlet (or LCI Inlet) and the Multiple CI module (or MCI module). This chapter describes how to install and operate the LCI inlet. Refer to the documentation included with the MCI module for installing and operating the module.
29
DO NOT flame dry the reagent gas lines with CI reagent gas present.
CI reagent gases may be hazardous. Use proper protection when installing the reagent gas.
1. Open System Control and click Manual Control. 2. Make sure that the electron multiplier, filament, and RF voltage are off. The Multiplier, Filament, and RF text should be red or black. NOTE: Two solenoid-operated valves control the flow of CI reagent gas into the manifold. The valves are opened and closed by clicking the CI button in Instrument Control. A needle valve controls the amount of reagent gas flowing into the manifold. Adjust the needle valve, behind MS door, manually using the knob labeled CI GAS. Turn the knob clockwise to increase the flow of reagent gas. See the Functional Block Diagram of the Vacuum System on page 12. 3. Verify that the CI gas solenoid valves are closed. When these valves are closed, the CI Gas icon to the left of the ion trap symbol is not green. (If the CI icon is green, click on the icon so that it turns to red or black.)
30
4. Install a two-stage pressure regulator on the reagent gas cylinder or lecture bottle. Tighten the connection securely. NOTE: A two-stage pressure regulator typically consists of the following components: Secondary valve, Pressure adjustment valve, Supply pressure gauge, and Delivery pressure gauge 5. Use the main valve to turn the gas on or off. The secondary valve on the pressure regulator is the coarse control of the flow of gas from the gas cylinder up to the pressure adjustment valve. The supply pressure gauge monitors the gas pressure in the bottle. The pressure adjustment sets the head pressure of the gas delivered to the MS. 6. Connect one end of the 1/8 in. OD gas supply line to the pressure regulator. 7. On the back of the MS, loosen the two screws that hold the plug in the CI Shutoff Manifold 2 to 3 turns. 8. Remove the plug by pulling straight out and twisting.
A
Plug
Screws
9. Use 1/8 in. OD tubing for the supply line between the gas cylinder and the CI shutoff manifold. No ferrule is required on the MS end of this tube. The seal is made with an elastomer O-ring. Inspect the end of the tubing and ensure that the surface finish is smooth. If there are scratches, cut off the damaged part or use 200-600 grit abrasive paper to refinish the sealing end of the tube. 10. Carefully insert the tube into the CI shutoff manifold hole (where the plug came out of) until it is firmly seated. Be careful not to scratch the tube. 11. Tighten the two screws. 12. Ensure that the secondary valve on the regulator on the gas cylinder is closed. 13. Open the main control valve on the lecture bottle. Next, open the secondary valve and adjust the pressure adjustment valve to approximately 5 psi so that reagent gas flows at a moderate rate through the gas line. 14. Open the MS door. Verify that the CI GAS needle valve is turned fully counterclockwise.
31
15. Flush the gas line of air and water vapor as follows: a. Turn the adjustment valve clockwise to reduce the pressure. b. Open the CI gas solenoid valves by clicking on the CI icon in the Control and Status field of the Manual Control tab dialog in System Control. When the valves are opened, the CI button is green. c. Evacuate the CI reagent supply line for about 30 minutes.
To flush excess water from the gas line, do the following: 1. Ensure that the electron multiplier, filament, and RF voltage are off. 2. Open the main valve on the lecture bottle. (The secondary valve on the pressure regulator should be open.) 3. Turn the CI needle valve fully counterclockwise. 4. Open the CI gas solenoid valves and allow the system to pump down for about 1 hour. 5. Close the main valve on the gas cylinder but keep the CI GAS solenoid valves open. Allow the system to pump down for about 15 minutes. 6. Recheck the air/water spectrum. If excess water is not present, continue to the next procedure, Setting CI Reagent Flow.
32
Methane
13 9 45 5000 2000 60 100
Isobutane
19 15 65 5000 2000 60 100
Ammonia
13 9 45 5000 2000 60 100
If you have the Liquid CI Inlet or the Multiple CI Module, use the following:
Acetonitrile d3-Acetonitrile
19 15 65 5000 2000 40 100 19 15 65 5000 2000 20* 100
Methanol
19 15 55 5000 2000 40 100
Use short reaction times for deuterated reagents. Longer reaction times allow more H/D exchange with background water and the resulting spectrum will show more [M+H]+ and less [M+D]+.
33
Isobutane
Ammonia
Acetonitrile
d3-Acetonitrile
Methanol
The reagent gas pressure in the ion trap will be approximately 1 to 2 x 10-5 Torr (about 1.3 to 2.6 x 10-3 Pa). The CI reagent molecules are about 1% of the gas pressure in the ion trap. Helium atoms from the column flow are present at 100 times this pressure.
Liquid CI Reagents
A liquid CI inlet assembly must be installed for the CI mode. Use the following instructions to install the liquid CI inlet assembly and, if necessary, to switch back to using gas CI reagents.
34
2. Remove the top cover and then attach the Liquid CI Inlet assembly to the back of the MS using the following instructions. Refer to the following drawings. a. From the back of the MS, remove one of the two screws that hold the CI shut-off block. Replace it, loosely, with a long screw supplied with the kit, part number 1222200625. b. Remove the other screw. c. Gently pull the free end of the liquid CI restrictor tube, part number 393002401 from the L-bracket where it attaches to the back of the MS, while leaving the other end of the restrictor tube attached to the Liquid CI Inlet block.
d. Loosely attach the Liquid CI Inlet assembly to the back of the MS using the L-bracket with the screw that was removed. e. Rotate the Liquid CI Inlet assembly out of the way to remove the remaining screw. f. Rotate the Liquid CI Inlet assembly back into position and loosely attach the liquid CI inlet assembly with the remaining long screw, part number 1222200625.
g. Reinsert the liquid CI restrictor tube through the L-bracket into the back of the MS. The restrictor tube must be inserted far enough to engage the O-ring in the CI shutoff block.
A B C D
Inlet Block (393002301) Viton O-ring (393010907) O-ring Retainer (393002501) Reservoir Bulb, same as Cal Gas Chamber (392027000)
E F G H
Reservoir Cover (393002601) Restrictor (393002401) L-Bracket (393002701) Viton O-Ring (393010904)
Liquid CI Inlet
35
3. Replace the long restrictor, part number 393059701, with 1/8" OD PEEK tubing, part number 393003701. a. With the liquid CI inlet mounting screws still loose, pull out the long restrictor tube from the CI shutoff block. b. Loosen the four screws on the top of the pneumatics manifold (at the front of the MS). c. Pull out the long restrictor tube from the bottom of the pneumatics manifold. Carefully pull the tube out of the front of the MS. Save this long restrictor to use with pressurized gases such as methane.
d. Feed the PEEK tube, part number 393003701, into position, starting from the front of the MS (occupies roughly the same space as the long restrictor tube). e. Gently install the PEEK tube end into the pneumatics manifold, being careful not to let the retaining plate scratch the tube. f. Do not retighten the 4 screws on the pneumatics manifold yet.
g. Insert the other end of the PEEK tube into the CI shutoff block and tighten the 2 screws from the rear of the MS. 4. Replace the front restrictor. a. Remove the existing short gas restrictor, part number 393059601, from the bottom of the pneumatics manifold. b. Install the front liquid CI restrictor, part number 393059602, into the same location in the pneumatics manifold. Do not let the retaining plate scratch the restrictor tube ends. c. Tighten the 4 screws on the pneumatics manifold.
36
Never force the cap onto the glass reservoir stemit can break.
Use safety glasses and protective gloves, especially when attempting to remove a cap from a filled reservoir.
Use a gentle, twisting/pushing motion to install the plastic cap onto the reservoir stem. Use a gentle twisting/pulling motion to remove the plastic cap from reservoir stem.
37
Be careful not to spill any liquid, especially the few drops that may be in the neck of the bulb.
38
39
40
MS Maintenance
This section provides procedures for the routine MS maintenance tasks listed in Quick Reference.
41
B A
C D
A B C D E
Gas Ballast Valve Air Inlet Exhaust Filler Plug Oil Level Sight Glass
F G H I
42
To purge the foreline pump oil: 1. Place an exhaust vent over the open exhaust port. 2. With the foreline pump running, turn the gas ballast valve counterclockwise to the open position. The pump will become noisy and emit oil vapor. 3. After 10 minutes, turn the gas ballast valve back to the closed position. 4. Remove the exhaust vent.
The pump weighs 25 kg (55 lb.). To prevent personal injury, use proper moving and lifting techniques.
6. Place an oil pan beneath the drain port to catch any spillage.
Hazardous chemicals may be present. Avoid contact with skin. Use proper eye and skin protection.
7. Remove the plastic cover and the filler plug on top of the pump. 8. Put the container where it can catch the oil and then slowly remove the drain plug in the front of the pump.
Toxic residues from MS samples build up in used pump oil. Dispose of all used pump oil in accordance with applicable regulations. Place a hazards warning label on the container, if necessary.
9. Tilt the pump forward and hold until oil flow ceases. 10. Return the pump to the horizontal and refit the plug. 11. Run the pump for approximately ten seconds with the intake port open. This will remove any residual oil from the pumping block.
43
Avoid breathing oil mist coming from the exhaust port during this operation.
12. Remove the plug, tilt the pump, and then drain the oil. 13. Return the pump to the horizontal position. 14. Wipe the oil residue from the drainage port and then refit the drain plug. 15. Fill the pump with fresh oil, part number 8829953800, through the filler port until the oil level reaches the maximum level in the sight glass. A funnel may be helpful.
Avoid breathing oil mist coming from the exhaust port during this operation.
6. Stop the pump, drain the flushing oil, and replace as described previously.
44
45
To reassemble the oil mist eliminator: 1. Install a new cartridge in Lower housing B. 2. Press gently to check that it is firmly seated. 3. Install Valve D with polished side toward cartridge. 4. Center Spring C over Valve D, and fit gasket, F in the groove. 5. Cover entire assembly with Upper housing B. 6. Tighten Upper and Lower housings B, using screws A. NOTE: After changing the cartridge several times, it may be necessary to replace the gasket and the centering O-ring gasket.
46
The cooling fans maintain an optimal temperature for the turbomolecular pump and the electronics modules. Without the cooling fans, the lifetime of the turbomolecular pump and temperature-sensitive electronic components would be shortened. To ensure proper operation of the cooling system, operate the MS with its covers in place. In addition, check the fans at least once each week. The MS has two fans on its rear panel. To check fan operation, do the following: 1. Make sure that the MS MAIN switch and SERVICE switch are turned ON. (See Error! Reference source not found. on page Error! Bookmark not defined. for a photo showing the locations of the main and SERVICE switch.) 2. Place a large sheet of paper over one of the fan guards. If the paper is sucked toward the fan guard, the fan is working. If it is not, the fan is broken. Contact your Varian Customer Support Representative to arrange for a replacement.
3. Check the second fan in the same manner. If the fans are excessively noisy, for example, if they whine or whir, a fan may be about to fail and it should be replaced. To identify which fan is about to fail, do the following: 1. Remove the top cover from the MS. If the noise continues, go to step 3. If the noise stops, go to step 2.
2. Turn off the electronics compartment fan using the SERVICE switch, and replace the top cover. (See Error! Reference source not found. on page Error! Bookmark not defined. for a photo showing the location of the SERVICE switch.) If the noise returns, it is coming from the turbomolecular pump cooling fan. Proceed to step 4. If the noise does not return, remove the cover and proceed to step 3.
3. This step is specific to the MS model. 210-MS or 220-MS only: Turn off the electronics compartment fan using the SERVICE switch. (See Error! Reference source not found. on page Error! Bookmark not defined. for a photo showing the location of the SERVICE switch.) If the noise continues, it is from the turbomolecular pump-cooling fan. Contact a Varian Customer Support Representative to replace it. If the noise stops, it is coming from the electronics compartment fan.
225-MS only: Contact a Varian Customer Support Representative replace of the IPS module.
47
A B C
D E F
To unsecure the turbomolecular pump, do the following: 1. Loosen each of the four clamping screws about 2 turns with a 3/16 in. ball head hex driver. NOTE: Do not completely unscrew the two inner clamping screws. (If you should unscrew them, put the screws back after you removed the turbomolecular pump from the MS.)
48
2. Remove the outside bottom clamping screw. 3. Remove the bottom clamp, while holding the turbomolecular pump in place. 4. Remove the outside top clamping screw (closest to the transfer line). 5. Remove the top clamp as you hold the turbomolecular pump in place. To replace and to secure the turbomolecular pump, do the following: 1. Pull the turbomolecular pump to the back and lift it clear of the MS. 2. Remove the large seal from the turbomolecular inlet, and place it on the inlet of the new turbomolecular pump, part number 393076401. The orientation of the seal is not important. Leave the red cap on the new turbomolecular exhaust port.
3. Carefully slide the new turbomolecular pump and seal into position on the end of the manifold. Make sure the electrical connection (turbomolecular cable) is tilted towards the bulkhead, for example, toward the left as viewed from the rear of the MS. Take care not to scratch the sealing surface on the manifold in front of the turbomolecular pump.
4. Insert the top clamp and loosely fasten it into place. 5. Insert the bottom clamp and loosely fasten it into place. 6. Tighten all four clamping screws until snug. To reconnect the turbomolecular pump to other components, do the following: 1. Reconnect the turbomolecular cable. Rotate the retaining ring clockwise with downward pressure to lock the cable into position. 2. Remove the red cap on the turbomolecular exhaust port. 3. Place the seal on the turbomolecular pump exhaust port. 4. Reconnect the vacuum hose elbow and clamp. 5. Reconnect the pneumatics exhaust tube. To finish installing the turbomolecular pump, do the following: 1. Make sure that the vent valve is closed. 2. Plug in the power cord. 3. Turn on the rear-panel main power switch. 4. Tighten the top and bottom clamp screws. 5. Monitor the turbomolecular pump speed using Diagnostics under Vacuum System Status. 6. Once the pump is running satisfactorily, replace the top cover and then slide the GC and MS back together. 7. Discard the old turbomolecular pump. Be sure to comply with all applicable health and safety regulations.
49
50
To turn off the cooled MS, do the following: 1. Shut off the turbomolecular pump, foreline pump, and all electronics by turning off the main power switch on the back panel. 2. Disconnect the MS power cord. 3. Open the front panel door and lift the toggle vent valve for 1 second to slow the turbomolecular pump down. 4. After the pump stops spinning down, open the vent valve. Leave it open until the system is fully vented, for example, about 5-10 minutes.
51
A B C D E F G
H I J K L M N
O-ring Transfer Line Tip Heating Cable Slot Nose Clip Bayonet Mount Analyzer Assembly Tongue Analyzer Assembly Lock-Down Tabs
NOTE: Be sure the transfer line is retracted. Otherwise, you cannot remove the analyzer assembly without damaging the analyzer. To prevent contamination, while handling the parts, wear latex or nitrile gloves. To remove the analyzer assembly, do the following: 1. Remove the top cover of the MS by grasping both sides and lifting up. 2. Unplug the trap heater harness located near the top of the instrument. 3. On the side of the analyzer assembly (near the transfer line), push out the locking tabs on the power ribbon cable. This releases the cable. 4. Pull the ribbon cable out and move it away from the analyzer. 5. Push down and spread the two analyzer release tabs.
52
NOTE: Some MS systems have a transfer line removal flap that blocks the locking tabs. If such a flap is present, tip it out of the way during the procedure and return it to its original position once the analyzer is replaced. 6. Tilt the rear end up carefully to remove the analyzer. 7. Move the analyzer assembly toward the rear to free the front tab. 8. Place the analyzer upside down on a flat surface.
53
A B C D E
Exit-End Cap Electron Multiplier Track Multiplier Signal Pin EM Grid Electron Multiplier Mount
F G H I
Multiplier Contacts Multiplier High Voltage Pin Transfer Line Alignment Transfer Line Entrance Hole
Electron Multiplier
54
A B C D E
Belleville Washer Ceramic Filament Assembly Filament Connectors Post Connectors for Filament Screw
F G H I J
Filament Retainer Analyzer Flange Transfer Line Alignment Tool Center Disk Feeler Disk
Filament Assembly
Do not rotate the assembly more than 2 degrees. Otherwise, you may damage the contact springs.
55
6. If you are only cleaning the electrodes, leave the gate parts in the oven. Otherwise, remove the gate, wavy spring washer, and gate conductor by turning the oven upside down. NOTE: The Silica-Coated Spacers have a shiny, mirror like finish on the inside surface.
56
A B C D E F G H I J K L M
A B C D E F G
Screw, 6/32 in. , 4 places Clamping Plate Exit-End Cap Quartz or Silica-Coated Spacer, 2 pieces RF Ring Electrode Filament-End Cap Electron Gate
H I J K L M
Wave Washer Gate Conductor Trap Oven, T is located this side. Filament Assembly Filament Clip Screw
57
Contaminants sometimes appear as dark or colored areas, but they may also be invisible. Clean each part thoroughly, even if there is no apparent contamination. After you clean a part, hold it under running water and use a clean applicator to remove the last visible traces of aluminum oxide.
2. Immediately place the clean part in a beaker containing a solution of detergent and warm water. NOTE: Do not let the slurry dry on the metal. Dried aluminum oxide is difficult to remove. 3. When you have finished cleaning all of the parts, place the beaker in an ultrasonic cleaner, and subject the beaker and its contents to ultrasound for about 1 minute. 4. Rinse each part with fresh water. 5. Using clean tools, place the parts in a beaker containing de-ionized water, and then subject the beaker and its contents to ultrasound for about 1 minute. If the water is cloudy afterwards, replace the deionized water and repeat.
6. Rinse the parts with methanol. 7. Place the parts in a beaker of fresh methanol. Subject the beaker and its contents to ultrasound for about 1 minute. NOTE: After the ion trap parts are clean, wear clean, lint-free gloves for succeeding procedures to prevent contamination. Do not wear vinyl gloves. 8. Remove the ion trap parts from the beaker, and place them on a clean, lintfree surface. Allow the parts to dry in air.
9. Inspect each part to make sure that all spots and particles have been removed. If you observe any contamination, clean the part again using the procedure described above.
NOTE: Clean any small parts, such as, the electron gate conductor, the gate, and wavy washer spring, by placing them along with the other parts in a beaker with methanol and subjecting them to ultrasound for 1 minute. Check the oven trap near the filament entrance hole for carbon deposits. Carbon buildup may decrease sensitivity and shorter filament lifetime. Remove the carbon stains only with a cotton swab and methanol. After cleaning, check filament entrance hole for particles and fibers. The area must be cleaned before reassembly.
58
DO NOT use Aluminum Oxide or other abrasives because this will remove the silica layer on the trap! DO NOT use harsh laboratory cleaners because this will remove the silica layer on the trap! Use only mild detergent (pH between 6 and 7.5).
For routine cleaning of the Silica-Coated electrodes, ultrasonicate the ion trap electrodes for 10 minutes in methylene chloride or methanol. Use separate beakers for each electrode to avoid scratching trap surfaces. Trap disassembly and reassembly is otherwise identical to the procedure in the Maintenance Section. If heavy matrix (dirty) samples are routinely run on the instrument and the electrodes are visibly discolored where the column enters the trap at the multiplier end cap, use a toothbrush and liquid hand soap or dish detergent (pH between 6 and 7.5) to gently scrub the trap parts. Rinse the trap and then sonicate it in water and then twice in methylene chloride or methanol. NOTE: The initial hydrocarbon background on the coated ion trap is higher than on the standard ion trap. To speed up the bakeout, you may want to bake out the ion trap overnight at 220 C. In the bakeout mode, the manifold is set to 120 C.
59
1. Replace the gate conductor, tab-down into position. 2. Replace the wavy washer on the gate conductor. The washer orientation is not important. 3. Replace the gate so that the flat, shiny surface faces the washer. 4. Replace the filament (single-hole) electrode in the oven. 5. Replace one of the quartz spacers so that the notch faces the filament (single-hole) electrode. 6. Replace the RF electrode, followed by a quartz spacer. The notch in the quartz spacer should face up towards the exit (seven-hole) electrode. NOTE: Make sure that the notch in the quartz spacer and the notch in the exitend cap are aligned. 7. Replace the exit (seven-hole) electrode so that the notch on this electrode faces the side of the trap labeled with the side-ways T. 8. Slide the clamping plate under the screws on the top of the trap oven assembly. 9. Visually check the transfer line hole, making sure that notches in the quartz spacer and exit-end cap electrode are aligned and centered in the trap oven. 10. Tighten the screws.
Do not rotate the assembly more than 2 degrees or, you may damage the contact springs.
2. To set transfer line hole height to the analyzer flange, place the nub of the center disk into the hole created by the notches in the quartz spacer and the exit (seven-hole) electrode. 3. Rotate the alignment tool so that the feeler disk touches or almost touches the analyzer flange. Proper alignment is achieved when the feeler disk touches the analyzer flange and the alignment tool is perpendicular to the flange. 4. Replace the Belleville washer so that the crown side is facing upwards. NOTE: When reinstalling the trap assembly, make sure that you orient the Belleville washer crown side up. Tighten the nut until the Belleville washer is flat, for example, until the nut bottoms out. 5. Replace and tighten the nut until it is snug. 6. Attach filament wires 1, 2, and C, to the flange feed-through pins.
60
61
6. Push the nose in, rotating it clockwise to lock it in place. 7. Connect the transfer line heater cable.
Turning On the MS
To turn on the MS, do the following: 1. Make sure that the power switch on the back of the MS is OFF. 2. Check that all heater cables are plugged in. 3. Plug in MS power cable. 4. Turn ON the MS power switch. 5. Open System Control on the computer. The most recently set instrument parameters are loaded into the MS. The software stays on the shut down page until the MS fully restarts.
6. Briefly press the analyzer assembly to ensure a good vacuum seal. 7. Replace the MS system cover.
62
Moving the MS
210-MS or 220-MS
To move the 210-MS or the 220-MS, do the following: 1. Shut down the GC and MS. 2. Turn off the GC and computer. Then unplug the GC, MS, and all other power cords. 3. Open the vent valve lever on the front of the MS for ten minutes. 4. Watch the capillary column inside the GC as you gently slide the MS away from the GC. Do not bend or kink the capillary column. 5. Use the alignment tool to prevent the transfer line from turning while you loosen the brass capillary nut connecting the column to the transfer line.
63
6. Cap the transfer line with a capillary nut and no-hole ferrule. 7. Place the capillary column and nut inside the GC oven to protect them from damage. 8. Turn off the carrier gas, and then disconnect the helium gas line connected to the GC filter. 9. Cap the filters with Swagelok plugs or caps. 10. Move the MS to its new location. Be sure the new location satisfies the power and environmental requirements described in the Pre-installation Instructions.
225-MS
To move the 225-MS, contact your Varian Service Representative.
64
Troubleshooting
Checking the GC
Run a test sample to check operational and performance factors, including the carrier gas supply, chromatographic characteristics, and sample-related problems. The test sample that is most frequently run is the COLTEST mixture. This multicomponent mixture is suited to troubleshooting injector and column problems. Please see Running the COLTEST Sample on page 83. See the GC manuals for information about fixing GC problems. Ensure that you are thoroughly familiar with all safety issues before you repair any electronics component.
Checking the MS
If your data system and GC are operating normally, the problem could be caused by the MS or by the communication channel between it and the data system. Typical problems with the ion trap include lack of response (no spectra), low response, poor resolution, and mass miss-assignment. Two procedures isolate problems associated with the MS. Auto Tune provides information about system performance. The diagnostics program tests the hardware and helps to isolate simple ion trap problems, for example, air leaks, burned-out filaments, and high contamination levels
65
NOTE: If diagnostics fail, after the problem is corrected, click the Reset button before doing further testing. Sometimes, you may need to separate the GC and MS to isolate an ion trap problem. In these cases, remove the column from the injector, and plug its end with a septum. This minimizes the input of air. Maintain the column and transfer line at ambient temperature to prevent degradation of the stationary phase. You do not need to vent the MS vacuum system to complete this procedure. To isolate the MS further, remove the column from the ion trap by shutting down the system and capping the transfer line with a no-hole ferrule.
No Spectrum
NOTE: Bake out the MS for at least 2 hours before doing troubleshooting. If a spectrum fails to appear when you click the ion trap icon in the Instrument Control Page, regardless of mass range, 1. Run Diagnostics to detect hardware problems. If you have done this, resolved any hardware problems, and the missingspectrum problem persists, continue with the following procedures. These procedures apply if the air/water or the cal gas peaks are missing. 2. Investigate the following: The method segment is a FIL/MUL Delay and ionization is EI (AUTO or FIXED) mode. During FIL/MUL Delay the trap icon is red. The filament is open. The turbomolecular pump has stopped. An RF adjustment is required. The instrument parameters are inappropriate. The trap has been incorrectly assembled. There is a problem with the electronics. The system has not finished baking out.
66
67
4. If you are unsure of appropriate levels, then reset the parameters by clicking the Defaults button in each section. a. Save your method file as Default. b. Activate Default file, turn on trap and Cal gas. Check for cal gas spectrum. 5. If the spectrum returns, note which parameter(s) were causing the problem. If no spectrum is present, and the trap was recently disassembled, check the trap.
1. If the axial modulation readback is near zero, the trap oven may be scratched and shorting out one of the end caps. Shut down the system, remove the trap oven, and use an ohm meter to check for continuity between the electrodes and ground. Use the screws holding the clamping plate as ground. If this test is done without removing the trap from the electronics assembly, there will be continuity to ground. 2. To check if there is a problem with the electron multiplier, do the following: a. Under Monitor States, select Multiplier. b. Under Acquisition System, check that the electron multiplier voltage is the same or close to the value displayed in the SetPoints tab in the Auto Tune section.
c.
If the electron multiplier voltage in the Diagnostics is only a few volts, the multiplier is shorted to ground. Shut down the system, and replace the electron multiplier or call a Varian Customer Support Representative.
68
NOTE: Bake out the MS for at least 2 hours before doing troubleshooting. If the problem persists, do the following: 1. Check for an air leak in Auto Tune section. 2. Check the RF ramp Adjustment. 3. Reduce trap temperature to 150 C. 4. Enter Method Builder and then check that the method contains EI AGC (Automatic Gain Control) ionization mode, and Default values for other parameters. NOTE: If, after performing these tests, you are cannot isolate the cause of the problem, contact your Varian Customer Support Representative.
69
NOTE: Bake out the MS for at least 2 hours before doing troubleshooting.
70
71
If the baseline ramps up, shut down the MS and then carefully clean the electrode surfaces with a lint-free cloth.
72
Often, major air leaks are accompanied by a hissing sound. These leaks may be due to extremely loose fittings, improperly seated O-rings, or open valves. If you suspect a major leak, do not turn on the electron multiplier, RF voltage, or filament. Using the Diagnostics section, confirm that the turbomolecular pump is operating at 100% speed. If it is not, there is a major air leak.
4. Open System Control and then click Auto Tune button. 5. Select Air/Water Check. 6. Click Start Auto Tune. 7. Compare your air/water spectra to the following:
If the peaks at masses 32 ( O 2 ), 28 ( N 2 ), and 18 ( H 2 O ) are severely broadened or undifferentiated, the system has a large air leak. Immediately turn off Air/Water Check.
73
Air/Water Spectrum from a System with a Very High Water Vapor Background
74
Air/Water Spectrum from a System with Excess Water Vapor and a Relatively Small Air Leak
If the ratio of the height of the peak of mass 18 (H2O+) to mass 19 (H3O+) is about 10:1, there is little water vapor in your system. If the ratio of peak height of mass 18 to mass 19 is less than 10:1 but greater than 5:1, additional bakeout maybe necessary. If the water vapor is not eliminated, sensitivity and performance may be less than optimal. If the ratio of the peak height of mass 18 to mass 19 is less than 10:1, your system contains excess water vapor.
Air/Water Spectrum Obtained from a System with No Significant Air Leaks and Little Water Vapor
This spectrum is indicated by: The peak at mass 18 (H2O+) may be the base (highest) peak. This is dependent on the level of water vapor. The ratio of the peak height at mass 18 (H2O+) to that at mass 19 (H3O+) is greater than or equal to 10:1. The 100% counts value is significantly lower than 500. The ratio of the peak height at mass 28 to that at mass 32 (O2+) is about 4:1.
8. If there are no air or water leaks, the following approximate values should be obtained. Note that these values vary from system to system. 100% value
<100
TIC
<1000
18:28 ratio
~ 1:1
19:18 ratio
10 to 15%
28 width
< 1 m/z
75
Air/Water Spectrum Obtained from a System with a Small Air Leak and Little Water Vapor
This spectrum is indicated by: The peak height at mass 28 is noticeably greater than that at mass 18. The ratio of the peak height at mass 28 to mass 32 is greater than 4:1. The 100% scale counts value increased to greater than 500. The ratio of the peak height at mass 18 to mass 19 is greater than or equal to 10:1.
Air/Water Spectrum Obtained from a System with a Moderate Air Leak and Little Water Vapor:
This spectrum is indicated by: The peak at 28 starts to overload. The 100% counts value may be several thousand counts. The peak height at mass 18 is greater than that at mass 19.
76
Air/Water Spectrum Obtained from a System with a Large Air Leak and Little Water Vapor:
This spectrum is indicated by: The peak at mass 32 is the base (highest) peak. The peaks at masses 18, 19, and 28 are broadened. As a leak increases, all peaks broaden and eventually become undifferentiated.
High water backgrounds may be observed after venting the system especially after cleaning the trap. Several hours of bakeout may be required for the water vapor to desorb from surfaces in the vacuum system, and for the water level to drop to a stable level. Never operate MS if the mass 18 and 19 peaks are the same height (or if the air/water check shows NO). After the system has baked out sufficiently (for example, overnight) and if water vapor in the system is still detected, there may be a contamination in the carrier gas tank or an air leak. Saturated filters on the GC may increase the air/water background. Replace the filters at regular intervals, and when moisture, or other background contaminantion from the GC, becomes a problem.
77
If the leak is at the injector, the system does not need to be vented. Wait until all GC zones are cool before beginning. If the leak is coming from the transfer line connection, shut down and vent the system before fixing it. NOTE: if you are using an argon leak-detection gas set the mass range from 35 to 50. If you are using a Freon leak-detection gas set the mass range from 80 to 110. Troubleshoot leaks using argon gas as a leak detecting gas. The mass peak of interest for argon is at mass 40. To reduce the risk of damaging the filaments or multiplier, develop a method file with the following parameters: 1. Set the electron multiplier 100 V below the 105 setting. 2. Turn off AGC and set the ion time to 100 sec. 3. Set the filament emission current to 10 A. 4. Set scan range from m/z of 35 to 50 (or 80 to 110). 5. Open System Control, and then activate the argon method for troubleshooting and turn the trap ON. NOTE: Do not spray argon indiscriminately around the fittings. Argon diffuses very rapidly from the fitting you are testing toward a true leak. This could lead you to mistakenly identify the fitting being tested as the leak source. To check for leaks: Spray a fine stream of argon on the transfer line closest to the analyzer. Examine the monitor for a response. If a peak at mass 40 does not appear, there is no leak. If a peak appears at mass 40, there is a leak. The transfer line O-ring may have particles on its surface. Shut down the system and check the O-ring.
Check the following gaskets and fittings for leaks, one at a time and in the following order. Tighten the fittings and/or flanges as needed. Wait a few seconds between subsequent applications of argon. 1. Cal gas tube fitting on the pneumatics manifold 2. Vent valve fitting on the manifold 3. Top vacuum manifold flange
78
If the brass nut on the transfer line is not tight enough, ensure the nut is tight, but do not over tighten the fittings. Otherwise, you may generate an even larger leak. Then, recheck the system If you cannot eliminate the leak, vent the system, and check the O-ring on the manifold and transfer line for particles. Wipe off the O-rings with lint-free paper. The turbomolecular pump will probably fail to achieve its 100% speed if there is a leak or poor seal at the turbo/manifold interface. Never attempt to operate the system under these conditions.
Checking GC Connections
NOTE: Check the GC Maintenance Section for additional information for trouble shooting leaks. To identify and correct a leak at the connections between the capillary column and the injector or transfer line, do the following: Make sure that the ferrules have the correct size, for example, 0.4 mm for 0.25-mm ID columns, and 0.5 mm for 0.32 mm ID columns. Make sure that the ferrule on the transfer line is a graphite/Vespel mixture. Most transfer line connection leaks occur on the high vacuum side such as around the transfer line O-ring. Graphite/Vespel ferrules: tighten each ferrule one-half turn beyond finger tightness. Graphite ferrules: tighten each ferrule three-quarters of a turn beyond finger tight.
79
Leaks at the septum may arise from loose injector nuts or a worn septum. Insert a new septum as part of your routine GC preventive maintenance program. To reduce the level of air bleeding into the system and background from the septum material, use good quality, low-bleed septa. Air leaks in the GC pneumatics are the most difficult leaks to detect and eliminate, because detection gases are not particularly effective for this purpose. Tighten all fittings, and then check for leaks using a solvent such as methanol. Saturated filters on the GC may increase the air/water background. Replace the filters regularly or when moisture or other background material from the GC becomes a problem.
80
4. Wrap the transfer line in clean aluminum foil and place it on a clean, dry surface. 5. Cover the analyzer opening with aluminum foil. To remove the capillary column from the GC injector, do the following: 1. Use a 5/16 in. wrench to loosen the capillary column nut that secures the column to the injector. 2. Carefully remove the nut, ferrule, and column from the injector. 3. Slide the column nut, along with the ferrule, off the end of the column. 4. Remove the ferrule from the column nut. Discard the ferrule. 5. Carefully lift the column support cage, along with the column, from the column hanger. Then, remove the support cage and column from the oven. 6. Seal the end of the column or insert the ends of the column into a septum. 7. Store the column and the support cage.
81
To position the column in the transfer line, do the following: 1. Install the brass nut on the end of the transfer line, but do not tighten the nut completely. 2. Keep an eye on the tip of the column and position it so that about 1 mm (1/32 in.) of the column projects from the transfer line tip. NOTE: As you tighten the nut, the position of the column in the transfer line may change. If this happens, loosen the nut and readjust the column until about 1 mm (1/32 in.) of the column projects from the transfer line tip. 3. Grasping the transfer line securely with the alignment tool, use a 5/16-in. wrench to tighten the brass nut. Tighten the nut until snug, but do not over tighten. 4. Rotate the transfer line so that the heater cable projects downward.
To install the transfer line in the manifold, do the following and refer to the Transfer Line Exploded View on page 16: 1. Position the transfer line in the manifold, and install the clip into the holes and slots. 2. Gently push the transfer line toward the manifold, and rotate the collar in the clockwise direction until the bayonet lock engages. 3. Reconnect the transfer line heating cable to the MS. 4. Gently push the MS toward the GC, until the transfer line boot fits snugly over the collar on the side of the GC oven. NOTE: The capillary column nut should be visible inside the GC column oven. 5. The MS is properly engaged when the bumpers on the left side of the spectrometer achieve full contact with the right side of the GC. 6. Replace the cover on the MS.
Troubleshooting the GC
NOTE: Please refer to the GC Operators Manual for information about GC troubleshooting and diagnostics procedures not described in this section. This section describes chromatographic troubleshooting. You will be able to see most of the problems addressed in this section by running the COLTEST mixture, part number 392027300. The following procedure describes the chromatographic conditions and the expected results when running the COLTEST sample with a 30-m DB-5 column (0.25 mm ID, 0.25 m film thickness).
82
Time
0.00 0.50
Event 1
On Off
Injector Mode
Splitless Split
83
Compound
decane 1-octanol undecane nonanal 2,6-dimethylphenol 2-ethylhexanoic acid 2,6-dimethylaniline decanoic acid, methyl ester undecanoic acid, methyl ester dicyclohexylamine dodecanoic acid, methyl ester hexachlorobenzene
Formula
C10H22 C8H18O C11H24 C9H18O C8H10O C8H16O2 C8H11N C11H22O2 C12H24O2 C12H23N C13H26O2 C6Cl6
Integer Weight
142 130 156 142 122 144 121 186 200 181 214 282
Quantitation Mass
57 69 71 67 107 73 106 74 87 138 143 284
The following is a typical chromatogram for this test mixture. Note that 2, 6dimethylphenol and 2-ethylhexanoic acid coelute normally on a DB-5 column, depending on column and injector.
84
The following figure demonstrates the resolving power of the MS for coeluting compounds.
Effectively separate the individual components in the mixture for subsequent data manipulation, such as library searches and quantitation. For details about plotting single ion chromatograms for ions specific to a single compound, please refer to the online help or section in the Software Reference manual.
Solution
Reinstall the column in the injector. Check the column seal with the insert in the SPI/1079 (on-column). Make sure you have a good cut on the column by examining the column under magnification. Check the 1077/1079 injector for insertion depth. Reduce the injection speed for the hot injectors. If possible, reduce the injector temperature. If you are using sandwich injection, reduce the solvent plug to 0.5 L. A typical setting for the SPI is 20 C below the solvent boiling point. The column temperature is set at the solvent boiling point. Hold the column at this temperature
85
until SPI has finished heating (usually about 2 min). Septum purge line plugged Check that the septum purge flow is 2 mL/min for a 1177 or 1079. If necessary, adjust the valve setting (depending on the injector configuration). For a splitless injection, the vent flow should be at least 70 mL/min. The injector should switch to the split mode 30 to 90 sec after the injection.
Solution
Change or clean the injector insert. If necessary, use a deactivated insert. Remove the front 15 cm of the column and reinstall it. Replace the column if the retention times change, or if cutting the column does not fix the problem,
Solution
Increase the temperature of the injector Condition the column at 30 C below its maximum operating temperature (320 C for DB-5). Switch to a high temperature column, (e.g., the SGE HT5), if conditioning does not help. Clean the ion trap as outlined in the Maintenance Section. Check Contamination Table for listing of potential contaminations. Lower the injector temperature and the injection speed. Check that the graphite ferrule in the 1079 is free of cracks, and that the septum support is tight. Increase the trap temperature in increments of 20 C.
High levels of silicone or other contamination coated on the ion trap surfaces Insufficient vaporization of the higher boiling point components
Symptom
Column overload due to injection of excessive amounts of a component Degradation of the stationary phase Carrier gas velocity too low
Solution
Dilute the sample, or perform a split injection. Change the column. Increase the carrier flow rate.
86
Solution
Modify the method (e.g., slow the column ramp rate) to improve the separation Decrease the carrier gas linear velocity to improve the resolution. Use a more polar column. Replace the column.
Solution
Visually check that the syringe is pulling up the sample. Check that the injector nut is tight. Flush the syringe with solvent. Heating the solvent in a hot injector may help if the syringe is plugged; otherwise, replace the syringe. Replace the septum regularly and ensure that the septum nut is tight. Check the installation of the column in the injector. Tighten the capillary column nut. Change the injector insert. Remove the front 15 cm of the column, or replace the column. Increase the injector temperature. Or, increase the maximum temperature to which the injector (1079) is programmed. Confirm that the switch time is chromatographically optimized.
Leak at the septum Improper installation of column in the injector, or a leak at the column inlet Sample being absorbed by active surfaces in the injector or column Incomplete vaporization of sample in the injector 1177 or 1079 splits too soon
Solution
Lower the injection temperature, or use a 1079 programmed injection.
Column temperature programming starts before Increase the initial column hold time 1079 has finished programming. until 1079 reaches its maximum temperature, (for example, typically at 2 min.). Column is cracked. A piece of septum is stuck in the injector insert. Re-cut and install the column. Replace the insert and septum.
87
Solution
Use high-temperature, low-bleed septa. Make sure that the septum purge flow is set to 2 mL/min for a 1177 or 1079 injector. Confirm impurities by running a solvent blank with a new syringe. Use certified sample vials, and keep the samples refrigerated. Check contamination table. Install or replace the carrier gas filters. Remove the column from the injector and bake it out at elevated temperature, (e.g., 350 C) using a purge of at least 20 mL/min. Confirm that this is indeed the case by running a blank or standard. Switch to a new septum type, lower the injection temperature, or reduce the injection volume. Use high purity grade solvents.
Impurities present in the sample Solvents extract impurities from the septum.
Solution
Check the pneumatics for leaks. If necessary, replace the flow controller/ regulator. Condition or replace the column. Replace the septum at regular intervals. Check that the septum nut and capillary column nut are tight.
88
Documents
The following documents have more information about the MS: Software Operation Manual, part number 395414500. Software Reference, part number 391496300. MS Workstation Tutorial Manual, part number 391498800. Pre-Installation Instructions, part number 395414200. Release Notes, part number 391496201.
Description
USB Cable PWA, POWER BOARD PWA, RF GENERATOR Cable, Controller to PWR 26 pins (Ribbon) Cable, Controller to PWR 64 pins (Ribbon) Replacement Spares Kit Assembly, Analyzer Flange Assembly, Transfer Line (115V) Assembly, Transfer Line (230V) Cable, Transfer Line heater (115V) Cable, Transfer Line heater (230V) Assembly, Vacuum Manifold (115V) Assembly, Vacuum Manifold (230V) Assembly, Ion Gauge Cable, Trap Heater (60V) PWA, Controller PWA, Lower Manifold PWA, Upper Manifold
89
Trap Components
NOTE: The Silica-Coated Spacers have a shiny, mirror like finish on the inside surface. Part Number
393055201 393055101 1492000900 393031501 393053501 393010801 393060191 393054901 393059191 393050001 393052401 393053502 1312200800 1499822800 393053901 393010904 1222200606 393010903 393010914 393010918
Description
Gate Conductor Gate Wavy Washer Assembly, Multiplier Spacer, Quartz Transfer Line Wrench/Analyzer Alignment Tool Assembly, Filament disk with wires Filament Clip Tip. Transfer Line (Ultra Clean) Trap Oven Clamping Plate Quartz Spacer, Silica-Coated Nut, 8-32 X 11/32" Belleville Washer, Large Thermo Well Thermo Well O-ring Trap Oven screw 6-32 X 3/8 O-ring, 1.112 ID Transfer Line Quad-ring, Viton Manifold Quad-ring, Viton Transfer Line
Description
DS-42 rotary vane pump, 120V DS-42 rotary vane pump, 240V Screen, Turbo Pump (V-81) Cable, Turbo Controller to turbo Turbo Controller 7 Length Tygon Tubing Turbo Molecular Pump (V-81) Premium Foreline Pump Oil (DS-42) DS-42 Oil Mist Eliminator Foreline Pump Oil (1L) for DS-102 Oil Mist Cartridge, 2/pk (DS-102) O-ring, Turbo Pump to Manifold IPS module for the 225-MS
90
GC Spares
Part Number
2869458001 394955100 200003400 190015800 390842300 7200008400 8850103100 CR298777 392534201 CR298713 392611927 392611936 391866308
Injector Type
1079 or 1177 1079 or 1177 1079 or 1177 1079 or 1177 1079 or 1177 1079 or 1177 1079 or 1177 1079 1079 1177 1177 1177 1177
Description
0.4 mm Graphite / Poly Ferrules Capillary Injector Nut Carrier Gas Line Assembly Ceramic Scoring Wafer Injector Nut Wrench Septa Extraction Tool Viton O-rings, 25/pk BTO Septa 11.5 mm, 50/pk Ferrule Insert Graphite 9 mm Septa Inlet Sleeve, Gooseneck 4 mm Open Inlet Sleeve, Gooseneck Glass Wool Screw Captive Micro Seal
Description
FC-43, Reservoir (Cal Gas Bulb) COLTEST sample OFN test sample Aluminum Oxide, 600 Grit GC/MS Calibration Compound, FC-43 Fuse, 5 x 20 mm, 0.5A Applicator, Cotton Tipped, pkg. 100 Solenoid, 3-way, Cal Gas Needle Valve, Cal Gas
CI Parts/Spares
Part Number
393010202 393059701 393059601 393010101 393002291 393010601
Description
Solenoid, 2-way, CI Restrictor, long, CI Restrictor, short, CI Needle Valve, CI Gas Liquid CI Inlet Kit CI Solenoid, 2-way, Chemrez
91
Varian Service
If you are unable to resolve a problem with your MS, call a Varian Customer Support Representative. When you call, you must provide the following information: MS serial number, which is inside the front panel. Installed options. Diagnostics test results.
If you are having problems with the gas chromatograph, provide the following information: GC model. AutoSampler model, if any Type of injector in use. Cryogenics (if applicable). Information about your GC column, (for example, the manufacturer, bonded phase, film thickness, and ID and length).
If you are having problems with your computer or software, provide the following information: Computer manufacturer and model. Windows version. Mouse driver version. Printer manufacturer and model. Network configuration. Printouts of your autoexec.bat and config.sys files. MS Workstation software version.
Observe the following guidelines when describing the problem to the Customer Support Representative: Tell the service representative which part of the software, you were using when the problem occurred. Tell the Support Representative which troubleshooting routines you used.
92