Methodology

This document outlines a methodology to test the accuracy and precision of pipet operators using an analytical balance. It involves pipetting various volumes of distilled water, recording the weights, repeating several times, and computing average weights and standard deviations. Accuracy is tested by pipetting different volumes and computing averages and deviations. Precision is tested by having multiple students pipette the same volumes and computing averages, percent errors, and standard deviations to determine variability between operators.

Uploaded by

Cara Jaen
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
29 views

Methodology

This document outlines a methodology to test the accuracy and precision of pipet operators using an analytical balance. It involves pipetting various volumes of distilled water, recording the weights, repeating several times, and computing average weights and standard deviations. Accuracy is tested by pipetting different volumes and computing averages and deviations. Precision is tested by having multiple students pipette the same volumes and computing averages, percent errors, and standard deviations to determine variability between operators.

Uploaded by

Cara Jaen
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
You are on page 1/ 5

METHODOLO

GY
TESTING
ACCURACY OF
OPERATORS
ANALYTICAL
BALANCE
- Turn on analytical balance
- Place parafilm/ aluminum foil on
- weighing balance
- Press TARE

Pipet 50l Pipet 100l Pipet 500l


distilled distilled distilled
water water water

- Record weight. - Record weight. - Record weight.


- Press TARE. - Press TARE. - Press TARE.
- Repeat five - Repeat five - Repeat five
times. times. times.
Compute for Compute for Compute for
AVERAGE AVERAGE AVERAGE
WEIGHT WEIGHT WEIGHT
STANDARD STANDARD STANDARD
DEVIATION DEVIATION DEVIATION
TESTING
PRECISION OF
OPERATORS
ANALYTICAL
BALANCE
- Turn on analytical balance
- Place parafilm/ aluminum foil on
- weighing balance
- Press TARE

Pipet 50l Pipet 100l Pipet 500l


distilled distilled distilled
water water water

- Record weight. - Record weight. - Record weight.


- Press TARE. - Press TARE. - Press TARE.
- Repeat for each - Repeat for each - Repeat for each
student student. student

Compute for Compute for Compute for


AVERAGE WEIGHT AVERAGE WEIGHT AVERAGE WEIGHT
PERCENT ERROR PERCENT ERROR PERCENT ERROR
STANDARD DEVIATION STANDARD DEVIATION STANDARD DEVIATION
STANDARD ERROR OF STANDARD ERROR OF STANDARD ERROR OF
MEAN MEAN MEAN