Validation of Analytical Methods
Validation of Analytical Methods
METHODS
Kelompok 10
1. Hani Hazarani (1311012022)
2. Novi Bakri (1311012025)
3. Hafsa Rahmi Latifa (1311012027)
4. Tri Ayuningtias (1311012029)
5. Poppy Agustin (1311012045)
Method Validation
analytical data.
• Validation is the formal and systematic proof that a method compiles with
the requirements for testing a product when
observing a defined procedures.
(G. Maldener, Chromatographia, July 1989)
2009 2
• Method validation is the process of demonstrating that analytical
procedures are suitable for their intended use and that they support the
i d e n t i t y, s t r e n g t h , q u a l i t y, p u r i t y a n d p o t e n c y o f t h e
2009 3
Examples of Methods That Require
Validation Documentation
2009 4
Considerations Prior to
Method Validation
Suitability of Instrument
• Status of Qualification and Calibration
Suitability of Materials
• Status of Reference Standards, Reagents, Placebo Lots
Suitability of Analyst
• Status of Training and Qualification Records
Suitability of Documentation
• Written analytical procedure and proper approved
protocol with pre-established acceptance criteria
2009 5
Validation Step
• Define the application, purpose and scope of the method.
• Analytes? Concentration? Sample matrices?
• Develop a analytical method.
• Develop a validation protocol.
• Qualification of instrument.
• Qualify/train operator
• Qualification of material.
• Perform pre-validation experiments.
• Adjust method parameters and/or acceptance criteria if necessary.
• Perform full validation experiments.
• Develop SOP for executing the method in routine analysis.
• Document validation experiments and results in the validation report.
2009 6
Purpose of Method Validation
2009 7
What is not Analytical Method Validation?
• Calibration
2009 8
• System Suitability
Test to verify the proper functioning of the operating system,
analytical operations.
2009 9
System Suitability
Validation
Calibration
Pump Injector
Detector Data System
Analyst Method
Sample
2009 10
Method Life Cycle
Validation
Development Optimization
2009 11
Verification vs. Validation
– Compendial methods-Verification
2009 12
Compendial Analytical Procedures
• The Analytical procedures in the USP 25/NF 20 are legally recognized under section
501(b) of the Federal Food, Drug and Cosmetic Act as the regulatory analytical
procedures for the compendial items. The suitability of these procedures must be
verified under actual conditions of use. When using USP 25/NF 20 analytical
f o l l o w i n g
characteristics:
– Intermediate precision
2009 13
Regulatory and Compliance
Requirements Review
USP 23 General
Information <1225>
2009 14
• The accuracy, sensitivity, specificity, and reproducibility
of test methods employed by the firm shall be
established and documented. Such validation and
documentation may be accomplished in accordance with .
2009 15
• The objective of validation of an analytical procedure
is to demonstrate that it is suitable
for its intended purpose
2009 16
• In practice, it is usually possible to design the experimental
work such that the appropriate validation characteristics can
be considered simultaneously to provide a sound, overall
knowledge of the capabilities of the analytical procedure, for
instance: Specificity, Linearity, Range, Accuracy, and
Precision.
2009 17
Today’s Validation Requirements
ICH/USP
GMPs
(legal) FDA
2009 18
ICH/USP Validation Requirements &
Parameters
USP ICH
• Specificity
Specificity
• Linearity
Linearity and Range
• Range
Accuracy
• Accuracy
Precision
• Precision
Limit of Detection
– Repeatability
Limit of Quantitation
– Intermediate Precision
Ruggedness
– Reproducibility
Robustness
• Limit of Detection
• Limit of Quantitation
2009 19
Specificity/Selectivity
• Ability of an analytical method to measure the analyte free from
interference due to other components.
2009 20
Specificity: Impurities Assay
• Chromatographic Methods
– Demonstrate Resolution
• Impurities/Degradants Available
– Spike with impurities/degradants
– Show resolution and a lack of interference
• Impurities/Degradants Not Available
– Stress Samples
– For assay, Stressed and Unstressed Samples should be compared.
– For impurity test, impurity profiles should be compared.
2009 21
Linearity
• Ability of an assay to
elicit a direct and
proportional
response to changes
in analyte
concentration.
2009 22
Linearity Should be Evaluated
2009 24
Accuracy
25
Accuracy
2009 26
Precision
sample.
2009 27
Precision… Considered at 3 Levels
• Repeatability
• Intermediate Precision
• Reproducibility
2009 28
Repeatability
2009 29
Intermediate Precision
confidence interval.
2009 30
Reproducibility
31
Detection Limit (LOD)/ Quantitation
Limit (LOQ)
• LOD LOQ
Lowest amount of analyte in a Lowest amount of analyte in a
Ratio of 10:1.
2009 32
LOD and LOQ Estimated by
2009 33
LOD and LOQ Estimated by
3.3s 10s
DL = QL =
S S
Ybl
LOD LOQ
2009 35
Robustness
2009 36
Ruggedness
2009 37
ICH/USP System Suitability
• ICH
– Definition: evaluation of equipment, electronic,
analytical operations and samples as a whole
– Determination: repeatability, tailing factor (T),
capacity factor (k’), resolution (R), and
theoretical Plates (N)
2009 38
Re-validation
• When
– Method parameters have been changed
– The scope of the method has been changed
– Synthetic methods have been changed
– Impurity profile has been changed
• What
– Preferably everything. Exceptions should be
scientifically justified
2009 39
References
• USP 23 General Information
• ICH Guideline for Industry Q2B, Validation of
Analytical Procedures: Methodology
• ICH Guideline for Industry Q2A, Text on Validation
of Analytical Procedures March 1995
• 21 CFR PART 211 - CURRENT GOOD
MANUFACTURING PRACTICE FOR FINISHED
PHARMACEUTICALS
THANK YOU