Physiotherapy and Electrotherapy Equipment
Physiotherapy and Electrotherapy Equipment
Equipment
Any short-wave therapy unit would give out the desired energy to the
patient only if & as long as, the unit is correctly tuned to the electrical
values of the part of the body.
Therefore, tuning must be carefully carried out at the beginning of the
treatment & continuously monitored during the treatment.
There is a possibility of the tuning getting affected due to unavoidable but
involuntary movements of the patients & the resultant fall of dosage.
In order to overcome the problem of making tuning adjustments during the
course of treatment, an additional circuit is fitted in the machine.
The RF current in the patient circuit changes a capacitor to a voltage, whose
polarity and magnitude is a measure of the detuning of the patient circuit.
This voltage accordingly moves a servo-motor, adjusting the tuning capacitor
so that resonance is restored.
Application Techniques of Short-wave
Therapy
Condenser Method
Why?
A severe limitation of diathermic machines is that they
direct continuous high frequency radio waves, & if a high
enough output of energy is sustained for even a brief time,
they can cause burns.
Thus, the wattage has to be lowered to tolerable limits.
Also, the heat resulting from diathermy has many
contraindications & limitations because the heat limits the
amount of energy that can be used.
The increase in energy output, while avoiding the dangers
of heat, achieved thru 'Pulsed Shortwave Therapy' (PSWT).
How Does It Works???
• Pulsed therapy apparatus works at 27.12 MHz, the
frequency of short-wave diathermy machine.
• However, the energy is delivered in the form of pulses of 65
ms with an interval between pulsations at a maximum
setting of 1600 ms.
• The rate of pulsations is adjustable in steps from 80-600
pulses per second.
• At this setting, the power is provided no more than 4% of
the total time during operation.
• The peak instantaneous wattage can be varied from 290 to
975 W.
• The effect of the rest periods is to reduce the output to a
maximum average of only 40 W.
• The result is an intermittent, relatively athermic,
electrotherapy
Despite high energy pulses, the heat is dissipated during the rest period, there is no
danger of burns or hyperthermic complications.
The depth of penetration depends upon the peak energy delivered, which is
adjustable.
Tuning adjustment is provided to obtain maximum efficiency at each wattage setting.
The control will enable the user to vary (a) the mean power delivered to the patient
and (b) the pulsing parameters governing the mode of delivery of the energy.
Two basic output :the electric field, comparable to the condenser (capacitor) field in traditional
shortwave diathermy & secondly, the magnetic field, comparable to inductothermy.
MICROWAVE DIATHERMY
• Microwave diathermy consists in irradiating the tissues of the patient's body
with very short wireless waves having frequency in the microwave region.
• The most commonly used microwave frequency for therapeutic heating is
2450 MHz corresponding to a wavelength of 12.25 cm.
• The heating effect is produced by the absorption of the microwaves in the
region of the body under treatment.
• Better results are obtained by using microwave compared to shortwave
diathermy technique.
• The microwaves are transmitted from an emitter, & directed towards the
portion of the body to be treated.
• The special design of the treatment heads & shapes focuses the field directly
at the target area. The whole device is used to direct the waves onto the
tissues.
• In this kind of treatment, the patient does not form a part of the circuit, so no
tuning is necessary as in short wave application.
• Microwave penetrates more than infrared rays but not deeply as shortwave
diathermy. Not suitable for deeply placed structures.
The microwaves are strongly absorbed by water, tissues with high fluid content are heated most,
Moreover, there is an appreciable heating of tissues having good blood supply such as muscle.
The Delay Circuit:
It is necessary for the magnetron to warm up for 3 to 4 minutes before power may be derived
from it.
A delay circuit is incorporated in the apparatus which connects the anode supply to the
magnetron only after this time elapses.
The arrangement is such that a lamp lights up after 4 minutes indicating that the apparatus is
ready for use.
The Magnetron Circuit:
The magnetron filament heating voltage is obtained directly from a separate secondary winding
of the transformer.
The filament cathode circuit contains interference-suppression filters.
The anode supply to the magnetron can be either DC or AC. A DC voltage is obtained by a full
wave rectifier followed by a voltage doubler circuit.
A high wattage variable resistance is connected in series which controls the current applied to
the anode of the magnetron.
When using AC, the voltage is applied to the anode of the magnetron through a series
connected thyratron so that the AC voltages of both tubes are equal in phase.
By shifting the phase of the control grid voltage with respect to the phase of the anode voltage,
the amount of current through the magnetron can be determined and thus the output power
can be varied.
The phase shift can be achieved by using a capacitor resistor network.
Safety Circuits & Precautions:
There are chances of the magnetron being damaged due to an excessive flow of current.
It is thus protected by inserting a fuse (500 mA) in the anode supply circuit of the magnetron. The
protection of both the patient and the radiator is ensured by the automatic selection of the
control range depending on the type of the radiator used.
Excessive dosage can cause skin burns, the skin should be dry , duration of irradiation 10-25 mins
ULTRASONIC THERAPY UNIT
The heating effect is produced by ultrasonic energy absorption property
of the tissues.
The effect of ultrasonics on the tissues -high speed vibration of micro-
massage.
Massage is used to treat tissue lesions.
Ultrasonic energy enables massage, firstly to a greater depth than
manual, & secondly (in acute injuries) when pressure cannot be exerted
by hand because of intolerable pain.
The thermal effects of ultrasound are dependent on:
the amount of energy absorbed,
the length of time of the ultrasound application
the frequency of the ultrasound generator.
The electrical power required is usually less than 3 W/cm2 of the
transducer area that is in contact with the part of the body to be
treated.
ULTRASONIC THERAPY UNIT
• Ultrasonic generators are constructed on the piezo-
electric effect.
• A high-frequency alternating current (e.g., 0.75-3.0
MHz) is applied to a crystal whose acoustic vibration
causes the mechanical vibration of a transducer head,
which itself is located directly in front of the crystal.
• These mechanical vibrations then pass through a metal
cap & into the body tissue through a coupling medium.
• The therapeutic ultrasonic intensity varies from 0.5 to
3.0 W/cm2 .
• Applicators range from 70 to 130 mm in diameter. The
larger the diameter of the applicator, the smaller would
be the angle of divergence of the beam & the less the
degree of penetration.
The heart of the system is a timed oscillator which produces the
electrical oscillations of the required frequency.
The oscillator output is given to a power amplifier which drives the
piezoelectric crystal to generate ultrasound waves.
Power amplification is achieved by replacing the transistor in
typical LC tuned Colpitt oscillator by four power transistors placed
in a bridge configuration.
The output of the oscillator can be controlled by either of the following
two methods:
• Using a transformer with a primary winding having multi-tapped
windings & switching the same as per requirement;
• Controlling the firing angle of a triac placed in the primary circuit
of the transformer, & thereby varying the output of the transformer.
These curves are determined by means of rectangular & triangular pulses in such a
manner that the threshold values are measured at progressively decreasing
stimulation durations.
The i-t curves have characteristic shapes & deviations from the standard form
which lead to an indication of the state of the tissues.
It also shows the shape of the curve for a totally denervated muscle having an
advanced state of degeneration and required excitability.
With degenerated muscle, the curve obtained is shifted to the right and upwards.
It makes use of a variable rate multi-vibrator (M1) to set the basic stimulus frequency.
The output from the free running multi-vibrator triggers a monostable multi-vibrator (M2) circuit which sets
the pulse width.
The output pulse from the monostable provides an interrupted galvanic output whose rate as well as
duration can be independently controlled.
Another astable multi-vibrator produces short duration pulses called faradic currents.
Faradic currents are usually modulated at the frequency set by the multi-vibrator Ml, in a mixer circuit
(M4).
Since the modulation of Faradic pulses takes place with a slow rate of increase & decrease, the output of M4
is surged Faradic currents.
By integrating the output of M2, the interrupted galvanic pulses can be modified to have an exponential rise
&fall.
The shape of these pulses is similar to a triangular waveform. Galvanic current is also made available by
Transcutaneous Electrical Nerve
Stimulator (TENS)
The electrical impulses required for electrotherapy to treat the pain
are provided by an instrument called TENS (Transcutaneous
Electrical Nerve Stimulator).
Investigations on a great variety of electrical impulse parameters
have indicated that two waveforms, the square wave & the spike
wave are optimally & equally effective in relieving pain.
Most stimulators feature adjustable settings to control the
amplitude (intensity) of stimulation by controlling voltage, current &
the width (duration) of each pulse.
Electrodes are placed at specific sites on the body for treatment of
pain.
The current travels through the electrodes & into the skin
stimulating specific nerve pathways to produce a tingling or
massaging sensation that reduces the perception of pain.
• The stimulator is based around a 500 ms spike pulse, adjustable
amplitude of 0 - 75 mA & adjustable frequency of 12 - 100 pulses
per second.
• Instruments having similar specifications except that they produce
square waveform, have a pulse frequency range of 20–200 Hz,
pulse width from 0.1 to 1.0 ms & pulse amplitude of 0–120 V with
maximum output current as 25 mA.
• The instrument powered by three standard flashlight batteries of
1.5 V each gives about 100 hours of continuous operation.
• Transcutaneous or skin surface application of electrical stimulus is
accomplished by application of the conducting pads to various
trigger- zone areas, acupuncture sites or even peripheral nerves.
• Skin irritation at the site of electrode application is
diminished by the use of carbonized rubber electrodes
applied with a tincture of Benzoin interface.
• The skin electrode system would minimize impedance
variations with motion, to conform to the body surface to
provide a uniform impedance across the surface of the
electrode and to have an adequate surface area.
• (TENS) electrodes are commonly molded from an elastomer
such as silicon rubber, loaded with carbon particles to
provide conductance.
• Conformability is achieved by making the electrode thin.
• Useful carbon-loaded silicon rubbers have a minimum
resistivity near 10 Ω cm.
• The frequency-dependence of the electrode performance
also has to be considered since the impedance between the
electrode & subcutaneous contains capacitance.
Spinal Cord Stimulator
Spinal cord stimulation is a term relating to the use of electrical stimulation of the human spinal
cord for the relief of pain.
This is accomplished through the surgical placement of electrodes close to the spinal cord, either
with leads extending through the skin, or chronically, with the leads connected to an implanted
source of electrical current.
The applied electrical impulses develop an electrical field in & around the spinal cord, which then
causes depolarization or activation of a portion of the neural system resulting in physiological
changes.
The stimulus source provides stimulation pulses at frequencies ranging from 10 - 1500 Hz, with
pulse widths from 100 - 600 μs & controllable amplitude from 1 to 15 mA delivered into a load
from 300 to 1500 Ω.
These parameters can be controlled when one is using an implant that derives power & control
through RF coupling from an externally power unit.
DEEP BRAIN STIMULATION (DBS)
The deep brain stimulation (DBS) therapy is a new treatment technique for a number of
neurologic disorders such as Parkinson's disease.
The system consists of three components: the implanted pulse generator (neurostimulator), the
electrode & the extension
The electrode or lead is a thin, insulated wire which is inserted through a small opening in the
skull & implanted in the brain.
The tip of the electrode is positioned within the targeted brain area.
The extension is an insulated wire that is passed under the skin of the head, neck, & shoulder,
connecting the lead to the neurostimulator which is usually implanted under the skin near the
collarbone.
The stimulator delivers a constant fast-frequency stimulus which interrupts a specific circuit in
the brain that is overactive in the disease state.
This interruption of the diseased overactive circuit can significantly improve the symptoms of
the disease.