DXH 600 - 800 Hardware CTS

The document provides an overview of the aspiration module on the Beckman Coulter DXH 600/800 hematology analyzer. It describes the key components of the aspiration module including the aspiration syringe pump assembly, blood sampling valve assembly, and blood detectors. It then explains the sample acquisition cycle and how blood is aspirated, transported, and detected during analysis.

Uploaded by

Reynaldo Macario
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
674 views52 pages

DXH 600 - 800 Hardware CTS

The document provides an overview of the aspiration module on the Beckman Coulter DXH 600/800 hematology analyzer. It describes the key components of the aspiration module including the aspiration syringe pump assembly, blood sampling valve assembly, and blood detectors. It then explains the sample acquisition cycle and how blood is aspirated, transported, and detected during analysis.

Uploaded by

Reynaldo Macario
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
You are on page 1/ 52

DXH 600/800

HARDWARE

Ver.1 November 2016 Beckman Coulter Technical Product Training . 1 of 52


MODULES
• Aspiration
• CBC
• VCSn
• Data plot Development
• Common Services
• Specimen Transport

Ver.1
Ver.1 November
Ver.1 November 2016Beckman
20162015
January Beckman
Beckman Coulter
Coulter Technical
Technical
Coulter Product
Product
Technical Training
Training
Product .
Training . . 2 of 52
Here is how to navigate this pro

ASPIRATION
MODULE

Ver.1 November 2016 Beckman Coulter Technical Product Training . 3 of 52


OBJECTIVES:

›Sample Aspiration Module


›Aspiration Syringe Pump Assembly
›Blood Sampling Valve Assembly
›Blood Detectors

Ver.1 November 2016 Beckman Coulter Technical Product Training . 4 of 52


SAMPLE ASPIRATION MODULE
• SAMPLE ACQUISITION CYCLE
• 2 sub-cycles
a. Aspiration sub- cycle
1. Air Gap
2. Venting
b. Transport Sub cycle
c. Blood Detection

Ver.1 November 2016 Beckman Coulter Technical Product Training . 5 of 52


SAMPLE ACQUISITION

Ver.1 November 2016 Beckman Coulter Technical Product Training . 6 of 52


Aspiration Syringe Pump Assembly

Ver.1 November 2016 Beckman Coulter Technical Product Training . 7 of 52


ASPIRATION SYRINGE PUMP

• The Aspiration Syringe Pump Module


is controlled by the Steps of a
Stepper Motor
• “Encoder”
• Keeps track of the motor steps
• Each motor step equals a
certain volume of fluid
(i.e. 1µL = 6 steps)
• The speed of the motor in steps
per second controls how fast
the fluids move

Ver.1 November 2016 Beckman Coulter Technical Product Training . 8 of 52


Blood Sampling Valve Module

Ver.1 November 2016 Beckman Coulter Technical Product Training . 9 of 52


BLOOD SAMPLING VALVE MODULE

• Blood Sampling Valve


Module:
• Blood Sampling Valve (BSV)
• Controls the pathways for
blood delivery for both CBC
& VCSn Analysis
• BSV Motor
• Moves the center section of
the BSV
• BSV Sensors
• Monitor the position of the
center section of the BSV
• BSV Animation
• BSV Animation

Ver.1 November 2016 Beckman Coulter Technical Product Training . 10 of 52


Blood Detectors

Ver.1 November 2016 Beckman Coulter Technical Product Training . 11 of 52


BLOOD DETECTORS

• Blood Detectors Monitor the


quality of aspiration
• Probe Blood Detector
• Aspiration Sub-Cycle
• First part of
Transport Sub-Cycle
• BSV Blood Detector
• Transport Sub-Cycle

Ver.1
Ver.1 November
November 20162016 Beckman
Beckman Coulter
Coulter Technical
Technical Product
Product Training
Training . . 12 of 52
BLOOD DETECTOR

Ver.1 November 2016 Beckman Coulter Technical Product Training . 13 of 52


SAMPLE ACQUISITION SUMMARY:
In summary:
1. Air Gap
2. Probe pierces tube and vents
3. Aspirate 165µL blood
4. Blood thru Probe Blood
Detector, stops
5. Transport begins
6. Blood pulled thru BSV then
BSV Blood Detector, stops
7. BSV segments, CBC blood is
pushed to CBC baths
8. Blood pushed back into
Probe, NRBC blood dispense
9. BSV home, DIFF & RETIC
blood dispense
10. Backwash and clean probe

Ver.1 November 2016 Beckman Coulter Technical Product Training . 14 of 52


Here is how to navigate this pro

CBC
MODULE

Ver.1 November 2016 Beckman Coulter Technical Product Training . 15 of 52


CBC TRANSDUCER PANEL

Ver.1 November 2016 Beckman Coulter Technical Product Training . 16 of 52


CBC TRANSDUCER PANEL

Ver.1 November 2016 Beckman Coulter Technical Product Training . 17 of 52


CBC WASTE/VENT ASSEMBLY

CBC Waste Chamber (VC115):


• Waste from CBC Baths,
Count Chambers, HGB
Cuvette
• Supplies Diluent/CBC Lyse
during “Shutdown/Remove
Cleaner”
Vent/Overflow Chamber (VC199):
• Supplies Vent for-
• CBC Waste, CBC Baths,
HGB Cuvette, Diluent
Buffer & Feeder, VCSn
Flow Cell Waste (VC218)

Ver.1 November 2016 Beckman Coulter Technical Product Training . 18 of 52


CBC Module
• CBC Waste Management
• VC115 (CBC Waste Chamber)
• VC199 (Vent/Overflow Chamber)

Ver.1 November 2016 Beckman Coulter Technical Product Training . 19 of 52


CBC Module
• WBC Correction:
• UWBC (uncorrected)
• WBC (corrected)
• Data from WBC Count, VCSn Analysis, WBC Histogram
• Proprietary Advanced Algorithms
• Digital pulse processing provides increased
discrimination using cell pulse characteristics
• Advanced UniCel DxH 800 CBC algorithms utilize
additional cellular measurements for:
• Elevated WBC
• Improved WBC
• Enhanced detection capabilities for identification
of cellular carryover and interferences
• Interference correction is performed
automatically

Ver.1 November 2016 Beckman Coulter Technical Product Training . 20 of 52


HISTOGRAM REVIEW

Ver.1 November 2016 Beckman Coulter Technical Product Training . 21 of 52


WBC Histogram

 X-Axis Lymphocytes
• Size Channels (0-450 fL)
• Most interference occurs at
the 35 fL threshold Granulocytes
 Y-Axis
• Relative number (# of Mononuclear
events)
 3 WBC Populations
Displayed:
• Lymphocytes (35-90 fL) LY GR
MO
• Mononuclear (90-160 fL)
• Granulocytes (160-450 fL)

Ver.1 November 2016 Beckman Coulter Technical Product Training . 22 of 52


PLT Histogram

› See platelet monograph Sample Fitted Curve


Preparation for more info:
DxH 600/800 Platelet Methodology

› Two Platelet Histograms:


» Raw data pulses in the size
range of 2-25 fL counted as Raw Data Curve
platelets
» Normal results- only the
“fitted curve” is displayed
» Abnormal results- only the
“raw data” curve is displayed

Ver.1 November 2016 Beckman Coulter Technical Product Training . 23 of 52


Here is how to navigate this pro

VCSn
MODULE

Ver.1 November 2016 Beckman Coulter Technical Product Training . 24 of 52


VCSn MODULE

A. Sample Flow
• Air Mix Temperature Control Module (AMTC)
• Sample Conditioning (Preparation)
• Distribution Valve (DV)
• Sheath/Sample Pressure
• Gains and Phase

Ver.1 November 2016 Beckman Coulter Technical Product Training . 25 of 52


VCSn MODULE

AMTC

Nice and CLEAN!!!!

Ver.1
Ver.1 November
Ver.1 November 2016Beckman
20162015
January Beckman
Beckman Coulter
Coulter Technical
Technical
Coulter Product
Product
Technical Training
Training
Product .
Training . . 26 of 52
VCSn MODULE

• Distribution Valve (DV)


• DV
• Stepper motor
• Rotates 360o
• 9 positions
• Dual Sensor
• Code Wheel
• DV Function
• Provides Pathways
• Position Sample in sample
line
• Clean pathways
• Remove waste
• Cleaning grooves
• 24 hour auto-cycle

Ver.1 November 2016 Beckman Coulter Technical Product Training . 27 of 52


DISTRIBUTION VALVE (DV)

Distribution Valve
• Common Problems:
• Freeze up
• Symptom:
• Loud grinding
• DV error messages
• Cause:
• Excessive build up
diluent/cleaner

Ver.1 November 2016 Beckman Coulter Technical Product Training . 28 of 52


VCSn MODULE

Distribution Valve (DV)


• Common Problems (cont’d):
• Counter measures
• Dx Tools
• Cycle DV- verifies the
DV motor, clutch, and
optical sensors (right
10 times, then left)
• Unlock DV- helps free
the distribution valve
when it is frozen and
cannot rotate
• DI H2O
• Send service

Ver.1 November 2016 Beckman Coulter Technical Product Training . 29 of 52


VCSn MODULE

• Sample/Sheath Pressure
• Sheath = 8.00 psi (7.98-8.02)
• Sample = typical 8.9-9.2 psi
(nominal = 9.1 psi)
• Advisor Case #K96323234
• Sheath tank
• Sample tank
• Hydrodynamic Focusing
• Processes by which
individual cells or particles
are focused in a stream of
diluent. Cells are aligned
and pass through the flow
cell in a singular fashion
Right Side of Instrument

Ver.1 November 2016 Beckman Coulter Technical Product Training . 30 of 52


VCSn MODULE

Ver.1 November 2016 Beckman Coulter Technical Product Training . 31 of 52


VCSn MODULE

• Gains/Phase
• Initial Gains/Phase
(DxH 800 FSM 4.200.20)
• Final Gains/Phase
(DxH 800 FSM 4.200.23)
WHY?
• Temperature
Compensation

Ver.1 November 2016 Beckman Coulter Technical Product Training . 32 of 52


VCS MODULE

B. Data Plot Development


• Thresholds (ET & SW)
• Impedance
• Light Analysis
• Populations

Ver.1 November 2016 Beckman Coulter Technical Product Training . 33 of 52


VCSn MODULE

Data Plot Development


• DF1
• Volume (impedance, 256
channels)
• MALS (median angle light
scatter, 256 channels)
• Two thresholds:
• Electronic (ET)
• Software (ST) ET
• Electronic (ET)
• Fixed impedance – total
events tracking (8192)

Ver.1 November 2016 Beckman Coulter Technical Product Training . 34 of 52


VCSn MODULE

Data Plot Development (cont’d)


• Software Threshold (ST)
• Discriminates WBCs from
Non-WBCs
• Everything above = WBCs
• Below to ET = Non-WBCs

Ver.1 November 2016 Beckman Coulter Technical Product Training . 35 of 52


VCSn MODULE

Data Plot Development (cont’d)


• The five WBC Differential
populations are:
• EO- Eosinophils: EO’s are a minor
MO
population and may not appear on NE EO
the scatterplot.
• NE- Neutrophils BA
LY
• MO- Monocytes
• LY- Lymphocytes
• BA- Basophils (behind the
Lymphs)

Ver.1 November 2016 Beckman Coulter Technical Product Training . 36 of 52


VCSn MODULE

NRBC Data Plot


• LALS (low angle light scatter)
• AL2 (axial light loss)
Grans
• Absorbance Debris

• Debris/Abnormal Platelets
LY

• Lymphocytes
• Granulocytes (MO, NE, EO)
• NRBCs
NRBCs

Ver.1 November 2016 Beckman Coulter Technical Product Training . 37 of 52


Here is how to navigate this pro

COMMON
SERVICES
MODULE

Ver.1 November 2016 Beckman Coulter Technical Product Training . 38 of 52


COMMON SERVICES

PNEUMATIC SERVICES MODULE


• Pneumatic Supply
Module (PSM)
• Raw Pressure (25 psi)
• Raw Vacuum (25 Hg
vacuum)
• Regulators
• 4 psi (mix pressure)
• 25 psi (Raw Pressure)
• Sample pressure
• Sheath pressure
• Low vacuum (6 Hg
vacuum)

Ver.1
Ver.1Ver.1 November
November
November 20162016
2016 Beckman
Beckman
Beckman Coulter
Coulter
Coulter Technical
Technical
Technical Product
Product
Product Training
Training . . .
Training 39 of 52
COMMON SERVICES

PNEUMATIC SERVICES MODULE


• Pneumatic Supply
Module (PSM)
• Manual PSM Power Up
(3 ways)
1. Dx Tools Maintenance
tab (Start Pneumatics)
2. System Monitor
VL303
3. PSM bypass switch on
control board under
top cover

Ver.1
Ver.1Ver.1 November
November
November 20162016
2016 Beckman
Beckman
Beckman Coulter
Coulter
Coulter Technical
Technical
Technical Product
Product
Product Training
Training . . .
Training 40 of 52
COMMON SERVICES

PNEUMATIC SERVICES MODULE


• 4 psi regulator (VCSn mix
pressure)

Ver.1
Ver.1Ver.1 November
November
November 20162016
2016 Beckman
Beckman
Beckman Coulter
Coulter
Coulter Technical
Technical
Technical Product
Product
Product Training
Training . . .
Training 41 of 52
Common Services Module
• Reagent Services Module

Ver.1 November 2016 Beckman Coulter Technical Product Training . 42 of 52


REAGENT SERVICES
MODULE REAGENT DETECTION
• Reagent Detection
• Sensing Electrode in buffer/reservoir
• Cycle Counter
• “Configure Low Level” (yellow alert), no
effect in/on determination of reagent
depletion
• Reagent Depleted
• Non-Diluent Reagent
• unable to fill the reagent buffer
within timeout
• Cycles typically 0, but not always
• All diluent supplies are depleted means
that both diluent supplies were consumed
logically (the counters decremented to 0 and
the diluent buffer always filled on time)
• Unable to sense diluent means that the
diluent buffer wasn't able to fill within the
expected timeout and most likely the diluent
manifold has ingested air

Ver.1
Ver.1Ver.1 November
November
November 20162016
2016 Beckman
Beckman
Beckman Coulter
Coulter
Coulter Technical
Technical
Technical Product
Product
Product Training
Training . . .
Training 43 of 52
REAGENT SERVICES
MODULE
DILUENT NOT SWITCHING
• Advisor case #K13237613
• Failure modes
• Pick up tube
• QD leaking/not connected
• Input tubing kinked
• Rescanned after software
update
• Operator error
• Not replacing when
depleted
• Pooling

Ver.1
Ver.1Ver.1 November
November
November 20162016
2016 Beckman
Beckman
Beckman Coulter
Coulter
Coulter Technical
Technical
Technical Product
Product
Product Training
Training . . .
Training 44 of 52
REAGENT SERVICES
MODULE

DILUENT NOT SWITCHING


• Failure modes (cont’d)
• Over consumption
• Diluent Threshold adjustment
Advisor case #K45444431
• 420 ml <200
specimens/day
• 530 ml > 200
specimens/day
• Threshold function
• Sets level at which diluent
switching occurs
• Dead volume

Ver.1
Ver.1Ver.1 November
November
November 20162016
2016 Beckman
Beckman
Beckman Coulter
Coulter
Coulter Technical
Technical
Technical Product
Product
Product Training
Training . . .
Training 45 of 52
Here is how to navigate this pro
SPECIMEN
TRANSPORT
MODULE
(STM)

Ver.1 November 2016 Beckman Coulter Technical Product Training . 46 of 52


Specimen Transport Module (STM)
• Cassette Flow
• Mixer

Ver.1 November 2016 Beckman Coulter Technical Product Training . 47 of 52


SPECIMEN TRANSPORT MODULE (STM)

Cassette Detection Sensor Locations- Advisor Case #K32701675

Ver.1 November 2016 Beckman Coulter Technical Product Training . 48 of 52


SPECIMEN TRANSPORT MODULE (STM)

Mixer Station
• Failure Modes
• Stuck
• Counter Measure- Advisor
Case #K32701675
• Test sensor function
• Clean Home Sensor with
DI H2O

Ver.1 November 2016 Beckman Coulter Technical Product Training . 49 of 52


REFERENCES:
• UniCel DxH
Hematology Analytical
Stations Field Service
Manual, Rev AL
• DxH Connectivity
Workstation Manual
(Airwatch)
• Advisor Cases

Ver.1 November 2016 Beckman Coulter Technical Product Training . 50 of 52


Ver.1 November 2016 Beckman Coulter Technical Product Training . 51 of 52
THE END