X-Ray Equipment Controls
X-Ray Equipment Controls
Equipment
Controls
University of Guyana
2
The control unit
Line compensator: This is where the voltage from the wall plug-in is measured and
then stabilized to 220 Volts for the x-ray circuit.
Autotransformer: Has a single winding and sends voltage to the filament and high
voltage circuit
A. Selects the major or minor kvp and steps the voltage up to
kilovolts
kvp meter: measures the electrical potential of the x-ray tube.
Exposure switch: used to complete the x-ray exposure. It regulates the length of the
exposure.
filament circuit: where we have the ammeter, or the mA selector, which selects the
tube current to heat the filament. 6
Diagram of x-ray unit
7
Current control (mA)
8
(Kvp) kilovolt peak
9
Timers
• Timers are used in the system so that radiographer starts the exposure and the timers stops it.
• The timer circuit is separate from the other main circuit of the x ray circuit of the x ray system.
Type of Timers
• Mechanical timers
• Electronic timers
Selecting time: control time it takes to charge the capacitor
Activating exposure also begin storage of charge in the capacitor
• MAs timers :a special type of timer that accurately control tube current and exposure time
• The product of mA and time(mAs) determine the number of x ray photon emitted and the density
on the film
• Monitors the product of mA and terminates the exposure when the desired mAs has been attained.
10
Meters, Switches and monitor
• Digital image the control unit is located in the control room and
all the parts was place together and there is automatic and
manual input.
15
References