BM800 Flow Diagram: General Description of Flowchart
BM800 Flow Diagram: General Description of Flowchart
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos1 V1 V2 V3 V4
6
5 7
Air pumpP+
4
BloodD 8
V10 V11
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
OT PD V15 R1
LVacuumP-
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos1 V1 V2 V3 V4
6
5 7
Air pumpP+
4
BloodD 8
V10 V11
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
OT PD V15 R1
LVacuumP-
Lower P- aspirates blood sample via V27, SV1-3, Process terminated by blood detector
P- aspirates from wash cup via V14 in the same time 3
Prepare for first dilution
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos2 V1 V2 V3 V4
6
5 7
Air pumpP+
4
BloodD 8
V10 V11
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
OT PD V15 R1
LVacuumP-
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos2 V1 V2 V3 V4
6
5 7
Air pumpP+
4
BloodD 8
V10 V11
4
3 3
9
CH2
2
10 MC1 MC2
1 11
V27 MPA
V12
V13 V14
PM
OT PD
V15 R1
LVacuumP-
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos1 V1 V2 V3 V4
6
5 7
Air pumpP+
4
BloodD 8
V10 V11
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
OT PD V15 R1
LVacuumP-
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos1 V1 V2 V3 V4
6
5 7
Air pumpP+
4
BloodD 8
V10 V11
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
OT PD V15 R1
LVacuumP-
Upper part P- aspirates dil to Dis1 via V-dil, V22, CH1, V5, CH2 by timing
7
Pipette wash
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos1 V1 V2 V3 V4
6
5 7
Air pumpP+
4
BloodD 8
V10 V11
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
OT PD V15 R1
LVacuumP-
P+ pushes dil from Dis1 to washing device via V9, CH1, V18, SV11-6, outside of needle
Lower part P- collects waste from inside needle via SV1-3, V27 and from wash cup via V14 8
Fill diluent pipette
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos1 V1 V2 V3 V4
6
5 7
Air pumpP+
4
BloodD 8
V10 V11
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
OT PD V15 R1
LVacuumP-
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos1 V1 V2 V3 V4
6
5 7
Air pumpP+
4
BloodD 8
V10 V11
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
OT PD V15 R1
LVacuumP-
P- empties M1 via V15, MC1 bypass, V12, Spa, V7, CH1, V5, CH2
P- empties M2 via V26, MC2 bypass, V13, SPb, V7, CH1, V5, CH2 10
Prepare for second dilution
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos1 V1 V2 V3 V4
6 T1
5 7
Air pumpP+
4 8
V10 V11
BloodD
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
T2
OT PD V15 R1
LVacuumP-
P- transfers 1:200 diluted blood from PM to MC2 via SV9-7, T1, V24, V13, T2, V11, CH2
Process stopped by BeakerD signal, around 0.5 ml remains in PM beaker 11
Second dilution
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos3 V1 V2 V3 V4
6
5 7
T1
Air pumpP+
4 8
V10 V11
BloodD
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
T2
OT PD V15 R1
LVacuumP-
P+ pushes diluent from Dis1 to MC1 via V9, CH1, V17, SV8-10, V12; lyze from Dis2 to MC2 via V21, T1,
V24, V13, T2 12
MC1 air vent via V3; MC2 air vent via V4, V28
Prepare PM beaker wash
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos1 V1 V2 V3 V4
6
5 7
Air pumpP+
4 8
V10 V11
BloodD
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
OT PD V15 R1
LVacuumP-
P- moves 2-3 ml Dis1 via CH2, V5 CH1, V22 and MV-Dil by timing
13
Wash PM beaker
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos1 V1 V2 V3 V4
6
5 7
T1
Air pumpP+
4 8
V10 V11
BloodD
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
T2
OT PD V15 R1
LVacuumP-
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos1 V1 V2 V3 V4
6
5 7
T1
Air pumpP+
4 8
V10 V11
BloodD
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
T2
OT PD V15 R1
LVacuumP-
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos3 V1 V2 V3 V4
6
5 7
T1
Air pumpP+
4 8
V10 V11
BloodD
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
T2
OT PD V15 R1
LVacuumP-
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos3 V1 V2 V3 V4
6
5 7
T1
Air pumpP+
4
BloodD 8
V10 V11
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
T2
OT PD V15 R1
LVacuumP-
P+ via Spa/b creates pressure in MC1/2, cells cross aperture, MC1 via V15, MC2 via V26, 270 ul counted
P- fills reagent pipettes Dis1/Dis2. Dil: MV-dil, V20; Lyz: MV-Lyz 17
Wash after count
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos1 V1 V2 V3 V4
6
5 7
Air pumpP+
4
BloodD 8
V10 V11
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
OT PD V15 R1
LVacuumP-
MC1 empties through V10, CH2, P-, air vent from V3. MC2 empties through V11, CH2, P-, air vent from
V4, V28. Vent line for MC1/2 are closed for a short time and vacuum are created. MC1 fills up with diluent 18
through V19, V12. MC2 fills up with diluent through V25, V13
Ready for next analysis
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos1 V1 V2 V3 V4
6
5 7
Air pumpP+
4 8
V10 V11
BloodD
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
OT PD V15 R1
LVacuumP-
After a cycle, instrument returns to initial state, ready for next analysis
4 pipettes full, 2 measuring chamber filled 19
PD sample aspiration (prediluted blood)
SPa SPb
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos1 V1 V2 V3 V4
6
5 7
Air pumpP+
4 8
V10 V11
BloodD
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
OT PD V15 R1
LVacuumP-
CH1
V28
M1 Dis1 Dis2 M2 V5 V6 V7 V8 V9
Shear valve Pos1 V1 V2 V3 V4
6
5 7
T1
Air pumpP+
4 8
V10 V11
BloodD
3
9
CH2
2
10 MC1 MC2
11
BeakerD
1
V27 MPA
V12
V13 V14
PM
T2
OT PD V15 R1
LVacuumP-
P+ pushes diluent from Dis1 to PM via V9, CH1, V16, MPA, brings blood sample in MPA to PM
Proceeds to “prepare to second dilution”. 21