38DLPlus Training Power Point
38DLPlus Training Power Point
Operation Training
Index
Click on the links below to go directly to the corresponding slide. (Please Note: you
have to be in slide show mode for the links to work.)
Power Up 38DLP Echo-to-Echo Feature ID Review
Do-ZERO-Standard Dual Transducers Grid View
Do-ZERO for THRU-COAT Transducers Auto Echo-to-Echo, Dual Element Transduc Creating Notes
ers
Selecting Units and Resolution Note Copy
Main Screen 38DLP Sending, Printing and Deleting dat
Manual Echo-to-Echo, Dual Element Transd
38DLP Standard Calibration ucers a
38DLP THRU-COAT Calibration Sending a File or Multiple Files
Temperature Compensation Feature Sending/Printing a Range of a File
Directly Enter Velocity, THRU-COAT Mode 38DLP Min/Average Setup Deleting Files
THRU-COAT Measurement Screen 38DLP B-Scan Mode File Copy
38DLP Single Element Transducer Calibr Additional Gage Features File Edit Rename (Non Grid Files)
ation File Edit Rename (Grid and Boiler Files)
38DLP Datalogger
Single Element Transducer Selection Selectable Text Editing Modes File Reports
Single Element Cal Vel/Cal Zero 38DLP Setup Menu
Datalogger Memory
Calibration Lock Setup Menu
File Open
38DLP Instrument Lock with Programma File Create Measurement Setup
ble Password
Incremental File Communication Setup
38DLP EMAT Transducer
Sequential File Display Setup
38DLP Waveform Controls
Sequential File with Custom Points 38DLP SP Menu
A-Scan Range Adjust
2D Grid File (Standard) 38DLP Single Element Transducer Setup
Delay Adjust
2D Grid File (EPRI Format) 38DLP Custom Dual Element Transducer
Setups
Echo Detection w/Dual Element Transducer 2D Grid File with Custom Points
38DLP Velocity and Time of Flight Measu
s 3D Grid File rements
Freeze Mode Boiler File 38DLP Software Options
Zoom ID# Entry/Edit
Display Brightness Adjust Saving Thickness/Waveforms
Manual Gain Adjust Dual Element Insert/Append ID# Into a File
Extended Blank for Dual Element
Power Up
The On/Off key is used to turn the
ON
OFF 38DL PLUS on and off.
Back to
Index
Model 38DL PLUS
Dual Element Do-ZERO
Back to
Index
Do-ZERO for Standard Dual Transducers
Uncouple the transducer and make “Do-ZERO” allows the gage to recognize
sure the tip is free of any couplant the transducer for optimal setup. It will also
layer. measures the time of flight through the
transducer to compensate for transducer
Press wear and changes in temperature.
Do-ZERO
CAL
2nd F then ZERO
Rx Tx
Delay Delay
Back to
Index
Do-ZERO for THRU-COAT Transducers
Uncouple the transducer and make
sure the tip is free of any couplant “Do-ZERO” allows the gage to recognize
the transducer for optimal setup. It will also
layer.
measures the time of flight through the
transducer to compensate for transducer
Do-ZERO
wear and changes in temperature.
Press: 2nd F then
CAL
ZERO
Back to
Index
Model 38DL PLUS
Calibration Standard Dual Transducers
Back to
Index
Standard Dual Element Cal Velocity
Couple transducer to the thick
sample press: Allows the user to calibrate for the
speed of sound of the material to be
VEL
tested. Usually done on a sample
CAL
VEL
representing the maximum of the
measurement range.
ENTER
Back to
Index
Standard Dual Element Cal Zero
Couple transducer to the A subtracted time measurement is
used to compensate for the transit
Do-ZERO
time through the delay line in the
CAL transducer and the couplant layer.
ZERO Cal Zero is usually done on a sample
thin sample press: representing the minimum of your
Once reading is steady press: measurement range.
ENTER
Back to
Index
Model 38DL PLUS
THRU-COAT Calibration
Back to
Index
THRU-COAT Cal Velocity
Couple transducer to the thick
sample press: Allows the user to calibrate for the
speed of sound of the material to be
VEL
tested. Usually done on a sample
CAL
VEL
representing the maximum of the
measurement range.
ENTER
Back to
Index
THRU-COAT Cal Zero
Couple transducer to the A subtracted time measurement used
to compensate for the transit time
Do-ZERO
through the delay line in the
CAL transducer and the couplant layer.
ZERO Cal Zero is usually done on a sample
thin sample press: representing the minimum of your
Once reading is steady press: measurement range.
ENTER
ENTER
Back to
Index
Directly enter the velocity in THRU-COAT Mode
Press VEL
Allows the user to directly enter the velocity
of the material to be tested. When using
CAL
2nd F then THRU-COAT the velocity of the coating can
VEL
also be entered.
Press
VEL
CAL
2nd F then VEL
again
Coating
Thickness
Metal Thickness
Back to
Index
Model 38DL PLUS
Calibration of Single Element
Transducers
Back to
Index
Single Element Transducer/Setup Selection
The 38DL PLUS has 21 preset default The user can select one of the
single element transducer setups default transducer setups or any
stored in its permanent memory and of the user defined custom setups.
up to 35 user defined custom setups
can be entered .
To select a default transducer setup
REF VALUE
press
XDCR
RECALL
ENTER
Back to
Index
Single Element Cal Zero
Couple transducer to the A subtracted time measurement used
to compensate for the transit time
through the delay line in the
Do-ZERO
transducer and the couplant layer.
CAL
ZERO Cal Zero is usually done on a sample
thin sample press: representing the minimum of your
Once reading is steady measurement range.
press:
ENTER
Back to
Index
Model 38DL PLUS
Instrument Lock with Programmable
Password
Back to
Index
Instrument Lock with Programmable Password
The instrument lock on the 38DL PLUS
allows the user to lock advanced features
and functions so that they are not
accidently altered or changed.
•Calibration
•Access to Setup Menu and SP Menu
•Transducer Recall
Note: If the user forgets the password •Datalogger (except for the save key)
please contact Olympus for the master •The ability to adjust the gain
password. •Waveform adjustment parameters
Back to
Index
Instrument Lock, setting a Password
SP MENU
Setting a password is an optional feature
Press SETUP and the 38DL PLUS can be locked with out
MENU
a password.
Back to
Index
Model 38DL PLUS
Using the EMAT Transducer
Back to
Index
EMAT Transducer
An EMAT (Electro Magnetic Acoustic Transducer)
uses the Magnetostrictive principle to generate
shear wave sound energy in ferrous metals that
are externally coated with high temperature oxide
scale.
PIPE
SCALE
Back to
Index
EMAT Transducer
•EMAT transducers are designed to be a quick way of
determining the approximate wall thickness without
removing the external oxide scale.
Back to
Index
Connecting the E110-SB EMAT Transducer
The 1/2XA/E110 filter adapter must be
plugged in to the transducer connectors
LCB-74-4 located at the top of the 38DL PLUS. The
E110-SB can then be connected to the
38DL PLUS 1/2XA/E110 adapter box using a standard
E110-SB
Lemo to BNC cable. (LCB-74-4)
Back to
Index
EMAT Transducer Calibration
For best accuracy it is recommended that
the 38DL PLUS be calibrated using
samples with external scale and known wall
thickness. Perform the same Cal Vel and
Cal Zero procedure that is outlined for any
single element transducer.
E110-SB
Oxide Scale
Known
Steel thickness
Back to
Index
Using Gain with the EMAT Transducer
When using the EMAT transducer it will often be necessary to adjust the gain level in order to
make proper echo detection. This is due to the variation in signal amplitude caused by changes
in the external oxide thickness and surface conditions.
Press then Use the [],[] keys to adjust the Gain Value
GAIN
Gain too low echo Gain set properly set for echo
not detected detection Back to
Index
Waveform Controls
Back to
Index
A-Scan Range Adjust
DELAY
Allows the operator to cycle through the
Press fixed ranges of the waveform display.
RANGE
The range should be set so that the
echo from your thickest material will be
Successive presses of the [RANGE] on screen.
key toggles through the different
ranges.
Note: Adjusting the range will not affect calibration. There are
fixed display ranges for each transducer type depending on
transducer frequency.
Back to
Index
Delay Adjust
The delay function is used to adjust the
Press
DELAY beginning (left side) of the waveform.
2nd F then RANGE The user can delay a portion of the
signal off screen so they can view the
important part of the waveform in
Then use the [],[] To adjust
greater detail.
the waveform delay
Delay
Back to
Index
Echo Detection with Dual Element Transducers
Echoes must be greater in
amplitude than 20% of screen
20% height in order to be detected.
Note: The Freeze function is useful for high temperature applications. This
function can be used to limit the transducer contact time and prevent
overheating the transducer. Freeze is also used in conjunction with
Min/Max mode to prevent capturing false couplant readings.
Back to
Index
Zoom, Single and Dual Element, Mode 1
E-TO-E
Zoom centers the measured echo on
Press ZOOM
the waveform display. The Zoom will
automatically track the measured echo
and assure that it remains in the center
60% of the screen.
Press [ZOOM] again to return
to the previously set range.
Note: Higher display brightness settings will decrease battery life battery:
4Hz and 0% brightness is 16 hours
4Hz and 100% 12.5 hours
Back to
Index
Manual Gain Adjust Dual Element Transducers
Press Allows the operator to adjust the receiver
Gain in 1 dB increments. Once an echo is
GAIN
above the 20% detection point, adjusting
gain will not effect the calibrated accuracy.
Back to
Index
Extended Blank for Dual Element Transducers
Allows the operator to extend a
Press blank from the sound entry point to
WAVE
Then use [],[] the max of the displayed range.
ADJ
Note: When the extended blank is set, the gage will measure to the first
valid echo after the end of the blank.
Back to
Index
Extended Blank
Extended
NOISE Blank PROPER
DETECTED DETECTION
Back to
Index
Model 38DL PLUS
Echo-to-Echo Mode
Back to
Index
Echo-to-Echo with Dual Element Transducers
E-TO-E
Back to
Index
Auto Echo-to-Echo, Dual Element Transducers
Echo-to-Echo
marker
Back to
Index
Manual Echo-to-Echo, Dual Element Transducers
Back to
Index
Echo-to-Echo Measurement Tips
E1 Blank E1 Blank
Backwall 1
Shear wave
Echo
Backwall 2
Note: Shear wave echoes can cause Echo-to-Echo measurement problems with the D790 in the
thickness range of .600”-1.00”. We recommend using the D797 for the thicker Echo-to-Echo measurements
or D7906 in Echo-to-Echo or in THRU-COAT mode. The delay material of the D7906 reduces shear wave
echo propagation
Back to
Index
D7906 in THRU-COAT or Echo-to-Echo Mode
Back to
Index
Echo-to-Echo, No Multiples
Back to
Index
Model 38DL PLUS
Temperature Compensation Feature
Temperature Compensation Feature
The sound velocity in steel changes approximately 1% per 1000F (550C) change in
temperature as recommended by ASME Standard E 797-95.
Back to
Index
Temperature Compensation Setup
Press SP MENU Temperature compensation allows the 38DL
SETUP PLUS to compensated for the change in sound
MENU velocity due to temperature. The sound velocity
of carbon steel changes about 1% per 100 0 F
Use the [],[] to highlight (55 0 C)
“Temp Compensation” then
press [ENTER]
Back to
Index
Temperature Compensation Mode
When Temperature
Compensation is activated and
in the measure mode
REF VALUE
Press XDCR
2nd F Then RECALL
Current Temp
Temperature Entered by user
compensated
thickness
Back to
Index
Model 38DL PLUS
Min/Average Setup
Back to
Index
Min/Average Setup
Allows user to save the Minimum or Average of
Press SP MENU 2, 3 or 4 thickness readings
SETUP
MENU
[SAVE]
Note: Pressing the [MEAS] key will clear all Min/Avg boxes.
Using the [ or ] keys allows the user to move to any of the
Min/Avg boxes and the value can be replaced by pressing the
save key.
Back to
Index
Model 38DL PLUS
B-Scan Mode
Back to
Index
38DL PLUS B-Scan
• Non-encoded B-Scan
• Draws thickness profile of wall thickness
• Shows live A-Scan while updating B-Scan
• Marks minimum or maximum thickness while scanning
• Select scan directions
• Marks areas of LOS (loss of signal)
• Allows user to review all thickness values on stored B-
Scan
• Can store up to 10,000 reading per B-Scan
Back to
Index
Live B-Scan Display Overview
A-Scan
B-Scan
B-Scan
Range
Current
Min/Max
Marker
Transducer
Direction
Min or Max Live Thickness
Back to
Thickness Index
B-Scan Setup
SP MENU
Back to
Index
B-Scan Setup
Use the [],[] to highlight the
parameter and [],[] to change the
parameter.
Press [MEAS] to return to the
measure mode with B-Scan active.
Back to
Index
B-Scan “LOS” Opt set to “STOP”
LOS
LOS Regions
Back to
Index
B-Scan Freeze Review
While collecting B-Scan data:
FREEZE
Press
Review
Marker
Back to
Index
B-Scan Freeze Review
Pressing
[],[]
Review
Marker
Thickness at
Review Marker
Back to
Index
Saving Thickness Readings in Freeze Mode
While B-Scan is Frozen (Freeze Review On)
WAVEFORM
Press SAVE
SEND
Review
Marker
Back to
Index
Saving B-Scan Screen or Entire B-Scan
Saving the Min or Max A-Scan Along with the Held B-Scan
WAVEFORM
SAVE
Press SEND
2nd F then
Back to
Index
Additional Gage Features
• Differential
• Reduction Rate Alarms
• Measurement Update Rate
• Min/Max Mode
• Alarm Mode
Back to
Index
Differential Mode
Displays the measured thickness deviation
SP MENU
Press from the differential set point.
SETUP
MENU
Back to
Index
Differential Mode Normal
Normal Differential Displays both the actual live thickness and the
absolute deviation from the differential reference value
Back to
Index
Differential Mode % Ratio
% Ratio Differential Displays both the actual live thickness and the
% deviation from the differential reference value
Back to
Index
Differential Mode Reduction Rate
Reduction Rate Differential is a special %Differential mode used in application of metal forming
or bending. It displays the former reference value, the actual live thickness and the % deviation
from the former value reference value. The user can select to display the actual thickness or
the % reduction in large fonts
Back to
Index
Measurement Update Rate
SP MENU Allows the user to select the display
Press SETUP measurement update rate. The user can
MENU select between (4, 8, 16, 20Hz or Max
Use the [],[] to highlight “Meas” then (approx 30 Hz).
press [ENTER]
Back to
Index
Min/Max Mode These two functions will allow the gage to scan
Press
SP MENU
and hold the Minimum, Maximum or Both Min
SETUP and Max thickness and waveform. Any time
MENU
the Min/Max modes are engaged, the gage will
Use the [],[] to highlight “Meas” then automatically go into Max Measurement update
press [ENTER] rate.
Back to
Index
Alarm Mode Standard Low and High Alarm Mode
Back to
Index
Alarm Mode Absolute Previous Thickness Alarm Mode
SP MENU
Press SETUP
Allows the user to compare current live
MENU thickness values to previously stored
thickness values at each ID# location, for
Use the [],[] to highlight “Alarm” then
the purpose of detecting large wall losses
press [ENTER] or growths.
Back to
Index
Alarm Mode
Previous Thickness – Absolute Differential
Previous
thickness
Current thickness is more than .050” Current thickness is more than .010”
thinner than the previous reading. thicker than the previous reading.
Back to
Index
Alarm Mode
Previous Thickness – Percent Differential
SP MENU Allows the user to compare current live
Press SETUP thickness values to previously stored thickness
MENU values at each ID# location, for the purpose of
detecting large wall losses or growths. The
Use the [],[] to highlight “Alarm” then
user can set a % Loss or % Growth from the
press [ENTER]
previously stored thickness.
Previous
thickness
thinner than the previous reading. thicker than the previous reading.
Back to
Index
B-Scan Alarm Mode
B-Scan must be active to set Allows the user to set High and Low
the B-Scan Alarm alarms for the B-Scan option. High
and low alarm lines will be shown on
SP MENU the B-Scan.
Press SETUP
MENU
Low/High
Alarm Lines
Alarm Flag
Back to
Index
38DL PLUS Datalogger
Back to
Index
SelectableText Editing Modes
SP MENU The 38DL PLUS has two text editing
Press SETUP modes: the new Virtual Keypad and the
MENU traditional text editing that was available on
our older generation thickness gages.
Use the [],[] to highlight “System”
and press [ENTER]
Back to
Index
Virtual Keypad Text Editing
When Virtual Keypad text editing mode is
selected the 38DL PLUS will display a
virtual keypad. This allows the user to
interface with the entire of text in one field.
Note: When an outer character is highlighted pressing the arrow to move outside the Virtual
keypad will cause the character on the opposite side of the Virtual keypad to highlight. This
keypad wrap around allows the user to quickly navigate through the character selection.
Back to
Index
Traditional Text Editing
When Traditional test editing mode is selected
the user uses the [,] key to select the
character (letters, Numbers and punctuation)
and [,] to move the curser. Pressing
[ENTER] to complete the editing.
Note:
[] key starts ABCD….#-…983210
[] Key starts 0123…#-… ZYX
[] moves the curser to the right
[] moves the curser to the left
[ENTER] complete the editing.
Back to
Index
Datalogger Memory
CLR MEM
Back to
Index
File Open Allows the user to open a previously
CLR MEM
created or downloaded file.
Press FILE
Use [,] to select sort by Name or Date then Note: The corresponding file header
press [ENTER] will be shown as you change your
Use [],[] to select the File then press file selection.
[ENTER]
Use [,] to select Open or Cancel and
[ENTER]
Back to
Index
File Create
Allows the user to create files. The
CLR MEM
38DLPlus has nine standard file formats.
Press FILE
Then [],[] to select “CREATE” then
press [ENTER]
•Thickness: Store standard single element, Dual element and Echo-to-Echo measurements
•THRU-COAT: Stores THRU-COAT measurements both the coating and metal thickness
•Temp COMP: Stores Temperature compensated Thickness and Temperature setting
•Oxide Layer: For optional Oxide software stores Tube and Internal oxide thickness
•Velocity: Stores velocity values when using velocity mode
•Time of Flight: Stores Time of Flight values when using Time of Flight mode
•Reduction Rate: Stores Actual thickness and Reduction % when using Reduction Rate Mode
•Soft Contact For optional Multilayer software, stores Sagittal Height, Radius and Thickness of lens
•%Total Thickness: For optional Multilayer software, stores thickness and % of total thickness readings
•Min/Max: Stores both Min and Max thickness as a single ID#
Back to
Index
Incremental File
Incremental files will start at the
entered point and increment from
the right when the save key is
pressed.
Example:
001
002
003
.
.
.
Enter the starting ID# using the editing controls then press 999
[ENTER]. Use [],[] to choose [Create] then press
[ENTER]
Back to
Index
Incremental File
Example 1 Example 2
Back to
Index
Sequential File
A sequential file starts at ID# 1 and
increment until ID# 2. ID’s will
increment in alphanumeric order from
Select Sequential File type, thickness mode and the right.
file protections and press [ENTER]. Use [],[]
to choose [CONTINUE] then press [ENTER].
Example:
ELBOW-12-A
ELBOW-12-B
ELBOW-12-C
ELBOW-12-D
.
.
.
ELBOW-12-H
Example:
001- LEFT
Example:
A01
B01
C01
D01
A02
B02
Enter all parameters using the editing controls then
.
press and select Standard for ID format and press
.
[ENTER]. Use [],[] to choose [Create] then press
M10
[ENTER].
Back to
Index
2D Grid File (EPRI Format)
Allows the user to build two-dimension grid
files by defining the starting column and row
as well as ending column and row. The user
Select 2D GRID File type, thickness mode and file can also choose the direction that the file
protections and press [ENTER]. Use [],[] to will increment. EPRI files use double letter
choose [CONTINUE] then press [ENTER]. columns, for columns past Z.
Example:
A01
B01
C01
.
Z10
AA01
Enter all parameters using the editing controls then AA02
press and select EPRI for ID format and press .
[ENTER]. Use [],[] to choose [CONTINUE] then FF10
press [ENTER] and enter header information then
choose {Done or Cancel}. Back to
Index
2D Grid File (Standard)
EXAMPLE:
Increment by: Increment by:
First ID# TANK-A01 Row Column
05
Back to
Index
2D Grid File with Custom Points
Allows to user to build three-dimension grid
files by defining starting/ending column and
row. A custom list of included points may
Select 2D +Custom PT File type, thickness mode also be entered. The user can also choose
and file protections and press [ENTER]. Use [], the order in which the parameters increment.
[] to choose [CONTINUE] then press [ENTER].
Example:
A01- TOP
A01- BOTTOM
A02- TOP
A02- BOTTOM
.
.
.
Enter each parameter using the editing controls then G05- BOTTOM
press [Enter]. Use [2nd F], [] or press [ENTER] on a
blank line to exit custom point. Choose [Create] and
then press [ENTER]. Back to
Index
3D Grid File
Allows the user to build three-dimension
grid files by defining the starting and ending
Column, Row, and Point. The user can also
Select 3D Grid File type, thickness mode and file choose the order in which the parameters
protections and press [ENTER]. Use [],[] to increment.
choose [CONTINUE] then press [ENTER].
Example:
A01A
A01B
A01C
A01D
A02A
.
.
.
Z15D
Enter all parameters using the editing controls then
press [ENTER]. Use [],[] to choose [Create]
then press [ENTER].
Back to
Index
3D Custom File
Allows the user to build three-dimension
grid files by defining starting and ending
column, a custom set of rows, and a custom
set of points. The user can also choose the
Select 3D Custom File type, thickness mode and order in which the parameters increment .
file protections and press [ENTER]. Use [],[]
to choose [CONTINUE] then press [ENTER].
Example:
01-Ring1-TOP
01-Ring1-BOTTOM
01-Ring2-TOP
01-Ring2-BOTTOM
01-Ring3-TOP
01-Ring3-
BOTTOM .
.
.
Enter all parameters using the editing controls then 10-Ring3-BOTTOM
press [ENTER]. Use [],[] to choose [Create]
then press [ENTER].
Back to
Index
Boiler File
Allows to user to build three-dimension
boiler files by defining the starting and
Select Boiler File type, thickness mode and file ending tube as well as custom points and
protections and press [ENTER]. Use [],[] to elevations. The user can also choose the
choose [CONTINUE] then press [ENTER]. order in which the parameters increment.
Enter each parameter using the editing controls then press [ENTER]. When in
Custom Point or Elevation use [2nd F], [] or press [ENTER] on a blank line to
exit the parameter. Choose [Create] and the [ENTER]
Back to
Index
Boiler File
EXAMPLE:
C C C
L R L R L R
0 ft
x C C C
Back to
Index
ID# Entry/Edit
Press
ID# ‘s can be 16 characters with letters,
NOTE
numbers, alphanumeric and ( , / * . # -
ID# Twice space)
Back to
Index
Saving Thickness Readings/Waveforms
Press WAVEFORM
Saves the thickness reading or thickness
SAVE
SEND readings and waveforms at the current ID#
location.
Back to
Index
Insert/Append ID# into a File
Press The user can insert a new ID# and reading
Note:
into an existing file or append a new ID#
and reading which will be add at the end of
ID# Twice the entire file.
Note: The user can jump [2ndF] [] to jump to the first ID# in a file
or or [2ndF],[] to jump to the last ID# in a file.
Note: The user can jump to a specific ID# location in the file by
using the ID# edit function.
Back to
Index
Grid View
Grid view allows the user to view grid
SP MENU
Press: file in a grid format and assigned us to
SETUP
MENU three color ranges. The user can select
to Shows half waveform/Grid file Full Grid,
Then use [],[] to highlight No A-Scan and also select to Reverse rows
DB Grid and press [ENTER] or columns, Transposed grid View grid in
linear form and set Data Flag
Inserted points
Inserted points
Back to
Index
Adding Comments to the Current ID# Location
NOTE
Up to four comments can be saved
Press 2nd F then ID# with each ID#. A thickness reading
can be saved along with the
comments or the comments can be
saved without a thickness reading.
Use [],[] to highlight the comment letter you want to use, then press [ENTER] to
select/un-select the comment. Press [MEAS] after comments are selected and choose
and the next time the [SAVE] key is pressed the comments will be saved with the
current ID#.
Back to
Index
Adding Comments to a Range of ID# Locations
NOTE
Up to four comments can be saved
Press 2nd F then ID# with each ID#. A thickness reading
can be saved along with the
comments or the comments can be
saved without a thickness reading.
Use [],[] to highlight the comment letter you want to use, then press [ENTER] to
select/un-select the comment. Press [2ndF], [] after comments are selected and edit
the start ID# location and then edit the ending DD# location. Use the [],[] to
highlight [SAVE] and press [ENTER]. The comments selected will be save to the
range of ID#’s.
Back to
Index
Note Copy Allows the user to copy a comment
CLR MEM code list from one file to another
Press selected file.
FILE
Use [],[] to select sort option and press [ENTER] Note: If a Note table is copied to a file
then use [],[] to select the source file then press with Note’s already stored, the new
[ENTER]. Use [],[] to select sort option and press copied Note: table will overwrite the
[ENTER] Use [],[] to select the destination file then notes previously stored.
press [ENTER] and press [ENTER] while “Copy” is
highlighted.
Back to
Index
38DL PLUS
Sending, Printing and
Deleting Data
Back to
Index
Sending a File or Multiple Files (RS-232 only)
Allows the user to send a single
CLR MEM
Press file, multiple files or all files to a
FILE computer or printer via RS-232.
This is function is not used with
GageView.
Back to
Index
Sending/Printing a Range of a File (RS-232 Only)
Note: Connector type must be set to RS-232 Press and Send/Save must be set to
“SEND”
Note: Only files that are not delete protected can be deleted
Back to
Index
Deleting a range of data in a file
Allows the user to Clear/Delete a range of
Press: CLR MEM data in a file or the entire open file. Both ID#
and thickness will be deleted in Incremental
2nd F then FILE
and Sequential files. Only the thickness
values will be deleted in Grid and Boiler
The Clear ID Range screen files.
will appear on the display
Press:
CLR MEM
Back to
Index
File Copy
CLR MEM
Press:
Allows the user to copy a file or the format of a
FILE file to another file. This is a quick way to create
multiple files with the same structure.
Use [],[] to select “Copy” then press
[Enter]
Use [],[] to select sort option and press [ENTER] then use [],
[] to select the file to copy then press [ENTER]
Use the editing functions to enter a file name for the new file then
press [ENTER]
Use [],[] to select Copy Data Yes or No and press [ENTER]
while “Copy” is highlighted.
Back to
Index
File Edit Rename (non grid files)
CLR MEM
Use [],[] to select sort option and press [ENTER] then use [],[] to
select the file to edit then press [ENTER]
Use the editing functions to edit the a File Name, Description, Inspector ID
and Notes, press [ENTER]
Use [],[] to change the file delete protection to On or Off and press
[ENTER]
Use [],[] to highlight UPDATE and press [ENTER] Back to
Index
File Edit Rename (Grid and Boiler Files)
CLR MEM
Allows the user to edit the file header, add Rows or
Press: columns to grid files and change the incrementing
FILE
direction.
Use [],[] to select “Edit” then press
[Enter]
FILE EDIT-RENAME
Use [],[] to select sort option and press [ENTER] then use [],[] to select the file to Edit
then press [ENTER]
Use the editing functions to edit the a File Name, Description, Inspector ID and Notes and
press [ENTER]
Use [],[] to change the file delete protection to On or Off and press [ENTER]
Use [],[] to highlight Continue and press [ENTER]
Use the editing functions to edit the End Column, End Row and Incrementing Directions and
Press [ENTER]
Back to
User the [],[] to highlight Update and press [ENTER] Index
File Reports
CLR MEM
Press:
FILE
Allows the user to generate four different
reports on the display of the 38DLPLUS.
Use [],[] to select “Reports” then press
Use the [],[] enter the Report selection box
Use [],[] select a report and press [ENTER]
Back to
Index
File Reports
File Summary
Use [],[] to select the sort option and press [ENTER This report will give a statistical summary of the
Use [],[] to select a file then press [ENTER] selected file.
Use [],[] to select report and press [ENTER]. And
the statistics will be displayed.
Back to
Index
File Reports
Min/Max Summary
This report lists the minimum and maximum
Use [],[] to select the sort option and press
thickness readings and their location for the
[ENTER Use [],[] to select a file then press
selected file.
[ENTER]
Use [],[] to select report and press [ENTER]. And
the Min/Max Summary will be displayed.
Back to
Index
File Reports
File Comparison
This report will compare the two selected files and
Use [],[] to select the sort option and press
show the maximum wall loss and locations,
[ENTER Use [],[] to select the Reference
average wall loss, and the locations of any wall
file then press [ENTER]. Use [],[] to select
growths.
the Comparison file then press [ENTER]. Use
[],[] to select report and press [ENTER].
Back to
Index
38DL PLUS Setup Menu
• Measurement
• System
• Communication
• Display
Back to
Index
Setup Menu
Used to change the setup parameters .
Press:
SP MENU By scrolling down the setup menu list the
SETUP sub parameters are visible and the user
MENU can navigate directly to any setup
parameter.
Back to
Index
Measurement Setup
Use [],[] to highlight the
parameter and [],[] to change the
parameter.
Back to
Index
Measurement Setup Continuation
Use [],[] to highlight the
parameter and [],[] to
change the parameter.
Back to
Index
System Setup
Use [],[] to highlight the
parameter and [],[] to
change the parameter
Save Data: Programs Save/Send Key to Save Thickness or THK and waveform
Default File Data Mode: Set the Default File data mode
Back to
Text Edit Mode: Set Editing Mode Virtual Keypad or Traditional Slewing Index
Communication Setup
Use [],[] to highlight the
parameter and [],[] to
change the parameter
RS-232 Device: Printer, Terminal (PC), Bar Code Reader, Digital Caliper, Fischer
Gage
Baud Rate: 1200, 2400, 4800, 9600, 19200, 38400, 57600 or 115200
Note: The following RS-232 parameters are fixed at the values below:
Data Bits: 8
Stop Bits: 1
Parity: None
Back to
Index
Display Setup
Use [],[] to highlight the
parameter and [],[] to
change the parameter
Display Brightness: Set brightness of display backlight 0%, 25%, 50% or 100%
Back to
Index
38DL PLUS SP Menu
• Clock Setup
• Language
• Gage Resets
• Diagnostic Tests
• Status
Back to
Index
SP Menu
Press: The SP Menu is used to change the instrument
SP MENU
parameters that are not often adjusted and to
2nd F Then SETUP activate software options and to run diagnostic
MENU instrument tests.
Back to
Index
Clock Setup
Back to
Index
Language
The user can import custom language
files from the external Micro SD memory
card or export a file to be translated and
converted into a language file in the gage.
Back to
Index
Gage Resets
Gage Resets is used to reset the gage
back to factory default conditions. The
user can choose to perform a
Measurement, Internal Memory
Communications or Master Reset.
Internal Memory Reset: Clears and resets the internal memory card
Dual XDCR Test: Reports RX and TX time of flight from Dual element transducer
Back to
Index
Software Diagnostic Test
Software diagnostic test reports and
software error messages will be
found under this tab .
Back to
Index
Status The status report lists the information
about the instrument including:
internal temperature, current battery
level, software version and hardware
version of the 38DL PLUS.
Back to
Index
38DL PLUS
Single Element Internal Setup and
Transducer Adjustment
Back to
Index
Setup Adjust for Single Element Transducers
Allows the user to adjust the pulser,
receiver, detection and blanking Caution: Making adjustments to the Setup
parameters. Parameters should be made by a qualified
individual who is familiar with ultrasonics
Press and the use of the 38DL PLUS.
WAVE
Adjustments made using this feature can
ADJ affect the measurements.
Adjustment
Parameter
Back to
Index
Mode 1
Mode 1 measures the time of flight between the Main Bang and the first
back wall echo, using direct contact transducers.
Main
Bang First
Back Wall
Back to
Index
Mode 2
Interface
Main
Bang
Back Wall
Echo
Back to
Index
Mode 3
First
Back Wall
Interface Second
Back Wall
Back to
Index
Setup Name
• Default or user defined name that identifies the application setup that is
currently selected
• Limited to 16 alphanumeric characters
Back to
Index
Measure Type
• For most single element transducer the Meas Type will be set to standard.
Standard is for Standard Mode 1, Mode 2 and Mode 3 measurements.
• First Peak is a special algorithm to measure the first peak of a group of
several peaks on either the positive or negative side of the waveform.
Common uses would be for gel coat on fiberglass measurements or bolt
elongation.
• Oxide Layer is a special measurement type used for displaying the
thickness of boiler tubes and internal oxide at the same time. This feature
is only available when the optional oxide software is activated
• Barrier is a special measurement algorithm used to measure the thickness
of the barrier layer thickness in multi-layer plastic material. Typical
applications are plastic fuel tanks, bottle preforms .
Back to
Index
Probe Type
Back to
Index
Pulser Power
Back to
Index
Pulser Power
Initial Initial
Pulse Pulse
Back to
Index
Max Gain
Max Gain
l o pe
GS
Initial Gain TD
Time Back to
Index
Initial Gain
Max Gain
l o pe
GS
Initial Gain TD
Time Back to
Index
TDG Slope
• The rate of increase of the receiver gain (time dependent) from Initial Gain to
Max Gain
• Slope can help suppress reflections from grain structure or fibers
• Slope can be adjusted from 0.0dB to 26.52 dB/µSec
• The point in time that the gain starts to slope is:
• The Main Bang Blank in Mode 1
• The End of the Interface Blank in Mode 2 and Mode 3
Max Gain
l o pe
GS
Initial Gain TD
Time Back to
Index
TDG Gain Mode 1
Max Gain
pe
Initial Slo
TDG
Gain
Back to
Index
TDG Gain Mode 2 and Mode 3
Max
Gain
S lope
Initial Gain TDG
MBB
Back to
Index
Main Bang Blank (MBBlank)
• A blank zone that protects the receiver from false readings generated
by the Main Bang
• Indicates the point in time where the gage begins to search for echoes
• Adjustment directly affects the minimum thickness the gage can
measure in Mode 1
• Set just past the Initial Pulse in Mode 1
• Set just before the Interface Echo in Modes 2 and 3
• Make sure that the Main Bang Blank is always set before the Interface
Echo from the shortest water path when using immersion transducers
Back to
Index
MBBlank in Mode 1
Back wall
Main Bang Echo
Blank
Initial
Pulse
Back to
Index
MBBlank in Mode 2 and Mode 3
Interface
Initial Main Bang Back wall
Echo
Pulse Blank Echo
Back to
Index
Echo Window
• The end of the Interface Blank and the Echo Window in Mode 2 or
Mode 3
• Indicates the selected period of time in which the receiver will search
for echoes
• Signal appearing after (i.e. outside of) the Echo Window will not be
detected by the gage
Back to
Index
Echo Window in Mode 1
Main Bang
Echo
Blank
Window
Back to
Index
Echo Window in Mode 2 and Mode 3
Echo
Interface Window
Echo
Back to
Index
Echo 1 Detect
• This indicates the detection polarity of the "first detected back wall
signal" in the Echo Window in Mode 1
• In Modes 2 &3 it refers to detection polarity of Interface (or delay line)
Echo
• Always choose the detection polarity on the singular lobe of the echo
• Echo polarity is affected by:
• Acoustic impedance
Back to
Index
Echo 1 Detect
Back to
Index
Echo 2 Detect (Modes 2 and 3 only)
• Indicates the detected polarity of the back wall echo in Mode 2 or the pair of
back wall echoes in Mode 3
• Always choose the detection polarity on the singular lobe of the echo
Echo 1 Detect
Positive
Echo 2 Detect
Negative
Back to
Index
Interface Blank (Modes 2 and 3 only)
• Sets the length of a blank that follows the detected interface echo
• Used to prevent the gage from detecting the trailing edge of the Interface
Echo in Mode 2
• Used to select which pair of back wall echoes will be detected in Mode 3
Back to
Index
Interface Blank Mode 2
Back to
Index
Interface Blank Mode 3
Back to
Index
Mode 3 Blank (Mode 3 only)
• Sets the length of a blank that follows the first detected back wall echo in Mode 3
• Used to prevent the gage from detecting the trailing edge of the first back wall echo
Back to
Index
Mode 3 Blank (Mode 3 only)
Back to
Index
38DL PLUS
Custom Dual Element Transducer
Setups
Back to
Index
Custom Transducer Setup for Non Standard Dual
Transducers
The 38DL PLUS offers probe recognition for the Olympus’ standard D79X
series dual element transducers. When one of these dual element
transducers is plugged into the 38DL PLUS the gage automatically recalls a
default setup and V-Path correction.
These D79X series transducers will provide superior performance when used
with the 38DL PLUS. The gage . The gage also has the ability to create and
store custom setups for non standard dual element transducers, including a
custom V-Path correction.
Back to
Index
Caution for Non Center Pin Duals
Olympus cannot guarantee measurement accuracy or performance for transducers other
than our standard D79X duals
Performance, accuracy and minimum/maximum thickness capability with other duals
needs to be verified by the user
Transducers that work in standard mode (first back wall) may not work very well in Echo-
to-Echo mode
The 38DL PLUS is designed to work with most dual element transducers that are used
for thickness gauging applications, but there is no guarantee that it will work with all dual
element transducers
D790 on 0.200 in. Block Other 5 MHz Dual on 0.200 in. Block Back to
Index
Non Standard Dual Element Transducers
When a user plugs in a dual element transducer that is not an Olympus D79X
transducer. The 38DL PLUS will display the generic dual element transducer setup
selection screen
Use [],[] to highlight “Default” to select from the list generic dual element transducer or
“Custom” to select for the list of previously stored custom dual element setups.
Note: When selecting a generic setup, choose the one that best matches the frequency of the
transducer being used.
Back to
Index
V-Path Builder (Create a custom V-Path)
Turn V-Path calibration on
Gage asks user if they want to do V-Path Cal when [CAL VEL] is pressed
Enter up to 10 points and save setup and V-Path as a custom dual element setup
The user can store up to 10 Custom Dual element setup
Back to
Index
Creating a Custom Dual Transducer Setup
Use the [],[] keys to select a generic setup
from the list of default dual element
transducers and press [MEAS]:
Do-ZERO
Note: the thickness measurement value at this point will most likely not be correct
Back to
Index
Two Point Calibration ( Cal VEL & Cal Zero)
•Couple to the thick step, get a steady reading and press [Cal VEL], then [ENTER],
then enter in the known thickness
•Couple to the thin step, get a steady reading press [Cal VEL], then [ENTER], then
enter in the known thickness and press [MEAS]
• Check the thickness on all steps if measurement accuracy is sufficient then save
the setup as a custom dual element transducer setup.
Note: If the accuracy is not acceptable then a custom V-Path should be performed.
Back to
Index
Turn V-Path Builder on
SP MENU
Press SETUP
MENU
V-Path builder requires at least three calibration points and allows the user to enter up to 10
calibration points to create a custom V-Path
Points should all be within the measurement range and should represent the min and max
thickness to be measured.
Back to
Index
Creating a custom V-Path table
Couple to the thickest standard and get a
steady thickness reading and press[ CAL VEL],
select “yes” for V-Path calibration, then press
[ENTER] and [ENTER] again Use the [],[]
keys to enter the know thickness of the step.
Back to
Index
Saving a custom V-Path
Back to
Index
38DL PLUS
Velocity and Time of Flight Measurements
Back to
Index
Velocity Measurements
The 38DL PLUS has the ability to be make thickness, Velocity or time of flight. The
user can select the measurement mode in the Setup menu
SP MENU
Press SETUP
MENU
Back to
Index
Making Velocity Measurements
When in velocity mode it is necessary to perform a two point thickness calibration to
properly set the zero offset for the transducer being used.
Press SETUP
MENU
Back to
Index
Making Time of Flight Measurements
When in time of flight mode it is necessary to perform a two point thickness
calibration to properly set the zero offset for the transducer being used.
Back to
Index
Activating Software Options
•The software options for the 38DL PLUS are activated using a licensed
option key. An option can be purchased at the time the unit is
purchased or the 38DL PLUS can be upgraded with the software
options in the field. If the unit was purchased along with software
options then they will be installed and activated when you receive it.
•If software options are needed after the unit is purchased then they will
have to be activated using a unique licensed option key.
•Please contact your Olympus representative for information concerning
activating software options for your 38DL PLUS
Back to
Index
Activating Software options
SP MENU
Press 2nd F Then SETUP Allows the user to activate one or several
MENU of the software options for the 38DL
PLUS. One single option key can activate
more then one software option.
Back to
Index
HR (High Resolution Software Option
The standard resolution for the 38DL PLUS is 0.001 in. (0.01 mm). The High
Resolution software option allows the gage to display thickness
measurements in High Resolution 0.0001 in (0.001 mm)
Note: High resolution is not available for Dual element transducer or HP (High
Penetration single element transducers.
When High Resolution had been activated it will then be one of the resolution
options available in the Measurement Setup Menu.
Back to
Index
38DL PLUS
Internal Oxide Software Option
Back to
Index
Oxide Software Option
• Oxide/Scale can build up on the inside of boiler tubes
• The oxide that builds up acts as an insulator.
• This effects the heat transfer from outside of the tube to the water inside the
tube
• The tubes runs at a higher temperature then they were designed for and they
start to thin faster.
• Knowing both the thickness of the steel and Internal Oxide the remaining
tube life can be more accurately predicted
Back to
Index
Oxide/Scale Measurement
• New measurement algorithm allows the user to measure scale or oxide build up on the
inside of boiler tubes
• Thickness of Oxide helps predict tube life
• Measures and displays Oxide and tube thickness at the same time
• Min scale thickness
•0.010 in. (0.25mm ) with M2017
•0.006 in. (0.150 mm) with M2091 Normal Incident Shear
Steel/Oxide Echo
Echo from
inside of oxide
Interface Echo
Data Window
Steel/Oxide
Echo Marker
Back to
Index
Oxide Thickness
Tube Thickness
Oxide setup for M2017 Transducer
Select the setup for the M2017
transducer setup
REF VALUE
Press
XDCR
RECALL
Back to
Index
Adjust the Initial Gain, Slope and Max Gain
Adjust the TVG Gain (Initial, Max
Gain and Slope) to make a mode 2
detection of the larger total back
wall.
Oxide/Steel
Interface Echo
Mode 2 detection
from delay line echo
to total back wall echo Back to
Index
Adjust Data Window1
Adjust the DataWin1 so it extends
far enough to cover the oxide steel
interface echo
Oxide/Steel
Interface Echo
Back to
Index
Oxide: Steel Cal Velocity
Couple transducer to thick steel sample Allows the user to calibrate for the speed of
without oxide sound of the material to be tested. Usually
Press VEL done on a sample representing the
CAL maximum of the measurement range.
VEL
Back to
Index
Oxide: Steel Cal Zero
Couple transducer to the
A subtracted time measurement used to
compensate for the transit time through the
Do-ZERO delay line in the transducer and the
couplant layer. Cal Zero is usually done on
Press CAL thin steel
ZERO a sample representing the minimum of
sample without oxide
your measurement range.
Press ENTER
Back to
Index
Oxide Coating Calibration
Couple transducer to a sample with Allows the user to calibrate for the sound
known oxide thickness speed of the of the oxide material.
Press VEL
CAL
VEL
Back to
Index
Directly Enter Velocity, Oxide Mode
VEL
CAL
Allows the user to directly enter the velocity
Press 2nd F Then VEL of the material to be tested. When using
oxide mode the velocity of the oxide can
also be entered
The current material velocity will be
displayed. Use the numeric keys
to enter the know velocity. Then press [MEAS]
Press
VEL
CAL
2nd F Then VEL Again
Back to
Index
38DL PLUS
HP Software Option
Back to
Index
HP transducer setups
REF VALUE
Press Once the HP (High Penetration) software has
XDCR
RECALL been activated the 38DL PLUS will add a list of
HP low frequency single element transducers to
the list of default single element transducers.
Back to
Index
Measurements with The M2008
VEL The M2008 is a special low frequency transducer
CAL that is used to measure thick fiberglass and
Press 2nd F Then VEL composite materials. Especially good for material
0.200 in. (6 mm) and greater. When using this
transducer it is necessary to Zero the delay line.
Back to
Index
Multi-Layer Software Option
The Multi-Layer software option is used to simultaneously display the thickness of up to four layers of a multi-
layer material.
Co-extruded plastic
Back to
Index
Multi-Measure Setup
SP MENU Once activated this software option allows the
Press
SETUP user to select up to four custom stored transducer
MENU setups and display all the thickness readings on
the display at one time. A custom transducer
setup has to be made for each layer of the
material to be measured.
Back to
Index
Multi-Measure Mode
Display Parameter (Waveform vs. Large Font)
Back to
Index
Multi-Measure Mode (Soft Contact)
Used to measure the layers of a soft
contact lens application
Back to
Index
Multi-Measure Mode (Percent Total Thickness)
Used to express the thickness of a
layer as a percentage of the total
thickness of a part
• Used in Mode 1 to setup a gate on the Barrier Layer echo. The gage
automatically centers DataWin1 on the detected Barrier Layer echo. The
width of DataWin1 should be set so the entire echo is within the gate.
Back to
Index
Data Window 2 (Barrier Layer only)
• In Mode 2 it is used to setup a gate on the Barrier Layer echo. The gage
automatically centers DataWin2 on the detected Barrier Layer echo. The width
of DataWin2 should be set so the entire echo is within the gate.
Back to
Index
Barrier Layer Measurement (Mode 1)
• Select the default setup for M116
• Enter the parameter adjust by pressing [WAVE ADJ]
• Use [] or [] to select MEAS Type and use [] or [] to select Barrier Layer
• Adjust Max Gain, Initial Gain and TDG Slope so that the Barrier Layer is being
detected as shown below
Barrier Layer
Echo
Back wall
Detection Echo
Marker
Back to
Index
Barrier Layer Measurement (Mode 1)
Use [] or [] to select DataWin1 and use [] or [] adjust the width so it is large
enough to include the entire Barrier Layer reflection, but not include the Main Bang
echo or Back wall echo.
Barrier Layer
Echo
Back wall
DataWin1 Echo
Back to
Index
Barrier Layer Measurement (Mode 2)
• Select the default setup for M208
• Enter the parameter adjust by pressing [WAVE ADJ]
• Use [] or [] and [] or [] to make sure Mode 2 is selected and the Meas Type
is set to Barrier Layer. Adjust Max Gain, Initial Gain, TDG Slope, Echo 1 and Echo 2
polarity so a proper detection is being made between the Interface echo and the
Barrier Layer echo as shown below.
Interface Barrier
Echo Echo
Back wall
Echo
Mode
Detection
Marker
Back to
Index
Barrier Layer Measurement (Mode 2)
• Use [] or [] to select DataWin1 and use [] or [] to adjust the width so it is
large enough to include the entire Interface echo but does not include the Main Bang
or the Barrier echo.
• Use [] or [] to select DataWin2 and use [] or [] to adjust the width so that the
width of the window is large enough to include the entire Barrier Layer reflection, but
does not include the Interface echo or Back wall echo.
Back wall
DataWin1 DataWin2 Echo
Back to
Index
38DL PLUS
Multi-Layer Calibration
Back to
Index
Cal Velocity (Standard Measurement Type)
Couple transducer to the thick
sample Allows the user to calibrate for the speed of
VEL
sound of the material to be tested. Usually
CAL
Press VEL
done on a sample representing the
maximum of the measurement range.
Press ENTER
Back to
Index
Cal Zero (Standard Measurement Type)
Couple transducer to the A subtracted time measurement used to
compensate for the transit time through the
Do ZERO delay line in the transducer and the
Press CAL
couplant layer. Cal Zero is usually done
ZERO thin sample on a sample representing the minimum of
your measurement range.
Once reading is steady
Press
ENTER
Back to
Index
Cal Mode 1 (Barrier Measurement Type)
VEL
Press CAL
VEL Allows the gage to store a reference
echo off a test block and calibrate for
Couple transducer to 0.50 in. or 15 mm test the sound speed of the barrier material
block to be tested.
Press ENTER
Back to
Index
Cal Mode 2 (Barrier Measurement Type)
VEL
Press CAL Allows the gage to calibrate for the
VEL sound speed of the barrier material
to be tested.
Couple to a sample with known barrier
thickness
Press
ENTER
Back to
Index