Gas Chromotograph - Daniel
Gas Chromotograph - Daniel
•Injection
•Separation
•Detection
•Integration / Calculations
•Reports
Brains of the GC “Controller”
•Integration / Calculations
•Reporting
Heart of the GC “Oven”
•Injection
•Separation
•Detection
GC Oven
The GC oven contains 3 main components mounted together in an
electrically heated heat-sink-oven:
SAMPLE
LOOP
6 PORT
VALVE
MEASUREMENT
DETECTORS
REFERENCE
DETECTOR
Flow Path in GC Oven
Analyzer Oven
Step 3: The sample valve turns back OFF to purge the next stream.
The backflush valve turns OFF reversing the carrier flow through
column 1 so that C6+ components elute first (all combined as a single
peak) to the measure detector. Note: by switching the direction of flow
through the first column, C6+ components bypass columns 2 and 3.
This helps to expedite the analysis.
Analyzer Oven
Given: Sample Stream gas is run, and the raw area of ethane
calculated by the controller is 200.
• Minimizes normalization.
•Minimizes cross-stream
contamination if solenoids do not
seal completely.
Sample Conditioning and Sample Transport
Calibration Standards
•Repeatability
•Accuracy
•Linearity
Repeatability and Accuracy can be checked by your calibration
Mole % = Area(comp x)
RF(Comp x)
Instrument Performance-Linearity Check
Using three standards
Note: Heating values are based on GPA 98 Engineering Data Book Volume II Section 23.
Using 60 F and 14.696 base conditions
Physical Property Calculations-Pressure Base
•During the analysis the GC uses columns to separate the gas into its
constituents
•The GC uses thermal conductivity detectors to quantify the mole %
of each component present
•A BTU is calculated for each constituent based on mole %
•By summing up each of these individual values we find the total
BTU value of the gas
Demon on Gas chromotograph
• Daniel C6+.exe