Pump Module: Flow Rate Accuracy
Pump Module: Flow Rate Accuracy
Pump Module
Flow Rate Accuracy:
The flow-rate accuracy of the pump can be evaluated simply by calculating the
time required to collect a predetermined volume of mobile phase at different
flow rate settings.
For example, the flow-rate accuracy at 2 mL/min can be verified by using a calibrated
stopwatch to measure the time it takes to collect 25 mL of effluent from the pump into a
25-mL volumetric flask. A calibrated flow meter can be used to determine the flow rate
as well.
A calibrated flow meter can be used to determine the flow rate as well.. A steady
backpressure may be required, depending on the requirement of the system.
1
Gradient Accuracy and Precision:
• The gradient operation precision can be assessed indirectly by monitoring the
relative standard deviation in retention time of peaks in the chromatographs from
repeated injections.
• The accuracy and linearity of the gradient solvent delivery can be verified
indirectly by monitoring the absorbance change when the composition of the two
solvents from two different channels changes.
• High-pressure gradient runs usually involve two solvent systems. Lower-
pressure gradient LC pumps are usually equipped with quaternary proportioning
valves, which can handle up to four solvents. The test will be performed for two
channels at a time.
2
Pump Module
Pressure Test:
The performance of the LC pump depends on the proper
functioning of the pump seal, check valves, and proper
connection of the tubing. Properly functioning pump seal check
valves and tubing connections are important in maintaining stable
mobile-phase flow and system pressure.
3
Pump Module
Pressure Test:
When performing the pump pressure test, make sure that the
system is well primed so that no bubbles are present in the pump
system. The pressure fluctuation caused by the presence of air
bubbles in the pump will lead to misinterpretation of the test
results.
4
Pump Module
Pressure Test:
The first step in the pressure test is to plug the outlet of the pump
using a dead-nut. The automatic pump shutdown pressure is set
to 4000 psi. The pump head pressure signal output is connected
to a recorder (this test can be performed without using a
recording device by visually monitoring the pressure reading as
the system is being pressurized).
5
Pump Module
Pressure Test:
Pressurize the pump by pumping methanol at 1 mL/min.
Methanol is less viscous than water and is more sensitive to leaks
in the system. The pressure inside the pump head increases
quickly as the outlet of the pump is blocked. As the pressure
increases to about 3000 psi, the flow rate is reduced to 0.1
mL/min.
6
Pump Module
7
Pump Module
Pressure Test:
The pressure in the pump head decreases slowly over time after
the automatic shutdown. Typically, the pressure drop is less than
10% over 5 min. A steep decrease in pressure over time implies
poor check-valve performance or leaks within the pumping
system.
8
Injector Module
Precision:
The ability of the injector to draw the same amount of sample in
replicate injections is crucial to the precision and accuracy for
peak-area or peak height comparison for external standard
quantitation. If the variability of the sample and standard being
injected into the column is not controlled tightly, the basic
principle of external standard quantitation is seriously
compromised.
9
Injector Module
Precision:
No meaningful comparison between the responses of the sample
and the standard can be made. The absolute accuracy of the
injection volume is not critical as long as the same amount of
standard and sample is injected.
10
Injector Module
Precision:
The volume precision of the injector can be demonstrated by
making at least six replicate injections from a sample. The relative
standard deviation (% RSD) of the response of the injections is
then calculated to evaluate the precision. A fast and sample
analysis should be considered for the test. The run condition for a
very simple LC analysis of caffeine is given below as an example.
11
Injector Module
Precision:
12
Injector Module
Linearity:
Most automated LC injectors are capable of varying the injection
volume. A variable volume of sample will be drawn into a sample
injection loop by a syringe or other metering device. The
uniformity of the sample loop and the ability of the metering
device to draw different amounts of sample in proper proportion
will affect the linearity of the injection volume.
13
Injector Module
Linearity:
Linearity is important for methods that require the use of variable
injection volumes, such as the high–low method in the
quantitation of impurities (see Chapter 3 for the validation of
related substances method). The linearity of the injector can be
demonstrated by making injections, typically 5,10, 20, 50, and
100 μL, to cover the range 0 to 100 μL.
14
Injector Module
Linearity:
The response of the injection is plotted against the injection
volume. The correlation coefficient of the plot is used in
evaluation of the injection linearity. Carryover:
15
Injector Module
Carryover:
To avoid cross-contamination from the preceding sample
injection, all the parts in the injector that come into contact with
the sample (the injection loop, the injection needle, and the
needle seat) have to be cleaned effectively after the injection. The
carryover can be evaluated by injecting a blank after a sample
that contains a high concentration of analyte.
16
Injector Module
Carryover:
The response of the analyte found in the blank sample expressed
as a percentage of the response of the concentrated sample can be
used to determine the level of carryover. Caffeine can be used for
the system carryover test for assessing the performance of an
injector and serves as a common standard for comparing the
performance of different injectors.
17
Resolution
Acceptance:
• ±1 nm in the UV range (200 to 380 nm) and ±3 nm in the visible range (380 to
800 nm). Three repeated scans of the same peak should be within ±0.5 nm.
Wavelength Accuracy
Stray Light