Lab Qa Compliance Verification

The document discusses compliance verification findings from a microbiology laboratory in March 2021. Key issues found include: 1) Lack of proper documentation such as missing signatures, dates and labeling on culture media and equipment logs. 2) Issues with microbiological practices like unlabeled pH meter, missing micropipettes and desynchronized incubators. 3) Deviations from good laboratory practices including unlabeled reagents and non-functional equipment. 4) Incomplete review of test data and documents with missing signatures and batch numbers. A total of 50 non-conformances were reported across documentation, microbiological work, laboratory practices and review practices.

Uploaded by

robit Siddiki
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
89 views18 pages

Lab Qa Compliance Verification

The document discusses compliance verification findings from a microbiology laboratory in March 2021. Key issues found include: 1) Lack of proper documentation such as missing signatures, dates and labeling on culture media and equipment logs. 2) Issues with microbiological practices like unlabeled pH meter, missing micropipettes and desynchronized incubators. 3) Deviations from good laboratory practices including unlabeled reagents and non-functional equipment. 4) Incomplete review of test data and documents with missing signatures and batch numbers. A total of 50 non-conformances were reported across documentation, microbiological work, laboratory practices and review practices.

Uploaded by

robit Siddiki
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PPTX, PDF, TXT or read online on Scribd
You are on page 1/ 18

Compliance Verification in

Microbiology Laboratory-
March 2021

Prepared By
Robit Siddiki
Executive QA

08/28/2021 1

Robit Siddiki Tirupati Lifesciences pvt. ltd.


Key Focus Points
 Good Documentation practices.
 Best Microbiological practices.
 Good Laboratory Practices.
 Good Review Practices.
 Analytical practices.
 Environmental Monitoring program.
 Validation/Qualification and Hold time study.
Robit Siddiki Tirupati Lifesciences pvt. ltd. 08/28/2021 2
Good Documentation practices.
 Some of dehydrated Culture media kept in accessories door for daily use
purpose w/o proper labelling i.e. :-
I. Frasser Broth base:- 0000421344 media opened on 11/02/2021 for
qualified its GPT but till date checked by sign not done in status label
affix on it.
II. EE Broth mossel :- 0000395413 Media opened on 22/02/2021 for qualified
its GPT but till date checked by sign not done in status label affix on it.
III. Muller hilton agar:- 0000390495 Valid upto 01/07/2020, Sabouraud
dextrose agar:- 0000380925 valid upto :- 23/04/2020, Nutrient agar:-
0000357976 valid upto :- 25/06/2020, Simmon citrate agar :- 0000357976
valid upto:- 29/01/2020. As per the status label media not used for
further analysis but till date 09/03/2021 kept in door.
Continued……

Robit Siddiki Tirupati Lifesciences pvt. ltd. 08/28/2021 3


 media consumption record of FBB TLS/QCM/003/F/02-01 Lot No:- 0000421344/01/10,
entry closed for page no. 305 on dated 08/03/2021 but till dated checked by sign not
done.
 19-Media plate reconciliation record for TSA (RODAC PLATES) used for PM on 01/03/2021
but till date 10/03/2021 checked by sign not done.
 Good Documentation practices SOP TRP/CQA/O/010-001 not followed during filling of
document ( i.e. more than two remarks are written in usages log book of Dual
chamber lower TLS/QCM/044/F/04-00 & Usages log book of Balance TLS/QCM/WB-03)
 Usages log book of incubator TLS/QCM/028/F/02-01 (TLS/QCM/BI-05) used for keeping
47 MLT pathogen tubes of MCB on dated 05/03/2021 at 17:35 and 43 MLT pathogen tubes
of MCB on dated 06/03/2021 at 16:59 but from 09/03/2021 to till date records not
updated.
 Usages log book of incubator TLS/QCM/028/F/02-01 (TLS/QCM/BI-03) used for keeping
SCDM for EEB bottle on dated 11/03/2021 at 10:38 hrs. but from this date to till date
record not updated while test completed on 14/03/2021.

Continued......

Robit Siddiki Tirupati Lifesciences pvt. ltd. 08/28/2021 4


 09 No's of Primary packaging material sampled on 08/03/2021 and analyzed
on 11/03/2021, in which 03 TDS found incomplete i.e. date of sampling,
sampled by, Date of Analysis, sample quantity and media detailed not
completed as on till date 16/03/2021.
 Usages log book of dual chamber (Lower) TLS/QCM/044/F/04-00, 12 plates of
BGBA (MLT) incubated on 12/03/2021 at 17:40 incubated completed on
14/03/2021 but till date record not update.
 Double door Autoclave thermal graph of dated 15/03/2021 kept in folder w/o
signing off.
 Destruction of used media/Vials culture and other accessories
TLS/QCM/026/F/01-01, Page No;- 507 strike out by NA without provided any
justification.
 Daily Verification record of balance TLS/QCM/036/F/01-01 reviewed by
pending from 11/03/2021 to till date.
 Miscellaneous sample entry register TLS/QCD/015/F/08-00 incomplete from
28/01/2021 to till date i.e. Remark's, Received by, Given By, not filled
properly.
Robit Siddiki Tirupati Lifesciences pvt. ltd. 08/28/2021 5
Continued…
 UV light burning hour record maintain for Pass box TLS/QCM/PB-01 as on
routine basic even though UV light not in working condition.
 During review of Raw material inward register TLS/QCM/013/F/07-00, its
observed that incomplete sample status has been made in inward register
while sample has been analyzed and released 03/02/2021 and 09/02/2021
respectively.
E.g.1:- Isomaltose:- 430360, 430361 and Maltodexterin:- 430756, 430757. and
A.R . No. not generated yet.
E.g.2:- Vitamin B12:- 400600903 analysis on 06/02/2021 but till date 27/03/2021
released by/results and status pending.
 Specimen signature not match with documented signature of Mrs. Rinky. i.e.
Usages log book of LAF-01 TLS/QCD/014/F/01-00 signature not verified.
 Daily Calibration record of pH meter TLS/QCM/062/F/03-00 loose format kept
in door and from 25/02/2021 format not used neither entry closed, if binding
log book has been in placed.
Continued…
Robit Siddiki Tirupati Lifesciences pvt. ltd. 08/28/2021 6
 Cleaning record of Autoclave TLS/QCM/039/F/01-00 entry not completed i.e.
cleaned by/Checked by sign pending on dated 27/02/2021 and from
20/03/2021 to till date document not used whereas frequency is weekly.

 Total Non conformance :- 17

Robit Siddiki Tirupati Lifesciences pvt. ltd. 08/28/2021 7


Best Microbiological Practices.
 pH Meter used for routine analysis having not any
calibration status label.
 10 ml micropipettes not available laboratory for routine
analysis of MLT.
 Time not synchronized for all incubators.
 Media approval and qualification procedure not
incorporate in SOP of Media Management.
 Excessive no. of oxygen bubble formed in water sample during
sampling

Robit Siddiki Tirupati Lifesciences pvt. ltd. 08/28/2021 8


 Media pour in analyzed plates of MLT & Water without checked its
temperature via I.R Gun. (Deviation of SOP)
 Purified water sample PW-59 fail in pH test parameter against the limit 5.50
to 7.50
 Water sample analyzed on 05/03/2021 and released om 10/03/2021 in which
some sampling point PW-05, PW-06, PW-47, PW-59, PW-31 having TNTC and
PW-30 having 1000 cfu & PW-17 having 180 cfu .
 Incubator door open for once 35 to 45 min and second 15-25 min for cleaning
purpose of floated sample of prohance mom. That lead to destruction of
incubation time of samples incubated inside incubator TLS/QCM/BI-04

 Total Non conformance :- 09

Robit Siddiki Tirupati Lifesciences pvt. ltd. 08/28/2021 9


Good Laboratory Practices
 Calibration status label not affix upon pressure gauge installed with Vertical
autoclave TLS/QCM/AT-02
 Many of the 70 % IPA spray bottle used for routine purpose without proper
labeling i.e. Date of Preparation and Use before date not legible.
 UV light burning hour record maintain for Pass box TLS/QCM/PB-01 as on
routine basic even though UV light not in working condition.
 GLP practices not followed during analysis i.e. LAF bench not clean frequently
during performing MLT.
 Two colony counter installed in microbiology laboratory for observing plates
count but both are not in working properly i.e.
(1) TLS/QCM/CC-01- beep continuously while working on it meanwhile base light
not emit properly for its correctness as required
(2) TLS/QCM/CC-02 - counting marker not available during operation is going on.
Robit Siddiki Tirupati Lifesciences pvt. ltd. Continued…
08/28/2021 10
 Weighing balance TLS/QCM/WB-02 observed is in off condition once operation
has been done. And also used for further purpose w/o verification done.

 Total Non conformance :- 06

Robit Siddiki Tirupati Lifesciences pvt. ltd. 08/28/2021 11


Good Review Practices.
 In MLT TDS Analyzed By Sign pending on confirmation test of S.aureus whereas
checked by sign and reviewed by sign done (Product Name:- Formula 1
Nutritional Shake Mix Batch No:- E81468G19)
 Wrong Batch No. mention in TDS sheet i.e. B4KOO001 instead of B4KOU001
(Product Name:- Reheptin Suspension Batch No:- B4KOU001)
 Incomplete Review checklist attached with MLT TDS i.e. Inference of Reviewer,
Remark and Reviewed by column not filled. (Product Name:- Afresh Peach Flv.
Batch No:-E81467C18)
 MLT TDS of Dinoshake chocolate Flv. Batch No. E81469G23; Checked by sign
pending for E.coli TDS whereas reviewed by sign done.
 Overwriting observed in Coliform TI Sheet for Date of Completion i.e. DOC not
legible whereas reviewed by sign done. Product Name :-Nitework Batch No:-
E81469B09

Robit Siddiki Tirupati Lifesciences pvt. ltd. 08/28/2021 12


 MLT TDS of Nitework Batch No:- E81469B09;A.R. No. Not match with MLT TDS
and FG CoA i.e. TLSHL-FPF-066/21 written instead of TLSHL-FPF-0661/21
 Incomplete TDS i.e. Sample Quantity, Date of Released, Checked by sign and
analytical record review checklist not filled or done on Platimax Tablet RT-06
Batch No. :-M PXT20003.

 Total Non conformance :- 07

Robit Siddiki Tirupati Lifesciences pvt. ltd. 08/28/2021 13


Analytical Practices
 Water sample analyzed on 05/03/2021 and released om 10/03/2021 in which some
sampling point PW-05, PW-06, PW-47, PW-59, PW-31 having TNTC and PW-30 having
1000 cfu & PW-17 having 180 cfu .
 Purified water sample PW-59 fail in pH test parameter against the limit 5.50 to 7.50
 Raw material Orange Pokoe batch No 8953 analyzed on 10/03/2021, TNTC observing
in both plates on dated 12/03/2021.
 Raw material Turmeric batch No 201 analyzed on 09/03/2021, TNTC observing in
both plates on dated 12/03/2021.
 Prohance mom Batch No. 63 enriched for salmonella spp. In two different
quantity i.e. 2000 gm & 175 gm in which due to gas formation one bottle of
2000 gm has been broken inside the incubator that lead fail to performed test
for further procedure.

 Total Non conformance :- 05


Robit Siddiki Tirupati Lifesciences pvt. ltd. 08/28/2021 14
Environmental Monitoring Program
 Active Air sampling has been done microbiology lab on dated 18/03/2021 and
released on 23/03/2021 found not satisfactory as there is no any expression
observed on to the sampled plate. Plates are similar to negative control.
 Air Sampling Plates of Class 100 Area having over count as per acceptance
criteria. i.e. Pass box-L3- 45 cfu, LAF-L2:- 36 cfu, LAF-L3:- 25 cfu, BSC-L2:- 10
cfu, Passbox-L2:- 11 cfu, Passbox-L3:- 17 cfu, Center:- 11 cfu.
 Trend data not prepared for EMP carried out in Microbiology laboratory.

 Total Non conformance :- 03


Robit Siddiki Tirupati Lifesciences pvt. ltd. 08/28/2021 15
Validation/Qualification and Hold time study
 Analyst qualification procedure not covered all the test parameter performed
in Microbiology Laboratory.
 Disinfectant validation not available in microbiology laboratory.
 Disinfectant efficacy test not available in microbiology laboratory.
 Garment &assoseries sterilization hold time study not available in
microbiology laboratory.
 Fumigation Validation not available in microbiology laboratory.
 UV Efficacy Test (effectiveness study) not available in microbiology
laboratory.
 Qualification of Biological indicator procedure not available in microbiology
laboratory.

 Total
Robit Siddiki Non pvt.
Tirupati Lifesciences conformance
ltd. :- 07 08/28/2021 16
Total No. of Non Com pliance

Trend analysis of Compliance verification in


18
Microbiology Laboratory

16

14

12

10

0
GDP BMP GLP GRP AP EMP VQHTS
Robit Siddiki Tirupati Lifesciences pvt. ltd. 08/28/2021 17
Non Compliance With respect to
Abbreviation

 GDP :- Good Documentation practices


 BMP:- Best Microbiological Practices
 GLP:- Good Laboratory Practices
 GRP :- Good Review Practices
 AP :- Analytical Procedure
 EMP :- Environmental Monitoring program
 VQHTS:- Validation & Qualification/Hold time study

Thanks
Robit Siddiki Tirupati Lifesciences pvt. ltd. 08/28/2021 18