eCL8000 Maintenance Guide
eCL8000 Maintenance Guide
Maintenance
Guidance
CONTENTS
1 Routine Maintenance
2 Site Maintenance
13 Case Study
1
PART 01
Routine Maintenance
--For end users
Daily
Maintenance
Clean S/R Probe and Aspiration Probe
Dirt on the S/R probe and aspiration probe may cause contamination and carryover, and affect results. Clean this
part
to prevent contamination.
Recommended Frequency: Everyday
Operator time: Approximately 5 minutes.
Precautions: Wear appropriate protective gloves and be careful not to damage the
probe;
All power switches must be OFF.
Materials Required
Clean Gauze obtain locally
Distilled or Deionized Water -------
70% Alcohol -------
en.lifotronic.com
Daily
Maintenance
Procedure:
Clean S/R Probe and Aspiration Probe
1 Turn off all power switches.
2 Move the probes to an area allowing easy operation(Move out the area of the cleaning cells)
3 Wipe the external surface of the probe with an alcohol-soaked gauze. Then wipe the probe with a gauze
soaked in distilled or deionized water.
4 Switch on the device and use it after normal startup.
en.lifotronic.com
Weekly Maintenance
Liquid leakage on the reaction tray may result in adhesion among assay cups, so it is necessary to clean
this component on a regular basis.
en.lifotronic.com
Weekly Maintenance
Procedure:
Clean Reaction Tray
1 Power off the device.
2 Remove the protective cover of reaction tray.
3 Use cotton swabs dipped with distilled water or deionized water to wipe assay cup holes of the reaction tray
one by one.
4 Wipe the surface of reaction tray with gauze soaked in alcohol; then wipe the surface again with gauze soaked
in distilled water or deionized water.
5 Put the reaction tray back in the original position.
6 Switch on the device.
Step 3 Step 4
en.lifotronic.com
Weekly Maintenance
There are a cleaning cell used to flush the reagent/sample probe and a cleaning cell used to flush the aspiration probe.
Dirt in the cleaning cells may result in residue. In order to prevent contamination, the cleaning cells should be
cleaned weekly.
Materials Required
Clean Gauze obtain locally
Distilled or Deionized Water -------
Cotton Swab -------
en.lifotronic.com
Weekly Maintenance
Procedure:
Clean the Cleaning Cells
en.lifotronic.com
Weekly Maintenance
Contamination in the measuring system may reduce the accuracy and precision of sample tests or even block
the measuring channel. In order to keep the measuring liquid channel clean and maintain the performance of
the measuring cell, the measuring cell should be cleaned at least once two weeks or 2500~3000 tests.
Materials Required
Measuring cell maintenance Buffer -------
en.lifotronic.com
Weekly Maintenance
Procedure:
Measuring Cell Maintenance
1 Remove the Auffer from the system reagent cartridge and put measuring cell maintenance buffer
2 Click [Maintenance] - [Maintenance Instructions] - [Measuring cell maintenance]. It takes 15 minutes for
the
performance to completed.
3 Replace the measuring cell maintenance buffer with Auffer.
4 Do the measuring cell preparation for 5 times as the instructions.
5 Complete the performance of the maintenance process.
en.lifotronic.com
Annual
Maintenance
4.Clean the special reagent cartridge
Recommended Frequency: 12 months
Operator time: Approximately 10 minutes.
Materials Required
Distilled Water or Deionized Water -------
Cotton Swab -------
Procedure:
1 Take out all the immunoreagents. Wipe the sides and the bottom of the cartridge with a cotton swab dipped in
distilled water or deionized
2 Wipe the drain of the cartridge.
en.lifotronic.com
PART 02
Site Maintenance
--For engineers
Quarterly
Maintenance
1. Replace the pinch valve hoses(pinch tube)
Liquid leakage may be caused if old pinch valve hoses are used, further affecting the accuracy of pipetting
volume. Therefore, the hoses should be replaced at least once every three months.
Materials Required
180mm pinch tube 2 pcs
en.lifotronic.com
Quarterly
Maintenance
Procedure:
Replace the pinch valve hoses
3 Use new pinch tube to connect the upper and lower two joints respectively.
4 Clamp the pinch tube to the inner side of the corresponding pinch valve.
5 Switch on the device. Go to [Application] – [Advanced] – [HD maintain]. Click [Replacement] to reset the
count.
en.lifotronic.com
Annual
Maintenance
2.Clean the channel and Add lubricating oil
Recommended Frequency: 12 months
Operator time: Approximately 10 minutes.
Materials Required
Clean Gauze obtain locally
Lubricating oil -------
Cotton Swab -------
Procedure:
1 Switch off the device. Use clean gauze or cotton swabs to remove dirt from the lubricating oil on the
track.
2 Wipe clean and apply the new oil.
3 Slowly move the 2D and 3D arms to evenly distribute the oil.
en.lifotronic.com
Annual
Maintenance
3.Clean the optocoupler
Recommended Frequency: 12 months
Operator time: Approximately 10 minutes.
Materials Required
Cotton Swab -------
Procedure:
1 Use clean cotton swab to clean the dust on the optocoupler. (3 optocouplers on 2D arms and 3D arms(X axis and
Y axis); 10 optocouplers on special reagent cartridge; 1 optocoupler on reaction tray)
Internal
structure of
measuring cell
Measuring cell is the most important part of machine, it must be in a good state, clean and healthy.
In order to make it clean and healthy, we must doing the measuring cell maintenance every two weeks or
over 2,000 tests.
Packages
Maintenance/Service package
Maintenance package
Service package
THANKS